Abstract
Here we report a method for improving the magnetic field sensitivity of an ensemble of Nitrogen-Vacancy (NV) centres in 12C-enriched diamond aligned along the [111] crystal axis. The preferentially-aligned NV centres are fabricated by a Plasma Enhanced Chemical Vapour Deposition (PECVD) process and their concentration is quantitatively determined by analysing the confocal microscopy images. We further observe that annealing the samples at high temperature (1500 °C) in vacuum leads to a conversion of substitutional nitrogen into NV centres. This treatment also increases the coherence time of the NV centres electron spins up to 40 μs, which corresponds to enhancement of the sensitivity by a factor of three. However, this procedure also leads to a loss of the preferential alignment by 34%.
Introduction
The Nitrogen-Vacancy (NV) colour centre in diamond is one of the most promising physical systems in the emerging field of quantum technology due to its unique properties1. It has been demonstrated that NVs can serve as universal sensors for electric2 and magnetic fields3,4, temperature5 and pressure6 with high sensitivity and nano-scale spatial resolution. This is achieved since the electron spin associated with a single NV can be manipulated and readout at ambient conditions. For magnetic field sensing with an ensemble of N NV centres, the sensitivity is given by7,
1 |
where h is the Planck constant, g the Landé factor, μB is the Bohr magneton and C the measurement contrast. The NV’s coherence time is represented by τ, set to T2 measured by a Hahn echo8 or with Δν being the NVs line width for ac and dc magnetometry, respectively. Most of the studies use single NVs, but in many cases an ensemble of NVs is required due to better sensitivity (as shown above) and due to the possibility to perform field imaging9–11 (imaging can be realized via scanning probe techniques3, but it is usually slower). For most of these studies, diamond crystals with a [100] crystal orientation of the surface have been used, where the NVs have four possible orientations. Often, a [111] orientation is preferred since in this case the optical dipole associated with the NV centres is parallel to the surface and thus it is the optimal orientation for the emission of fluorescence. Often, it is difficult to work with several orientations and preferentially-aligned NVs are required, as the number of collected photons increases and therefore the signal-to-noise ratio in sensing experiments.
In previous studies, it has been shown that single NVs can be aligned along two different orientations12, followed by the possibility to align up to 94% of single NV centres along the [111] crystal orientation13–15, when choosing a [111] oriented diamond substrate. In recent reports, an alignment ratio of an NV ensemble above 99% was demonstrated16 and also shallow ensembles approximately 10 nm away from the surface were created17. More recently the thermal stability of aligned NVs in thick layers (≈ 25 μm) was studied18, where it was observed that the alignment can be broken and even the NV centres can be destroyed. All NV ensembles in these studies often show a relatively short Hahn echo coherence time going up to 6 μs.
In the current work we achieve an order of magnitude longer T2 for ensembles of NV centres with comparable concentrations where we show that high temperature annealing of CVD grown diamond in vacuum leads to enhanced magnetic field sensitivity without loosing NV centres.
Results
Sample Analysis
In this work we report on engineering nitrogen doped isotopically enriched diamond layers for magnetometry applications with NV centres preferentially-aligned along the [111] crystal axis. We use purified 12CH4 gas (99.999%) for all processes, which already contains nitrogen with a concentration of approximately 70 ppm according to the supplier (Cambridge Isotope Laboratories). In the first experiment we fabricated a 40 nm thick diamond layer with a high concentration of NV centres using Plasma Enhanced Chemical Vapour Deposition (PECVD) on a IIa [111] diamond substrate (Element Six Group). Optically Detected Magnetic Resonance (ODMR) measurements revealed that all NVs were oriented along the [111] crystal axis, see Fig. 2(d). With this sample (referred to as Sample I) and applying the method described in19 we were able to detect hydrogen Nuclear Magnetic Resonance (NMR) signal coming from the immersion oil used in our confocal microscope and also to determine the average depth of the NVs to be 8 nm (see Supplementary Information). The coherence time measured via Hahn echo decay was found to be T2 = 4 μs, but we expected a much longer value for the given 12C concentration as reported previously20. This value is even shorter for a diamond with natural abundance of 13C nuclear spins, where it is expected that T2 ≈ 500 μs21. Thus we can conclude that T2 is limited by the presence of paramagnetic defects, probably nitrogen related ones such as substitutional nitrogen atoms (P1 centres), NVH and NVN (H3), but also other vacancy related defects (R5, R6 or R10)22.
In order to study the fabrication process in more detail and to improve the spin properties, we fabricated two additional samples at different conditions (diamond substrates type IIa from Applied Diamond Inc., with an intrinsic nitrogen concentration below 1 ppm). The layers were grown in our home-built PECVD reactor with different microwave powers which results in different plasma densities, growth temperatures and therefore different growth rates. The latter are important for tuning the nitrogen incorporation and thus the NV formation in the diamond. The parameters for the individual growth processes are summarized in Table 1. The grown layers are investigated with a home-built confocal microscope set-up where we observe a high NV concentration (>4.8 · 1014/cm3) in all samples. We have developed a method for analysing the confocal images to determine the concentration of NV centres on an area of 20 μm × 20 μm. First we take a confocal image as shown in Fig. 1(a), where NVs appear as bright spots. In a next step, a calibration of the fluorescence signal has to be performed, therefore we measure a single NV centre, confirmed by a second order auto correlation function g(2) experiment23, and average the photon time trace, to receive the count rate of a single NV. The count rate of each pixel of the confocal image is then normalized with this value and since the spatial resolution of the fluorescence signal of our sample is diffraction limited, we obtain a count rate per confocal volume (defined simply as spot). The point spread function is distributed over several pixels. In a last step the number of NV centres per spot is calculated for the whole confocal image and is plotted as a histogram in Fig. 1(b). This distribution is fitted with a Gaussian function to obtain the averaged number of NV centres per confocal volume. The analysis method is repeated for several confocal images for better statistics. It is interesting to note that the standard deviation of the fit is nearly the square root of the mean value, indicating that the emitters (NV centres) in the sample follow a Poisson distribution.
Table 1.
Sample | MW Power, Watt | Temp., °C | ν, nm/hour | δ, nm | NV conc. as-grown, ppb | NV conc. anneal., ppb | T2 as-grown, μs | T2 anneal., μs |
---|---|---|---|---|---|---|---|---|
Sample I | 1200 | 760* | 20** | 40** | 640.0 | —*** | 4 | —*** |
Sample A | 1200 | 930 | 210 | 420 | 10.3 | 12.5 | 10.1 ± 2.6 | 32.3 ± 4.2 |
Sample B | 720 | 875 | 120 | 240 | 3.8 | 17.3 | 6 ± 3.3 | 26 ± 4.4 |
Conditions of the growth processes are: growth time t = 2 h, microwave (MW) frequency f = 2.46 GHz, pressure p = 22.5 mbar, methane concentration c0 = 0.5% for Samples A and B and cI = 0.05% for Sample I with respect to hydrogen which was applied with a flow of 200 sccm.
*Temperature of Sample I was measured with a thermocouple whereas the others were measured via an infra-red pyrometer.
**Estimated from SIMS results of another sample with same growth conditions.
***No annealing procedure was performed on this sample.
NV centre analysis and coherence property study
The above described technique is used for both Samples (A and B) and we observe a good separation of the corresponding histograms, as shown in Fig. 2(c). The data shows that the NV density increases with microwave power, which is in contradiction with previous studies16. The temperatures during the growth processes are measured by a pyrometer focused on the diamond’s surface and listed in Table 1. Unfortunately we could not run the CVD process at a microwave power lower than 720 W since the plasma becomes unstable. It is also worth noting that our CVD system is equipped with a graphite heated stage which keeps the temperature at 750 °C even when the source of microwave plasma is completely switched off. ODMR measurements of all samples confirm the presence of NV− centres and prove preferential alignment along the [111] orientation. In other words, the NV quantization axis (the line connecting the nitrogen atom and the vacancy) is perpendicular to the diamond surface, as illustrated in Fig. 2(a). These measurements are performed in a static magnetic field with the B-field vector pointing along the [111] direction. In this configuration, NV centres oriented along the other three directions make the same angle (≈ 70.5°) to the applied magnetic field, and hence the resonance frequencies overlap. Figure 2(d), shows the simulated and measured ODMR spectrum at approximately 33 G, exemplary taken from Sample A. The simulation shows the relative contribution of different NV orientations compared with maximum ODMR contrast (≈ 30%). The simulation is performed with the assumption that all four NV orientations are equally present in the sample, which is markedly different from the experimental observation. From the measured data, we estimate that about 99% of the NV centres have the same orientation.
We further investigated the spin properties of the NV ensembles, where we find that the average coherence time T2, of more than ten measured ensembles in Sample A, is about 10 μs, see Table 1 (see also Supplementary Information for individual measurements). Again we find that T2 is limited by some other spin defects, probably in the form of P1 centres or electron traps. Motivated by previous reports24,25 on annealing of diamonds with NV centres at high temperatures, we decided to perform similar treatment here. Both samples were annealed at a temperature of 1500 °C in ultra high vacuum (below 10−7 mbar) for two hours. Under these conditions vacancies and vacancy related defects as described above, created during the crystal growth, become mobile and start to diffuse. Some of them are destroyed at the surface, but a certain amount is captured by nitrogen atoms and form additional NV centres. This effect can be seen in the analysis data shown in Fig. 3(a), where the Gaussian fit of the histograms after the annealing is shifted to higher NV numbers. The increase for Sample A (high MW power) is from 66 to 81 NVs per confocal volume whereas the increase for Sample B (low MW power) is from 14 to 64 NVs per confocal volume. Furthermore, we observe an increase of the Hahn echo coherence time T2 as shown in Fig. 3(c). For the as-grown NVs in Sample A, we measured an average Hahn echo coherence time of 10 μs, while after the annealing procedure we find T2 values up to 40 μs (detailed overview of Hahn Echo measurements for different ensembles are shown in the Supplementary Information). This effect has been reported previously as mentioned above and it is believed that at those temperatures various paramagnetic defects are destroyed, thus leading to an increase of NVs’ T2. In the ODMR spectra of the NV centres after the annealing treatment from Sample A, shown in Fig. 3(d), two additional spectral lines are found (two lines instead of one are found, as the external magnetic field is not perfectly aligned along the [111] crystal axis but with a slight off angle of approximately 3 degrees). This observation is a result of breaking the preferential alignment or in other words, there are NV centres which are not aligned along the [111] crystal axis. The proposed mechanism for diffusion of NV centres at high temperatures26 can explain this effect. Another explanation would be that during the annealing new NV centres are formed, which are randomly aligned along the four possible orientations. By analysing the intensity of the lines in Fig. 3(d), we find that only 66% of the NVs are preferentially-aligned along the [111] crystal axis and 34% have other orientations, which decreases the ODMR contrast and therefore the magnetic field sensitivity by the same value.
In Table 2 we compare the sensitivity of ensembles created by different methods, keeping in mind that equation (1) has to be interpreted carefully as a larger detector size does not automatically translate into better sensitivity. Depending on the application, the number of NV sensors has to be optimized as for example in the case of magnetometry applications, NV centres in equal distance to the signal source are required, as the signal decays strongly with distance. Hence, a better benchmark for evaluating the quality of a sample is to give a sensitivity per root volume and therefore we modify equation (1), for a pulsed measurement scheme to a volume normalised sensitivity ηV:
2 |
where CO is the measurement contrast, nB = N/V the sensor density (with N: number of NVs, V: volume), pB the average rate of detected photons and τL the duration of the readout laser pulse7. The estimated sensitivity of our samples are similar to previously reported values for preferentially aligned CVD grown NVs (calculated from17). In comparison to ion implantation technique9 the sensitivities of CVD grown NVs are better, as it is not possible to produce preferentially aligned NVs by ion bombardment. The best values reported so far are achieved with electron irradiated [111] oriented diamond fabricated via High Pressure High Temperature method (HPHT)27 where an expected sensitivity of 100 using a pulsed detection scheme with discrete readout steps is used. Independently of the overall sensitivity in our samples, the improvement factors resulting from our vacuum annealing treatment are 1.3 and 2.9, respectively for Samples A and B.
Table 2.
Volume normalised sensitivity ηV in | ||||||
---|---|---|---|---|---|---|
Sample A | Sample B | Sample I | CVD grown* | HPHT sample27 | ||
as-grown | annealed | as-grown | annealed | as-grown | as-grown | e-irradiated |
9.6 | 7.3 | 20.4 | 6.9 | 0.4 | 5.8 | 8.3 |
Secondary Ion Mass Spectroscopy (SIMS)
In order to gain more information about the composition of the diamond layers we performed secondary ion mass spectrometry (SIMS) measurements, shown in Fig. 4. The first important information from the SIMS data is that the growth rates for both processes are different. The growth time was set to two hours in both cases and the calculated growth rates are given in Table 1. Another interesting result is the varying nitrogen density within the layer, see Fig. 4(a). At the beginning of the growth process there is a maximum of the nitrogen concentration and then it decays monotonically towards the surface. Since no nitrogen gas was introduced in the chamber, it could be concluded that there is some limited nitrogen source, which is used up during the growth. Another possible source of nitrogen would be contamination of the methane gas, however this would mean a constant nitrogen concentration within the diamond layers. The total amount of nitrogen in the layers, after the annealing processes, was determined to be 1.07 × 1012 and 1.02 × 1012 atoms, from where we calculated the conversion efficiency from nitrogen to NV centres to be 0.25% and 0.21%, for Samples A and B respectively. These results are in good agreement to previous studies where the NV− concentration with respect to the total nitrogen content is approximately 1:300 (conversion efficiency 0.33%)12. The hydrogen SIMS data (Fig. 4(b)) show about three orders of magnitude higher hydrogen concentration in the layer compared to the substrate. From the data we calculate that the concentration varies within the layer from 10 to 3000 ppm, values which are still puzzling for us, though in earlier reports similar concentrations have been reported for polycrystalline CVD layers28,29. The hydrogen is either bond only to carbon to form the H129,30 and H229 defect centres or it is bound to an NV building the NVH31 centre. All these defects have been also observed in mono-crystalline diamond layers30,31, they are paramagnetic and stable at temperatures similar to our treatment here28,30. The high concentration of these spin impurities could explain the short values of the coherence time of the NVs compared to the case where the limiting factor is the presence of 13C nuclear spins.
Conclusion
In conclusion we have shown that the creation efficiency of NV ensembles in [111] oriented CVD grown diamond layers can be improved by high temperature annealing in vacuum. Moreover, this treatment leads to an increase of the coherence time as well as to a loss of the preferential alignment, where by using equation (2), this results in an enhancement of the magnetic field sensitivity by a factor of 2.9. However, the coherence times in those samples are shorter than the one expected for a limit posed by the dynamics of the 13C nuclear spin bath. This limitation is due to paramagnetic defects present in the grown layer, where possible candidates could be H1 and H2 centres suggested by the high hydrogen concentration revealed by the SIMS measurements.
Methods
CVD Diamond growth process
The diamond layers were produced in a home-built microwave assisted chemical vapour deposition chamber which consists of an external cavity (the vacuum chamber itself), a conical microwave coupling structure, a heated substrate holder and an inner cavity for plasma creation32. The microwave generator (Muegge electronic, type MW-GPEOM77B-5K-04\85) operates at an output frequency of 2.46 GHz and an output power up to 6 kW. The substrate holder is equipped with a graphite heater running at 1.5 kW which leads to a diamond substrate temperature of approximately 750 °C, measured with an infra-red pyrometer (Optris, type CTlaser 1 M\2 M) which operates at a wavelength of 1.6 μm. All gases used in this study are additionally purified. The hydrogen (chemical purity 99.99999%) by a palladium filter (Johnson Matthey, type Hydrogen Purifier HP-25) and the isotopically purified methane (isotopical purity 99.999%) by a heated getter device (MonoTorr, type PS4-MT3-531).
Confocal microscopy analysis
All experiments have been performed on a home built confocal microscope set-up equipped with a 515 nm pulsed diode laser (Toptica Photonics, type iBeam-smart-515-S). The electron spin manipulation was realised using a continuous wave MW source (Rohde und Schwarz, type SMIQ04B), a MW switch (Mini circuits, type ZASWA-2-50DR+) and an arbitrary waveform generator to form MW pulses (Tektronix, AWG7122C). The constant magnetic field has been provided by a permanent magnet which was mounted on a rotational 3D stage (Thorlabs, type PT03 and CR01). The experiments were controlled with the qudi software package33.
Annealing process
The diamond samples were annealed in a high vacuum chamber (<10−7 mbar), equipped with a boron-nitride heater (Tectra, type boralectric heater HTR-1001) with a maximum temperature of 1500 °C.
Supplementary information
Acknowledgements
We are grateful to Prof. S. Strehle for the experimental support, to S. Schmitt for fruitful discussions. This work has been supported by DFG, Volkswagenstiftung, EU (STREP Project DIADEMS, ERC Synergy Grant BioQ), and IQST. T.T. acknowledges the support of JSPS KAKENHI (Grant Number 26220903 and 16H06326), JST CREST Grant Number JPMJCR1773 and MEXT Q-LEAP (Quantum metrology & sensing), Japan. B.N. is grateful to the Bundesministerium für Bildung und Forschung (BMBF) for the ARCHES award.
Author Contributions
C.O., B.N. and F.J. designed the experiments. C.O., M.M. and J.L. performed the measurements. P.B. performed the theoretical calculations and T.T. performed the SIMS measurements. C.O., P.B. and B.N. wrote the manuscript which was reviewed by the other authors. All authors discussed the research.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Supplementary information accompanies this paper at 10.1038/s41598-019-42314-7.
References
- 1.Doherty MW, et al. The nitrogen-vacancy colour centre in diamond. Physics Reports. 2013;528:1–45. doi: 10.1016/j.physrep.2013.02.001. [DOI] [Google Scholar]
- 2.Dolde F, et al. Electric-field sensing using single diamond spins. Nature Phys. 2011;7:459. doi: 10.1038/nphys1969. [DOI] [Google Scholar]
- 3.Balasubramanian G, et al. Nanoscale imaging magnetometry with diamond spins under ambient conditions. Nature. 2008;455:648. doi: 10.1038/nature07278. [DOI] [PubMed] [Google Scholar]
- 4.Maze JR, et al. Nanoscale magnetic sensing with an individual electronic spin in diamond. Nature. 2008;455:644. doi: 10.1038/nature07279. [DOI] [PubMed] [Google Scholar]
- 5.Kucsko G, et al. Nanometre-scale thermometry in a living cell. Nature. 2013;500:54. doi: 10.1038/nature12373. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Doherty MW, et al. Electronic properties and metrology applications of the diamond nv-center under pressure. Phys. Rev. Lett. 2014;112:047601. doi: 10.1103/PhysRevLett.112.047601. [DOI] [PubMed] [Google Scholar]
- 7.Rondin L, et al. Magnetometry with nitrogen-vacancy defects in diamond. Reports on Progress in Physics. 2014;77:056503. doi: 10.1088/0034-4885/77/5/056503. [DOI] [PubMed] [Google Scholar]
- 8.Jelezko F, Gaebel T, Popa I, Gruber A, Wrachtrup J. Observation of coherent oscillations in a single electron spin. Phys. Rev. Lett. 2004;92:076401. doi: 10.1103/PhysRevLett.92.076401. [DOI] [PubMed] [Google Scholar]
- 9.Steinert S, et al. High sensitivity magnetic imaging using an array of spins in diamond. Review of Scientific Instruments. 2010;81:043705. doi: 10.1063/1.3385689. [DOI] [PubMed] [Google Scholar]
- 10.Hall, L. T. et al. High spatial and temporal resolution wide-field imaging of neuron activity using quantum nv-diamond. Scientific Reports2, 10.1038/srep00401 (2012). [DOI] [PMC free article] [PubMed]
- 11.Barry JF, et al. Optical magnetic detection of single-neuron action potentials using quantum defects in diamond. Proceedings of the National Academy of Sciences. 2016;113:14133–14138. doi: 10.1073/pnas.1601513113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Edmonds AM, et al. Production of oriented nitrogen-vacancy color centers in synthetic diamond. Phys. Rev. B. 2012;86:035201. doi: 10.1103/PhysRevB.86.035201. [DOI] [Google Scholar]
- 13.Michl J, et al. Perfect alignment and preferential orientation of nitrogen-vacancy centers during chemical vapor deposition diamond growth on (111) surfaces. Applied Physics Letters. 2014;104:102407. doi: 10.1063/1.4868128. [DOI] [Google Scholar]
- 14.Lesik M, et al. Perfect preferential orientation of nitrogen-vacancy defects in a synthetic diamond sample. Applied Physics Letters. 2014;104:113107. doi: 10.1063/1.4869103. [DOI] [Google Scholar]
- 15.Tahara K, Ozawa H, Iwasaki T, Mizuochi N, Hatano M. Quantifying selective alignment of ensemble nitrogen-vacancy centers in (111) diamond. Applied Physics Letters. 2015;107:193110. doi: 10.1063/1.4935709. [DOI] [Google Scholar]
- 16.Ozawa H, Tahara K, Ishiwata H, Hatano M, Iwasaki T. Formation of perfectly aligned nitrogen-vacancy-center ensembles in chemical-vapor-deposition-grown diamond (111) Applied Physics Express. 2017;10:045501. doi: 10.7567/apex.10.045501. [DOI] [Google Scholar]
- 17.Ishiwata H, et al. Perfectly aligned shallow ensemble nitrogen-vacancy centers in (111) diamond. Applied Physics Letters. 2017;111:043103. doi: 10.1063/1.4993160. [DOI] [Google Scholar]
- 18.Ozawa H, Ishiwata H, Hatano M, Iwasaki T. Thermal stability of perfectly aligned nitrogen-vacancy centers for high sensitive magnetometers. physica status solidi (a) 2018;215:1800342. doi: 10.1002/pssa.201800342. [DOI] [Google Scholar]
- 19.Staudacher T, et al. Nuclear magnetic resonance spectroscopy on a (5-nanometer)3 sample volume. Science. 2013;339:561–563. doi: 10.1126/science.1231675. [DOI] [PubMed] [Google Scholar]
- 20.Balasubramanian G, et al. Ultralong spin coherence time in isotopically engineered diamond. Nature Materials. 2009;8:383. doi: 10.1038/nmat2420. [DOI] [PubMed] [Google Scholar]
- 21.Mizuochi N, et al. Coherence of single spins coupled to a nuclear spin bath of varying density. Phys. Rev. B. 2009;80:041201. doi: 10.1103/PhysRevB.80.041201. [DOI] [Google Scholar]
- 22.Yamamoto T, et al. Extending spin coherence times of diamond qubits by high-temperature annealing. Phys. Rev. B. 2013;88:075206. doi: 10.1103/PhysRevB.88.075206. [DOI] [Google Scholar]
- 23.Beveratos, A. et al. Bunching and Antibunching from Single NV Color Centers in Diamond, 261–267 (Springer US, Boston, MA, 2002).
- 24.Naydenov B, et al. Enhanced generation of single optically active spins in diamond by ion implantation. Applied Physics Letters. 2010;96:163108. doi: 10.1063/1.3409221. [DOI] [Google Scholar]
- 25.Tetienne J-P, et al. Spin properties of dense near-surface ensembles of nitrogen-vacancy centers in diamond. Phys. Rev. B. 2018;97:085402. doi: 10.1103/PhysRevB.97.085402. [DOI] [Google Scholar]
- 26.Pinto H, et al. On the diffusion of nv defects in diamond. physica status solidi (a) 2012;209:1765–1768. doi: 10.1002/pssa.201200050. [DOI] [Google Scholar]
- 27.Wolf T, et al. Subpicotesla diamond magnetometry. Phys. Rev. X. 2015;5:041001. doi: 10.1103/PhysRevX.5.041001. [DOI] [Google Scholar]
- 28.Jia H, Shinar J, Lang DP, Pruski M. Nature of the native-defect esr and hydrogen-dangling-bond centers in thin diamond films. Phys. Rev. B. 1993;48:17595–17598. doi: 10.1103/PhysRevB.48.17595. [DOI] [PubMed] [Google Scholar]
- 29.Zhou X, Watkins GD, McNamara Rutledge KM, Messmer RP, Chawla S. Hydrogen-related defects in polycrystalline cvd diamond. Phys. Rev. B. 1996;54:7881–7890. doi: 10.1103/PhysRevB.54.7881. [DOI] [PubMed] [Google Scholar]
- 30.Mizuochi, N., Watanabe, H., Isoya, J., Okushi, H. & Yamasaki, S. Hydrogen-related defects in single crystalline cvd homoepitaxial diamond film studied by epr. Diamond and Related Materials13, 765–768, 14th European Conference on Diamond, Diamond-Like Materials, Carbon Nanotubes, Nitrides and Silicon Carbide, 10.1016/j.diamond.2003.11.028 (2004).
- 31.Glover C, Newton ME, Martineau P, Twitchen DJ, Baker JM. Hydrogen incorporation in diamond: The nitrogen-vacancy-hydrogen complex. Phys. Rev. Lett. 2003;90:185507. doi: 10.1103/PhysRevLett.90.185507. [DOI] [PubMed] [Google Scholar]
- 32.Silva, F., Bonnin, X., Scharpf, J. & Pasquarelli, A. Microwave analysis of pacvd diamond deposition reactor based on electromagnetic modelling. Diamond and Related Materials19, 397–403, Proceedings of Diamond 2009, The 20th European Conference on Diamond, Diamond-Like Materials, Carbon Nanotubes and Nitrides, Part 1, 10.1016/j.diamond.2009.10.032 (2010).
- 33.Binder JM, et al. Qudi: A modular python suite for experiment control and data processing. SoftwareX. 2017;6:85–90. doi: 10.1016/j.softx.2017.02.001. [DOI] [Google Scholar]