Detection of Fentanyl Analogs and Synthetic Opioids in Real Hair Samples

Alberto Salomone 1,, Joseph J Palamar 2,3, Rachele Bigiarini 4, Enrico Gerace 1, Daniele Di Corcia 1, Marco Vincenti 1,4
PMCID: PMC6460334  PMID: 30462247

Abstract

Novel synthetic opioids include various analogs of fentanyl and emerging non-fentanyl compounds with different chemical structures, such as AH-7921, MT-45 and U-47700. In recent years, these drugs have rapidly emerged on the drug market, and their abuse has been increasing worldwide. The motivations for use of these new compounds include their legal status, ready availability, low cost, users’ curiosity or preference for their particular pharmacological properties and the intention to avoid detection. Furthermore, more common drugs like heroin are now increasingly being replaced or cut with fentanyl or new designer opioids; thus, many drug users are unintentionally or unknowingly using synthetic fentanyl analogs. In this scenario, the detection of new psychoactive substances in hair can provide insight into their current diffusion among the population and social characteristics of these synthetic drug users. In this manuscript, we describe a simple, fast, specific and sensitive UHPLC–MS-MS method able to detect 13 synthetic opioids (including fentanyl analogs) and metabolites in hair samples. Furthermore, the method includes the detection of 4-anilino-N-phenethyl-piperidine (4-ANPP), which is considered both a precursor and a metabolite of several fentanyl analogs. The method was applied to 34 real hair samples collected in New York City from subjects who had reported past-year non-medical opioid and/or heroin use. In total, 17 samples tested positive for at least one target analyte, with oxycodone (nine samples) and tramadol (eight samples) being the most common. Among these, the method was able to quantify furanyl-fentanyl and fentanyl in the pg/mg range in two samples. Simultaneously, also 4-ANPP was detected, giving evidence for the first time that this compound can be selected as a marker of fentanyl analogs use via hair testing. In conclusion, this study confirmed the increasing diffusion of new synthetic opioids and “fentalogs” with high potency among non-medical opioid users.

Introduction

For many years, the spectrum of illegal psychoactive drugs amounted to relatively few substances, commonly referred to as common/traditional drugs of abuse. More recently, a large upsurge of new psychoactive substances (NPS) has been observed (1). Although most NPS fall into the classes of synthetic cannabinoids and designer cathinones (1), novel synthetic opioids (NSO) have also appeared recently on the drug market with a progressive increase in their consumption (2) Many of the NSO are derivatives of a well-known pharmaceutically used substance, fentanyl. However, new synthetic opioids such as AH-7921, MT-45 and U-47700, have chemical structures distinct from any known therapeutic or recreational drug. These substances are often sold as “research chemicals” mainly through internet shops and used as substitutes of controlled opioids, but they are also increasingly sold under the guise of heroin posing a high risk of severe intoxication to unaware users (3). The motivations for the use of these new drugs include their legal status, availability, low cost, user curiosity or preference for their particular pharmacological properties and the intention to avoid detection (2). Furthermore, major public health concern is raised by the evidence that common drugs like heroin are now frequently replaced or cut with new designer opioids (4, 5). While heroin is the most common drug to be adulterated with these novel opioids, these compounds have been also detected in cocaine, methamphetamine, ketamine, ecstasy and counterfeit Xanax® and Percocet® pills (69). Therefore, many drug users are unintentionally or unknowingly using synthetic fentanyl analogs (10). Since the composition and potency of these compounds on the “street” market is highly variable, their use may lead to severe intoxications and overdose fatalities—particularly among opioid-naïve individuals in which overdoses are not as easily reversed with naloxone (8). Several cases of intoxications and deaths related to the uptake of NSO have been reported to the United Nations Office on Drugs and Crime (UNODC) and have been summarized in the Global SMART Update on Fentanyl analogs (11). Fentanyl has been particularly problematic in the USA, where it was detected in over half (56%) of opioid-related deaths in 2016, and its analogs were detected in 14% of fatal cases (12). As a consequence of increasing prevalence and emergence of NPS, forensic and clinical laboratories worldwide are continuously requested to update their analytical procedures for the identification and quantification of these new drugs in various biological matrices. Few preliminary methods have been published for their detection in conventional matrices such as urine (13) and blood (13, 14), and more generally in biological specimens (15). Recent review papers have summarized the cases of fatal overdoses and the analytical approaches used for the investigation of such cases (16, 17).

To the best of our knowledge, comprehensive methods for testing fentanyl and/or its analogs in hair have not been published yet. Furthermore, the detection of new synthetic opioids in hair samples can serve as a practical means to provide preliminary information on black market penetration of these analogs in specific territories and populations, similarly to how recent testing for synthetic cannabinoids and synthetic cathinones has been utilized (1825). The keratin matrix incorporates the parent NPS consumed once ore repeatedly over extended time periods, providing access to a much wider diagnostic window than urine and saliva (26). This feature, combined with the analytical performances of the last generation ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC–MS-MS) instruments, allows researchers to obtain significant information about past exposure, even a single intake, of any targeted NPS, with older periods of use corresponding to the hair segments more distant from the hair root (27, 28).

In the present study, we focused on the specific population of electronic dance music (EDM) nightclub and festival attendees in New York City (NYC), as a high-risk population for both intentional and unintentional exposure to common drugs (29) and various NPS (30, 31). Inclusion of both common opioids/opiates and novel opioids in our screening allowed us to distinguish different consumption patterns, including co-use, occasional NSO intake, and frequent use. To our knowledge, this is the first study to examine hair results of new opioids use in a selected population.

Experimental

Reagents and standards

Furanyl-fentanyl and 4-ANPP were produced by Chiron (Trondheim, Norway). Acetyl-fentanyl, remifentanil and carfentanil were produced by Toronto Research Chemicals (North York, Canada). Tramadol, oxycodone, hydrocodone and norfentanyl were purchased from Sigma-Aldrich (Milan, Italy). Alfentanil, U-47700, fentanyl, sufentanil and the deuterated internal standards (norfentanyl-D5, fentanyl-D5 and oxycodone-D6) were produced by Cerilliant (Round Rock, Texas, USA). Ultra-pure water was obtained using a Milli-Q® UF-Plus apparatus (Millipore, Bedford, MA, USA). All stock standard solutions were prepared in methanol at 1 mg/mL and stored at −20°C until used. Working solutions were prepared at the final concentration of 1,000 ng/mL by dilution with methanol.

Sample preparation

Briefly, ~25 mg of hair was twice-washed with dichloromethane and then methanol (2 mL, vortex mixed for 3 min). After complete removal of solvent washes, the hair was dried at room temperature by a gentle nitrogen flow and subsequently grinded with a ball mill (Precellys 24, Bertin Instruments, Montigny-le-Bretonneux, France). Hair samples were fortified with 2.5 μL of an internal standards mixture yielding a final concentration of 0.01 ng/mg. After the addition of 1 mL of methanol, the samples were incubated at 55°C for 15 h without stirring. Lastly, the organic phase was collected and an aliquot of 2 μL was directly injected into the UHPLC–MS-MS system. Whenever the real samples concentrations were found to exceed the highest calibration point, the final extracts were diluted with methanol and re-injected into the system.

Instrumentation

All analyses were performed using a Shimadzu LC-30A Series system (Shimadzu, Duisburg, Germany), interfaced to an API 5500 triple quadrupole mass spectrometer (Sciex, Darmstadt, Germany) equipped with an electrospray Turbo Ion source operating in positive-ion mode. A CORTECS UPLC C18 plus column 100 mm × 2.1 mm, 1.6 μm (Waters Corporation, Italy), protected by a C18 waters VanGuard pre-column, was used for the target analytes separation. The column oven was maintained at 45°C and the elution solvents were water/formic acid 5 mM (Solvent A) and acetonitrile plus formic acid 5 mM (Solvent B). After an initial isocratic elution at 95% A for 0.5 min, the mobile-phase composition was varied by a linear gradient (A:B; v/v) from 95:5 to 50:50 in 3.5 min; then isocratic elution at 50% B was maintained for 0.5 min. The flow rate was 0.5 mL/min and the total run time was 6 min including re-equilibration at the initial conditions before each injection. MS-MS detection was executed in the selected reaction monitoring (SRM) mode. In order to establish appropriate SRM conditions, each analyte was individually infused into the electrospray ionization (ESI) capillary while the declustering potential (DP) was adjusted to maximize the intensity of the protonated molecular species [M+H]+. The collision energy (CE) was set so as to preserve ~10% of precursor ion and the cell exit potentials (CXP) were also optimized. The SRM transitions were monitored during a time window of ±17.5 s around the expected retention time, and the cycle time of the SRM program was 0.100 s. Optimal signals were obtained using a source block temperature of 600°C and an ion-spray voltage of 1,900 V. Gas pressures were set as follows: curtain gas 35 psi, ion source gas (1) 45 psi and ion source gas (2) 50 psi. SRM transitions and potentials for the analytes and internal standards are presented in Table I.

Table I.

SRM transitions and corresponding potentials for the target compounds and internal standard detection

Compound RT (min) Precursor Ion [M+H]+ DP (V) Target Qualifier 1 Qualifier 2
Fragment CE (V) CXP (V) Fragment CE (V) CXP (V) Fragment CE (V) CXP (V)
1 Oxycodone 2.1 316.0 70 241.1 38 11 298.1 25 14 256.1 34 11
2 Hydrocodone 2.1 300.0 76 199.1 42 9 171.1 51 7 128.0 74 15
3 Norfentanyl 2.4 233.0 50 84.1 24 9 150.1 22 14 55.0 50 8
4 Tramadol 2.6 264.1 35 58.1 46 11
5 Remifentanil 2.8 377.1 43 317.2 22 14 228.0 27 8 116.1 37 14
6 Acetyl-Fentanyl 3.0 323.0 73 188.2 38 15 105.0 38 14 77.0 86 10
7 Alfentanil 3.3 417.0 52 268.3 24 8 197.2 35 9 165.0 47 16
8 U-47700 3.3 330.9 47 286.1 24 11 206.1 34 8
328.9 47 204.1 36 7
9 Fentanyl 3.3 337.1 57 188.2 32 6 105.0 49 5 132.1 42 6
10 4-ANPP 3.3 281.0 72 188.2 24 8 105.0 41 12 103.0 63 16
11 Furanyl-Fentanyl 3.4 375.0 55 188.2 28 14 105.0 52 13 103.0 82 16
12 Carfentanil 3.5 395.0 50 335.2 25 11 246.1 34 9 113.0 34 6
13 Sufentanil 3.7 387.0 43 238.1 26 10 355.2 26 15 111.0 46 12
IS1 Oxycodone-d6 2.0 322.0 70 247.1 38 11
IS2 Norfentanyl-d5 2.4 238.0 50 84.1 24 9
IS3 Fentanyl-d5 3.3 342.0 57 188.2 32 6

DP, Declustering potential; CE, Collision Energy; CXP, Cell Exit Potential.

Method validation

The following parameters were investigated: selectivity, specificity, linearity range, detection and quantification limits (LOD and LOQ), intra-day and interday precision and trueness. Carry-over effect, recovery and matrix effects were also investigated. One qualifying SRM transition was monitored, in addition to the primary fragmentation (see Table I). The relative abundances of mass transitions intensities were considered acceptable if they did not differ by more than ±20%, with respect to the corresponding relative abundances of the same transitions acquired from a positive control sample. Selectivity was determined on five blank head hair samples. The signal-to-noise ratio (S/N) was measured on the less intense mass transition at the expected analyte retention time. A S/N < 3 was considered satisfactory in order to verify the method’s selectivity.

The calibration process was conducted with an optimized procedure, requiring the preparation of three replicates of the calibration curves spiked with the working solution at six concentration levels (1, 2, 5, 10, 20 and 50 pg/mg and 10, 20, 50, 100, 200 and 500 pg/mg for Oxycodone), for three consecutive days for a total of nine calibration curves. The data from each specific calibration curve were quantified using a calibration curve obtained in a different day, allowing to manage each set of data as independent. Therefore, nine samples (from nine different batches) per each calibration level were employed to estimate the following validation parameters: linearity, LODs, LOQs, intra-assay precision and trueness. The calibration was completed by internal standardization. The linear calibration parameters were evaluated using the least squares regression method; several significance tests were performed to evaluate linearity, including lack-of-fit test and analysis of variance (ANOVA). Determination coefficient (R2), adjusted determination coefficient (Adj R2), relative standard deviation of the slope, normality of the standardized residuals and deviation from back-calculated concentrations were also evaluated. The LODs were estimated using the Hubaux–Vox approach (32), and the LOQs were then approximated as three times the LOD values. In-house developed spreadsheets were employed to evaluate these parameters and perform their relative significance tests. For all analytes, intra- and interday precision (expressed as percent variation coefficient, CV%) and trueness (expressed as bias %) were evaluated at three concentration levels. The samples prepared for the evaluation of linearity were used, thus consisting of nine replicates (prepared and analyzed in three different days) of blank hair samples spiked with the standard solutions at the lowest calibration point (close to the estimated LOQ values), at intermediate calibration level and the highest calibration point. Intra-day and interday precision parameters were considered satisfactory when CV% values turned below 15%. Satisfactory trueness (in terms of bias %) was achieved when the experimentally determined average concentration lied within ±20% from the expected value. Extraction recoveries were determined by comparing the responses obtained from samples (five replicates) initially spiked with the analytes at a concentration of 20 pg/mg (200 pg/mg for Oxycodone) and subsequently extracted and processed as usual, with the responses of blank samples in which the analytes were added at the same concentration after the extraction step. The matrix effect was calculated, by comparing the peak area ratio between analyte and ISTD obtained from hair sample extracts spiked after the extraction step, with the corresponding ratio obtained from a pure methanol solution, at the same concentrations. In this case, the matrix effect is expected to be partly compensated by a well-matched internal standard, i.e., the isotopically-marked analyte, whenever possible, or the one having the closest RT to the analyte, and possibly suffering similar interferences from the matrix. The matrix effect was calculated as the mean value obtained from five different hair sources. The percent difference represented either matrix suppression (values below 100%) or matrix enhancement (values above 100%). The possible presence of carry-over effects was evaluated by injecting an alternate sequence of five blank head hair samples and five extracts from blank head hair samples spiked with all analytes at the highest concentration point of the calibration curve (50 pg/mg and 500 for Oxycodone). To ensure the absence of any carry-over effect, the signal-to-noise ratio had to be lower than three for each monitored transition.

The effect of dilution when a real sample exceeded the highest calibration point was validated as follows: 10 spiked hair samples at 500 pg/mg (5,000 pg/mg for oxycodone) were 10-fold diluted and then injected. Trueness and precision were therefore evaluated with 50 pg/mg (500 pg/mg for oxycodone) as target value and considered acceptable if within ±15%.

Analysis of real samples

Subjects were eligible if they identified as age 18–40 and were about to attend the randomly selected party in NYC. After providing informed consent, subjects were surveyed about use of dozens of drugs including 18 common and novel opioids (33). At the end of the survey, participants were asked if they were willing to donate a hair sample to be analyzed at a future date. Collection was conducted from July through September of 2017. This study focused on the participants who reported any past-year non-medical opioid or heroin use and provided a useable hair sample for analysis. If the participant agreed, the trained recruiter collected the sample by cutting a small lock of hair (~100 hairs) from as close to the participant’s scalp as possible. Hair was cut with a clean scissor (wiped with an alcohol-wipe after each use), folded up in a piece of tin foil, and stored at room temperature in a small envelope labeled with the participant’s anonymous study ID number (34). A total of 34 real hair samples was collected. All samples were analyzed in their full length. This study was approved by the New York University Langone Medical Center Institutional Review Board.

Results and Discussion

The optimized UHPLC–MS-MS method allowed the simultaneous determination of 13 fentanyl analogs and NSO in hair samples. The whole chromatographic run, comprehensive of the time required for column re-equilibration before the following injection, was completed in 6 min. Retention times ranged between 2.0 min (oxycodone) and 3.70 min (sufentanil). Tramadol produced only one product ion (m/z 58.1) Figure 1 shows the SRM chromatograms recorded from a blank hair spiked with all analytes at 5 pg/mg concentration (50 pg/mg for oxycodone).

Figure 1.

Figure 1.

SRM chromatograms recorded from blank hair sample spiked with all analytes at 5 pg/mg concentration (50 pg/mg for oxycodone). For each analyte, labeled by the progressive number assigned in Table I, only the target transition is shown.

Validation

All the validation results are reported in Table II. No carry-over effect was observed under the conditions described in the experimental section. Selectivity tests proved successful, i.e., SRM chromatograms from negative head hair samples showed no interfering signals at the retention time where the analytes were expected to elute. The trend of the variance associated to the tested calibration points (i.e., homoscedasticity vs. heteroscedasticity) was evaluated. A weighting factor equal to (1/x2), where x represents the concentration of the calibration point, was adopted for all analytes to compensate for heteroscedastic distribution of the data points. The lack-of-fit and ANOVA tests suggested a linear regression model for all analytes. Table IIa reports a summary of the observed results, including the Adj R2 values of the regression models. LOD and LOQ values estimated by Hubaux–Vos’ methodology are reported, too. In particular, LOD values were in the range 0.1–0.3 pg/mg for all analytes, with the exception of oxycodone which was 1.5 pg/mg. LOQ values lied between 0.3–0.9 pg/mg and 1.5 pg/mg, respectively. The calculated LODs were experimentally confirmed with one blank hair sample spiked at the estimated LODs concentrations. Extraction recoveries were mostly close to 100% and always in the interval 100 ± 30%, as estimated from samples spiked at 20 pg/mg concentration (200 pg/mg for oxycodone). The effect of the real hair matrix appeared to be significant for some of the analytes tested, particularly when the absolute peak areas are considered (ion suppression was observed for oxycodone and tramadol). After peak area correction using the internal standards (Table IIb), ion suppression and ion enhancement were mostly compensated. The variability among different hair samples turned out acceptable (CV% <25%) for all the analytes, provided that the results are normalized using the ISTD. Furthermore, the good linearity observed in the calibration plots supports the observation of constant percent matrix effect, which in fact does not depend on the analytes’ concentration. Intra-day precision and trueness were satisfactory for all analytes at all calibration level (1, 10 and 50 pg/mg and 10, 100, 500 for oxycodone), while modest deviation was observed in the evaluation of interday precisionfor remifentanil and sufentanil. Finally, precision and trueness were acceptable also in the tests that investigated the dilution effect.

Table IIa.

Range of calibration, linearity, LODs and LOQs values, linearity tests over the calibration interval for all analytes

Compound Linearity range (pg/mg) Internal standard (IS) Linearity (Adj R2) LOD (pg/mg) LOQa (pg/mg) Lack-of-Fit (Fcrit = 2.37) ANOVA test (Ftab = 3.69)
1 Oxycodone 10–500 Oxycodone-d6 0.9999 1.5 4.5 0.08 0.06
2 Hydrocodone 1–50 Fentanyl-d5 0.9994 0.1 0.3 0.27 0.37
3 Norfentanyl 1–50 Norfentanyl-d5 0.9995 0.1 0.3 0.37 0.18
4 Tramadol 1–50 Norfentanyl-d5 0.9996 0.1 0.3 0.35 0.20
5 Remifentanil 1–50 Fentanyl-d5 0.9968 0.3 0.9 1.93 0.70
6 Acetyl-Fentanyl 1–50 Fentanyl-d5 0.9980 0.2 0.6 1.57 0.34
7 Alfentanil 1–50 Fentanyl-d5 0.9992 0.1 0.3 0.65 0.21
8 U-47700 1–50 Norfentanyl-d5 0.9998 0.1 0.3 0.09 0.28
9 Fentanyl 1–50 Fentanyl-d5 0.9993 0.1 0.3 0.53 0.27
10 4-ANPP 1–50 Fentanyl-d5 0.9994 0.1 0.3 0.48 0.21
11 Furanyl-Fentanyl 1–50 Fentanyl-d5 0.9991 0.1 0.3 0.59 0.41
12 Carfentanil 1–50 Fentanyl-d5 0.9983 0.2 0.6 0.76 0.72
13 Sufentanil 1–50 Fentanyl-d5 0.9961 0.3 0.9 1.46 0.90

aCalculated LOQ.

Table III.

Results from samples which tested positive for at least one NSO (concentrations are given in pg/mg)

Subject Detected analytes Heroin metabolites
Case Length (cm) Fentanyl Furanyl-fentanyl 4-ANPP Oxycodone Hydrocodone Tramadol Morphine 6-AM Codeine
1 3.5 3 44 3 8700 >10,000 580
2 7.5 6 1 260 3060 59
3 2.0 240 4
4 1.5 780
5 3.5 530
6 2.0 60
7 2.0 13
8 12.0 110 14
9 8.0 52 71
10 2.5 95
11 12.0 33
12 2.0 48
13 1.0 71 2
14 4.0 110
15 12.0 3700
16 6.0 13 22
17 5.0 13 48
Positive findings 2 1 2 9 4 8 2 2 2

Note. All samples were collected from participants’ heads. 6-AM: 6-acetylmorphine.

Table IIb.

Precision, trueness, recovery and matrix effect for all analytes

Compound Intra-day precision (CV%) Interday precision (CV%) Trueness (bias%) Recoverya (%) Matrix effectb Matrix effectc
1 pg/mg 10 pg/mg 50 pg/mg 1 pg/mg 10 pg/mg 50 pg/mg 1 pg/mg 10 pg/mg 50 pg/mg Mean (±%) CV (%) Mean (±%) CV (%)
1 Oxycodone 10 4 2 9 9 3 9 1 −0.2 95 −31 33 3 18
2 Hydrocodone 14 9 5 7 12 8 15 8 1 85 11 29 9 6
3 Norfentanyl 6 3 3 12 10 2 6 −6 −0.3 93 5 12 4 0.5
4 Tramadol 4 4 1 10 7 4 2 −5 −0.3 98 −30 16 10 7
5 Remifentanil 14 4 2 15 11 16 −19 −7 −1 71 4 21 5 11
6 Acetyl-Fentanyl 3 2 3 14 5 4 6 −3 −1 91 −8 13 −3 13
7 Alfentanil 4 4 3 14 13 1 −13 −5 −0.6 99 −4 10 9 2
8 U-47700 10 2 2 11 3 13 16 2 −0.2 112 3 20 8 7
9 Fentanyl 3 3 3 14 5 7 −13 −3 −0.6 96 4 31 10 11
10 4-ANPP 7 4 2 13 10 4 10 −8 −0.2 88 −3 22 3 22
11 Furanyl-Fentanyl 7 2 3 12 7 12 17 −4 −1 96 5 28 10 17
12 Carfentanil 6 3 3 15 14 8 −6 −3 −1 96 −9 5 5 0.5
13 Sufentanil 15 5 4 21 14 13 −19 −5 −2 92 −14 10 2 3

aRecovery evaluated at 20 pg/mg for all analytes except for Oxycodone tested at 200 pg/mg.

bMatrix effect was evaluated using five different sources of hair (2 pg/mg for all analytes except for oxycodone tested at 20 pg/mg).

cMatrix effect was evaluated on the same sources of hair, but in this case the effect was compensated by the use of ISTD.

CV%, percent variation coefficient.

Analysis of real samples

A total of 34 samples was considered to verify the applicability of the developed method. Among these, 17 samples (50%) tested positive for at least one opioid compound (Table III). The molecules detected were oxycodone (nine samples, range of concentrations: 13–780 pg/mg), tramadol (eight samples, range 2.0–3700 pg/mg), hydrocodone (four samples, range 13–71 pg/mg), fentanyl (two samples, respectively, 3 and 6 pg/mg), furanyl-fentanyl (one sample, 44 pg/mg), and 4-ANPP (two samples, respectively, 2 and 1 pg/mg). Since the hair samples were analyzed in their full length, the cited quantitative results refer to averaged values. Specific hair segments may possibly contain much higher local concentration of the analytes. The other compounds included in the panel of target analytes were not detected in any of the selected samples. Fentanyl was measured at the level of low pg/mg, suggesting low rate of incorporation into the keratin matrix or, more likely, the exposure to low quantities of the drug. Like other classes of NPS (35), more data about the detection of NSO in hair are needed before any clear interpretation of analytical findings can be given. All samples were also tested for common opiates (i.e., heroin) with a previously published method for traditional drugs of abuse (36). Remarkably, only the two samples positive to fentanyl also showed a high concentrations of heroin metabolites. Unfortunately, it was not possible to perform a segmental analysis to discriminate if fentanyl and heroin had been taken in the same period, due to the low amount of available hair. Therefore, it is impossible to discriminate if fentanyl was taken alone or as a heroin cutting agent. However, one of the two subjects whose hair tested positive for fentanyl admitted known past-year use of fentanyl, while the other denied any voluntary past use of fentanyl. In the latter case, it is likely that NSO were consumed unknowingly. The metabolite norfentanil could not be detected in the two samples positive to fentanyl at low concentrations. This is not surprising, since it is well-known that metabolites are usually incorporated in hair samples at a lower rate than the parent compound. On the other hand, 4-anilino-N-phenethyl-piperidine (4-ANPP) was detected in one sample positive to fentanyl and fluorofentanyl, with comparable concentrations. The 4-ANPP is considered both a precursor and a metabolite of several fentanyl analogs (14, 3739). Therefore, its detection in biological samples can possibly be identified as a promising marker for some fentanyl analogs intake, even when the (newly introduced) drug is not included in the panel of the target analytes. More research is needed in order to interpret the 4-ANPP concentration in hair, in particular with respect to fentanyl concentration. In this perspective, the ratio 4-ANPP/norfentanyl will likely be useful to speculate if only fentanyl was taken, or instead other analogs had contributed to the 4-ANPP concentration.

Conclusions

The present study demonstrated that 13 new synthetic opioids can be determined in the keratin matrix with high sensitivity and specificity, allowing wide-range monitoring of drug intake over extended periods of time. In the future, further pharmacologically-related substances—either analogs of fentanyl or synthetic opioids with chemical structures unrelated to fentanyl—will be added to the panel of the target analytes. As a matter of fact, the evolution of the black market is unpredictable; thus, a comprehensive screening method for NSO is needed in order to draw a realistic picture of consumption in specific populations and at different times. In this scenario, the use of 4-ANPP as a common marker for several fentanyl analogs appears to be extremely promising.

Furthermore, hair testing using large NSO screening panels that takes into account the timely fluctuation of the market can finally be applied to the individuals undergoing periodic control (e.g., in workplace control or driving re-licensing) to check how frequently these compounds are used, either knowingly or unknowingly.

Funding

Research reported in this publication was supported by the National Institute on Drug Abuse of the National Institutes of Health under Award Number K01DA038800. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.

Conflict of Interest

The authors declare no conflict of interest.

References

  • 1.(2017) EMCDDA, European Drug Report 2017: trends and development, http://www.emcdda.europa.eu/edr2017_en (accessed May 24, 2018)
  • 2.(2017) Drug Enforcement Administration. Counterfeit prescription pills containing fentanyls: a global threat [DEA Intelligence Brief]. Available at: https://www.dea.gov/docs/Counterfeit%20Prescription%20Pills.pdf. (accessed May 28, 2017).
  • 3.(2017) European Monitoring Centre for Drugs and Drug Addiction. European Drug Report. Trends and developments 2015. http://www.emcdda.europa.eu/system/files/publications/974/TDAT15001ENN.pdf. (accessed May 24, 2018)
  • 4. Prekupec M.P., Mansky P.A., Baumann M.H. (2017) Misuse of novel synthetic opioids: a deadly new trend. Journal of Addiction Medicine, 11, 256–265. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Misailidi N., Papoutsis I., Nikolao P., Dona A., Spiliopoulou C., Athanaselis S. (2018) Fentanyls continue to replace heroin in the drug arena: the cases of ocfentanil and carfentanil. Forensic Toxicology, 36, 12–32. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Arens A.M., van Wijk X.M., Vo K.T., Lynch K.L., Wu A.H., Smollin C.G. (2016) Adverse Effects From Counterfeit Alprazolam Tablets. JAMA internal medicine, 176, 1554–1555. [DOI] [PubMed] [Google Scholar]
  • 7. McCall Jones C., Baldwin G.T., Compton W.M. (2017) Recent increases in cocaine-related overdose deaths and the role of opioids. American Journal of Public Health, 107, 430–432. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Sutter M.E., Gerona R.R., Davis M.T., Roche B.M., Colby D.K., Chenoweth J.A., et al. (2017) Fatal fentanyl: one pill can kill. Academic Emergency Medicine, 24, 106–113. [DOI] [PubMed] [Google Scholar]
  • 9. Armenian P., Olson A., Anaya A., Kurtz A., Ruegner R., Gerona R.R. (2017) Fentanyl and a novel synthetic opioid U-47700 Masquerading as Street “Norco” in Central California: a case report. Annals of Emergency Medicine, 69, 87–90. [DOI] [PubMed] [Google Scholar]
  • 10. Mars S.G., Ondocsin J., Ciccarone D. (2018) Sold as heroin: perceptions and use of an evolving drug in Baltimore, MD. Journal of Psychoactive Drugs, 50, 167–176. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.(2017) United Nations Office on Drugs and Crime. UNODC Global Smart Update, Fentanyl and its analogues—50 years on, vol. 17. Available from: https://www.unodc.org/documents/scientific/Global_SMART_Update_17_web.pdf (accessed May 24, 2018)
  • 12. O’Donnell J.K., Halpin J., Mattson C.L., Goldberger B.A., Gladden R.M. (2017) Deaths involving fentanyl, fentanyl analogs, and U-47700–10 States, July-December 2016. Morbidity and Mortality Weekly Report, 66, 1197–1202. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Marchei E., Pacifici R., Mannocchi G., Marinelli E., Busardò F.P., Pichini S. (2018) New synthetic opioids in biological and non-biological matrices: a review of current analytical methods. Trends in Analytical Chemistry, 102, 1–15. [Google Scholar]
  • 14. Fogarty M.F., Papsun D.M., Logan B.K. (2018) Analysis of fentanyl and 18 novel fentanyl analogs and metabolites by LC–MS-MS, and report of fatalities associated with methoxyacetylfentanyl and cyclopropylfentanyl. Journal of Analytical Toxicology. 10.1093/jat/bky035. [DOI] [PubMed] [Google Scholar]
  • 15.(2017) UNODC, Recommended methods for the Identification and Analysis of Fentanyl and its analogues in Biological Specimens, https://www.unodc.org/…/Recommended_methods_for_the_identification_and_Analysis_of_Fentanyl.pdf (accessed May 24, 2018)
  • 16. Drummer O.H. (2018) Fatalities caused by novel opioids: a review. Forensic Sciences Research. 10.1080/20961790.2018.1460063. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Gerace E., Salomone A., Vincenti M. (2018) Analytical approaches in fatal intoxication cases involving new synthetic opioids. Current Pharmaceutical Biotechnology, 19, 113–123. 10.2174/1389201019666180405162734. [DOI] [PubMed] [Google Scholar]
  • 18. Salomone A., Gerace E., D’Urso F., Di Corcia D., Vincenti M. (2012) Simultaneous analysis of several synthetic cannabinoids, THC, CBD and CBN, in hair by ultra-high performance liquid chromatography tandem mass spectrometry. Method validation and application to real samples. Journal of Mass Spectrometry, 47, 604–610. [DOI] [PubMed] [Google Scholar]
  • 19. Salomone A., Luciano C., Di Corcia D., Gerace E., Vincenti M. (2014) Hair analysis as a tool to evaluate the prevalence of synthetic cannabinoids in different populations of drug consumers. Drug Testing and Analysis, 6, 126–134. [DOI] [PubMed] [Google Scholar]
  • 20. Hutter M., Kneisel S., Auwärter V., Neukamm M. (2012) Determination of 22 synthetic cannabinoids in human hair by liquid chromatography-tandem mass spectrometry. Journal of Chromatography B, 903, 95–101. [DOI] [PubMed] [Google Scholar]
  • 21. Gottardo R., Sorio D., Musile G., Trapani E., Seri C., Serpelloni G., et al. (2014) Screening for synthetic cannabinoids in hair by using LC-QTOF MS: a new and powerful approach to study the penetration of these new psychoactive substances in the population. Medicine, Science, and the Law, 54, 22–27. [DOI] [PubMed] [Google Scholar]
  • 22. Rust K.Y., Baumgartner M.R., Dally A.M., Kraemer T. (2012) Prevalence of new psychoactive substances: a retrospective study in hair. Drug Testing and Analysis, 4, 402–408. [DOI] [PubMed] [Google Scholar]
  • 23. Salomone A., Gazzilli G., Di Corcia D., Gerace E., Vincenti M. (2016) Determination of cathinones and other stimulant, psychedelic, and dissociative designer drugs in real hair samples. Analytical and Bioanalytical Chemistry, 408, 2035–2042. [DOI] [PubMed] [Google Scholar]
  • 24. Palamar J.J., Salomone A., Vincenti M., Cleland C.M. (2016) Detection of “bath salts” and other novel psychoactive substances in hair samples of ecstasy/MDMA/“Molly” users. Drug and Alcohol Dependence, 161, 200–205. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Palamar J.J., Salomone A., Gerace E., Di Corcia D., Vincenti M., Cleland C.M. (2017) Hair testing to assess both known and unknown use of drugs amongst ecstasy users in the electronic dance music scene. International Journal of Drug Policy, 48, 91–98. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Silverstein J.H., Rieders M.F., McMullin M., Schulman S., Zahl K. (1993) An analysis of the duration of fentanyl and its metabolites in urine and saliva. Anesthesia & Analgesia, 76, 618–621. [DOI] [PubMed] [Google Scholar]
  • 27. Vincenti M., Salomone A., Gerace E., Pirro V. (2012) Application of mass spectrometry to hair analysis for forensic toxicological investigations. Mass Spectrometry Reviews, 32, 1–21. [DOI] [PubMed] [Google Scholar]
  • 28. Kintz P., Salomone A., Vincenti M. (eds).. Hair Analysis in Clinical and Forensic Toxicology. Elsevier–Academic Press: San Diego, CA, USA, 2015. [Google Scholar]
  • 29. Palamar J.J., Griffin-Tomas M., Ompad D.C. (2015) Illicit drug use among rave attendees in a nationally representative sample of US high school seniors. Drug and Alcohol Dependence, 152, 24–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Palamar J.J., Acosta P., Sherman S., Ompad D.C., Cleland C.M. (2016) Self-reported use of novel psychoactive substances among attendees of electronic dance music venues. American Journal of Drug and Alcohol Abuse, 42, 624–632. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Palamar J.J., Barratt M.J., Ferris J.A., Winstock A.R. (2016) Correlates of new psychoactive substance use among a self-selected sample of nightclub attendees in the United States. The American Journal on Addictions, 25, 400–407. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Hubaux A., Vos G. (1970) Decision and detection limits for calibration curves. Anaytical Chemistry, 42, 849–855. [Google Scholar]
  • 33. Palamar J.J., Le A., Cleland C.M. (2018) Nonmedical opioid use among electronic dance music party attendees in New York City. Drug and Alcohol Dependence, 186, 226–232. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Salomone A., Tsanaclis L., Agius R., Kintz P., Baumgartner M.R. (2016) European guidelines for workplace drug and alcohol testing in hair. Drug Testing and Analysis, 8, 996–1004. [DOI] [PubMed] [Google Scholar]
  • 35. Salomone A., Vincenti M., Gerace E. (2017) Interpretation of NPS results in real hair samples. Toxicologie Analytique et Clinique, 29, 4–10. [Google Scholar]
  • 36. Di Corcia D., D’Urso F., Gerace E., Salomone A., Vincenti M. (2012) Simultaneous determination in hair of multiclass drugs of abuse (including THC) by ultra-high performance liquid chromatography–tandem mass spectrometry. Journal of Chromatography B, 899, 154–159. [DOI] [PubMed] [Google Scholar]
  • 37. Goggin M.M., Nguyen A., Janis G.C. (2017) Identification of unique metabolites of the designer opioid furanyl fentanyl. Journal of Analytical Toxicology, 41, 367–375. [DOI] [PubMed] [Google Scholar]
  • 38. Moody M.T., Diaz S., Shah P., Papsun D., Logan B.K. (2018) Analysis of fentanyl analogs and novel synthetic opioids in blood, serum/plasma, and urine in forensic casework. Drug Testing and Analysis, 10, 1358–1367. 10.1002/dta.2393. [DOI] [PubMed] [Google Scholar]
  • 39. Martucci H.F.H., Ingle E.A., Hunter M.D., Rodda L.N. (2018) Distribution of furanyl fentanyl and 4-ANPP in an accidental acute death: a case report. Forensic Science International, 283, e13–e17. [DOI] [PubMed] [Google Scholar]