Abstract
Introduction
Long lesions contribute to a significant number of percutaneous coronary interventions.
Aim
To assess the efficacy and safety of a novel long-tapered drug-eluting stent (DES) at a 12-month follow-up (FU) in patients with long coronary atherosclerotic lesions
Material and methods
A prospective clinical cohort study was conducted in 32 patients who underwent percutaneous coronary intervention using a BioMime Morph tapered stent (Meril Life Sciences, India). The patients were followed for 3, 6, and 12 months. The safety endpoints were death, myocardial infarction (MI), target lesion revascularization (TLR), target vessel revascularization (TVR), and MACE and/or major bleeding.
Results
Mean lesion length was 48 mm (range: 35–70 mm) measured via quantitative coronary analysis (QCA). In most cases, the target lesion was located in the LAD (68.75%). A GuideLiner catheter (Vascular Solutions Inc., MN, USA) was used in 12.5% of procedures; buddy-wire technique in 9.4% of cases. Bifurcation lesions were treated in 40.6% of cases. Additional stent implantation was needed in 56% of the procedures (25% of cases due to proximal or distal dissection, or due to insufficient stent length in 31% of cases). On 12-month FU we observed 1 TLR (3.1%), 1 TVR (3.1%), and 1 non-cardiovascular death.
Conclusions
The long sirolimus-eluting stent with tapered structure was characterized by good deliverability in long coronary lesions, although in some cases “buddy wire” or extension microcatheter use was necessary. Follow-up at 3, 6, and 12 months showed no significant major adverse cardiovascular events related to the device.
Keywords: long-tapered stent, drug-eluting stent, long coronary lesion, coronary artery disease
Summary
The long sirolimus-eluting stent with tapered structure was characterized by good deliverability in long coronary lesions, although in some cases “buddy wire” or extension microcatheter use was necessary. Previous lesion preparation even with non-compliant balloon is important for successful stent introduction. Operator experience plays an important role. Follow-up at 3, 6, and 12 months showed no significant major adverse cardiovascular events related to the device.
Introduction
Atherosclerosis is a progressive inflammatory disease and among the aging population becomes a challenge for interventional cardiologists [1–3]. Percutaneous treatment is recommended in most patients with significant coronary artery disease (CAD) [4]. Long lesions contribute to a significant number of percutaneous coronary interventions (PCI), which requires the use of multiple overlapping stents, mostly in the context of insufficient stent length or challenging deliverability of longer stents [5]. However, the strategy of using overlapping stents can result in stent misplacement as well as increased risk of restenosis [5]. The newer-generation drug-eluting stents (DES) are designed with reduced strut thickness and have been shown to be efficacious in treating very long coronary lesions [6, 7]. BioMime Morph (Meril Life Sciences Pvt. Ltd., Vapi, India) is a long-tapered sirolimus-eluting stent (SES) system designed for the treatment of very long coronary lesions including PCI of chronic total occlusions [6]. This long-tapered stent has been demonstrated to be associated with good procedural and 1-year clinical outcomes in complex patients with long and diffuse lesions [8].
Aim
The aim of our study was to assess clinical aspects, efficacy, and short-term follow-up safety of the novel long DES with tapered structure, tailored to fit the coronary anatomy in patients with long coronary atherosclerotic lesions.
Material and methods
In this prospective clinical cohort study, we included 32 consecutive patients admitted to the Department of Interventional Cardiology, Jagiellonian University College of Medicine and the John Paul II Hospital, Krakow, Poland between May 2016 and July 2017, who underwent PCI and in whom implantation of a long, tapered DES was performed due to the presence of coronary lesions longer than 35 mm. BioMime Morph is an open-cell cobalt-chromium sirolimus-eluting stent with closed cells on its tapered end and was implanted in all patients who qualified for inclusion in the study. The PCI procedures were performed according to current practice guidelines [4]. All patients provided written informed consent prior to the procedure. Interventional and pharmacological strategies were left to the discretion of the operators. Distal-maximal and proximal-maximal diameter (Dmax) were calculated via quantitative coronary analysis (QCA) as diameters of the vessel on the proximal and distal edge of the baseline lesion, respectively. Optimal PCI result was defined as full lesion coverage without coronary artery dissection and with proper apposition of the stent (less than 20% residual stenosis) in two perpendicular angiographic projections. The clinical outcomes were assessed at 3-month, 6-month, and 12-month follow-ups.
At the 12-month observation, the following endpoints were assessed: death, myocardial infarction, target lesion revascularization (TLR), target vessel revascularization (TVR), and major cerebrovascular event and/or major bleeding, adjudicated by independent researchers blinded to patient characteristics. Major bleeding was defined as bleeding with a need for hospitalization and/or was fatal to the patient. All study procedures were performed according to local bioethics committee guidance as well as the Helsinki Declaration.
Statistical analysis
We expressed continuous variables as mean with standard deviation or median with interquartile range for non-normal distributions. Discrete variables were presented as percentages. All analyses were performed in SPSS version 24 software (IBM).
Results
Patient characteristics
The study population consisted of 32 patients presenting with typical CAD risk factor profiles, with a high percentage of diabetic or impaired glucose tolerance patients, positive history of myocardial infarction (MI) or PCI, and suboptimal treatment of dyslipidemia. 68.8% of the patients were male, with a mean age of 68 ±8 years. Most procedures were performed in stable coronary artery disease (SCAD), with 5 procedures performed in acute coronary syndromes (ACS).
Procedure
In most cases, the target lesion was located in the left anterior descending artery (68.75%). About 60.0% of the procedures were performed using a transradial approach. The long-tapered BioMime Morph DES was successfully implanted in all 32 patients. Stents with a diameter of 3.0–2.5 mm (proximal to distal) were the most frequently used. We observed a need for deployment of an additional stent in 56% of the procedures (31% proximal and 25% distal to the implanted stent), in 25% of cases due to proximal (15.6%) or distal (9.4%) dissection, or due to insufficient stent length in 31% of cases. Mean lesion length was 48 mm, with a range of 35 to over 70 mm. Stent to baseline minimal artery diameter ratio at the distal edge of lesion was 1.26 ±0.27, and was similar in groups with single stent treatment (1.30 ±0.29) and in patients who needed additional stent implantation (1.24 ±0.26; p = 0.60 for t-test). Mean deployed to nominal stent diameter ratio at the distal edge was 0.90 ±0.13. A GuideLiner catheter was used in 12.5% of procedures, while the buddy-wire technique was used in 9.4% of cases. Bifurcation lesions were treated in 40.6% of cases, with predominant anatomy in MEDINA classification 1,1,0. No major side branch, defined as greater than 1 mm diameter, was lost. A transradial approach was used in 60% of the PCI procedures. The long-tapered BioMime Morph drug-eluting stent (40–60 mm) was successfully implanted in all patients. Detailed characteristics of patients and procedures are presented in Table I. Stent implantation images and the QCA analysis prior to and post-stent implantation are presented in Table II and in Figures 1 and 2.
Table I.
Clinical characteristics of patients and procedures
| Risk factors | Value |
|---|---|
| Male sex | 22 (69%) |
| Age, mean ± SD [years] | 68.0 ±8.0 |
| BMI, mean ± SD [kg/m2] | 29.6 ±4.1 |
| Medical history: | |
| Arterial hypertension | 30 (93.8%) |
| Systolic blood pressure, mean ± SD [mm Hg] | 131 ±15 |
| Diabetes | 11 (34.4%) |
| Impaired glucose tolerance | 4 (12.5%) |
| Dyslipidemia | 28 (87.5%) |
| Prior PCI | 14 (45.2%) |
| Prior CABG | 2 (6.3%) |
| Prior myocardial infarction | 14 (43.8%) |
| Indication for invasive coronary angiography: | |
| Stable coronary artery disease | 27 (84.4%) |
| Non-ST elevation myocardial infarction | 3 (9.4%) |
| ST elevation myocardial infarction | 1 (3.1%) |
| Unstable coronary artery disease | 1 (3.1%) |
| Artery treated: | |
| Left anterior descending | 22 (68.8%) |
| Left circumflex | 4 (12.5%) |
| Right coronary artery | 6 (18.8%) |
| Bifurcation lesions | 13 (40.6%) |
| MEDINA (0,1,1) | 3 (9.4%) |
| MEDINA (1,0,1) | 1 (3.1%) |
| MEDINA (1,1,0) | 5 (15.6%) |
| MEDINA (1,1,1) | 4 (12.5%) |
| Implanted stent length: | |
| 40 mm | 10 (31.3%) |
| 50 mm | 13 (40.6%) |
| 60 mm | 9 (28.1%) |
| Implanted stent diameter: | |
| 2.75–2.25 mm | 6 (18.8%) |
| 3.0–2.5 mm | 18 (56.3%) |
| 3.5–3.0 mm | 7 (21.9) |
| 4.0–3.5 mm | 1 (3.1%) |
| Procedure information: | |
| NC balloon pre-dilatation | 8 (25.0%) |
| Stiff guiding wire used | 11 (34.4%) |
| “Buddy-wire” technique | 3 (9.4%) |
| GuideLiner microcatheter use | 4 (12.5%) |
| Additional stent needed | 18 (56.3%) |
| Dissection present | 8 (25.0%) |
| Dissection proximal to first stent | 3 (9.4%) |
| Dissection distal to first stent | 5 (15.6%) |
| Stent length too short | 10 (31.3%) |
| Post-dilatation NC balloon use | 22 (68.8%) |
| IVUS used | 2 (6.3%) |
| Final TIMI 3 flow | 32 (100.0%) |
| Contrast medium, median (IQR) [ml] | 260 (200–370) |
| Radiation dose, median (IQR) [mGy] | 1597 (761–2941) |
BMI – body mass index, SD – standard deviation, Dmax – maximal diameter measured in QCA, PCI – percutaneous coronary intervention, CABG – coronary artery bypass grafting, IQR – interquartile range.
Table II.
QCA analysis of coronary lesions
| QCA analysis | Value |
|---|---|
| Prior to angioplasty: | |
| MLD prior to PCI, mean (SD) [mm] | 0.68 (0.25) |
| Percent diameter stenosis, mean (SD) [%] | 78.0 (13.6) |
| Lesion length, median (min.–max.) [mm] | 50.93 (38.4–80.1) |
| Reference diameter | 2.65 (0.34) |
| Dmax proximal, mean (SD) [mm] | 2.97 (0.30) |
| Dmax distal, mean (SD) [mm] | 2.16 (0.49) |
| After angioplasty: | |
| MLD after PCI, mean (SD) [mm] | 2.37 (0.45) |
| Residual stenosis after PCI [%] | 9.1 (9.9) |
| Dmax proximal, mean (SD) [mm] | 3.13 (0.38) |
| Dmax distal, mean (SD) [mm] | 2.34 (0.46) |
| Proximal stent to vessel lumen diameter ratio, mean (SD) | 0.99 (0.07) |
| Distal stent to vessel lumen diameter ratio, mean (SD) | 1.14 (0.18) |
| Lesion length to stent length ratio, mean (SD) | 1.08 (0.19) |
Figure 1.
Long-tapered stent implantation (40–60 mm)
Figure 2.
Quantitative coronary analysis (QCA) of lesion before (A) and after PCI (B)
Clinical follow-up
During the 3-month follow-up period, 1 (3.1%) patient was readmitted due to the reoccurrence of symptoms, in whom angiographic progression proximally to the previously deployed stent was revealed. The patient required target lesion revascularization (TLR) and no other major adverse cardiac events (MACE) were observed. During the 6-month follow-up period no MACE were observed, although 1 patient was hospitalized for minor lower gastrointestinal bleeding. Data on 12-month follow-up were obtained from 30 patients. The TVR was observed in 1 (3.3%) patient in the setting of stable CAD and another patient (3.3%) experienced ACS with non-target vessel as the culprit lesion. One patient died 8 months after PCI due to multi-organ-failure (MOF) with no evidence of ACS as the primary cause.
Discussion
The new-generation DES have become a routine choice in the treatment of CAD. In current guidelines, full lesion coverage is the preferred PCI strategy [4]. Complete revascularization in this setting often requires using multiple stents in a single vessel, which is a direct consequence of the maximum stent lengths currently available. There is still a lack of data about the treatment of long diffuse coronary lesions using the single or multiple stent strategies.
Many researchers have investigated the problem of stented segment length, demonstrating that longer stent lengths are associated with a significantly higher risk of stent thrombosis and major coronary events [5–8]. Since the length of the segment requiring PCI remains a non-addressable issue, we tried to assess the impact of overlap reduction. It has been shown that stent overlap may be associated with impaired endothelialization. Increased strut density imposes a greater risk of side branch compromise, and additionally increases the risk of local stent fractures and arterial wall injury [9]. Given these factors, stent overlap compared with no overlap is associated with impaired clinical and angiographic outcomes during long-term follow-up [8, 10]. Moreover, standard new-generation DES hold their diameter with the stent length, which in cases of very long lesions may be a cause of stent size mismatch with the distal coronary lumen [11]. Using a long single-tapered DES addresses these aforementioned issues, theoretically giving a reduced risk of complications.
In our experience, device implantation appears to be efficacious independently of clinical presentation [12, 13]. In our case series, an all comers population was treated. We successfully deployed the stent in 100% of the procedures. However, long lesion angioplasty has a significant learning curve, especially when diffused atherosclerosis is present, and operator experience is of tremendous significance.
We observed one TVR in the 3-month follow-up, and another one on 12-month follow-up, with one death (due to MOF) and no major bleeding. Furthermore, no other MACE were seen at 12 months, which in our opinion is an encouraging result.
One should note a need for additional stent implantation in 56% of the procedures, in 25% of cases due to dissection presence (both proximal or distal) and in the remaining 31% of cases due to insufficient length of the first implanted stent. The latter was probably due to a tendency for the operator to underestimate the lesion length or occurrence of unanticipated problems with long-stent insertion into the distal part of the vessel. Given the good deliverability of the device, even if in some cases “buddy wire” technique or extension catheter was needed, it might suggest the use of longer, even 60 mm tapered stents, to avoid overlap in the future. In addition, the expansion ratio of distal artery diameter, over 1.2 times compared to baseline, was associated with a low percentage of distal dissection, which was present in 15% of patients.
The notably low radial access rate, only in 60% of cases, might have been due to the anticipated complexity of procedure, caused by anatomy and length of lesions. However, there was a 100% success rate of stent delivery in both femoral and radial access groups, irrespective of clinical setting. We used the GuideLiner catheter in only 12.5% of cases, while the buddy-wire technique was used in 9.4% of cases. In our opinion, these observations demonstrate acceptable deliverability of the device in an all comers population. Obviously, proper preparation of the lesion with high pressure inflations is crucial to facilitate stent delivery and decrease the need for additional technique application (e.g. “buddy wire” technique, extension microcatheter).
Contrast medium usage in our case series was similar to those in recently emerging clinical studies with long stents [14, 15]. We hypothesize that a lower number of implanted stents should result in lower contrast medium and radiation use. However, there is still a lack of published data involving a direct comparison of multiple stenting versus the long single-stent strategy in terms of contrast expenditure or radiation exposure.
Study limitations
The present study included a relatively small, all-comers patient population and requires a larger cohort study for replication of these results. A second limitation is the lack of a control group to compare the use of contrast medium and radiation exposure.
Conclusions
The long sirolimus-eluting stent with tapered structure was characterized by good deliverability in long coronary lesions, although in some cases “buddy wire” or extension microcatheter use was necessary. Previous lesion preparation even with a non-compliant balloon is important for successful stent introduction. Operator experience in complex PCI plays an important role. Follow-up at 3, 6, and 12 months showed no significant major adverse cardiovascular events associated with the device. Further investigational clinical trials are needed to compare multiple vs. single-stent strategies in patients with long coronary lesions.
Conflict of interest
The authors declare no conflict of interest.
References
- 1.Piepoli MF, Hoes AW, Agewall S, et al. 2016 European Guidelines on cardiovascular disease prevention in clinical practice. Eur Heart J. 2016;37:2315–81. doi: 10.1093/eurheartj/ehw106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Podolec J, Kopec G, Niewiara L, et al. Chemokine RANTES is increased at early stages of coronary artery disease. J Physiol Pharmacol. 2016;67:321–8. [PMC free article] [PubMed] [Google Scholar]
- 3.Podolec J, Niewiara L, Skiba D, et al. Higher levels of circulating naïve CD8 + CD45RA + cells are associated with lower extent of coronary atherosclerosis and vascular dysfunction. Int J Cardiol. 2018;259:26–30. doi: 10.1016/j.ijcard.2018.01.079. [DOI] [PubMed] [Google Scholar]
- 4.Kolh P, Windecker S, Alfonso F, et al. 2014 ESC/EACTS Guidelines on myocardial revascularization. Eur J Cardiothoracic Surg. 2014;46:517–92. [Google Scholar]
- 5.Aoki J, Ong ATL, Rodriguez Granillo GA, et al. “Full metal jacket” (stented length > or =64 mm) using drug-eluting stents for de novo coronary artery lesions. Am Heart J. 2005;150:994–9. doi: 10.1016/j.ahj.2005.01.050. [DOI] [PubMed] [Google Scholar]
- 6.Ruchin PE, Trabattoni D, Fabbiocchi F, et al. Use of multiple overlapping sirolimus-eluting stents for treatment of long coronary artery lesions: results from a single-center registry in 318 consecutive patients. Int J Cardiol. 2009;134:231–7. doi: 10.1016/j.ijcard.2008.02.012. [DOI] [PubMed] [Google Scholar]
- 7.Caputo RP, Goel A, Pencina M, et al. Impact of drug eluting stent length on outcomes of percutaneous coronary intervention (from the EVENT Registry) Am J Cardiol. 2012;110:350–5. doi: 10.1016/j.amjcard.2012.03.031. [DOI] [PubMed] [Google Scholar]
- 8.Räber L, Jüni P, Löffel L, et al. Impact of stent overlap on angiographic and long-term clinical outcome in patients undergoing drug-eluting stent implantation. J Am Coll Cardiol. 2010;55:1178–88. doi: 10.1016/j.jacc.2009.11.052. [DOI] [PubMed] [Google Scholar]
- 9.Lim SY, Jeong MH, Hong SJ, et al. Inflammation and delayed endothelization with overlapping drug-eluting stents in a porcine model of in-stent restenosis. Circ J. 2008;72:463–8. doi: 10.1253/circj.72.463. [DOI] [PubMed] [Google Scholar]
- 10.Degertekin M, Arampatzis CA, Lemos PA, et al. Very long sirolimus-eluting stent implantation for de novo coronary lesions. Am J Cardiol. 2004;93:826–9. doi: 10.1016/j.amjcard.2003.12.018. [DOI] [PubMed] [Google Scholar]
- 11.Timmins LH, Meyer CA, Moreno MR, Moore JE. Mechanical modeling of stents deployed in tapered arteries. Ann Biomed Eng. 2008;36:2042–50. doi: 10.1007/s10439-008-9582-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Podolec J, Niewiara Ł, Baran J, et al. First in Poland, unique 60-mm long single drug eluting tapered stent implantation in a patient with unstable angina. Kardiol Pol. 2017;75:78. doi: 10.5603/KP.2017.0003. [DOI] [PubMed] [Google Scholar]
- 13.Podolec J, Baran J, Niewiara Ł, et al. Intravascular ultrasonography guided 60-mm long drug-eluting tapered stent implantation in a long calcified lesion in a patient with stable coronary artery disease. Adv Interv Cardiol. 2017;13:88–9. doi: 10.5114/aic.2017.66196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Valero E, Consuegra-Sánchez L, Mińana G, et al. Initial experience with the novel BioMime 60 mm-long sirolimus-eluting tapered stent system in long coronary lesions. Eurointervention. 2018;13:1591–4. doi: 10.4244/EIJ-D-17-00616. [DOI] [PubMed] [Google Scholar]
- 15.Premchand RK, Kumar YS. A report of successful procedural, clinical, and angiographic outcomes with a tapered stent of a patient in naturally tapered coronary vessel. J Clin Diagn Res. 2017;11:OD06–7. doi: 10.7860/JCDR/2017/22241.9288. [DOI] [PMC free article] [PubMed] [Google Scholar]


