Thermal Ablation Dose Administration in Swine

Experimental group Acoustic output power (W) Duration R frontal or temporal lobe (s) Duration L frontal or temporal lobe (s) Predicted ablation volume (cm3)
Surgery 1 Acute 4 180 N/A 4.26
Surgery 2 Acute 4 180 120 L: 4.26; R: 2.21
Surgery 3 Acute 6 180 N/A 5.20
Surgery 4a Acute 3 N/A 150 1.63
Surgery 5a Survival 4b 120 N/A
Surgery 6a Survival 3 100 N/A 0.44

aUsed ablator probe with 7-mm element instead of 10-mm element.

bTheraVision software system became unresponsive during HIFU treatment and was manually shut off. Watts delivered during that time is unknown.