Abstract
Purpose
To study the relationship between liquid nitrogen loss and temperature in cryostorage dewars and develop an early-warning alarm for impending tank failure.
Methods
Cryostorage dewars were placed on custom-engineered scales, and weight and temperature data were continuously monitored in the setting of slow, medium, and fast rate-loss of LN2 to simulate three scenarios of tank failure.
Results
LN2 Tank weights and temperatures were continuously monitored and recorded, with a calculated alarm trigger set at 10% weight loss and temperature of − 185 °C. With an intact tank, a 10% loss in LN2 occurred in 4.2–4.9 days. Warming to − 185 °C occurred in 37.8–43.7 days, over 30 days after the weight-based alarm was triggered. Full evaporation of LN2 required ~ 36.8 days. For the medium rate-loss simulation, a 10% loss in LN2 occurred in 0.8 h. Warming to − 185 °C occurred in 3.7–4.8 h, approximately 3 h after the weight-based alarm was triggered. For the fast rate-loss simulation, a 10% weight loss occurred within 15 s, and tanks were depleted in under 3 min. Tank temperatures began to rise immediately and at a relatively constant rate of 43.9 °C/h and 51.6 °C/h. Temperature alarms would have sounded within 0.37 and 0.06 h after the breech.
Conclusions
This study demonstrates that a weight-based alarm system can detect tank failures prior to a temperature-based system. Weight-based monitoring could serve as a redundant safety mechanism for added protection of cryopreserved reproductive tissues.
Electronic supplementary material
The online version of this article (10.1007/s10815-019-01402-3) contains supplementary material, which is available to authorized users.
Keywords: Cryostorage, Cryopreservation, Embryos, Oocytes, Weight
Introduction
Safeguarding cryopreserved embryos and gametes is one of the critical responsibilities of fertility centers and tissue banks. Cryopreservation in liquid nitrogen (LN2) preserves the viability of biological materials by halting their molecular processes [1], enabling their long-term storage [2–4]. Failure to maintain adequate LN2 within a storage tank can result in stored samples thawing and, ultimately, in their total loss. In the case of stored embryos or eggs, such a loss can be catastrophic for affected patients, their partners, and the fertility center [5]. Thus, it is of utmost importance to develop and perfect methods to prevent the depletion of LN2 from cryopreservation tanks.
Cryopreserved samples are typically stored in cryostorage tanks that utilize LN2 to maintain a temperature of − 196 °C [6]. Small storage tanks hold ~ 30–75 L of LN2 within double-walled, insulated metal containers with vacuum chambers between their two walls. Incidents of unintentional thaws of cryopreserved tissues are usually caused by one of three mechanisms [7]: (A) slow rate-loss: in the case of “tank abandonment,” the cryopreservation tank functions normally and LN2 evaporates at the expected rate, but the tank is not refilled in a timely manner, allowing LN2 to be depleted and the temperature to rise; (B) medium rate-loss: a breach to the insulation of the tank, due to either damage to the vacuum seal or someone forgetting to replace the tank’s foam insulation, leads to a faster than normal rate of LN2 evaporation and the supply being depleted prior to the next scheduled refilling; or (C) fast rate-loss: a catastrophic tank failure causes direct leakage of LN2.
Existing methods to prevent unintentional thawing of cryopreserved tissues consist of a combination of manual checks and automated monitoring systems [8]. The most common form of automated monitoring involves using temperature probes, which are placed within tanks and communicate wirelessly to a central monitoring station. When temperatures rise, the probes send an audible alert and, typically, automatically call a predetermined phone number. While this system is helpful, it has some inherent limitations. Firstly, temperatures only start to rise once the LN2 is nearly depleted. There is, in turn, a very limited timeframe for which the LN2 can be replaced before the specimens are completely lost. Secondly, since the sensors note a rise in temperature, the lowest possible storage temperature will have, by definition, been passed before the alarm is sounded. Thirdly, this method requires that sensitive electronic equipment be placed within the harsh conditions of a cryopreservation tank, thereby limiting the lifespan of the probes and increasing the cost for monitoring. Finally, because current monitors require that probes be placed within the LN2, the introduction of a second probe for additional monitoring of samples is cumbersome and represents a replicative rather than truly redundant method. There is, as such, a need for a monitoring method that is redundant to the temperature probe, is positioned externally to the tank, and provides early detection of impending tank failure to safeguard the integrity of cryopreserved tissues.
Regular weight measurements to determine LN2 evaporation rates have been used in the assessment of cryostorage tank viability, but such efforts have yet to be applied to in-use specimen storage tanks where improved quality control practices are needed [8]. We hypothesized that continuous observation of tank weight could be a simple, safe, low-cost method for real-time, automated monitoring of cryostorage tanks. In contrast to temperature alarms, weight-based alarms can be calibrated to sound when a given proportion of LN2 is lost. An alert would thus be triggered before the LN2 supply is completely depleted and the tank’s temperature rises, giving staff more time to refill the tank or transfer samples to a different tank. The scale would also be external to the tank, thereby obviating the need for electronics that could withstand the supercooled environment within the tanks. Finally, a weight-based alarm system could provide a redundancy to the existing safety mechanisms.
Here, we tested the ability of custom-built, web-enabled scales to monitor cryostorage dewars in slow, medium, and fast rate-loss tank failure simulations and demonstrate that a weight-based system provides an effective early-warning and redundant safeguard for cryopreserved specimens.
Materials and methods
Scale design
For all simulations, custom-engineered and manufactured scales were used to measure and record tank weight. The scales had a 400 lb. capacity and data output capabilities, and were attached to a terminal that was connected via Ethernet to a central computer monitoring station. TCP Wedge with custom scripts was used for collection of six fields of data from the networked device. A programmable alarm system was used to record real-time weight information (Fig. 1). Alarms were set to trigger after a 10% weight loss. All experiments were performed with 22.4 L MDD 93/42/EEC compliant passive cryostorage tanks (MVE xc 22/5; Chart Biomedical) containing empty canisters and no patient samples. The tanks involved in this study were fully functioning, in-use tanks made available by consolidating their samples into other tanks. Tanks were placed on the scales and the scales were tared. The tanks were then filled with LN2, and topped-off until the temperature reached equilibrium and LN2 boiling ceased. The temperature within the tanks was simultaneously monitored with probes (Safepoint Scientific) placed near the top of the tanks with calculated alarms using a − 185 °C threshold, as per standard operating practice. Ambient room temperatures were 23–25 °C.
Tank failure scenarios
For the “slow rate-loss” simulations, designed to mimic what would occur if a fully functioning tank were left unattended, two LN2 dewars (Tank A and Tank B) were filled to capacity with LN2, topped-off until temperature reached equilibrium and LN2 boiling ceased, and then left undisturbed while the tank weights and temperatures were monitored continuously, in usual operating conditions. LN2 was allowed to evaporate at the normal rate until all LN2 had evaporated and temperature returned to room temperature (RT). For the “medium rate-loss” simulation, designed to mimic what would occur if the insulating capacity of a tank were compromised, the insulating foam core of the tank neck was removed and the insulating vacuum was eliminated by the creation of a 1/16″ hole in just the outer tank wall. Dewars were then filled to capacity with LN2, topped-off until the temperature reached equilibrium and LN2 boiling ceased and then left alone while the tank weight and temperature were monitored continuously. For the “fast rate-loss” simulation, designed to mimic what would occur if a catastrophic tank breech were to cause LN2 to rapidly spill from the tanks, the 1/16″ hole was continued through the inner tank wall. Dewars were then filled to capacity with LN2, topped-off until the temperature reached equilibrium and LN2 boiling ceased, and LN2 was released at a rate of 0.15 L/s. All simulations were performed in duplicate.
Results
Weight and temperature data collected during the simulations of slow rate-loss from Tanks A and B are presented in Fig. 2a, b, respectively, and in Table 1. The time required for a 10% loss in LN2 was 4.2 and 4.9 days for Tanks A and B, respectively (Table 1). Warming to − 185 °C occurred in 37.8 and 43.7 days for Tanks A and B, respectively. Thus, the weight-based system was able to detect a 10% weight loss between 33.7 and 38.8 days before the temperature alarm threshold was reached (Fig. 2a, b, Table 1). The rate of LN2 loss was 1.05 kg/day and 0.89 kg/day for Tanks A and B, respectively. Because Tank B retained LN2 better, Tank A was used for subsequent studies to minimize the differences between weight- and temperature-based monitoring. Full evaporation of Tank A required 36.8 days (Fig. 2a). Tank temperatures, as recorded by a temperature probe placed a few centimeters below the tank neck, remained stable until LN2 was > 95% evaporated (Fig. 2b). Thereafter, the rate of warming was nearly constant occurred at a rate of and 60.5 °C/day (Fig. 2b). The rate of weight loss for the intact tank can be described by the formula W = − 2.778 t + 101.48, where W is the weight as a percentage of the full tank and t is the time in days (R2 = 0.9999).
Table 1.
Slow rate Tank A (Days) |
Slow rate Tank B (Days) | Medium rate Run 1 (Hours) |
Medium rate Run 2 (Hours) |
Fast rate Run 1 (Hours) | Fast rate Run 2 (Min) | |
---|---|---|---|---|---|---|
Time to 10% ↓ wt | 4.2 | 4.9 | 0.8 | 0.8 | NA | NA |
Time to − 185 °C | 37.8 | 43.7 | 3.7 | 4.8 | 0.37 | 0.06 |
Δ 10% ↓wt and − 185 °C | 33.7 | 38.8 | 3.0 | 4.0 | NA | NA |
For the medium rate-loss simulation, the time required for a 10% loss in LN2 was 0.8 h for runs 1 and 2 (Table 1, Fig. 3a). Warming to − 185 °C occurred in 3.7 and 4.8 h for runs 1 and 2, respectively (Table 1). Thus, the weight-based system was able to detect a 10% weight loss between 3.0 and 4.0 h before the temperature alarm threshold was reached (Fig. 3a, Table 1). The rate of LN2 loss was significantly higher than in the trials using the uncompromised tanks and was the fastest for the loss of the initial 50% of LN2: the first 50% of the volume was lost in 4.3 and 4.7 h for runs 1 and 2, respectively. The remaining 50% of tank volume was lost in 7.5 and 7.8 h for runs 1 and 2, respectively (Fig. 3a and Supplementary Fig. 1). In contrast to the results of the uncompromised tank study, temperatures in the compromised tanks began rising much sooner—just about 1 hour after 10% of LN2 volume was lost. The rate of temperature rise was bi-phasic with a slower rate of warming (3 °C/h and 2.3 °C/h for runs 1 and 2, respectively) while LN2 was present and a much faster rate of warming (48 °C/h and 58.3 °C/h for runs 1 and 2, respectively) after LN2 was depleted (Fig. 3a and Supplementary Fig. 1). The rate of weight loss for the breeched tank can be approximated by the formula W = 0.4447 t2–13.661 t + 100.07 where W is the weight as a percentage of the full tank and t is the time in hours (R2 = 0.9997, calculated from run 1).
For the fast rate-loss simulation, a 10% weight loss occurred within 15 s and tanks were completely depleted in under 3 min (Fig. 3b, Table 1, Supplementary Fig. 2). As shown in Fig. 3b, tank temperatures began to rise immediately and at a relatively constant rate of 43.9 °C/h and 51.6 °C/h. Temperature alarms would have sounded within 0.37 and 0.06 h after the breech (Fig. 3b, Table 1, Supplementary Fig. 2).
Discussion
Existing methods to protect cryopreserved reproductive tissue are reliable and failures are extremely rare [5]. However, as recent cases have highlighted, catastrophic failures are still possible. Thus, there is a need to reevaluate safeguards and consider innovative quality management practices [8].
This study demonstrates that a weight-based, automated alarm system can detect tank failures prior to a temperature-based alarm system. The time interval between the weight-based and temperature-based alarms varies depending on the threshold set for each and the rapidity and nature of the loss. Using standard temperature alarm settings and a 10% weight loss threshold, the weight-based alarm detected anomalies as far as 33 days in advance of a temperature rise. Such a system directly correlates and more precisely determines LN2 level measurements. In cases where there was a higher rate of LN2 loss, the interval was smaller.
When the integrity of the vacuum insulation was left intact, tank temperatures remained below − 190 °C until the LN2 was essentially completely depleted (Fig. 2a) whereas, when the insulating capacity was breached, temperatures rose above the critical point of − 185 °C about 8 days before depletion, when 60% of the LN2 supply still remained (Fig. 3a). This is likely because the intact vacuum layer allowed lingering nitrogen vapor to maintain low tank temperatures during evaporation, while that ability failed when ambient air entered the annular surrounding the inner tank. This is of practical implication because the normal buffer time available to replenish LN2 is drastically and disproportionately shortened in cases of an abrupt loss of insulating capacity.
It should be noted that the anticipated application of this technology is not as a replacement for existing safety monitoring systems, including temperature, but as an additional and redundant safety mechanism. Since existing tanks and their probes can be placed directly on scales, this technology can be readily adopted. For centers that place tanks on wheeled dollies, the scale can be designed to fit onto the dolly and beneath the dewar, or integrated into the dolly itself. In cases of cryopreservation of less critical tissues (e.g., animal cell lines), a weight-based alarm system may be a cost-effective alternative to temperature-based alarms. Given the ability to accurately record the rate of loss of LN2, the scales can also be used for QC testing of tanks, including the qualification of new tanks [8], to confirm that they maintain LN2 as specified. In clinical scenarios, as with temperature alarms, QC procedures with periodic testing of the scale alarm would be necessary.
There were several limitations to this study. The first is that it was conducted with a single brand and make of dewar. There is likely variation in the rate of LN2 loss and temperature rise depending on the particular type, model, and insulation integrity of the dewar. The tank size used (22.4 L) was among the smaller sizes available, and the specific kinetics would likely differ depending on the tank size. The time interval between when weight-based and temperature-based alarms are triggered also depends on where within the tank the temperature probe is placed. Nonetheless, the overall finding that weight anomalies will proceed temperature anomalies is likely to hold true with dewars of differing size and manufacturer. In our experiments, temperature probes were placed near the top of the tank. Placing the temperature probe in a lower position within the tank would undoubtedly result in even more LN2 lost before a temperature probe would detect a change. A similar effect would occur if warmer threshold temperature (e.g., − 175 °C) had been selected [8]. In order to minimize the interval between the weight and temperature-based alarms, the threshold selected for triggering the temperature alarm in this study was − 185 °C. In clinical practice, the normal rate of evaporation of LN2 from each tank would need to be determined in order to select an appropriate threshold weight at which to trigger an alarm.
Performance of the scale-based system in real-world conditions and over time will need to be evaluated, but should significantly improve the response time interval as shown in this study. Any reduction in the interval between alarm and response time could prove critical in limiting the possibility of compromising samples.
Conclusions
Our data show that continuous monitoring of cryostorage tank weight enabled detection of anomalies in advance of temperature rises. As part of an automated alarm system, this would provide earlier detection of a problem and thereby give increased time to remedy the situation before cryopreserved reproductive tissue is lost. When used in conjunction with existing safety systems, including temperature monitoring, a weight-based system would also provide a redundant safety mechanism. While risk can never be completely eliminated, our hope is that by adding additional safeguards to the existing manual and automated safety mechanisms, cryopreservation can become even safer.
Electronic supplementary material
Acknowledgements
The authors thank the members of Columbia University Fertility Center for their assistance in conducting the trial.
Funding information
ZW is partially supported through funding from the National Institutes of Health (grants HD068546 and U19CA179564). ZW is listed as an inventor on patent applications filed by Columbia University relating to this system.
Footnotes
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Cohen J, Inge KL, Wiker SR, Wright G, Fehilly CB, Turner TGJ. Duration of storage of cryopreserved human embryos. J In Vitro Fert Embryo Transf. 1988;5:301–303. doi: 10.1007/BF01132183. [DOI] [PubMed] [Google Scholar]
- 2.Glenister PH, Whittingham DG, Lyon MF. Further studies on the effect of radiation during the storage of frozen 8-cell mouse embryos at -196 degrees C. J Reprod Fertil. 1984;70:229–234. doi: 10.1530/jrf.0.0700229. [DOI] [PubMed] [Google Scholar]
- 3.Goldman KN, Kramer Y, Hodes-Wertz B, Noyes N, McCaffrey C, Grifo JA. Long-term cryopreservation of human oocytes does not increase embryonic aneuploidy. Fertil Steril. 2015;103:662–668. doi: 10.1016/j.fertnstert.2014.11.025. [DOI] [PubMed] [Google Scholar]
- 4.Ueno S, Uchiyama K, Kuroda T, Yabuuchi A, Ezoe K, Okimura T, Okuno T, Kobayashi T, Kato K. Cryostorage duration does not affect pregnancy and neonatal outcomes: a retrospective single-centre cohort study of vitrified-warmed blastocysts. Reprod BioMed Online. 2018;36:614–619. doi: 10.1016/j.rbmo.2018.03.008. [DOI] [PubMed] [Google Scholar]
- 5.Bulletin A. ASRM statement on second report of storage tank failure. In: ASRM Office of Public Affairs Vol 20, 2018.
- 6.Tomlinson M, Morroll D. Risks associated with cryopreservation: a survey of assisted conception units in the UK and Ireland. Hum Fertil (Camb) 2008;11:33–42. doi: 10.1080/14647270701553280. [DOI] [PubMed] [Google Scholar]
- 7.Pomeroy KO. Liquid nitrogen storage tank failure: can we improve the current system? Fertil Steril, www.fertilstertdialog.com, May 19, 2018.
- 8.Schiewe MC, Freeman M, Whitney JB, VerMilyea MD, Jones A, Aguirre M et al. Comprehensive assessment of cryogenic storage risk and quality management concerns: best practice guidelines for ART labs. J Assist Reprod Genet 2018; 35(12):in press. 10.1007/s10815-018-1310-6. [DOI] [PMC free article] [PubMed]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.