Abstract
Rationale and Objectives
To assess if a liquid tissue surrogate for the liver (LTSL) can emulate contrast-enhanced liver parenchyma and lesions and quantify the impact of reduced-kV imaging as a function of lesion contrast, phase of enhancement, and phantom size.
Materials and Methods
First, CT attenuation of LTSL- and water-iodine solutions was measured as a function of iodine concentration and tube potential. For each solution, the iodine concentration was determined to emulate liver parenchyma at 120 kV. CT attenuation for both solutions was predicted for different tube potentials and compared to published patient data. Second, liver parenchyma in late arterial phase (LA: +92HU at 120kV) and portal venous phase (PV: +112HU at 120kV) was emulated using LTSL-iodine and a two-size phantom. Fourteen setups of hyper- and hypoattenuating lesions (lesion-to-parenchyma contrast (CLP) = -50 to +50HU) were created. Each combination of CLP, phase, and size was imaged at 80, 100, 120, and 140 kV at constant radiation dose. CT attenuation, CLP, and lesion-to-parenchyma contrast-to-noise ratio (CNRLP) were assessed and compared to a theoretical model.
Results
LTSL-iodine more accurately emulated the CT attenuation of liver parenchyma across different tube potentials compared to water-iodine solutions. The theoretical model was confirmed by the empirical measurements using LTSL-iodine solutions: attenuation, CLP, and CNRLP increased when the tube potential decreased (P<0.001). This trend was independent of lesion contrast, phase, and size. The absolute improvement in CLP and CNRLP, however, was inversely related to the magnitude of CLP at 140kV.
Conclusion
LTSL accurately emulated the energy-dependent CT attenuation characteristics of contrast-enhanced liver parenchyma and lesions. The relative improvement in CLP and CNRLP by applying reduced-kV imaging was independent of lesion contrast, phase, and size while the absolute improvement decreased for low-contrast lesions.
Keywords: Computed tomography, Liver, Lesion, Tube Potential, Liquid tissue surrogate
Introduction
Contrast-enhanced computed tomography (CT) has been a central tool in oncologic imaging which facilitates lesion detection, lesion characterization and assessment of tumor treatment response. In CT, lesions are usually detected and characterized based on their CT attenuation in Hounsfield units (HU), their lesion-to-parenchyma contrast (CLP), and/or their attenuation changes during different phases of parenchymal enhancement. Effort has been made to define CT attenuation thresholds that improve lesion characterization and help to differentiate benign from malignant disease 1. Traditionally, these thresholds are based on CT scans at 120 kV.
In recent years, there has been a trend towards imaging at tube potentials lower than the typical 120 kV in contrast-enhanced CT (reduced-kV imaging). Reduced tube potentials increase the photoelectric effect and decrease Compton scattering and therefore increase the CT attenuation of iodinated structures 2,3. This results in increased contrast between tissues with different iodine uptake, for example between a lesion and the surrounding parenchyma. Because the CT attenuation of contrast-enhanced structures is energy dependent, reduced-kV imaging leads to changes in the CT attenuation of organ parenchyma as well as in the CLP and contrast-to-noise ratio CNRLP between the parenchyma and lesions. This may lead to mischaracterization and misinterpretation in clinical practice, if the diagnosis is based on the quantitative measurement of these parameters and traditional attenuation thresholds are applied. This is particularly important in liver imaging, where reduced-kV imaging has been shown to improve lesion detectability and conspicuity 2–6. Furthermore, the effect of reduced-kV imaging might not be the same for different phases of enhancement of liver parenchyma (e.g. late arterial phase versus portal venous phase) as well as for hyperattenuating versus hypoattenuating lesions.
Therefore, it is desirable to estimate the effect of reduced-kV imaging on the attenuation characteristics of liver parenchyma and lesions as a function of lesion contrast and phase of parenchymal enhancement.
In phantom studies, parenchyma or lesions are often represented using a solution of distilled water or saline with iodinated contrast medium 2,7–10. However, this method does not reflect the energy dependence of the CT attenuation of liver parenchyma. FitzGerald et al. have recently proposed a method for producing liquid tissue surrogate (LTS) materials which more faithfully represent human tissues 11. These surrogates mimic the elemental composition and density of several human soft tissues, including the liver parenchyma (LTSL), and could improve the scientific investigation of kV-dependent imaging 12.
Therefore, the purpose of our study was to validate the use of LTSL for reduced-kV imaging by assessing the quantitative effect of reduced-kV imaging on HU, CLP, and CNRLP as a function of tube potential, lesion contrast, and phase of parenchymal enhancement compared to a theoretical model and published patient data.
Materials and methods
Study Design
Two different phantom experiments were conducted. In the first experiment, we measured the CT attenuation of LTSL-iodine and water-iodine solutions as a function of iodine concentration and tube potential. Based on these measurements, we calculated the iodine concentration of LTSL-iodine and water-iodine solutions required to produce the CT attenuation of patients’ liver parenchyma at 120 kV. We then calculated the predicted CT attenuation for each of those solutions at tube potentials of 80, 100, 120, and 140 kV and compared them to published patient data 2,13,14.
In the second experiment, based on the results of the first experiment, we emulated the CT attenuation of contrast-enhanced liver parenchyma in each of the late arterial and portal venous phases using LTSL-iodine solutions. We serially scanned a phantom filled with LTSLiodine representing liver parenchyma in each phase together with additional LTSL-iodine solutions representing relevant lesions of varying CLP: hyperattenuating lesions for the late arterial phase and hypoattenuating lesions for the portal venous phase. The effect of reducedkV imaging on CLP and CNRLP was assessed and compared to a theoretical model.
Experiment 1:
CT attenuation characteristics of water-iodine and LTSL-iodine solutions
A liquid tissue surrogate for liver (LTSL) was created by our Department of Chemistry using the formula described by FitzGerald et al. 11. In a first step, we investigated the CT attenuation measured at different tube potentials resulting from different concentrations of iodinated contrast medium in either distilled water or the proposed LTSL. Distilled water and LTSL were each mixed with iopamidol (300 mg iodine per mL, Isovue®-300, Bracco Diagnostics, Princeton, NJ) to achieve eight iodine concentrations ranging from non-iodinated water or LTSL to approximately 8 mg iodine per mL solutions of water or LTSL. We separately placed the 16 samples in 50-mL vials and simultaneously scanned them on a 64-detector-row CT scanner (Discovery CT 750 HD, GE Healthcare, Chicago, IL, USA) in single-energy mode at 80, 100, 120 and 140 kV using the large body bow-tie filter and helical mode. The tube current-time product (mAs) was adjusted for each tube potential to achieve a constant volume CT dose index (CTDIvol) of 6.2 mGy. This dose value was considered to be sufficient enough to image the vials in air. The scanner was calibrated in air immediately before scanning each setup. Images were reconstructed at a 1.5-mm slice thickness using filtered back projection and a standard kernel. CT attenuation was measured from the reconstructed images by drawing circular regions of interests (ROIs) in the center of each vial. The diameter of each ROI was 4 mm (area = 12.6 mm2). For each solution and tube potential, the mean value of the measurements from ten adjacent slices in the center of the vial was calculated. Using Excel (Microsoft Office 365 ProPlus, Microsoft, Redmond, WA, USA), we fitted linear equations to the averaged measurements of the CT attenuation versus iodine concentration for the LTSLiodine solutions and for the water-iodine solutions at each tube potential. The resulting equations allowed us to estimate the iodine concentration required to produce a given CT attenuation at a given tube potential using distilled water or LTSL.
Experiment 2:
Emulation of liver parenchyma and liver lesions
A custom-made phantom (Fig. 1) was fabricated using a 3-D printer (Connex 3, Objet 350, Stratasys, Eden Prairie, MI, USA). This phantom consists of a fillable chamber (1.9 liter) and 15 separately fillable cylinders of different sizes (diameters of 5 mm – 20 mm) and locations. The chamber emulates the hepatic parenchyma and is shaped to fit within one or two (7.7-cm and 15.6-cm total thickness, respectively) adipose-equivalent plastic rings to emulate a medium-sized (effective diameter 28.3 cm) and large-sized (effective diameter 36.2 cm) patient. The effective diameter was estimated as a geometric mean of the lateral and anterior-posterior dimensions of the phantom 15.
Figure 1.
Photo (left) and axial CT image (right) of the custom-made liver phantom with a fillable chamber (X) and 15 fillable cylinders (white arrow). The chamber was filled with LTSL-iodine solutions to emulate liver parenchyma in late arterial phase (shown, right) or portal venous phase. Eleven of the fifteen cylinders were filled with LTSL-iodine solutions of slightly higher or lower iodine concentrations to emulate hyper- or hypoattenuating liver lesions with different lesion-to-background contrast values (shown, right, with hyperattenuating lesions with CLP of +50 HU). Three cylinders were filled with neutral lesions (N) and one with distilled water (W). 5-mm, 6-mm, neutral, and water lesions were excluded from statistical analysis. One or two rings of fat-equivalent plastic were added to emulate medium- and large-sized patients (one ring is shown).
Based on the linear equations of experiment 1, we created LTSL-iodine solutions that mimic the CT attenuation of in vivo liver parenchyma. For the late arterial phase, we aimed to achieve the previously-published CT attenuation of 120, 106, 91, and 86 HU at 80, 100, 120, and 140 kV, respectively 2. For the portal venous phase, published data on CT numbers at 80 kV and 100 kV are not available because 120 kV is currently used as the routine in most clinical practices. Therefore, we aimed for the average CT attenuation of 112 HU at 120 kV that is within the range of published data 13,14.
First, the main chamber of the phantom was filled with the LTSL-iodine solution that emulates liver parenchyma in the late arterial phase (setup A). We created additional LTSL-iodine solutions with slightly higher iodine concentration than the parenchymal solution to emulate hyperattenuating lesions with a nominal CLP of +5, +10, +15, +20, +30, +40, and +50 HU. The lesion-representing cylinders of the phantom were filled with these additional solutions. Of the 15 cylinders, 3 were filled with the same solution as the main chamber to emulate “neutral” control lesions (CLP nominally 0), 1 cylinder was filled with distilled water, and, sequentially, the remaining 11 cylinders were all filled with the same one of the 7 solutions described above to emulate hyperattenuating lesions (setups A1-A7).
Second, the main chamber of the phantom was filled with the LTSL-iodine solution formulated to emulate the portal-venous phase (setup B). We created additional LTSL-iodine solutions mimicking hypoattenuating lesions with a nominal CLP of -5, -10, - 15, -20, -30, -40, and -50 HU. We filled the phantom’s lesion-representing cylinders using the same approach as for setup A (setups B1 to B7).
CT Scanning
Each of the overall 14 different setups was scanned separately using the medium and the large phantom sizes at 80, 100, 120, and 140 kV, resulting in 112 different CT scans. The scan conditions of Experiment 1 were used except that the tube current-time product (mAs) was adjusted to achieve a constant CTDIvol of 10 mGy in the medium-sized phantom and 15 mGy in the large-sized phantom. These radiation dose values were chosen according to the recently published dose ranges for abdominal CT of the ACR Dose Index Registry 16. The gantry rotation time as well as the pitch had to be reduced for certain reduced-kV settings to reach the desired CTDIvol. The variable scanning parameters are summarized in Table 1. The scanner was calibrated in air immediately before scanning each setup. All images were reconstructed at a slice thickness of 5 mm and an increment of 2.5 mm using an adaptive statistical iterative reconstruction (ASiR, GE Healthcare, Chicago, IL, USA) algorithm at 40%blending with filtered back projection (FBP).
Table 1.
CT scan parameters for each tube potential using the medium (M) and large (L) patient-sized phantom.
| 80 kV | 100 kV | 120 kV | 140 kV | |||||
|---|---|---|---|---|---|---|---|---|
| M | L | M | L | M | L | M | L | |
| X-ray tube current (mA) | 690 | 605 | 590 | 570 | 370 | 575 | 260 | 405 |
| Gantry rotation time (s) | 0.8 | 1.0 | 0.5 | 0.8 | 0.5 | 0.5 | 0.5 | 0.5 |
| Helical pitch | 1.375 | 0.984 | 1.375 | 1.375 | 1.375 | 1.375 | 1.375 | 1.375 |
Quantitative Assessment of CLP and CNRLP
CT attenuation was measured by drawing ROIs in each of the 11 lesion-emulating cylinders and in the adjacent parenchyma. The measurements were performed using ten consecutive slices at the center of the phantom length. A MATLAB (The Mathworks, Natick, MA, USA) script was developed and used to automate the measurements and standardize the position and size of the ROIs among all images. The diameter of each ROI was 4 mm (area = 12.6 mm2) for the cylinders and 20 mm (area = 314 mm2) for the adjacent parenchyma. We chose to exclude the 5-mm and 6-mm cylinders from analysis in order to minimize potential measurement errors produced by the wall of the cylinders. For each cylinder of each dataset, lesion-to-parenchyma contrast (CLP) was calculated as abs(ROIL – ROIP), where ROIL is the mean CT attenuation in the lesion ROIs averaged over ten slices and ROIP is the mean CT attenuation of the adjacent parenchyma ROIs averaged over ten slices. Lesion-to-parenchym contrast-to-noise ratio (CNRLP) of each cylinder was calculated as CLP / N, where N is the standard deviation of the CT attenuation in the parenchyma ROIs, averaged over ten slices.
Theoretical impact of reduced-kV imaging on CLP and CNRLP
CT attenuation is proportional to iodine concentration 4 and directly related to the x-ray energy spectrum, which is dependent on the tube potential 17. Therefore, the CT attenuation of a lesion (L) at a certain tube potential VL can be described by the linear equation:
where m represents the slope of the CT attenuation-to-iodine concentration function in HU/mg iodine/mL for a certain VL, IL represents the iodine concentration of the lesion in mg/mL, and bL represents the zero-iodine-concentration intercept coefficient for the lesion tissue in HU. Similarly, the CT attenuation of the adjacent parenchyma at VL can be described as:
where IP represents the iodine concentration of the parenchyma in mg iodine/mL, and bP represents the zero-iodine-concentration intercept coefficient for the parenchymal tissue in HU. Here, we assume that the slope m for liver parenchyma and liver lesion is the same (which is most likely reasonable for well-differentiated hepatocellular carcinomas).
For a given case, CLP at a given tube potential VL is therefore:
In other words, CLP at a certain tube potential depends on i) the slope of the CT attenuation-to-iodine concentration curve for that specific tube potential, ii) the difference in iodine concentration in the lesion and the parenchyma, and iii) the difference between the unenhanced attenuation of the lesion and the parenchyma. For liver tumors for which (iii) is nearly 0, the b terms cancel. This is the case in our experiment, because unenhanced parenchyma and lesions are identical – LTSL.
Based on our fitted linear equations of the LTSL solutions, we created a theoretical model estimating the change in CLP at reduced tube potentials of 80, 100, and 120 kV compared to the baseline of 140 kV. Assuming that image noise is rather constant at different tube potentials at constant radiation dose, the same model can be applied to estimate the change in CNRLP. Further, the measured change (compared to 140 kV) in CLP and CNRLP was calculated for all the measured data. Linear fits were done to determine if the measured data corresponded to the theoretical change in CLP and CNRLP that would be expected for 80, 100, and 120 kV compared to 140 kV.
Statistics
A multi-variate statistical model (linear mixed-effects) was applied to all measured data to determine how measured CLP and CNRLP were affected by tube potential, nominal contrast, acquisition phase, and phantom size. Fixed terms were used for tube potential, nominal contrast, acquisition phase, and phantom size, and a conditional random term (tube potential given nominal contrast, assumed normal distribution) was also included. The statistical model was fitted to the data using MATLAB’s statistics and machine learning toolbox (fitglme function, maximum pseudo likelihood fitting method). P-values less than 0.00625 were considered significant (0.00625 was chosen using a Bonferroni correction of 0.05 / 8, due to having 2 models with 4 terms each).
Results
Experiment 1:
CT attenuation characteristics of water-iodine and LTSL-iodine solutions
We found that the CT attenuation of LTSL-iodine solutions (Table 2) and distilled water-iodine solutions (Table 3) as a function of iodine concentration and tube potential were substantially different. Figure 2 shows the linear fits for the LTSL-iodine and water-iodine solutions at all four tube potentials. The baseline CT attenuation of LTSL without the addition of iodine was 62 HU at 120 kV which closely matches the CT attenuation of liver parenchyma in non-contrast CT.
Table 2.
CT attenuation of LTSL-iodine solutions at various iodine concentrations and each tube potential.
| Iodine concentration (mg iodine /mL solution) |
CT attenuation (HU) of LTSL-iodine solutions |
|||
|---|---|---|---|---|
| 80 kV | 100 kV | 120 kV | 140 kV | |
| 0.00 | 77 | 68 | 62 | 62 |
| 0.82 | 123 | 105 | 92 | 86 |
| 1.11 | 138 | 115 | 101 | 94 |
| 1.64 | 163 | 135 | 118 | 108 |
| 2.26 | 194 | 160 | 138 | 126 |
| 4.40 | 296 | 240 | 204 | 181 |
| 6.36 | 385 | 311 | 262 | 232 |
| 8.53 | 477 | 382 | 322 | 283 |
Table 3.
CT attenuation of distilled water-iodine solutions at various iodine concentrations and each X-ray tube potential.
| Iodine concentration (mg iodine /mL solution) |
CT attenuation (HU) of distilled water-iodine solutions |
|||
|---|---|---|---|---|
| 80 kV | 100 kV | 120 kV | 140 kV | |
| 0.00 | 15 | 11 | 6 | 4 |
| 0.60 | 50 | 35 | 24 | 19 |
| 1.08 | 72 | 53 | 40 | 33 |
| 1.74 | 105 | 78 | 61 | 52 |
| 2.20 | 126 | 94 | 75 | 64 |
| 3.91 | 209 | 160 | 129 | 110 |
| 6.32 | 320 | 248 | 201 | 173 |
| 8.06 | 400 | 310 | 252 | 217 |
Figure 2.
Plot of CT attenuation of LTS-iodine solutions (brown) and distilled water-iodine solutions (blue) at tube potentials of 80, 100, 120, and 140 kV. Measured data are plotted as markers; the lines are the fitted data, and the coefficients for the linear fits to each curve are shown in the legend. R2 values were 0.999 for all linear fits.
We calculated the required iodine concentration of a LTSL-iodine and water-iodine solution to emulate the CT attenuation of patients’ liver parenchyma at 120 kV and assessed how the CT attenuation changes at different tube potentials (at constant iodine concentration). We found that the LTSL-iodine solution closely matched the CT attenuation of published patient data at different tube potentials whereas the water-iodine solution led to a higher attenuation values at reduced tube potentials. Figure 3 shows CT attenuation as a function of kV for LTSL-iodine and water-iodine solutions (at a fixed iodine concentration) to illustrate this result. As can be seen, higher iodine concentrations must be used for water-iodine solutions compared to LTSL-iodine solutions to emulate the CT attenuation of liver parenchyma at a desired baseline tube potential. This increase in iodine concentration will then cause additional overestimation of the CT attenuation at reduced-kV setting.
Figure 3.
Plot showing the CT attenuation of liver parenchyma in patients (black) and of LTS-iodine solutions (brown) and distilled water-iodine solutions (blue) at different tube potentials as calculated from calibration equations. The iodine concentration in these hypothetical solutions was chosen to produce 91 HU at 120 kV, thereby matching the published value for liver in patients 2.
Experiment 2:
Theoretical impact of reduced-kV imaging on CLP and CNRLP
Using the fitted linear equations of experiment 1, we estimated the theoretical change in CLP and CNRLP at reduced tube potentials of 80, 100, and 120 kV compared to the CLP at 140 kV (model for CLP shown in Figure 4). These data show that CLP and CNRLP increase at reduced tube potentials and the increase is greater for lesions with a high CLP at 140 kV (e.g. 50 HU) compared to lesions with a low CLP at 140 kV (e.g. 5 HU).
Figure 4.
Plot showing theoretical change in lesion-to-background contrast (CLP) at 80 (purple), 100 (light blue), and 120 kV (yellow) compared to 140 kV (baseline) versus the nominal CLP. CLP improves by reducing the tube potential. This improvement is proportional to the nominal CLP of the lesion at 140 kV.
Quantitative assessment of CLP and CNRLP
The results of the quantitative analysis confirmed the realism of our theoretical model, which predicts that CLP and CNRLP improve when reducing the tube potential and that the absolute improvement is inversely related to the lesion contrast. Our first finding from the empirical result was that tube potential had a significant impact on CLP and CNRLP (P<0.001 for both parameters): both parameters increased when reducing the tube potential from 140 kV to 80 kV. When averaged over both phantom size and all lesion sizes, CLP at 140, 120, 100, and 80 kV was 16, 19, 24, and 31 HU, respectively, for the late arterial phase and 15, 18, 22, and 29 HU, respectively, for the portal venous phase. Similarly, CNRLP at 140, 120, 100, and 80 kV was 0.7, 0.9, 1.1, and 1.4, respectively, for the late arterial phase and 0.7, 0.8, 1.0, and 1.3, respectively, for the portal venous phase. Our second finding was that the increase in CLP and CNRLP by reducing the tube potential from 140 kV to 80 kV was higher for high-contrast lesions compared to low-contrast lesions. Figure 5 and Figure 6 show the measured CLP and the measured CNRLP for each tube potential, nominal lesion contrast (defined at 120 kV), phase of enhancement, and phantom size. For both acquisition phases and phantom sizes, the increase in CLP and CNRLP when reducing the tube potential from 140 kV to 80 kV increases with nominal CLP. In Figure 7, we further plotted the increase in CNRLP for 80, 100, and 120 kV compared to 140 kV versus the measured CLP; we applied a linear fit of the data points for each tube potential. The linear fits demonstrate that the change in CNRLP at reduced tube potentials is proportional to CLP at 140 kV.
Figure 5.
Box-and-whisker plot showing median values and quartiles of measured lesion-to-background contrast CLP for each tube potential, nominal lesion contrast (defined at 120 kV), phase of enhancement, and phantom size. For all lesions, the median CLP increased by lowering the tube potential from 140 kV to 80 kV. However, this median increase in CLP is higher for lesions with an initial high nominal contrast (for example 50 HU) compared to lesions with an initial low nominal contrast (for example 5 HU).
Figure 6.
Box-and-whisker plot showing median values and quartiles of measured absolute CNRLP for each tube potential, nominal lesion contrast, phase of enhancement, and phantom size. For all lesions, the median CNRLP increased by reducing the tube potential from 140 kV to 80 kV. This trend is similar in the medium phantom results.
Figure 7.
Plot showing the measured change in lesion-to-background CNRLP at 80, 100, and 120 kV relative to 140 kV for each measured lesion. The “CNR change” data are calculated for each measured sample as CNR (measured) – CNR (mean for same conditions at 140 kV); the data in this plot are prior to taking the absolute value, as was done for all other data in this report. The lines depict linear fits to all the data points for each of 80, 100, and 120 kV. Each data point represents the measured change in CNRLP for one image slice and one cylinder of the phantom.
Discussion
Our study investigated the utility of a liquid tissue surrogate to emulate the CT attenuation, CLP, and CNRLP of liver parenchyma and lesions as a function of tube potential, lesion contrast, phase of enhancement, and phantom size. Our results demonstrated that LTSL closely matched the CT attenuation of liver parenchyma in patients across different tube potentials. Because the unenhanced attenuation of water (0 HU) is lower than that of liver (approximately 60 HU), a higher iodine concentration is required when using a water-iodine solution to emulate the CT attenuation of liver parenchyma at 120 kV. Consequently, a water-iodine solution of the same iodine concentration would result in unrealistically increased CT attenuations at lower tube potentials. Therefore, the use of water-iodine solutions could lead to an overestimation and misinterpretation of the potential benefit of reduced-kV imaging in research studies and consequently to an inappropriate translation into clinical practice. On the other hand, LTSL solution emulates more accurate CT attenuation across tube potentials using only a single iodine concentration. This is preferable to investigate the impact and benefits of reduced-kV imaging and to optimize contrast-enhanced CT protocols. LTSs have important implications in research because solutions for various human body tissues can be created and various different phases of enhancement can be emulated by mixing LTS with various iodine concentrations. In addition, this also has potential important clinical implications because, to date, the characterization of certain pathologies is based on CT number thresholds which were defined at 120 kV. For example, an enhancing renal lesion is defined as a lesion that demonstrates an increase in CT attenuation of more than 15 HU from unenhanced to enhanced phase 18. However, this threshold is only applicable at 120 kV. Taking into account the fact that CT attenuation of iodinated structures is energy-dependent, this threshold would shift if reduced tube potentials are used. Therefore, current guidelines which apply CT attenuation thresholds suffer from the dilemma that these thresholds are not adjusted to the emerging application of reduced-kV imaging. LTS solutions offer the possibility to emulate specific tissue and tissue pathologies like lesions and to investigate and define preliminary energy-dependent CT attenuation thresholds which could then be validated in clinical studies. In addition, LTS solutions are more cost-effective compared to plastic tissue-equivalent phantoms, especially in the evaluation of contrast-enhanced tissues. Furthermore, LTS solutions could be used to perform dynamic experiments like perfusion studies.
In the second part of our study we assessed the feasibility of LTSL in the quantification of CLP and CNRLP of lesions at different lesion contrast and different phases of parenchymal enhancement as a function of tube potential and compared these results to a theoretical predictive model. We investigated the improvement of CNRLP relative to 140 kV as the baseline because 140 kV had the theoretical lowest CNR values. The empirical quantitative measurements using LTSL demonstrated equal trends compared to the predictive model. The results confirmed that reduced-kV imaging improved CLP and CNRLP for both hyperattenuating lesions in the late arterial phase and for hypoattenuating lesions in the portal venous phase. Interestingly, this improvement was independent of the phase of enhancement of the liver parenchyma and also independent of whether a lesion was hyper- or hypoattenuating, but depended only on the difference in iodine concentration between a lesion and the surrounding parenchyma. For example, the improvement in CNRLP by reducing the tube potential from 140 kV to 80 kV was equal for a hyperattenuating lesion of +40 HU at 140 kV in late arterial phase and for a hypoattenuating lesion of -40 HU at 140 kV in portal venous phase because both lesions have the same difference in iodine concentration compared to their surrounding parenchyma.
Former studies have described the improvement in quantitative image quality parameters like CLP and CNRLP by reduced-kV imaging 2,3. Compared to these studies, we also investigated the impact of different lesion characteristics and parenchymal phases of enhancement. Our findings demonstrate that the absolute improvement in CLP and CNRLP when using reduced-kV imaging depends on the initial CLP at the baseline tube potential. For instance, a lesion with a CLP of 5 HU at 140 kV shows less improvement in CLP with reducedkV imaging compared to a lesion with a CLP of 20 HU. If we interpret this finding in a clinical context, the visibility of lesions that have high CLP at 140 kV improves more when using reduced-kV imaging compared to the visibility of lesions with low CLP at 140 kV. This is especially interesting considering that one of the main purposes of reduced-kV imaging is to improve the visibility and conspicuity of lesions which are difficult to depict at higher tube potentials. In other words, lesions that are difficult to visualize because they have a low CLP at higher tube potentials of 120 kV or more may not demonstrate a significant improvement in conspicuity when using reduced tube potential imaging. Based on our work, further studies are needed to define the threshold of initial CLP for which reduced tube potential imaging is beneficial on lesion’s visibility.
Several potential limitations of our study merit consideration. First, we did not have access to patient data on CT attenuation of the liver parenchyma in the portal venous phase at 80 kV and 100 kV because 120 kV is routinely used in most clinical practices. However, we adjusted the CT attenuation of our emulated LTSL solution to match patient data at 120 kV from a previous study population. Second, we used LTSL-iodine solutions as a straightforward alternative to water-iodine solutions. LTS for blood is also available 11. A future study could use an appropriate mixture of LTS liver and LTS blood for unenhanced liver, and add appropriate concentrations of iodine to represent contrast-enhanced liver parenchyma and lesions. Because LTS liver and LTS blood have similar linear attenuation coefficients for clinically-used CT spectra 11, we believe this would be a second-order effect in our study. Third, we did not perform a human observer or model observer study to grade the improvement of visibility and conspicuity of the lesions by reducing the tube potential. Previous studies have shown that the detectability of liver lesions in clinical routine is not necessarily correlated to quantitative image quality parameters alone 19–22. Model observers can be extremely useful in providing a quantitative measure of conspicuity. They are particularly useful when comparing images which have simultaneous differences in noise texture (noise power spectrum) and resolution. In this study, the noise texture and resolution are not drastically different between the lesions or images being compared. Thus, CNR should be a relatively good surrogate for conspicuity in this case. Furthermore, the main purpose of our study was to validate the use of LTSL to emulate energy-dependent CT attenuation characteristics. Fourth, we did not assess the impact of potential beam hardening at lower tube potentials on CT attenuation (and consequently CLP and CNRLP) or lesion visibility. Especially bone structures can cause beam hardening at low tube potentials in patients and therefore alter image quality parameters. A future phantom study could include bone structures to account for this.
Conclusions
In conclusion, our phantom study confirmed that liquid tissue surrogates for the liver accurately emulate the energy-dependent CT attenuation characteristics of liver parenchyma and liver lesions at different tube potentials. Our results indicate that the relative improvement in CLP and CNRLP at reduced tube potentials is independent of lesion contrast, phase of enhancement, and phantom size while the absolute improvement is inversely related to the lesion contrast. Visibility of low-contrast lesions, which are most difficult to detect in clinical routine, may be increased far less than high-contrast lesions when reduced-kV imaging is used. Future studies should evaluate the impact of reduced-kV imaging on the accuracy of lesion detection dependent on lesions’ contrast properties.
Acknowledgements
Research reported in this publication was supported in part by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Award Number R01EB015476. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. Andre Euler received financial research support by the Swiss Society of Radiology.
Source of Funding:
Andre Euler receives a research fellowship supported by GE Healthcare and research support by the Swiss Society of Radiology.
Paul FitzGerald is an employee of GE Global Research.
Ehsan Samei receives research funding from GE Healthcare and Siemens Healthineers and is on the advisory board of medInt Holdings, LLC, for projects unrelated to the present study.
Rendon C. Nelson is a medical consultant to GE Healthcare.
Research reported in this publication was supported in part by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under Award Number R01EB015476.
Abbreviations
- CT
Computed tomography
- HU
Hounsfield unit
- CLP
Lesion-to-parenchyma contrast
- CNRLP
Lesion-to-parenchyma contrast-to-noise ratio
- LTS
Liquid tissue surrogate
- LTSL
Liquid tissue surrogate for liver
- CTDIvol
Volume CT dose index
Footnotes
Conflicts of Interest
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Patel NS, Poder L, Wang ZJ, et al. The characterization of small hypoattenuating renal masses on contrast-enhanced CT. Clin Imaging. 2009;33(4):295–300. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Schindera ST, Nelson RC, Mukundan S Jr.,, et al. Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection--phantom study. Radiology. 2008;246(1):125–132. [DOI] [PubMed] [Google Scholar]
- 3.Robinson E, Babb J, Chandarana H, Macari M. Dual source dual energy MDCT: comparison of 80 kVp and weighted average 120 kVp data for conspicuity of hypo-vascular liver metastases. Invest Radiol. 2010;45(7):413–418. [DOI] [PubMed] [Google Scholar]
- 4.Marin D, Nelson RC, Samei E, et al. Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection--initial clinical experience. Radiology. 2009;251(3):771–779. [DOI] [PubMed] [Google Scholar]
- 5.Schindera ST, Hareter LF, Raible S, et al. Effect of tumor size and tumor-to-liver contrast of hypovascular liver tumors on the diagnostic performance of hepatic CT imaging. Invest Radiol. 2012;47(3):197–201. [DOI] [PubMed] [Google Scholar]
- 6.Lv P, Liu J, Zhang R, Jia Y, Gao J. Combined Use of Automatic Tube Voltage Selection and Current Modulation with Iterative Reconstruction for CT Evaluation of Small Hypervascular Hepatocellular Carcinomas: Effect on Lesion Conspicuity and Image Quality. Korean J Radiol. 2015;16(3):531–540. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Chandarana H, Megibow AJ, Cohen BA, et al. Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses. AJR Am J Roentgenol. 2011;196(6):W693–700. [DOI] [PubMed] [Google Scholar]
- 8.Karlo C, Lauber A, Gotti RP, et al. Dual-energy CT with tin filter technology for the discrimination of renal lesion proxies containing blood, protein, and contrast-agent. An experimental phantom study. Eur Radiol. 2011;21(2):385–392. [DOI] [PubMed] [Google Scholar]
- 9.Mileto A, Nelson RC, Samei E, et al. Dual-energy MDCT in hypervascular liver tumors: effect of body size on selection of the optimal monochromatic energy level. AJR Am J Roentgenol. 2014;203(6):1257–1264. [DOI] [PubMed] [Google Scholar]
- 10.Mileto A, Barina A, Marin D, et al. Virtual Monochromatic Images from Dual-Energy Multidetector CT: Variance in CT Numbers from the Same Lesion between Single-Source Projection-based and Dual-Source Image-based Implementations. Radiology. 2016;279(1):269277. [DOI] [PubMed] [Google Scholar]
- 11.FitzGerald PF, Colborn RE, Edic PM, Lambert JW, Bonitatibus PJ Jr., Yeh BM. Liquid tissue surrogates for X-ray and CT phantom studies. Med Phys. 2017. doi: 10.1002/mp.12617. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Photon, Electron, Proton and Neutron Interaction Data for Body Tissues.: International Commission on Radiation Units and Measurements;1992.
- 13.Fujigai T, Kumano S, Okada M, et al. Optimal dose of contrast medium for depiction of hypervascular HCC on dynamic MDCT. Eur J Radiol. 2012;81(11):2978–2983. [DOI] [PubMed] [Google Scholar]
- 14.Paparo F, Garello I, Bacigalupo L, et al. CT of the abdomen: degree and quality of enhancement obtained with two concentrations of the same iodinated contrast medium with fixed iodine delivery rate and total iodine load. Eur J Radiol. 2014;83(11):1995–2000. [DOI] [PubMed] [Google Scholar]
- 15.Brady SL, Kaufman RA. Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation. Radiology. 2012;265(3):832–840. [DOI] [PubMed] [Google Scholar]
- 16.ACR Dose registry. Executive summary report, Jul-Dec2017. https://nrdrsupport.acr.org/helpdesk/attachments/11027512192 Accessed July 28 2018.
- 17.FitzGerald PF, Colborn RE, Edic PM, et al. CT Image Contrast of High-Z Elements: Phantom Imaging Studies and Clinical Implications. Radiology. 2016;278(3):723–733. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Graser A, Becker CR, Staehler M, et al. Single-phase dual-energy CT allows for characterization of renal masses as benign or malignant. Invest Radiol. 2010;45(7):399–405. [DOI] [PubMed] [Google Scholar]
- 19.Fletcher JG, Yu L, Li Z, et al. Observer Performance in the Detection and Classification of Malignant Hepatic Nodules and Masses with CT Image-Space Denoising and Iterative Reconstruction. Radiology. 2015;276(2):465–478. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Schindera ST, Odedra D, Raza SA, et al. Iterative reconstruction algorithm for CT: can radiation dose be decreased while low-contrast detectability is preserved? Radiology. 2013;269(2):511–518. [DOI] [PubMed] [Google Scholar]
- 21.Schindera ST, Odedra D, Mercer D, et al. Hybrid iterative reconstruction technique for abdominal CT protocols in obese patients: assessment of image quality, radiation dose, and low-contrast detectability in a phantom. AJR Am J Roentgenol. 2014;202(2):W146–152. [DOI] [PubMed] [Google Scholar]
- 22.Euler A, Stieltjes B, Szucs-Farkas Z, et al. Impact of model-based iterative reconstruction on lowcontrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages. Eur Radiol.2017;27(12):5252–5259. [DOI] [PubMed] [Google Scholar]