Abstract
The combined analysis of circulating cell-free (tumor) DNA (cfDNA/ctDNA) and circulating cell-free (tumor) RNA (cfRNA/ctRNA) shows great promise in determining the molecular profile of cancer patients. Optimization of the workflow is necessary to achieve consistent and reproducible results. In this study, we compared five centrifugation protocols for the optimal yield of both cfDNA/ctDNA and cfRNA/ctRNA. These protocols varied in centrifugation speed, ambient temperature, time, and number of centrifugation steps. Samples from 33 participants were collected in either BD Vacutainer K2EDTA (EDTA) tubes or cell-free DNA BCT® (Streck) tubes. cfDNA concentration and fragment size, and cfRNA concentration were quantitated in all samples by digital droplet PCR (ddPCR) and quantitative PCR (qPCR). The KRAS-mutated ctDNA and ctRNA fraction was determined via ddPCR. In EDTA tubes, the protocol generating both plasma and platelets was found to produce high quality cfDNA and cfRNA concentrations. Two-step, high-speed centrifugation protocols were associated with high cfDNA but low cfRNA concentrations. High cfRNA concentrations were generated by a one-step, low-speed protocol. However, this coincided with a high amount of genomic DNA (gDNA) contamination. In Streck tubes, two-step, high-speed centrifugation protocols also generated good quality, high cfDNA concentration. However, these tubes are not compatible with cfRNA analysis.
Keywords: liquid biopsy, cell-free DNA, cell-free RNA, centrifugation protocol, ddPCR
1. Introduction
The analysis of circulating cell-free (tumor) DNA (cfDNA/ctDNA) as a liquid biopsy has shown tremendous potential in many aspects of oncology, ranging from minimally invasive molecular profiling and treatment monitoring to the detection of resistance mutations [1]. Several limitations of ctDNA-based liquid biopsies have been addressed over the last few years. Complications include the short half-life of cfDNA [2] as well as wild type (WT) DNA contamination from leukocyte lysis [3]. In order to detect the often-low concentration and highly variable ctDNA fraction, several techniques with high analytical sensitivity and specificity have been developed [4]. Furthermore, blood preservation tubes have been designed, which stabilize white blood cells and inhibit nuclease activity [5]. ctDNA has been established as an excellent source of mutation and methylation analysis [4], however, it might be less appropriate for the detection of gene rearrangements. Even though the group of Russo et al. were successful in detecting gene-fusions in ctDNA [6], extensive deep sequencing or break-point encompassing targeted assays are necessary to detect rearrangements. Circulating cell-free (tumor) RNA (cfRNA/ctRNA) analysis could complement ctDNA in determining the molecular profile. Analysis of cfRNA in circulation or embedded in vesicles or tumor-educated platelets (TEPs) has shown great potential in the detection of several cancer-associated aberrations [7,8,9].
Standardization of the liquid biopsy workflow is necessary for clinical implementation [10]. Several studies have been performed to determine optimal pre-analytical conditions for cfDNA [11,12,13] and, to a lesser extent, for cfRNA analysis [14]. However, studies which address the generation of both cfDNA and cfRNA are lacking. In this study, we compared five centrifugation protocols with varying centrifugation speed, temperature, time, and number of centrifugation steps for the optimal cfDNA and cfRNA yield. These protocols were evaluated in two of the most commonly used blood collection tubes, namely BD Vacutainer K2EDTA tubes (EDTA tubes; BD, Erembodegem, Belgium) and cell-free DNA BCT® tubes (Streck tubes; Streck, Biomedical Diagnostics, Antwerp, Belgium). To this purpose, both cfDNA and cfRNA were isolated from healthy volunteers and patients with KRAS-mutated metastatic cancer. CfDNA concentration and fragment size, and cfRNA concentration were quantitated in all samples. The fraction of KRAS-mutated ctDNA and ctRNA was also determined.
2. Results
In this study, five centrifugation protocols were compared for the optimal yield of both cfDNA/ctDNA and cfRNA/ctRNA in either EDTA or Streck tubes (Table 1).
Table 1.
ID | Protocol | Specifications | Details | Temperature | Matrix |
---|---|---|---|---|---|
CPBasic | Basic (Biobank UZA/UAntwerpen) | 1. 10’ – 400 g | RT | Plasma | |
CPAdBasic_P | Basic (adapted) | 1. 10’ – 400 g 2. 1’ – max speed |
2nd centrifugation step after storage at −80 °C | RT | Pellet |
CPAdBasic | Plasma | ||||
CPPlat | Platelet | 1. 20’ – 120 g 2. 20’ – 360 g 3. 5’ – 360 g |
Platelets are washed with PBS in 3rd centrifugation step | RT | Plasma |
CPPlat_P | Platelets | ||||
CPStreck | Streck | 1. 10’ – 1600 g 2. 10’ – 6000 g |
RT | Plasma | |
CPCEN | CEN | 1. 10’ – 1900 g 2. 10’ – 16,000 g |
Blood samples on ice immediately after blood collection | 4 °C | Plasma |
CPAdCEN | CEN (adapted) | 1. 10’ – 1900 g 2. 10’ – 16,000 g |
RT | Plasma |
A smooth braking profile was used in all centrifugation protocols to prevent disruption of the buffy coat layer. Plasma (and platelets) from each centrifugation protocol were snap frozen and stored at −80 °C. CEN: European Committee for Standardization; cfDNA: circulating cell-free DNA; cfRNA: circulating cell-free RNA; CP: centrifugation protocol; Platelets: only RNA isolation; RT: room temperature; 1: first; 2: second, and; 3: third centrifugation step, respectively.
2.1. Total cfDNA Concentration and Characteristics
The cfDNA yield, as well as genomic DNA (gDNA) contamination, were examined per centrifugation protocol. Significant differences in total cfDNA concentration were observed by digital droplet PCR (ddPCR) (p < 0.001 and p = 0.004) and quantitative PCR (qPCR) analysis (p = 0.001 and p = 0.017) in both EDTA and Streck tubes, respectively. Centrifugation protocols were also found to influence DNA integrity in EDTA (p < 0.001) and Streck tubes (p < 0.001), as determined by fragment size analysis. The individual differences between these protocols was further investigated (Figure 1; individual p-values are provided in Table S1).
2.1.1. cfDNA Yield
DdPCR analysis revealed that the highest cfDNA concentrations in EDTA tubes were generated by the basic (CPBasic; one-step, low-speed) centrifugation protocol and the pellet of the adapted basic protocol (CPAdBasic_P). No significant differences were observed between the other centrifugation protocols (Figure 1A). Similar results were observed by qPCR analysis (Figure 1B).
In Streck tubes, ddPCR analysis revealed that the basic centrifugation protocol (CPBasic) alone was responsible for the highest cfDNA concentration. The pellet of the adapted basic protocol (CPAdBasic_P) generated a very variable cfDNA yield and did not differ from the other centrifugation protocols (Figure 1A). More differences were observed with qPCR (Figure 1B). cfDNA concentration seemed to be similar between the basic protocol (CPBasic), the Streck (CPStreck), and adapted CEN (CPAdCEN) protocols. However, while the basic protocol had a significantly higher cfDNA yield than the platelet-generating protocol (CPplat), the plasma (CPAdBasic), and pellet (CPAdBasic_P) of the adapted basic protocol, only the latter (CPAdBasic_P) was found to be lower than that of the Streck (CPStreck) and adapted CEN (CPAdCEN) protocols.
2.1.2. cfDNA Integrity
Next, we investigated DNA integrity (Figure 1C). In EDTA tubes, the pellet of the adapted basic protocol (CPAdBasic_P) generated the highest concentration of long fragments (gDNA contamination), followed by the basic protocol (CPBasic) itself. The plasma of the adapted basic protocol generated significantly fewer long fragments, but still retained higher concentrations than the other centrifugation protocols. Interestingly, no differences in DNA integrity were detected between the platelet generating (CPPlat) and (adapted) CEN (CP(Ad)CEN) protocols, despite their differences in centrifugation speed and time.
Similar results were detected in Streck tubes. The highest concentration of long fragments was detected in the pellet of the adapted basic protocol (CPAdBasic_P), followed by the basic protocol (CPBasic). However, there was no significant difference in DNA integrity between the plasma of the adapted basic (CPAdBasic) and the platelet-generating (CPPlat) protocol. As seen in EDTA tubes, the two-step, high-speed (adapted CEN (CPAdCEN), Streck (CPStreck)) protocols, as well as the two-step, low-speed, platelet-generating protocol (CPPlat), were found to generate good-quality cfDNA (low amounts of long fragments).
Even though comparing the performance of EDTA and Streck tubes was not part of the aim of this study, a significant difference in cfDNA concentration and DNA integrity per centrifugation protocol was detected between these tubes, with the exception of the tube specific protocols (CPCEN and CPStreck). The pellet of the adapted basic protocol (CPAdBasic_P) generated significantly higher cfDNA concentrations in EDTA tubes according to ddPCR (p = 0.015) and qPCR (p < 0.001) analysis. Similar distributions were observed in the basic protocol (CPBasic), although, only by qPCR analysis (p = 0.013). In contrast, DNA integrity was found to be significantly different in all centrifugation protocols, except for the pellet of the adapted basic protocol (CPAdBasic_P) (Figure 2; individual p values are provided in Table S2). Finally, a higher percentage of long fragments was detected in EDTA tubes compared to Streck tubes.
2.2. KRAS-Mutated ctDNA
To investigate the effect of the centrifugation protocol on the ctDNA fraction, we quantified the KRAS-mutated tDNA and ctDNA in spiked samples from healthy volunteers and KRAS-mutated metastatic cancer patients, respectively. In a first phase, spiked KRAS-mutated tDNA was evaluated by ddPCR analysis in samples from 8 participants collected in EDTA (4) and Streck (4) tubes. Next, the same number of samples were collected from KRAS-mutated metastatic cancer patients. As the results from spiked and cancer patient samples differed significantly, it was opted to proceed with the data from cancer patient samples only (Figure 3). All samples were investigated for the presence of KRAS mutations. In 7 out of 8 patients, KRAS-mutated ctDNA could not be detected in the pellet of the adapted basic protocol (CPAdBasic_P). The exception was patient 4, who had a very high mutated KRAS allele frequency (AF; ranging from 20.6 to 51.9% in the other protocols). No significant differences in KRAS ctDNA concentration or AF could be observed between the other centrifugation protocols. However, the sample generated by the basic protocol (CPBasic) of patient 2 was defined as KRAS WT, while KRAS-mutated ctDNA was detected in the plasma generated by the other centrifugation protocols. Despite the number of patients, it seems obvious that both the basic protocol (CPBasic) and the pellet of the adapted basic protocol (CPAdBasic_P) are not suitable for ctDNA analysis.
2.3. Total cfRNA Concentration
Next, the total cfRNA yield was determined using ddPCR (Figure 4). In EDTA tubes, significant differences in cfRNA concentration were observed between the different centrifugation protocols (p < 0.0001; individual p-values are provided in Table S3). As expected, platelets (CPPlat_P) provided the highest cfRNA yield, while the platelet-rich plasma (PRP) basic protocol (CPBasic) still generated more cfRNA than the other centrifugation protocols. No differences could be observed between the pellet (CPAdBasic_P) and the platelet-poor plasma (PPP) of the adapted protocol (CPAdBasic) and the platelet-generating protocol (CPPlat). These protocols still generated higher cfRNA concentrations compared to the adapted and original CEN protocols (CP(Ad)CEN), which were also found to be similar in cfRNA yield. In Streck tubes, no cfRNA or only very low concentrations could be detected across all centrifugation protocols.
2.4. KRAS-Mutated ctRNA
To further investigate the effect of the centrifugation protocol on the ctRNA fraction, the samples were also screened for the presence of KRAS mutations. In EDTA tubes, KRAS-mutated ctRNA could only be detected in the platelets (CPPlat_P) (Figure 5) in 3 out of 4 KRAS-mutated metastatic cancer patients (as depicted in Figure 3). In Streck tubes, no KRAS-mutated ctRNA could be detected in any centrifugation protocol, as only extremely low cfRNA concentrations, or no cfRNA at all, were generated.
3. Discussion
The purpose of this study was to determine the optimal centrifugation protocol for the generation of high-quality cfDNA and cfRNA in two commonly used blood collection tubes; EDTA and Streck tubes. Despite the relatively small sample size of this study, significant differences in cfDNA and cfRNA yield and quality could be observed. In EDTA tubes, the plasma and platelets from the platelet-generating protocol were found to produce both good quality cfDNA and high cfRNA concentration. cfDNA analysis in Streck tubes generated similar results in all centrifugation protocols, however, at best, very low levels of cfRNA could be generated.
In EDTA tubes, the cfDNA characteristics, namely cfDNA concentration and contamination with long DNA fragments, from the platelet-generating protocol were found to be similar to those of the adapted and original CEN protocols. KRAS-mutated ctDNA concentration and AF also seemed similar. While the platelet-generating protocol consists of two slow-speed centrifugation steps, the (adapted) CEN protocols have a two-step, high-speed procedure. Most studies recommend a double centrifugation protocol, consisting of an initial slow centrifugation to separate plasma from buffy coat, followed by a fast centrifugation to clear any remaining cellular material and generate high quality cfDNA [12,13,15]. Although there is no consensus with regards to actual centrifugation speed [16,17], both centrifugation steps described in these studies, together with the adapted and original CEN protocols, are much faster in comparison to the platelet-generating protocol. We speculate that the longer centrifugation time (20 min versus 10 min) might be responsible for the low contamination with long DNA fragments, despite the slow centrifugation speed. In contrast, the plasma of the adapted basic protocol, generated by a first slow-speed centrifugation step, followed by a 1-min high-speed centrifugation post-thaw, was found to contain significantly more long DNA fragments. Hence, the interplay between centrifugation speed and time probably has a stronger influence on cfDNA generation than these other factors alone. We did not detect an influence of temperature, as both cfDNA characteristics and ctDNA detection were similar between the adapted and original CEN protocols. Furthermore, no differences between these four centrifugation protocols could be observed in mutated ctDNA concentration and AF. It is important to note that, despite our limited number of patients with detectable KRAS-mutated ctDNA, the basic protocol and the pellet of the adapted basic protocol were found to be less suitable for cfDNA analysis. The pellet of the adapted basic protocol seems to consist almost exclusively of long DNA fragments. This is to be expected, as previous research highlights that the second centrifugation step removes any remaining cellular material [12]. Hence, the resulting plasma will mostly contain the cfDNA fraction. Our data support this, as KRAS-mutated ctDNA was only detected in the pellet of a patient with a very high ctDNA concentration. The mutated KRAS ctDNA AF in the other centrifugation protocols ranged between 20.6% and 50.9%. This is highly unusual, as ctDNA is most often present in very low (<1%) percentages [2]. The plasma of the basic protocol contains both the remaining cellular material as well the cfDNA fraction, as indicated by the high percentage of long fragments. In this manner, the ctDNA is diluted by WT DNA, resulting in the classification of one patient sample as WT despite the use of the highly sensitive ddPCR technique [18]. Furthermore, there are several disorders and physiological processes, such as autoimmune disorders, exhaustive exercise, or trauma, that are known to contribute to the total cfDNA concentration [19]. This might explain the low AF detected in all centrifugation protocols from patient 7 (Figure 3). In these cases, further dilution of the mutated ctDNA with WT gDNA would be detrimental to analytical success.
In Streck tubes, cfDNA analysis revealed similar results as in EDTA tubes. Interestingly, no difference in the generated plasma volume was observed in EDTA or Streck tubes. This was unexpected, especially in the case of the Streck tubes. In a multicenter study, we observed a negative influence on plasma volume upon longer storage of blood samples in Streck tubes. Increased centrifugation speed was found to negate this effect [20,21], which is recommended by Streck. As plasma volume is directly associated with cfDNA concentration [22], it is doubtful whether protocols based on low centrifugation speed, such as the platelet-generating and (adapted) basic protocol, will generate sufficient material for the detection of low abundant aberrations. Even though there were no significant differences in cfDNA characteristics detected between the platelet-generating protocol and the Streck and adapted CEN protocol, it seems that there was still a slight shift towards more long fragments in the former. This might explain the fact that no significant differences were observed between this protocol and the plasma of the adapted basic protocol. The Streck and adapted CEN protocols seem to be the most appropriate centrifugation protocols for cfDNA analysis in Streck tubes. Both are two-step, high-speed protocols, with a first centrifugation step of <2000 g for 10 min. However, the second step differs significantly (6000 vs. 16,000 g). This corresponds to previously reported data in which no difference in cfDNA concentration was observed between a second centrifugation step of 3000 g or >10,000 g [17]. Hence, there is no need to have access to a full spectrum of centrifuges with higher centrifugation speeds. Interestingly, significant differences in percentage of long fragments were observed between centrifugation protocols of EDTA vs. Streck tubes, with the exception of the pellet of the adapted basic protocol. Van Dessel et al. reported a background level of leukocyte lysis, which is responsible for low levels of gDNA [11]. Hence, leukocyte lysis might be diminished by the cell stabilizing effect of Streck tubes [21].
In EDTA tubes, the highest cfRNA concentration, as well as KRAS-mutated ctRNA, were detected in the platelet fraction of the platelet-generating protocol. Our data corresponds to previous research in which platelets were highlighted to be an excellent source for tumor RNA (tRNA) [7,8]. Nilsson et al. also reported a higher sensitivity for the detection of tumor-associated aberrations in platelet RNA than in the plasma of the platelet-generating protocol. Even though no KRAS-mutated ctRNA could be retrieved from the other centrifugation protocols, the basic protocol might be the best alternative in the absence of platelets. The proposed workflow for PPP consists of two centrifugation steps prior to freezing; namely to remove the bulk of circulating cells and the residual platelets [23]. Hence, the basic protocol most likely still contains some platelets due to its single slow-speed centrifugation procedure. Furthermore, a single freeze/thaw cycle of plasma samples containing (residual) platelets has been demonstrated to release significant quantities of miRNA and platelet microparticles [24]. In contrast, the adapted and original CEN protocols both generate PPP, resulting in the lowest cfRNA concentrations of all protocols. Similar cfRNA yields were detected between the plasma of the platelet-generating protocol, and the pellet and plasma of the adapted basic protocol, despite differences in centrifugation speed and time. These findings also highlight the importance of both factors on cfDNA and cfRNA generation. It is important to note that in this study, the entire platelet fraction was used for cfRNA isolation. This fraction was generated from the entire blood sample (10 mL). In the case of plasma samples, only 200 µL of plasma could be used, which corresponds to approximately 500 µL of blood. The pellet of the adapted basic protocol was re-dissolved in 1 mL of nuclease-free water, of which 200 µL was also is used. Hence, this represents approximately 2 mL of blood. By increasing the input volume, higher cfRNA concentrations might be obtained [25]. Recently, RNA isolation kits with increased input volumes have been developed and are currently being tested, among other pre-analytical variables, in a Belgian extracellular RNA quality control (exRNAQC) study [26]. These findings will aid in the standardization of the cfDNA- and cfRNA-based liquid biopsy workflow. Another issue that needs to be addressed is the low KRAS-mutated ctRNA concentration and AF compared to ctDNA. This might be the result of the reactivity of isolated platelets, or contamination with residual cells and microparticles [27]. Red blood cells have been shown to contain a plethora of RNAs [28], which might interfere with these results. Furthermore, centrifugation alone might not be sufficient to generate a pure platelet fraction. The group of Rikkert et al. demonstrated that the pellet of the platelet generating protocol not only contains platelets (71%), but also large (tumor-derived) extracellular vesicles (EVs; 22%), circulating tumor cells (CTCs), smaller (tumor-derived) EVs, and cfDNA (<3%) [29]. Even though we were able to detect KRAS-mutated ctRNA in platelets in 3 out of 4 patients with detectable KRAS-mutated ctDNA, further research is necessary to increase the quality of cfRNA from platelets and plasma.
However, the advantage of the platelet-generating protocol is the generation of sufficient cfRNA concentration and high quality cfDNA from 10 mL of blood. In this manner, the amount of molecular information would be maximized with limited material from the patient. The true potential of cfRNA analysis lies in the detection of gene fusions and amplifications, for which cfDNA is a less appropriate matrix [7]. An example of the potential of this approach is the real-time follow up of Anaplastic Lymphoma Kinase (ALK) rearranged non-small cell lung cancer (NSCLC). ALK-rearranged NSCLC is a validated molecular target of ALK tyrosine kinase inhibitors (TKIs) [30]. Sadly, ALK TKI resistance is an issue, with acquired ALK mutations responsible for a large subset of resistant cases [31]. In these cases, it is an advantage to have the cfRNA as well as the cfDNA. The high quality cfDNA can be screened for any resistance mutations, while translocation can be detected in the platelet cfRNA by targeted assays [7] or via sequencing [32].
In contrast, little to no cfRNA was generated in Streck tubes, irrespective of centrifugation protocol. It is likely that the cell-free DNA BCT® (Streck) tubes do not sufficiently stabilize cfRNA. This could be expected, as cell-free RNA BCT® tubes have already been developed [33]. The low yield might also indicate that the preservative of the Streck tubes is not compatible with the cfRNA isolation kit we used. Presently, there are studies ongoing in which the efficiency of several cfRNA tubes and isolation techniques are being evaluated [26,34]. A blood collection tube which preserves both cfDNA and cfRNA would maximize the detection of several cancer-associated aberrations based on one blood sample.
4. Materials and Methods
4.1. Study Design
In total, 15 patients with KRAS-mutated advanced cancer and 18 healthy volunteers were included. Each participant contributed five 10 mL blood samples, which were collected in either EDTA (n = 95) (BD, Erembodegem, Belgium) or Streck (n = 70) (Streck, Biomedical Diagnostics, Antwerp, Belgium) tubes. In the former, samples were processed within 2 h after blood collection, whereas Streck tubes were left at room temperature for 24 h before plasma generation. The study was conducted according to the Declaration of Helsinki, and ethical committee approval (B300201422715) was obtained from the ethical committee of Antwerp University Hospital (UZA). All participants provided written informed consent.
This study consisted of three phases (Table 2). Healthy volunteers were included in phase one. In phase two, the blood samples from healthy volunteers were spiked with KRAS-mutated tumor DNA in order to simulate ctDNA. DNA was isolated from KRAS-mutated NSCLC formalin-fixed paraffin embedded (FFPE) tissue samples, because in contrast to high-quality cell line-derived DNA, the fragmented and partly degraded properties of FFPE-derived DNA are a more suitable ctDNA alternative. The resulting spike mix consisted of 900 mutated DNA copies per µL, of which 1 µL was injected into the blood samples immediately after blood collection. The third phase consisted of known KRAS-mutated NSCLC, colorectal cancer (CRC), and pancreatic ductal adenocarcinoma (PDAC) patients to validate our findings. Total cfDNA and cfRNA concentration and cfDNA fragment analysis were investigated in all phases.
Table 2.
Phase | Participants | EDTA (N = 19) | Streck (N = 14) | ||
---|---|---|---|---|---|
cfDNA Samples | cfRNA Samples | cfDNA Samples | cfRNA Samples | ||
1 | Healthy volunteers | n° = 48 | n° = 55 | n° = 12 | n° = 7 |
| |||||
2 | Healthy volunteers—spiked with KRAS-mutated DNA | n° = 24 | n° = 28 | n° = 24 | n° = 28 |
| |||||
3 | KRAS-mutated metastatic cancer patients | n° = 42 | n° = 47 | n° = 48 | n° = 53 |
|
cfDNA: circulating cell-free DNA; cfRNA: circulating cell-free RNA; ctDNA: circulating cell-free tumor DNA; ctRNA: circulating cell-free tumor RNA; N: number of subjects; n°: number of samples; tDNA: tumor DNA.
Plasma was generated from matched blood samples with different centrifugation protocols varying in centrifugation speed, temperature, time, and number of centrifugation steps (Table 1). This study was performed in collaboration with the Biobank UZA/UAntwerpen. Therefore, the standard (CPBasic) protocol from the Biobank UZA/UAntwerpen was included. As most studies recommend two centrifugation steps [12,13,15], an adapted version of this protocol was also added, of which both the plasma (CPAdBasic) and resulting pellet (CPAdBasic_P) were analyzed. We also selected the “Isolated circulating cell-free DNA from plasma protocol, CEN/TS 16835-3” as advised by the European committee for standardization (CEN; CPCEN). This protocol requires for blood samples to be put on ice immediately after blood collection and centrifugation at 4 °C. The adapted protocol (CPAdCEN) can be completely performed at room temperature. As Streck tubes are not compatible with low temperatures, we included the manufacturer’s instructed protocol for Streck tubes (CPStreck) [21]. Lastly, we selected a centrifugation protocol which produces both plasma (CPPlat) and platelets (CPPlat_P) [7]. Plasma samples were stored at −80 °C by the Biobank UZA/UAntwerpen (Antwerp, Belgium; ID: BE71030031000; Belgian Virtual Tumourbank funded by the National Cancer Plan, BBMRI-ERIC; No. Access: 1; Last: 13 June 2018) [35].
4.2. cfDNA Isolation and Analysis
cfDNA isolation was performed using the Maxwell RSC ccfDNA large volume plasma kit (Promega, Leiden, The Netherlands) according to the manufacturer’s protocol. Detailed procedures are described elsewhere [20]. Plasma samples were processed immediately after thawing, with the exception of the adapted basic protocol. Prior to processing, the plasma was centrifuged at maximum speed for one minute. The plasma was separated from the resulting pellet, which was hereafter dissolved in 1 mL nuclease-free water. For cfRNA, 0.5 mL of all plasma samples and the dissolved pellet were reserved.
Samples were screened with the ddPCR KRAS Screening Multiplex Kit (Bio-Rad, Temse, Belgium) to determine total cfDNA concentration and KRAS mutational status. In total, 10 µL of template was used per sample. Detailed procedures can be found elsewhere [36]. The concentration of these samples was not determined prior to ddPCR analysis. Total cfDNA was also quantified by a qPCR assay consisting of an 82-base pair (bp) Long Interspersed Nuclear Elements (LINE-1) sequence [37]. The samples were brought to the same concentration for cfDNA sample quality assessment using a 425:82 bp ratio. The 425 bp amplicon was designed to encompass the 82 bp amplicon; representing mainly genomic DNA released from blood cells and cfDNA, respectively. Hence, a high ratio is indicative of gDNA contamination [38]. A long fragment contamination curve consisting of cell line and digested FFPE tissue material was used to determine the percentage of long fragment contamination. All samples were run in duplicate. qPCR mixtures were assembled by 4 µL LightScanner Mastermix (Bioké, Leiden, The Netherlands), 1 µL of 1 µM primer sets, and 5 µL (diluted) cfDNA. Cycle conditions were 95 °C for 2.5 min, then 95 °C for 30 s, 58 °C for 30 s, 72 °C for 1 min cycled 45 times, and 72 °C for 3 min. The LightCycler 480 SW (Roche, Vilvoorde, Belgium) was used to calculate the quantitation cycle (Cq).
4.3. cfRNA Isolation and Analysis
cfRNA was isolated from 200 µL of plasma or directly from platelets and converted to complement DNA (cDNA) using the miRNeasy Serum/Plasma Kit and miScript II RT Kit, respectively, as per manufacturer’s instructions (Qiagen, Antwerp, Belgium). Samples were analyzed using the ddPCR KRAS Screening Multiplex kit in the same manner as the cfDNA samples, with the exception of the platelet cDNA. Only 1 µL of template was used instead of 10 µL.
4.4. Statistical Analysis
The Kruskal–Wallis test was performed to ascertain whether there was a significant difference between the centrifugation protocols and blood collection tubes for total cfDNA and cfRNA yield, percentage of long DNA fragments, and KRAS-mutated ctDNA fraction. The Mann–Whitney U test was used to determine the effect of the individual centrifugation protocols. Statistical analysis was performed using IBM SPSS Statistics version 24.0 (IBM, Brussels, Belgium). p-values of <0.05 were considered statistically significant. Figures were made with Graphpad Prism7 (Graphpad Software Inc., La Jolla, CA, USA).
5. Conclusions
To conclude, the platelet-generating protocol provides high quality cfDNA and cfRNA concentrations in EDTA tubes. However, the selected centrifugation protocol will be dependent on the downstream analysis. Therefore, we recommend this protocol when looking at cfRNA or multiple aberration types (e.g., translocations and acquired resistance mutations). In the case of cfDNA screening alone, the adapted CEN protocol (two-step, high-speed) would be more appropriate, as it is a shorter and less laborious procedure. In terms of biobanking, where blood samples can be collected without a definite downstream analysis, the basic protocol (one-step, low-speed) might offer the most possibilities. In this case, the generated platelet-rich plasma will permit cfRNA isolation, while a second high-centrifugation step will clear away most of the remaining cellular material. Cell-free DNA BCT® (Streck) tubes should only be used for cfDNA analysis. Both the Streck and adapted CEN protocols generate high quality cfDNA. Further research might reveal blood preservation tubes which generate both good quality cfDNA and cfRNA from one blood sample.
Acknowledgments
The authors would like to thank everybody involved in this study, especially the patients who were willing to provide these blood samples. Special thanks to Franscica Kohl, Peggy De Clercq, Sven De Keersmaeker, Tim Mertens, and all the staff from Oncology for collecting these samples. Furthermore, the authors would also like to thank the Biobank UZA/UAntwerpen for the close collaboration. Lastly, special thanks to Dr. Abraham Lin for proofreading our manuscript.
Supplementary Materials
The following are available online at https://www.mdpi.com/2072-6694/11/4/458/s1, Table S1: Statistical analysis of total cfDNA, Table S2: cfDNA integrity of EDTA vs. Streck tubes, Table S3: Statistical analysis of total cfRNA in EDTA tubes.
Author Contributions
Conceptualization: L.S., K.Z. and P.P.; Methodology: L.S., K.Z., J.J. and K.D.W.; Software: L.S., K.Z., J.J. and K.D.W.; Validation: L.S., K.Z., J.J. and K.D.W.; Formal Analysis: L.S., K.Z., J.J. and K.D.W.; Investigation: L.S.; Resources: L.S. and K.Z.; Data Curation: L.S., K.Z., J.J. and K.D.W.; Writing—Original Draft Preparation: L.S.; Writing—Review and Editing: K.Z., J.J., K.D.W., S.G., P.R., K.V.C., E.A., F.L., G.R., M.P., J.V.M., C.R. and P.P.; Visualization: L.S.; Supervision: K.Z., J.J., K.D.W. and P.P.; Project Administration: L.S. and K.Z.; Funding Acquisition: L.S., K.Z. and P.P.
Funding
This research was supported by the Belgian Foundation Against Cancer (Grant Number: FA/2014/237) and Stand up to Cancer, the Flemish Cancer Society (Grant Number: 00000000116000000177).
Conflicts of Interest
P.P. received research grants from AstraZeneca, Boehringer-Ingelheim, Pfizer, and Roche.
References
- 1.Wan J.C.M., Massie C., Garcia-Corbacho J., Mouliere F., Brenton J.D., Caldas C., Pacey S., Baird R., Rosenfeld N. Liquid biopsies come of age: Towards implementation of circulating tumour DNA. Nat. Rev. Cancer. 2017;17:223–238. doi: 10.1038/nrc.2017.7. [DOI] [PubMed] [Google Scholar]
- 2.Diehl F., Schmidt K., Choti M.A., Romans K., Goodman S., Li M., Thornton K., Agrawal N., Sokoll L., Szabo S.A., et al. Circulating mutant DNA to assess tumor dynamics. Nat. Med. 2008;14:985–990. doi: 10.1038/nm.1789. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Norton S.E., Lechner J.M., Williams T., Fernando M.R. A stabilizing reagent prevents cell-free DNA contamination by cellular DNA in plasma during blood sample storage and shipping as determined by digital pcr. Clin. Biochem. 2013;46:1561–1565. doi: 10.1016/j.clinbiochem.2013.06.002. [DOI] [PubMed] [Google Scholar]
- 4.Siravegna G., Marsoni S., Siena S., Bardelli A. Integrating liquid biopsies into the management of cancer. Nat. Rev. Clin. Oncol. 2017;14:531–548. doi: 10.1038/nrclinonc.2017.14. [DOI] [PubMed] [Google Scholar]
- 5.Warton K., Yuwono N.L., Cowley M.J., McCabe M.J., So A., Ford C.E. Evaluation of streck bct and paxgene stabilised blood collection tubes for cell-free circulating DNA studies in plasma. Mol. Diagn Ther. 2017;21:563–570. doi: 10.1007/s40291-017-0284-x. [DOI] [PubMed] [Google Scholar]
- 6.Russo M., Misale S., Wei G., Siravegna G., Crisafulli G., Lazzari L., Corti G., Rospo G., Novara L., Mussolin B., et al. Acquired resistance to the trk inhibitor entrectinib in colorectal cancer. Cancer Discov. 2016;6:36–44. doi: 10.1158/2159-8290.CD-15-0940. [DOI] [PubMed] [Google Scholar]
- 7.Nilsson R.J., Karachaliou N., Berenguer J., Gimenez-Capitan A., Schellen P., Teixido C., Tannous J., Kuiper J.L., Drees E., Grabowska M., et al. Rearranged eml4-alk fusion transcripts sequester in circulating blood platelets and enable blood-based crizotinib response monitoring in non-small-cell lung cancer. Oncotarget. 2016;7:1066–1075. doi: 10.18632/oncotarget.6279. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Best M.G., Wesseling P., Wurdinger T. Tumor-educated platelets as a noninvasive biomarker source for cancer detection and progression monitoring. Cancer Res. 2018;78:3407–3412. doi: 10.1158/0008-5472.CAN-18-0887. [DOI] [PubMed] [Google Scholar]
- 9.Normanno N., Cervantes A., Ciardiello F., De Luca A., Pinto C. The liquid biopsy in the management of colorectal cancer patients: Current applications and future scenarios. Cancer Treat. Rev. 2018;70:1–8. doi: 10.1016/j.ctrv.2018.07.007. [DOI] [PubMed] [Google Scholar]
- 10.Stewart C.M., Kothari P.D., Mouliere F., Mair R., Somnay S., Benayed R., Zehir A., Weigelt B., Dawson S.J., Arcila M.E., et al. The value of cell-free DNA for molecular pathology. J. Pathol. 2018;244:616–627. doi: 10.1002/path.5048. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Van Dessel L.F., Beije N., Helmijr J.C., Vitale S.R., Kraan J., Look M.P., de Wit R., Sleijfer S., Jansen M.P., Martens J.W., et al. Application of circulating tumor DNA in prospective clinical oncology trials—Standardization of preanalytical conditions. Mol. Oncol. 2017;11:295–304. doi: 10.1002/1878-0261.12037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Sherwood J.L., Corcoran C., Brown H., Sharpe A.D., Musilova M., Kohlmann A. Optimised pre-analytical methods improve kras mutation detection in circulating tumour DNA (ctdna) from patients with non-small cell lung cancer (nsclc) PLoS ONE. 2016;11:e0150197. doi: 10.1371/journal.pone.0150197. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Van Ginkel J.H., van den Broek D.A., van Kuik J., Linders D., de Weger R., Willems S.M., Huibers M.M.H. Preanalytical blood sample workup for cell-free DNA analysis using droplet digital pcr for future molecular cancer diagnostics. Cancer Med. 2017;6:2297–2307. doi: 10.1002/cam4.1184. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Fernando M.R., Norton S.E., Luna K.K., Lechner J.M., Qin J. Stabilization of cell-free rna in blood samples using a new collection device. Clin. Biochem. 2012;45:1497–1502. doi: 10.1016/j.clinbiochem.2012.07.090. [DOI] [PubMed] [Google Scholar]
- 15.Chiu R.W., Poon L.L., Lau T.K., Leung T.N., Wong E.M., Lo Y.M. Effects of blood-processing protocols on fetal and total DNA quantification in maternal plasma. Clin. Chem. 2001;47:1607–1613. [PubMed] [Google Scholar]
- 16.El Messaoudi S., Rolet F., Mouliere F., Thierry A.R. Circulating cell free DNA: Preanalytical considerations. Clin. Chim. Acta. 2013;424:222–230. doi: 10.1016/j.cca.2013.05.022. [DOI] [PubMed] [Google Scholar]
- 17.Risberg B., Tsui D.W.Y., Biggs H., Ruiz-Valdepenas Martin de Almagro A., Dawson S.J., Hodgkin C., Jones L., Parkinson C., Piskorz A., Marass F., et al. Effects of collection and processing procedures on plasma circulating cell-free DNA from cancer patients. J. Mol. Diagn. 2018;20:883–892. doi: 10.1016/j.jmoldx.2018.07.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Demuth C., Spindler K.G., Johansen J.S., Pallisgaard N., Nielsen D., Hogdall E., Vittrup B., Sorensen B.S. Measuring kras mutations in circulating tumor DNA by droplet digital pcr and next-generation sequencing. Transl. Oncol. 2018;11:1220–1224. doi: 10.1016/j.tranon.2018.07.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Volckmar A.L., Sultmann H., Riediger A., Fioretos T., Schirmacher P., Endris V., Stenzinger A., Dietz S. A field guide for cancer diagnostics using cell-free DNA: From principles to practice and clinical applications. Genes Chromosomes Cancer. 2018;57:123–139. doi: 10.1002/gcc.22517. [DOI] [PubMed] [Google Scholar]
- 20.Sorber L., Zwaenepoel K., De Winne K., Van Casteren K., Augustus E., Jacobs J., Zhang X.H., Galdermans D., De Droogh E., Lefebure A., et al. A multicenter study to assess egfr mutational status in plasma: Focus on an optimized workflow for liquid biopsy in a clinical setting. Cancers. 2018;10:290. doi: 10.3390/cancers10090290. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Streck . Cell-Free DNA bct: Instructions for Use. Streck; Omaha, NE, USA: 2014. [Google Scholar]
- 22.Devonshire A.S., Whale A.S., Gutteridge A., Jones G., Cowen S., Foy C.A., Huggett J.F. Towards standardisation of cell-free DNA measurement in plasma: Controls for extraction efficiency, fragment size bias and quantification. Anal. Bioanal. Chem. 2014;406:6499–6512. doi: 10.1007/s00216-014-7835-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Cheng H.H., Yi H.S., Kim Y., Kroh E.M., Chien J.W., Eaton K.D., Goodman M.T., Tait J.F., Tewari M., Pritchard C.C. Plasma processing conditions substantially influence circulating microrna biomarker levels. PLoS ONE. 2013;8:e64795. doi: 10.1371/journal.pone.0064795. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Mitchell A.J., Gray W.D., Hayek S.S., Ko Y.A., Thomas S., Rooney K., Awad M., Roback J.D., Quyyumi A., Searles C.D. Platelets confound the measurement of extracellular mirna in archived plasma. Sci. Rep. 2016;6:32651. doi: 10.1038/srep32651. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Cerkovnik P., Perhavec A., Zgajnar J., Novakovic S. Optimization of an rna isolation procedure from plasma samples. Int. J. Mol. Med. 2007;20:293–300. doi: 10.3892/ijmm.20.3.293. [DOI] [PubMed] [Google Scholar]
- 26.Decock A., De Wever O., Dhondt B., D’huyvetter T., Helsmoortel H., Hendrix A., Philippron A., Verniers K., Mestdagh P., Vandesompele J. LKI Symposium, Leuven, Belgium—Abstract; http://hdl.handle.net/1854/LU-8552649 . Ghent University Library; Ghent, Belgium: 2018. The extracellular rna quality control (exrnaqc) study: Testing and controlling pre-analytical variables of rna sequencing on liquid biopsies. [Google Scholar]
- 27.Wrzyszcz A., Urbaniak J., Sapa A., Wozniak M. An efficient method for isolation of representative and contamination-free population of blood platelets for proteomic studies. Platelets. 2017;28:43–53. doi: 10.1080/09537104.2016.1209478. [DOI] [PubMed] [Google Scholar]
- 28.Doss J.F., Corcoran D.L., Jima D.D., Telen M.J., Dave S.S., Chi J.T. A comprehensive joint analysis of the long and short rna transcriptomes of human erythrocytes. BMC Genom. 2015;16:952. doi: 10.1186/s12864-015-2156-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Rikkert L.G., van der Pol E., van Leeuwen T.G., Nieuwland R., Coumans F.A.W. Centrifugation affects the purity of liquid biopsy-based tumor biomarkers. Cytom. A. 2018;93:1207–1212. doi: 10.1002/cyto.a.23641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Lin J.J., Riely G.J., Shaw A.T. Targeting alk: Precision medicine takes on drug resistance. Cancer Discov. 2017;7:137–155. doi: 10.1158/2159-8290.CD-16-1123. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Gainor J.F., Dardaei L., Yoda S., Friboulet L., Leshchiner I., Katayama R., Dagogo-Jack I., Gadgeel S., Schultz K., Singh M., et al. Molecular mechanisms of resistance to first- and second-generation alk inhibitors in alk-rearranged lung cancer. Cancer Discov. 2016;6:1118–1133. doi: 10.1158/2159-8290.CD-16-0596. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Best M.G., Sol N., Kooi I., Tannous J., Westerman B.A., Rustenburg F., Schellen P., Verschueren H., Post E., Koster J., et al. Rna-seq of tumor-educated platelets enables blood-based pan-cancer, multiclass, and molecular pathway cancer diagnostics. Cancer Cell. 2015;28:666–676. doi: 10.1016/j.ccell.2015.09.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Qin J., Williams T.L., Fernando M.R. A novel blood collection device stabilizes cell-free rna in blood during sample shipping and storage. BMC Res. Notes. 2013;6:380. doi: 10.1186/1756-0500-6-380. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.George N. Evaluation of available blood collection tubes for use in stabilizing concentrations of extracellular vesicles/exosomes and associated cell-free rna; Proceedings of the American Association for Cancer Research Annual Meeting 2018; Chicago, IL, USA. 14–18 April 2018. [Google Scholar]
- 35.Goethals S., De Wilde A., Lesage K., Smits E., Pauwels P., Peeters M. Tumorbank@uza: A collection of tissue, fluid samples and associated data of oncology patients for the use in translational research. Open J. Bioresour. 2018;5:4. doi: 10.5334/ojb.30. [DOI] [Google Scholar]
- 36.Sorber L., Zwaenepoel K., Deschoolmeester V., Roeyen G., Lardon F., Rolfo C., Pauwels P. A comparison of cell-free DNA isolation kits: Isolation and quantification of cell-free DNA in plasma. J. Mol. Diagn. 2017;19:162–168. doi: 10.1016/j.jmoldx.2016.09.009. [DOI] [PubMed] [Google Scholar]
- 37.Rago C., Huso D.L., Diehl F., Karim B., Liu G., Papadopoulos N., Samuels Y., Velculescu V.E., Vogelstein B., Kinzler K.W., et al. Serial assessment of human tumor burdens in mice by the analysis of circulating DNA. Cancer Res. 2007;67:9364–9370. doi: 10.1158/0008-5472.CAN-07-0605. [DOI] [PubMed] [Google Scholar]
- 38.Diaz I.M., Nocon A., Mehnert D.H., Fredebohm J., Diehl F., Holtrup F. Performance of streck cfdna blood collection tubes for liquid biopsy testing. PLoS ONE. 2016;11:e0166354. doi: 10.1371/journal.pone.0166354. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.