Abstract
Radiochromic film dosimetry has been widely employed in most of the applications of radiation physics for over twenty years. This is due to a number of appealing features of radiochromic films, such as reliability, accuracy, ease of use and cost. However, current radiochromic film reading techniques, based on the use of commercial densitometers and scanners, provide values of dose only after the exposure of the films to radiation. In this work, an innovative methodology for the real-time reading of radiochromic films is proposed for some specific applications. The new methodology is based on opto-electronic instrumentation that makes use of an optical fiber probe for the determination of optical changes of the films induced by radiation and allows measurements of dose with high degree of precision and accuracy. Furthermore, it has been demonstrated that the dynamic range of some kinds of films, such as the EBT3 Gafchromic films (intensively used in medical physics), can be extended by more than one order of magnitude. Owing to the numerous advantages with respect to the commonly used reading techniques, a National Patent was filed in January 2018.
Introduction
Radiochromic films (RCFs) are dosimeters based on the property of modifying the structural characteristics of their crystalline sensitive element when exposed to ionizing radiations. The interaction of ionizing radiation with the film produces a polymerization process in the monomers of the sensitive element1. This microscopic phenomenon is reflected in a color change at macroscopic level and the color of the film can be related to the radiation dose. The first work on the employment of solid solutions showing radio-synthesis after the exposure to radiations dates back to 1965 by Mc Laughing et al.2. The subsequent works on RCFs can be attributed to the National Institute of Standard and Technology (NIST)1,3,4. The first prototypes of RCFs were sensitive in the range of dose from a few to hundreds of kGy and therefore these films were employed for high-dose applications only. A kind of radiochromic medium sensitive to lower doses (up to 5 Gy), known as Gafchromic film, was developed in 1986 by the International Speciality Products Inc. (ISP)1.
Nowadays RCF dosimetry is considered as a reliable technique for accurate dose assessment and quality checks in many applications of radiation physics. Several types of films are currently available on the market, covering a very wide range of dose from a few mGy to hundreds of kGy5. The main application is in medical physics, although RCFs are commonly used also in other fields, e.g. studies of radiation damage of electronic devices, beam diagnostics, radiation processing and radiation-induced sterilization (food and medical instrumentation). Such a wide use is consequence of the intrinsic characteristics of the films. Today RCFs are reliable, easy to use, cheap, portable and non-invasive instruments, providing accurate and permanent values of dose6. It has to be noticed that this last feature is very important for legal aspects related to dosimetry. RCFs have very high spatial resolution and moreover most of RCFs types are made of tissue-equivalent materials. These characteristics fit perfectly for the medical physics, in particular for the determination of percentage depth-dose curves in radiotherapy with photon and ion beams7–11. RCFs are sensitive to all kinds of ionizing radiation employed in the applications (photons, electrons, protons and heavy ions). The response of the films is almost completely independent of dose-rate, while the energy dependence is more complicated, although it was studied in depth in several works reported in literature. More details on the energy dependence of RCFs are given in the next section.
Despite such appealing characteristics, RCFs cannot be employed as dosimeters or detectors if real-time information are needed. The current RCFs reading techniques, mainly based on commercial densitometers, scanners and spectrophotometers12–14, provide dose measurements only after the exposure of the films to radiation. In the practice, the reading is carried out after the withdrawal of the film from the irradiation site. Therefore, the current reading techniques don’t allow measurements of the trend of dose in time, but only integrated dose values. In the applications this important limitation is overcome (when needed) by using other detectors providing the response in real-time. As an example, gas-filled chambers, scintillation detectors, phosphor screens and semiconductor detectors are usually employed for this purpose; however, RCFs may be used in combination with these detectors. The most widespread reading technique based on flat-bed scanners is affected by reproducibility issues, including the warm-up, uniformity and resolution of the scanner, as well as the dependence on the film size, orientation and region of interest for the analysis, lateral non-uniformity and polarization effect. All these aspects have been widely discussed in literature, leading to uncertainties that normally are around 3–6%15–17.
In this work, an innovative method, based on opto-electronic instrumentation, is proposed for the real-time reading of RCFs. Previous works on this topic are reported by Mignani et al. and Rink et al.18,19. together with a U.S. patent20. The method proposed here has several new interesting features with respect to the existing RCF reading techniques. Because it makes direct use of commercial RCFs (cut in small pieces), it avoids carrying out complicated operations of deposition of radiochormic media on the tip of the fiber. Moreover, with respect to the above-quoted RCF real-time techniques, based on single-use probes, this method uses an optical fiber probe, in which the RCF is disposable and can be quickly replaced with other RCFs in order to make other measurements. This allows the preservation of all the above-discussed characteristics of the films. The advantages of using an optical-fiber-based instrumentation instead of commercial RCF reading tools, not only allow the performance of the reading of the dose in real-time, but also the exploitation of most of the potential of the RCF dosimeters. In fact, as will be demonstrated in this work, measurements of dose can be performed with a very high degree of precision and accuracy. Furthermore, the range of sensitivity of some types of films can be extended by several orders of magnitude, with respect to that declared by the manufacturer. The originality, the technological advances with respect to the existing RCF reading methods and the potential distribution on the market made the invention patentable. For these reason a National Patent was filed in January 2018.
In the next section an overview of the main characteristics of RCFs will be discussed: the operating principle and the response of the films to radiation type, energy and dose-rate. Moreover, different types of films will be shown in relation to their specific applications. Following this, the principle and the experimental set-up of the innovative RCF reading method will be shown and discussed. Finally, in the last section, the operation of the invented method will be demonstrated for two types of films widely employed in the applications: EBT3 and XR-QA2 Gafchromic films exposed to 60Co gamma rays at two dose-rates.
Radiochromic films
This section gives an overview of the main characteristics and application fields of the RCFs. The operating principle and the dependences on relevant physical quantities such as radiation type, energy and dose-rate are briefly described, while more details can be found in the referenced publications. The characteristics discussed here have made RCF dosimetry such a widely spread dosimetry technique in several application fields. Therefore, the proposed RCF real-time reading method adds a significant value to the features of this dosimeter.
Operating principle and radiation type, energy and dose-rate dependence
RCFs consist of a single or double layer of radiation-sensitive material on a thin polyester base with a coating. The last could be transparent or opaque5. The ionizing radiation produces a polymerization process in the monomers of the active layer, resulting in a modification of the optical properties of the film. The materials responsible for RCFs coloration are the dyacetilenes, particular type of crystalline polyacetylenes that show excellent sensitivity to ionizing radiation. Differently from silver halide films, RCFs are self-developing, namely they don’t need chemical or physical processing after irradiation9,15.
RCFs are sensitive to almost any type of ionizing radiation (photons, electrons and charged particles over a wide range of energy and dose), being at the same time almost totally insensitive to room light. The response of most of RCFs shows very low dependence on the radiation quality, at least for low-LET (Linear Energy Transfer) radiation (e.g. MeV-energy electrons and photons) usually employed in radiotherapy and radiation hardness applications21–23. RCFs energy dependence to high-LET radiation is more complicated. Nevertheless the response of the films to different LET-components of charged particle beams can be taken into account by applying relative effectiveness factors. For example, some types of RCFs, such as EBT2 and EBT3 Gafchromic films, intensively used in radiotherapy with photon, electron and also ion beams, are known for underestimate the dose in the Bragg Peak region. This is known in literature as “quenching effect”. However, RCFs are successfully and commonly employed for the determination of percentage depth dose profiles of photon and ion beam in radiotherapy24–28. Concerning RCFs dose-rate dependence, no difference in response has been observed. The sensitivity to neutrons is very low: to date results about RCFs neutron sensitivity are reported by Bazioglou et al.29. In the practice, RCFs are not used for measuring neutron dose and neutron beam profiles as it is usually done for the other radiations.
Types of films and related applications
The types of RCFs available on the market cover a wide range of dose, from a few mGy to several hundreds of kGy. As a consequence of this and of the above-mentioned characteristics in comparison to other dosimeters and detectors, RCFs are currently employed in many applications of radiation physics. Figure 1 shows the application fields in which RCFs are use as a function of the dose in Gy. All dose values reported in this work are referred to water, that is the standard for the medical physics. RCFs are used in the dose range from a few tens to hundreds of mGy for quality checks of typical diagnostic instrumentation such as X-ray tube for radiography and Computed Tomography (CT). For these purposes, the ISP provides the Gafchromic series XR. In particular, XR-QA2 films are suited for radiology quality assurance tests, XR-CT2 measure beam slice width in CT scanner, XR-M2 films are dedicated to mammography tests and XR-RV3 are used for peak skin dose measurement30. These films are made of an active layer of about 25 μm between a layer of yellow polyester of 97 μm and a layer of white polyester (opaque layer). It has to be noticed that the structure of the films is an important parameter for the real-time reading method, as will be explained in the following section. RCFs are used in the range from hundreds of mGy to tens of Gy for measuring doses typical of external beam therapy with photon and ion beams. The film dosimetry has been adopted as successful method by several hospitals, clinics and centres for cancer therapy31. The brachytherapy application deals with doses up to hundreds of Gy and RCFs are employed for dose verification also in this field32–34. EBT3 Gafchromic films are intensively used in the medical physics for the determination of depth-dose curves of photon, electron and ion beams. These films are made of an active layer of about 28 μm sandwiched between two polyester layers of 120 μm. These films are available as sheets of 8 × 10 inches in order to allow the use when large radiation fields are needed. However, if the determination of dose in small areas is needed, it is possible to cut the films in smaller pieces without losing their characteristics. For very high-dose applications, such as industrial application, radiation processing, radiation hardness quality assurance tests, beam diagnostics and high energy physics (HEP) application, other kinds of films are available. Gafchromic HDV2 films are sensitive in the dose range from 10 to 1000 Gy. Beyond the kGys, thin foils films (active layer 12 μm and polyester substrate 97 μm) are used. GEX Corporation trades the B3 films (active layer: 18 μm), sensitive in the range 1–150 kGy; the Far West Technology makes available FWT60 films (active layer: 18 μm) for application in the range 0.5–200 kGy35. It is notable that the employment of RCFs as dosimeter in HEP field is a topic under study. Owing to their characteristics, RCFs can be a perfect tool for dosimetry monitoring of such a harsh environment like that of HEP facilities: no electronics typical of other detectors and therefore no interaction with the electric and magnetic fields. However, a mixed radiation field of such facilities is very complicated to study with RCFs, being composed of different radiation types and energies.
Radiochromic films reading methods
To measure the dose, a physical quantity that represent the darkening of the film has to be related to the radiation dose. The relationship between this physical quantity and the dose of a set of films exposed to different doses of a known radiation field is the calibration. In the following, the principles of the commercial RCF reading techniques (scanners and densitometers) are explained, by showing an example of calibration of a set of films with an EPSON V800 Scanner. Therefore, the innovative RCF reading instrumentation is presented.
Commercial techniques
The physical quantity that represent the film darkness can be the Optical Density (OD) or any OD-based-functions such as netOD1,8. Densitometers usually provide the OD calibrated with respect to certified step-tablets. The relationship between the physical quantity representing the darkening of the films and the dose is generally non-linear; however the calibration data fit quite well to polynomial functions. The calibration performed with densitometers is therefore the OD as a function of the dose. The reading of RCFs with commercial scanners is usually performed by digitizing the image of the film. The Pixel Value (PV) can be evaluated with suited software of analysis of images and the calibration can be carried out as PV as a function of the dose or, often, as OD as a function of the dose. The OD is calculated from the PV12,36. As an example, Fig. 2 shows the calibration of a set of EBT3 Gafchromic films exposed to 1 MeV electron beam from ILU-6 LINAC accelerator of Institute of Nuclear Chemistry and Technology at Warsaw (Poland). The optical density increases as the dose increases. In particular, the growth is sharp up to about 10 Gy, where the sensitivity is very high. For higher doses, the optical density slowly increases until a saturation condition is reached.
The innovative set-up
Figure 3 shows the schematics of the experimental apparatus employed for the real-time reading of RCFs. It basically consists of an optical fiber bundle with six illuminating fibers positioned on the sides of the bundle and one light-collecting fiber in the centre, a RCF, a light source and a spectrum analyser connected to a computer for data analysis. The six illuminating fibers transport the light from the source to the tip of the probe where the RCF is positioned. After passing into the RCF, the light is first backscattered by an appropriate material, then collected by the central fiber and therefore sent back to the spectrometer. Changes of optical properties of the film due to the darkening induced by ionizing radiation can be in this way detected in real-time. The choice of the backscattering material affects the integration-time of the spectrometer. This parameter is fundamental, being related to the sensitivity of the method: it determinates the minimum amount of time in which the spectrometer can detect sensitive changes of the light. Figure 4 shows the schematics of the elements of the terminating part of the innovative set-up. This part was designed and carried out with a twofold aim: to ensure the interaction of ionizing radiation with the RCF and to allow the backscattering of the light with enough intensity to be detected by the spectrometer. With reference to the nomenclature of Fig. 4, the element “2” is both a support for the optical fibers bundle and a holder for backscattering materials. These materials can be plastics such as ABS or Teflon, in the form of cylinders (height 1 mm and diameter 6 mm) or thin foils of Mylar (of thickness 1.5 μm). It has to be noticed that the designed material-holder allows the use of the RCF itself as backscattering material, if needed. This is very useful if the RCF has an opaque layer; this is the case, e.g. of Gafchromic XR- models. The element “4” allows the hosting of the plastic cylinder. This can be screwed and fixed on the element “2”. If the thin foils are used, the element “3” can be inserted in the material-holder “4”. The elements 2-3-4 are holed. For the case of thin transparent RCFs such as EBT3 Gafchromic film model, the only absorber is the plastic cylinder or the 1.5 μm Mylar foil that operate as light mirror. For opaque films such as as XRQA2 Gafchromic film model, the elements 3 and 4 are omitted so that there is no absorber between the RCF and the incident radiation. The bundle for the optical fibers can be made of different materials such as stainless steel, plastics (less interacting with ionizing radiation) or, if necessary, the single fibers can be used. In this work the stainless steel probe was used. During the operation, the whole apparatus is positioned in the irradiation area. The light source used for this prototype is the “AvaLight-DH-S-BAL Balanced Power” and the spectrometer “AvaSpec-ULS2048XL”. The latter is controlled via computer. The RCF is positioned in the material holder as shown in Fig. 5(a). Figure 5(b) shows the top view of the optical fiber bundle with the Mylar foil and the red spot light. The effective area of the RCF samples involved in the dose reading depends on the size and position of the illuminating and collecting fibers within the terminating part of the probe; in particular, referring to pictures shown in Fig. 5, the effective area corresponds to that of a circle with a diameter of three fibers having a core of 200 μm. Nevertheless, in order to allow a practical and reliable insertion into the readout system, RCF samples of 8 × 20 mm2 were employed.
Results
The innovative dosimetry method was successfully tested with several types of RCFs. In order to demonstrate the operating principle of the method, the results, in terms of light spectra and dose calibration curves, of irradiations of EBT3 and XR-QA2 Gafchromic films with 60Co gamma rays will be discussed in the following. Several physical physical quantities can be exploited for the calibration of the films, depending on the dose range to be investigated. This property gives a high level of versatility to this method, allowing the extension of the dynamic range of some types of films by more than one order of magnitude.
Experimental
EBT3 Gafchromic film samples were exposed to 60Co-gamma rays from ISOF-CNR gamma facility at Bologna (Italy). ISOF-CNR is equipped with two Gammacell220, providing High Dose-Rate (HDR) of 2.59 Gy/min and Low Dose-Rate (LDR) of 0.10 Gy/min from the decay of the radioactive isotope 60Co (half-life: 5.272 y). This undergoes β− decay to the first and second excited states of 60Ni, which in turn decays via gamma emission with energies of 1.17 and 1.32 MeV. The two irradiators consist in self-shielded vertical cylinders with the centre surrounded by 60Co vertical rods. The dose-rates of the irradiators are determined by the activity of the 60Co rods. An opening in the upper part of the irradiators allows the positioning of the samples to be irradiated. During the irradiation, the fiber bundle and the film, i.e. the terminating part of the RCF reading set-up shown in Fig. 5 were positioned in the centre of the irradiator, while the light source and the spectrometer were remotely controlled. The radiation dose in the position of the RCF was previously measured with conventional RCF dosimetry. Since the radiation field of 60Co Gammacell is uniform and omnidirectional, the orientation of the detector for this specific irradiation is not important. In the more general case of a directional field, the new RCF reading method operates with the fiber bundle coaxial to the radiation beam (or the RCF perpendicular to the radiation beam). The film was continuously exposed to radiation and the readout performed at times fixed by the integration time. A picture of the set-up during operation is shown in Fig. 6.
Figure 7 shows the spectra corresponding to the exposure of an EBT3 Gafchormic film in the HDR-irradiator for ten minutes irradiation. In this first experiment the holder with the Teflon cylinder was used, implying an integration-time of the spectrometer of 1.5 s. The spectra are zoomed in the wavelength region 500–700 nm. The spectrometer acquires one spectrum for each integration-time (1.5 s). This amount of time corresponds to a dose on the film of 64 mGy. The effect of the ionizing radiation on the film consists in an integral reduction of the counts. As a consequence of this, the different curves represent the trend in time of the darkening of the film, that is the trend of the dose absorbed by the film. The first curve of Fig. 7 (blank) corresponds to 0 Gy on the film (no exposure to ionizing radiation). The second curve, with a reduced number of counts, corresponds to a dose of 0.5 Gy. The third curve (1.0 Gy) has a subsequent reduction of the counts and so on. The maximum dose achieved in this experiment was 30 Gy. The counts recorded by the spectrometer are affected by experimental uncertainties mainly related to the temperature of the light source. The contribution of the uncertainty of the counts due to the reproducibility was estimated to be less than 0.1% for a 2 hour acquisition made after 40 minutes from the switching on of the light source (the time needed for the warm-up). However, in order to take into account any source of uncertainty (e.g. possible instabilities of the light source) the overall uncertainty on the counts was considered the square root of the counts, in agreement with the Poisson statistics. This conservative approach overestimates the uncertainties of the counts of all calibrations reported in this work, without affecting the precision of the method. In fact, the uncertainty of the counts was evaluated to be in any case less than 1%.
Dose calibration of EBT3 Gafchromic films
The observation of the spectra of Fig. 7 suggests that the calibration can be performed by means of any reasonable physical quantity. For example, the wavelength λlow = 635 nm, corresponding to the valley of the spectra of Fig. 7, can be used for an optimal discrimination of low doses.
Figure 8 shows the counts of the curves of Fig. 7 corresponding to the wavelength λlow as a function of the dose. The uncertainty of dose values (evaluated to be less then 0.2% for all the calibration plots) was obtained by the propagation of the statistical uncertainties of the integration-time and dose-rate (both less than 0.1%). The physical quantity λlow is very sensitive up to about 2 Gy. In fact, with reference to Fig. 7, the gap of the counts between the first and second curve (blank and 0.5 Gy) is much bigger than that between the second and the third curve. This physical quantity saturates above 4 Gy; beyond this value the determination of dose changes is not possible.
In order to get a good discrimination of higher doses, the wavelength λhigh = 663 nm can be used for the calibration. Figure 9 shows the calibration of the film obtained by recording the counts of the curves of Fig. 7 corresponding to the wavelength λhigh. This physical quantity allows very precise discrimination of dose in a wide dose range, from 1 Gy to more than 30 Gy. The red curve is the exponential function (Eq. 1) that best fits to the experimental data.
1 |
Eq. 1 relates the counts (C) with the dose (D) by means of the fitting parameters P0, P1 and P2. The goodness of the fit was quantified by means of the chi-squared (χ2) test. The values of the reduced chi-square (χ2/ndf, where ndf is the number of degrees of freedom) and χ2-probability, respectively 0.2 and 1, of the calibration of Fig. 9 express the excellent agreement of the fitting function function with the experimental data. The inset of Fig. 9 reports the values of χ2, χ2-probability and the best estimates of the fitting parameters and of the statistical uncertainties. It is noticeable that the value of the reduced chi-square less than 1 implies an overestimation of the variances, accordingly to the conservative hypothesis of the Poisson statistics.
The wavelength λlow is not useful for the determination of doses greater than 2 Gy, although it is a very powerful physical quantity for the discrimination of low doses. The minimum amount of dose detected by the spectrometer is determined by two factors: the dose-rate and the spectrometer integration-time (1.5 s). This value is fixed by the material employed for backscattering the light, i.e. the Teflon cylinder. With the aim of investigating a method for a better discrimination of low doses, an irradiation of an EBT3 Gafchromic film with the LDR-irradiator (dose-rate: 0.10 Gy/min) was carried out. Furthermore, in order to reduce the integration-time, the 1.5 μm Mylar foil was used as backscattering material. This set-up allowed the setting of the integration-time of the spectrometer to 246 ms. In this way, the spectrometer detects changes of the darkening of the film in a smaller time with respect to that with the Teflon cylinder.
Figure 10 shows the calibration of the EBT3 Gafchromic film obtained by recording the counts of the curves of the spectra corresponding to the wavelength λlow as a function of the dose. The amount of dose absorbed by the film in one integration-time is 410 μGy. The dose-trend of the experimental data of Fig. 10 is the same of Figs 8 and 9, i.e. the data are very well represented (see values of χ2/ndf and χ2-probability in the inset of 10) by an exponential function (red curve in Fig. 10).
Dose calibration of XR-QA2 Gafchromic films
As already mentioned in the previous sections, some types of RCFs have a particular structure which is suitable for this innovative method. For example, XR-QA2 Gafchromic films have an opaque layer, allowing the use of the film itself as backscattering material. For this reason, an irradiation of an XR-QA2 Gafchromic film in the LDR-irradiator was carried out.
Figure 11 shows the spectra corresponding to the exposure of an XR-QA2 Gafchormic film in the LDR-irradiator for ten minutes irradiation. The spectra are zoomed in the wavelength region 530–640 nm.
The integration-time was set to 630 ms. The dose absorbed by the film in one integration-time is therefore 1 mGy. Figure 12 shows the calibration of the XR-QA2 Gafchromic film obtained by recording the counts of the curves of Fig. 11 corresponding to the wavelength λlow. It is noticeable that the Gafchromic film XR-QA2 shows characteristic spectra and calibration curves similar to that of the EBT3 model. In particular, the fitting curve is the sum of two exponential functions of the type of Eq. 1. The values of the fitting parameters are shown in the inset of Fig. 12. The values of χ2/ndf and χ2-probability are 0.8 and 1 respectively.
Validation of the new RCF real-time reader
In order to validate the obtained calibration curves, additional films (from the same batch) were exposed to known doses. The doses obtained with the calibration curves of Figs 9–12 were compared to the nominal doses, allowing the evaluation of the corresponding uncertainties. Figure 13 shows the output plot of the new reader (Dose on vertical axis; Counts on the horizontal axis) for an exposure of an EBT3 Gafchromic film to the HDR irradiatior. The doses (blue curve) were computed by reversing the calibration curve of Eq. 1 and compared to the nominal doses (black dots).
The maximum uncertainty is within 2% from 2.5 Gy on. Similar procedures were carried out for the data of Figs 10 and 12 leading to maxima uncertainties within 4% from 100 mGy and within 8% from 30 mGy respectively. These uncertainty values can be reduced by optimizing the integration-time of the spectrometer to the wavelengths of interest for the calibration. Moreover, the integration-time can be further reduced by using thin high-reflecting materials. These procedures provide also the value of the minimum detectable dose obtained with this method. This point will be object of next studies.
Discussion
In this work an innovative method, based on opto-electronic instrumentation, is proposed for the real-time reading of RCFs. The method is based on the use of optical fibers that transport the light from a source, such as a halogen or deuterium light source, up to the RCF. During the exposure to ionizing radiation, the optical changes of the film are detected in real-time by a light-collecting fiber and sent back to a spectrometer for data analysis. A proper scattering material can be used, if needed, to increase the amount of light in the collecting fiber. This allows the reduction of the integration-time of the spectrometer, affecting ultimately the minimum detectable dose.
The method was tested for several RCF types, showing excellent performances in terms of accuracy and precision. The results of the irradiations of two types of films (EBT3 and XR-QA2 Gafchromic model) to 60Co-gamma rays at two dose-rates (2.59 and 0.10 Gy/min) are reported in this work. The versatility of the method relies on many factors, among which the capability of performing the dose calibration by means of more criteria of data analysis. For example, the counts corresponding to the characteristic valley of the spectra, i.e. the wavelength λlow = 635 nm, can be put in relation to the dose on the RCF. The subsequent calibration, based on a proper choice of the backscattering material, shows that the dynamic range of some types of films is higher with respect to what declared by the manufacturer. In fact, EBT3 Gafchromic films, intensively employed in medical physics for dosimetry quality checks in radiation therapy, are sensitive up to dozens of Gy with the analysis with commercial reading instruments. The proposed dosimetry method clearly allows the detection of doses of hundreds of mGy with EBT3 Gafchromic films with precision of 4%. These uncertainties can be further reduced for lower doses by optimizing the integration-time and by using high-reflecting back-scattering materials. Different data analysis techniques can be exploited for the investigation of a specific dose range. For example, by recording the counts of the spectra corresponding to the wavelength λhigh = 663 nm, it is possible to perform a calibration for better discriminate the high doses. It should be noted that the spectra shown in Figs. 7 and 11 can be converted in principle into optical density (OD) spectra, provided that they are normalized to the first acquired spectrum corresponding to a zero dose. By way of example, Fig. 15 shows the spectra of Fig. 7 up to 3 Gy converted into OD. The OD was calculated according to the following expression:
where I0(λ) is the light intensity measured before the exposure and I(λ) is that measured during the exposure to radiation. The OD spectra exhibit two main peaks respectively at 635 nm and 585 nm, revealing high similarity to those shown in the literature37 as proof of the consistence of the proposed method with standard light-transmission-based approaches. The evaluation of measurement uncertainties has been made by means of a precautionary approach, based on the Poisson statistics, allowing uncertainties within 1% for all the calibration plots (Figs 9–11).
The advantages of the proposed method with respect to the state of the art are manifold. Firstly, the real-time dosimetry with RCFs can be performed by monitoring the dose from a remote position with an optical fiber probe, exploiting the potential of optical fibers such as micrometric thickness, low attenuation coefficient and no interaction with electromagnetic fields. The irradiation of small areas (less than 1 mm2) of commercial RCF allows multiple exposures and readings of the same RCF up to the saturation. As a result a National Patent was filed in January 2018. RCF dosimetry finds applications in any field of the radiation physics, each one with a specific target. Current author’s research efforts are devoted to the study of the miniaturization of the present probe, in order to make it competitive with other dosimeters and detectors in some specific application field, such as the dosimetry of the irradiations of the electronic devices for space and terrestrial radiation hardness applications. Finally, the miniaturization of the present prototype up to sub-millimetre dimensions can pave the way to interesting applications of the proposed method such as in-vivo dosimetry.
Acknowledgements
The authors would like to acknowledge Prof. Raffaele Velotta and Prof. Giovanni Chiefari for the precious scientific advices provided, Mr. Mario Borriello and the staff of INFN Mechanical Workshop of Napoli for the help provided in the manufacturing of mechanical components, Mila D’Angelantonio for the help provided during the irradiations at ISOF-CNR.
Author Contributions
P.C. and L.C. equally contributed to this work; M.C., A. Cus. and A. Cut. designed the opto-electronic set-up and supervised the optical data analysis; S.B., L.C. and G.B. conceived, designed and supervised the radiation tests; L.C. conceived the idea and performed the radiation tests; F.D.C., F.F., P.C. and P.V. performed radiation tests and data analysis; P.C. designed the mechanical probe and wrote the manuscript. All authors reviewed the manuscript.
Data Availability
The data analysed during the current study are available from the corresponding author on reasonable request.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Pierluigi Casolaro and Luigi Campajola contributed equally.
Contributor Information
Pierluigi Casolaro, Email: casolaro@na.infn.it.
Marco Consales, Email: consales@unisannio.it.
References
- 1.Williams, M. & Metcalfe, P. Radiochromic film dosimetry and its applications in radiotherapy. In AIP Conference Proceedings, vol. 1345, 75–99 (AIP, 2011).
- 2.McLaughlin WL, Chalkley L. Measurement of radiation dose distributions with photochromic materials. Radiol. 1965;84:124–125. doi: 10.1148/84.1.124. [DOI] [PubMed] [Google Scholar]
- 3.Humphreys JC. Nist high-dose calibration services. Nucl. Instruments Methods Phys. Res. Sect. B: Beam Interactions with Mater. Atoms. 1989;40:1173–1177. doi: 10.1016/0168-583X(89)90565-X. [DOI] [Google Scholar]
- 4.McLaughlin WL, et al. Novel radiochromic films for clinical dosimetry. Radiat. protection dosimetry. 1996;66:263–268. doi: 10.1093/oxfordjournals.rpd.a031731. [DOI] [Google Scholar]
- 5.Soares CG. Radiochromic film dosimetry. Radiat. measurements. 2006;41:S100–S116. doi: 10.1016/j.radmeas.2007.01.007. [DOI] [Google Scholar]
- 6.Al-Haj A, Lobriguito A, Chantziantoniou K, Lagarde C, Iqeilan N. Evaluation of the calibration parameters of radiochromic films for patient dosimetry in interventional radiology. Radiat. protection dosimetry. 2010;142:228–237. doi: 10.1093/rpd/ncq220. [DOI] [PubMed] [Google Scholar]
- 7.Fletcher CL, Mills JA. An assessment of gafchromic film for measuring 50 kv and 100 kv percentage depth dose curves. Phys. Medicine & Biol. 2008;53:N209. doi: 10.1088/0031-9155/53/11/N02. [DOI] [PubMed] [Google Scholar]
- 8.Butson MJ, Peter K, Cheung T, Metcalfe P. Radiochromic film for medical radiation dosimetry. Mater. Sci. Eng. R: Reports. 2003;41:61–120. doi: 10.1016/S0927-796X(03)00034-2. [DOI] [Google Scholar]
- 9.Devic S. Radiochromic film dosimetry: past, present, and future. Phys. medica. 2011;27:122–134. doi: 10.1016/j.ejmp.2010.10.001. [DOI] [PubMed] [Google Scholar]
- 10.Vatnitsky SM. Radiochromic film dosimetry for clinical proton beams. Appl. radiation isotopes. 1997;48:643–651. doi: 10.1016/S0969-8043(97)00342-4. [DOI] [PubMed] [Google Scholar]
- 11.Bilge H, Cakir A, Okutan M, Acar H. Surface dose measurements with gafchromic ebt film for 6 and 18 mv photon beams. Phys. Medica. 2009;25:101–104. doi: 10.1016/j.ejmp.2008.05.001. [DOI] [PubMed] [Google Scholar]
- 12.Devic S, et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Med. physics. 2005;32:2245–2253. doi: 10.1118/1.1929253. [DOI] [PubMed] [Google Scholar]
- 13.Lewis, D. Performance of the vidarR red led dosimetry pro advantageTM: A scanner optimized for use with gafchromicR ebt dosimetry film. ISP Article (2007).
- 14.Devic S, et al. Absorption spectroscopy of ebt model film. Med. physics. 2007;34:112–118. doi: 10.1118/1.2400615. [DOI] [PubMed] [Google Scholar]
- 15.Devic S, Tomic N, Lewis D. Reference radiochromic film dosimetry: review of technical aspects. Phys. Medica. 2016;32:541–556. doi: 10.1016/j.ejmp.2016.02.008. [DOI] [PubMed] [Google Scholar]
- 16.Giaddui T, et al. Characteristics of gafchromic xrqa2 films for kv image dose measurement. Med. physics. 2012;39:842–850. doi: 10.1118/1.3675398. [DOI] [PubMed] [Google Scholar]
- 17.Marroquin EYL, Herrera Gonzalez JA, Camacho Lopez MA, Barajas JEV, García-Garduño OA. Evaluation of the uncertainty in an ebt3 film dosimetry system utilizing net optical density. J. applied clinical medical physics. 2016;17:466–481. doi: 10.1120/jacmp.v17i5.6262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Mignani, A. G., Romano, S., Fusi, F. & Mencaglia, A. A. Radiation dosimetry in radiotherapy: a model for an extrinsic optical fiber sensor. In European Workshop on Optical Fibre Sensors, vol. 3483, 99–103 (International Society for Optics and Photonics, 1998).
- 19.Rink A, Vitkin IA, Jaffray DA. Suitability of radiochromic medium for real-time optical measurements of ionizing radiation dose. Med. physics. 2005;32:1140–1155. doi: 10.1118/1.1877832. [DOI] [PubMed] [Google Scholar]
- 20.Rink, A. et al. Fiber optic radiochromic dosimeter probe and method to make the same US Patent 9,000,401 (2015).
- 21.Glaser, M., Ravotti, F. & Moll, M. Dosimetry assessments in the irradiation facilities at the cern-ps accelerator. In Radiation and Its Effects on Components and Systems, 2005. RADECS 2005. 8th European Conference on, PI5–1 (IEEE, 2005).
- 22.Wie BS, et al. Evaluation and application of us medical proton facilities for single event effects test. IEEE Transactions on Nucl. Sci. 2015;62:2490–2497. doi: 10.1109/TNS.2015.2496329. [DOI] [Google Scholar]
- 23.Massillon-JL G, Chiu-Tsao S-T, Domingo-Muñoz I, Chan MF. Energy dependence of the new gafchromic ebt3 film: dose response curves for 50 kv, 6 and 15 mv x-ray beams. Int. J. Med. Physics, Clin. Eng. Radiat. Oncol. 2012;1:60. doi: 10.4236/ijmpcero.2012.12008. [DOI] [PubMed] [Google Scholar]
- 24.Avila-Rodriguez M, Wilson J, McQuarrie S. The use of radiochromic films to measure and analyze the beam profile of charged particle accelerators. Appl. Radiat. Isot. 2009;67:2025–2028. doi: 10.1016/j.apradiso.2008.10.009. [DOI] [PubMed] [Google Scholar]
- 25.Kirby D, et al. Let dependence of gafchromic films and an ion chamber in low-energy proton dosimetry. Phys. Medicine & Biol. 2009;55:417. doi: 10.1088/0031-9155/55/2/006. [DOI] [PubMed] [Google Scholar]
- 26.Buenfil A, et al. Response of radiochromic dye films to low energy heavy charged particles. Nucl. Instruments Methods Phys. Res. Sect. B: Beam Interactions with Mater. Atoms. 2002;197:317–322. doi: 10.1016/S0168-583X(02)01488-X. [DOI] [Google Scholar]
- 27.Sorriaux J, et al. Evaluation of gafchromicR ebt3 films characteristics in therapy photon, electron and proton beams. Phys. Medica. 2013;29:599–606. doi: 10.1016/j.ejmp.2012.10.001. [DOI] [PubMed] [Google Scholar]
- 28.Castriconi R, et al. Dose–response of ebt3 radiochromic films to proton and carbon ion clinical beams. Phys. Medicine & Biol. 2016;62:377. doi: 10.1088/1361-6560/aa5078. [DOI] [PubMed] [Google Scholar]
- 29.Bazioglou M, Kalef-Ezra J. Dosimetry with radiochromic films: a document scanner technique, neutron response, applications. Appl. Radiat. Isot. 2001;55:339–345. doi: 10.1016/S0969-8043(01)00060-4. [DOI] [PubMed] [Google Scholar]
- 30.Tomic N, et al. Characterization of calibration curves and energy dependence gafchromictm xr-qa2 model based radiochromic film dosimetry system. Med. physics. 2014;41:062105. doi: 10.1118/1.4876295. [DOI] [PubMed] [Google Scholar]
- 31.Fiandra C, et al. Clinical use of ebt model gafchromicTM film in radiotherapy. Med. physics. 2006;33:4314–4319. doi: 10.1118/1.2362876. [DOI] [PubMed] [Google Scholar]
- 32.Palmer AL, Nisbet A, Bradley D. Verification of high dose rate brachytherapy dose distributions with ebt3 gafchromic film quality control techniques. Phys. medicine & biology. 2013;58:497. doi: 10.1088/0031-9155/58/3/497. [DOI] [PubMed] [Google Scholar]
- 33.Palmer AL, Lee C, Ratcliffe AJ, Bradley D, Nisbet A. Design and implementation of a film dosimetry audit tool for comparison of planned and delivered dose distributions in high dose rate (hdr) brachytherapy. Phys. Medicine & Biol. 2013;58:6623. doi: 10.1088/0031-9155/58/19/6623. [DOI] [PubMed] [Google Scholar]
- 34.Palmer, A., Nisbet, A. & Bradley, D. Semi-3d dosimetry of high dose rate brachytherapy using a novel gafchromic ebt3 film-array water phantom. In Journal of Physics: Conference Series, vol. 444, 012101 (IOP Publishing, 2013).
- 35.Barrett J, et al. A high-dose intercomparison study involving red 4034 perspex and fwt-60-00 radiochromic dye films. Int. J. Radiat. Appl. Instrumentation. Part C. Radiat. Phys. Chem. 1990;36:505–507. doi: 10.1016/1359-0197(90)90042-G. [DOI] [Google Scholar]
- 36.Schoenfeld AA, Poppinga D, Harder D, Doerner K-J, Poppe B. The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers. Phys. Medicine & Biol. 2014;59:3575. doi: 10.1088/0031-9155/59/13/3575. [DOI] [PubMed] [Google Scholar]
- 37.León-Marroquín EY, et al. Spectral analysis of the ebt3 radiochromic film irradiated with 6 mv x-ray radiation. Radiat. Meas. 2016;89:82–88. doi: 10.1016/j.radmeas.2016.03.007. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data analysed during the current study are available from the corresponding author on reasonable request.