A comparison of static and dynamic ∆B 0 mapping methods for correction of CEST MRI in the presence of temporal B 0 field variations

Esau Poblador Rodriguez 1, Philipp Moser 1, Barbara Dymerska 2, Simon Robinson 1, Benjamin Schmitt 3, Andre van der Kouwe 4, Stephan Gruber 1, Siegfried Trattnig 1,5, Wolfgang Bogner 1,
PMCID: PMC6563466  PMID: 30924210

Abstract

Purpose

To assess the performance, in the presence of scanner instabilities, of three dynamic correction methods which integrate ∆B 0 mapping into the chemical exchange saturation transfer (CEST) measurement and three established static ∆B 0‐correction approaches.

Methods

A homogeneous phantom and five healthy volunteers were scanned with a CEST sequence at 7 T. The in vivo measurements were performed twice: first with unaltered system frequency and again applying frequency shifts during the CEST acquisition. In all cases, retrospective voxel‐wise ∆B 0‐correction was performed using one intrinsic and two extrinsic [prescans with dual‐echo gradient‐echo and water saturation shift referencing (WASSR)] static approaches. These were compared with two intrinsic [using phase data directly generated by single‐echo or double‐echo GRE (gradient‐echo) CEST readout (CEST‐GRE‐2TE)] and one extrinsic [phase from interleaved dual‐echo EPI (echo planar imaging) navigator (NAV‐EPI‐2TE)] dynamic ∆B 0‐correction approaches [allowing correction of each Z‐spectral point before magnetization transfer ratio asymmetry (MTRasym ) analysis].

Results

All three dynamic methods successfully mapped the induced drift. The intrinsic approaches were affected by the CEST labeling near water (∆ω < |0.3| ppm). The MTRasym contrast was distorted by the frequency drift in the brain by up to 0.21%/Hz when static ∆B 0‐corrections were applied, whereas the dynamic ∆B 0 corrections reduced this to <0.01%/Hz without the need of external scans. The CEST‐GRE‐2TE and NAV‐EPI‐2TE resulted in highly consistent MTRasym values with/without drift for all subjects.

Conclusion

Reliable correction of scanner instabilities is essential to establish clinical CEST MRI. The three dynamic approaches presented improved the ∆B 0‐correction performance significantly in the presence of frequency drift compared to established static methods. Among them, the self‐corrected CEST‐GRE‐2TE was the most accurate and straightforward to implement.

Keywords:B0 mapping, chemical exchange saturation transfer (CEST), dynamic ∆B0 correction, frequency drift, scanner instabilities

1. INTRODUCTION

High and ultrahigh static magnetic fields (B 0) provide advantages that are critical for the quality of CEST MRI results. Not only the SNR, but also the chemical specificity are significantly improved, because the spectral separation between the resonances of interest increases. The CEST also benefits from a prolonged storage of saturation in bulk water due to longer T1, which facilitates its sensitivity. Furthermore, CEST works particularly well in the slow to intermediate exchange regime (k sw < ∆ω S, where k sw is the exchange rate from the labile proton pool to the bulk water pool and ∆ω S the solute proton pool frequency offset (which is proportional to B 0)). Under this condition, spectral resonances of interest can be much better distinguished from the direct saturation of water. This regime is increasingly met at higher B 0 even for rapidly exchanging protons.1, 2

On the other hand, local B 0 inhomogeneities (∆B 0) are more severe at higher B 0, causing regionally dependent frequency shifts in Z‐spectra. This complicates quantification in CEST. The applied frequency‐selective CEST saturation pulses at ∆ω RF are shifted away from the targeted nominal frequency offset (∆ω) by δω = ∆ω RF ‐ ∆ω = ∆B 0/γ (γ being the gyromagnetic ratio), which is proportional to the local ∆B 0.2 Even with optimized B 0 shimming protocols prior to the experiments, a retrospective ∆B 0 correction is generally needed, in which the Z‐spectrum of each voxel is centered at the water resonance (0 ppm). A highly accurate ∆B 0 correction is particularly important when studying exchanging compounds very close to water (e.g., glucose, lactate, myo‐inositol).3, 4, 5, 6 Zaiss et al. recently reported substantial pseudo‐CEST effects for dynamic CEST at 3 T from B 0 alterations (~1% per 7‐Hz drift).7

The simplest approach for a ∆B 0 correction is to determine the water resonance frequency from the Z‐spectrum (from here on this approach is termed “CEST‐minZ”).8, 9, 10 This works well only when the applied saturation power is low and the magnetization transfer contrast from semisolid macromolecules and CEST effect close to the water resonance can be neglected.8, 9, 11 In addition to this limitation, CEST‐minZ requires the full sampling of a high‐spectral‐resolution Z‐spectrum, which increases the scan time (i.e., the spectral range covering the water peak cannot be excluded). In vivo CEST measurements with higher saturation power and more accurate ∆B 0 correction can be achieved by acquiring an external ∆B 0 map. Water saturation shift referencing is probably the most widely used method to acquire such an external ∆B 0 map for CEST data processing. In WASSR, a matching CEST pulse sequence is acquired as a prescan, but with low saturation power and targeting only a narrow range of offsets around 0 ppm with high spectral resolution.12 Alternatively, a prescanned ∆B 0 map can be obtained from the phase difference of at least two gradient‐echo images acquired at different echo times (here termed “GRE‐2TE”).13, 14, 15, 16, 17, 18

All these three established ∆B0 correction methods, (A) CEST‐minZ, (B) WASSR, and (C) GRE‐2TE, estimate the water resonance frequency for a single time point and correct every z‐spectral point by applying the same shift (δω). This shift may not necessarily be representative for all points of the CEST spectrum, as they are acquired at different times. Consequently, all three methods share the drawback of being prone to errors due to temporal changes in the B 0 field over the course of a CEST experiment. Such temporal ∆B 0 fluctuations may arise from system instabilities [primarily heating of magnet’s gradient coils by heavy duty cycles19 or heating of passive shims20], from cardiac or respiratory effects21 or subject movement.22 Some previous studies performed at 3 T have reported drifts ranging from of 1.2 Hz/min to 5 Hz/min after a series of functional MRI or diffusion weighted imaging scans.23, 24, 25 Larger drifts are expected at higher B 0 (e.g., at 7 T periodic B 0 fluctuations of up to ~4 Hz due to respiration are already detectable even far away from the lungs).26, 27, 28, 29

In this paper, three dynamic ∆B 0 correction methods, which integrate the ∆B 0 mapping as part of the CEST sequence, are proposed; two of them via the phase generated by the CEST readout itself and the third by an interleaved 2D EPI navigator. They allow temporal fluctuations in B 0 to be mapped and compensated for each individual Z‐spectral point separately. The accuracy of this B 0 mapping and its impact on CEST correction are compared for these dynamic approaches and to three established static methods (i.e., CEST‐minZ, WASSR, and GRE‐2TE) for CEST analysis close to water.

2. THEORY

2.1. Chemical exchange in the presence of B 0 inhomogeneities

At present, the most common CEST quantification metric is the asymmetric magnetization transfer ratio (MTRasym). The purpose of CEST asymmetry analysis is to separate the asymmetric CEST contribution from the symmetric components (e.g., direct water saturation).30, 31, 32 This analysis is performed by subtraction of the magnitude signal on one side of the Z‐spectrum from its mirrored side1, 2:

MTRasymΔω=MTRΔω-MTR-Δω=Ssat-Δω-Ssat+ΔωS0, (1)

where S 0 and S sat correspond to reference and labeled magnitude signals, respectively. Here the MTRasym would be equal to the proton transfer ratio expression considered in the two‐pool exchange model if the water saturation were caused purely by exchange and in the absence of ∆B 0.33, 34, 35 However, when the frequency is shifted by δω, the proton transfer ratio can be redefined as36:

PTRΔω=η·PTRΔω=η·PTRasymΔω-ΔMTR, (2)

where η is a modulation factor that fully compensates the proton transfer ratio and ∆MTR is the MTR offset used to compensate MTRasym. The factor η can be further derived as:

η=αB1,ΔωS·1-σB1,ΔωSαB1,ΔωS+δω·1-σB1,ΔωS+δω (3)

It can be seen from Equation (3) that the modulation comes from the labeling coefficient α and spillover factor σ. In the denominator, both terms are dependent on the frequency shift δω with respect to the solute resonance frequency offset ∆ω S, making η strongly sensitive to ΔB 0. It has previously been shown that inaccurate ∆B 0 corrections can be erroneously interpreted as CEST effects, especially at frequencies close to the water resonance where the slope of the Z‐spectrum is steep because of direct saturation of bulk water.37

2.2. Static ∆B 0 correction

The three state‐of‐the‐art static approaches for ∆B 0 correction compared in this study are:

(A) CEST‐minZ

The water resonance frequency is intrinsically determined from the Z‐spectrum of each voxel (i.e., using magnitude images from a CEST experiment), by finding the minimum signal intensity value after a smoothing‐splines interpolation in the spectral domain.9, 10, 38

(B) WASSR

The WASSR uses an additional CEST sequence (i.e. prescan) with high spectral resolution over a narrow frequency range and low saturation power, in which CEST and MTC contributions are considered negligible, allowing the reference frequency to be estimated from the dominating direct water saturation. Similar to CEST‐minZ, WASSR utilizes only magnitude images, but the water resonance is determined by fitting the water peak with a Lorentzian curve, from which the central frequency of water can be estimated.12

(C) GRE‐2TE

The GRE‐2TE approach calculates the ∆B 0 maps from the difference of two gradient‐echo‐based phase images acquired at echo times TE1 and TE2 13:

ΔB0=ΔΦγ·ΔTE=ΦTE2GRE-ΦTE1GREγ·TE2-TE1, (4)

where ΦTEj,lGRE is the phase image acquired at echo time (j), by channel (l), in this case from the gradient‐echo pre‐scan (see Figure 1). For multichannel coils, the coil combination can be performed calculating the sum over the channels of the weighted channel‐wise phase difference (i.e., the sum Hermitian‐inner product)14:

ΔB0=l=1ChannelsMTE2,lGRE·MTE1,lGRE·eiΦTE2,lGRE-ΦTE1,lGREγ·TE2-TE1, (5)

Figure 1.

Figure 1

An overview of the sequence scheme for prescans and CEST measurement. The name of each ∆B 0 mapping method appears on the top face of each block. The blocks represent image acquisition or labeling modules. The static correction methods WASSR and GRE‐2TE use only prescans (labeled correspondingly) to calculate ∆B 0 maps. The CEST measurement is composed of a multishot 2D EPI navigator scan with dual‐echo readout; a CEST preparation in which the saturation RF pulses are applied at a range of frequency offsets depicted in the frequency axis of the Z‐spectra; and a gradient‐echo with dual‐echo readout. Magnitude and phase images from the interleaved navigator are used by the dynamic method NAV‐EPI‐2TE to generate a set of ∆B 0 maps per ∆ω. In addition to their use for CEST quantification, CEST‐weighted images generated by the postlabeling gradient‐echo readout are used for self‐correction by three correction methods: 1) calculating one ∆B 0 map using the static method CEST‐minZ (from the averaged magnitude images at TE1 and TE2); 2) computing a ∆B 0 map for each ∆ω by the dynamic method CEST‐GRE‐2TE (from magnitude and phase images at echo times TE1 and TE2), and 3) calculating a ∆B 0 map for each ∆ω by the dynamic method CEST‐GRE‐1TE (from magnitude and phase images at only the first echo time TE1). Note that CEST‐GRE‐1TE requires phase offset maps per channel (ΦRX,I) from the GRE prescan (indicated by dotted arrow). CEST, chemical exchange saturation transfer; CEST‐GRE‐2TE, chemical exchange saturation transfer‐gradient echo readout‐2TE; NAV, navigator

where symbolizes the angle of the complex data, MTE1,lGRE and MTE2,lGRE are the magnitude images, and ΦTE1,lGRE and ΦTE2,lGRE are the phase images for a given channel acquired at echo times TE1 and TE2.

2.3. Dynamic ∆B 0 correction

The three methods that we propose to use for CEST ∆B 0 correction integrate dynamic B0 estimation as part of the CEST sequence, construct a ∆B 0 map for each saturation frequency offset (∆ω) applied in the CEST labeling module, and allow the independent ∆B 0 correction of each individual Z‐spectral point.

(D) CEST‐GRE‐2TE

The CEST‐GRE‐2TE method calculates the ∆B 0 maps from multichannel dual‐echo data in the same way as the static method GRE‐2TE. The main difference is that the commonly used single‐echo readout of the CEST sequence is replaced by a dual‐echo readout. Thereby, the magnitude/phase from the prescan in Equation (5) is replaced by the CEST data, which are intrinsically generated immediately after the CEST‐labeling module (see Figure 1).

(E) CEST‐GRE‐1TE

Similar to CEST‐GRE‐2TE, this method derives ∆B 0 maps from phase data from the CEST sequence, although in this case only a single echo is needed. This is possible if we assume that the phase offset per channel (ΦRX,l) is time‐invariant, in which case a single estimation of ΦRX,l, calculated from a dual‐echo prescan (e.g., the same used for method GRE‐2TE) is sufficient. This has been demonstrated to hold even at 7 T and where there is significant head motion.39, 40 The phase offsets are calculated using:

ΦRX,l=TE1·ΦTE2,lGRE-TE2·ΦTE1,lGRETE1-TE2 (6)

This approach enables the calculation of dynamic ∆B 0 maps from single‐echo readouts by correcting the phase images per channel before calculating the weighted averaged phases over the channels. The ∆B 0 maps can hence be derived as:

ΔB0Δω=l=1ChannelsMTE1,lCEST-GREΔω·eiΦTE1,lCEST-GREΔω-ΦRX,lγ·TE1, (7)

where MTE1,lCEST-GREΔω and ΦTE1,lCEST-GREΔω are the magnitude and phase images per channel, acquired from the CEST readout at a single echo time for each frequency offset.

(F) NAV‐EPI‐2TE

In contrast to using the intrinsic magnitude/phase information of the CEST‐weighted images, the NAV‐EPI‐2TE method uses dual‐echo data in Equation (5), which are additionally acquired before each saturation module via a 2D multishot EPI navigator (see Figure 1).

3. METHODS

The accuracy of all six ∆B 0 correction methods and any possible bias on the intrinsic dynamic mapping methods (i.e., CEST‐GRE‐2TE and CEST‐GRE‐1TE) was first investigated in a homogeneous polydimethylsiloxane oil phantom (Siemens AG, Munich, Germany). In such a phantom, MTRasym should ideally be 0% in the absence of magnetization transfer (MT) from semisolid or CEST agents. Subsequently, each method was tested on five healthy volunteers (three males, two females; mean age 34 ± 4 years) after Ethics Committee approval by the Medical University of Vienna and informed consent was obtained.

All phantom scans were performed for a single slice with 1.7 × 1.7 × 6 mm3 spatial resolution over a field of view of 220 × 220 mm2 with a spectral resolution of 0.11 ppm. For the volunteer scans, the field of view was 270 × 270 mm2 with a resolution of 2.1 × 2.1 × 6 mm3, and the frequency offset increments were 0.15 ppm. Details on imaging parameters are listed in the following and in Table 1. Imaging parameters were matched wherever possible.

Table 1.

Comparison of main scan parameters for each of the ∆B 0 acquisition methods

B 0 mapping TR [ms] TE1 [ms] TE2 [ms] BW [Hz/Px] k‐space lines / shot Miscellaneous
STATIC CEST‐minZ 9.5 1.74 780 1 Ts = 700 ms, B1rms = 2.0 µT, ∆ω = 0.15/0.11 ppm
WASSR 4.5 1. 74 780 1 Ts = 100 ms, B1rms = 0.2 µT, ∆ω = 0.05/0.01 ppm
GRE‐2TE 9.5 1.74 5.16 780 1
DYNAMIC CEST‐GRE‐2TE 9.5 1.74 5.16 780 1 Ts = 700 ms, B1rms = 2.0 µT, ∆ω = 0.15/0.11 ppm
CEST‐GRE‐1TE 9.5 1.74 780 (390)* 1 Ts = 700 ms, B1rms = 2.0 µT, ∆ω = 0.15/0.11 ppm
NAV‐EPI‐2TE 15 5.4 9.0 2442 4
*

The experimentally used readout BW of 780Hz/Px stated here for CEST‐GRE‐1TE, could be in practice halved to 390Hz/Px, thereby matching the readout duration of two echoes of the CEST‐GRE‐2TE readout. Averaging these two echoes of the CEST‐GRE‐2TE readout with 780Hz/Px BW, should result in very similar SNR as the single CEST‐GRE‐1TE readout with 390Hz/Px BW.

3.1. Static ∆B 0 corrections

  • A

    To minimize the sensitivity to temporal instabilities, the Z‐spectral points were sampled with alternating saturation frequency offsets, decreasing from the maximum frequency to those close to water.

  • B

    For WASSR, a high spectral resolution with ∆ω steps of 0.05 ppm (≈15 Hz) over a frequency range of ±0.8 ppm was chosen, resulting in a total acquisition time of <4 min for in vivo and ∆ω = 0.01 ppm (≈3.6 Hz) with a total acquisition time <6 min for phantom measurements.

  • C

    For GRE‐2TE, the prescan was acquired in <1 s. To prevent possible phase errors from mistiming of readout gradients and the acquisition, both echoes were sampled under gradients of the same polarity (a.k.a. “monopolar”).

3.2. Dynamic ∆B 0 corrections

  • D

    For CEST‐GRE‐2TE, the more commonly used single‐echo readout of the CEST sequence was replaced by two readouts with doubled receiver bandwidth (e.g., 780 Hz/Px instead of 390 Hz/Px).

  • E

    The CEST‐GRE‐1TE could be acquired with no change to the conventional single‐echo readout of the CEST sequence. However, to prevent bias in the comparison, data from TE2 of the GRE readout post CEST labeling were simply ignored and the ∆B 0 mapping used only data from TE1. The required coil offset maps, which were assumed to be time‐invariant,39, 40 were obtained from the GRE pre scan.

  • F

    For NAV‐EPI‐2TE, a navigator was placed before each CEST‐labeling module as shown in Figures 1 and 2. The number of k‐space lines (i.e., echoes) collected in each shot of the EPI navigator was set to 4 to minimize geometric distortions and a readout bandwidth of 2442 Hz/Px used to reduce fat‐water chemical shift to 0.4 mm, resulting in a ~1‐s navigator duration. This could be further shortened by increasing the number of k‐space lines per shot (e.g., from 4 up to 128). The delay between the ∆B 0 mapping with the navigator and CEST data sampling illustrated in Figure 2 was assumed to be negligible.

Figure 2.

Figure 2

An overview of the experimental protocol followed for the comparison of six ∆B 0 correction methods (A). Scheme of the CEST measurement (B), in which a delay for T1 recovery of water signal is interleaved with S 0 and S sat acquisitions, each of them consisting of three blocks (C): an EPI navigator with dual‐echo readout scan generating magnitude and phase images for each TE (D); a CEST labeling period in which the magnetization is saturated by a train of saturation pulses at a frequency offset ∆ω for each Z‐spectral point; and a GRE sequence with dual‐echo readout scan generating magnitude and phase images for each echo time (E). CEST, chemical exchange saturation transfer; EPI, echo planar imaging; GRE, gradient‐echo;WASSR, water saturation shift referencing

3.3. Measurement protocols

All experiments were performed on a whole‐body 7 T MR Magnetom system (Siemens Healthcare, Erlangen, Germany) with a 1H 32‐channel head coil (Nova Medical, Wilmington, Massachusetts, USA). As illustrated in Figure 2A, after localization and B 0 shimming, three prescans were performed: 1) One WASSR measurement, in which the magnetization preparation was performed using one saturation pulse (B1rms = 0.2 µT) and 2) two monopolar dual‐echo gradient‐echo scans (duration <1 s), one prior to each CEST measurement, which generated magnitude and phase images for each channel and echo time.

To investigate the performance of the dynamic ∆B 0 correction methods versus the static ones in the presence of scanner instabilities, a scan‐rescan experiment was defined. The first CEST acquisition was performed with no deliberate changes to the imaging system, while in the second one a linear frequency drift of 60 Hz [consistent with previously reported drifts up to 5 Hz/min at 3 T 23, 24, 25]) was induced over the duration of the scan. The sequence was modified to apply a drift by updating the reference frequency in the analog‐to‐digital converters (ADCs) blocks of the navigator and CEST readouts for each ∆ω loop.

Each CEST scan (of duration 13 min) comprised three blocks (Figure 2): 1) a multishot EPI navigator with dual‐echo readout “NAV‐EPI‐2TE”; followed by 2) the CEST‐labeling module; and 3) a train of monopolar dual‐echo gradient‐echo readouts that covered the entire k‐space in one step (i.e., “CEST‐GRE‐2TE”). A subsequent delay of 10 s ensured T1 recovery of the water signal between acquisitions of different Z‐spectral points. The CEST labeling was executed by a train of four Gaussian pulses of 100‐ms duration, duty cycle of 50%, and B1rms = 2.0 µT. To study the bias of inaccurate ΔB 0 correction (which should be particularly strong closer to water) 61 spectral offsets were equidistantly distributed in the range from −4.5 ppm to 4.5 ppm (−4.5 ppm, +4.5 ppm, −4.35 ppm, +4.35 ppm, …, 0 ppm) for in vivo and 41 spectral offsets from −2.2 ppm to 2.2 ppm for phantom measurements. Magnitude and phase images were saved separately for each channel and echo.

3.4. Data analysis

All MR images were saved in DICOM format and data processing and evaluation were conducted retrospectively with MATLAB (R2017b, MathWorks, Natick, MA USA).

The resonance frequency of bulk water was determined voxel‐wise for the static methods CEST‐minZ and WASSR, as the minimum of the spline‐interpolated z‐spectra,9 and by least squares Lorentzian fitting [MATLAB function b0wassr.m available at http://godzilla.kennedykrieger.org/CEST/], respectively. For the ∆B 0 mapping methods (C)–(F) phase images from different coils were combined by applying Eqs. (5), (6), (7) and subsequently unwrapped via fast 2D phase unwrapping41 available at https://github.com/mfkasim91/unwrap_phase/. The phase offset maps required for the method CEST‐GRE‐1TE were masked and smoothed using a discretized spline smoother [MATLAB function smoothn.m42] to provide reliable results even at the brain’s boundaries, as has been shown previously for coil combination and distortion correction.40, 43, 44, 45 Finally, as the ∆B 0 maps were not masked, they were smoothed by a spatial Hamming filter before being used for CEST correction.

Each pair of magnitude images (i.e., from the CEST double‐echo readout) were averaged and ∆B 0 correction and subsequent MTRasym analysis [Equation (1)] were performed voxel‐wise. For static ∆B 0 correction methods (A)–(C), the estimated frequency shift δω was applied to center the entire Z‐spectra to 0 ppm, whereas for the dynamic methods (D)–(F) a time‐dependent δω(t) was applied to correct each Z‐spectral point independently.

To evaluate the B 0 estimation performance between different methods, a region of interest (ROI) was defined to compute ROI‐averaged B 0 and MTRasym curves (Figure 6). To evaluate the effect of ∆B 0 on the MTRasym maps, a ROI was manually drawn along phantom and volunteer’s brain boundaries and MTRasym mean and standard deviation were derived from voxels contained within these ROIs.

Figure 6.

Figure 6

Dynamic field mapping and its impact on the MTRasym within an ROI located in the white matter (WM) section of volunteer V3 (E). The ROI‐averaged B 0 estimation without (A) and with (B) an induced frequency drift of ~4.6 Hz/min. The static methods are represented by stars, while dynamic methods are depicted by solid lines. The times on the x axes are the times elapsed after the “GRE‐2TE” prescan. For illustration the time point of the WASSR acquisition was also set to t = 0. The offsets on the x axis ( ∆ω) refer to the frequencies at which saturation pulses were applied at the time instant when each B 0 sample was determined. Differences in MTRasym curves with different ∆B 0‐correction methods under unaltered (C) and drifted (D) conditions are shown in the bottom row. CEST‐GRE‐2TE (purple), CEST‐GRE‐1TE (green), and NAV‐EPI‐2TE (light blue) succeeded in mapping and correcting the induced drift ∆B 0(t). A good estimate for the precision of the B 0 mapping approaches can be derived from the standard deviation of the B 0 measurements in absence of an induced frequency drift (A). For CEST‐GRE‐2TE, assuming a Gaussian distribution, it can be stated that in 95% of all cases a B 0 change of ±1.92 Hz can be corrected. CEST‐GRE‐1TE, chemical exchange saturation transfer‐gradient echo‐1TE; MTRasym, asymmetric magnetization transfer ratio; NAV‐EPI‐2TE, navigator‐echoplaner imaging‐2TE; ROI, region of interest; WASSR, water saturation shift referencing

4. RESULTS

The accuracy of dynamic B 0 estimation via CEST‐GRE‐2TE and CEST‐GRE‐1TE was compromised when the CEST‐labeling pulses were applied close to water. Figure 3 shows how the ∆B 0 maps of phantom experiments were apparently corrupted by saturation RF trains applied at ∆ω < |0.33| ppm. On the other hand, NAV‐EPI‐2TE presented unaffected ∆B 0 maps over the whole saturation ∆ω range.

Figure 3.

Figure 3

Color‐coded ∆B 0 maps for the proposed dynamic methods CEST‐GRE‐2TE (first row), CEST‐GRE‐1TE (second row), and NAV‐EPI‐2TE (third row) acquired at seven saturation frequency offsets around the water resonance (0 ppm) at ∆ω = [‐0.33 to 0.33] ppm (columns). NAV‐EPI‐2TE is the only dynamic method insensitive to the RF saturation pulses applied for CEST labeling. Methods CEST‐GRE‐1TE and CEST‐GRE‐2TE are consistent with these results distant to the water resonance, but suffer from corrupted maps for ∆ω < |0.33| ppm (i.e., close to water). CEST‐GRE‐1TE, chemical exchange saturation transfer‐gradient‐echo‐1TE; EPI, echo planar imaging; GRE, gradient‐echo; NAV‐EPI‐2TE, navigator‐echoplanar imaging‐2TE

Figure 4 presents the accuracy of ∆B 0 maps for the different correction methods by evaluating deviations from the 0% MTRasym that was expected for a phantom containing no MTC or CEST agents. The GRE‐2TE generated the most homogeneous MTRasym map (i.e., although comparable to WASSR, the one that had the lowest spatial variability) among the static methods with minimal offset from 0% MTRasym compared to WASSR. Among the dynamic methods, CEST‐GRE‐2TE resulted in the most accurate MTRasym maps, with spatial variability similar to GRE‐2TE. The methods, CEST‐GRE‐1TE and NAV‐EPI‐2TE, on the other hand, led to slight spatial gradients in MTRasym maps, and ~40% to 60% higher variability than for CEST‐GRE‐2TE. The third row of Figure 2 illustrates that most of these differences become negligible for ∆ω = ±[1–2] ppm.

Figure 4.

Figure 4

The ∆B 0 mapping accuracy in terms of MTRasym in a homogeneous phantom: The first row presents the color‐coded ∆B 0 maps determined by the static methods (CEST‐minZ, WASSR, and GRE‐2TE) and the first ∆B 0 map from the dynamic methods (CEST‐GRE‐1TE, CEST‐GRE‐2TE, and NAV‐EPI‐2TE). Below color‐coded maps of ∆ωB 0‐corrected MTRasym values are shown for the integration range ∆ω of ± [0.3–1.0] ppm (second row) and ±[1–2] ppm (third row). Values within the black circular delineated ROI are expressed as mean ± standard deviation. Accurate ∆ωB 0 correction is indicated by homogeneous MTRasym maps with low values, as the phantom containing no MTC or CEST agents should result in MTRasym = 0%. Methods GRE‐2TE and CEST‐GRE‐2TE provided the most accurate MTRasym values among static and dynamic methods with similar low spatial variability. CEST, chemical exchange saturation transfer; EPI, echoplanar imaging; GRE, gradient‐echo; MTC, magnetization transfer contrast; MTRasym, asymmetric magnetization transfer ratio; NAV, navigator; ROI, region of interest; WASSR, water saturation shift referencing

For in vivo experiments, MT effects and other confounding factors cannot be neglected, so ∆B 0 mapping accuracy cannot be evaluated. However, rows 2 to 4 of Figure 5 show how the differences between correction methods gradually diminish the further the integration ranges ∆ω are from the water resonance. This indicates that in vivo CEST is sensitive to ∆B 0 over a wider range of ∆ω than the phantom experiments. The ∆B 0 corrections via static methods were generally inferior to those using the dynamic methods for ∆ω = ±[0.3–1.0] ppm (Figure 5). The CEST‐GRE‐1TE slightly underestimated B 0 compared to CEST‐GRE‐2TE, thereby producing artificial MTRasym increases, while the MTRasym maps corrected by NAV‐EPI‐2TE showed a gradient from posterior left to anterior right direction (consistent with the phantom experiments).

Figure 5.

Figure 5

Performance of each ∆B 0 correction method under unaltered conditions for volunteer V2: The first row shows the color‐coded ∆B 0 maps of the static correction methods (CEST‐minZ, WASSR, and GRE‐2TE) and the first ∆B 0 map from the dynamic methods (NAV‐EPI‐2TE, CEST‐GRE‐1TE, and CEST‐GRE‐2TE). Below, color‐coded maps of ∆B 0‐corrected MTRasym values for the frequency ranges |0.3–1.0|, |1–2|, and |3–4| ppm are displayed in the second, third, and fourth rows, respectively. The outline of the brain, defining voxels included in the mean MTRasym (see Section “Data analysis”), is indicated in black. The dynamic method CEST‐GRE‐1TE appears to estimate slightly lower B 0 compared to method CEST‐GRE‐2TE, producing a higher‐valued MTRasym map. The NAV‐EPI‐2TE shows the opposite effect (overestimation of B 0 compared to CEST‐GRE‐2TE) in the frontal region. CEST‐minZ, determination of the water resonance frequency from the Z‐spectrum; GRE‐2TE, MTRasym, gradient echo‐2TE‐asymmetric magnetization transfer ratio; NAV‐EPI‐2TE,navigator‐echoplanar imaging‐2TE; WASSR, water saturation shift referencing

Figure 6 shows the variation in B 0 over time and the effect of this on the corrected MTRasym curves within a ROI placed in a white matter (WM) region of volunteer V3, for both minimum field change and the imposed frequency drift. The dynamic mapping methods successfully followed the B 0 evolution and corrected each Z‐spectral point independently before the MTRasym analysis. All static methods, on the other hand, resulted in severely underestimated (~1/3 for CEST‐minZ) or overestimated (~3 times for WASSR and GRE‐2TE) MTRasym values.

The ΔB 0‐corrected color‐coded MTRasym maps derived from the scan‐rescan experiment are presented in Figure 7 for volunteer V4. For the CEST sequence with induced frequency drift, the MTRasymcontrast after ΔB 0 correction via CEST‐minZ was underestimated, while WASSR and GRE‐2TE ΔB 0 correction led to the opposite effect. On the contrary, all dynamic methods compensated for this drift efficiently. The CEST‐GRE‐2TE and NAV‐EPI‐2TE achieved MTRasymmaps with the most similar contrasts with and without induced frequency drift (ROI‐averaged errors of 0.001%/Hz and 0.002%/Hz drift, respectively), while CEST‐GRE‐1TE showed slightly higher deviation (e.g., error of 0.006%/Hz), but was still superior to all the static approaches.

Figure 7.

Figure 7

Effect of a known linear frequency drift on the MTRasym maps of volunteer V4: The top row shows color‐coded maps for the different correction methods in unaltered conditions (an apparent drift of ~6 Hz was measured between the GRE‐2TE prescan and the CEST acquisition at ∆ω = ±1 ppm); the bottom row depicts maps for which a ~4.6‐Hz/min drift was induced. The static method CEST‐minZ overestimated B 0, resulting in decreased MTRasym values, while WASSR and GRE‐2TE show the opposite effect, since both scans were performed before the field drift was applied. Among the dynamic methods, CEST‐GRE‐2TE and NAV‐EPI‐2TE compensate for this drift most efficiently, providing very close MTRasym values within the delineated ROI (in black) between acquisitions, followed by CEST‐GRE‐1TE. CEST‐GRE‐2TE, chemical exchange saturation transfer‐gradient echo‐2TE; CEST‐minZ; determination of the water resonance frequency from the Z‐spectrum; MTRasym, asymmetric magnetization transfer ratio; NAV‐EPI‐2TE, navigator echoplanar imaging‐2TE; ROI, region of interest; WASSR, water saturation shift referencing

The results of the comparison among all five healthy volunteers (V1‐V5) for the dynamic ΔB 0 correction methods CEST‐GRE‐2TE and NAV‐EPI‐2TE are presented in Figure 8. The MTRasym maps derived by CEST‐GRE‐2TE and NAV‐EPI‐2TE were highly consistent between acquisitions with/without artificially induced frequency drift for all subjects.

Figure 8.

Figure 8

Comparison of the correction performance of the proposed dynamic methods CEST‐GRE‐2TE (rows 1 and 2) and NAV‐EPI‐2TE (rows 3 and 4) for subjects V1‐V5: ∆B 0‐corrected MTRasym (0.3–1.0 ppm) maps in the absence (first and third rows) and in the presence (second and fourth rows) of the induced B 0 drift during each CEST measurement. A ROI along each volunteer's brain border was manually drawn (delineated in black) and the MTRasym mean ± standard deviation values within this ROI are shown at the bottom of each map. Among the dynamic methods, both CEST‐GRE‐2TE and NAV‐EPI‐2TE generated highly consistent MTRasym contrasts between scans in all cases. CEST‐GRE‐2TE, chemical exchange saturation transfer‐gradient echo‐2TE; MTRasym, asymmetric magnetization transfer ratio; NAV‐EPI‐2TE, navigator‐echoplanar imaging‐2TE

5. DISCUSSION

In this study, we have investigated the applicability of three dynamic ∆B 0 mapping methods for correction of CEST, which integrate ∆B 0 mapping in the CEST measurement. In contrast to static correction methods, each Z‐spectral point can be adjusted independently to compensate for temporal B 0 changes such as those arising from system instabilities. Static and dynamic correction performance were evaluated and compared focusing on the ΔB 0 mapping accuracy and bias of MTRasym maps in the human brain at 7 T, first in the absence and then in the presence of a frequency drift.

Previous studies have evaluated the accuracy of static ∆B 0 correction of CEST‐weighted maps by multiecho methods,15, 16, 17, 18 while studies proposing dynamic methods to correct for temporal ∆B 0 changes have only recently emerged.46, 47

Windschuh et al. proposed a method to correct each Z‐spectral point independently retrospectively, combining phase images from a single‐echo GRE CEST readout and a prescan to calculate relative ∆B 0 maps (WASABI).46 However, development work is still needed to improve the stability of this approach, which was significantly affected by the RF saturation pulses even at ∆ω ≈ 1 ppm from the water resonance. In contrast, the intrinsic dynamic correction methods we propose here (CEST‐GRE‐2TE and CEST‐GRE‐1TE) were affected by the CEST labeling only within a narrow range of ∆ω < |0.3| ppm, most likely because of low SNR of the saturated CEST images.

Windschuh et al. estimated the error of the MTRasym (1.1 ppm) to be on average 0.18%/Hz drift. This is in good agreement with the deviations of MTRasym (0.3–1.1 ppm) between acquisitions with and without induced B 0 drift found in our study (e.g., 0.21%/Hz and 0.18%/Hz corrected by WASSR and GRE‐2TE, respectively). The spatial inhomogeneity cannot be directly compared, since the surface of the ROI over the measured cartilage on the knee in the study by Windschuh et al. was much smaller than that covered here, in the brain.

Simegn et al. proposed a prospective motion and ∆B 0 correction of glycoCEST by updating the zero‐order and first‐order shim gradients for each CEST offset acquisition using a 3D version similar to that of our navigator.47 However, they did not show any ∆B 0 map or provide any information about the remaining local inhomogeneities of higher than first‐order within the CEST volume after correction, and hence did not attempt to apply any further postprocessing steps.

The metric MTRasym, which is highly sensitive to frequency shift errors close to the water resonance,12 has been used to assess the quality of the ∆B 0 correction methods in a similar way to the previously proposed Symmetric Analysis of Z‐Spectra (SAS).48 The static GRE‐2TE and the dynamic CEST‐GRE‐2TE methods resulted in MTRasym maps with lowest spatial variability and mean values closest to 0% for the phantom experiments. This indicates that these two methods lead to the most accurate ∆B 0 correction among the static and dynamic approaches.

The in vivo results showed comparable corrected CEST‐weighted signal distribution between the static methods GRE‐2TE and WASSR. In contrast, the dynamic methods showed the following divergences: 1) the MTRasym maps corrected by GRE‐CEST‐2TE achieved similarly homogeneous distribution to the static methods; and 2) corrections performed by CEST‐GRE‐1TE and NAV‐EPI‐2TE resulted in an overall positive MTRasym offset and a slight spatial gradient in the anterior‐posterior direction (frequency encoding direction) relative to GRE‐CEST‐2TE, respectively. These effects could arise from delays of the applied gradients and could be corrected by acquiring the same scan with opposite image readout orientation.49, 50

The scan‐rescan experiment revealed very high consistency between ΔB 0‐corrected MTRasym maps with and without frequency drift for all subjects when using GRE‐CEST‐2TE and NAV‐EPI‐2TE for all volunteers. The B 0 estimation by GRE‐CEST‐1TE corrected the frequency drift, but less efficiently than the other two dynamic methods. Further investigations would be necessary to identify the source of this deviation; however, the known nonlinear phase evolution in white matter (due to specific tissue microstructure) could be a potential contributor.51

In our study, the accuracy of multiecho ∆B 0 mapping was dependent not only on the ∆TE between the two echoes, as reported previously,16 but also on the actual values set for each TE and their receiver bandwidths. We had to optimize the TE settings experimentally to eliminate erroneous B 0 offsets and spatially linear B 0 gradients (mostly in the readout encoding direction) in phantoms prior to the CEST experiments. For routine use, it will be important to achieve accurate ∆B 0 mapping for any CEST sequence setting without previous optimization. For similar reasons we also used only monopolar readout gradients, although GRE‐CEST‐2TE and GRE‐2TE should benefit from bipolar readout gradients. We have also refrained from performing additional corrections for any other confounding effects such as motion, B 1 inhomogeneities, semisolid MT, water relaxation, T2‐dependent spillover or nuclear Overhauser enhancement exchange to prevent introducing factors that could make it difficult to isolate B 0‐related effects.52, 53, 54, 55, 56, 57, 58, 59, 60 Of course these should be used when applying accurate CEST quantification in (patient) studies.

In the future, the proposed dynamic methods could be additionally combined with real‐time motion correction by extending the navigator to 3D as previously applied in MRI and MRSI.47, 52, 53, 54, 55 Thereby, artifacts due to motion and B 0‐instabilities could be simultaneously mitigated. Although not investigated here, other CEST quantification routines such as Lorentzian or Bloch fitting61, 62, 63 would also be likely to benefit from the presented dynamic ∆B 0 correction, since B 0 is usually included as a fitting parameter.

6. CONCLUSION

We have presented three dynamic ∆B 0 correction methods for CEST MRI that successfully mapped and compensated for B 0 changes for each individual Z‐spectral point. Improved correction performance in the presence of frequency drift was demonstrated by comparison with established static approaches. Among them, the self‐correcting properties of CEST‐GRE‐2TE made it the most reliable and easiest to implement. Implementation of an interleaved navigator (NAV‐EPI‐2TE) was more complicated, but allowed improved dynamic ∆B 0 correction even close to water, but not better than CEST‐GRE‐2TE for typically investigated frequency ranges.

Dynamic B 0 corrections for CEST are another important step toward more reliable clinical CEST MRI without the need for (lengthy) prescans or for acquisition of additional Z‐spectral points near water (as required for CEST‐minZ), lending itself particularly to dynamic CEST MRI.

ACKNOWLEDGMENTS

This study was supported by the Austrian Science Fund (FWF): KLI‐718, the Austrian National Bank (OeNB Jubiläumsfonds) under Grant 16911, and the Christian Doppler Laboratory for Clinical Molecular MR Imaging. Barbara Dymerska was additionally supported by a Marie Skłodowska‐Curie Individual Fellowship. We thank Vladimir Mlynarik for his invaluable advice.

Poblador Rodriguez E, Moser P, Dymerska B, et al. A comparison of static and dynamic ∆B 0 mapping methods for correction of CEST MRI in the presence of temporal B 0 field variations. Magn Reson Med. 2019;82:633–646. 10.1002/mrm.27750

REFERENCES

  • 1. Kogan F, Hariharan H, Reddy R. Chemical exchange saturation transfer (CEST) imaging: description of technique and potential clinical applications. Curr Radiol Rep. 2013;1:102–114. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Zhou J, van Zijl PC. Chemical exchange saturation transfer imaging and spectroscopy. Prog Nucl Magn Reson Spectrosc. 2006;48:109–136. [Google Scholar]
  • 3. DeBrosse C, Nanga RPR, Bagga P, et al. Lactate chemical exchange saturation transfer (LATEST) imaging in vivo: a biomarker for LDH activity. Sci Rep. 2016;6:19517. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Haris M, Cai K, Singh A, Hariharan H, Reddy R. In vivo mapping of brain myo‐inositol. Neuroimage. 2011;54:2079–2085. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Walker‐Samuel S, Ramasawmy R, Torrealdea F, et al. Arstad E and others. In vivo imaging of glucose uptake and metabolism in tumors. Nat Med. 2013;19:1067–1072. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Wang J, Weygand J, Hwang K‐P, et al. Magnetic resonance imaging of glucose uptake and metabolism in patients with head and neck cancer. Sci Rep. 2016;6:30618. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Zaiss M, Herz K, Deshmane A, et al. Possible artifacts in dynamic CEST MRI due to motion and field alterations. J Magn Reson. 2019;298:16–22. [DOI] [PubMed] [Google Scholar]
  • 8. Zhou J, Payen JF, Wilson DA, Traystman RJ, Van Zijl P. Using the amide proton signals of intracellular proteins and peptides to detect pH effects in MRI. Nat Med. 2003;9:1085–1090. [DOI] [PubMed] [Google Scholar]
  • 9. Stancanello J, Terreno E, Castelli DD, Cabella C, Uggeri F, Aime S. Development and validation of a smoothing‐splines‐based correction method for improving the analysis of CEST‐MR images. Contrast Media Mol Imaging. 2008;3:136–149. [DOI] [PubMed] [Google Scholar]
  • 10. Zaiss M, Schmitt B, Bachert P. Quantitative separation of CEST effect from magnetization transfer and spillover effects by Lorentzian‐line‐fit analysis of z‐spectra. J Magn Reson. 2011;211:149–155. [DOI] [PubMed] [Google Scholar]
  • 11. Dula AN, Arlinghaus LR, Dortch RD, et al. Amide proton transfer imaging of the breast at 3 T: establishing reproducibility and possible feasibility assessing chemotherapy response. Magn Reson Med. 2013;70:216–224. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Kim M, Gillen J, Landman BA, Zhou J, van Zijl PC. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn Reson Med. 2009;61:1441–1450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Jezzard P, Balaban RS. Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med. 1995;34:65–73. [DOI] [PubMed] [Google Scholar]
  • 14. Robinson S, Jovicich J. B0 mapping with multi‐channel RF coils at high field. Magn Reson Med. 2011;66:976–988. [DOI] [PubMed] [Google Scholar]
  • 15. Sun PZ, Farrar CT, Sorensen AG. Correction for artifacts induced by B(0) and B(1) field inhomogeneities in pH‐sensitive chemical exchange saturation transfer (CEST) imaging. Magn Reson Med. 2007;58:1207–1215. [DOI] [PubMed] [Google Scholar]
  • 16. Wei W, Jia G, Flanigan D, Zhou J, Knopp MV. Chemical exchange saturation transfer MR imaging of articular cartilage glycosaminoglycans at 3 T: accuracy of B0 Field Inhomogeneity corrections with gradient echo method. Magn Reson Imaging. 2014;32:41–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Togao O, Hiwatashi A, Keupp J, et al. Scan‐rescan reproducibility of parallel transmission based amide proton transfer imaging of brain tumors. J Magn Reson Imaging. 2015;42:1346–1453. [DOI] [PubMed] [Google Scholar]
  • 18. Togao O, Keupp J, Hiwatashi A, et al. Amide proton transfer imaging of brain tumors using a self‐corrected 3D fast spin‐echo dixon method: comparison with separate B0 correction. Magn Reson Med. 2017;77:2272–2279. [DOI] [PubMed] [Google Scholar]
  • 19. Felblinger J, Kreis R, Boesch C. Effects of physiologic motion of the human brain upon quantitative 1H‐MRS: analysis and correction by retro‐gating. NMR Biomedicine. 1998;11:107–114. [DOI] [PubMed] [Google Scholar]
  • 20. Henry P‐G, van de Moortele P‐F, Giacomini E, Nauerth A, Bloch G. Field‐frequency locked in vivo proton MRS on a whole‐body spectrometer. Magn Reson Med. 1999;42:636–642. [DOI] [PubMed] [Google Scholar]
  • 21. Felblinger J, Kreis R, Boesch C. Effects of physiologic motion of the human brain upon quantitative 1H‐MRS: analysis and correction by retro‐gating. NMR . Biomedicine. 1998;11:107–114. [DOI] [PubMed] [Google Scholar]
  • 22. Bhattacharyya PK, Lowe MJ, Phillips MD. Spectral quality control in motion‐corrupted single‐voxel J‐difference editing scans: an interleaved navigator approach. Magn Reson Med. 2007;58:808–812. [DOI] [PubMed] [Google Scholar]
  • 23. Harris AD, Glaubitz B, Near J, et al. Impact of frequency drift on gamma‐aminobutyric acid‐edited MR spectroscopy. Magn Reson Med. 2014;72:941–948. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Rowland BC, Liao H, Adan F, Mariano L, Irvine J, Lin AP. Correcting for frequency drift in clinical brain MR spectroscopy. J Neuroimaging. 2017;27:23–28. [DOI] [PubMed] [Google Scholar]
  • 25. Lee CY, Choi IY, Lee P. Prospective frequency correction using outer volume suppression‐localized navigator for MR spectroscopy and spectroscopic imaging. Magn Reson Med. 2018;80:2366–2373. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Van de Moortele PF, Pfeuffer J, Glover GH, Ugurbil K, Hu X. Respiration‐induced B0 fluctuations and their spatial distribution in the human brain at 7 Tesla. Magn Reson Med. 2002;47:888–895. [DOI] [PubMed] [Google Scholar]
  • 27. Duerst Y, Wilm BJ, Wyss M, et al. Utility of real‐time field control in T2 *‐weighted head MRI at 7T. Magn Reson Med. 2016;76:430–439. [DOI] [PubMed] [Google Scholar]
  • 28. Zeller M, Kraus P, Muller A, Bley TA, Kostler H. Respiration impacts phase difference‐based field maps in echo planar imaging. Magn Reson Med. 2014;72:446–451. [DOI] [PubMed] [Google Scholar]
  • 29. Zahneisen B, Assländer J, LeVan P, et al. Quantification and correction of respiration induced dynamic field map changes in fMRI using 3D single shot techniques. Magn Reson Med. 2014;71:1093–1102. [DOI] [PubMed] [Google Scholar]
  • 30. Mulkern RV, Williams ML. The general solution to the Bloch equation with constant rf and relaxation terms: Application to saturation and slice selection. Med Phys. 1993;20:5–13 [DOI] [PubMed] [Google Scholar]
  • 31. Hua J, Jones CK, Blakeley J, Smith SA, van Zijl PC, Zhou J. Quantitative description of the asymmetry in magnetization transfer effects around the water resonance in the human brain. Magn Reson Med. 2007;58:786–793. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Swanson SD, Malyarenko DI, Fabiilli ML, Welsh RC, Nielsen JF, Srinivasan A. Molecular, dynamic, and structural origin of inhomogeneous magnetization transfer in lipid membranes. Magn Reson Med. 2017;77:1318–1328. [DOI] [PubMed] [Google Scholar]
  • 33. Goffeney N, Bulte J, Duyn J, Bryant LH, van Zijl P. Sensitive NMR detection of cationic‐polymer‐based gene delivery systems using saturation transfer via proton exchange. J Am Chem Soc. 2001;123:8628–8629. [DOI] [PubMed] [Google Scholar]
  • 34. Zaiss M, Bachert P. Chemical exchange saturation transfer (CEST) and MR Z‐spectroscopy in vivo: a review of theoretical approaches and methods. Phys Med Biol. 2013;58:R221–269. [DOI] [PubMed] [Google Scholar]
  • 35. Zhou J, Wilson DA, Sun PZ, Klaus JA, Van Zijl PC. Quantitative description of proton exchange processes between water and endogenous and exogenous agents for WEX, CEST, and APT experiments. Magn Reson Med. 2004;51:945–952. [DOI] [PubMed] [Google Scholar]
  • 36. Sun PZ, van Zijl PC, Zhou J. Optimization of the irradiation power in chemical exchange dependent saturation transfer experiments. J Magn Reson. 2005;175:193–200. [DOI] [PubMed] [Google Scholar]
  • 37. Singh A, Haris M, Cai K, et al. Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T. Magn Reson Med. 2012;68:588–594. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Zhou J, Payen J‐F, Wilson DA, Traystman RJ, van Zijl P. Using the amide proton signals of intracellular proteins and peptides to detect pH effects in MRI. Nat Med. 2003;9:1085. [DOI] [PubMed] [Google Scholar]
  • 39. Robinson S, Grabner G, Witoszynskyj S, Trattnig S. Combining phase images from multi‐channel RF coils using 3D phase offset maps derived from a dual‐echo scan. Magn Reson Med. 2011;65:1638–1648. [DOI] [PubMed] [Google Scholar]
  • 40. Dymerska B, Poser BA, Barth M, Trattnig S, Robinson SD. A method for the dynamic correction of B0‐related distortions in single‐echo EPI at 7T. Neuroimage. 2018;168:321–331. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Herráez MA, Burton DR, Lalor MJ, Gdeisat MA. Fast two‐dimensional phase‐unwrapping algorithm based on sorting by reliability following a noncontinuous path. Appl Opt. 2002;41:7437–7444. [DOI] [PubMed] [Google Scholar]
  • 42. Garcia D. Robust smoothing of gridded data in one and higher dimensions with missing values. Comput Stat Data Anal. 2010;54:1167–1178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Dymerska B, Poser BA, Bogner W, et al. Correcting dynamic distortions in 7T echo planar imaging using a jittered echo time sequence. Magn Reson Med. 2016;76:1388–1399. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Eckstein K, Dymerska B, Bachrata B, et al. Computationally efficient combination of multi‐channel phase data from multi‐echo acquisitions (ASPIRE). Magn Reson Med. 2018;79:2996–3006. [DOI] [PubMed] [Google Scholar]
  • 45. Lima Cardoso P, Dymerska B, Bachratá B, et al. The clinical relevance of distortion correction in presurgical fMRI at 7T. NeuroImage. 2018;168:490–498. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Windschuh J, Zaiss M, Ehses P, Lee JS, Jerschow A, Regatte RR. Assessment of frequency drift on CEST MRI and dynamic correction: application to gagCEST at 7 T. Magn Reson Med. 2019;81:573–582. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Simegn GL, Van der Kouwe A, Robertson FC, Meintjes EM, Alhamud A. Real‐time simultaneous shim and motion measurement and correction in glycoCEST MRI using double volumetric navigators (DvNavs). Magn Reson Med. 2019;81:2600–2613. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Fairney JL, G, Shmueli, K , Golay, X . SAS: Symmetric Analysis of Z‐Spectra, a Method to Evaluate B0 Correction Techniques for CEST Data in Clinical Systems Using Non‐Exchanging Phantoms In: Proceedings of the 23rd Annual Meeting of ISMRM, Toronto, Canada, 2015. Abstract 3344. [Google Scholar]
  • 49. Reeder SB, Atalar E, Faranesh AZ, McVeigh ER. Referenceless interleaved echo‐planar imaging. Magn Reson Med. 1999;41:87–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50. Strasser B, Chmelik M, Robinson SD, et al. Coil combination of multichannel MRSI data at 7 T: MUSICAL. NMR Biomedicine. 2013;26:1796–1805. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51. Sood S, Urriola J, Reutens D, et al. Echo time‐dependent quantitative susceptibility mapping contains information on tissue properties. Magn Reson Med. 2017;77:1946–1958. [DOI] [PubMed] [Google Scholar]
  • 52. Hess AT, Tisdall MD, Andronesi OC, Meintjes EM, van der Kouwe AJ. Real‐time motion and B0 corrected single voxel spectroscopy using volumetric navigators. Magn Reson Med. 2011;66:314–323. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Hess AT, Andronesi OC, Tisdall MD, Sorensen AG, van der Kouwe AJ, Meintjes EM. Real‐time motion and B0 correction for localized adiabatic selective refocusing (LASER) MRSI using echo planar imaging volumetric navigators. NMR Biomedicine. 2012;25:347–358. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Bogner W, Hess AT, Gagoski B, et al. Real‐time motion and B0‐correction for LASER‐localized spiral‐accelerated 3D‐MRSI of the brain at 3T. Neuroimage. 2014;88:22–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55. Alhamud A, Taylor PA, van der Kouwe AJ, Meintjes EM. Real‐time measurement and correction of both B0 changes and subject motion in diffusion tensor imaging using a double volumetric navigated (DvNav) sequence. Neuroimage. 2016;126:60–71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56. Zaiss M, Zu Z, Xu J, et al. A combined analytical solution for chemical exchange saturation transfer and semi‐solid magnetization transfer. NMR Biomed. 2015;28:217–230. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57. Xu J, Chan KW, Xu X, Yadav N, Liu G, van Zijl PC. On‐resonance variable delay multipulse scheme for imaging of fast‐exchanging protons and semisolid macromolecules. Magn Reson Med. 2017;77:730–739. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58. Windschuh J, Zaiss M, Meissner JE, et al. Correction of B1‐inhomogeneities for relaxation‐compensated CEST imaging at 7 T. NMR Biomed. 2015;28:529–537. [DOI] [PubMed] [Google Scholar]
  • 59. Peterson P, Olsson E, Svensson J. T2 relaxation time bias in gagCEST at 3T and 7T: comparison of saturation schemes. Magn Reson Med. 2018;81:1044–1051. [DOI] [PubMed] [Google Scholar]
  • 60. Khlebnikov V, Windschuh J, Siero JC, et al. On the transmit field inhomogeneity correction of relaxation‐compensated amide and NOE CEST effects at 7 T. NMR Biomedicine. 2017;30:e3687. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61. Desmond KL, Stanisz GJ. Understanding quantitative pulsed CEST in the presence of MT. Magn Reson Med. 2012;67:979–990. [DOI] [PubMed] [Google Scholar]
  • 62. Zaiss M, Xu J, Goerke S, et al. Inverse Z‐spectrum analysis for spillover‐, MT‐, and T1 ‐corrected steady‐state pulsed CEST‐MRI–application to pH‐weighted MRI of acute stroke. NMR Biomed. 2014;27:240–252. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63. Zhang X‐Y, Wang F, Li H, et al. Accuracy in the quantification of chemical exchange saturation transfer (CEST) and relayed nuclear Overhauser enhancement (rNOE) saturation transfer effects. NMR Biomed. 2017;30:e3716. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley