Abstract
Background
Shear wave elastography (SWE) imaging have been proposed for characterization of focal liver lesions. We conducted a meta-analysis to evaluate the accuracy and clinical utility of SWE imaging for differentiation of malignant and benign hepatic lesions.
Methods
PubMed, Embase, Web of Science, and the Cochrane Library were systematically reviewed to search for studies published between January 1, 1990, and November 30, 2018. The studies published in English relating to the evaluation the diagnostic accuracy of SWE imaging for distinguishing malignant and benign liver lesions were retrieved and examined for pooled sensitivity, specificity, likelihood ratios, and diagnostic odds ratios, using bivariate random-effects models. The hierarchical summary receiver operating characteristic (HSROC) curve was estimated to assess the SWE imaging accuracy. The clinical utility of SWE imaging for differentiation of malignant liver lesions was evaluated by Fagan plot.
Results
A total of 15 studies, involving 1894 liver lesions in 1728 patients, were eligible for the meta-analysis. The pooled sensitivity and specificity for identification of malignant liver lesions were 0.82 (95% CI: 0.77–0.86) and 0.82 (95% CI: 0.76–0.87), respectively. The AUC was 0.89 (95% CI: 0.86–0.91). When the pre-test probability was 50%, after SWE imaging measurement over the cut-off value (positive result), the corresponding post-test probability for the presence of malignant liver lesions was 82%; the post-test probability was 18% after negative measurement.
Conclusions
SWE imaging showed high sensitivity and specificity in differentiating malignant and benign liver lesions and may be promising for noninvasive evaluation of liver lesions.
Trial registration
The review was registered in the International Prospective Register of Systematic Reviews (PROSPERO): CRD42018104510.
Electronic supplementary material
The online version of this article (10.1186/s12876-019-0976-2) contains supplementary material, which is available to authorized users.
Keywords: Liver neoplasms, Ultrasonography, Elasticity imaging techniques, Diagnosis, Meta-analysis
Background
The development and extensive application of imaging technology has resulted in increased detection of focal liver lesions (FLLs) [1]. Liver cancer is the second most common cause of death from cancer worldwide [2]. Therefore, it is crucial to differentiate malignant from benign liver lesions, despite how extremely challenging that might be amid the wide variation of FLLs.
Ultrasonography is commonly used as the first imaging technique for detecting and distinguishing focal liver lesions because of its availability, low cost, and safety. Contrast-enhanced ultrasound (CEUS) has been shown to be a practicable and accurate method, because it can increase the sensitivity and specificity of detection of focal liver lesions detected by ultrasound to above 90 and 80%, respectively [3, 4]. However, some nodules are still difficult to distinguish, especially in the context of liver cirrhosis; moreover, the adverse effects of contrast agents limits the use of this technology to some extent. Liver biopsy has always been regarded as the gold standard for differentiating malignant and benign lesions. Despite its strengths, liver biopsy is an invasive procedure and could give rise to several complications such as pain, bleeding, and risks of mortality [5, 6].
Ultrasound elastography (USE) is a noninvasive method for the determination of tissue stiffness and the measurement value is usually altered by specific pathological or physiological processes of soft tissues (such as malignancy, inflammation, et al) [7]. Quantitative ultrasound elastography methods currently include acoustic radiation force impulse (ARFI) and transient elastography (TE) techniques [8]. The term “shear wave elastography” (SWE) refers to the technique of detecting shear-wave velocity (SWV) excited by acoustic radiation forces [9]. Both point shear-wave elastography (pSWE) and two-dimensional shear-wave elastography (2D-SWE) rely on the ARFI technique, which uses focused, short-duration acoustic pulses to deform localized tissue and generate shear waves [10]. Although both pSWE and 2D-SWE use ARFI to generate shear waves, pSWE is often referred to as ARFI elastography in some literature and 2D-SWE is referred to as real-time two-dimensional SWE (RT-2D-SWE).
A series of studies evaluate the performance of USE in quantifying tumor stiffness to characterize focal liver lesions [11–13]. In this study, we performed a systematic review and meta-analysis to assess the diagnostic accuracy and clinical utility of SWE imaging in differentiating malignant and benign FLLs.
Methods
The review was registered in the International Prospective Register of Systematic Reviews (PROSPERO, http://www.crd.york.ac.uk/PROSPERO): CRD42018104510. We reported this study in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) of Diagnostic Test Accuracy Studies [14].
Literature search
We searched PubMed, Embase, Web of Science, and the Cochrane Library for studies published between January 1, 1990, and November 30, 2018, to identify articles evaluating SWE for distinguishing malignant and benign liver lesions. The following search strategy including Medical Subject Heading (MeSH) terms and a series of relevant keywords was used: ((liver lesion) OR (liver neoplasm) OR (liver cancer) OR (hepatic lesion) OR (hepatic tumor)) AND ((shear wave elastography) OR (SWE) OR (acoustic radiation force impulse) OR (ARFI) OR (virtual touch tissues quantification) OR (VTQ) OR (ultrasound elastography)) AND ((diagnosis) OR (differentiation) OR (evaluation) OR (distinguishing) OR (discriminate)). We also retrieved the reference lists of related studies manually and searched for other studies that might be omitted in electronic retrieval. The search was limited to journal articles written in English.
Selection criteria
The included studies were required to fulfill the following inclusion criteria: (1) evaluated the performance of SWE imaging for differentiation of malignant and benign liver lesions; (2) used an appropriate reference standard for the diagnosis, such as cytology/histology acquired by biopsy or surgical specimens, or clinical imaging findings (CEUS or computed tomography/magnetic resonance imaging [CT/MRI]); (3) reported data sufficient to calculate the diagnostic accuracy results of SWE imaging (true positive, false positive, false negative, and true negative) for distinguishing liver lesions. The appropriate author was contacted by e-mail if such data were unavailable, and the study would be excluded if no author’s reply. (4) The study included at least 30 patients for the purpose of attaining good reliability. Studies were excluded if they were in a language other than English or were animal experiments. Only the most recent study could be included if the publications used an overlapping cohort of patients.
Data extraction and quality assessment
The studies were retrieved and assessed independently by two reviewers; conflicts were resolved by consulting with a third investigator. The data were extracted independently by two investigators according to the predefined protocol. The following data were extracted from included studies: author, year of publication, country, study design, elastography modality, ultrasonic instrument, number of patients, number of liver lesions, number of the malignant liver lesions, invalid measures, lesion types, nodule size, reference standard for the diagnosis, proportion of cirrhosis, proportion of chronic liver disease, and the cut-off values. In addition, true positive (TP), false positive (FP), false negative (FN), and true negative (TN) could be extracted directly or calculated indirectly. Two investigators independently assessed the quality of the included studies by the Quality Assessment of Diagnostic Accuracy Studies 2 (QUADAS-2) [15], with divergences resolved by consensus.
Statistical analysis
Summary measures of the SWE imaging accuracy (sensitivity, specificity, likelihood ratios, and diagnostic odds ratios) were calculated using a bivariate random-effects model. A hierarchical summary receiver operating characteristic (HSROC) curve was also plotted, and the area under the ROC curve (AUC) was calculated using bivariate model. The inconsistency index (I2) and Cochrane Q statistic were used to estimate the heterogeneity across studies. I2 value greater than 50% or a P value less than 0.10 suggested substantial heterogeneity. Sensitivity analysis was performed by removing studies in which the cut-off values of shear wave velocity (SWV) were less than 2.0 m/s to estimate whether undue influence of a single study was possible.
In addition, univariate meta-regression analyses were performed to explore the sources of potential heterogeneity among studies. The covariates included the following: elastography modality (pSWE vs. 2D-SWE), study location (Asian vs. European or North American), gold standard (histopathology only vs. histopathology and/or others), number of liver lesions (≥100 vs.<100), prevalence of malignant liver lesions (≥50% vs. < 50%), blinded from the results of the reference standard before interpretation of SWE (blinded vs. not blinded), and attrition rate (≥10% vs.<10%). Furthermore, groups were divided into subgroups based on the heterogeneity between studies. We performed a separate analysis of the studies that reported the cut-off value of the SWV/elasticity in FLL, as well as the cut-off value of the SWV/elasticity ratio (FLL to surrounding liver parenchyma).
We performed the Fagan plot to assess the clinical utility of the SWE imaging [16]. We calculated pre-test probabilities of 25, 50, and 75% versus corresponding post-test probabilities, following a positive or negative measurement of SWE. The potential publication bias was inspected by examining a Deeks funnel plot asymmetry test, with P < 0.1 for the slope coefficient indicating significant asymmetry [17].
All statistical analyses were conducted using Stata version 13.0 (StataCorp, College Station, TX) with midas and metandi modules.
Results
Search results and study characteristics
Based on the predefined search strategies, a total of 473 studies were retrieved initially, with 54 duplicates. After eliminating 336 irrelevant studies on review of the titles and abstracts, 83 potentially relevant studies underwent full-text review to determine their eligibility for inclusion. Sixty-eignt articles were excluded for the following reasons: not diagnostic accuracy study (n = 49), not in English (n = 5), overlapping cohort of patients (n = 1), insufficient data (n = 6), the modality of strain elastography (n = 3), small sample size (n = 2), and focus on the comparison between benign tumors (n = 2) [13, 18]. Finally, 15 articles that fulfilled the inclusion criteria were included in this meta-analysis. The flow chart of the study selection appears in Fig. 1.
The characteristics of the 15 included studies are summarized in Table 1 and Additional file 1: Table S1. A total of 1894 liver lesions (673 benign, 1221 malignant) in 1728 patients were investigated. The cut-off value of shear wave speed in pSWE ranged from 1.82 m/s to 2.5 m/s, the cut-off value of Young modulus in 2D-SWE ranged from 20.7 Kpa to 24.43 Kpa, and the cut-off value of SWV/elasticity ration (FLL to surrounding liver parenchyma) ranged from 1.3 to 1.67. All of the 15 included studies were prospective in design. Twelve studies [19–30] used pSWE as the diagnostic imaging modality; the other 3 studies [31–33] performed 2D-SWE imaging modality. Ten studies [19, 20, 22, 23, 26, 28, 29, 31–33] exclusively reported the cut-off value of the SWV/elasticity in FLL, whereas 4 studies [21, 24, 25, 30] reported the cut-off value of the SWV/elasticity and the cut-off value of the SWV ratio. In addition, 1 study [27] reported the sum of the SWV (FLL and surrounding liver parenchyma). There were respectively four articles [22, 23, 30, 33] and six articles [19, 20, 23, 26, 30, 31] reporting the proportion of patients with cirrhosis and the proportion of chronic liver disease in the enrolled population. Eleven studies [19, 20, 22–25, 27, 28, 31–33] reported the number of patients who were excluded from the study due to SWE measurement failure for the reason of the presence of deep-seated lesion (greater than 8 cm from the skin); or with lesion near the heart and large blood vessels; or with lesion smaller than the size of the sampling box (smaller than 1 cm) for SWV measurement; or the patients’ inability to hold their breath properly. The percentage of SWE nonfeasible due to the technical limitations ranged from 1.2 to 26.3%, with the mean percentage of 12.7%. Eleven studies were conducted in Asia (7 in China, 3 in Korea, 1 in India), 3 in Europe, and 1 in North America (Canada).
Table 1.
Author,year | Country | Elastography modality | Instrument | No.of patients | No.of liver lesions (Malignant) | Invalid measures(%) | Lesion types | Reference standard | Cut-off value (SWE/SWV ratio) |
---|---|---|---|---|---|---|---|---|---|
Cho et al., 2010 [19] | Korea | pSWE | Siemens, ACUSON S2000 | 51 | 60(43) | 21.7% | HCC,CCC,LM,haemangioma | histopathology, CT/MRI | 2 m/s |
Davies et al., 2011 [21] | Holland | pSWE | Siemens, ACUSON S2000 | 37 | 45(10) | NR | haemangioma,LM | biopsy,CT/MRI | 2.5 m/s (1.6) |
Shuang-Ming et al., 2011 [20] | China | pSWE | Siemens, ACUSON S2000 | 116 | 128(68) | 22.7% | HCC, CCC,LM, haemangioma,FNH,RN,FFS,FFC,abscess,adenoma,SNN | histopathology, CT/MRI | 2.22 m/s |
Yu et al., 2011 [23] | Canada | pSWE | Siemens, ACUSON S2000 | 89 | 105(41) | 6.1% | HCC,LM,haemangioma,FNH,FFS,FFD,adenoma | biopsy, CT/MRI | 1.9 m/s |
Kapoor et al., 2011 [22] | India | pSWE | Siemens, ACUSON S2000 | 42 | 42(27) | 12.5% | HCC,LM,haemangiomas,lymphomas,FNH,sarcoid,abscesses,FFS | biopsy | 2.5 m/s |
Kim et al., 2013 [26] | Korea | pSWE | Siemens, ACUSON S2000 | 74 | 101(73) | NR | HCC,CCC,LM, hemangioma | pathology, CT/MRI | 2.73 m/s |
Park et al., 2013 [27] | Korea | pSWE | Siemens, ACUSON S2000 | 47 | 47(39) | 20% | HCC,CCC,LM,haemangioma,FNH | histopathology, CT/MRI | 1.82 m/s |
Zhang et al., 2014 [28] | China | pSWE | Siemens, ACUSON S2000 | 156 | 170(112) | 9.7% | HCC,CCC,LM,hemangioma,FNH,RN | histopathology, CT/MRI,DSA | 2.16 m/s |
Guo et al., 2015 [25] | China | pSWE | Siemens, ACUSON S2000 | 134 | 134(55) | 11.0% | HCC,CCC,LM,hemangioma,focal fatty degeneration,FFS,FNH, abscess | pathology, CT/MRI | 2.13 m/s (1.37) |
Lu et al., 2015 [24] | China | pSWE(Elast PQ) | Philips iu22 | 259 | 259(201) | 1.2% | HCC,ICC,LM,hemangioma, FNH, CN | histopathology | 13Kpa (1.3) |
Wu et al., 2016 [29] | China | pSWE | Siemens, ACUSON S2000 | 46 | 55(27) | NR | HCC,LM,hemangioma,FFS,RN | pathology, CEUS, CT/MRI | 2.22 m/s |
Dong et al., 2017 [30] | China | pSWE(ElastPQ) | Philips EPIQ7 | 154 | 154(129) | NR | HCC,ICC,CHC,sarcoma,LM,AML,FNH,hemangioma,IPT | histopathology | 2.06 m/s (1.67) |
Wen-Shuo et al., 2016 [31] | China | 2D-SWE(SuperSonic Imagine) | Aixplorer | 221 | 229(164) | 6.6% | HCC,ICC,CHC,LM,sarcomaneuroendocrine cancer,Unclassified cancer,FNH,hemangioma,angioleiomyolipoma,cavernous vascular tumor,abscess,Inflammatory pseudo-tumor,inflammation,solitary fibroma | histopathology,CEUS | Emax39.60Kpa/ Emean24.43 Kpa |
Gerber et al., 2017 [32] | Germany | 2D-SWE(SuperSonic Imagine) | Aixplorer | 106 | 106(64) | 24.3% | haemangioma,FNH, adenoma,FFS, RN,cholangiofibroma, HCC,CCC, LM | histopathology,CEUS, CT/MRI | Emean20.7 Kpa |
Grgurevic et al., 2018 [33] | Switzerland | 2D-SWE(SuperSonic Imagine) | Aixplorer | 196 | 259(168) | 26.3% | HCC,CCC,LM,haemangioma, FNH | biopsy, CT/MRI | Emean22.3 Kpa |
No number, HCC hepatocellular carcinoma, ICC intrahepatic cholangiocellular carcinoma, CCC cholangiocarcinoma, CHC Combined hepatocellular carcinoma and intrahepatic cholangiocarcinoma, LMN liver metastasis, FNH focal nodular hyperplasia, FFS focal fatty sparing/ nodule of focal fatty sparing, FFD focal fat deposits, FFC focal fatty change, SNN solitary necrotic nodule, RN regenerative node, CN cirrhotic nodule, AML hepatic angiomyolipoma, IPT inflammatory pseudotumor, CT computer tomography, MRI magnetic resonance imaging, CEUS contrast enhanced ultrasound, MRI; DSA digital subtraction angiography, NR not reported
Quality assessment of the included studies
The included studies conformed to the majority of the criteria of the QUADAS-2 and the overall quality of the studies was moderate (Fig. 2 and Additional file 1: Table S2). However, the risk of bias in patient selection was unclear in 3 studies [28, 29, 32] because it was not clear whether patients were consecutively enrolled in the cohort. Risk of bias in the index test was unclear in 3 studies [23, 25, 30] because they did not report whether SWE imaging was performed blinded to the reference standard. In 8 studies [19, 21, 25–30], the risk of bias in the reference standard was unclear because it was unclear whether the reference standard results were interpreted without knowledge of the measurement of SWE. The time interval between SWE imaging and the reference standard was not described in 3 studies [19, 20, 31].
Pooled analysis of diagnostic accuracy and heterogeneity assessment
The pooled sensitivity and specificity of SWE in differentiation of benign and malignant liver lesions were 0.82 (95% CI: 0.77–0.86) and 0.82 (95% CI: 0.76–0.87), respectively (Fig. 3). The positive likelihood ratio (PLR) and negative likelihood ratio (NLR) were 4.57 (95% CI: 3.28–6.38) and 0.22 (95% CI: 0.16–0.29), respectively; the summary diagnostic odds ratio (DOR) was 21.06 (95% CI: 12.14–36.59); and the area under the HSROC was 0.89 (95% CI 0.86–0.91) (Fig. 4a). There was statistically significant heterogeneity among the studies in pooling sensitivity (SEN) (I2 = 71.36%, P < 0.01), specificity (SPE) (I2 = 67.32%, P < 0.01), and DOR (I2 = 99.96%, P < 0.01). We performed a sensitivity analysis in which two studies [23, 27] (the cut-off values of SWV less than 2.0 m/s) were removed, whereas the results were not influenced seriously. We cannot separately calculate the diagnostic accuracy of SWE imaging in differentiation of liver lesions in patients with cirrhosis and non-cirrhosis, because the articles did not report the data or the data could not be extracted.
Accuracy of SWV ratio for the differentiation of benign and malignant liver lesions
The pooled sensitivity, specificity, PLR, and NLR, of the SWV ratio (FLL to surrounding liver parenchyma) for the differentiation of malignant and benign liver lesions were 0.72 (95% CI: 0.59–0.83), 0.82 (95% CI: 0.43–0.97), 4.08 (95% CI: 0.88–18.89), and 0.33 (95% CI: 0.19–0.60), respectively; the summary DOR was 12.17 (95% CI: 1.62–91.31), and the area under the SROC was 0.78 (95% CI: 0.74–0.81) (Fig. 4b).
Meta-regression and subgroup analyses
Univariate meta-regressions were performed to examine the sources of potential heterogeneity in sensitivity and specificity. The results showed that elastography modality, study location, gold standard, blinded interpretation of SWE, and attrition rate were significantly associated with the heterogeneity of sensitivity, whereas number of liver lesions, prevalence of malignant liver lesions, blinded interpretation of SWE, and attrition rate were significantly associated with the heterogeneity of specificity (Additional file 1: Figure S1). We performed subgroup analysis according to the elastography modality. The sensitivity of 2D-SWE was slightly higher compared with pSWE (84% vs. 82%, P < 0.01), whereas there was no significant difference in the specificity for the two modalities (P = 0.18). The sensitivity and the specificity of high attrition rate (≥10%) was higher than low attrition rate (<10%) (82% vs. 80%, P < 0.01; 81% vs. 78%, P < 0.05). Summary data stratified into several subgroups are shown in Table 2.
Table 2.
Covariates | Subgroup | No.of studies | Pooled SEN (95% CI) |
P value | Pooled SPE (95% CI) |
P value | Meta-Regression Joint P Value |
---|---|---|---|---|---|---|---|
modality | 1-pSWE | 12 | 0.82(0.76–0.87) | <0.01 | 0.83(0.77–0.89) | 0.18 | 0.70 |
0-2D-SWE | 3 | 0.84(0.75–0.93) | 0.79(0.66–0.92) | ||||
location | 1-Asian | 11 | 0.82(0.77–0.88) | 0.02 | 0.82(0.76–0.89) | 0.05 | 0.99 |
0- European or North America | 4 | 0.82(0.72–0.92) | 0.81(0.71–0.92) | ||||
gold standard | 1-histopathology | 3 | 0.82(0.72–0.92) | 0.02 | 0.85(0.73–0.96) | 0.22 | 0.89 |
0- histopathology and/or other | 12 | 0.82(0.77–0.88) | 0.81(0.75–0.88) | ||||
number of lesions | 1- ≥ 100 | 10 | 0.84(0.79–0.88) | 0.06 | 0.80(0.73–0.86) | <0.01 | 0.10 |
0-<100 | 5 | 0.76(0.66–0.87) | 0.88(0.80–0.97) | ||||
prevalence of malignant lesions | 1- ≥ 50% | 11 | 0.84(0.79–0.88) | 0.14 | 0.81(0.74–0.88) | 0.02 | 0.29 |
0-<50% | 4 | 0.75(0.63–0.87) | 0.84(0.74–0.94) | ||||
blinded | 1-yes | 11 | 0.82(0.77–0.88) | 0.02 | 0.81(0.74–0.88) | 0.02 | 0.79 |
0-unclear | 4 | 0.82(0.72–0.92) | 0.85(0.75–0.95) | ||||
attrition rate | 1- ≥ 10% | 7 | 0.82(0.77–0.87) | <0.01 | 0.81(0.74–0.88) | 0.04 | <0.01 |
0-<10% | 4 | 0.80(0.75–0.86) | 0.78(0.69–0.87) |
SEN sensitivity, SPE specificity
Assessment of clinical utility of SWE for differentiation of liver lesions
The Fagan plot demonstrated that SWE imaging was fairly effective in distinguishing benign from malignant liver lesions when the pre-test probability was 50%, with 82% probability of malignant disease following a positive measurement, and the probability reduced to 18% when a negative measurement occurred (Fig. 5b). However, when the pre-test probability was 25%, probability was only 60% to differentiate malignant liver lesions correctly following a positive measurement of SWE imaging (Fig. 5a). In addition, the probability of a correct diagnosis rate reached 93% for malignant liver lesions following a positive measurement when the pre-test probability was 75%; nevertheless, the incidence of malignant lesions could reach 39% with a negative measurement (Fig. 5c).
Publication bias
The Deeks’ funnel plot showed that the studies included in the meta-analysis were distributed symmetrically (P = 0.87, Additional file 1: Figure S2), indicating no clear evidence of publication bias.
Discussion
Ultrasound elastography has been explored for application in many fields, it has shown good performance in the evaluation of liver fibrosis and the characterization of superficial organ lesions [34–36]. Similarly, focal liver lesions differ mechanically from surrounding tissues and show changes in their elasticity, with the tendency to increase stiffness in malignant lesions [8, 37]. Previous research have demonstrated that strain elastography is helpful for differentiation of benign and malignant liver lesions [38, 39], and enable us to distinguish accurately between HCC and metastatic adenocarcinoma [40]. SWE have been studied rencently to characterise focal liver lesions and have been proved to have certain clinical value in differential diagnosis [41, 42]. In this systematic evaluation and meta-analysis, we conducted an evidence-based summary of the performance characteristics of SWE in identifying malignant liver lesions.
The meta-analysis synthetic results indicate that SWE imaging has a high accuracy to discriminate malignant liver lesions from benign ones, with sensitivity, specificity, and AUC of 0.82 (95% CI: 0.77–0.86), 0.82 (95% CI: 0.76–0.87), and 0.89 (95% CI: 0.86–0.91), respectively. Furthermore, the results of the Fagan plot show that SWE imaging is valuable in differentiating liver lesions. When the pre-test probability was 50%, the correct diagnosis of malignant liver lesions increased to 82% after the SWE measurement was positive, whereas when the measurement was negative, malignant liver lesions were present in only 18% of patients. Therefore, it was reasoned that SWE imaging is promising and would play an important role in clinical practice. SWE imaging is easily and inexpensively integrated into the ultrasound systems and can be performed with one conventional probe so that the operator can visualize the liver directly, and the region of interest can be positioned manually at the specific location [43]. In addition, SWE imaging is rarely affected by ascites and obesity, because the generated shear waves originate inside the liver rather than from the body surfaces [44]. The results of a prior meta-analysis was also encouraging [45], with the sensitivity, the specificity, and the AUC of 0.86, 0.89, and 0.94, respectively, yet only 8 studies assessing diagnostic the performance of pSWE were included. Another meta-analysis [46] discussing the efficiency of SWE imaging for detecting malignant lesions of the liver also showed good results, with sensitivity, specificity and AUC of 0.82, 0.80 and 0.87, respectively; however, the 9 studies included were all to evaluate the diagnostic performance of pSWE while 2D-SWE studies were not included in this meta-analysis.
FLLs may occur on different liver backgrounds; shear wave velocity (SWV) of SWE imaging in the same type of focal lesion is variable with different surrounding parenchyma. Cirrhosis is an important cause of increased liver parenchyma stiffness. It is noteworthy that the liver is surrounded by a stiff, expandable capsule (Glisson’s capsule), so liver stiffness would increase following any increase in its volume caused by inflammation, cholestasis, or steatosis [44, 47]. Some viewpoints speculate that the ratio of SWV values (FLL to the surrounding liver parenchyma) can more accurately differentiate malignant lesions [48]. However, our meta-analysis found that the cut-off value of the SWV/elasticity in FLL showed superior performance compared to the cut-off value of the SWV ratio, with an AUROC of 0.89 vs. 0.78. The diagnostic performance for the sum of SWV values (FLL and the surrounding liver parenchyma) to differentiate liver lesions is reported in only one article [27], with an AUROC of 0.853.
Despite the results of our study showing promising results, there were still some technical limitations in SWE imaging. First, the maximum detection depth of SWE is limited. The transmission of an acoustic radiation impulse was allowed only up to 10 cm from the skin in pSWE, owing to safety concerns [49]. The maximum detection depth of 2D-SWE would be deeper than pSWE but would depend on the type of instrument and probe. It is difficult to detect the shear wave of a deep-seated lesion because of high attenuation of the signal as it propagates, whereas a shear wave attenuates more slowly due to the Mach cone’s effect in 2D-SWE. Second, the shear wave’s speed that SWE measured was susceptible to motion-related factors, and the accuracy of the results decreased when the lesion was close to the heart and large vessels, or when patients had poor breath-holds. Third, there are wide ranges of stiffness values for focal liver lesions, and some values overlap between the benign and malignant lesions [50]. It seems likely that malignant liver lesions are usually stiffer than benign ones, especially for tumours with pronounced desmoplastic stroma reaction such as intrahepatic cholangiocellular carcinoma, but internal hemorrhage or necrosis in malignant lesions would decrease stiffness. The stiffness of some benign lesions with a high proportion of fibrous tissue such as focal nodular hyperplasia can increase due to fibrous septa and central scar [34]. Finally, there is a high potion of the measurement failure among the included studies (from 1.2 to 26.3%), the diversity of attrition rate may be related to the patient inclusion criteria of the original studies and the proficiency of the operator. In general, patients with FLLs in the right liver lobe and a proximal edge located < 7 cm from the body’s surface would be easily detected, and patients with successful measurement had a lower body mass index (BMI) as compared to patients in which SWE measurement failed. A few studies demonstrate that the performance of SWE for FLL characterization appears limited due to the overlap of the stiffness values between the benign and malignant lesions [50, 51].
There are several potential limitations in our study that should be taken into consideration. First, a considerable amount of heterogeneity was detected among the included studies. Our subgroup analyses found that potential sources of heterogeneity included elastography modality, study location, prevalence of malignant liver lesions, blinded interpretation of SWE, attrition rate, and several other differences, which were unrecorded in these studies, might also contribute to the heterogeneity. Second, because of the limited number of included studies, only 3 studies assessed 2D-SWE, and our study included some relatively small samples of studies; large-sample and multicenter studies on the different kinds of elastography modality in different liver parenchymal settings (such as cirrhosis and non-cirrhosis) are still needed. Third, the summary measurement of diagnostic accuracy pooled the optimal results from each study, with diverse cut-off values, which also may result in overestimation of the performance of SWE imaging. Nevertheless, in clinical practice, one criterion of cut-off value would be required for differentiating liver malignant lesions. Fourth, original studies included in our meta-analysis did not provide size-stratified subgroup analysis results which in turn make it hard to identify appropriate size of lesion to do SWE imaging. Future studies are needed to do subgroup analysis according to tumor size to investigate the proper size. In addition, only English-language articles were included in our study; thus, language bias may have influenced the results. Therefore, considering these limitations, our findings should be interpreted with caution.
Conclusions
In conclusion, our meta-analysis suggests that SWE imaging has favorable diagnostic value for differentiating malignant liver lesions from benign ones. SWE imaging is a promising method undergoing rapid development [52], which could give additional important information to conventional ultrasound, with high sensitivity and specificity in differential diagnosis of liver lesions. It should be emphasized that the SWE imaging assessment of a liver lesion should be interpreted in the context of the patient’s clinical background. Future large-scale studies are required to evaluate the performance of SWE imaging in differentiation of malignant liver lesions and to determine an optimal cut-off value.
Additional file
Acknowledgements
Not applicable.
Funding
This work was supported by the Chinese Government 13th Five-Year Plan (2017ZX10201101), Major Project of Beijing Municipal Science and Technology Committee (D161100000416003, D171100000517003), the National Natural Science Foundation of China (No. 81701984), the National Science and Technology Major Project (2018ZX10302205–005), the NSFC-NIH Biomedical collaborative research program (81761128001), the Capital Health Research and Development of Special Fund (2016–1-2182), Beijing Key Laboratory (No. BZ0089), and the Scientific Research Project of You’an Hospital, CCMU, 2018 (No. YNKTTS20180106).
Availability of data and materials
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Abbreviations
- 2D-SWE
two-dimensional shear wave elastography
- ARFI
acoustic radiation force impulse
- AUC
area under the ROC curve
- BMI
body mass index
- CEUS
contrast-enhanced ultrasound
- FLL
focal liver lesion
- HSROC
hierarchical summary receiver operating characteristic
- NAFLD
Non-alcoholic fatty liver disease
- pSWE
point shear wave elastography
- SWE
shear wave elastography
- SWV
shear wave velocity
- TE
transient elastography
Authors’ contributions
All authors have contributed to the manuscript and have no conflicts of interest to declare. XH, XJH, HW, and FKM conceived and designed the protocol and study. XH and LD identified eligible articles and extracted data. XH and TZ assessed the study quality. XH and JHH performed the analysis with assistance from HC. XH and XJH composed the manuscript and WYL and AXS revised the manuscript. All authors read and approved the final manuscript.
Ethics approval and consent to participate
This is a systematic review and meta-analysis, thus ethics approval and consent to participant are not applicable.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Hao Wu, Phone: +86-010-83997599, Email: whdoc@sina.com.
Fankun Meng, Phone: +86-010-83997599, Email: mengfankun818@126.com.
References
- 1.Marrero JA, Ahn J, Rajender Reddy K. Americal College of G. ACG clinical guideline: the diagnosis and management of focal liver lesions. Am J Gastroenterol. 2014;109(9):1328–1347. doi: 10.1038/ajg.2014.213. [DOI] [PubMed] [Google Scholar]
- 2.Ferlay J, Soerjomataram I, Dikshit R, Eser S, Mathers C, Rebelo M, et al. Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012. Int J Cancer. 2015;136(5):E359–E386. doi: 10.1002/ijc.29210. [DOI] [PubMed] [Google Scholar]
- 3.Strobel D, Seitz K, Blank W, Schuler A, Dietrich C, von Herbay A, et al. Contrast-enhanced ultrasound for the characterization of focal liver lesions--diagnostic accuracy in clinical practice (DEGUM multicenter trial) Ultraschall Med. 2008;29(5):499–505. doi: 10.1055/s-2008-1027806. [DOI] [PubMed] [Google Scholar]
- 4.Seitz K, Strobel D, Bernatik T, Blank W, Friedrich-Rust M, Herbay A, et al. Contrast-enhanced ultrasound (CEUS) for the characterization of focal liver lesions - prospective comparison in clinical practice: CEUS vs. CT (DEGUM multicenter trial). Parts of this manuscript were presented at the ultrasound Dreilandertreffen 2008, Davos. Ultraschall Med. 2009;30(4):383–389. doi: 10.1055/s-0028-1109673. [DOI] [PubMed] [Google Scholar]
- 5.Lindor KD, Bru C, Jorgensen RA, Rakela J, Bordas JM, Gross JB, et al. The role of ultrasonography and automatic-needle biopsy in outpatient percutaneous liver biopsy. Hepatology (Baltimore, Md) 1996;23(5):1079–1083. doi: 10.1002/hep.510230522. [DOI] [PubMed] [Google Scholar]
- 6.Atwell TD, Smith RL, Hesley GK, Callstrom MR, Schleck CD, Harmsen WS, et al. Incidence of bleeding after 15,181 percutaneous biopsies and the role of aspirin. AJR Am J Roentgenol. 2010;194(3):784–789. doi: 10.2214/AJR.08.2122. [DOI] [PubMed] [Google Scholar]
- 7.Shiina T, Nightingale KR, Palmeri ML, Hall TJ, Bamber JC, Barr RG, et al. WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 1: basic principles and terminology. Ultrasound Med Biol. 2015;41(5):1126–1147. doi: 10.1016/j.ultrasmedbio.2015.03.009. [DOI] [PubMed] [Google Scholar]
- 8.Kennedy P, Wagner M, Castera L, Hong CW, Johnson CL, Sirlin CB, et al. Quantitative Elastography methods in liver disease: current evidence and future directions. Radiology. 2018;286(3):738–763. doi: 10.1148/radiol.2018170601. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Bamber J, Cosgrove D, Dietrich CF, Fromageau J, Bojunga J, Calliada F, et al. EFSUMB guidelines and recommendations on the clinical use of ultrasound elastography. Part 1: basic principles and technology. Ultraschall Med. 2013;34(2):169–184. doi: 10.1055/s-0033-1335205. [DOI] [PubMed] [Google Scholar]
- 10.Sarvazyan AP, Rudenko OV, Swanson SD, Fowlkes JB, Emelianov SY. Shear wave elasticity imaging: a new ultrasonic technology of medical diagnostics. Ultrasound Med Biol. 1998;24(9):1419–1435. doi: 10.1016/s0301-5629(98)00110-0. [DOI] [PubMed] [Google Scholar]
- 11.Ronot M, Di Renzo S, Gregoli B, Duran R, Castera L, Van Beers BE, et al. Characterization of fortuitously discovered focal liver lesions: additional information provided by shearwave elastography. Eur Radiol. 2015;25(2):346–358. doi: 10.1007/s00330-014-3370-z. [DOI] [PubMed] [Google Scholar]
- 12.Park HS, Kim YJ, Yu MH, Jung SI, Jeon HJ. Shear wave Elastography of focal liver lesion: Intraobserver reproducibility and elasticity characterization. Ultrasound quarterly. 2015;31(4):262–271. doi: 10.1097/RUQ.0000000000000175. [DOI] [PubMed] [Google Scholar]
- 13.Brunel T, Guibal A, Boularan C, Ducerf C, Mabrut JY, Bancel B, et al. Focal nodular hyperplasia and hepatocellular adenoma: the value of shear wave elastography for differential diagnosis. Eur J Radiol. 2015;84(11):2059–2064. doi: 10.1016/j.ejrad.2015.07.029. [DOI] [PubMed] [Google Scholar]
- 14.McInnes MDF, Moher D, Thombs BD, McGrath TA, Bossuyt PM. The P-DTAG, et al. preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies: the PRISMA-DTA statement. Jama. 2018;319(4):388–396. doi: 10.1001/jama.2017.19163. [DOI] [PubMed] [Google Scholar]
- 15.Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155(8):529–536. doi: 10.7326/0003-4819-155-8-201110180-00009. [DOI] [PubMed] [Google Scholar]
- 16.Hellmich M, Lehmacher W. A ruler for interpreting diagnostic test results. Methods Inf Med. 2005;44(1):124–126. [PubMed] [Google Scholar]
- 17.Deeks JJ, Macaskill P, Irwig L. The performance of tests of publication bias and other sample size effects in systematic reviews of diagnostic test accuracy was assessed. J Clin Epidemiol. 2005;58(9):882–893. doi: 10.1016/j.jclinepi.2005.01.016. [DOI] [PubMed] [Google Scholar]
- 18.Taimr P. Klompenhouwer AJ. Can point shear wave elastography differentiate focal nodular hyperplasia from hepatocellular adenoma. 2018;46(6):380–385. doi: 10.1002/jcu.22603. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Cho SH, Lee JY, Han JK, Choi BI. Acoustic radiation force impulse elastography for the evaluation of focal solid hepatic lesions: preliminary findings. Ultrasound Med Biol. 2010;36(2):202–208. doi: 10.1016/j.ultrasmedbio.2009.10.009. [DOI] [PubMed] [Google Scholar]
- 20.Shuang-Ming T, Ping Z, Ying Q, Li-Rong C, Ping Z, Rui-Zhen L. Usefulness of acoustic radiation force impulse imaging in the differential diagnosis of benign and malignant liver lesions. Acad Radiol. 2011;18(7):810–815. doi: 10.1016/j.acra.2011.01.026. [DOI] [PubMed] [Google Scholar]
- 21.Davies G, Koenen M. Acoustic radiation force impulse elastography in distinguishing hepatic haemangiomata from metastases: preliminary observations. Br J Radiol. 2011;84(1006):939–943. doi: 10.1259/bjr/97637841. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Kapoor A, Kapoor A, Mahajan G, Sidhu BS, Lakhanpal VP. Real-time elastography in differentiating metastatic from nonmetastatic liver nodules. Ultrasound Med Biol. 2011;37(2):207–213. doi: 10.1016/j.ultrasmedbio.2010.11.013. [DOI] [PubMed] [Google Scholar]
- 23.Yu H, Wilson SR. Differentiation of benign from malignant liver masses with acoustic radiation force impulse technique. Ultrasound quarterly. 2011;27(4):217–223. doi: 10.1097/RUQ.0b013e318239422e. [DOI] [PubMed] [Google Scholar]
- 24.Lu Q, Ling W, Lu C, Li J, Ma L, Quan J, et al. Hepatocellular carcinoma: stiffness value and ratio to discriminate malignant from benign focal liver lesions. Radiology. 2015;275(3):880–888. doi: 10.1148/radiol.14131164. [DOI] [PubMed] [Google Scholar]
- 25.Guo LH, Wang SJ, Xu HX, Sun LP, Zhang YF, Xu JM, et al. Differentiation of benign and malignant focal liver lesions: value of virtual touch tissue quantification of acoustic radiation force impulse elastography. Med Oncol. 2015;32(3):68. doi: 10.1007/s12032-015-0543-9. [DOI] [PubMed] [Google Scholar]
- 26.Kim JE, Lee JY, Bae KS, Han JK, Choi BI. Acoustic radiation force impulse elastography for focal hepatic tumors: usefulness for differentiating hemangiomas from malignant tumors. Korean J Radiol. 2013;14(5):743–753. doi: 10.3348/kjr.2013.14.5.743. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Park H, Park JY, Kim DY, Ahn SH, Chon CY, Han KH, et al. Characterization of focal liver masses using acoustic radiation force impulse elastography. World J Gastroenterol. 2013;19(2):219–226. doi: 10.3748/wjg.v19.i2.219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Zhang P, Zhou P, Tian SM, Qian Y, Li JL, Li RZ. Diagnostic performance of contrast-enhanced sonography and acoustic radiation force impulse imaging in solid liver lesions. Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine. 2014;33(2):205–214. doi: 10.7863/ultra.33.2.205. [DOI] [PubMed] [Google Scholar]
- 29.Wu JP, Shu R, Zhao YZ, Ma GL, Xue W, He QJ, et al. Comparison of contrast-enhanced ultrasonography with virtual touch tissue quantification in the evaluation of focal liver lesions. Journal of clinical ultrasound : JCU. 2016;44(6):347–353. doi: 10.1002/jcu.22335. [DOI] [PubMed] [Google Scholar]
- 30.Dong Y, Wang WP, Xu Y, Cao J, Mao F, Dietrich CF. Point shear wave speed measurement in differentiating benign and malignant focal liver lesions. Medical ultrasonography. 2017;19(3):259–264. doi: 10.11152/mu-1142. [DOI] [PubMed] [Google Scholar]
- 31.Tian WS, Lin MX, Zhou LY, Pan FS, Huang GL, Wang W, et al. Maximum value measured by 2-D shear wave Elastography helps in differentiating malignancy from benign focal liver lesions. Ultrasound Med Biol. 2016;42(9):2156–2166. doi: 10.1016/j.ultrasmedbio.2016.05.002. [DOI] [PubMed] [Google Scholar]
- 32.Gerber L, Fitting D, Srikantharajah K, Weiler N, Kyriakidou G, Bojunga J, et al. Evaluation of 2D- shear wave Elastography for characterisation of focal liver lesions. Journal of gastrointestinal and liver diseases : JGLD. 2017;26(3):283–290. doi: 10.15403/jgld.2014.1121.263.dsh. [DOI] [PubMed] [Google Scholar]
- 33.Grgurevic I, Bokun T, Salkic NN, Brkljacic B, Vukelic-Markovic M, Stoos-Veic T, et al. Liver elastography malignancy prediction score for noninvasive characterization of focal liver lesions. Liver international : official journal of the International Association for the Study of the Liver. 2018;38(6):1055–1063. doi: 10.1111/liv.13611. [DOI] [PubMed] [Google Scholar]
- 34.Sigrist RMS, Liau J, Kaffas AE, Chammas MC, Willmann JK. Ultrasound Elastography: review of techniques and clinical applications. Theranostics. 2017;7(5):1303–1329. doi: 10.7150/thno.18650. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Cosgrove D, Piscaglia F, Bamber J, Bojunga J, Correas JM, Gilja OH, et al. EFSUMB guidelines and recommendations on the clinical use of ultrasound elastography. Part 2: clinical applications. Ultraschall Med. 2013;34(3):238–253. doi: 10.1055/s-0033-1335375. [DOI] [PubMed] [Google Scholar]
- 36.Shiha G, Ibrahim A, Helmy A, Sarin SK, Omata M, Kumar A, et al. Asian-Pacific Association for the Study of the liver (APASL) consensus guidelines on invasive and non-invasive assessment of hepatic fibrosis: a 2016 update. Hepatol Int 2017;11(1):1–30. [DOI] [PubMed]
- 37.Sun XL, Yao H, Men Q, Hou KZ, Chen Z, Xu CQ, et al. Combination of acoustic radiation force impulse imaging, serological indexes and contrast-enhanced ultrasound for diagnosis of liver lesions. World J Gastroenterol. 2017;23(30):5602–5609. doi: 10.3748/wjg.v23.i30.5602. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Onur MR, Poyraz AK, Ucak EE, Bozgeyik Z, Ozercan IH, Ogur E. Semiquantitative strain elastography of liver masses. Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine. 2012;31(7):1061–1067. doi: 10.7863/jum.2012.31.7.1061. [DOI] [PubMed] [Google Scholar]
- 39.Gheorghe L, Iacob S, Iacob R, Dumbrava M, Becheanu G, Herlea V, et al. Real time elastography - a non-invasive diagnostic method of small hepatocellular carcinoma in cirrhosis. Journal of gastrointestinal and liver diseases : JGLD. 2009;18(4):439–446. [PubMed] [Google Scholar]
- 40.Kato K, Sugimoto H, Kanazumi N, Nomoto S, Takeda S, Nakao A. Intra-operative application of real-time tissue elastography for the diagnosis of liver tumours. Liver international : official journal of the International Association for the Study of the Liver. 2008;28(9):1264–1271. doi: 10.1111/j.1478-3231.2008.01701.x. [DOI] [PubMed] [Google Scholar]
- 41.Gallotti A, D'Onofrio M, Romanini L, Cantisani V, Pozzi MR. Acoustic radiation force impulse (ARFI) ultrasound imaging of solid focal liver lesions. Eur J Radiol. 2012;81(3):451–455. doi: 10.1016/j.ejrad.2010.12.071. [DOI] [PubMed] [Google Scholar]
- 42.Guibal A, Boularan C, Bruce M, Vallin M, Pilleul F, Walter T, et al. Evaluation of shearwave elastography for the characterisation of focal liver lesions on ultrasound. Eur Radiol. 2013;23(4):1138–1149. doi: 10.1007/s00330-012-2692-y. [DOI] [PubMed] [Google Scholar]
- 43.Berzigotti A, Ferraioli G, Bota S, Gilja OH, Dietrich CF. Novel ultrasound-based methods to assess liver disease: the game has just begun. Digestive and liver disease : official journal of the Italian Society of Gastroenterology and the Italian Association for the Study of the Liver. 2018;50(2):107–112. doi: 10.1016/j.dld.2017.11.019. [DOI] [PubMed] [Google Scholar]
- 44.Ferraioli G, Filice C, Castera L, Choi BI, Sporea I, Wilson SR, et al. WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 3: liver. Ultrasound Med Biol. 2015;41(5):1161–1179. doi: 10.1016/j.ultrasmedbio.2015.03.007. [DOI] [PubMed] [Google Scholar]
- 45.Ying L, Lin X, Xie ZL, Tang FY, Hu YP, Shi KQ. Clinical utility of acoustic radiation force impulse imaging for identification of malignant liver lesions: a meta-analysis. Eur Radiol. 2012;22(12):2798–2805. doi: 10.1007/s00330-012-2540-0. [DOI] [PubMed] [Google Scholar]
- 46.Jiao Y, Dong F, Wang H, Zhang L, Xu J, Zheng J, et al. Shear wave elastography imaging for detecting malignant lesions of the liver: a systematic review and pooled meta-analysis. Medical ultrasonography. 2017;19(1):16–22. doi: 10.11152/mu-925. [DOI] [PubMed] [Google Scholar]
- 47.Tang A, Cloutier G, Szeverenyi NM, Sirlin CB. Ultrasound Elastography and MR Elastography for assessing liver fibrosis: part 2, diagnostic performance, confounders, and future directions. AJR Am J Roentgenol. 2015;205(1):33–40. doi: 10.2214/AJR.15.14553. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Hasab Allah M, Salama RM, Marie MS, Mandur AA, Omar H. Utility of point shear wave elastography in characterisation of focal liver lesions. Expert review of gastroenterology & hepatology. 2018;12(2):201–207. doi: 10.1080/17474124.2018.1415144. [DOI] [PubMed] [Google Scholar]
- 49.Fahey BJ, Nightingale KR, Nelson RC, Palmeri ML, Trahey GE. Acoustic radiation force impulse imaging of the abdomen: demonstration of feasibility and utility. Ultrasound Med Biol. 2005;31(9):1185–1198. doi: 10.1016/j.ultrasmedbio.2005.05.004. [DOI] [PubMed] [Google Scholar]
- 50.Frulio N, Laumonier H, Carteret T, Laurent C, Maire F, Balabaud C, et al. Evaluation of liver tumors using acoustic radiation force impulse elastography and correlation with histologic data. Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine. 2013;32(1):121–130. doi: 10.7863/jum.2013.32.1.121. [DOI] [PubMed] [Google Scholar]
- 51.Heide R, Strobel D, Bernatik T, Goertz RS. Characterization of focal liver lesions (FLL) with acoustic radiation force impulse (ARFI) elastometry. Ultraschall Med. 2010;31(4):405–409. doi: 10.1055/s-0029-1245565. [DOI] [PubMed] [Google Scholar]
- 52.Dong J, Qi X. Liver imaging in precision medicine. EBioMedicine. 2018;32:1–2. doi: 10.1016/j.ebiom.2018.05.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.