. 2019 May 8;2019:3489634. doi: 10.1155/2019/3489634

Multiresidue Pesticides Analysis of Vegetables in Vietnam by Ultrahigh-Performance Liquid Chromatography in Combination with High-Resolution Mass Spectrometry (UPLC-Orbitrap MS)

Nam Vu-Duc 1,, Trung Nguyen-Quang 1, Thuy Le-Minh 1,2, Xuyen Nguyen-Thi 1,2, Tri Manh Tran 2, Hai Anh Vu 2, Lan-Anh Nguyen 2, Tien Doan-Duy 3, Bui Van Hoi 4, Cam-Tu Vu 4, Dung Le-Van 5, Lan-Anh Phung-Thi 6, Hong-An Vu-Thi 7, Dinh Binh Chu 7,
PMCID: PMC6530207  PMID: 31205797

Abstract

An ultrahigh-performance liquid chromatography in combination with high-resolution mass spectrometry Thermo Q-Extractive Focus Orbitrap MS has been introduced for analysis of multiclass pesticides in vegetable samples collected in Hanoi, Vietnam. Multiclass pesticides were separated on the Thermo Hypersil Gold PFP column utilizing a gradient of the mobile phase consisting of 5 mM ammonium formate, 0.1% formic acid in deionized water, and methanol. The target analytes were detected in the full-scan mode on Thermo Scientific Q-Exactive Focus Orbitrap MS for quantitation at the optimum operating conditions. These conditions included, but not limit to, the resolution of 70000 at the full width at half maximum in both positive and negative mode, mass range from 80 to 1000 m/z, and optimized parameters for the heated electrospray ionization source. The identification of the analytes in real samples was based on retention times, mass to charge ratios, mass accuracies, and MS/MS spectra at the confirmation mode with the inclusion list of target analytes. The mass accuracies of target analytes were from −4.14 ppm (dinotefuran) to 1.42 ppm (cinosulfuron) in the neat solvent and from −3.91 ppm (spinosad D) to 1.29 ppm (cinosulfuron) in the matrix-matched solution. Target analytes in the vegetable-based matrix were extracted by the QuEChERS method. Some critical parameters of the analytical method such as linearity, repeatability, limit of detection, and limit of quantitation have been evaluated and implemented. Excellent LOD and LOQ of the developed method were achieved at the range of 0.04–0.85 and 0.13–2.9 μg·kg−1, respectively. Intraday and interday repeatability of the analytical signal (peak area, n=6) of the developed method were below 3% and 10%, correspondingly. The matrix effect, extraction recovery, and overall recovery were fully investigated by spiking experiments. Experimental results demonstrated that the ionization suppression or enhancement was the main contribution on the overall recoveries of target analytes. Finally, the in-house validated method was applied to pesticides screening in vegetables samples in local villages in Hanoi, Vietnam. The concentrations of all target analytes were below limit of quantitation and lower than US-FDA or EU maximum residue levels.

1. Introduction

The residual of pesticides in the foods is global, concerning in the context of food safety. Pesticide is used in the agriculture in order to protect crops. Therefore, pesticide presents as residue in food, especially in vegetables. The maximum residue level of pesticide in vegetables is lower than 0.01 mg·kg−1 according to EU regulation [1] and 0.01 mg·kg−1 according to US-FDA regulation as well [2]. Other regions have their own standards, regulations, requirements, and maximum residual levels of pesticides in their food production and consumption [3, 4]. In Vietnam, the pesticide residue level in food in general and in vegetables was also monitored. According to Regulation TT50/2016/TT-BYT of the Ministry of Health, Vietnam, the maximum residue level of pesticides ranged from 0.01 to 100 mg·kg−1, depending on each kind of pesticides or vegetables [5]. Ministry of Agriculture and Rural Development, Vietnam (MARD), also regulated maximum levels used of more than 1700 different pesticides. Besides, more than 30 different pesticides were banned in agriculture according to Regulation TT03/2018/TT-BNNPTNT of MARD [6].

Recently, many mass spectrometry techniques have been proposed for analysis of pesticides in environmental samples such as GC-EI-MS, GC-EI-MS/MS, GC-CI-MS, LC-ESI-MS, and LC-ESI-MS/MS [713]. Among these methods, LC-MS/MS has been the most popular for pesticide analysis, especially for analysis of multiclass pesticides [14, 15]. In addition, the high-resolution mass spectrometry-based methods with several advantages, namely, excellent accurate mass and high sensitivity, have been introduced for screening of residual pesticides in food matrices [9]. Several sample preparation procedures have been reported, for example, liquid-liquid extraction and solid-phase extraction for such kind of analytes [16]. Besides, QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) sample preparation method has been known as the most powerful for removing isobaric and nonisobaric interference compounds in the food matrices. This method could be proposed as a “golden standard” for the sample preparation, especially in multiclass residue analysis. In the recent reports, a QuEChERS has been in combination with high-resolution mass spectrometry for screening of pesticides, their metabolites, or multiclass compounds in various food matrices [1719], for instance, in baby food [20], in fruits, in vegetables [21], and in surface water [22]. Therefore, a fast, sensitivity, and reliable analytical method is necessary for quality control of vegetable-based food in domestic, imported, or exported food matrices (vegetable, fruit, and meat).

In this work, a QuEChERS sample preparation and reversed-phase liquid chromatography in combination with high-resolution Orbitrap MS have been introduced for analysis of multiclass pesticides in vegetable samples. The ionization suppression or enhancement has been also investigated by spiking experiments. The in-house validation of the developed method including limit of detection (LOD), limit of quantitation (LOQ), linearity, short- and long-term stability, matrix effect, and overall recovery has been performed via spiking experiments, according to the guidelines of the USA-FDA, EU [23, 24]. Finally, the developed method was applied to screen multiresidues in vegetable samples, which were collected from several villages in Hanoi, Vietnam.

2. Materials and Methods

2.1. Chemicals and Reagents

Fifty-three pesticide standards (high purity grade, >90%) were purchased from Dr. Ehrenstorfer GmbH. Acetonitrile (ACN, LC-MS grade), methanol (MeOH, LC-MS grade) for HPLC with 99.80% of purity, and formic acid (FA, ACS reagent) were from Merck (Merck, Singapore). Ammonium formate (LC-MS grade, Sigma-Aldrich, Singapore) was also used for preparation of the mobile phase. Standard compounds were classified into 30 groups and used to prepare individual stock solutions around 1000 μg·mL−1 in appropriate solvents such as acetone (GC-MS grade), methanol, n-hexane (GC-MS grade), acetonitrile, and ethanol (ACS reagent) in amber vials. The mixed standard solutions of all target analytes (10 μg·mL−1) were prepared and diluted with acetonitrile. Stock standards were stored in the amber LC vial at 4°C. The working standard solutions were daily prepared by diluting the mixed standard solution in the mobile phase. The mobile phase was daily prepared by dissolving appropriate amount of ammonium formate in methanol/deionized water (Milli-Q Integral 3, Merck Millipore, France) containing 0.1% formic acid. The mobile phase was degassed in the ultrasonic bath (S 100H, Elma, Germany) to eliminate dissolved gas.

2.2. Instrumentation

An ultrahigh-performance liquid chromatography (UPLC) system including the column oven and thermostat autosampler (Ultimate 3000, Thermo Fisher Scientific, Bremen, Germany) in combination with the Thermo Scientific Q-Exactive Focus Orbitrap MS (Thermo Fisher Scientific, Bremen, Germany) was used for data acquisition.

For liquid chromatographic separation, Hypersil GOLD PFP column (150 × 2.1 mm, 3 μm, Thermo Fisher Scientific, USA) was used for separation of target analytes at the temperature of 40°C. The binary mobile phases were 0.1% FA + 5 mM HCOONH4 in H2O (A) and 0.1% FA + 5 mM HCOONH4 in MeOH (B). The gradient elution started at 2% B in 0.25 minutes, raised to 30% B in 0.75 minutes, and linearly increased to 100% B in 24 minutes (held for 5 minutes). In the end, the eluent was restored to initial conditions in 0.5 minutes and held for 7.0 minutes to reequilibrate the column for the next injection. Total time for chromatographic separation was 37.5 minutes. The flow rate was constantly kept at 0.3 mL·min−1 during the whole chromatographic analysis process. Both samples and standard solutions were kept at 10°C in the sample tray. A 5 µL of standard or samples was injected into LC-Q-Exactive Focus Orbitrap MS system via an autosampler. The needle and the sample loop in the autosampler were washed triplicate, using the mixture of methanol and deionized water (1 : 1, v : v).

A high-resolution mass spectrometer Thermo Scientific Q-Exactive Focus Orbitrap MS equipped with heated electrospray ionization (HESI) and working at 70000 full width at half maximum (FWHM) resolution (at 200 Da) was operated in both positive and negative electrospray ionization modes. The mass spectrometer was calibrated before each batch of measurement by using Pierce positive/negative ion mass calibration solution (Thermo Fisher, USA). This system was optimized by the direct infusion experiment, using the mixture of pesticides standard solutions in the mobile phase. Optimum operating conditions were achieved with following parameters: sheath gas pressure at 32 psi; auxiliary gas flow rate at 7 L·min−1; sweep gas flow rate at 1 L/min; spray voltage +2800 V; and −2500 V for positive and negative ionization mode, respectively; capillary temperature at 320°C; vaporizer temperature at 295°C; and S-lens RF level at 50 V. The HRMS was acquired under full MS mode (resolution 70000-FWHM at 200 Da) over the mass range m/z of 80–1000 for both positive and negative ionization mode, and it was conducted to measure the target ions of precursors. The full MS/dd-MS2 (full-scan and data-dependent MS/MS mode) could simultaneously record the MS/MS (fragmentation) spectra for the precursors. Besides, the full MS/confirmation mode (with an inclusion list of target analytes) was also used to confirm fragments of the selected precursors. The dd-MS2 with confirmation mode conditions was set up with the following parameters: resolution 17500 FWHM; mass isolation window 1.0 Da; maximum and minimum automatic gain control (AGC) target 8 × 103 and 5 × 103, respectively; normalized collision energy (NCE) 30%; spectrum data format for confirmation: centroid. All the parameters of the UHPLC-HRMS system were controlled through Thermo Scientific Xcalibur software version 4.0 (Thermo Scientific, Bremen, Germany).

2.3. Sample Preparation

Vegetable samples were collected from local villages in Hanoi, Vietnam, and stored at 4°C until analysis. In brief, 10 g of each homogenized sample was weighed into a 50 mL QuEChERS centrifuge tube (Thermo Scientific, USA), which contained 4 g MgSO4, 1 g NaCl, 1 g Na3 citrate, and 0.5 g Na2 citrate. After that, 10 mL of ACN was added into the QuEChERS tube. The sample was placed in an ultrasonic bath for 5 minutes before centrifuging at 4450 × g for 15 minutes (Z 326 k, Hermle, Germany). The supernatant was filtrated through 0.45 μm Titan 3, PTFE membrane (Thermo Scientific, USA). The clear aqueous solution was collected into the 2 mL amber LC vial (Thermo Scientific, USA) and then injected into UPLC-Q-Exactive Focus Orbitrap MS system via the autosampler at the optimum experimental conditions. For recovery testing, the experiment was conducted with three groups of pooled samples, which were the mixtures of six types of homogenized vegetables (containing the same amount of cabbage, white mustard, Chinese spinach, green mustard, water morning spinach, and edible chrysanthemum in mixture) to control the effect of matrix. The individual matrix was also investigated in the same manner. The first set of pooled samples was performed following the above procedure. The second set and the last set of pooled samples were spiked preextraction and postextraction, respectively, at the concentration of 50 ng·mL−1 for all target analytes in the final solution, as shown in Figure 1. In addition, the same experiments were conducted with three vegetables: cabbage, white mustard, and edible chrysanthemum, which are planted most popularly in the winter time in Vietnam. Besides, real vegetable samples containing cabbage, white mustard, Chinese spinach, green mustard, water morning spinach, and star gooseberry were prepared as the same manner as above for quantitation of pesticides. The concentrations of target analytes in real samples were calculated by external calibration curves.

Figure 1.

Figure 1

Experimental setup for assessment of recovery, matrix effect, and entire sample preparation procedure.

2.4. Matrix Effect in LC-Orbitrap MS

In liquid chromatography tandem mass spectrometry, the matrix is the most important factor that affects the reproductivity of the analytical method. For the assessment of the matrix effect, several designs of experiments were proposed, for example, postextraction spiking experiments, matrix-match calibration curves, and isotopic labelled internal standard spiking experiments [25, 26]. In this study, the matrix effect was investigated by postextraction spiking experiments. Pesticide standards were spiked pre- and postextraction during the sample preparation procedure. A set of 15 pooled samples was used for such experiments as in Figure 1.

Matrix effect (ME), in terms of signal suppression/enhancement, was investigated and assessed by comparing analytical signals of analytes in samples postextraction spike with analytes to those in the “neat” solvent. The matrix effect was calculated by the following equation:

ME %=100peak  area  of  target analytepostextraction  spikepeak  area  of  target analytein  standard  solution. (1)

Ideally, a value of 100% means the absence of the matrix effect on the MS measurement. ME lower than 100% and higher than 100% are indicated to be ionization suppression and enhancement, respectively. ME is acceptable when its absolute value ranged from 80% to 120%. Any value of ME that is outside that range denotes matrix effect occurs in the MS measurement [10, 27]. In addition, the losing of analytes or contamination in the sample preparation step was also addressed in this study. In order to figure out the contribution of each step in the entire sample preparation procedure, spiking experiments were carried out before and after sample preparation. Therefore, the recovery of the extraction step (RE) (R) is calculated as follows:

RE %=100peak  area  of  target analytepreextraction  spikepeak  area  of  target analytepostextraction  spike. (2)

The recovery of the entire of sample preparation is calculated as follows:

R %=100peak  area  of  target  analytepreextraction  spikepeak  area  of  target  analytein  standard  solution. (3)

By using three above equations, the recovery of each step could be investigated and assessed. The recovery of each step is very noteworthy and important for development and validation of a new analytical method, in terms of overall extraction efficiency. The LOD, LOQ, MQLs, and MDLs were assessed by spiking standards in the real matrix in pooled samples and individual matrix as well. Therefore, the matrix effect, extraction efficiency, and overall recovery were taken into account for calculation of LOD and LOQ.

2.5. Fragmentation Pathway and Mass Accuracy

The mass accuracy of all target analytes in standard solution and matrix-match solution was assessed by injecting independently five times of 50 ng·mL−1 of all analytes in mobile phase/matrix-match solution into the LC-Q-Exactive Orbitrap MS. The mass of the protonated molecular ion (in the positive ionization mode) or deprotonated molecular ion (in the negative ionization mode) was extracted by Thermo qualitative software version 4.0. The theoretical mass was calculated by online enviPat Web 2.2 from Eawag aquatic research (https://www.envipat.eawag.ch/index.php). The mass accuracy was calculated by subtracting the theoretical mass from experimental one and then dividing the result by the theoretical mass and presented in part per million mass accuracy. The fragmentation pathway of all target analyte was performed at 30% NCE (normalized collision energy) at resolution 17500 FWHM and crosschecked using Mass Frontier software version 7.0 (Thermo Scientific, USA). Mass fragmentation pathway is also an interesting topic, especially in the context of the nontargeted analysis. The fragmentation pathway of pesticides has been reviewed on the work of Niessen [28]. The product ions with associate mass accuracy of all target analytes are listed in Table S5 in Supplemental Materials. In addition, identification points (IPs) for confirmation of target analytes presenting in the samples were calculated according to European Council Directive 96/23/EC for MS-based methods [29]. The identification of the analytes was based on the protonated/deprotonated molecular mass to charge ratio, at least two daughter ions in MS/MS confirmation modes, isotopic pattern, and mass accuracy of these ions. Therefore, the minimum IPs number achieved 5 in this work.

2.6. Quality Control

The Thermo Orbitrap Q-Exactive Focus MS was calibrated by using Pierce positive/negative ion mass calibration solution (Thermo Fisher, USA) before each batch of samples. Before each batch, the quality control (QC) sample was injected five-time repeatability for checking intensity and retention time. After each 10 samples, three blank samples (mobile phase solution) were injected in triplicate for assessment of the carry-over effect. In total, the number of method blank and quality control samples was approximately 20% of the total number of injections on LC-Orbitrap MS, as a proposal from Peters et al. [30].

2.7. Data Evaluation

Quantification of pesticides in vegetables was performed by Thermo Trace Finder version 3.3 (Thermo Scientific, Bremen, Germany) using external calibration curves. The mass accuracy of quantitative ions was set at 5 ppm for all pesticides in both negative and positive ionization modes. Weighted and nonweighted linearity and quaternary of the peak areas plotted as a function of concentration of the target analytes were used for quantification. The confirmation of analytes presenting in the real samples was based on the retention times, mass accuracies, isotopic pattern, and MS/MS spectrum at 30% of the normalized collision energy. The peak area was extracted and integrated without a smooth factor in 5 ppm mass resolution window for confirmation. Other important parameters of the developed method are LOD and LOQ. The LOD and LOQ of the developed method were assessed by spiking experiments at low concentrations (at level 10 ng·g−1) and then injecting into the LC-Orbitrap MS at the optimized operating conditions. The signal to noise ratio, which was determined according to European Pharmacopoeia guideline, was taken into account for the calculation of LOD (3 ∗ S/N) and LOQ (10 ∗ S/N) [3033].

3. Results and Discussion

3.1. Chromatographic Separation of All Target Analytes

Because of the nature of multiclass pesticides in the standard solutions, reversed-phase column was the best choice for separation. Many reversed-phase columns with different kinds of stationary phases have been introduced for pesticides analysis. In this work, the Thermo Hypersil GOLD PFP column (silica-based pentafluorophenyl endcapped stationary phase) has been used for chromatographic separation of multiclasses pesticides. The advantages of the PFP stationary phase for retaining of polar compounds were assessed by Si-Hung et al. [34]. Separation mechanism of the PFP column is multi-interactions, such as hydrogen bonding, dipole-dipole interactions, and hydrophobic between functional groups on the surface of stationary phase and analytes, especially the π-π interaction of the aromatic-like compounds with the stationary phase [35, 36]. For testing, several available columns in our lab were used such as Hypersil ODS (C18, 125 × 2.1 mm, 5 μm, Thermo Fisher Scientific, USA) and HyperClone™ 5 μm ODS 120 Å (C18, 125 × 4.0 mm, 5 μm, Phenomenex, USA). Total ion chromatograms in the positive ionization mode of the mixture standard (500 ng·mL−1) on three LC columns: Thermo Hypersil GOLD PFP, Hypersil ODS, and HyperClone™ were depicted in Figures 2(a)2(c), respectively. It is clearly shown that the Hypersil GOLD PFP LC column was the best separation for multiclass pesticides in terms of separation efficiency and peak shape. Therefore, this column has been chosen for the further experiments.

Figure 2.

Figure 2

Total chromatogram of multiclass pesticides on the different stationary phases: (a) PFP column; (b) Hypersil ODS; (c) HyperClone™ 5 μm ODS 120 Å.

In addition, components of mobile phase, organic modifiers, and other factors also have influenced the chromatographic separation efficiency. In this work, the mobile phase containing methanol, ammonium formate, and formic acid was used as the suggestion from Lee et al. with some modifications [37]. For optimization of chromatographic conditions, the effects of various method parameters such as mobile phase, flow rate, and solvent ratio were evaluated; the chromatographic parameters such as peak asymmetric factor, resolution, and column efficiency were calculated. The best results were obtained with a gradient mobile phase composition of 5 mM ammonium formate, 0.1% formic acid, and methanol, at a flow rate of 0.3 mL·min−1 as mentioned above. Adding 5 mM ammonium formate in the mobile phase improved the chromatographic peak shape of the target analyte, especially the compounds that were eluted in the end of chromatogram [38].

Six independent standard solutions were prepared and injected into the LC-Orbitrap MS in triplicate at the optimum operating conditions as above. The short-term (in 3 hours of continuous measurement) and long-term (in 24 hours of continuous measurement) stabilities of retention time in the mobile phase and in matrix-match solution were assessed and presented by relative standard deviation (RSD). The relative standard deviation of short- and long-term stabilities of retention time of these compounds are calculated and presented in Table 1. Stability of all target analytes in terms of retention time was calculated and is listed in Table S1 of Supplemental Materials.

Table 1.

Short-term and long-term stabilities of retention time of three target analytes in the standard solution and in the real matrix.

No Analytes Retention time (min) Standard solution Matrix-match solution
RSD (short term, n=6) (%) RSD (long term, n=6) (%) RSD (short term, n=6) (%) RSD (long term, n=6) (%)
1 Carbofuran 8.03 0.18 0.53 0.19 0.31
2 Azoxystrobin 13.49 0.06 0.25 0.27 0.36
3 Pyridaben 19.54 0.09 0.18 0.10 0.22

As can be seen from Table 1, the excellent repeatability of retention time of three selected analytes was achieved, and relative standard deviation of short- and long-term stabilities in standard solution were below 0.18 and 0.53%, respectively. The repeatability of retention time of the target analytes was also investigated in the real sample by postextraction spiking experiments. In the matrix-match solution, relative standard deviations of short- and long-term stabilities of retention time of these compounds were below 0.27 and 0.36%, respectively. The total ion chromatogram (TIC) and extracted ion chromatogram (EIC) of three representative pesticides in the standard solution and the matrix-match solution are shown in Figures 3(a) and 3(b), respectively.

Figure 3.

Figure 3

The total ion and extracted ion chromatogram of the target analytes in the standard solution (a) and in the matrix-match solution (b) on the Thermo Hypersil GOLD PFP column, and other operating conditions are mentioned in Section 2.2. The first and second numbers on the right upper are mass-to-charge ratio and retention time, respectively.

It is clear from Figure 3 that the retention times of three selected analytes were observed a little earlier in the matrix-match solution. However, relative standard deviations of three selected analytes were below 0.3% for both short- and long-term stabilities in terms of retention time. It should be concluded that the developed method was excellent and stable for chromatographic separation of the multiclass pesticides. In addition, sample matrix was not affected on the chromatographic separation. The short- and long-term stabilities of retention time of all target analytes on the Thermo HyperGold PFP column are listed in Table S1 in Supplemental Materials. Relative standard deviations of short- and long-term stabilities of retention time of all target analytes were in the range of 0.07–0.2% and 0.13–0.29%, respectively. The total ion chromatogram and extracted ion chromatograms of all target analytes in standard solution and matrix-match solution with both ionization modes are depicted in Figures S1S4 in Supplemental Materials, respectively.

3.2. Fragmentation Pathway and Mass Accuracy

For assessment of mass accuracy, five independent mixtures of target analytes in the mobile phase and matrix-match solutions were injected on UPLC-Orbitrap MS at the optimum operating conditions. The mass accuracies of target analytes are calculated and presented as parts per million (ppm) in Table 2.

Table 2.

Mass and mass accuracy of three representative pesticides in the standard solution and in the matrix-match solution.

No Analytes Chemical formula Theoretical mass (Da, m/z) Standard solution Matrix-match solution
Experimental mass (Da), mean value, n=5 Mass accuracy (ppm) Experimental mass (Da), mean value, n=5 Mass accuracy (ppm)
1 Carbofuran C12H14NO3 222.1125 222.1122 −1.4 222.1120 −1.8
2 Azoxystrobin C22H17N3O5 404.1241 404.1234 −2.0 404.1231 −2.2
3 Pyridaben C19H25ClN2OS 365.1449 365.1443 −1.6 365.1441 −2.5

Theoretical mass was calculated through online enviPAT software (https://www.envipat.eawag.ch/).

As clearly demonstrated in Table 2, the mass accuracy of three target analytes was achieved below 2.5 ppm. It should be noted that the excellent mass accuracy was achieved in both pure solvent and matrix-match solutions. The mass accuracies of all target analytes are calculated and presented in Table S1 in Supplemental Materials. Mass accuracies of all target analytes were below ±3 ppm for both matrices: pure solvent and matrix-match solution. In addition, the tandem mass spectra of all target analytes in standard solution and in the matrix-match solution have been performed at 17500 FWHM resolution, 1 Da isolated mass window, and 30% normalized collision energy. Mass accuracy of all product ions is listed in Table S5 in Supplemental Materials. It was clearly shown that mass accuracy of all product ions was achieved below 5 ppm, except for dinotefuran (−11.5 ppm). The lower mass accuracy of product ions in comparison with the precursor ion should be attributed by lower resolution setting in the confirmation MS/MS mode (17500 FWHM at 200 Da).

3.3. Validation of the Developed Method

3.3.1. Stability of the Analytical Signal

Short term and long term of the analytical signal play as a critical role in terms of measurement uncertainty of the developed analytical method. For assessment of the analytical signal, two sets of five solutions containing all target analytes at a concentration of 50 ng·mL−1 were prepared in solvent and in matrix-match solution, respectively. These solutions were injected on the LC-LC Q-Exactive Focus Orbitrap MS at the optimum conditions. The short-term and long-term stabilities were performed in three and twenty-four hours of continuous measurement, respectively. The peak area was integrated on TraceFinder version 3.3 (Thermo Scientific, Bremen, Germany) with 5 ppm mass accuracy window. The peak area and relative standard deviation of peak are shown in Table 3.

Table 3.

Short-term and long-term repeatability of the analytical signal (peak area) of target analytes in UPLC-Orbitrap MS.

No Analytes Standard solution, RSD of peak area (%) Matrix-match solution, RSD of peak area (%)
Short term, n=5 Long term, n=5 Short term, n=5 Long term, n=5
1 Carbofuran 0.70 4.22 0.77 0.99
2 Azoxystrobin 0.96 5.78 1.17 6.31
3 Pyridaben 1.71 5.54 3.68 4.40

As can be seen from Table 3, excellent repeatability of the analytical signal was achieved for both short term and long term. RSD of the peak area of short term and long term was below 3.68% and 5.78% for both standard and matrix-match solutions, respectively. It was worthy to note that the good repairability of the analytical signal was achieved in this study. The relative standard deviation of the analytical signal of all target analytes is listed in Table S6 in Supplemental Materials. All RSD values of the analytical signal were lower than the acceptable value according to Horwitz [39].

3.3.2. Linearity, LOD, and LOQ

Seven independent standard solutions (with concentrations of 10, 25, 50, 100, 250, 500, and 1000 μg·L−1 for all target analytes) of all targeted analytes were prepared by dilution stock solution in the mobile phase and injected triplicate into the LC-Orbitrap MS at the optimum operating conditions. The peak area of the target analyte was taken into account for quantitation. The average peak area was fitted either linearity or as a quadratic function of concentration of target analyte. The nonweighted linearity/quadratic or weighted linearity/quadratic of peak area (Y) as a function of concentration (X) was performed as a proposal from Gu et al. [40].

As shown in Table 4, the good correlation between analytical signal (peak area) and concentration of analyte (R 2 > 0.998 for three representative analytes) was achieved. The calibration equations and correlation coefficients of all target analytes are listed in Table S2 in Supplemental Materials.

Table 4.

Calibration curves and correlation of three representative analytes on the LC-Q-Exactive Focus Orbitrap MS.

No Analytes t R (min) Polarity Regression equation Correlation coefficient (R 2)
1 Carbofuran 8.03 + Y = 1.557e7 ∗ X + 8.567e7 0.9989
2 Azoxystrobin 13.49 + Y = 1.503e7 ∗ X + 8.688e7 0.9980
3 Pyridaben 19.54 + Y = 2.855e6 ∗ X + 6.367e6 0.9993

As can be clearly seen from Table 5, LOD and LOQ of the developed method were high enough for directly analyzing multiclass pesticides in vegetable samples according to US-FDA and European Commission [23, 24]. The LOD and LOQ of all analytes are listed in Table S3 in Supplemental Materials. It should be noted that the developed LOD and LOQ of the method were also comparable with recent publications [4145].

Table 5.

Analytical figures of merit of the LC-Q-Exactive Orbitrap MS method for three representative pesticides.

No Analytes t R (min) LOD (ng·mL−1) LOQ (ng·mL−1) MDL (pg) MQL (pg)
1 Carbofuran 8.03 0.09 0.30 0.5 1.5
2 Azoxystrobin 13.48 0.02 0.07 0.1 0.3
3 Pyridaben 19.54 0.09 0.30 0.5 1.5

Absolute method of detection (MDL) and method of quantitation (MQL) are defined as the absolute amount of analyte injected into the LC column.

3.3.3. Overall Recovery and Matrix Effect

Matrix effect (ionization suppression/enhancement) is the most important factor in liquid chromatography in combination with tandem mass spectrometry, especially in electrospray ionization (ESI) mass spectrometry, because it influences the robustness and ruggedness of the analytical method. Ideally, a value of 100% means the absence of the matrix effect on the MS measurement. ME is acceptable when it is in range of 80% to 120%. Any value of ME that is outside that range denotes that the matrix effect occurs in the MS measurement [46]. Due to the fact that there is no vegetable-based matrix-certified reference materials for multiclass pesticides commercially available at the moment, the assessment of the matrix effect and recovery of the sample preparation were performed by spiking experiments in this study as the proposal from B.K. Matuszewski with some modifications [47]. The matrix effect during validation of the developed method was investigated by comparison of the analytical signal of a target analyte (peak area or peak height normally) in the postextraction spiked solution and that of the same target analyte in the mobile phase.

The matrix effect, recovery of sample preparation, and overall recovery were calculated by equations (1)–(3). ME, RE, and R of three representative analytes are listed in Table 6.

Table 6.

Matrix effect, recovery of extraction, and overall recovery of three representative pesticides in pooled samples.

No Analytes Class t R (min) ME ± RSD (%) (n=5) RE ± RSD (%) (n=5) R ± RSD (%) (n=5)
1 Azoxystrobin Strobilurin 13.42 140.1 ± 13.8 60.7 ± 11.4 85 ± 17.9
2 Carbofuran Carbamate 7.96 97.1 ± 1.6 89.9 ± 7.4 87.3 ± 7.5
3 Pyridaben Pyridazinone 19.47 121 ± 8.7 72.3 ± 10.8 87.5 ± 13.8

As can be seen from Table 6, the main factors affected on the overall recoveries of azoxystrobin and pyridaben were both the matrix effect and extraction recovery. However, the ionization enhancement was observed in both cases. ME and RE in case of carbofuran were in an acceptable range (from −20% to 20%). In case of azoxystrobin, a strong enhancement of the analytical signal was observed (40.1% higher than the peak area of this compound in the neat solvent). In addition, low extraction recovery of such compounds was also observed. Interestingly, the overall recovery of this compound was still in an acceptable range (from 80 to 120%), according to SANTE Guideline of European Commission [23, 48]. The experimental results indicated that the matrix effect (ionization suppression/enhancement) had a major contribution to overall recovery of three representative analytes. The matrix effect, extraction recovery, and overall recovery of all target analytes are listed in Table S4a in Supplemental Materials. ME, RE, and R of all target analytes were in the range of 50–293%, 52–94%, and 38–209%, respectively. It should be concluded that the major contribution to the overall recoveries of almost all analytes was ionization suppression/enhancement. In addition, extraction recovery, matrix effect, and overall recovery of all target analytes in three individual vegetable matrices (cabbage, white mustard, and edible chrysanthemum) that are mostly planted in the winter in Vietnam were investigated in the same manner as the pooled sample. The results of ME, RE, and R are listed in Table S4b in Supplemental Materials. The experimental results indicated that ME, RE, and R in an individual matrix sample obtained the same characteristic as in the pooled sample. Therefore, the pooled sample of the individual matrix in the same commodity group could be proposed for assessment of ME, RE, and R.

Overall recovery of all target analytes in the pooled sample is shown in Figure 4. As clearly shown that an overall recovery of 70% of target analytes was in an acceptable range (from 80 to 120% overall recovery). Overall recovery of 18% and 12% of total number of target analytes was higher and lower than 120% and 80%, respectively.

Figure 4.

Figure 4

Overall recovery of all targeted analytes in pooled samples analyzed by UPLC-Orbitrap MS. Columns and error bars present the overall recovery and relative standard deviation (n=3), respectively. Red color indicates the recovery of the analyte out of an acceptable range (from 80 to 120%). (−) denotes an analyte that is measured in the negative electrospray ionization mode.

3.4. Analysis of Real Samples

For application, the in-house validated method was used for analysis of pesticides in vegetable samples collected from local villages in Hanoi, Vietnam. The QuEChERS sample preparation procedure was used for vegetable samples. Total ion chromatogram and extracted ion chromatograms of some pesticides found in the real samples are shown in Figure 5.

Figure 5.

Figure 5

TIC and EICs of pesticides found in edible chrysanthemum (a) and green salad (b).

As can be seen from Figure 5, only a few analytes were detected in the real samples. For instant, concentration of oxadiazon and pendimethalin in edible chrysanthemum and spinosad D in green salad was found to be 3.8 ng·g−1, 9.4 ng·g−1, and 12.5 ng·g−1, respectively. In addition, some pesticides were detected at below the limit of quantitation. However, the concentration of all detected analytes in the vegetable samples was below the maximum residual level according to USA-FDA, EU, and Vietnamese national regulation as well. Total ion and extracted ion chromatograms of detected analytes in some real samples are also shown in Figures S5 and S6 in Supplemental Materials.

4. Conclusion

An UHPLC-Q-Exactive Focus Orbitrap MS analytical method was successfully developed for multiclass pesticides (53 compounds and 30 classes) analysis in vegetable samples using a PFP column. The critical parameters of the analytical method such as linearity, correlation coefficients, LOD, and LOQ have been investigated and implemented. The excellent LOQ of the developed method was achieved. The ionization suppression or enhancement effect was also addressed. The experimental results indicated that the main contribution to overall recovery was the ionization effect on heated electrospray ionization sources. The developed method was applied to screen pesticides in some kinds of vegetables that were collected in local villages in Hanoi, Vietnam. Only few pesticides were found in these samples. Screening of pesticides and their metabolites in the vegetable samples will be focused by using in-house high-resolution mass spectra and MS/MS spectra in the next steps. The further sample clean-up steps will also be taken into account in order to minimize the matrix effect on the liquid chromatography high-resolution mass spectrometry.

Acknowledgments

This research was funded by the Vietnam Academic of Science and Technology (VAST) under Component 2 “Development of Identification Method for Toxics in Foods” (Subproject number: VAST.TĐ.TP.02/16-18) and Component 4 “Study of Methods for Testing Some Food Additives” (Subproject number VAST.TĐ.TP.04/16-18) in the research project “Research for Toxics Determination in Food Preservation and Process” (Project number: VAST.TĐ.TP/16-18).

Contributor Information

Nam Vu-Duc, Email: namvd@yahoo.com.

Dinh Binh Chu, Email: binh.chudinh@hust.edu.vn.

Data Availability

The LC-Orbitrap MS and MS/MS spectra data used to support the findings of this study are available from the corresponding authors upon request.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Supplementary Materials

Supplementary Materials

Table S1: mass accuracy and retention time of target analytes in neat solvent and matrix-matched solutions. Table S2: calibration curves and correlation of analytes on the UPLC-Orbitrap Q-Exactive MS. Table S3: analytical figures of merit of the UPLC-Q-Exactive Orbitrap MS method for pesticide analysis. Table S4a: matrix effect, recovery of extraction, and overall recovery of pesticides in pooled samples. Table S4b: matrix effect, recovery of extraction, and overall recovery of pesticides in three vegetable matrices (cabbage, white mustard, and edible chrysanthemum). Table S5: mass and mass accuracy of analytical product ions. Table S6: short- and long-term stabilities of the analytical signal in standard solution and in matrix-match solution. Figure S1: the total ion chromatogram of the target analytes in the standard solution (positive mode). Figure S2: the total ion chromatogram of the target analytes in the standard solution (negative mode). Figure S3: the total ion chromatogram and extracted ion chromatogram of the target analytes in the matrix-match solution (positive mode). Figure S4: the total ion chromatogram and extracted ion chromatogram of the target analytes in the matrix-match solution (negative mode). Figure S5: the total ion chromatogram and extract ion chromatogram of the target analytes in several real samples (positive mode). Figure S6: the total ion chromatogram and extracted ion chromatogram of the target analytes in several real samples (negative mode).

References

  • 1. EU legislation on MRLs - European Commission, https://ec.europa.eu/food/plant/pesticides/max_residue_levels/eu_rules/mrls_2019_en.
  • 2. US-FDA, Chemicals, Metals and Pesticides in Food, https://www.fda.gov/food/chemicals-metals-pesticides-food/pesticides.
  • 3. C. for F. S. and A. Nutrition, “Laboratory Methods - Pesticide Analytical Manual (PAM)”, https://www.fda.gov/food/laboratory-methods-food/pesticide-analytical-manual-pam.
  • 4. EU, “Legislation on Maximum Residue Levels – 2018 - European Commission, https://ec.europa.eu/food/plant/pesticides/max_residue_levels/eu_rules/mrls_2018_en.
  • 5. V. Ministry of Health, “Thông tư 50/2016/TT-BYT Quy định mức tối đa tồn dư thuốc bảo vệ thực vật trong thực phẩm, http://www.fsi.org.vn/pic/files/thong-tu-quy-dinh-gioi-han-toi-da-du-luong-thuoc-bao-ve-thuc-vat-trong-thuc-phamcompressed.pdf.
  • 6. MARD Vietnam, “Danh muc thuocs bảo vệ thực vật 2018,” 2018, https://www.mard.gov.vn/Pages/vbpq-toanvan.aspx?ItemID=130767.
  • 7.Ochiai N., Sasamoto K., Kanda H., et al. Optimization of a multi-residue screening method for the determination of 85 pesticides in selected food matrices by stir bar sorptive extraction and thermal desorption GC-MS. Journal of Separation Science. 2005;28(9-10):1083–1092. doi: 10.1002/jssc.200500017. [DOI] [PubMed] [Google Scholar]
  • 8.Taha S. M., Gadalla S. A. Development of an efficient method for multi residue analysis of 160 pesticides in herbal plant by ethyl acetate hexane mixture with direct injection to GC-MS/MS. Talanta. 2017;174:767–779. doi: 10.1016/j.talanta.2017.06.080. [DOI] [PubMed] [Google Scholar]
  • 9.Rajski Ł., Gómez-Ramos M. M., Fernández-Alba A. R. Application of LC-time-of-flight and orbitrap-MS/MS for pesticide residues in fruits and vegetables. Applications of Time-of-Flight and Orbitrap Mass Spectrometry in Environmental, Food, Doping, and Forensic Analysis. 2016;71:119–154. doi: 10.1016/bs.coac.2016.02.001. [DOI] [Google Scholar]
  • 10.Kruve A., Künnapas A., Herodes K., Leito I. Matrix effects in pesticide multi-residue analysis by liquid chromatography-mass spectrometry. Journal of Chromatography A. 2008;1187(1-2):58–66. doi: 10.1016/j.chroma.2008.01.077. [DOI] [PubMed] [Google Scholar]
  • 11.Kamel A. Refined methodology for the determination of neonicotinoid pesticides and their metabolites in honey bees and bee products by liquid chromatography−tandem mass spectrometry (LC-MS/MS) Journal of Agricultural and Food Chemistry. 2010;58(10):5926–5931. doi: 10.1021/jf904120n. [DOI] [PubMed] [Google Scholar]
  • 12.Picó Y., Blasco C., Font G. Environmental and food applications of LC-tandem mass spectrometry in pesticide-residue analysis: an overview. Mass Spectrometry Reviews. 2004;23(1):45–85. doi: 10.1002/mas.10071. [DOI] [PubMed] [Google Scholar]
  • 13.Dušek M., Jandovská V., Olšovská J. Analysis of multiresidue pesticides in dried hops by LC-MS/MS using QuEChERS extraction together with dSPE clean-up. Journal of the Institute of Brewing. 2018;124(3):222–229. doi: 10.1002/jib.490. [DOI] [Google Scholar]
  • 14.Lehotay S. J., Sapozhnikova Y., Mol H. G. J. Current issues involving screening and identification of chemical contaminants in foods by mass spectrometry. TrAC Trends in Analytical Chemistry. 2015;69:62–75. doi: 10.1016/j.trac.2015.02.012. [DOI] [Google Scholar]
  • 15.Alder L., Greulich K., Kempe G., Vieth B. Residue analysis of 500 high priority pesticides: better GC-MS or LC-MS/MS? Mass Spectrometry Reviews. 2007;25(6):838–865. doi: 10.1002/mas.20091. [DOI] [PubMed] [Google Scholar]
  • 16.Zanella R., Prestes O. D., Bernardi G., Adaime M. B. Advanced Sample Preparation Techniques for Pesticide Residues Determination by HRMS Analysis. Amsterdam, Netherlands: Elsevier; 2017. [Google Scholar]
  • 17.Gómez-Pérez M. L., Romero-González R., Vidal J. L. M., Frenich A. G. Identification of transformation products of pesticides and veterinary drugs in food and related matrices: use of retrospective analysis. Journal of Chromatography A. 2015;1389:133–138. doi: 10.1016/j.chroma.2015.02.052. [DOI] [PubMed] [Google Scholar]
  • 18.Polgár L., García-Reyes J. F., Fodor P., et al. Retrospective screening of relevant pesticide metabolites in food using liquid chromatography high resolution mass spectrometry and accurate-mass databases of parent molecules and diagnostic fragment ions. Journal of Chromatography A. 2012;1249:83–91. doi: 10.1016/j.chroma.2012.05.097. [DOI] [PubMed] [Google Scholar]
  • 19.López A., Yusà V., Millet M., Coscollà C. Retrospective screening of pesticide metabolites in ambient air using liquid chromatography coupled to high-resolution mass spectrometry. Talanta. 2016;150:27–36. doi: 10.1016/j.talanta.2015.11.068. [DOI] [PubMed] [Google Scholar]
  • 20.Jia W., Chu X., Ling Y., Huang J., Chang J. High-throughput screening of pesticide and veterinary drug residues in baby food by liquid chromatography coupled to quadrupole Orbitrap mass spectrometry. Journal of Chromatography A. 2014;1347:122–128. doi: 10.1016/j.chroma.2014.04.081. [DOI] [PubMed] [Google Scholar]
  • 21.Portolés T., Mol J. G. J., Sancho J. V., López F. J., Hernández F. Validation of a qualitative screening method for pesticides in fruits and vegetables by gas chromatography quadrupole-time of flight mass spectrometry with atmospheric pressure chemical ionization. Analytica Chimica Acta. 2014;838:76–85. doi: 10.1016/j.aca.2014.06.006. [DOI] [PubMed] [Google Scholar]
  • 22.Inostroza P. A., Wicht A.-J., Huber T., Nagy C., Brack W., Krauss M. Body burden of pesticides and wastewater-derived pollutants on freshwater invertebrates: method development and application in the Danube river. Environmental Pollution. 2016;214:77–85. doi: 10.1016/j.envpol.2016.03.064. [DOI] [PubMed] [Google Scholar]
  • 23.The European Comission. Sante/11813/2017 Analytical Quality Control and Method Validation Procedures for Pesticide Residues Analysis. Brussels, Belgium: The European Comission; 2018. [Google Scholar]
  • 24.Department of Health and Human Services Food and Drug Administration U.S. Guidance for Industry Bioanalytical Method Validation. Silver Spring, MD, USA: U.S. Department of Health and Human Services Food and Drug Administration; 2001. [Google Scholar]
  • 25.George R., Haywood A., Khan S., Radovanovic M., Simmonds J., Norris R. Enhancement and suppression of ionization in drug analysis using HPLC-MS/MS in support of therapeutic drug monitoring: a review of current knowledge of its minimization and assessment. Therapeutic Drug Monitoring. 2018;40(2):p. 284. doi: 10.1097/ftd.0000000000000471. [DOI] [PubMed] [Google Scholar]
  • 26.Trufelli H., Palma P., Famiglini G., Cappiello A. An overview of matrix effects in liquid chromatography-mass spectrometry. Mass Spectrometry Reviews. 2009;30:491–509. doi: 10.1002/mas.20298. [DOI] [PubMed] [Google Scholar]
  • 27.Kong C., Wang Y., Huang Y., Yu H. Multiclass screening of >200 pharmaceutical and other residues in aquatic foods by ultrahigh-performance liquid chromatography-quadrupole-Orbitrap mass spectrometry. Analytical and Bioanalytical Chemistry. 2018;410(22):5545–5553. doi: 10.1007/s00216-018-1124-5. [DOI] [PubMed] [Google Scholar]
  • 28.Niessen W. M. A. Group-specific fragmentation of pesticides and related compounds in liquid chromatography-tandem mass spectrometry. Journal of Chromatography A. 2010;1217(25):4061–4070. doi: 10.1016/j.chroma.2009.09.058. [DOI] [PubMed] [Google Scholar]
  • 29.96/23/Ec Commission Decision. 96/23/EC COMMISSION DECISION of 12 August 2002 Implementing Council Directive 96/23/EC Concerning the Performance of Analytical Methods and the Interpretation of Results (Notified under Document Number C(2002) 3044) (Text with EEA Relevance) (2002/657/EC) 96/23/Ec Commission Decision, 2002, https://publications.europa.eu/en/publication-detail/-/publication/ed928116-a955-4a84-b10a-cf7a82bad858/language-en. [Google Scholar]
  • 30.Peters F. T., Drummer O. H., Musshoff F. Validation of new methods. Forensic Science International. 2007;165(2-3):216–224. doi: 10.1016/j.forsciint.2006.05.021. [DOI] [PubMed] [Google Scholar]
  • 31.Kruve A., Rebane R., Kipper K., et al. Tutorial review on validation of liquid chromatography-mass spectrometry methods: part I. Analytica Chimica Acta. 2015;870(1):29–44. doi: 10.1016/j.aca.2015.02.017. [DOI] [PubMed] [Google Scholar]
  • 32.EMA. Guideline on bioanalytical method validation. Committee for Medicinal Products for Human Use. 2011-2012;44:1–23. [Google Scholar]
  • 33.Kruve A., Rebane R., Kipper K., et al. Tutorial review on validation of liquid chromatography-mass spectrometry methods: Part II. Analytica Chimica Acta. 2015;870:8–28. doi: 10.1016/j.aca.2015.02.016. [DOI] [PubMed] [Google Scholar]
  • 34.Si-Hung L., Causon T. J., Hann S. Comparison of fully wettable RPLC stationary phases for LC-MS-based cellular metabolomics. Electrophoresis. 2017;38(18):2287–2295. doi: 10.1002/elps.201700157. [DOI] [PubMed] [Google Scholar]
  • 35.Ashu-Arrah B. A., Glennon J. D., Albert K. Spectroscopic and chromatographic characterisation of a pentafluorophenylpropyl silica phase end-capped in supercritical carbon dioxide as a reaction solvent. Journal of Chromatography A. 2013;1298:86–94. doi: 10.1016/j.chroma.2013.05.024. [DOI] [PubMed] [Google Scholar]
  • 36.Bell D. S., Jones A. D. Solute attributes and molecular interactions contributing to “U-shape” retention on a fluorinated high-performance liquid chromatography stationary phase. Journal of Chromatography A. 2005;1073(1-2):99–109. doi: 10.1016/j.chroma.2004.08.163. [DOI] [PubMed] [Google Scholar]
  • 37.Lee J., Shin Y., Lee J., Lee J., Kim B. J., Kim J.-H. Simultaneous analysis of 310 pesticide multiresidues using UHPLC-MS/MS in brown rice, orange, and spinach. Chemosphere. 2018;207:519–526. doi: 10.1016/j.chemosphere.2018.05.116. [DOI] [PubMed] [Google Scholar]
  • 38.Casado J., Santillo D., Johnston P. Multi-residue analysis of pesticides in surface water by liquid chromatography quadrupole-Orbitrap high resolution tandem mass spectrometry. Analytica Chimica Acta. 2018;1024:1–17. doi: 10.1016/j.aca.2018.04.026. [DOI] [PubMed] [Google Scholar]
  • 39.Horwitz W. Evaluation of analytical methods used for regulation. Analytical Chemistry. 1982;54(1):525–530. doi: 10.1021/ac00238a765. [DOI] [Google Scholar]
  • 40.Gu H., Liu G., Wang J., Aubry A.-F., Arnold M. E. Selecting the correct weighting factors for linear and quadratic calibration curves with least-squares regression algorithm in bioanalytical LC-MS/MS assays and impacts of using incorrect weighting factors on curve stability, data quality, and assay performance. Analytical Chemistry. 2014;86(18):8959–8966. doi: 10.1021/ac5018265. [DOI] [PubMed] [Google Scholar]
  • 41.da Silva M. C., Goncalves de Oliveira M. L., Augusti R., Faria A. F. Simultaneous extraction of pesticides and polycyclic aromatic hydrocarbons in Brazilian cachaça using a modified QuEChERS method followed by GC-MS/MS quantification. Journal of Agricultural and Food Chemistry. 2018;67(1):399–405. doi: 10.1021/acs.jafc.8b04682. [DOI] [PubMed] [Google Scholar]
  • 42.Lee J., Shin Y., Lee J., Lee J., Kim B. J., Kim J.-H. Simultaneous analysis of 310 pesticide multiresidues using UHPLC-MS/MS in brown rice, orange, and spinach. Chemosphere. 2018;207:519–526. doi: 10.1016/j.chemosphere.2018.05.116. [DOI] [PubMed] [Google Scholar]
  • 43.Kunzelmann M., Winter M., Å.nte M., Hellenäs K.-E., Rosén J. Non-targeted analysis of unexpected food contaminants using LC-HRMS. Analytical and Bioanalytical Chemistry. 2018;410(22):5593–5602. doi: 10.1007/s00216-018-1028-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Hakme E., Lozano A., Uclés S., Fernández-Alba A. R. Further improvements in pesticide residue analysis in food by applying gas chromatography triple quadrupole mass spectrometry (GC-QqQ-MS/MS) technologies. Analytical and Bioanalytical Chemistry. 2018;410(22):5491–5506. doi: 10.1007/s00216-017-0723-x. [DOI] [PubMed] [Google Scholar]
  • 45.Lopardo L., Rydevik A., Kasprzyk-Hordern B. A new analytical framework for multi-residue analysis of chemically diverse endocrine disruptors in complex environmental matrices utilising ultra-performance liquid chromatography coupled with high-resolution tandem quadrupole time-of-flight mass spectrometry. Analytical and Bioanalytical Chemistry. 2018;411(3):689–704. doi: 10.1007/s00216-018-1483-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Schwanz T. G., Carpilovsky C. K., Weis G. C. C., Costabeber I. H. Validation of a multi-residue method and estimation of measurement uncertainty of pesticides in drinking water using gas chromatography-mass spectrometry and liquid chromatography-tandem mass spectrometry. Journal of Chromatography A. 2019;1585:10–18. doi: 10.1016/j.chroma.2018.11.058. [DOI] [PubMed] [Google Scholar]
  • 47.Matuszewski B. K., Constanzer M. L., Chavez-Eng C. M. Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC−MS/MS. Analytical Chemistry. 2003;75(13):3019–3030. doi: 10.1021/ac020361s. [DOI] [PubMed] [Google Scholar]
  • 48.Heywang W., Cross L. E. Guidelines on good laboratory practice in pesticide residue analysis. Giornale Italiano di Nefrologia. 2012;29(4):1–5. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Materials

Table S1: mass accuracy and retention time of target analytes in neat solvent and matrix-matched solutions. Table S2: calibration curves and correlation of analytes on the UPLC-Orbitrap Q-Exactive MS. Table S3: analytical figures of merit of the UPLC-Q-Exactive Orbitrap MS method for pesticide analysis. Table S4a: matrix effect, recovery of extraction, and overall recovery of pesticides in pooled samples. Table S4b: matrix effect, recovery of extraction, and overall recovery of pesticides in three vegetable matrices (cabbage, white mustard, and edible chrysanthemum). Table S5: mass and mass accuracy of analytical product ions. Table S6: short- and long-term stabilities of the analytical signal in standard solution and in matrix-match solution. Figure S1: the total ion chromatogram of the target analytes in the standard solution (positive mode). Figure S2: the total ion chromatogram of the target analytes in the standard solution (negative mode). Figure S3: the total ion chromatogram and extracted ion chromatogram of the target analytes in the matrix-match solution (positive mode). Figure S4: the total ion chromatogram and extracted ion chromatogram of the target analytes in the matrix-match solution (negative mode). Figure S5: the total ion chromatogram and extract ion chromatogram of the target analytes in several real samples (positive mode). Figure S6: the total ion chromatogram and extracted ion chromatogram of the target analytes in several real samples (negative mode).

Data Availability Statement

The LC-Orbitrap MS and MS/MS spectra data used to support the findings of this study are available from the corresponding authors upon request.


Articles from Journal of Analytical Methods in Chemistry are provided here courtesy of Wiley