Objective:
To investigate the implications of using human tissue compositions for dose calculations in mice for kilovoltage photon beams.
Methods:
Dual energy CT (DECT) images of 9 female mice were used to extract the effective atomic number Zeff and the relative electron density ρe for each voxel in the images. To investigate the influence of the tissue compositions on the absorbed radiation dose for a typical kilovoltage photon beam, mass energy-absorption coefficients μen/ρ were calculated for 10 different tissues in each mouse.
Results
Differences between human and murine tissue compositions can lead to errors around 7.5 % for soft tissues and 20.1 % for bone tissues in μen/ρ values for kilovoltage photon beams. When considering the spread within tissues, these errors can increase up to 17.5 % for soft tissues and 53.9 % for bone tissues within only a single standard deviation away from the mean tissue value.
Conclusion:
This study illustrates the need for murine reference tissue data. However, assigning only a single mean reference value to an entire tissue can still lead to large errors in dose calculations given the large spread within tissues of μen/ρ values found in this study. Therefore, new methods such as DECT and spectral CT imaging need to be explored, which can be important next steps in improving tissue assignment for dose calculations in small animal radiotherapy.
Advances in knowledge:
This is the first study that investigates the implications of using human tissue compositions for dose calculations in mice for kilovoltage photon beams.
Introduction
Dedicated small animal treatment planning systems such as SmART-ATP can accurately calculate dose distributions in both tumors and surrounding healthy tissues for kilovoltage photon beams used in image-guided precision irradiation platforms.1–3 However, to accurately calculate a dose distribution in a medium using first-principles methods (e.g. Monte Carlo simulations), the elemental composition of the medium should be taken into account, since dose deposition is influenced by the atomic number Z. In the field of small animal radiotherapy, it is particularly important to use correct tissue compositions due to the low energies of the photons in the treatment beam.4 For low energy photons, photoelectric absorption is the dominant interaction. The atomic cross-section for photoelectric absorption depends strongly on Z (∝ Z4~5) and thus, in the case of tissues, the elemental composition.5 The Z dependence of the atomic cross-section for photoelectric absorption results in a Z dependence of the absorbed radiation dose (∝ Z3~4).
It is common practice to use human tissue compositions for dose calculations in mice, since detailed murine tissue compositions are not available in the literature. Very few studies have been published that present quantitative data on the elemental compositions of murine tissues. Most of these studies focus on trace elements,6–9 which only make up a small part of the total tissue composition. In addition to elemental compositions, mass densities ρ are another crucial factor in dose calculations. These mass densities are usually assigned using a cone beam CT (CBCT) image that is acquired before the treatment. For this method, a predetermined calibration curve is used to convert CT numbers to mass densities. Another method is to assign mass densities using the human reference data. The well-established reference data for human tissue compositions and mass densities are listed in several reports by the International Commission on Radiation Units and Measurements (ICRU) and the International Commission on Radiological Protection (ICRP).10–12
In this study, the dosimetric implications of using human tissue compositions for dose calculations in mice are investigated using dual energy CT (DECT) imaging, which is a novel imaging modality in the field of small animal radiotherapy that allows for in vivo extraction of the effective atomic number Zeff and the relative electron density ρe. Both Zeff and ρe can be resolved for each voxel, which also allows for an assessment of the spread in tissue composition and density within tissues. Zeff is defined as and ρe is defined as . ω and A represent the elemental weight fraction and mass number, respectively, for element i. ρ is the mass density, ρe,water is the electron density of water (≈ 0.555 mole electrons/cm3) and the exponent n is defined as 3.3, which was determined to be optimal in a previous study.13 To investigate the influence of tissue compositions on the absorbed radiation dose for a typical kilovoltage photon beam, the mass energy-absorption coefficient μen/ρ was calculated for 10 different tissues in a series of 9 mice. μen/ρ can be multiplied by the photon energy fluence to calculate the dosimetric quantity collision kerma, which is closely related to absorbed radiation dose. Assuming an approximately linear relationship between μen/ρ and the absorbed radiation dose for kilovoltage photon beams (i.e. the potential photon fluence perturbation by the different media is ignored), differences in μen/ρ are proportional to differences in absorbed radiation dose.
Methods and materials
Figure 1 shows the workflow of the methods used in this study. First, CBCT images were acquired with two different X-ray spectra, which were then registered before applying the DECT methods to obtain Zeff and ρe images. These images were used to calculate the means and standard deviations for both Zeff and ρe for 10 different tissues in each mouse. A noise correction was applied to the standard deviations which were then used to calculate the ±1 standard deviation ranges of Zeff and ρe to represent the spread in tissue composition and density, respectively. The mean tissue composition and its spread were then used to calculate the mean μen/ρ and its spread. The following sections explain these steps in greater detail.
Figure 1.
Workflow of the methods used in this study. The top half shows how the elemental compositions are calculated from the 50 and 90 kVp cone beam CT images and the bottom half shows how the mass-energy absorption coefficients are calculated from these compositions.
Image acquisition
The X-RAD 225Cx system (Precision X-ray, North Branford, CT) was used to image 9 female Naval Medical Research Institute nude mice.14 For each mouse, two CBCT images were acquired with two different peak voltages: 50 and 90 kVp. This combination was determined to be optimal in terms of Zeff and ρe errors, as shown in a previous study.15 The acquired images were reconstructed using the Feldkamp backprojection algorithm that is implemented in an in-house developed software platform for preclinical CBCT image reconstruction, built using the open source software RTK (Creatis, Lyon, France).16,17 The DECT imaging protocols are summarized in Table 1.
Table 1.
DECT imaging protocols
| Parameter | Low energy CT | High energy CT |
| Peak voltage (kVp) | 50 | 90 |
| Exposure (mAs) | 670.8 | 249.6 |
| Frame rate (s−1) | 10 | 5 |
| Panel mode | Mid Gain 1 × 1 | Low Gain 1 × 1 |
| Imaging dose (cGy) | 30 | 30 |
| Number of voxels | 1024 × 1024×1024 | |
| Voxel dimensions (μm3) | 100 × 100×100 | |
Image registration
Since the 50 and 90 kVp CBCT images were acquired consecutively, image registration was required to correct for motion of the animal between the two scans. A multiresolution rigid registration was applied first, followed by a multi resolution deformable registration. The open source software Elastix (Image Sciences Institute, Utrecht, Netherlands) was used to determine the deformation fields and the open source software Transformix (Image Sciences Institute, Utrecht, Netherlands) was used to apply them.18 The 50 kVp CBCT images were selected to be the fixed images and the 90 kVp CBCT images were selected to be the moving images in the image registration process. The image registration parameters are listed in appendix A (table A1, Supplementary Material 1).
Contouring
A total of 10 tissues were contoured by a radiation oncologist using the clinical treatment planning software Eclipse (Varian Medical Systems, Palo Alto, CA). The contoured tissues were brain, cortical bone, eye, femur, heart, kidney, liver, marrow, muscle and rib. The cortical bone and marrow tissues were contoured in the femur and the muscle tissue was contoured in the gluteal region. One rib was contoured for each mouse. On average, this was the 10.3th rib when counting from the head towards the tail. Separate contours were created for each tissue in each mouse. All tissues were contoured on the 50 kVp CBCT images, since these images provided better soft tissue contrast than the 90 kVp CBCT images.
Image processing
The tissue contours were converted into bitmasks using the poly2mask function in MATLAB R2016b (MathWorks, Natick, MA). An erosion with a spherical structuring element (radius 2 pixels) was applied to the bitmasks for all tissues in all mice to remove boundary voxels that might be affected by partial volume effects. The bitmasks were applied to the Zeff and ρe images to obtain Zeff and ρe values for each voxel in the contoured tissue. Mean values and standard deviations were calculated for the Zeff and ρe voxels in each bitmask of 10 tissues in all of the 9 mice.
Calibrations
The procedure to calibrate for Zeff and ρe was adopted from a previous study.15 A mouse-sized phantom with 12 cylindrical inserts (SmART Scientific Solutions, Maastricht, Netherlands) was used to establish a calibration curve for both Zeff and ρe. Zeff images were calculated using the tissue substitute method described by Landry et al and ρe images were calculated using a method described by Saito.13,19 Equation 1 was used to fit Zeff and equation 2 was used to fit ρe. In equation 1, μ is the mean linear attenuation coefficient for each phantom insert calculated from the measured CT numbers, Zeff is the reference effective atomic number calculated from the elemental compositions provided by the manufacturer and A, B, C, D, E, F, n and m are fit coefficients. In equation 2, HU is the mean CT number for each phantom insert, ρe is the reference relative election density calculated from the mass densities and elemental compositions provided by the manufacturer and a, b and α are fit parameters.
Zeff and ρe were combined to obtain elemental weight fractions using a method described by Hünemohr et al.20 In this method, weight fractions for the ith element were derived by a linear fit of Zeff, ρe and an interaction of both. The weight fractions for the individual elements were combined to obtain tissue compositions. Negative weight fractions were replaced by zero and the summed weight fractions were normalized to unity. In contrast to the fitting procedure for Zeff and ρe, (which was applied to the phantom tissues), the fitting procedure for the elemental weight fractions was applied to the 71 human reference tissues listed by Schneider et al.21 However, the lung tissue was excluded from the fitting procedure due to its aberrant values. A calibration curve (equation 3) was fitted through the 70 remaining points. A separate fit was performed for the soft tissues (Zeff <8.2) and the bone tissues (Zeff ≥8.2). In this equation, a, b, c and d are fit parameters. The index i iterates over the six most abundant elements in the human body, which are hydrogen (H), carbon (C), nitrogen (N), oxygen (O), phosphorus (P) and calcium (Ca).
Standard deviations
The standard deviations in Zeff and ρe represent the spread in tissue composition and density. However, the measured spread in Zeff and ρe is not solely caused by the heterogeneity of the tissues. When acquiring measurements of homogeneous materials such as phantom inserts, a small spread in Zeff and ρe can be observed, which is mainly caused by imaging noise in the CBCT images. Since this study focuses on the actual heterogeneity of the tissues (σtissue), the contribution of noise to the spread in Zeff and ρe needs to be removed first. To this end, data from a previous study were used in which the same imaging protocols were used and thus the same noise levels can be expected.15 From these data, standard deviations for eight homogeneous validation phantom inserts were obtained (σnoise). The validation phantom in the previous study contained 3 out of 12 empty insert positions and a Teflon insert which yielded aberrant results. The empty insert positions and the Teflon insert were left out. The standard deviations were then quadratically subtracted from the standard deviations that were obtained in the present study (σmeasured) using equation 4. In this equation, the mean standard deviation σnoise equals 0.30 for Zeff and 0.034 for ρe.
Mass energy-absorption coefficients
To investigate the influence of tissue compositions on the absorbed radiation dose, μen/ρ was calculated using the NIST tables of photon mass attenuation coefficients and mass energy-absorption coefficients.22 The 225 kVp spectrum for the X-RAD 225Cx system including 0.3 mm copper filtration was calculated using SpekCalc.23–25 This spectrum, which is assumed to be invariant throughout the mouse, has a mean photon energy of 86.1 keV and a half value layer of 0.975 mm copper or 11.6 mm aluminum. μen/ρ values for the 225 kVp spectrum were calculated for the individual elements (H, C, N, O, P and Ca), which were then multiplied by the elemental weight fractions and summed to obtain μen/ρ for the different tissue compositions.
Reference data
The human reference data used in this study were taken from ICRU and ICRP reports.10–12 In addition, two publications from the 1980’s by Woodard and White, which formed the basis for these ICRU and ICRP reports, were used in this study.26,27 Two entries for marrow tissues are listed in the reference data: one for yellow marrow and one for red marrow. For this study, the assumption from Schneider et al that marrow consists of a 1:1 mixture of yellow and red marrow was adopted.21 The reference data that are listed for the femur applies to the total bone, which is a mixture of bone and marrow. For the rib tissues, two entries are listed in the reference data: one for the 2nd and 6th rib and one for the 10th rib. The reference data for the 10th rib was used.
Results
The distributions of Zeff and ρe voxels in the examined murine tissues are shown in Figures 2 and 3, respectively. The crosses and diamonds represent the extracted mean values and the circles represent the reference values. The corrected standard deviations are indicated by the black error bars and the uncorrected standard deviations are indicated by the gray error bars. The black and gray error bars overlap for most of the bone tissues. The spread (indicated by the ±1 standard deviation ranges) within tissues is considerably larger than the spread of mean values between mice. This can be observed in both Figure 2 for Zeff and Figure 3 for ρe. For Zeff, a good agreement (<2.5% difference) was found between the extracted mean values and the reference values for brain, eye, heart, kidney, liver and muscle. Larger differences (>5.0%) were found for cortical bone, femur, rib and especially marrow (17.4%). The difference between extracted mean Zeff and reference Zeff was found to be positive for some tissues and negative for other tissues. A more one-sided difference was found for ρe. All tissues examined in this study, with the exception of cortical bone and rib, have a higher extracted mean ρe than reference ρe. On average, the ρe difference equals 6.8%. The largest errors were found for cortical bone (16.6%), marrow (12.6%) and rib (18.7%). A large spread in both Zeff and ρe can be observed between the different voxels of each tissue examined in this study. On average, the ±1 standard deviation ranges were found to be 1.5 for Zeff and 0.2 for ρe.
Figure 2.
Distribution of the effective atomic number. Top: mean ± 1 standard deviation ranges for the mice with the most extreme mean values for each tissue. Black and gray error bars indicate the corrected and uncorrected standard deviations, respectively. Bottom: comparison of the spread in effective atomic number (indicated by the ±1 standard deviation ranges) between mice and within tissues.
Figure 3.
Distribution of the relative electron density. Top: mean ± 1 standard deviation ranges for the mice with the most extreme mean values for each tissue. Black and gray error bars indicate the corrected and uncorrected standard deviations, respectively. Bottom: comparison of the spread in relative electron density (indicated by the ±1 standard deviation ranges) between mice and within tissues.
Figure 4 shows the mean μen/ρ values and their mean spread for the tissues examined in this study. For the soft tissues, μen/ρ was found to be larger for the extracted data than for the reference data. The opposite was found for the bone tissues. An average difference of 11.3% was found between the extracted data and the reference data, more specifically 7.5% for soft tissues (4.0% excluding marrow) and 20.1% for bone tissues. When considering the spread within tissues, errors ranging from −5.5% (eye) to +17.5% (marrow) were found for the soft tissues and errors ranging from –53.9% to +38.0% (both femur) were found for bone tissues. Both error ranges apply to the ±1 standard deviation interval, which only makes up 68% of the voxels in the case of a normal distribution.
Figure 4.
Distribution of the mass-energy absorption coefficients. The crosses represent the mean values (±1 standard deviation ranges) extracted using the dual energy CT method. The circles represent the mass-energy absorption coefficients calculated from the reference data.
Discussion
This work shows that differences between human and murine tissue compositions can lead to errors around 7.5% for soft tissues and 20.1% for bone tissues in μen/ρ values. Assuming an approximately linear relationship between μen/ρ and the absorbed radiation dose for kilovoltage photon beams (i.e. the potential photon fluence perturbation by the different media is ignored), differences in μen/ρ are proportional to differences in absorbed radiation dose. Therefore, errors around 7.5% for soft tissues and 20.1% for bone tissues can be expected in dose calculations for kilovoltage photon beams. When considering the spread within tissues, these errors can increase up to 17.5% for soft tissues and 53.9% for bone tissues within only a single standard deviation away from the mean tissue value. Even larger errors can be expected for softer X-ray spectra (i.e. when using lower tube voltages or thinner filters for the treatment beam) in which μen/ρ differs even more between different tissues.2 Although an overall good agreement was found between the extracted and reference Zeff, the strong dependence of the atomic cross-section for photoelectric absorption on Z leads to large differences in μen/ρ and thus in the absorbed radiation dose. The one-sided difference that was found between the extracted (murine) and reference (human) ρe could lead to errors in methods that rely on the reference data for densities. For example, errors in CT number predictions could occur when using the stoichiometric method described by Schneider et al.28 When assigning densities using the human reference data, the differences in ρe could also lead to errors in dose calculations.
The accuracy of the extracted Zeff and ρe is influenced by a number of different factors. One of these factors is imaging noise. In this study, the extracted standard deviations are corrected for the contribution of noise based on measurements in a phantom with homogeneous tissue-equivalent inserts. Another factor is beam hardening, which could be a possible explanation for the large difference found between the extracted and reference values for marrow. The marrow tissue is surrounded by dense cortical bone tissue, which could substantially harden the X-ray spectrum and thus influence the CT numbers from which Zeff and ρe are calculated. Other imaging artifacts, such as scatter, partial volume effects and motion of the animal could also compromise the accuracy of CT numbers. More research is required to investigate the influence of these factors on Zeff and ρe derived from DECT.
The discrepancies for the marrow tissue could also be caused by the fact that the assumption that marrow consists of a 1:1 mixture of yellow and red marrow is not accurate for mice. For the femur, any discrepancies between the measured and reference values could be caused by a difference in bone/marrow ratio between mice and humans. Rib tissue discrepancies could be caused by the fact that the different ribs in the thorax contain different amounts of cartilage. This could also explain the large difference between the mouse with the lowest value and the mouse with the highest value in Figures 2 and 3. Aside from being different to human tissues, murine tissues might also vary between different strains, genders and ages among others.
Uncertainties in tissue compositions lead to an additional uncertainty in dose calculations. As a first step to investigate the magnitude of this uncertainty, μen/ρ was calculated for 10 different murine tissues and compared to μen/ρ values that were calculated from the human reference data. To investigate the accumulated dose difference to an actual tumor, dose calculations are required to account for differences in absorption and attenuation of the treatment beam which are caused by differences in the elemental compositions and densities of the tissues through which the beam passes.
Conclusion
This study illustrates the need for murine reference tissue data. However, a large collection of data would be required to include all different strains, genders and ages among others. Moreover, assigning only a single mean reference value to an entire tissue can still lead to large errors in dose calculations given the large spread within tissues of μen/ρ values found in this study. Therefore, new methods such as DECT and spectral CT imaging need to be explored, which can be important next steps in improving tissue assignment for dose calculations in small animal radiotherapy.29
Footnotes
Competing interests: FV is a founder of SmART Scientific Solutions (Maastricht, Netherlands), which works with the company Precision X-ray (North Branford, CT) on commercial development of small animal treatment planning software.
Funding: AV acknowledges financial support from CAPES Scholarship and Science without Borders program from Brazil (BEX 12030/13–9).
Contributor Information
Lotte EJR Schyns, Email: lotte.schyns@maastro.nl.
Daniëlle BP Eekers, Email: danielle.eekers@maastro.nl.
Brent van der Heyden, Email: brent.vanderheyden@maastro.nl.
Isabel P Almeida, Email: isabel.dealmeida@maastro.nl.
Ana Vaniqui, Email: ana.vaniqui@maastro.nl.
Frank Verhaegen, Email: frank.verhaegen@maastro.nl.
REFERENCES
- 1.van Hoof SJ, Granton PV, Verhaegen F. Development and validation of a treatment planning system for small animal radiotherapy: SmART-Plan. Radiother Oncol 2013; 109: 361–6. doi: 10.1016/j.radonc.2013.10.003 [DOI] [PubMed] [Google Scholar]
- 2.Verhaegen F, van Hoof S, Granton PV, Trani D. A review of treatment planning for precision image-guided photon beam pre-clinical animal radiation studies. Z Med Phys 2014; 24: 323–34. doi: 10.1016/j.zemedi.2014.02.004 [DOI] [PubMed] [Google Scholar]
- 3.Verhaegen F, Granton P, Tryggestad E. Small animal radiotherapy research platforms. Phys Med Biol 2011; 56: R55–R83. doi: 10.1088/0031-9155/56/12/R01 [DOI] [PubMed] [Google Scholar]
- 4.Bazalova M, Graves EE. The importance of tissue segmentation for dose calculations for kilovoltage radiation therapy: Tissue segmentation for kilovoltage radiotherapy. Med Phys 2011; 38: 3039–49. doi: https://doi.org/10.1118/1.3589138 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.White DR. An analysis of the Z-dependence of photon and electron interactions. Phys Med Biol 1977; 22: 219–28. doi: 10.1088/0031-9155/22/2/003 [DOI] [PubMed] [Google Scholar]
- 6.Debernardi N. Microcalcifications in atherosclerotic lesion of apolipoprotein E-deficient mouse: Calcium deposits in murine atherosclerosis. Int J Exp Pathol 2010; 91: 485–94. doi: 10.1111/j.1365-2613.2010.00729.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Bäurle J, Kučera J, Frischmuth S, Lambertz M, Kranda K. Dynamics of trace element concentration during development and excitotoxic cell death in the cerebellum of Lurcher mutant mice. Brain Pathol 2009; 19: 586–95. doi: 10.1111/j.1750-3639.2008.00200.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Dey S, Arjun J, Das M, Bhattacharjee CR, Dkhar PS. Effect of prenatal lead toxicity on surface ultrastructural features, elemental composition and infrared absorption characteristics of the skin of albino mice. Cytobios 2001; 106(Suppl 2): 245–54. [PubMed] [Google Scholar]
- 9.Anniko M, Wroblewski R. Elemental composition of the developing inner ear. Ann Otol Rhinol Laryngol 1981; 90(1 Pt 1): 25–32. doi: 10.1177/000348948109000107 [DOI] [PubMed] [Google Scholar]
- 10.White DR, Booz J, Griffith RV, Spokas JJ, Wilson IJ. ICRU Report 44: tissue substitutes in radiation dosimetry and measurement. International Commission on Radiation Units and Measurements 10.1093/jicru/os23.1.Report44 1989. [DOI] [Google Scholar]
- 11.White DR, Griffith RV, Wilson IJ. ICRU Report 46: photon, electron, proton, and neutron interaction data for body tissues. International Commission on Radiation Units and Measurements 1992. doi: 10.1093/jicru/os24.1.Report46 [DOI] [Google Scholar]
- 12.Valentin J. Basic anatomical and physiological data for use in radiological protection: reference values: ICRP Publication 89: Approved by the Commission in September 2001. Annals of the ICRP 2002; 32: 1–277. doi: 10.1016/S0146-6453(03)00002-2 [DOI] [PubMed] [Google Scholar]
- 13.Landry G, Seco J, Gaudreault M, Verhaegen F. Deriving effective atomic numbers from DECT based on a parameterization of the ratio of high and low linear attenuation coefficients. Phys Med Biol 2013; 58: 6851–66. doi: 10.1088/0031-9155/58/19/6851 [DOI] [PubMed] [Google Scholar]
- 14.Clarkson R, et al. Characterization of image quality and image-guidance performance of a preclinical microirradiator: Characterization of a preclinical microirradiator. Med Phys 2011; 38: 845–56 10.1118/1.3533947. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Schyns LE, Almeida IP, van Hoof SJ, Descamps B, Vanhove C, Landry G, et al. Optimizing dual energy cone beam CT protocols for preclinical imaging and radiation research. Br J Radiol 2017; 90: 20160480. doi: 10.1259/bjr.20160480 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am A 1984; 1: 612. doi: 10.1364/JOSAA.1.000612 [DOI] [Google Scholar]
- 17.Rit S, Vila Oliva M, Brousmiche S, Labarbe R, Sarrut D, Sharp GC. The Reconstruction Toolkit (RTK), an open-source cone-beam CT reconstruction toolkit based on the Insight Toolkit (ITK). J of Phy: Conf Ser 2014; 489: 012079. doi: 10.1088/1742-6596/489/1/012079 [DOI] [Google Scholar]
- 18.Klein S, Staring M, Murphy K, Viergever MA, Pluim JP. elastix: a toolbox for intensity-based medical image registration. IEEE Trans Med Imaging 2010; 29: 196–205. doi: 10.1109/TMI.2009.2035616 [DOI] [PubMed] [Google Scholar]
- 19.Saito M. Potential of dual-energy subtraction for converting CT numbers to electron density based on a single linear relationship: Conversion of energy-subtracted CT number to electron density. Med Phys 2012; 39: 2021–30. doi: 10.1118/1.3694111 [DOI] [PubMed] [Google Scholar]
- 20.Hünemohr N, Paganetti H, Greilich S, Jäkel O, Seco J. Tissue decomposition from dual energy CT data for MC based dose calculation in particle therapy: Tissue decomposition from DECT for MC-based dose calculation. Med Phys 2014; 41061714. doi: 10.1118/1.4875976 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Schneider W, Bortfeld T, Schlegel W. Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions. Phys Med Biol 2000; 45: 459–78. doi: 10.1088/0031-9155/45/2/314 [DOI] [PubMed] [Google Scholar]
- 22.Hubbell JH, Seltzer SM. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients from 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest. National Institute of Standards and Technology (NIST). 2004. Available from: https://www.nist.gov/pml/x-ray-mass-attenuation-coefficients.
- 23.Poludniowski G, Landry G, DeBlois F, Evans PM, Verhaegen F. SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes. Phys Med Biol 2009; 54: N433–N438. doi: 10.1088/0031-9155/54/19/N01 [DOI] [PubMed] [Google Scholar]
- 24.Poludniowski GG, Evans PM. Calculation of x-ray spectra emerging from an x-ray tube. Part I. electron penetration characteristics in x-ray targets. Med Phys 2007; 34: 2164–74. doi: 10.1118/1.2734725 [DOI] [PubMed] [Google Scholar]
- 25.Poludniowski GG. Calculation of x-ray spectra emerging from an x-ray tube. Part II. X-ray production and filtration in x-ray targets. Med Phys 2007; 34: 2175–86. doi: 10.1118/1.2734726 [DOI] [PubMed] [Google Scholar]
- 26.Woodard HQ, White DR. The composition of body tissues. Br J Radiol 1986; 59: 1209–18. doi: 10.1259/0007-1285-59-708-1209 [DOI] [PubMed] [Google Scholar]
- 27.White DR, Woodard HQ, Hammond SM. Average soft-tissue and bone models for use in radiation dosimetry. Br J Radiol 1987; 60: 907–13. doi: 10.1259/0007-1285-60-717-907 [DOI] [PubMed] [Google Scholar]
- 28.Schneider U, Pedroni E, Lomax A. The calibration of CT Hounsfield units for radiotherapy treatment planning. Phys Med Biol 1996; 41: 111–24. doi: 10.1088/0031-9155/41/1/009 [DOI] [PubMed] [Google Scholar]
- 29.Trojanova E, Schyns LEJR, Ludwig D, Jakubek J, Pape AL, Sefc L, Dubois L, Le Pape A, et al. Tissue sensitive imaging and tomography without contrast agents for small animals with Timepix based detectors. J Instrum 2017; 12: C01056. doi: 10.1088/1748-0221/12/01/C01056 [DOI] [Google Scholar]




