Abstract.
This erratum corrects an error in “Measuring temporal stability of positron emission tomography standardized uptake value bias using long-lived sources in a multicenter network,” by D. Byrd et al.
This article [J. Med. Imag. 5(1), 011016 (2018), doi: 10.1117/1.JMI.5.1.011016] was originally published online on 4 January 2018 with the following error. The Pearson correlation coefficients were not computed and reported in a consistent fashion, most notably in Table 10, where values labeled “” were instead the values of without appropriate squaring. Figure 8 contained a similar inconsistency, though not all values were affected.
Original (erroneous) and corrected correlation coefficients from Table 10 are enlisted here with their corresponding measurements:
| Original (erroneous) | Corrected | ||
|---|---|---|---|
| Measurements | Correlation coefficient | Measurements | Correlation coefficient |
| 12 | 0.966 | 12 | 0.934 |
| 12 | 0.956 | 12 | 0.915 |
| 10 | 0.934 | 10 | 0.872 |
| 7 | 0.867 | 7 | 0.752 |
| 10 | 0.583 | 10 | 0.340 |
| 20 | 0.993 | 20 | 0.985 |
| 15 | 0.851 | 15 | 0.725 |
| 8 | 0.866 | 8 | 0.749 |
| 15 | 0.378 | 15 | 0.143 |
| 6 | 0.976 | 6 | 0.953 |
| 8 | 0.594 | 8 | 0.353 |
| 3 | 0.387 | 3 | 0.150 |
| 8 | 0.944 | 8 | 0.892 |
| 11 | 0.002 | ||
The original and the corrected Figure 8 are depicted below:
Original
Corrected
This article was corrected online on 23 March 2019. It appears correctly in print.