Abstract
Importance
The real prevalence and clinical burden of severe neonatal jaundice are undefined due to difficulties in measuring total serum bilirubin (TSB) outside secondary and tertiary clinical centers.
Objective
To assess the diagnostic performance of the point-of care Bilistick System (BS) in identifying neonatal jaundice patients requiring treatment.
Design
Between April 2015 and November 2016, 1911 neonates, were recruited to participate in the study. Blood samples were simultaneously collected for the TSB determination by BS and by hospital laboratory (Lab). Data were collected and sent to the Bilimetrix headquarter in Trieste where statistical analysis was performed. Newborns with neonatal jaundice were treated with phototherapy according to each center's guidelines.
Setting
17 hospitals from Nigeria, Egypt, Indonesia, and Viet Nam.
Participants
1911 newborns were included, of which 1458 (76·3%) fulfilled the inclusion criteria.
Results
TSB level measured by BS agreed (p < .0001) with the lab result in all four countries. The diagnostic performance of BS showed a positive predictive value (PPV) of 92·5% and a negative predictive value (NPV) of 92·8%.
Conclusions and Relevance
BS is a reliable system to detect neonatal jaundice over a wide range of bilirubin levels. Since Bilistick is a point-of-care test, its use may provide appropriate and timely identification of jaundiced newborns requiring treatment.
Keywords: Neonatal jaundice, Severe hyperbilirubinemia, Neonatal screening, Bilirubin, Bilistick System, Point-of-care system, Diagnostic accuracy study, STARD, Low-medium income countries
Research in context.
Evidence before this study
Severe neonatal jaundice remains in low-income settings a life-threating condition, and the real dimension of the problem is largely unknown. We searched peer rewired articles in PubMed using the search terms “neonatal jaundice” OR “hyperbilirubinemia” OR “acute bilirubin encephalopathy" OR "kernicterus spectrum disorders" OR "kernicterus" OR "cerebral palsy" AND “epidemiology” OR “prevalence” OR “frequency” OR “surveillance” OR “screening” OR “low income” OR “resource poor” OR "mortality". We only included English-language studies. Although we found many recent studies that have reported important data on severe hyperbilirubinemia, ABE, and CBE in poor resource countries, the real prevalence and clinical burden of severe neonatal jaundice remain still undefined mainly because the information reported from LMICs are heterogeneous and largely drawn from tertiary hospitals. In the secondary and primary levels of health care, the dimension of the problem remains unknown due to the lack of affordable and reliable tools to diagnose of NNJ. This inability to promptly estimate total bilirubin level is a major obstacle to carry out an appropriate and timely treatment and eliminate this tragic and preventable life-threating condition.
Added value of this study
Our prospective and multicenter cohort study involved 17 hospitals from Nigeria, Egypt, Indonesia, and Vietnam between April 2015 and November 2016. We compared total serum bilirubin (TSB) assessed in clinical laboratory (Lab) with that measured by the Bilistick System (BS) a low-cost point-of-care assay. Our objective was to validate the performance of BS on measuring TSB and determining the prevalence of neonatal jaundice and need of treatment in real-world condition across different countries.
Implications of all the available evidence
The present study reveals the BS as a reliable point-of-care assay to detect neonatal jaundice over a wide range of bilirubin levels and determine appropriate treatment for jaundiced newborns to prevent development of acute bilirubin encephalopathy and its consequences.
Alt-text: Unlabelled Box
1. Introduction
A common condition in newborn infants is neonatal jaundice (NNJ) [1] characterized by high levels of Total Serum Bilirubin (TSB) concentrations. Under normal conditions, it resolves without treatment and with a benign outcome. International medical guidelines [2], [3], [4], [5] however recommend assessing the risk of developing severe hyperbilirubinemia in each newborn [6], [7] in order to identify those who need treatment to prevent development of acute bilirubin encephalopathy (ABE) and its long-term consequences, kernicterus spectrum disorders (KSD) [8], [9], [10].
The number of cases of severe hyperbilirubinemia in high-income countries (HICs) has decreased drastically since the implementation of these guidelines [11]. On the contrary, several studies have shown that severe hyperbilirubinemia remains a life-threating condition in many areas of the world, especially low middle-income countries (LMICs), though the real dimension of the problem is largely unknown [12].
In many LMICs there is paucity of accessible timely effective laboratory support services to promptly estimate bilirubin levels in jaundiced newborns [13], especially in the secondary and primary levels of health care. In many instances, results are obtained ≥ 24 h after testing leading to unnecessary delays in needed treatment, increasing the possibility of the patient needing an exchange transfusion (ET) or developing ABE. The introduction of inexpensive, easy to use, accurate point-of-care devices to measure bilirubin TSB that allow rapid, appropriate treatment of severe NNJ is an essential bridge in eliminating a major gap treatment of this disease problem [14].
The Bilistick System (Bilimetrix srl, Italy - BS) is a low cost point-of-care (POC) system that can be used as a screening diagnostic tool to estimate the TSB concentration, simplifying the triage of newborns and the evaluation of hyperbilirubinemia risk following discharge from birthing centers [15], [16]. One of the main advantages of the BS is the ability to carry the instrument where the TSB needs to be performed, i.e. physician offices, rural healthcare settings including village clinics and even homes [17], [18], [19].
The aim of the study was to test the diagnostic accuracy of the BS for the evaluation of the TSB concentration on whole blood of newborn infants in Africa and Asia, comparing it to the laboratory-based assay.
2. Methods
The study was conducted according to the STARD (Standards for the Reporting of Diagnostic Accuracy Studies) guidelines [20], [21]. We performed a prospective, multicenter, cohort study in 17 medical centers serving ethnically different populations distributed in Africa (1 in Egypt and 3 in Nigeria) and in Asia (7 in Indonesia and 6 in Viet Nam), for a total period of 20 months from April 1st, 2015 through November 30th, 2016 (eTable 1 in the Supplement).
Samples and clinical data were collected from consecutive inborn and outborn (referred) healthy ≥ 35 weeks gestation age and < 28 days of age, presenting NNJ at visual examination or ABE signs thus requiring a laboratory TSB measurement, or when TSB level was measured as part of routine screening [2], [8]. In these newborns, blood samples for TSB assay by both BS and clinical laboratory are simultaneously performed with the same sample for both tests. Age, sex, weight and hematocrit data were also recorded. No formal calculation of sample size was performed.
Participants without TSB result (BS and/or Lab TSB) or demographic data were excluded from the analysis. Other exclusion criteria were hematocrit outside the limits of the machine (< 25 or > 65%), TSB by Bilistick bilirubin level > 40 mg/dL or when technical problems occurred during the test.
The study was approved by each hospital ethics committee, and written or oral informed consent was obtained from the parents of the study newborns.
2.1. Test Methods
The Bilistick System (BS) is a simple, rapid, minimally invasive bilirubin assay that does not require the use of reagents, thus simplifying the process of measurement and reducing costs. It is able to give accurately measures blood samples with bilirubin level < 40 mg/dL and hematocrit (HCT) ranging from 25% to 65%. Operationally, nursing/medical staff collects 25 μL of blood, which is applied on a test strip previously inserted in the Bilistick reader. The strip separates the plasma from corpuscular components of the blood allowing the flow of serum onto the nitrocellulose membrane by capillarity. After saturation of the membrane, the TSB concentration is determined by reflectance spectroscopy within 2 min from loading (Video) [15], [16]. The technical problems which may occur during the testing are due to: 1) the fact that HCT is outside the 25%–65% range; and/or 2) the incorrect handling of sample resulting in haemolysis or undersaturation of the strip membrane or blood coagulation. The Bilistick Reader is programmed to detect when technical problems occur, and the error message appears on display.
The gold standard method for TSB measurement in serum samples is high performance liquid chromatography (HPLC) [22], [23], [24]. However, this method is used primarily in research labs due to its technical complexity which renders it unfeasible for clinical use [24], [25]. Clinically, the TSB is usually assessed by the diazo reaction method or direct spectrophotometry [5], [23], [26], [27]. Since these methods were most prevalent around the world, they were taken as reference standards for this study, according to the method available and normally used in each hospital laboratory (Lab) (eTable 1 in the Supplement). Quality controls of laboratory instruments were made according to each hospital's guidelines and clinical practice. Interlaboratory control was not performed.
The blood samples for Bilistick test and for the laboratory TSB determination were collected simultaneously from the same newborn usually by heel prick: the Bilistick TSB determination was performed immediately by the health personnel while the other sample was sent to the hospital laboratory. In order to prevent bilirubin photo-conversion, standard precautions were used to protect specimens from light exposure. The results obtained were registered for subsequent statistical analysis on a specific format. The laboratory did not know about the result obtained by BS.
2.2. Statistical Analysis
Comparison of TSB results given by Bilistick and laboratory assay were made on each pair of samples. Most variables were not Gaussian-distributed and all are reported as percentiles. Bland–Altman plots of the bias vs. the average were used to evaluate the presence of a proportional bias [28], [29]. Proportional bias was minimal or non-existent as detected by Pearson's correlation coefficient so that the Bland–Altman limits of agreement (LOA) were calculated. The bias was Gaussian-distributed and is reported as mean and standard deviation. Lin's concordance coefficient was also calculated to evaluate inter-method agreement.
To investigate the performance of Bilistick in predicting treatment needed, each TSB result was subject to analysis according to treatment threshold table proposed by NICE CG98 full guideline [2]. This table takes into account the bilirubin level and age of the neonate and suggests whether the level of jaundice requires treatment with phototherapy or exchange transfusion. TSB values obtained by laboratory assay were used as control for treatment prediction. Sensitivity, specificity, predictive values, likelihood ratios and ROC Area for the detection of neonates requiring phototherapy by Bilistick were calculated. Statistical analysis was performed using Stata 14.2 (Stata Corporation, College Station, TX, USA).
3. Results
3.1. Participants
Fig. 1 shows how subjects were enrolled and selected in each country. A total of 1911 infants presenting clinically with neonatal jaundice were selected for screening and 1854 were found eligible. A total of 1458 (75.3%) were analyzed. Three hundred and ninety-six subjects were excluded because they lacked TSB measured by Bilistick (n = 130, 33% of excluded), two for TSB > 40 mg/dL, 27 for technical reasons, and 237 (60% of excluded) for lack of laboratory data. Out of the remaining 1458 (76·3% of total), 79·5% (n = 1159) did not require treatment, 15.5% (n = 226) required phototherapy, and 5.0% (n = 73) required exchange transfusion.
Fig. 1.
Diagram reporting flow of participants through the study.
Table 1 reports the population stratified according to geographical origin (continent, country, and hospital), age and outcome of treatment if managed according to NICE CG98 guidelines [2], although actual treatment levels were determined by the treating doctor at each hospital. The majority of the newborns originated from Asia (79.6%) of which 54% from Viet Nam and 46% from Indonesia with the remaining 19.4% from Africa with 11.5% from Nigeria and 8.9% from Egypt.
Table 1.
Stratification of the population according to categorical measures.
| Total (n = 1458) |
||
|---|---|---|
| Continent | ||
| Africa | 298 (20.4%) | |
| Asia | 1160 (79.6%) | |
| State | ||
| Egypt | 130 (8.9%) | |
| Nigeria | 168 (11.5%) | |
| Indonesia | 530 (36.4%) | |
| Viet Nam | 630 (43.2%) | |
| Hospital | ||
| Egypt | E-CUCH | 130 (8.9%) |
| Nigeria | N-ABUTH | 58 (4.0%) |
| N-AKTH | 34 (2.3%) | |
| N-JUTH | 76 (5.2%) | |
| Indonesia | I-BADH | 76 (5.2%) |
| I-CIPTO | 154 (10.6%) | |
| I-JGH | 24 (1.6%) | |
| I-KDH | 103 (7.1%) | |
| I-KMH | 42 (2.9%) | |
| I-PRDH | 77 (5.3%) | |
| I-TDH | 54 (3.7%) | |
| Viet Nam | V-MCRGH | 92 (6.3%) |
| V-NBOPH | 154 (10.6%) | |
| V-QTPGH | 40 (2.7%) | |
| V-SPH | 87 (6.0%) | |
| V-VPOPH | 185 (12.7%) | |
| V-VSH | 72 (4.9%) | |
| AGE (NICE cut-point in hours) | ||
| 0 | 26 (1.8%) | |
| 6 | 4 (0.3%) | |
| 12 | 7 (0.5%) | |
| 18 | 11 (0.8%) | |
| 24 | 69 (4.7%) | |
| 30 | 17 (1.2%) | |
| 36 | 33 (2.3%) | |
| 42 | 18 (1.2%) | |
| 48 | 156 (10.7%) | |
| 54 | 29 (2.0%) | |
| 60 | 27 (1.9%) | |
| 66 | 10 (0.7%) | |
| 72 | 254 (17.4%) | |
| 78 | 15 (1.0%) | |
| 84 | 14 (1.0%) | |
| 90 | 10 (0.7%) | |
| 96 | 758 (52.0%) | |
| Nice management according to laboratory | ||
| Not requiring treatment | 1159 (79.5%) | |
| Perform phototherapy or ET | 299 (20.5%) | |
| Nice management according to Bilistick System | ||
| Not requiring treatment | 1230 (84.4%) | |
| Perform phototherapy or ET | 228 (15.6%) | |
All data are presented as frequency (percentage).
The demographic characteristics of the neonates are shown in Table 2 expressed as 50th percentile [25th–75th percentile]. The median age of participants was 96 h with a range of 1 h to 28 days, 56% of the newborns were male. TSB mean values (M ± SD) measured by either laboratory or Bilistick was 13.5 ± 5.0 and 13.0 ± 4.8 mg/dL, respectively.
Table 2.
Main measurements of the study subjects.
| Africa |
Asia |
Total |
|||
|---|---|---|---|---|---|
| Egypt |
Nigeria |
Indonesia |
Viet Nam |
||
| (n = 130) | (n = 168) | (n = 530) | (n = 630) | (n = 1458) | |
| Male-to-female ratio | 1.55: 1 | 1.30: 1 | 1.12: 1 | 1.36: 1 | 1.27: 1 |
| Age (hours) | 144 [78–195] |
96 [48–120] |
96 [72–168] |
72 [48–110] |
96 [58–144] |
| Weight (g) | 3200 [3000–3400] |
2900 [2500–3338] |
2470 [1811–3070] |
3000 [2700–3400] |
2950 [2405–3300] |
| Hematocrit (%) | 35.0 [31.9–40.1] |
44.0 [40.0–48.0] |
44.1 [38.3–48.7] |
43.0 [39.0–48.2] |
42.7 [38.0–48.0] |
| Bilirubin by Laboratory (mg/dL) | 12.9 [9.7–16.5] |
12.7 [8.8–15.5] |
13.6 [10.8–16.9] |
13.4 [11.2–16.4] |
13.4 [10.7–16.5] |
| Bilirubin by Bilistick (mg/dL) | 11.6 [8.2–14.5] |
12.0 [8.2–14.9] |
12.9 [10.2–16.1] |
12.5 [10.3–15.7] |
12.5 [10.0–15.7] |
All data are presented as 50th percentile [25th–75th percentile].
The prevalence of newborns needing treatment according to the NICE CG98 full guideline [2] varied according to country ranging from 15% in Egypt to 26% in Viet Nam (Table 3).
Table 3.
Estimates of diagnostic accuracy of Bilistick System.
| Africa |
Asia |
Total |
||||
|---|---|---|---|---|---|---|
| Egypt |
Nigeria |
Indonesia |
Viet Nam |
|||
| (n = 130) | (n = 168) | (n = 530) | (n = 630) | (n = 1458) | ||
| Prevalence | Pr(A) | 15.0% | 20.0% | 16.0% | 26.0% | 21.0% |
| [9.7%–22.8%] | [14.0%–26.5%] | [13.0%–19.4%] | [22.0%–29.2%] | [18.0%–22.7%] | ||
| Sensitivity | Pr(+|A) | 60.0% | 69.7% | 67.1% | 73.9% | 70.6% |
| [36.1%–80.9%] | [51.3%–84.4%] | [56.0%–76.9%] | [66.4%–80.5%] | [65.0%–75.7%] | ||
| Specificity | Pr(−|N) | 98.2% | 97.8% | 99.8% | 97.7% | 98.5% |
| [93.6%–99.8%] | [93.6%–99.5%] | [98.8%–100%] | [95.8%–98.8%] | [97.7%–99.1%] | ||
| ROC Area | (Sens. + Spec.)/2 | 0.791 | 0.837 | 0.834 | 0.858 | 0.846 |
| [0.680–0.902] | [0.757–0.918] | [0.784–0.884] | [0.823–0.893] | [0.819–0.872] | ||
| Likelihood ratio (+) |
Pr(+|A)/Pr(+|N) | 33.0 | 31.4 | 298.0 | 31.5 | 48.1 |
| [8.0–136.0] | [10.0–98.2] | [41.9–2126.0] | [17.4–56.9] | [29.8–77.6] | ||
| Likelihood ratio (−) |
Pr(−|A)/Pr(−|N) | 0.41 | 0.31 | 0.33 | 0.27 | 0.30 |
| [0.24–0.70] | [0.19–0.52] | [0.24–0.45] | [0.21–0.35] | [0.25–0.36] | ||
| Odds ratio | LR(+)/LR(−) | 81.0 | 101.0 | 904.0 | 118.0 | 161.0 |
| [16.7– .] | [27.1–371.0] | [151.0– .] | [59.4–234.0] | [94.3–275.0] | ||
| Positive predictive value |
Pr(A |+) | 85.7% | 88.5% | 98.3% | 91.5% | 92.5% |
| [57.2%–98.2%] | [69.8%–97.6%] | [90.8%–100%] | [85.4%–95.7%] | [88.3%–95.6%] | ||
| Negative predictive value |
Pr(N |−) | 93.1% | 93.0% | 94.1% | 91.6% | 92.8% |
| [86.9%–97.0%] | [87.4%–96.6%] | [91.5%–96.0%] | [88.8%–93.9%] | [91.3%–94.2%] | ||
All data are presented as value [95% Confidence Interval].
3.2. Test results
The comparison of TSB measured by Bilistick and Lab by Bland–Altman analysis is shown in Fig. 2. Overall the TSB measured by Bilistick was slightly lower than TSB by the laboratory. This difference was constant in the range of TSB measure (1 to 40 mg/dL). The LOA were in Indonesia − 4.0 to 2.3 mg/dL, with a mean ± SD of − 0.8 ± 1.6) mg/dL; in Viet Nam − 5.5 to 3.7 mg/dL, with a mean ± SD of − 0.9 ± 2.3 mg/dL; Egypt − 7.4 to 4.0 mg/dL, with a mean ± SD of − 1.7 ± 2.9 mg/dL, and in Nigeria − 7.5 to 6.0 mg/dL, with a mean ± SD of − 0.8 ± 3.4 mg/dL.
Fig. 2.
Bland–Altman analysis of Bilistick vs TSB. The lines are the mean difference and the limits of agreement (LOA).
The Lin's concordance coefficient showed a significant correlation (p < .0001) in Indonesia (0.94 [95% CI 0.93 to 0.95]), in Viet Nam (0·84 [95% CI 0.82 to 0.86]), Egypt (0.80 [95% CI 0.74 to 0.85]) and Nigeria (0.78 [95% CI 0.72 to 0.84]).
The diagnostic accuracy and precision analysis expressed as value [95% CI] of Bilistick for bilirubin measurement divided by country are reported in Table 3. The prediction of Bilistick to determine the need of phototherapy showed a sensitivity of 70.6% [65.0 to 75.7%], specificity of 98·5% [97.7 to 99.1%], positive predictive value (PPV) of 92.5% [88.3 to 95.6%] and negative predictive value (NPV) of 92.8% [91.3 to 94.2%].
4. Discussion
This study is an expansion of the data previously reported in Egypt and includes a large numbers of newborns from 4 different countries in Africa and Asia [16]. As in the previous study, the measurements of TSB were obtained at the same time by Bilistick and laboratory making possible assessment of the performance of the POC BS in different scenarios.
We observed that Bilistick underestimates TSB by about 1 mg/dL and this underestimation is stable over a wide range of TSB values ranging from 1 to 40 mg/dL. This bias may explain the rather low sensitivity (70.6%) we observed. We are now working to remove this systematic error which relates primarily to an error in the calibration curve. Never-the-less we confirmed Bilistick is a valuable and reliable POC device which may be used to determine TSB when laboratory measurements are either unavailable or the results excessively delayed.
The good concordance of results across different nations and continents is a major strength of this study. With all the local limitations and different health care systems, the reproducibility of TSB obtained with Bilistick confirmed that this device works well regardless of the environments where it is used. Based on these results Bilistick may be used in different field studies to define the actual prevalence of severe NNJ worldwide – Information that is still largely unknown [12], [30].
Our study has some limitations which must be considered. The first limitation is the lack of a comparable and standardized method for the measurement of TSB by the laboratories of the different hospitals in the different countries. It is possible that the reproducibility among and within the laboratories was less than ideal. The second limitation of our study is the lack of a priori determination of sample size due to the absence of reliable information of the prevalence of severe neonatal jaundice in most of the countries involved in the study. This preliminary information is essential to calculate positive and negative predictive values. According to the accepted guidelines [31], we did not perform a post-hoc sample size calculation, but we report instead the 95% confidence of the estimates as a means to evaluate their precision. The third limitation is the intrinsic limitation of Bilistick method. Additionally, the threshold maximum hematocrit of 65% (1.8%, eTable 2 in the Supplement) above which there is insufficient amount of plasma in the membrane for bilirubin determination and the skill of the operators in using the Bilistick was problematic in some of the newborns. Most of the newborns excluded for technical reasons (1.4%, eTable 2 in the Supplement) were enrolled early in the study suggesting that operators quickly became comfortable and proficient in the use of the reader and strip suggesting that the Bilistick is user friendly.
The following are the supplementary data related to this article.
Different operational steps to perform TSB measurement by BS.
Supplementary tables
Supplementary material
Abbreviations
- ABE
Acute Bilirubin Encephalopathy
- BS
Bilistick System
- ET
Exchange Transfusion
- HCT
Hematocrit
- HICs
High-Income Countries
- HPLC
High Performance Liquid Chromatography
- KSD
Kernicterus Spectrum Disorders
- LAB
Bilirubin test method available and normally used in each hospital laboratory
- LMICs
Low Middle-Income Countries
- LOA
Bland–Altman Limits Of Agreement
- NNJ
Neonatal Jaundice
- POC
Point-Of-Care
- STARD
Standards for the Reporting of Diagnostic Accuracy Studies
- TSB
Total Serum Bilirubin
Hospitals' abbreviations
- E-CUCH
Cairo University Children Hospital, Cairo, Egypt
- N-JUTH
Jos University Teaching Hospital, Jos, Nigeria
- N-ABUTH
Ahmadu Bello University Teaching Hospital, Zaria, Nigeria
- N-AKTH
Aminu Kano Teaching Hospital, Kano, Nigeria
- I-CIPTO
Cipto Mangunkusumo General Hospital, Jakarta, Indonesia
- I-BADH
Budhi Asih District Hospital, Jakarta, Indonesia
- I-KDH
Koja District Hospital, Jakarta, Indonesia
- I-TDH
Tarakan District Hospital, Jakarta, Indonesia
- I-PRDH
Pasar Rebo District Hospital, Jakarta, Indonesia
- I-KMH
RSUP Prof. Dr. R.D. Kandou Manado Hospital, Manado, Indonesia
- I-JGH
Prof. Dr. W.Z. Johannes General Hospital, Kupang, Indonesia
- V-QTPGH
Quang Tri Provincial General Hospital, QuangTri, Viet Nam
- V-MCRGH
Moc Chau Referral General Hospital, SonLa, Viet Nam
- V-SPH
Saint Paul Hospital, Hanoi, Viet Nam
- V-VPOPH
Vinh Phuc Obgyn & Pediatrics Hospital, VinhPhuc, Viet Nam
- V-NBOPH
Ninh Binh Obgyn & Pediatrics Hospital, NinhBinh, Viet Nam
- V-VSH
Vietnam–Sweden Hospital, QuangNinh, Viet Nam
Acknowledgments
Acknowledgment
The studies performed in Nigeria and Indonesia were possible through the generous support of the Saving Lives at Birth partners: the United States Agency for International Development (USAID), the Government of Norway, the Bill & Melinda Gates Foundation, Grand Challenges Canada, and the UK Government (FOG award N° AID-OAA-F-13-00072). The study in Viet Nam was realized with the contribution of Eric Hemel and Barbara Morgen (Scarsdale, USA), private donors of the East meets West Foundation (Oakland, CA, USA).
Contributors
C.G. conceptualized and designed the study, carried out the initial analyses, drafted the initial manuscript and reviewed and revised the manuscript. I.F.I. and S.Z.E.H. supervised data collection in Egypt and reviewed and revised the manuscript. R.R. and L.R. supervised data collection in Indonesia and reviewed and revised the manuscript. W.N.O. and A.O.D.O. supervised data collection in Nigeria and reviewed and revised the manuscript. L.M. and N.T.X.H. supervised data collection in Viet Nam and reviewed and revised the manuscript. G.B. designed and carried out the statistical analyses and reviewed and revised the manuscript. C.T. conceptualized and designed the study and critically reviewed the manuscript. C.D.C.Z. conceptualized and designed the study, coordinated and supervised data collection and critically reviewed the manuscript. All the other authors of the study team participated in the collection of the data. All authors approved the final manuscript as submitted and agree to be accountable for all aspect of the work.
Additional Members of the Study Team
EGYPT:
-
•
Dina M. Akmal
Department of Pediatrics, Cairo University Children Hospital, Cairo, Egypt
-
•
Dalia A. Khairy
Department of Pediatrics, Cairo University Children Hospital, Cairo, Egypt
INDONESIA:
-
•
Radhian Amandito
Neonatal Intensive Care Unit, Pondok Indah General Hospital, South Jakarta, DKI Jakarta, Indonesia
-
•
Marcelina Heidy Limanto
Neonatal Unit Dr. Cipto Mangunkusumo, Jakarta, Indonesia
-
•
Rosida Sihombing
Neonatal Unit Budhi Asih District Hospital, Indonesia
-
•
Meiharty Bahar Zulkifli
Neonatal Unit Budhi Asih District Hospital, Indonesia
-
•
Dewi Iriani
Neonatal Unit Koja General Hospital, Jakarta, Indonesia
-
•
Opy Dyah Paramita
Neonatal Unit Tarakan General Hospital, Jakarta, Indonesia
-
•
Ellen Sianipar
Neonatal Unit Pasar Rebo General District Hospital, Jakarta, Indonesia
-
•
Rocky Wilar
Neonatal Unit Prof. Dr. RD Kandou Hospital, Manado, Indonesia
-
•
Johnny Rompis
Neonatal Unit Prof. Dr. RD Kandou Hospital, Manado, Indonesia
-
•
Magdalena Kristi Daradjati Saudale
Neonatal Unit WZ Johannes General Hospital, Kupang, Indonesia
-
•
Woro Indri Padmosiwi Purba
Neonatal Unit WZ Johannes General Hospital, Kupang, Indonesia
NIGERIA:
-
•
Stephen Oguche
Department of Pediatrics, University of Jos, Jos University Teaching Hospital, Jos, Nigeria
-
•
Christian Isichei
Department of Chemical Pathology, University of Jos, Jos University Teaching Hospital, Jos, Nigeria.
-
•
Bose O. Toma
Department of Paediatrics, University of Jos, Jos University Teaching Hospital, Jos, Nigeria
-
•
David D. Shwe
Department of Paediatrics, University of Jos, Jos University Teaching Hospital, Jos, Nigeria
-
•
Fidelia Bode-Thomas
Department of Paediatrics, University of Jos, Jos University Teaching Hospital, Jos, Nigeria
-
•
Udochukwu Michael Diala
Department of Paediatrics, University of Jos, Jos University Teaching Hospital, Jos
-
•
Christopher Sabo Yilgwan
Department of Paediatrics, University of Jos, Jos University Teaching Hospital, Jos
-
•
Isa Abdulkadir
Neonatal Unit, Department of Paediatrics, Ahmadu Bello University/Teaching Hospital, Zaria, Nigeria
-
•
Laila Hassan
Neonatal Unit, Department of Paediatrics, Ahmadu Bello University/Teaching Hospital, Zaria, Nigeria
-
•
Fatima Abdullahi
Neonatal Unit, Department of Paediatrics, Ahmadu Bello University/Teaching Hospital, Zaria, Nigeria
-
•
Zubaida L Farouk
Department of Paediatrics Bayero University, Kano/Aminu Kano Teaching Hospital, Kano, Nigeria
-
•
Idris Yahaya Muhammed
Department of Chemical Pathology, Bayero University, Kano/Aminu Kano Teaching Hospital, Kano, Nigeria
-
•
Shehu Umar
Department of Paediatrics, Bayero University, Kano/Aminu Kano Teaching Hospital, Kano, Nigeria
-
•
Abdussalam Muhammed
Department of Paediatrics Bayero University, Kano/Aminu Kano Teaching Hospital, Kano, Nigeria
-
•
Fatima Tsiga-Ahmed
Department of Community Medicine, Bayero University Kano/Aminu Kano Teaching Hospital, Kano, Nigeria.
-
•
Laila Umar
Department of Paediatrics, Murtala Muhammed Specialist Hospital, Kano, Nigeria
VIET NAM:
-
•
Nguyen Thuy Duong
Newborn Department, Saint Paul Hospital, Hanoi, Viet Nam
-
•
Vu Thi La
Vinh Phuc Obgyn & Pediatrics Hospital, Vinh Phuc province, Viet Nam
-
•
Khuat Thanh Binh
Pediatrics Department, Moc Chau Referral General Hospital, Son La, Viet Nam
-
•
Nguyen Thi Kim Ngoc
Pediatrics Department, Ninh Binh Obgyn & Pediatrics Hospital, Ninh Binh, Viet Nam
-
•
Nguyen Thi Luyen
Newborn Department, Quang Tri Provincial General Hospital, Quang Tri, Viet Nam
-
•
Vu Thi Hao
Newborn Department, Vietnam Sweden Hospital, Quang Ninh province, Viet Nam.
Data availability
The full study protocol can be available as supplementary material. Dr. Coda Zabetta had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.
Conflicts of Interest
Dr. Tiribelli is the President and CEO of Bilimetrix s.r.l. and Drs. Greco and Coda Zabetta are employees of Bilimetrix s.r.l. In addition, Dr. Tiribelli and Dr. Coda Zabetta have a patent IT 2016000094056 pending, a patent IT 2016000094078 pending, and a patent PCT/IB2017/055657 pending. The other authors declare no conflicts of interest.
References
- 1.Smitherman H., Stark A.R., Bhutani V.K. Early recognition of neonatal hyperbilirubinemia and its emergent management. Semin Fetal Neonatal Med. 2006;11:214–224. doi: 10.1016/j.siny.2006.02.002. [DOI] [PubMed] [Google Scholar]
- 2.NICE National Institute for Health and Care Excellence Neonatal jaundice. 2010. https://www.nice.org.uk/guidance/cg98/evidence/full-guideline-245411821 [published online May]
- 3.Barrington D.K.J., Sankaran K., Canadian Paediatric Society Fetus and Newborn Committee Guidelines for detection, management and prevention of hyperbilirubinemia in term and late preterm newborn infants (35 or more weeks' gestation) – summary. Paediatr Child Health. 2007;12:401–407. doi: 10.1093/pch/12.5.401. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.World Health Organization . World Health Organization; 2013. Pocket book of Hospital Care for Children: guidelines for the management of common childhood illnesses. [PubMed] [Google Scholar]
- 5.Subcommittee on Hyperbilirubinemia. Management of Hyperbilirubinemia in the newborn infant 35 or more weeks of gestationPediatrics. 2004;114:297–316. doi: 10.1542/peds.114.1.297. [DOI] [PubMed] [Google Scholar]
- 6.AAP Subcommittee on Neonatal Hyperbilirubinemia Neonatal jaundice and kernicterus. Pediatrics. 2001;108:763–765. doi: 10.1542/peds.108.3.763. [DOI] [PubMed] [Google Scholar]
- 7.Bhardwaj K., Locke T., Biringer A. Newborn bilirubin screening for preventing severe hyperbilirubinemia and bilirubin encephalopathy: a rapid review. Curr Pediatr Rev. 2017;13:67–90. doi: 10.2174/1573396313666170110144345. [DOI] [PubMed] [Google Scholar]
- 8.Ives N.K. Management of neonatal jaundice. Paediatr Child Health. 2011;21:270–276. [Google Scholar]
- 9.Ramachandran A. Neonatal hyperbilirubinaemia. Paediatr Child Health. 2016;26:162–168. [Google Scholar]
- 10.Le Pichon J.-B., Riordan S.M., Watchko J., Shapiro S.M. The neurological sequelae of neonatal hyperbilirubinemia: definitions, diagnosis and treatment of the kernicterus spectrum disorders (KSDs) Curr Pediatr Rev. 2017;13:199–209. doi: 10.2174/1573396313666170815100214. [DOI] [PubMed] [Google Scholar]
- 11.Bhutani V.K., Johnson L. Kernicterus in the 21st century: frequently asked questions. J Perinatol. 2009;29(Suppl. 1):S20–S24. doi: 10.1038/jp.2008.212. [DOI] [PubMed] [Google Scholar]
- 12.Greco C., Arnolda G., Boo N.-Y. Neonatal jaundice in low- and middle-income countries: lessons and future directions from the 2015 Don Ostrow Trieste yellow retreat. Neonatology. 2016;110:172–180. doi: 10.1159/000445708. [DOI] [PubMed] [Google Scholar]
- 13.Anticona Huaynate C.F., Pajuelo Travezaño M.J., Correa M. Diagnostics barriers and innovations in rural areas: insights from junior medical doctors on the frontlines of rural care in Peru. BMC Health Serv Res. 2015;15:454. doi: 10.1186/s12913-015-1114-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Olusanya B.O., Ogunlesi T.A., Slusher T.M. Why is kernicterus still a major cause of death and disability in low-income and middle-income countries? Arch Dis Child. 2014;99:1117–1121. doi: 10.1136/archdischild-2013-305506. [DOI] [PubMed] [Google Scholar]
- 15.Coda Zabetta C.D., Iskander I.F., Greco C. Bilistick: a low-cost point-of-care system to measure total plasma bilirubin. Neonatology. 2013;103:177–181. doi: 10.1159/000345425. [DOI] [PubMed] [Google Scholar]
- 16.Greco C., Iskander I.F., Akmal D.M. Comparison between Bilistick System and transcutaneous bilirubin in assessing total bilirubin serum concentration in jaundiced newborns. J Perinatol. 2017;37:1028–1031. doi: 10.1038/jp.2017.94. [published online June 15] [DOI] [PubMed] [Google Scholar]
- 17.Ben M.D., Gazzin S., Tiribelli C. Neonatal hyperbilirubinemia. Ital J Pediatr. 2014;40:A10. [Google Scholar]
- 18.Olusanya B.O., Ogunlesi T.A., Kumar P. Management of late-preterm and term infants with hyperbilirubinaemia in resource-constrained settings. BMC Pediatr. 2015;15:39. doi: 10.1186/s12887-015-0358-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Arnolda G., Nwe H.M., Trevisanuto D. Risk factors for acute bilirubin encephalopathy on admission to two Myanmar national paediatric hospitals. Matern Health Neonatology Perinatol. 2015;1 doi: 10.1186/s40748-015-0024-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Bossuyt P.M., Reitsma J.B., Bruns D.E. STARD 2015: an updated list of essential items for reporting diagnostic accuracy studies. BMJ. 2015;h5527:351. doi: 10.1136/bmj.h5527. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Cohen J.F., Korevaar D.A., Altman D.G. STARD 2015 guidelines for reporting diagnostic accuracy studies: explanation and elaboration. BMJ Open. 2016;6 doi: 10.1136/bmjopen-2016-012799. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Bhutani V.K., Gourley G.R., Adler S., Kreamer B., Dalin C., Johnson L.H. Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia. Pediatrics. 2000;106 doi: 10.1542/peds.106.2.e17. [DOI] [PubMed] [Google Scholar]
- 23.Fevery J. Bilirubin in clinical practice: a review. Liver Int. 2008;28:592–605. doi: 10.1111/j.1478-3231.2008.01716.x. [DOI] [PubMed] [Google Scholar]
- 24.Kazmierczak S.C., Robertson A.F., Catrou P.G., Briley K.P., Kreamer B.L., Gourley G.R. Direct spectrophotometric method for measurement of bilirubin in newborns: comparison with HPLC and an automated diazo method. Clin Chem. 2002;48:1096–1097. [PubMed] [Google Scholar]
- 25.Randeberg L.L., Roll E.B., Nilsen L.T.N., Christensen T., Svaasand L.O. In vivo spectroscopy of jaundiced newborn skin reveals more than a bilirubin index. Acta Paediatr. 2005;94:65–71. doi: 10.1111/j.1651-2227.2005.tb01790.x. [DOI] [PubMed] [Google Scholar]
- 26.El-Beshbishi S.N., Shattuck K.E., Mohammad A.A., Petersen J.R. Hyperbilirubinemia and transcutaneous bilirubinometry. Clin Chem. 2009;55:1280–1287. doi: 10.1373/clinchem.2008.121889. [DOI] [PubMed] [Google Scholar]
- 27.Okwundu C.I., Uthman O.A., Suresh G., Smith J., Wiysonge C.S., Bhutani V.K. Cochrane database of systematic reviews. John Wiley & Sons, Ltd.; 2017. Transcutaneous bilirubinometry versus total serum bilirubin measurement for newborns. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Bland J.M., Altman D.G. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–160. doi: 10.1177/096228029900800204. [DOI] [PubMed] [Google Scholar]
- 29.Carstensen B. John Wiley & Sons; 2011. Comparing clinical measurement methods: A practical guide. [Google Scholar]
- 30.Slusher T.M., Zamora T.G., Appiah D. Burden of severe neonatal jaundice: a systematic review and meta-analysis. BMJ Paediatr Open. 2017;e000105:1. doi: 10.1136/bmjpo-2017-000105. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Walters S.J. Consultants' forum: should post hoc sample size calculations be done? Pharm Stat. 2009;8:163–169. doi: 10.1002/pst.334. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Different operational steps to perform TSB measurement by BS.
Supplementary tables
Supplementary material
Data Availability Statement
The full study protocol can be available as supplementary material. Dr. Coda Zabetta had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.