Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2019 Jul 1.
Published in final edited form as: Magn Reson Med. 2017 Nov 20;80(1):101–111. doi: 10.1002/mrm.27009

Phase-Sensitive B1 Mapping: Effects of Relaxation and RF Spoiling

Jie Wen 1, Alexander L Sukstanskii 1, Dmitriy A Yablonskiy 1,*
PMCID: PMC6433377  NIHMSID: NIHMS1012432  PMID: 29159883

Abstract

Purpose:

To develop a phase-based B1 mapping technique accounting for the effects of imperfect RF spoiling and magnetization relaxation.

Theory and Methods:

The technique is based on a multi-gradient-echo sequence with 2 successive orthogonal radiofrequency (RF) excitation pulses followed by the train of gradient echoes measurements. We have derived a theoretical expression relating the MR signal phase produced by the 2 successive RF pulses to the B1 field and B0-related frequency shift. The expression takes into account effects of imperfections of RF spoiling and T1 and T2* relaxations.

Results:

Our computer simulations and experiments revealed that imperfections of RF spoiling cause significant errors in B1mapping if not accounted for. By accounting for these effects along with effects of magnetization relaxation and frequency shift, we demonstrated the high accuracy of our approach. The technique has been tested on spherical phantoms and a healthy volunteer.

Conclusion:

In this paper, we have proposed, implemented, and demonstrated the accuracy of a new phase-based technique for fast and robust B1mapping based on the measured MR signal phase, frequency, and relaxation. Because imperfect RF spoiling effects are accounted for, this technique can be applied with short TRs and therefore substantially reduces the scan time.

Keywords: MRI, B1 mapping, RF spoiling, flip angle mapping, frequency shifts

INTRODUCTION

An accurate measurement of B1 radiofrequency (RF) field is important for many MRI applications, such as quantitative mapping of MR relaxation parameters, RF pulse design for parallel transmission (15), and mapping electric tissue properties (68). For example, mapping of a longitudinal relaxation rate constant R1 often relies on MRI with variable flip angles (VFA) (916). However, the accuracy of VFA methods not only depends on the models (1622) used to analyze data but also highly depends on accuracy of B1 measurements.

Up to date, several methods have been proposed to measure B1, either based on the magnitude (2333) or phase (3438) of the MR signal. Initially, the doubleangle methods (23,24) were used for B1 measurement. In these methods, long repetition times (TR > 5T1) are necessary to eliminate the T1-dependence of the signal magnitude, therefore usually requiring long acquisition time. Substantial reduction of the dependence of B1 evaluation on the system relaxation parameters has been achieved in the methods proposed by Yarnykh (30) and Nehrke et al. (32,33). As an alternative to magnitude-based methods, the MR signal phase-based methods have been proposed to measure B1 field (3437). To generate the MR signal phase dependence on flip angle, Morell (35) and later Chang (37) proposed to use a pair of successive orthogonal RF pulses, whereas Sacolick et al. (36) proposed to use the Bloch-Siegert phase shift effect (39).

In this paper, we further explore a phase-sensitive method for B1 mapping that is based on a pair of successive orthogonal RF pulses encoding information on B1 field strength. We demonstrate that the MR signal phase generated by the pair of successive orthogonal RF pulses depends not only on B1 field strength but also on the MR signal relaxation properties (T1 and T2* ) and (most importantly) on imperfections of RF spoiling. These effects have not been considered previously. We have developed a theory accounting for contribution of the above mentioned effects (T1, T2* and RF spoiling) as well as the B0-induced frequency shift to the MR signal phase and proposed experimental method for B1 mapping relying on a multi-gradient-echo sequence with 2 successive orthogonal RF pulses used for signal excitation. The use of the multi-gradient-echo sequence provides simultaneous measurements of the B1-encoded MR signal phase, signal transverse relaxation rate constant R2* and the B0-dependent MR signal frequency shift. The feasibility and accuracy of the method are examined using computer simulations and demonstrated on a phantom and on a healthy volunteer.

THEORY

Two consecutive RF pulses (corresponding to the nominal flip angles α and β, respectively) with orthogonal directions (x and y) tilt magnetization from the longitudinal B0 direction to the x,y-plane and form a magnetization phase in the x,y-plane. This phase depends on the flip angles α and β and, therefore, can be used for inferring information on B1 field. However, the measured phase of MR signal deviates from the magnetization phase–it depends also on different instrumental factors (RF coil mutual locations, cable length, etc.) (40). To eliminate these effects, Morell (35) proposed to use the second measurement with another pair of RF pulses (α, –β), and then to calculate the difference between the phases in these 2 measurements (in the original paper by Morell [35], the pairs [2α, α] and [2α, –α] were used). In Supporting Information (SI) to this paper, we analyze a choice of the flip angles (α, β). Our analysis is based on 2 main principles: first, the flip angles should be chosen in such a way to minimize an uncertainty in their determination from experimental data; second, experimental values of flip angles that can be used in human experiments are restricted by the RF power deposition characterized by the specific absorption rate. The preferences dictated by these 2 principles lead to different requirements for the optimal flip angles (see Supporting Fig. S1), and an optimal choice is a trade-off between these requirements. As a result, we found (and used in our experiments) that the best choice corresponds to α = β ~ 90°.

The quantity Δφ corresponding to the difference of the MR signal phases generated by the pairs of RF pulses (α, α) and (α, –α) can be presented as

Δφ=arctan[Q(α,p)],Q(α,p)=2Ω2(p2+α2cosΩ)α2(1cosΩ)(α2+(p2+Ω2)cosΩ), [1]

where Δω = 2π · Δf is the B0-related frequency shift, p = Δω · τ, Ω=(α2+p2)1/2 and τ is the duration of each RF pulse.

However, Equation [1] does not take into account magnetization relaxation effects and details of MRI pulse sequence used for data acquisition. These effects are considered below.

Effects of Relaxation on Signal Phase

To elucidate effects of relaxation properties, we need to solve the Bloch equations for magnetization M, accounting for the RF field B1, the B0-related frequency shift Δω and longitudinal and transverse relaxations described by R1 and R2* relaxation rate constants, respectively. For our purpose, it is convenient to present the Bloch equations in the rotating frame in the following matrix form:

Mt=A^M+P,A^=(R2*ΔωγB1yΔωR2*γB1xγB1yγB1xR1),P=(0,0,R1·M0)T, [2]

where M0 is the nominal magnetization and γ is the gyromagnetic ratio.

The solution of Equation [2] with the initial magnetization M(0) is

M(t)=exp(A^·t)·(M(0)V)+V,V=A^1·P. [3]

In what follows, we consider a sequence comprising a train of double-RF pulses. In each pair, the first RF α-pulse of duration τα and strength B1x(α=γ·B1x·τα) along the x-axis is immediately followed by the second RF β-pulse of duration τβ and strength B1y(β=γ·B1y·τβ) about the y-axis.

In the interval between the pairs of RF pulses, the longitudinal and transverse components of magnetization evolve as:

Mn+1()=exp(A^0·(TRτατβ))·(Mn(+)V0)+V0, [4]

where Mn() and Mn(+) are the magnetizations before and after the n-th pair of the double RF pulses, respectively (the index n numerates pairs of the double RF pulses in the train), the matrix A^0 is given by Equation [3] with B1X,Y = 0, V0 = A^01  P, and TR is the repetition time. In what follows, we will only consider RF pulses with equal duration τα=τβ=τ.

The magnetization Mn(+) after the n-th pair of the double RF pulses can be presented as

Mn(+)=exp(A^βτβ)·exp(A^ατα)·(Mn(-)Vα)+exp(A^βτβ).(VαVβ)+Vβ,Vα,β=A^α,β1P, [5]

where the matrices A^α and A^β are given by Equation [2] with B1y = 0 and B1x = 0, respectively.

Although the matrix-form expression in Equation [5] for the magnetization Mn(+) looks rather simple, the corresponding explicit expressions for Mn(+) are very cumbersome and we do not provide them. However, Equations [4] and [5] can be readily used for obtaining analytical results under some simplifications and for numerical calculations in the framework of any matrix-operating software, (e.g., Wolfram “Mathematica,” Wolfram Research, Champaign, IL, and MATLAB, The MathWorks, Natick, MA).

In the brain, longitudinal and transverse relaxation times are ~1 s and ~20–50 ms, respectively, that is much longer than typical durations of RF pulses τ (<1 ms). Therefore, for sufficiently short TRs, the following inequalities take place:

R1·τR1·TR1,R2*·τ1. [6]

In this case, using Equations [4] and [5], it can be demonstrated that under steady-state condition and ideal spoiling (the transverse magnetization is completely annihilated before the RF pulses), in the linear approximation with respect to the small parameters R1·τ, R2*·τ,τ/TR, and R1 · TR, the difference of the phases for 2 pairs of the RF pulses (α, α) and (α, _α) can be presented as

Δφ=arctan[Q(α,p)+(τTR)·Q1(α,p)+(R2*·τ)·Q2(α,p)], [7]

where the function Q(α,p) is given in Equation [1]. Importantly, in this approximation, the relaxation rate constant R1 does not enter Equation [7] (this constant appears only in the second-order terms with respect to the small parameters mentioned above). The parameter (R2*·TR), which is not small, does not enter Equation [7] because of the assumption of ideal spoiling (this issue will be addressed below). Note, however, that this equation should be modified for longer TR ~ T1. The expressions for the function Q1,2(α,p) for an arbitrary value of p=Δω·τ are given in Equations [A1] and [A2] in Appendix.

Effects of Imperfect RF Spoiling on Signal Phase

All the above considerations were based on the assumption of a perfect spoiling of the transverse magnetization in each TR cycle. A standard and broadly used RF spoiling approach (41) comprises a RF phase cycling scheme (each RF pulse is accompanied by a phase shift Ψk+1=Ψk+k·Ψ0, where Ψ0=117°,, and k is the RF pulse number) and an additional spoiling gradient preceding the next RF pulse. However, computer simulations (see below) revealed that for double RF pulses used in our approach, the effects of this standard RF spoiling turns out to be not sufficient to eliminate the transverse magnetization before the next RF pulse, which may lead to artifacts in B1 mapping.

A development of a new RF spoiling scheme applicable to double RF pulses is beyond the scope of the present paper. Instead, in this study, we propose another approach based on introducing a correction to the expression for the phase difference Δφ, Equation [7], accounting for incomplete spoiling of the transverse magnetization.

To examine the effect of RF spoiling of the transverse magnetization on signal phase in our approach, we used computer simulations. For this purpose, using Equations [4] and [5], the Bloch equations were numerically solved for the train of consecutive TR cycles. To simulate the effect of the spoiling gradient, 50 isochromats were included in the simulations. An additional phase assigned to each isochromat was added to the isochro-mat’s phase at the end of each TR cycle to simulate effects of spoiling gradient. These additional phases were uniformly distributed over the interval [0, 2π] among isochromats. By the end of the simulation, the magnetization of the 50 isochromats were averaged. For TR = 40 ms and relaxation parameters typical for gray matter (T1 = 1200 ms, T2* = 70 ms) and white matter (T1 = 800 ms, T2* = 50 ms), the magnetization was shown to reach the steady- state within 60 steps, respectively. For CSF (T1 = 4000 ms, T2* = 400 ms), ~250 steps are required.

The simulations were done for 2 cases: (1) perfect spoiling (ps): transverse magnetizations were artificially set to zero before each pair of RF pulses, and (2) standard experimental RF spoiling: using the scheme described above. All simulations were done in MATLAB (The MathWorks) for equal durations and flip angles in the RF pulse pair (α = β = 90°). The results of simulations and their analysis are illustrated in Figure 1.

FIG. 1.

FIG. 1.

Effect of imperfect RF spoiling on signal phase. (a) The difference between Δφ simulated with the standard RF spoiling with Ψ0=117°(Δφexp) and that simulated with the perfect spoiling (Δφps) as a function of the frequency shifts Δf for different T1(T2*=60 ms, TR = 40 ms τ=0.4ms). (b) The simulated difference Δφexp - Δφps as a function of the frequency shifts Δf for different T2* (symbols) (T1=1s,TR=40ms,τ=0.4ms); the lines, fitting by the function in Equation [8]. (c) The fitting parameter κ (appearing in Equation [8]) as a function of the dimensionless product, (R2*TR) for different RF pulse duration T (symbols); the line, fitting by the function in Equation [8].

Obviously, for sufficiently long repetition times compared to T*2 relaxation time constant, (TR > T2*), the transverse magnetization before the next RF pulse becomes negligible, and no spoiler is required. However, for shorter TR, Δφexp deviates from Δφps at non-zero frequency shifts Δf (see Fig. 1). As demonstrated in Figures 1a and 1b, the difference δφ = Δφexp – Δφps is practically independent of the longitudinal relaxation time T1, but sensitive to the transverse relaxation time T2*. The dependence of δφ on the frequency shift Δf is, generally, rather complicated (Fig. 1a). However, in brain applications, frequency shifts caused by the macroscopic field inhomogeneity usually do not exceed ±300 Hz because of the good shimming capability of modern MRI scanners (42). In this interval, the simulations show that for the relaxation rate constant R2* typical for brain tissues the difference δφ depends mainly on the dimensionless parameters R2*·TR and p=2π·Δf·τ and can be well approximated by the following empirical expression:

ΔφexpΔφps=tanh(κ·p), [8]

where

κ=2.38exp[2.83(R2*TR)0.65] [9]

It should be underlined that this expression is not universal and valid only in the intervals, |Δf|·τ<0.15,0.2<R2*TR<5.5.

Taking into account Equation [8], Equation [7] can be modified as follows:

Δφexp=arctan[Q(α,p)+(τTR)·Q1(α,p)+(R2*·τ)·Q2(α,p)]+tanh(κp), (10)

where the functions Q(α, p) and Q1,2(α, p) are given in Equation [1] and Equations [14] and [15], respectively. This expression can be used to account for the imperfect spoiling effect.

The importance of this correction is demonstrated in Figure 2, where the dependences of the parameter q (the ratio of the measured and input flip angles) on TR and Δf are shown for the values of T1 and T2* characteristic for WM (T1 = 800 ms, T2* = 50 ms), GM (T1 = 1200 ms, T2* = 70 ms), and CSF (T1 = 4000 ms, T2* = 400 ms). The top panels illustrate these dependencies for the results calculated from Equation [10] without the last term, whereas the lower panels illustrate the results obtained from Equation [10] with the last term.

FIG. 2.

FIG. 2.

Maps of the parameter q (the ratio of the measured and input flip angles) as a function of Δf and TR before (top) and after (bottom) correction (accounting for imperfection of RF spoiling by means of the last term in [Eq. 10]). Simulations were performed for 3 sets of parameters typical for different tissue types: WM (T1 = 800 ms T2*=50MS, GM, (T1=1200ms,T2*=70ms), and CSF, T1=4000ms,T2*=400ms).

It should be emphasized that the phenomenological expression for the correction term in Equation [10] is derived for the specific value of the angle, Ψ0 = 117°. However, different scanner manufacturers use different values of this parameter, for example, ,Ψ0 = 117°, (VARIAN), Ψ0 = 50° (SIEMENS), and Ψ0 = 115.4° (GE). Although we cannot provide a general expression for the correction factor for arbitrary RF spoiling scheme, the analysis of the cases with Ψ0 = 50° and Ψ0 = 115.4° revealed that they still can be described in terms of Equation [10] but with the slightly different dependence of the coefficient κ on the product, R2*×TR:

ψ0=50°:κ=7.53exp[3.99(R2*TR)0.46]ψ0=115.4°:κ=3.95exp[3.27(R2*TR)0.49]. [11]

The correction factors for other Ψ0 can be derived on a case-by-case basis in a manner similar to described above (the authors will be happy to assist for any other specific value of Ψ0).

METHODS

MR Experiment

All experiments were performed on 3T SIEMENS TRIO (VB17) MRI scanner. In our experimental settings, we used 2 orthogonal rectangular RF pulses (flip angle α = 90°) with total duration of 800 μs and a 20 μs gap in between allowing for RF phase shift. The RF duration was selected as the shortest under the specific absorption rate limitation (number of echoes = 6, TE1 = 2 ms, and ΔTE = 4 ms). A navigator echo was collected during each TR to correct for artifacts induced by physiological fluctuation, as described previously (43). In all scans, the body coil was used as a transmitter. A 3D multi-gradient recalled echo (GRE) sequence with a combined non- selective RF excitation pulse consisted of the (αx,α±y) pairs was used as depicted in Figure 3.

FIG. 3.

FIG. 3.

Schematic diagram of the pulse sequence. RF, RO, and PE represent radio frequency, read-out, and phase encoding, respectively. As in a standard RF spoiling procedure, for each subsequent TR, the phase of each RF pulse in the (α,α)-pair and the phase of receiver are changed by the increments of ΔΨk=kΨ0, where k is the phase encoding step number and Ψ0 is a hardware-based parameter (in our Siemens scanner Ψ0 = 50°). The dephasing gradient is applied to destroy residual signal at the end of each TR block.

Phantoms were prepared by adding 3.75 g of NiSO4 × 6H2O and 5 g NaCl per 1L of water. Experiments were run on: (1) a small spherical phantom with a radius of 1.5 cm using a 15-channel knee coil and image resolution of 1 × 1 × 1 mm3; (2) a spherical phantom with a radius of 2.5 cm using a 12-channel head coil and image resolution of 4 × 4 × 4 mm3; and (3) a healthy volunteer using a 32-channel head coil and image resolution of 2 × 2 × 2 mm3.Human studies were approved by the local Institutional Review Board. The healthy volunteer provided informed consent.

For phantom experiments, the frequency shifts were introduced by (1) changing the B0 frequency shift, and (2) changing the shimming profile along 1 phaseencoding direction. Two different TRs (30 ms and 200 ms) were used to compare the RF spoiling effect. To minimize effects resulting from the presence of voxels with coexisting fat and water compartments in human experiments, an additional Gaussian-type fat-suppression RF pulse preceding the phase-sensitive RF excitation pulse was added (bandwidth, 375 Hz; duration, 5120 μs). With TR = 40 ms, a resolution of 2 × 2 × 2 mm3 and GRAPPA (44) algorithm (acceleration factor of 2 and 24 autocalibrating lines in each phase encoding direction), each scan on the healthy volunteer required 1 min 40 s to cover the whole brain and the upper spinal cord.

To demonstrate the B1 effect on quantitative measurements of tissue relaxation parameters, we have also collected simultaneous multi-angular relaxometry of tissue (SMART) MRI data (22) with multiple flip angles of 5°, 10°, 20°, 40°, and 60°. This approach is based on (1) collecting GRE MRI data with multiple gradient echoes (we used gradient echo times TE = 2.26, 6.19, and 10.12 ms) and multiple flip angles (in this experiment we used 5°, 10°, 20°, 40°, and 60°), and (2) a new theoretical expression for GRE signal that takes into account the presence in the tissue of “free-” and “bound-water” pools (see details in 22). By using a high resolution of 1 × 1 × 1mm3, TR = 18 ms and GRAPPA, the total acquisition time of the SMART sequence was 17min 30 s. Images from B1 scans were registered to SMART images using “flirt” (45,46) tool in “FSL.” SMART data was then processed using the same procedure as described in (22). The maps of the SMART parameters–R1app (apparent longitudinal relaxation rate constant), R1f (longitudinal relaxation rate constant of the free-water compartment), and kapp (apparent magnetization exchange rate constant)–were calculated with and with-out B1 correction.

Data Analysis

Raw data were loaded into MATLAB (The MathWorks) for processing that required several steps.

  1. Using a strategy developed in (47,48), data from different channels are combined for each voxel in a single data set S(TE):
    S(TEn)=1M·m=1Mλm·Sm*(TE1)·Sm(TEn)λm=1Mσm2·l=1Mσl2, [12]
    where the sum is over all the channels (M), S* denotes complex conjugate of S, λm are the weighting parameters, and σm are the noise amplitudes (the index corresponding to the voxel position is omitted for clarity). This algorithm allows for the optimal estimation of quantitative parameters, such as MR signal frequency and decay rate constants and also removes the initial phase incoherence between the channels.
  2. The frequency shift (Δω = 2π . Δf and R2* maps are calculated by fitting a mono-exponential function of echo time to the signal (generated from Equation [12]).

  3. To get the phase difference Δφexp, data from different channels and different scans are combined for each voxel according to Equation [13]:
    Δφexp=arg[1M·n=1Nm=1Mλm·Sm(1)(TEn)·Sm(2)*(TEn)], [13]
    where the sum is over all channels and all echoes (N). The superscripts (1) and (2) correspond to the first and the second scans, respectively. This procedure is similar to the previously proposed by Bernstein et al. (40) for phase contrast MR angiography. It removes the initial phase φ0 in the channels and increases the signal-to-noise ratio (SNR) in the Δφ image.
  4. The measured flip angles αexper are calculated by solving Equation [10] (using the results of steps 2 and 3) on the voxel-by-voxel basis. The results of flip angle measurements are represented by a parameter q=αexper/αinput.

RESULTS

The accuracy of our method was first tested on a small phantom (1.5 cm in radius) that was placed at the isocenter of the scanner. In this phantom, that had a size much smaller than the characteristic size of RF transmit coil (body coil), no significant inhomogeneities were found (see Fig. 4): the parameter q was equal to 1.00 ± 0.03 (mean ± STD). This result provides experimental validation of our method as it shows an excellent agreement with the result obtained by the standard SIEMENS RF transmitter calibration procedure (4951).

FIG. 4.

FIG. 4.

Map and histogram of the parameter q obtained from a small spherical phantom (1.5 cm in radius) using the sequence with TR = 27 ms. The histogram shows very narrow distribution with q = 1.00 ± 0.03 (mean ± STD) therefore validating the accuracy of our approach against a gold standard RF calibration.

The theoretical results for different frequency shifts were tested on a spherical phantom (2.5 cm in radius). As shown in Figure 5 (lower panels), no imperfect spoiling effects were observed for long TR (200 ms). However, when TR was reduced to 30 ms, the estimated flip angles αexper (and corresponding histograms of the parameter q) obtained by using Equation [1] were shifted in the direction depending on the sign of the frequency shift Δf, that associated with the imperfect spoiling of the transverse magnetization. However, when αexper was calculated from Equation [10], the q-maps and corresponding histograms become practically identical with the maps and histograms obtained for long TR, as shown in Figures 5 and 6, respectively.

FIG. 5.

FIG. 5.

Maps of the parameter q obtained from a spherical phantom (2.5 cm in radius) with 2 different TRs (30 and 200 ms) and B0-related frequency shifts (0 and ±200 Hz). For short TR (30 ms), q maps (3 left columns) were calculated without (top) and with (middle) correction according to Equation [10] with the correction factor in Equation [11] for our Siemens scanner (Ψ0 = 50°).

FIG. 6.

FIG. 6.

Histograms of the parameter q for the same experimental parameters as in Figure 5. (a) TR = 30 ms (before correction). (b) TR = 30 ms (after correction). (c) TR = 200 ms. Red, Δf =−200 Hz; green, Δf =0; blue, Δf =+200 Hz.

The validity of our correction approach has also been demonstrated in another experiment in which inhomoge-neous frequency shift was introduced by changing the shimming profile along 1 phase-encoding direction (see Fig-7).

FIG. 7.

FIG. 7.

Results obtained from a spherical phantom (2.5cm in radius) with different TRs (200 and 30 ms). Frequency shifts were introduced by changing the shimming profile along 1 phase-encoding direction. Calculated Δf and q maps are shown on the left. Histograms of q with different TR are shown on the right panel (green, TR = 200 ms; red, TR = 30 ms before correction; blue, TR = 30 ms after correction).

Note that before correction, the q map is strongly asymmetric along the phase-encoding direction and deviates from the result obtained for long TR when RF spoiling does not create artifacts. After correction, results for short TR practically coincide with the result for long TR.

The maps of the SMART parameters–R1app (apparent longitudinal relaxation rate constant), R1f (longitudinal relaxation rate constant of free-water compartment), and kapp (apparent magnetization exchange rate constant)– calculated without and with correction are shown in Figure 8. The corresponding histograms of the SMART parameters are shown in Figure 9 along with the difference maps (before and after B1 correction). The B1 correction of the flip angles (i.e., accounting for the deviation of the actual flip angles from their input scanner values) changes the values of the SMART parameters by up to 50%, 85%, and 50% for R1app, R1f, and kapp, respectively. With the strongest deviations in the areas of the large differences between assigned and actual flip angles (compare images in Fig. 9 with B1 map in Fig. 8)

FIG. 8.

FIG. 8.

Comparison of SMART parameters (R1app, R1f, and kapp) maps without (second row) and with correction (third row) for imperfect RF spoiling in B1 mapping. Top row: R2*, frequency shift (Δf, in Hz) and q maps. Units for R2*, R1app, R1f, and kapp colorbars are s–1. R1app, apparent longitudinal relaxation rate constant; R1f, longitudinal relaxation rate constant of the free-water compartment; kapp, apparent magnetization exchange rate constant (see details in (22)).

FIG. 9.

FIG. 9.

The difference maps and comparison of the histograms of SMART parameters (R1app, R1f, and kapp) without (dashed lines) and with correction (solid lines) for imperfect RF spoiling in B1 mapping. The difference maps were calculated by subtracting the SMART parameters before correction from the corresponding parameters after correction. After correction, the SMART parameters become more uniform across the entire brain, compared to those images before corrections The biggest changes are in the brain areas with strongest actual B1 deviation from the assigned B1 value. The contrast between WM and GM also becomes higher. The SMART maps are shown in Figure 8 above.

DISCUSSION

In this study, we focused on the effects of imperfect RF spoiling and relaxation on the phase-based B1 mapping technique with 2 orthogonal excitation RF pulses used for encoding B1 information in the signal phase. We demonstrated that the MR signal phase generated by the pair of successive orthogonal RF pulses depends not only on B1 field strength but on the MR signal relaxation properties (T1 and T2* ) and (most importantly) on imperfections of RF spoiling as well. These effects have not been considered previously. To account for these effects, we proposed to append a GRE echo train with multiple gradient echoes to the double RF pulse signal excitation. The gradient echo train of the GRE sequence allows mapping the local frequency shifts and transverse relaxation of the MR signal. Combining the phase, frequency and R2* relaxation time constant measurements provided by the proposed sequence with the theoretical expression derived for the MR signal phase, makes it possible to calculate the local B1 field strength. Because the MR signal phase depends on instrumental factors and is diverse across different RF channels in the phased-array coil, we used the previously developed optimized methods for combining multi-channel data (40,48), allowing for optimal parameters’ estimation.

Our computer simulation revealed that in the presence of B0-related frequency shifts, the standard RF spoiling procedure does not completely destroy the transverse magnetization excited by the pair of orthogonal RF pulses. Such incomplete spoiling substantially affects the accuracy of B1 mapping (if not accounted for) when using short TRs. As shown in Figure 2 (upper row), in the presence of frequency shifts Δf without accounting for RF spoiling artifact, the calculated q-values (the ratio of measured and input flip angles) for all 3 tissue types dramatically deviate from unity. The deviations are more severe for short TRs and higher frequency shifts. In this study, we propose a new method that allows B1 mapping even in the presence of imperfect RF spoiling. As shown in Figure 2 (bottom row), after applying the new approach the calculated q-values become very close to 1.

As expected, for long TRs (>100 ms), the imperfect spoiling effect is negligible for GM and WM. However, for short TRs, without correction, the parameter q sub-stantially deviates from 1, especially for CSF that has much longer T1 and T2* than WM and GM. The lower panels in Figure 2 clearly show that the correction procedure works well for WM and GM. Even for CSF, in which the imperfect spoiling effect is more severe than in other tissues because of its longer T1 and T2* the correction procedure substantially decreases the deviation of the parameter q from 1: |q – 1| < 0.1.

The new method was tested on spherical phantoms (Figs. [4 and 5], and 7). As described above, the computer simulation results have shown that frequency shifts, imperfect spoiling of the transverse magnetization and relaxation had significant adverse effect on B1 mapping results when using double RF pulses and short TRs. These effects were also revealed in the phantom results. As shown in Figures 5 and 7, the calculated q-values were severely distorted in the regions with high frequency shifts. After corrections, the q-values became consistent with those acquired with longer TRs, for which the effect of imperfect RF spoiling was minimal. For human studies, significant changes of SMART parameters were observed after applying corrections, which indicated that the correction accounting for imperfect RF spoiling and relaxation is crucial for phase- based B1 mapping techniques using double RF pulses.

In our study, we used a phased-array RF receive coil to map the B1 field generated by the transmit coil in each imaging voxel. To optimize this measurement, the signals from different channels of the phased-array were combined according to Equations [12] and [13], as proposed in (47,48). However, our approach, based on using 2 orthogonal excitation RF pulses, can also be used for mapping the field from each individual channel in transmit-receive coils. This is important for more accurate implementations of parallel transmission approaches (13).

Our approach has certain restrictions because it relies on an assumption that the system under consideration can be described in terms of a single compartment model (or a multi-compartment model where all tissue compartments in the voxel have the same MR frequency). This, however, does not create a problem for most biological tissues possessing sufficiently small frequency differences between the compartments. However, the situation is substantially different in the regions containing both the fat and the water components because the frequency shift between them is high enough (~430Hz at 3T). An accurate quantification of B1 in such areas requires an approach allowing for separation of the fat-water components or suppression of one of them. In our experiments, we use a Gaussian type fat-suppression RF pulse preceding the phase sensitive RF excitation pulse to minimize this problem. The current studies use non- selective RF pulses, which restricts this method to 3D applications. A slice-selective gradient in 2D applications could cause large frequency shifts, therefore bringing in additional artifacts.

CONCLUSION

In this study, we proposed and validated a phase-based mapping method that can be used to get fast, accurate B1 maps using multi-channel phased-array coil. Our theoretical consideration includes effects of MR signal longitudinal and transverse relaxation that were not considered in previous studies. We have also demonstrated that the standard RF spoiling is not sufficient for the sequence with 2 orthogonal RF pulses and might create systematic errors in the B1 mapping if not accounted for. To deal with this adverse effect, we proposed a new method that allows substantially improve B1 estimation even for short TR. We also demonstrated that the effect of imperfect RF spoiling becomes negligible for TR longer than 200 ms for human brain at 3T. The feasibility of this method is demonstrated on phantoms and a healthy volunteer.

APPENDIX

The functions Q1;2(α, β) entering Equation [7]

Q1(α,p)=N1(α,p)D1(α,p)N1(α,p)=2{p2Ω(3α4+10α2p2+4p4)++4p2cosΩ[Ω(2α4+3α2p2+2p4)α4sinΩ]++(α2+2p2)[(3α44α2p24p4).sinΩ+α2cos2Ω(p2Ω+α2sinΩ)]}D1(α,p)=(1cosΩ)α2Ω[α2+(α2+2p2)cosΩ]2 [A1]
Q2(α,p)=N2(α,p)D2(α,p)N2(α,p)=α2Ω(α44p4)cosΩ[7α4+100α2p2+8p4+2(3α4+8α2p2+4p4)cosΩ]++α4(α22p2)(α2+2p2)2Ωcos2Ω4α4(α2+4p2)(3α2+4p2)Ω4cotΩ212α2(α22p2)(7α6+12α4p28p6)Ωcsc2Ω2+2(2α43α2p2+2p4)Ω7csc4Ω2++2α2(3α8+8α6p2+14α4p4+24α2p6+24p8)sinΩ++2α2(α2+7p2)Ω6csc4Ω2sinΩ+α4(α6108α4p212α2p4+8p6)sin2Ω}D2(α,p)=2Ω(1+cosΩ)α4[α2+(α2+2p2)cosΩ]3 [A2]
Q1(α,0)=2(3+cos2α)αsinα(1+cosα)Q1(α,0)=(3+cos2α)(1+αcotα)2αsin3α [16]

Supplementary Material

Fig. S1.

Estimation of optimal flip angles. (a) The second flip angle, β* = β*(α), corresponding to the minimum of the function F(α,β) with respect to β at a given α. (b) The plot of the estimation error δq, corresponding to the flip angles pair (α,β*(α)). (c) The dependence of the specific absorption rate on the flip angle α for the flip angles pair (α,β*(α)). Both the pulses are assumed to have the same duration. The system and sequence parameters are: T1 = 1 s, TR=30 ms, and SNRd=50 for the sequence with α = β = 90 ΰ

ACKNOWLEDGMENT

The authors are grateful to Dr. Yulin Chang for providing a custom program for GRAPPA reconstruction. Jie Wen is a former fellow of National Multiple Sclerosis Society.

Footnotes

SUPPORTING INFORMATION

Additional Supporting Information may be found in the online version of this article.

REFERENCES

  • 1.Katscher U, Bornert P, Leussler C, van den Brink JS. Transmit SENSE. Magn Reson Med 2003;49:144–150. [DOI] [PubMed] [Google Scholar]
  • 2.Zhu Y Parallel excitation with an array of transmit coils. Magn Reson Med 2004;51:775–784. [DOI] [PubMed] [Google Scholar]
  • 3.Xu D, King KF, Zhu Y, McKinnon GC, Liang ZP. Designing multichannel, multidimensional, arbitrary flip angle RF pulses using an optimal control approach. Magn Reson Med 2008;59: 547–560. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Zhang Z, Yip CY, Grissom W, Noll DC, Boada FE, Stenger VA. Reduction of transmitter B1 inhomogeneity with transmit SENSE slice-select pulses. Magn Reson Med 2007;57:842–847. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Sharma A, Bammer R, Stenger VA, Grissom WA. Low peak power multiband spokes pulses for B1 (+) inhomogeneity-compensated simultaneous multislice excitation in high field MRI. Magn Reson Med 2015;74:747–755. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Voigt T, Katscher U, Doessel O. Quantitative conductivity and permittivity imaging of the human brain using electric properties tomography. Magn Reson Med 2011;66:456–466. [DOI] [PubMed] [Google Scholar]
  • 7.Zhang XT, Zhu SA, He B. Imaging electric properties of biological tissues by RF field mapping in MRI. IEEE Trans Med Imaging 2010;29: 474–481. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Zhang XT, van de Moortele PF, Schmitter S, He B. Complex B1 mapping and electrical properties imaging of the human brain using a 16-channel transceiver coil at 7T. Magn Reson Med 2013;69:1285–1296. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Fram EK, Herfkens RJ, Johnson GA, Glover GH, Karis JP, Shimakawa A, Perkins TG, Pelc NJ. Rapid calculation of T1 using variable flip angle gradient refocused imaging. Magn Reson Imaging 1987;5:201–208. [DOI] [PubMed] [Google Scholar]
  • 10.Deoni SC, Peters TM, Rutt BK. High-resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DES-POT2. Magn Reson Med 2005;53:237–241. [DOI] [PubMed] [Google Scholar]
  • 11.Christensen KA, Grant DM, Schulman EM, Walling C. Optimal determination of relaxation-times of Fourier-transform nuclear magnetic-resonance - determination of spin-lattice relaxation- times in chemically polarized species. J Phys Chem 1974;78: 1971–1977. [Google Scholar]
  • 12.Homer J, Beevers MS. Driven-equilibrium single-pulse observation of T1 relaxation - a reevaluation of a rapid new method for determining NMR spin-lattice relaxation-times. J Magn Reson 1985;63:287–297. [Google Scholar]
  • 13.Homer J, Roberts JK. Conditions for the driven equilibrium single pulse observation of spin-lattice relaxation-times. J Magn Reson 1987; 74:424–432. [Google Scholar]
  • 14.Wang HZ, Riederer SJ, Lee JN. Optimizing the precision in T1 relaxation estimation using limited flip angles. Magn Reson Med 1987;5: 399–416. [DOI] [PubMed] [Google Scholar]
  • 15.Homer J, Roberts JK. Routine evaluation of M0 ratios and T1 values from driven-equilibrium NMR-spectra. J Magn Reson 1990;89:265–272. [Google Scholar]
  • 16.Deoni SCL, Rutt BK, Peters TM. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magn Reson Med 2003;49:515–526. [DOI] [PubMed] [Google Scholar]
  • 17.Ernst RR, Anderson WA. Application of Fourier transform spectroscopy to magnetic resonance. Rev Sci Instrum 1966;37:93–102. [Google Scholar]
  • 18.Ou X, Gochberg DF. MT effects and T1 quantification in singleslice spoiled gradient echo imaging. Magn Reson Med 2008;59:835–845. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Mossahebi P, Yarnykh VL, Samsonov A. Analysis and correction of biases in cross-relaxation MRI due to biexponential longitudinal relaxation. Magn Reson Med 2014;71:830–838. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Bieri O, Scheffler K. On the origin of apparent low tissue signals in balanced SSFP. Magn Reson Med 2006;56:1067–1074. [DOI] [PubMed] [Google Scholar]
  • 21.Liu F, Block WF, Kijowski R, Samsonov A. Rapid multicomponent relaxometry in steady state with correction of magnetization transfer effects. Magn Reson Med 2016;75:1423–1433. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Sukstanskii AL, Wen J, Cross AH, Yablonskiy DA. Simultaneous multi-angular relaxometry of tissue with MRI (SMART MRI): theoretical background and proof of concept. Magn Reson Med 2017;77: 1296–1306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Hornak JP, Szumowski J, Bryant RG. Magnetic-field mapping. Magn Reson Med 1988;6:158–163. [DOI] [PubMed] [Google Scholar]
  • 24.Insko EK, Bolinger L. Mapping of the radiofrequency field. J Magn Reson A 1993;103:82–85. [Google Scholar]
  • 25.Akoka S, Franconi F, Seguin F, Lepape A. Radiofrequency map of an NMR coil by imaging. Magn Reson Imaging 1993;11:437–441. [DOI] [PubMed] [Google Scholar]
  • 26.Cunningham CH, Pauly JM, Nayak KS. Saturated double-angle method for rapid B1 + mapping. Magn Reson Med 2006;55:1326–1333. [DOI] [PubMed] [Google Scholar]
  • 27.Dowell NG, Tofts PS. Fast, accurate, and precise mapping of the RF field in vivo using the 180 degrees signal null. Magn Reson Med 2007;58:622–630. [DOI] [PubMed] [Google Scholar]
  • 28.Jiru F, Klose U. Fast 3D radiofrequency field mapping using echoplanar imaging. Magn Reson Med 2006;56:1375–1379. [DOI] [PubMed] [Google Scholar]
  • 29.Stollberger R, Wach P. Imaging of the active B1 field in vivo. Magn Reson Med 1996;35:246–251. [DOI] [PubMed] [Google Scholar]
  • 30.Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med 2007;57:192–200. [DOI] [PubMed] [Google Scholar]
  • 31.Helms G, Finsterbusch J, Weiskopf N, Dechent P. Rapid radiofrequency field mapping in vivo using single-shot STEAM MRI. Magn Reson Med 2008;60:739–743. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Nehrke K, Bornert P. DREAM-- a novel approach for robust, ultrafast, multislice B(1) mapping. Magn Reson Med 2012;68:1517–1526. [DOI] [PubMed] [Google Scholar]
  • 33.Nehrke K, Versluis MJ, Webb A, Bornert P. Volumetric B1 (+] mapping of the brain at 7T using DREAM. Magn Reson Med 2014;71: 246–256. [DOI] [PubMed] [Google Scholar]
  • 34.Oh CH, Hilal SK, Cho ZH, Mun IK. Radio frequency field intensity mapping using a composite spin-echo sequence. Magn Reson Imaging 1990;8:21–25. [DOI] [PubMed] [Google Scholar]
  • 35.Morrell GR. A phase-sensitive method of flip angle mapping. Magn Reson Med 2008;60:889–894. [DOI] [PubMed] [Google Scholar]
  • 36.Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch- Siegert shift. Magn Reson Med 2010;63:1315–1322. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Chang YV. Rapid B1 mapping using orthogonal, equal-amplitude radio-frequency pulses. Magn Reson Med 2012;67:718–723. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Sharma A, Tadanki S, Jankiewicz M, Grissom WA. Highly-accelerated Bloch-Siegert |B1+| mapping using joint autocalibrated parallel image reconstruction. Magn Reson Med 2014;71:1470–1477. [DOI] [PubMed] [Google Scholar]
  • 39.Bloch F, Siegert A. Magnetic resonance for nonrotating fields. Phys Rev 1940;57:522–527. [Google Scholar]
  • 40.Bernstein MA, Grgic M, Brosnan TJ, Pelc NJ. Reconstructions of phase contrast, phased array multicoil data. Magn Reson Med 1994; 32:330–334. [DOI] [PubMed] [Google Scholar]
  • 41.Zur Y, Wood ML, Neuringer LJ. Spoiling of transverse magnetization in steady-state sequences. Magn Reson Med 1991;21:251–263. [DOI] [PubMed] [Google Scholar]
  • 42.Koch KM, Rothman DL, de Graaf RA. Optimization of static magnetic field homogeneity in the human and animal brain in vivo. Prog Nucl Magn Reson Spectrosc 2009;54:69–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Wen J, Cross AH, Yablonskiy DA. On the role of physiological fluctuations in quantitative gradient echo MRI: implications for GEPCI, QSM, and SWI. Magn Reson Med 2015;73:195–203. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang JM, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA]. Magn Reson Med 2002;47:1202–1210. [DOI] [PubMed] [Google Scholar]
  • 45.Jenkinson M, Smith S. A global optimisation method for robust affine registration of brain images. Med Image Anal 2001;5:143–156. [DOI] [PubMed] [Google Scholar]
  • 46.Jenkinson M, Bannister P, Brady M, Smith S. Improved optimization for the robust and accurate linear registration and motion correction of brain images. Neuroimage 2002;17:825–841. [DOI] [PubMed] [Google Scholar]
  • 47.Quirk JD, Sukstanskii AL, Bretthorst GL, Yablonskiy DA. Optimal decay rate constant estimates from phased array data utilizing joint Bayesian analysis. J Magn Reson 2009;198:49–56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Luo J, Jagadeesan BD, Cross AH, Yablonskiy DA. Gradient echo plural contrast imaging--signal model and derived contrasts: T2*, T1, phase, SWI, Tlf, FST2*and T2*-SWI. Neuroimage 2012;60:1073–1082. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Perman WH, Bernstein MA, Sandstrom JC. A method for correctly setting the RF flip angle. Magn Reson Med 1989;9:16–24. [DOI] [PubMed] [Google Scholar]
  • 50.Carlson JW, Kramer DM. Rapid radiofrequency calibration in MRI. Magn Reson Med 1990;15:438–445. [DOI] [PubMed] [Google Scholar]
  • 51.Feiweier T, Heubes P, Speckner T. Method and magnetic resonance system for adjustment of the field strength of rf pulses. US Patents: US20070145975A1. June 28, 2007. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Fig. S1.

Estimation of optimal flip angles. (a) The second flip angle, β* = β*(α), corresponding to the minimum of the function F(α,β) with respect to β at a given α. (b) The plot of the estimation error δq, corresponding to the flip angles pair (α,β*(α)). (c) The dependence of the specific absorption rate on the flip angle α for the flip angles pair (α,β*(α)). Both the pulses are assumed to have the same duration. The system and sequence parameters are: T1 = 1 s, TR=30 ms, and SNRd=50 for the sequence with α = β = 90 ΰ

RESOURCES