Skip to main content
. 2019 Mar 27;24(7):1203. doi: 10.3390/molecules24071203

Table 5.

Multiple reaction monitoring-optimized parameters for quantitation in rat plasma.

Mode Compounds Capillary Voltage (kV) Nebulizer Gas (L/h) Desolvation Gas (L/h) Cone Voltage (V) Quantitation Transition (m/z, Collision (eV)1) Confirmation Transition (m/z, Collision (eV) 1)
ESI+ Gly 2.5 50 800 14 75.97 > 30.19 (8) 75.97 > 48.14 (6)
Asp 2.5 50 800 12 133.97 > 74.03 (14) 133.97 > 88.07 (10)
Glu 2.5 50 800 14 147.97 > 84.11 (16) 147.97 > 130.03 (8)
GABA 2.5 50 800 20 103.97 > 86.99 (10) 103.97 > 68.03 (14)
NE 2.5 50 800 6 169.97 > 152.04 (8) 169.97 > 107.3 (18)
ACh 2.5 50 800 22 146.03 > 87.04 (12) 146.01 > 60.11 (10)
DA 2.5 50 800 12 153.97 > 137 (10) 153.97 > 91.08 (22)
5-HT 2.5 50 800 10 176.97 > 160.02 (8) 176.97 > 132.06 (20)
9-CRA 3 40 500 20 301.29 > 123.17 (22) 301.29 > 161.40 (22)
ESI AA 2.6 50 650 32 303.35 > 259.29 (12) 303.35 > 205.21 (16)
LA 2.8 40 500 36 279.35 > 261.23 (18) 279.35 > 59.13 (18)
cholesterol sulfate 2.6 50 650 62 465.48 > 97.03 (36) 465.48 > 80.00 (80)