Real-World Hypoglycemia Avoidance with a Continuous Glucose Monitoring System's Predictive Low Glucose Alert

Sarah Puhr 1,, Mark Derdzinski 1, John B Welsh 1, Andrew Scott Parker 1, Tomas Walker 1, David A Price 1
PMCID: PMC6477579  PMID: 30896290

Abstract

Background: Programmable and fixed auditory and/or vibratory threshold alerts are essential features of real-time continuous glucose monitoring (rtCGM) systems that provide users time to intervene before the onset of clinical hypoglycemia or hyperglycemia. A sixth-generation rtCGM system from Dexcom, Inc. (G6) includes a new alert that is triggered when an algorithm predicts that an estimated glucose value ≤55 mg/dL will occur within 20 min, allowing users more time to act to avoid hypoglycemia. We examined whether this predictive low glucose alert provided added benefit to traditional low threshold alerts.

Methods: We analyzed glucose values from an anonymized sample of 1424 patients who transitioned to G6 from the preceding fifth-generation system (G5) with no predictive alert. Users with the low threshold alert setting of 70 or 80 mg/dL were evaluated separately. Receiver users, those who disabled the predictive low glucose alert, or those with <30 days of data immediately before or after the transition to G6 were excluded.

Results: Percent time <54, ≤55, <70, and >250 mg/dL fell significantly after the transition to G6, independent of low threshold alert setting. Time in range improved for G6 users with a low threshold alert setting of 70 mg/dL.

Conclusions: Advance warning provided by predictive low glucose alerts may further reduce hypoglycemia among rtCGM-experienced users.

Keywords: Alarms, Alerts, Continuous glucose monitoring, Hypoglycemia, Predictive alerts

Introduction

Real-time Continuous Glucose Monitoring (rtCGM) systems are important diabetes management tools that transmit glucose measurements at regular intervals from a wearable body sensor to a nearby receiver or mobile device through Bluetooth, providing users with actionable information on historic and current glucose concentration and velocity of glucose change. Currently available rtCGM systems offer programmable and fixed auditory and/or vibratory alerts that are activated in response to existing or impending hypoglycemia or hyperglycemia or to rapidly changing glucose concentrations. RtCGM systems differ from the currently available intermittently scanned continuous glucose monitoring (isCGM) system, which provides historic and current glucose data and velocity of glucose change only upon active scanning of a wearable body sensor by a handheld reader. Unlike rtCGM systems, the FDA-approved version of the isCGM system does not provide auditory or vibratory alerts or alarms.1

Randomized controlled clinical trials that examined the efficacy of modern rtCGM in participants with type 1 diabetes (T1D) treated with multiple daily injections (MDI), such as the DIAMOND2–5 and HypoDE trials,6 enabled device alerts and alarms. Given the individual variability in hypoglycemia awareness and risk for hypoglycemia, participant threshold alert settings were personalized by clinicians for these trials. Median (interquartile range) time in measured hypoglycemia at study end was reduced to 4 (0–13) min <50 mg/dL daily in DIAMOND2 and 3.8 (1.1–11.9) min <54 mg/dL daily in HypoDE.6 In contrast, IMPACT studied the efficacy of isCGM—absent alerts and alarms—in participants with T1D treated with MDI and demonstrated that mean (standard deviation) time in measured hypoglycemia was 48.0 (57.6) min daily <55 mg/dL, 18.6 (25.8) min of which occurred at night.7 This difference in time spent in hypoglycemia between rtCGM and isCGM system use supports the premise that low threshold alerts are important components of rtCGM systems.

The goal of low threshold alerts is to warn patients and their caregivers of hypoglycemic events before neuroglycopenia develops. Although the glucose threshold at which cognitive impairment is first detected or observed may vary between individuals, neuroglycopenic symptoms commonly occur at or <55 mg/dL.8 Consensus statements have recommended that low threshold alerts be set at 70 mg/dL to provide patients enough time to intervene before neuroglycopenia/clinical hypoglycemia (i.e., glucose <55 mg/dL).9 Further research demonstrated that a higher low threshold alert setting of 80 mg/dL provided additional time for interventions before the onset of clinical hypoglycemia compared with a low threshold alert setting of 70 mg/dL10; more advance warning may be particularly beneficial to those with impaired awareness of hypoglycemia or a history of problematic hypoglycemia. Davey et al. showed that short-term rtCGM use with a low threshold alert enabled at 80 mg/dL reduced the incidence of hypoglycemia ≤65 mg/dL by 44% in patients with T1D compared with rtCGM use with alerts disabled.11 Analysis of real-world rtCGM data supported these findings and concluded that pediatric and adult patients with a low threshold alert set at ≥80 mg/dL had a higher mean glucose than those with a low threshold alert set at <80 mg/dL.12,13 However, higher low threshold alert settings are only practical in more accurate rtCGM systems, where sensitivity and specificity are high; inaccurate “nuisance” alerts have been cited as a primary reason for rtCGM therapy discontinuation.14–17 The incorporation of a predictive low glucose alert into a clinically accurate rtCGM system may allow a user to set a lower low threshold alert, while decreasing nuisance alerts at “normal” glucose levels and minimizing clinical hypoglycemia. Earlier in silico studies support this hypothesis.18,19

We examined whether a predictive low glucose alert could provide additional advance warning to rtCGM users before the onset of clinical hypoglycemia by evaluating estimated glucose values (EGVs) from an anonymized convenience sample of 1424 patients before and after their transition from an earlier fifth-generation (G5) system with no predictive alert to a comparably accurate20,21 sixth-generation (G6) system with predictive “Urgent Low Soon” (ULS) alert. The ULS alert is enabled by default and triggered when an EGV ≤55 mg/dL is predicted in the next 20 minutes; the activation of a ULS alert inhibits low threshold alert activations in the following 30 minutes to limit alarm fatigue.

Methods

We examined EGVs from an anonymized convenience sample of patients who used the G5 mobile and transitioned to the G6 (both from Dexcom, Inc., San Diego, CA) between May 1, 2018, and August 31, 2018. Patients were included if they had uploaded one or more EGVs during each of the 30 days of data immediately before and after the transition to G6. Only patients using Internet-connected mobile devices, which passively and continuously upload data to the Dexcom Cloud, were included. Users with a low threshold alert setting of 70 or 80 mg/dL (default) were included. Data from patients who disabled the ULS alert on their G6 systems, or those from patients who varied their low threshold alert setting at any time during the study interval, were excluded. A total of 1424 patients met the selection criteria.

Data from users with a low threshold alert setting of 70 mg/dL (n = 658) or 80 mg/dL (n = 766) were evaluated separately. Hypoglycemia exposure, hyperglycemia exposure, and time in range (TIR) were calculated as the percentage of EGVs below, above, and within certain thresholds, respectively. Hypoglycemia exposure was evaluated as the percentage of EGVs ≤55 mg/dL, to match the ULS alert setting, as well as the percentage of EGVs <54 mg/dL, to match consensus guidelines.22 P-values were computed using a two-sided Welch's unequal variance t-test between population means.

Results

The ULS alert remained enabled among >97% of G6 users (not shown) and was triggered less than once daily on average (Table 1). Users who had a low threshold alert set at 80 mg/dL trended toward having fewer ULS notifications daily than those with a low threshold alert set at 70 mg/dL (0.6 ± 0.6 vs. 0.9 ± 0.8 times daily, respectively).

Table 1.

Urgent Low Soon Activations and Mean (Standard Deviation) Continuous Glucose Monitoring–Derived Glycemic Metrics of G5 Users Who Transitioned to G6 with the Urgent Low Soon Alert Enabled

  Low threshold alert setting of 70 mg/dL (n = 658) Low threshold alert setting of 80 mg/dL (n = 766)
G5 G6 P-value G5 G6 P-value
ULS activations/day 0.9 (0.8) 0.6 (0.6)
Mean glucose (mg/dL) 165.4 (33.0) 163.0 (31.4) 0.18 174.2 (30.3) 172.2 (29.8) 0.19
Time <54 mg/dL (%) 1.0 (1.5) 0.6 (1.0) <0.001 0.6 (1.1) 0.4 (0.8) <0.001
Time ≤55 mg/dL (%) 1.2 (1.7) 0.8 (1.2) <0.001 0.7 (1.3) 0.5 (0.9) <0.001
Time <70 mg/dL (%) 3.7 (3.5) 3.0 (3.0) <0.001 2.3 (2.8) 2.0 (2.5) 0.006
TIR (70–180 mg/dL) (%) 61.1 (17.9) 63.0 (18.1) 0.05 57.5 (17.5) 58.7 (18.1) 0.18
Time >180 mg/dL (%) 35.2 (19.1) 34.0 (19.1) 0.23 40.2 (18.4) 39.4 (18.9) 0.37
Time >250 mg/dL (%) 12.6 (12.1) 11.1 (11.2) 0.02 14.4 (11.8) 13.0 (11.5) 0.02

G5, fifth-generation; G6, sixth-generation; TIR, time in range; ULS, “Urgent Low Soon”.

The transition to G6 was associated with significantly reduced biochemical (<70 mg/dL) and clinical (<54 mg/dL) hypoglycemia, independent of low threshold alert setting (Table 1). Compared with intervals of G5 use, the extent of clinical hypoglycemia fell by 40.0% and 33.3% during G6 use for users with a low threshold setting of 70 and 80 mg/dL, respectively. Exposure to biochemical and clinical hypoglycemia was slightly lower overall for users with a threshold alert setting of 80 mg/dL compared with those with a threshold alert setting of 70 mg/dL.

Mean glucose was not significantly different before and after the transition to G6, either for users with a low threshold alert setting of 70 or 80 mg/dL. Mean glucose was slightly higher overall for users with a low threshold alert setting of 80 mg/dL compared with users with a low threshold alert setting of 70 mg/dL, supporting prior studies.12,13 TIR improved for users who had a threshold alert setting of 70 mg/dL (P = 0.05), but was not significantly different before and after the transition to G6 for users who had a threshold alert setting of 80 mg/dL (P = 0.18). Although time spent >180 mg/dL was not significantly different, time spent >250 mg/dL fell significantly after the transition to G6 for users of both low threshold alert settings, suggesting that reduced hypoglycemia did not come at the expense of increased severe hyperglycemia.

Discussion and Conclusion

A pilot randomized controlled study demonstrated that rtCGM but not isCGM therapy reduced exposure to hypoglycemia in patients with T1D and impaired awareness of hypoglycemia, suggesting that threshold alerts unique to rtCGM systems are important for hypoglycemia avoidance.23,24 The current data demonstrate that the incorporation of a predictive low glucose alert into a G6 CGM system further diminishes hypoglycemia exposure, significantly reducing hypoglycemia relative to a CGM system without the predictive alert, independent of threshold alert setting. It is not surprising that users with a threshold alert setting of 80 mg/dL tended to spend less time in hypoglycemia overall compared with those with a threshold alert setting of 70 mg/dL; having a low threshold alert of 80 mg/dL provides more advanced warning to intervene before hypoglycemia.10 Users with a low threshold alert of 80 mg/dL also tended toward fewer ULS notifications daily, likely because they treated impending hypoglycemia after the primary low threshold alert notification. The low rate at which the ULS alert was disabled suggests that it was well tolerated and perceived as beneficial, rather than a nuisance.

The ULS alert-driven reduction in hypoglycemia did not come at the expense of increased hyperglycemia. In fact, more severe hyperglycemia (>250 mg/dL) decreased significantly, which was unanticipated but may be attributable to prevention of symptomatic hypoglycemia and/or fewer episodes of disproportionate carbohydrate intake after ULS alert notification. That said, the methods used to treat hypoglycemia in this study—omitting or reducing insulin doses, eating, or both—are unknown. This is in contrast to low glucose suspend (LGS) or predictive low glucose suspend (PLGS) automated insulin delivery systems, which suspend insulin in response to existing or impending hypoglycemia, respectively. Benefits of LGS and PLGS systems compared with sensor-augmented pump therapy have been demonstrated.25–28

A strength of this study is that it tracked the same 1424 patients for 30 days before and after their transition to G6, making it unlikely that other nondevice-related behaviors were responsible for the observed reductions in hypoglycemia and severe hyperglycemia. However, other device settings, such as high threshold alerts, were not evaluated and may have contributed to the observed results. This study was limited by the narrow population of Dexcom users who were evaluated; among patients who transitioned to G6 immediately after the launch were those whose experience with CGM may not be representative of the population at-large. Moreover, excluding individuals who accessed their CGM data on a receiver alone may have biased the study population. Additional studies are needed to understand the benefit in subpopulations, such as patients with impaired awareness of hypoglycemia, pediatric patients, or patients on Medicare.

Author Disclosure Statement

The authors are employees of Dexcom, Inc.

References

  • 1. Edelman SV, Argento NB, Pettus J, Hirsch IB: Clinical implications of real-time and intermittently scanned continuous glucose monitoring. Diabetes Care 2018;41:2265–2274 [DOI] [PubMed] [Google Scholar]
  • 2. Beck RW, Riddlesworth T, Ruedy K, et al. : Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial. JAMA 2017;317:371–378 [DOI] [PubMed] [Google Scholar]
  • 3. Polonsky WH, Hessler D, Ruedy KJ, et al. : The impact of continuous glucose monitoring on markers of quality of life in adults with type 1 diabetes: further findings from the DIAMOND randomized clinical trial. Diabetes Care 2017;40:736–741 [DOI] [PubMed] [Google Scholar]
  • 4. Ruedy KJ, Parkin CG, Riddlesworth TD, et al. : Continuous glucose monitoring in older adults with type 1 and type 2 diabetes using multiple daily injections of insulin: results from the DIAMOND trial. J Diabetes Sci Technol 2017;11:1138–1146 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Beck RW, Riddlesworth TD, Ruedy KJ, et al. : Effect of initiating use of an insulin pump in adults with type 1 diabetes using multiple daily insulin injections and continuous glucose monitoring (DIAMOND): a multicentre, randomised controlled trial. Lancet Diabetes Endocrinol 2017;5:700–708 [DOI] [PubMed] [Google Scholar]
  • 6. Heinemann L, Freckmann G, Ehrmann D, et al. : Real-time continuous glucose monitoring in adults with type 1 diabetes and impaired hypoglycaemia awareness or severe hypoglycaemia treated with multiple daily insulin injections (HypoDE): a multicentre, randomised controlled trial. Lancet 2018;391:1367–1377 [DOI] [PubMed] [Google Scholar]
  • 7. Bolinder J, Antuna R, Geelhoed-Duijvestijn P, et al. : Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet 2016;388:2254–2263 [DOI] [PubMed] [Google Scholar]
  • 8. Cryer PE: Hypoglycemia, functional brain failure, and brain death. J Clin Invest 2007;117:868–870 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. American Diabetes Association Workgroup on Hypoglycemia: Defining and reporting hypoglycemia in diabetes. Diabetes Care 2005;28:1245–1249 [DOI] [PubMed] [Google Scholar]
  • 10. Peyser TA, Nakamura K, Price D, et al. : Hypoglycemic accuracy and improved low glucose alerts of the latest Dexcom G4 Platinum continuous glucose monitoring system. Diabetes Technol Ther 2015;17:548–554 [DOI] [PubMed] [Google Scholar]
  • 11. Davey RJ, Jones TW, Fournier PA: Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus. J Diabetes Sci Technol 2010;4:1457–1464 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Nakamura K, Walker T, Balo A: Patient data from RT-CGM suggests use of threshold alerts impacts glycemic control. Diabetes 2016;65(S1):A223–A224 [Google Scholar]
  • 13. Welsh JB, Walker T, Liu X: Continuous glucose monitoring screen views and alert settings affect glycemic control in children and adolescents. Diabet Med 2017;34(Suppl. s1):129 [Google Scholar]
  • 14. Tsalikian E, Fox L, Weinzimer S, et al. : Feasibility of prolonged continuous glucose monitoring in toddlers with type 1 diabetes. Pediatr Diabetes 2012;13:301–307 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Tansey M, Laffel L, Cheng J, et al. : Satisfaction with continuous glucose monitoring in adults and youths with type 1 diabetes. Diabet Med 2011;28:1118–1122 [DOI] [PubMed] [Google Scholar]
  • 16. Ramchandani N, Arya S, Ten S, Bhandari S: Real-life utilization of real-time continuous glucose monitoring: the complete picture. J Diabetes Sci Technol 2011;5:860–870 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Polonsky WH, Hessler D: Perceived accuracy in continuous glucose monitoring: understanding the impact on patients. J Diabetes Sci Technol 2015;9:339–341 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Zecchin C, Facchinetti A, Sparacino G, Cobelli C: Reduction of number and duration of hypoglycemic events by glucose prediction methods: a proof-of-concept in silico study. Diabetes Technol Ther 2013;15:66–77 [DOI] [PubMed] [Google Scholar]
  • 19. Facchinetti A, Sparacino G, Guerra S, et al. : Real-time improvement of continuous glucose monitoring accuracy: the smart sensor concept. Diabetes Care 2013;36:793–800 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Welsh JB, Gao P, Derdzinski M, et al. : Accuracy, utilization, effectiveness comparisons of different continuous glucose monitoring systems. Diabetes Technol Ther 2019;21:128–132 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Agiostratidou G, Anhalt H, Ball D, et al. : Standardizing clinically meaningful outcome measures beyond HbA1c for type 1 diabetes: a consensus report of the American Association of Clinical Endocrinologists, the American Association of Diabetes Educators, the American Diabetes Association, the Endocrine Society, JDRF International, The Leona M. and Harry B. Helmsley Charitable Trust, the Pediatric Endocrine Society, and the T1D Exchange. Diabetes Care 2017;40:1622–1630 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Reddy M, Jugnee N, El Laboudi A, et al. : A randomized controlled pilot study of continuous glucose monitoring and flash glucose monitoring in people with Type 1 diabetes and impaired awareness of hypoglycaemia. Diabet Med 2018;35:483–490 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Reddy M, Jugnee N, Anantharaja S, Oliver N: Switching from flash glucose monitoring to continuous glucose monitoring on hypoglycaemia in adults with type 1 diabetes at high hypoglycaemia risk–the extension phase of the I HART CGM study. Diabetes Technol Ther 2018;20:751–757 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Bergenstal RM, Klonoff DC, Garg SK, et al. : Threshold-based insulin-pump interruption for reduction of hypoglycemia. N Engl J Med 2013;369:224–232 [DOI] [PubMed] [Google Scholar]
  • 25. Beato-Vibora PI, Quiros-Lopez C, Lazaro-Martin L, et al. : Impact of sensor-augmented pump therapy with predictive low-glucose suspend function on glycemic control and patient satisfaction in adults and children with type 1 diabetes. Diabetes Technol Ther 2018;20:738–743 [DOI] [PubMed] [Google Scholar]
  • 26. Forlenza GP, Li Z, Buckingham BA, et al. : Predictive low-glucose suspend reduces hypoglycemia in adults, adolescents, and children with type 1 diabetes in an at-home randomized crossover study: results of the PROLOG trial. Diabetes Care 2018;41:2096–2104 [DOI] [PubMed] [Google Scholar]
  • 27. Beato-Vibora PI, Gil-Poch E, Galan-Bueno L, et al. : The incremental benefits of the predictive low-glucose suspend function compared to the low-glucose suspend function as automation against hypoglycemia in sensor-augmented pump therapy. J Diabetes Sci Technol 2018;12:1241–1243 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Diabetes Technology & Therapeutics are provided here courtesy of Mary Ann Liebert, Inc.