Reliability of the Polar T31 Uncoded Heart Rate Monitor in Free Motion and Treadmill Activities

JEFFREY MONTES 1,, JAMES W NAVALTA 1,
PMCID: PMC6533094  PMID: 31156753

Abstract

The Polar T31 uncoded heart rate monitor (T31) is currently accepted as a precision measurement device that can be used in lieu of electrocardiography. However, minimal literature exists to verify reliability when used for this purpose. The purpose of this study is to evaluate the reliability of heart rate (HR) measurements obtained from the T31. Forty volunteers participated in a three-day test-retest protocol. Participants gave anthropometric data on the first day. On the second day, they performed two 5-minute self-paced free motion walks (FMW) and two 5-minute self-paced free motion jogs (FMJ). Finally, on the third day, they performed two 5-minute treadmill walks (TW) and two 5-minute treadmill jogs (TJ). Treadmill speeds for each TW and TJ were determined by the distances traveled during the FMWs and FMJs on day 2. HR values at rest (HRR) and while in motion (minutes 1–5) were evaluated (HRM). Values calculated used Cronbach’s α (≥0.70) for reliability with significance accepted at p<0.05. FMW HRR (all, α=0.96; male, α=0.97; female, α=0.94), FMJ HRR (all, α=0.98; male, α=0.98; female, α=0.99), TW HRR (all, α=0.96; male; α=0.92; female, α=0.98), TJ HRR (all, α=0.97; male, α=0.98; female, α=0.96), FMW HRM (all, α=0.93; male, α=0.93; female, α=0.91), FMJ HRM (all, α=0.93; male, α=0.94; female, α=0.90), TW HRM (all, α=0.95; male, α=0.92; female, α=0.95), TJ HRM (all, α=0.94; male, α=0.95; female, α=0.92). All p-values < 0.001. The T31 provided reliable measures at rest and for walking and jogging in both a free motion and treadmill setting.

Keywords: Test-retest, wearable technology, fitness

INTRODUCTION

The Polar T31 uncoded chest worn heart rate monitor (T31; Lake Success, NY, USA) is generally accepted as a precision heart rate (HR) measurement device. Many studies have utilized it for exercise evaluation purposes (6, 19), occupational applications (9, 35), medical research (31, 32), and for validating alternate heart rate monitors (4, 20, 21, 28). While the use of the T31 is common, the question remains whether it is a viable substitute for an electrocardiograph instrument (ECG). The ECG is the gold standard by which HR measurements are compared (16, 22, 37, 40). While considered the gold standard for laboratory use, ECGs are not feasible in all locations due to size, financial cost, or availability of trained personnel. For an alternate heart rate monitor to be used in lieu, certain criteria need to be established. First, it must measure what it claims to measure within an acceptable range. Second, it must be consistent. Third, it must be financially reasonable, and fourth, it needs to be user friendly. Finally, its use should minimally alter the behavior of the user as to not influence its measurement (23). Because the T31 is smaller, more cost effective, portable, and user friendly when compared to an ECG system, it has become the typically acceptable method for measuring HR when an ECG is not an option.

While the T31 is commonly used by investigators in many areas of study, there is only one known study that has assessed its accuracy. Bouts et al. compared the HR measured by a T31 and an ECG for 15 exercising individuals (4). The HR at nine different intensities of stationary cycling were compared. The Pearson’s Product Moment Correlation (r) between the two ranged from 0.976 to 1.00 with all p values < 0.05. While this comparison showed that the T31 can be considered accurate, to our knowledge no known research has shown that it is consistent in its measurements.

The purpose of this study was to determine if the T31 HR values could be considered reliable at rest (HRR) and during subsequent motion-based activities (HRM). We hypothesized that the T31 would be reliable for both HRR and HRM in free motion and treadmill activities for both the male and female genders.

METHODS

Participants

Young (N=40, male n=20, female n=20), healthy (identified as low risk according to the ACSM pre-participation screening questionnaire), individuals volunteered for this investigation (descriptive characteristics are provided in Table 1). Participants filled out an informed consent form that was approved by the institutional review board (#885569-3).

Table 1.

Participants characteristics.

Age (yrs) Height (cm) Mass (kg) BMI (kg/m2)
All participants 25.09±7.17 169.64±11.18 77.19±19.2 26.43±5.19
Male participants 24.46±3.07 176.47±8.32 83.56±15.02 26.27±4.26
Female participants 25.72±5.76 162.81±6.25 70.82±12.51 26.59±4.13

BMI = Body Mass Index

Protocol

On day one, participants reported to the University of Nevada, Las Vegas Exercise Physiology Lab (UEPL) and provided anthropometric data. Participant age (years) was self-reported, height (cm) was measured with a Health-o-meter wall mounted height rod (Pelstar LLC/Health-o-meter, McCook, IL), mass (kg) and Body Mass Index (BMI) was provided by a hand-and-foot bioelectric impedance analyzer (seca mBCA 514 Medical Body Composition Analyzer, Seca North America, Chino, CA).

On day two, participants were fitted with a T31 and a Polar CE0537 watch (PW) (Lake Success, NY). They then reported to a 200-foot section of building hallway that was marked off with cones. After a 5-minute seated rest period, HRR was recorded. Participants then completed the first 5-minute self-paced free motion walk (FMW-1) back and forth between the cones with HRM recorded every minute until completion. Following a 5-minute seated rest period, HRR was recorded and subjects then completed the first 5-minute self-paced free motion jog (FMJ-1) with HRM again being recorded every minute. Participants rested in a seated position for 10-minutes. They then performed a second self-paced 5-minute free motion walk (FMW-2) and jog (FMJ-2) with HRR and HRM recorded in the same manner. Distance traveled for FMW-1, FMW-2, FMJ-1, and FMJ-2 was measured. Speed in meters per minute (m•min−1) was calculated and rounded to the nearest 0.1 for all four conditions (miles per hour; [mph] noted after).

On day three, participants reported to the UEPL for treadmill-based walking and jogging. Both were performed on a Trackmaster treadmill (Full Vision, Inc. Newton, KS). Participants were again fitted with a T31 and PW. After a 5-minute seated rest period, HRR was recorded. They then completed the first 5-minute treadmill walk (TW-1) at the speed calculated from FMW-1. HRM was recorded every minute until completion. Following a 5-minute seated rest period, HRR was recorded and subjects then completed the first 5-minute treadmill jog (TJ-1) at the speed calculated from FMJ-1. HRM was again recorded every minute. Participants then rested in a seated position for 10-minutes. They then performed a second 5-minute treadmill walk (TW-2) and jog (TJ-2) with HRR and HRM recorded in the same manner as the first treadmill activities. Speeds for TW-2 and TJ-2 were calculated from FMW-2 and FMJ-2. The grade for all treadmill testing was set to 0%.

Statistical Analysis

Test-retest reliability was determined through Cronbach’s α using the Intraclass Correlation function in IBM SPSS (IBM Statistics version 24.0, Armonk, NY). Significance was accepted at the p < 0.05 level and considered to have acceptable reliability when ICC > 0.70. β was set at 0.80. Effect size was calculated using G*Power statistical software (G*Power version 3.1.9.2, Universität Kiel, Kiel, Germany). At the time of this study, there was no previous research data to calculate an “n” size. However, using the indicated α, β, and actual “n”, the calculated effect size for HR was 0.40.

RESULTS

When walking or jogging in either a free motion or treadmill setting was evaluated, all measurements were observed to have significant test-retest reliability and met the minimum threshold with HRR ranging from 0.92 to 0.98 (see Table 2) and HRM (minutes 1–5) ranging from 0.90 to 0.95 (see Table 3).

Table 2.

Heart rate at rest test-retest values.

HRR 1 HRR2 Cronbach’s α p-value
FMW 92±15 95±15 0.96 <0.001
All FMJ 94±17 97±17 0.98 <0.001
TW 88±13 92±14 0.96 <0.001
TJ 98±16 101±16 0.97 <0.001

FMW 87±15 90±16 0.97 <0.001
Male FMJ 89±16 91±15 0.98 <0.001
TW 83±9 86±11 0.92 <0.001
TJ 95±15 97±14 0.98 <0.001

FMW 97±14 100±13 0.94 <0.001
Female FMJ 100±16 105±16 0.99 <0.001
TW 94±14 99±14 0.98 <0.001
TJ 102±17 105±17 0.96 <0.001

HRR = Heart Rate at Rest, FMW = Free Motion Walk, FMJ = Free Motion Jog, TW = Treadmill Walk, TJ = Treadmill Jog

Table 3.

Heart rate while moving test-retest values. Minutes 1–5

HRM 1 Average Speed (1st) m•min−1 [mph] HRM 2 Average Speed (2nd) m•min−1 [mph] Cronbach’s α p-value
FMW 109±16 74.6±6.6 [2.8±0.2] 115±17 75.6±7.0 [2.8±0.3] 0.93 <0.001
All FMJ 156±15 130.0±17.7 [4.9±0.7] 160±14 130.9±19.6 [4.9±0.7] 0.93 <0.001
TW 107±16 74.6±6.6 [2.8±0.2]1 115±17 75.6±7.0 [2.8±0.3]3 0.95 <0.001
TJ 151±17 130.0±17.7 [4.9±0.7]2 160±14 130.9±19.6 [4.9±0.7]4 0.94 <0.001

FMW 102±14 73.8±7.1 [2.8±0.3] 108±16 75.2±7.6 [2.8±0.3] 0.93 <0.001
Male FMJ 153±15 128.0±15.9 [4.8±0.6] 158±15 128.9±17.5 [4.8±0.7] 0.94 <0.001
TW 102±12 73.8±7.1 [2.8±0.3]1 108±16 75.2±7.6 [2.8±0.3]3 0.92 <0.001
TJ 145±17 128.0±15.9 [4.8±0.6]2 158±15 128.9±17.5 [4.8±0.7]4 0.95 <0.001

FMW 116±14 75.3±5.9 [2.8±0.2] 122±16 76.1±6.3 [2.8±0.2] 0.91 <0.001
Female FMJ 159±13 132±19.5 [4.9±0.7] 163±13 132.8±21.5 [5.0±0.8] 0.90 <0.001
TW 113±18 75.3±5.9 [2.8±0.2]1 122±16 76.1±6.3 [2.8±0.2]3 0.95 <0.001
TJ 157±16 132±19.5 [4.9±0.7]2 163±13 132.8±21.5 [5.0±0.8]4 0.92 <0.001

HRM = Heart Rate in Motion, FMW = Free Motion Walk, FMJ = Free Motion Jog, TW = Treadmill Walk, TJ = Treadmill Jog.

1

Treadmill walking speed 1 based on the corresponding free motion speed from the first walk.

2

Treadmill jogging speed 2 based on the corresponding free motion speed from the first jog.

3

Treadmill walking speed 1 based on the corresponding free motion speed from the second walk.

4

Treadmill jogging speed 2 based on the corresponding free motion speed from the second jog.

DISCUSSION

The primary aim of this study was to examine the reliability of the T31 to measure HR at rest and while performing various activities in different settings. It is the first to demonstrate that the T31 is reliable when used by the general population under conditions such as walking and jogging, in both a free motion and treadmill setting. When evaluating all participants, the lowest HRR result was α=0.96 and lowest HRM was α=0.93. However, there are different physical characteristics when comparing males and females. On average, males weigh about 15% more than females and are taller by about 15cm (29). Male skeletons are also generally more massive and wider than females, especially in the rib cage (8). Because the T31 is worn on the sternum, measurements taken on a male may have been affected by these larger anatomical proportions. It was important to categorize our analysis by sex to account for these variances. Even when reliability was separated out to account for sex, we observed the lowest HRR for males and females to be α=0.92 and α=0.94 respectively. The lowest HRM for males and females was α=0.92 and α=0.90 respectively. We believe this study provides strong evidence that the reliability of the T31 is excellent for use by all.

To date, four manuscripts in 2018 utilizing the Polar T31 monitor are available through PubMed and the National Library of Medicine search, and 336 hits on articles in 2018 were found using Google Scholar. Previous research includes numerous studies involving the measurement of HR with a T31 for a multitude of purposes including human physiological research (2, 7, 13, 41), occupational related applications (12, 14, 24, 36), fitness assessment (1, 5, 17, 25), athlete evaluation (3, 15, 26, 30), setting exercise intensity (11, 27, 33, 34), and validation of alternate HR instruments (10, 18, 38). The popularity and use of the T31 make the results of the current investigation both timely and necessary.

While it is widely accepted as a precision measurement device for HR, we are aware of no other research, except for Bouts et al., that has reported on the validity or reliability of the T31 (4). As an ECG system is costly and requires some training to operate, the use of an acceptable substitute that is cost efficient, user-friendly, and easily purchased can be straightforwardly argued as desirable by those that do not have access to an ECG system. The T31 is lightweight, water-resistant, slender, and comes with a medium strap designed to fit chest sizes from 63.5cm to 137 cm (25″ to 54″). It currently retails for $39.95 (39). It is for these reasons that it had become one of the preferred alternates for the ECG.

The lack of published literature regarding the specific accuracy and consistency of the T31 is evident when scientific reference searches are performed. This is an important point since its use is widespread in many areas of research. Therefore, it was viewed as imperative by the authors to evaluate and report the current findings. This gives those that do use the T31 confidence based on solid statistical analysis. In concert with the findings reported by Bouts et al. researchers can continue to use the T31 as it is both valid and reliable in returning heart rate measures (4).

REFERENCES

  • 1.Alansare A, Afford K, Lee S, Church T, Jung HC. The effects of high-intensity interval training vs. moderate-intensity continuous training on heart rate variability in physically inactive adults. Int J Environ Res Public Health. 2018;15(7) doi: 10.3390/ijerph15071508. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Bartholomae EM, Moore J, Ward K, Kressler J. Sex differences in postprandial glucose response to short bouts of exercise: A randomized controlled trial. J Sci Med Sport. 2018 doi: 10.1016/j.jsams.2018.07.009. ahead of print. [DOI] [PubMed] [Google Scholar]
  • 3.Behan FP, Willis S, Pain MTG, Folland JP. Effects of football simulated fatigue on neuromuscular function and whole-body response to disturbances in balance. Scand J Med Sci Sports. 2018;28(12):2547–2557. doi: 10.1111/sms.13261. [DOI] [PubMed] [Google Scholar]
  • 4.Bouts AM, Brackman L, Martin E, Subasic AM, Potkanowicz ES. The accuracy and validity of iOS-based heart rate apps during moderate to high intensity exercise. Int J Exerc Sci. 2018;11(7):533–540. [PMC free article] [PubMed] [Google Scholar]
  • 5.Canino MC, Cohen BS, Redmond JE, Sharp MA, Zambraski EJ, Foulis SA. The relationship between soldier performance on the Two-Mile Run and the 20-m Shuttle Run Test. Military Medicine. 2018;183(5–6):E182–E187. doi: 10.1093/milmed/usx081. [DOI] [PubMed] [Google Scholar]
  • 6.Christopher JM, Beato M, Hulton AT. Manipulation of exercise to rest ratio within set duration on physical and technical outcomes during small-sided games in elite youth soccer players. Hum Mov Sci. 2016;(48):1–6. doi: 10.1016/j.humov.2016.03.013. [DOI] [PubMed] [Google Scholar]
  • 7.Cooke S, Samual TJ, Cooper SM, Stoohr EJ. Adaptation of myocardial twist in the remodelled athlete’s heart is not related to cardiac output. Exp Physiol. 2018;103(11):1456–1468. doi: 10.1113/EP087165. [DOI] [PubMed] [Google Scholar]
  • 8.Delavier F. Women’s strength training anatomy. 1st ed. Champaign: Human Kinetics; 2003. [Google Scholar]
  • 9.Dewhurst S, Bargh M, Davidson C, Bampouras T. Loaded and unloaded marching: Implications for fluid replacement. Open Sport Sci J. 2014;7(S1:M4):16–21. [Google Scholar]
  • 10.Dieli-Conwright CM, Jensky NE, Battaglia GM, McCauley SA, Schroeder ET. Validation of the 11. CardioCoachCO2 for submaximal and maximal metabolic exercise testing. J Strength Cond Res. 2009;23(4):1316–1320. doi: 10.1519/JSC.0b013e3181a3c5e8. [DOI] [PubMed] [Google Scholar]
  • 11.Dionne A, Perusse L, Comtois AS, Leone M. Acute cardiorespiratory responses in participants with heart disease during cycling at different immersion levels. Clin Physiol Funct Imaging. 2018;38(1):100–107. doi: 10.1111/cpf.12389. [DOI] [PubMed] [Google Scholar]
  • 12.Faghy M, Blacker S, Brown PI. Effects of load mass carried in a backpack upon respiratory muscle fatigue. Eur J Sport Sci. 2016;16(8):1032–1038. doi: 10.1080/17461391.2016.1202326. [DOI] [PubMed] [Google Scholar]
  • 13.Ferrandi PJ, Fico BG, Whitehurst M, Zourdos MC, Bao F, Dodge KM, Rodriguez AL, Pena G, Huang CJ. Acute high-intensity interval exercise induces comparable levels of circulating cell-free DNA and Interleukin-6 in obese and normal-weight individuals. Life Sci. 2018;(202):161–166. doi: 10.1016/j.lfs.2018.04.007. [DOI] [PubMed] [Google Scholar]
  • 14.Foulis SA, Canino MC, Hydren JR, Lacrom K, Sauers SE, Walker LA, Redmond JE, Frykman PN, Peter N, Smith LA, Warr BJ, Cohen BS, Zambraski EJ, Sharp MA. U.S. Army Research Institute of Environmental Medicine (0704–0188) 2018. A compendium of the physiological demands of the 32 critical physically demanding tasks of combat arms soldiers; pp. 1–93. [Google Scholar]
  • 15.Fox JL, Stanton R, Scanlan AT. A comparison of training and competition demands in semi-professional male basketball players. Res Q Exerc Sport. 2018;89(1):103–111. doi: 10.1080/02701367.2017.1410693. [DOI] [PubMed] [Google Scholar]
  • 16.Georgiou K, Larentzakis AV, Khamis NN, Alsuhaibani GI, Alaska YA, Giallafos EJ. Can wearable devices accurately measure heart rate variability? A systematic review. Folia Med (Plovdiv) 2018;60(1):7–20. doi: 10.2478/folmed-2018-0012. [DOI] [PubMed] [Google Scholar]
  • 17.Grace F, Herbert P, Elliot AD, Richards J, Beaumont A, Sculthorpe NF. High intensity interval training (HIIT) improves resting blood pressure, metabolic (MET) capacity and heart rate reserve without compromising cardiac function in sedentary aging men. Exp Gerontol. 2018;(109):75–81. doi: 10.1016/j.exger.2017.05.010. [DOI] [PubMed] [Google Scholar]
  • 18.Hiremath SV, Ding D. Evaluation of activity monitors in manual wheelchair users with paraplegia. J Spinal Cord Med. 2011;34(1):110–117. doi: 10.1179/107902610X12911165975142. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Johnson AM, Kurti SP, Smith JR, Rosenkranz SK, Harms CA. Effects of an acute bout of moderate-intensity exercise on postprandial lipemia and airway inflammation. Appl Physiol Nutr Metab. 2016;41(3):284–291. doi: 10.1139/apnm-2015-0314. [DOI] [PubMed] [Google Scholar]
  • 20.Johnstone JA, Ford PA, Hughes G, Watson T, Garrett AT. Bioharness multivariable monitoring device: part. I: validity. J Sports Sci Med. 2012;11(3):400–408. [PMC free article] [PubMed] [Google Scholar]
  • 21.Johnstone JA, Ford PA, Hughes G, Watson T, Mitchell AC, Garrett AT. Field based reliability and validity of the Bioharness multivariable monitoring device. J Sports Sci Med. 2012;11(4):643–652. [PMC free article] [PubMed] [Google Scholar]
  • 22.Kisilevsky BS, Brown CA. Comparison of fetal and maternal heart rate measures using electrocardiographic and cardiotocographic methods. Infant Behav Dev. 2016;(42):142–151. doi: 10.1016/j.infbeh.2015.12.004. [DOI] [PubMed] [Google Scholar]
  • 23.LaPorte RE, Montoye HJ, Caspersen CJ. Assessment of physical activity in epidemiologic research: problems and prospects. Public Health Rep. 1985;100(2):131–146. [PMC free article] [PubMed] [Google Scholar]
  • 24.Mac VV, Tovar-Aguilar JA, Flocks J, Economos E, Hertzberg VS, McCauley LA. Heat exposure in Central Florida fernery workers: Results of a feasibility study. J Agromedicine. 2017;22(2):89–99. doi: 10.1080/1059924X.2017.1282906. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Manttari A, Suni J, Sievanen H, Husu P, Vaha-Ypya H, Valkeinen H, Tokola K, Vasankari T. Six-minute walk test: a tool for predicting maximal aerobic power (VO2 max) in healthy adults. Clin Physiol Funct Imaging. 2018;38(6):1038–1045. doi: 10.1111/cpf.12525. [DOI] [PubMed] [Google Scholar]
  • 26.Milioni F, Leite JVM, Beneke R, de Poli RAB, Papoti M, Zagatto AM. Table tennis playing styles require specific energy systems demands. PLoS One. 2018;13(7):e0199985. doi: 10.1371/journal.pone.0199985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Millslagle D, DeLaRosby A, VonBank S. Incremental exercise in dynamic visual acuity. Percept Mot Skills. 2005;101(2):657–664. doi: 10.2466/pms.101.2.657-664. [DOI] [PubMed] [Google Scholar]
  • 28.Montes J, Stone TM, Manning JW, McCune D, Tacad DK, Young JC, Debeliso M, Navalta JW. Using Hexoskin wearable technology to obtain body metrics during trail hiking. Int J Exerc Sci. 2015;8(4):425–430. [PMC free article] [PubMed] [Google Scholar]
  • 29.Ogden CL, Fryer CD, Carroll MD, Flegal KM. Mean body weight, height, and body mass index, United States 1960–2002. Adv Data. 2004;(347):1–17. [PubMed] [Google Scholar]
  • 30.Peart AN, Nicks CR, Mangum M, Tyo BM. Evaluation of seasonal changes in fitness, anthropometrics, and body composition in collegiate division II female soccer players. J Strength Cond Res. 2018;32(7):2010–2017. doi: 10.1519/JSC.0000000000002578. [DOI] [PubMed] [Google Scholar]
  • 31.Radespiel-Troger M, Rauh R, Mahlke C, Gottschalk T, Muck-Weymann M. Agreement of two different methods for measurement of heart rate variability. Clin Auton Res. 2003;13(2):99–102. doi: 10.1007/s10286-003-0085-7. [DOI] [PubMed] [Google Scholar]
  • 32.Saemann L, Lachner K, Wenzel F. Cardiac frequency and cutaneous microcirculation during and after exercising in the view of physical condition. Clin Hemorheol Microcirc. 2017;67(3–4):221–227. doi: 10.3233/CH-179203. [DOI] [PubMed] [Google Scholar]
  • 33.Slater LV, Simpson AS, Blemker SS, Hertel J, Saliba SA, Weltman AL, Hart JM. Biomechanical adaptations during running differ based on type of exercise and fitness level. Gait Posture. 2018;60:35–40. doi: 10.1016/j.gaitpost.2017.11.012. [DOI] [PubMed] [Google Scholar]
  • 34.Slusher AL, Zuniga TM, Acevedo EO. Aerobic fitness alters the capacity of mononuclear cells to produce pentraxin 3 following maximal exercise. Eur J Appl Physiol. 2018;118(7):1515–1526. doi: 10.1007/s00421-018-3882-3. [DOI] [PubMed] [Google Scholar]
  • 35.Smith DM, McAuliffe J, Johnson MJ, Button DC, Behm DG. Seated inversion adversely affects vigilance tasks and suppresses heart rate and blood pressure. Occup Ergo. 2013;11(4):153–163. [Google Scholar]
  • 36.Steinman Y, van den Oord M, Frings-Dresen MHW, Sluiter JK. Flight performance during exposure to acute hypobaric hypoxia. Aerosp Med Hum Perform. 2017;88(8):760–767. doi: 10.3357/AMHP.4789.2017. [DOI] [PubMed] [Google Scholar]
  • 37.Svennberg Y, Stridh M, Engdahl J, Al-Khalili F, Friberg L, Frykman V, Rosenqvist M. Safe automatic one-lead electrocardiogram analysis in screening for atrial fibrillation. Europace. 2017;19(9):1449–1453. doi: 10.1093/europace/euw286. [DOI] [PubMed] [Google Scholar]
  • 38.Tanner EA, Montes J, Manning JW, Taylor JE, Debeliso M, Young JC, Navalta JW. Validation of Hexoskin biometric shirt to COSMED K4 b2 metabolic unit in adults during trail running. Sports Tech. 2016;8(3–4):118–123. [Google Scholar]
  • 39.T31 Transmitter. 2018. [09/24/2018]; Available from: https://www.polar.com/us-en/products/accessories/ T31 transmitter.
  • 40.Usadel L, Haverkamper G, Hermann S, Lober R, Weiss K, Opgen-Rhein B, Berger F, Will JC. Arrhythmia Detection in pediatric patients: ECG quality and diagnostic yield of a patient-triggered Einthoven lead-I event recorder (Zenicor EKG-2) Pediatr Cardiol. 2016;37(3):491–496. doi: 10.1007/s00246-015-1304-4. [DOI] [PubMed] [Google Scholar]
  • 41.Vosselman MJ, van der Lans AA, Brans B, Wierts R, van Baak MA, Schrauwen P, van Marken Lichtenbelt WD. Systemic beta-adrenergic stimulation of thermogenesis is not accompanied by brown adipose tissue activity in humans. Diabetes. 2012;61(12):3106–3113. doi: 10.2337/db12-0288. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from International Journal of Exercise Science are provided here courtesy of Western Kentucky University