Abstract
Oxidative stress is a risk for and cause of various disease, however, measurements of oxidative stress are either time-consuming or non-specific. Here, we established a rapid method of using high performance liquid chromatography (HPLC) to measure serum oxidized albumin in a rat model. We optimized HPLC conditions for rat oxidized albumin. To validate our method, three-week-old male Sprague-Dawley rats were uninephrectomized and treated normal diet, high salt diet or high salt diet with Tempol, a superoxide dismutase (SOD) mimetic. After 4 weeks of treatment, we analyzed serum oxidized albumin. The main findings are listed as below. (i) Our method of oxidized albumin measurement only takes 16 minutes, with an intra-day and inter-day deviation within 1% and a detection limit concentration of 6.4 mg/ml. (ii) Oxidized albumin levels were significantly higher in the high salt diet group than in the normal salt diet group, and this effect was reversed by Tempol. (iii) Oxidized albumin levels also correlated with urinary protein and 8-isoprostane levels. In conclusion, we have established a simple method for evaluating rat serum oxidized albumin using HPLC. Our method is rapid and has an advantage over conventional methods and may be useful for animal models of oxidative stress.
Subject terms: Animal disease models, Experimental models of disease
Introduction
Oxidative stress has elicited high levels of interest in the field of biology for a long time. It has been reported that oxidative stress plays pivotal roles in a variety of disease conditions as well as in aging1–5. To measure oxidative stress levels, many biomarkers has been used, such as 8-isoprostane, malondialdehyde (MDA), nitrotyrosine levels, and serum antioxidant capacity. Each of them has distinct characteristics, but does not necessarily reflect a ubiquitous oxidative stress level. To overcome these problems, we and others have investigated the levels of the oxidized form of albumin as a new marker of oxidative stress6–9. Albumin is a mixture of reduced albumin (mercaptalbumin) and oxidized albumin (non-mercaptalbumin) in extracellular fluid such as serum. Reduced albumin has one free sulfhydryl group in Cys-34, while oxidized albumin has a ligand bound to the sulfhydryl group in Cys-349. Reduced albumin is bound with its mixed disulphide with cysteine or glutathione8. Oxidized albumin has more oxidized products such as sulfenic (-SOH), sulfinic (-SO2H) or sulfonic (-SO3H) states9.
A former study reported that oxidized rat serum albumin cannot be clearly separated from reduced albumin by conventional HPLC method8. Hayashi T et al. has developed a method by HPLC which can separate both reduced and oxidize albumin in rat serum9. However, Hayashi’s method takes about one hour to measure one sample in room temperature. Here, we established a simple and rapid method for measuring oxidized albumin in rat serum. We validated this method by using an established rat model of high oxidative stress which also demonstrated proteinuria and hypertension10–12.
Results
Separation of rat oxidized and reduced albumin by using HPLC
By systematically screening of measurement conditions, we eventually determined the optimal condition for measuring rat oxidized and reduced albumin: 25 mM phosphoric acid buffer with 60 mM sodium sulfate, plus 1.5% ethanol (solution I, pH 5.3) and 1000 mM magnesium chloride (solution II). After balancing the column, the flow rate was set to 1 ml/min. The oven temperature was set to 40 °C, and the samples volume was 3 microliters. The whole process of the serum analysis lasted for 12 minutes, and the linear gradient was 0–65% from solution I to solution II. Thus, only a total time of 16 minutes was needed to analyze one sample (including the column balancing and sample analysis times). A representative chromatograph is shown in Fig. 1. Compared with the measurement conditions of human serum albumin, the major differences are ethanol, pH value and linear gradient time. The comparison is summarized in the supplemental Table 2.
Reproducibility of analysis method
The reproducibility of our analysis method is summarized in Supplemental Table 1. The CV values of reproducibility for inter-day and intra-day were 0.77% and 0.81%, respectively. We performed multiple dilutions, and the minimal detectable concentration of oxidized albumin was 6.4 mg/ml (Fig. 2a).
Our result showed that 20 minutes of incubation at room temperature caused a significant increase in the percentage of oxidized albumin in a sample, and that the oxidation of albumin is time-dependent (Fig. 2b). We named this process “auto-oxidation”. Serum albumin could be auto-oxidized at room temperature after 20 minutes (Fig. 2b).
We also conducted an “interference evaluation” study. In brief, we used commercially available oxidized albumin standards to calculate the percentage of oxidized albumin, and demonstrated a positive correlation between standard albumin/total albumin and oxidized albumin (Fig. 2c).
Validation of method in a rat model of proteinuria and hypertension
We validate our method of measuring oxidized albumin in an established rat model of proteinuria and hypertension10. As expected, high salt-loading resulted in significantly higher systolic blood pressure (176.1 ± 34.6 mmHg) in rats with uninephrectomy (UNx) compared to normal salt treated UNx rats (126.4 ± 9.4 mmHg), and additional Tempol in drinking water attenuated high salt loading-induced elevation in systolic blood pressure (140.4 ± 15.0 mmHg) in rats (Fig. 3a).
The urinary protein level in the high salt diet group was significantly higher (78.75 ± 87.13 mg /day) than that from normal salt diet group (8.59 ± 8.95 mg/day), and after treatment with Tempol, the effect of high salt diet on urinary protein was abolished (4.82 ± 5.29 mg /day) (Fig. 3b).
Serum oxidized albumin in high salt diet group (35.43% ± 4.39%) was significantly higher compared to that from normal salt diet group, and Tempol significantly reversed this effect of high salt loading (26.98% ± 2.19%) (Fig. 3c).
Consistent with previous reports13–15, the 8-isoprostane level in high salt diet group (36.7 ± 45.9 ng/day) was significantly higher than that in normal salt diet group (10.5 ± 7.4 ng/day), and Tempol reversed this effect (11.0 ± 16.6 ng/day) (Fig. 3d).
There are positive correlations between oxidized albumin% and both proteinuria (Fig. 4a) and 8-isoprostane (Fig. 4b). Based on the ROC curve, the areas under the curve (AUCs) were 0.643 for urinary 8-isoprostane and 0.917 for oxidized albumin, indicating that oxidized albumin is a new biomarker with greater sensitivity for evaluating oxidative stress, compared to a traditional marker, urinary 8-isoprostane (p < 0.01) (Fig. 4c).
Discussion
In the present study, we described a simple, adapted method for the measurement of rat oxidized albumin. The total time taken to measure oxidized albumin with our method was only a short 16 minutes, with an intra-day and inter-day deviation within 1% and a detection limit at a concentration of 6.4 mg/ml. Our method is sensitive and rapid, and has an advantage over conventional methods and may be useful for future studies of animal models of oxidative stress.
The gradual increase in oxidized albumin in the intra-day reproducibility analysis may resulted from auto-oxidation. Our experiments have shown that auto-oxidation occurs as quickly as 20 minutes, making it necessary to measure the samples over a short period of time. The total time taken to analyze a sample using our method is as short as 16 minutes (column equilibrating time included). Our method is rapid and capable of separating peaks clearer than former reports8,9.
For the peaks to be sharp and be clearly separated, we needed optimal concentrations of magnesium chloride and ethanol. The acidity of the solutions are also important when combining solutions I and II, whereby the acidic II solution was added gradually to solution I to obtain an optimized linear gradient. The pH values of solution I are 6.0 and 5.3 in the human and rat, respectively, which is due to the isoelectric point difference between humans and rats. Both the ion-exchange and hydrophilic interaction of the resin can contribute to the separation of oxidized albumin and reduced albumin by the column, which is further enhanced by varying magnesium concentration and the pH of solution I. It is also noteworthy that there are also limitations of our method. It is reported that both in human and rat cases, oxidized albumin is the mixture of NA-1(Non-mercaptalbumin-1) and NA-2(Non-mercaptalbumin-2)16–18.These two states of oxidized albumin may represent different pathophysiology19. In our model, the new method has not shown separation of these two peaks. Further research in different model should be conducted.
For the interference study, we did an “interference evaluation”. As standard oxidized albumin (100% oxidized) increased by 8%, the percentage of oxidized albumin increased by 7.42% (AVE). There is a positive correlation between the percentage of standard albumin/total albumin and oxidation albumin. Our result showed that our method for measuring oxidized albumin closely reflect the actual amount of standard albumin. Based on the intra-day and inter-day reproducibility, the method was sensitive enough to measure rat oxidized albumin at a concentration as low as 6.4 mg/ml.
Previous reports have shown that many diseases are associated with high oxidative stress levels in both humans6,20,21 and animal models22–26. In vitro studies have also shown that oxidized albumin itself is toxic27. Previous animal study has also shown oxidative changes in the blood and serum albumin in rats with monoarthritis and polyarthritis28. To validate our method, we utilized an established rat model of high in oxidative stress which is associated with proteinuria and hypertension10. The results from our study showed that the percentage of oxidized albumin was significantly higher in the disease model than in the control group and that this change was reversed by the anti-oxidant drug Tempol. These results were consistent with human renal function data7,29. To further confirm our results, we have compared serum oxidized albumin with urinary 8-isoprostane levels, which is a commonly used marker to indicate the oxidative stress. The high salt group also showed a significantly higher 8-isoprostane level than the other two groups, and it positively correlated with our measurement of serum oxidized albumin. These findings suggest that our method of measuring oxidized albumin is in accordance with the currently using marker 8-isoprostane and has the potential to be applied to other disease models. Our correlation data between oxidized albumin and urinary protein also suggests that measured oxidized albumin may be a better marker for oxidative stress-related organ damage.
8-Isoprostane is formed by the peroxidation of arachidonic acid30, which is found in cell membrane phospholipids and is a target of reactive oxygen species31, whilst urinary protein is a representative marker to evaluate renal damage10,32 and reflects glomerular damage. A previous report shows that 8-isoprostane is not a prerequisite for renal damage33. Taken together, oxidized albumin is a useful marker for oxidative stress-induced kidney disease, and appear to be an advanced marker compared to 8-isoprostane. A reduction in the percentage of oxidized albumin may also be a target for the prevention of renal diseases.
In conclusion, we have established a simple and rapid method for measuring oxidative stress levels with oxidized albumin. We validated this method by using an established rat model of proteinuria and hypertension which demonstrate high levels of oxidized albumin and 8-isoprostane. Our method is sensitive and rapid, and has an advantage over conventional methods and may be useful for future studies of animal models of oxidative stress.
Materials and Methods
Animals
Three-week-old male Sprague Dawley (SD) rats (body weights ranging from 40 g to 50 g) were subjected to left uninephrectomy and then randomly divided into the following 3 groups: the normal salt diet (UNx + NS, 0.3% NaCl, n = 20); high salt diet (UNx + HS, 8% NaCl, n = 18) and high salt diet treated with a superoxide-dismutase mimetic, namely, 4-hydroxy-2,2,6,6-tetramethyl-piperidine-N-oxyl, also known as Tempol (UNx + HST, 8% NaCl, 1 mmol/kg/day in drinking water, n = 17). After 4 weeks of treatment, blood samples were collected by cardiac puncture.
Rats were maintained in a humidity- (60 ± 5%), temperature- (23 ± 1.5 °C), and light cycle-regulated (0700–1900 h) room and had free access to food and drinking water. Reasearch Ethics Committee of the University of Tokyo approved this investigation (No:P14–145), which was conducted according to the guidelines for the care and use of laboratory animals of the University of Tokyo, Graduate School of Medicine.
Body weight and systolic blood pressure (SBP) were measured prior to sacrifice. SBP was measured using the tail-cuff method (P-98A; Softron, Tokyo, Japan). Twenty-four-hour urine samples were collected at the end of the 4-week treatment using metabolic cages to measure both the urinary protein and the 8-isoprostane levels.
Condition establishment for HPLC
The HPLC system (LabSolutions system; Shimazu Co., Ltd, Kyoto, Japan) comprised of a degasser (DGU20A3R), two pumps (LC-20AT), an autosampler (SIL30AC), a thermostatic oven (CTO-20AC), a fluorescence detector (RF-20Axs) and a system controller (CBM-20 A).
The polyvinyl alcohol cross-linked gel (9 µm in diameter) (Asahipak GS-520; AsahiKasei Co., Ltd, Tokyo, Japan) which was dried in a vacuum for more than 16 hours was then suspended in 10 ml of dimethyl sulfoxide (DMSO; Tokyo Chemical Industry Co., Ltd, Tokyo, Japan) per gram of the dried gel. Next, to suspend the gel, 20 mmol of epichlorohydrin were added to per gram of the dried gel for 20 hours at 30 °C. The activated gel was then filtrated, followed by an addition of 10% aqueous solution of diethyl amine (Wako Pure Chemical Industries, Ltd, Osaka, Japan) for another 20 hours. Finally, the gel was packed in a stainless column (50 × 7.6 mm I.D.). By using the above prepared column which is of high resolution and anion exchanged, the analysis condition to separate rat reduced albumin and oxidized albumin was screened by optimizing the factors—buffers, pH value, ethanol concentration, flow rate, the linear gradient of magnesium concentration, oven temperature and sample volume.
All of the reagents were either HPLC grade or a special grade (Wako Pure Chemical Industries, Ltd, Osaka, Japan).
The fluorescence detection wavelengths were 280 nm and 340 nm. Two eluent buffers contained 60 mM sodium sulfate and 25 mM phosphoric acid buffer. Three microliters of serum or plasma samples was directly injected into the system. The flow rate was 1 ml/min, and the analytical temperature was 40 °C. The peaks for reduced albumin and oxidized albumin were then confirmed. Based on the peak areas, the results were expressed as Oxidized albumin% = Oxidized albumin/(Reduced albumin + Oxidized albumin) × 100%7,34,35.
Evaluation of method performance
The intra-day reproducibility was checked 10 times continuously in one day, and inter-day reproducibility was checked on 10 continuous days.
Sensitivity of the analysis method
To study the detection limit of rat oxidized albumin in our method, we used serum samples from healthy rats and diluted them with PBS. Oxidized albumin level was measured as described above.
Evaluation of auto-oxidation
To investigate the time-dependent auto-oxidation of albumin at room temperature, we measured control samples of serum albumin every 20 minutes for a total 120 minutes.
Interference study
We added standard albumin (100% oxidized albumin, Supplemental Fig. 1) to the rat serum samples following a linear concentration gradient from 0% to 48% (the sample concentration gradient increased by 8%) and then measured oxidized albumin level using HPLC.
Method for evaluation of the 8-isoprostane concentration
Urine samples of the rats were centrifuged at 10,000 rpm for 5 minutes. The urine samples from rats fed normal salt diets were diluted 5 times, whilst samples from rats fed a high salt diet or high salt diet plus Tempol were not diluted.
According to the manufacturer’s instructions, we measured 8-isoprostane levels with the 8-iso-PGF2α Elisa kit (Enzo Life Sciences, Inc. Farmingdale, New York, USA). Briefly, the plate was incubated with the sample at room temperature on a plate shaker at 120 rpm for 120 minutes and then washed 5 times using washing solution. After incubating at room temperature with a solution of p-nitrophenyl phosphate in buffer for 45 minutes without shaking and adding stop solution, the optical density was measured at 420 nm. The concentration of 8-isoprostane was calculated. Urine samples and standard samples were all run in duplicate.
Statistical analysis
Data are expressed as the mean ± SEM. All urinary protein levels were logarithmically transformed prior to statistical analysis by ANOVA with Tukey’s post hoc test because they had a heteroscedastic distribution. Statistical analysis among the 3 groups was performed by one-way ANOVA followed by Tukey’s test. The receiver operating characteristic (ROC) curve for oxidized albumin and 8-isoprostane was drawn to compare the diagnostic value. P < 0.05 was considered to be statistically significant.
Supplementary information
Acknowledgements
The authors acknowledge the financial support from Kakenhi 17K09724 (Shimosawa T.) and the Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering (Yatomi Y.).
Author Contributions
B.L. conceptualized the study, analyzed the data for this research, and wrote the manuscript. T.S. contributed on study design, as well as provided substantial edits to the manuscript. S.S.K., K.Y., Y.Y., edited the manuscript and provided insight for additional discussion. A.Y., L.R. and C.W. provided insight for discussion. All authors contributed to manuscript read and approved the submitted version.
Data Availability
The authors confirm that the data supporting the findings of this study are available within the article and its supplementary materials.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Supplementary information accompanies this paper at 10.1038/s41598-019-45134-x.
References
- 1.Sies H, Berndt C, Jones DP. Oxidative Stress. Annu. Rev. Biochem. 2017;86:715–748. doi: 10.1146/annurev-biochem-061516-045037. [DOI] [PubMed] [Google Scholar]
- 2.Poprac P, et al. Targeting free radicals in oxidative stress-related human diseases. Trends in Pharmacological Sciences. 2017;38:592–607. doi: 10.1016/j.tips.2017.04.005. [DOI] [PubMed] [Google Scholar]
- 3.Bouayed Jaouad. Anxiety Disorders. 2011. Relationship Between Oxidative Stress and Anxiety: Emerging Role of Antioxidants Within Therapeutic or Preventive Approaches. [Google Scholar]
- 4.Zhao MX, et al. Salusin-β contributes to oxidative stress and inflammation in diabetic cardiomyopathy. Cell Death Dis. 2017;8:1–10. doi: 10.1038/s41419-017-0042-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Wang YL, et al. Protective effect of curcumin against oxidative stress-induced injury in rats with parkinson’s disease through the Wnt / β -catenin signaling pathway. Cell. Physiol. Biochem. 2017;43:2226–2241. doi: 10.1159/000484302. [DOI] [PubMed] [Google Scholar]
- 6.Nagumo K, et al. Cys34-cysteinylated human serum albumin is a sensitive plasma marker in oxidative stress-related chronic diseases. PLoS One. 2014;9:e852116. doi: 10.1371/journal.pone.0085216. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Yasukawa K, Shimosawa T, Okubo S, Yatomi Y. A simple, rapid, and validated high-performance liquid chromatography method suitable for clinical measurements of human mercaptalbumin and nonmercaptalbumin. Ann. Clin. Biochem. 2017;55:121–127. doi: 10.1177/0004563217693257. [DOI] [PubMed] [Google Scholar]
- 8.Nishimura K, Harada K, Masuda S, Sugii A. High-performance liquid chromatography of serum albumins on an N-methylpyridinium polymer-based column. J. Chromatography. 1990;525:176–182. doi: 10.1016/S0378-4347(00)83391-5. [DOI] [PubMed] [Google Scholar]
- 9.Hayashi T, Suda K, Imai H, Era S. Simple and sensitive high-performance liquid chromatographic method for the investigation of dynamic changes in the redox state of rat serum albumin. J. Chromatography. 2002;772:139–146. doi: 10.1016/s1570-0232(02)00068-5. [DOI] [PubMed] [Google Scholar]
- 10.Kawarazaki H, et al. Mineralocorticoid receptor activation: A major contributor to salt-induced renal injury and hypertension in young rats. Am J Physiol Ren. Physiol. 2011;300:1402–1409. doi: 10.1152/ajprenal.00691.2010. [DOI] [PubMed] [Google Scholar]
- 11.Sugimoto K, Tsuruoka S, Matsushita K, Fujimura A. Enalapril on oxidative stress in 5/6 nephrectomized rats. Nephron. 1999;83:282–283. doi: 10.1159/000045527. [DOI] [PubMed] [Google Scholar]
- 12.Nath KA, Croatt AJ, Hostetter TH. Oxygen consumption and oxidant stress in surviving nephrons. Am. J. Physiol. 1990;258:F1354–62. doi: 10.1152/ajprenal.1990.258.5.F1354. [DOI] [PubMed] [Google Scholar]
- 13.Kitiyakara C, et al. Salt intake, oxidative stress, and renal expression of NADPH oxidase and superoxide dismutase. J. Am. Soc. Nephrol. 2003;14:2775–2782. doi: 10.1097/01.ASN.0000092145.90389.65. [DOI] [PubMed] [Google Scholar]
- 14.Meng S, Roberts LJ, Cason GW, Curry TS, Manning RD. Superoxide dismutase and oxidative stress in Dahl salt-sensitive and -resistant rats. Am. J. Physiol. Regul. Integr. Comp. Physiol. 2002;283:R732–8. doi: 10.1152/ajpregu.00346.2001. [DOI] [PubMed] [Google Scholar]
- 15.Cruz A, et al. Effects of clofibrate on salt loading-induced hypertension in rats. J. Biomed. Biotechnol. 2011;2011:1–8. doi: 10.1155/2011/469481. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Matsuyama Y, Terawaki H, Terada T, Era S. Albumin thiol oxidation and serum protein carbonyl formation are progressively enhanced with advancing stages of chronic kidney disease. Clin Exp Nephrol. 2009;13:308–315. doi: 10.1007/s10157-009-0161-y. [DOI] [PubMed] [Google Scholar]
- 17.Terawaki Hiroyuki, Hayashi Tomoya, Murase Takayo, Iijima Ryutaro, Waki Kaito, Tani Yoshihiro, Nakamura Takashi, Yoshimura Kazunobu, Uchida Shunya, Kazama Junichiro James. Relationship between Xanthine Oxidoreductase Redox and Oxidative Stress among Chronic Kidney Disease Patients. Oxidative Medicine and Cellular Longevity. 2018;2018:1–6. doi: 10.1155/2018/9714710. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Terawaki H, et al. Oxidative stress is enhanced in correlation with renal dysfunction: Examination with the redox state of albumin. Kidney Int. 2004;66:1988–1993. doi: 10.1111/j.1523-1755.2004.00969.x. [DOI] [PubMed] [Google Scholar]
- 19.Wada Y, et al. Increased ratio of non-mercaptalbumin-1 among total plasma albumin demonstrates potential protein undernutrition in adult rats. Front. Nutr. 2018;5:1–9. doi: 10.3389/fnut.2018.00064. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Colombo G, et al. Redox albuminomics: oxidized albumin in human diseases. Antioxid. Redox Signal. 2012;17:1515–1527. doi: 10.1089/ars.2012.4702. [DOI] [PubMed] [Google Scholar]
- 21.Suzuki Y, Suda K, Matsuyama Y, Era S, Soejima A. Close relationship between redox state of human serum albumin and serum cysteine levels in non-diabetic CKD patients with various degrees of renal function. Clin. Nephrol. 2014;82:320–5. doi: 10.5414/CN108040. [DOI] [PubMed] [Google Scholar]
- 22.Herbet M, et al. Chronic variable stress is responsible for lipid and DNA oxidative disorders and activation of oxidative stress response genes in the brain of rats. Oxid. Med. Cell. Longev. 2017;2017:1–10. doi: 10.1155/2017/7313090. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Zhang Y, et al. A new role for oxidative stress in aging: The accelerated aging phenotype in Sod1-/- mice is correlated to increased cellular senescence. Redox Biol. 2017;11:30–37. doi: 10.1016/j.redox.2016.10.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Park KT, Yun CH, Bae CS, Ahn T. Decreased level of albumin in peripheral blood mononuclear cells of streptozotocin-induced diabetic rats. J. Vet. Med. Sci. 2014;76:1087–1092. doi: 10.1292/jvms.13-0631. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Bjorntorp P. The oxidation of fatty acids combined with albumin by isolated rat liver mitochondria. J. Biol. Chem. 1966;241:1537–1543. [PubMed] [Google Scholar]
- 26.Sivonová M, et al. Relationship between antioxidant potential and oxidative damage to lipids, proteins and DNA in aged rats. Physiol. Res. 2007;56:757–64. doi: 10.33549/physiolres.931094. [DOI] [PubMed] [Google Scholar]
- 27.Miyamura S, et al. Comparison of posttranslational modification and the functional impairment of human serum albumin in commercial preparations. J. Pharm. Sci. 2016;105:1043–1049. doi: 10.1016/j.xphs.2015.12.015. [DOI] [PubMed] [Google Scholar]
- 28.Bracht A, et al. Oxidative changes in the blood and serum albumin differentiate rats with monoarthritis and polyarthritis. Springerplus. 2016;5:1–14. doi: 10.1186/s40064-016-1671-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Nakatani S, et al. Non-mercaptalbumin,oxidized form of serum albumin,significantly associated with renal function and anemia in chronic kidney disease patients. Sci. Rep. 2018;8:1–9. doi: 10.1038/s41598-018-35177-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Morrow JD, et al. A series of prostaglandin F2-like compounds are produced in vivo in humans by a non-cyclooxygenase, free radical-catalyzed mechanism. Proc. Natl. Acad. Sci. USA. 1990;87:9383–9387. doi: 10.1073/pnas.87.23.9383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Costa LG, Giordano G, Guizzetti M. In vitro neurotoxicology: An introduction. Methods Mol. Biol. 2011;758:1–9. doi: 10.1007/978-1-61779-170-3_1. [DOI] [PubMed] [Google Scholar]
- 32.Jia L, et al. Using an isolated rat kidney model to identify kidney origin proteins in urine. PLoS One. 2013;8:e66911. doi: 10.1371/journal.pone.0066911. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Green ML, Freel RW, Hatch M. Lipid peroxidation is not the underlying cause of renal injury in hyperoxaluric rats. Kidney Int. 2005;68:2629–2638. doi: 10.1111/j.1523-1755.2005.00735.x. [DOI] [PubMed] [Google Scholar]
- 34.Era S, et al. Further studies on the resolution of human mercapt- and nonmercaptalbumin and on human serum albumin in the elderly by high-performance liquid chromatography. I n t. J. Pept. Protein Res. 1988;31:435–442. doi: 10.1111/j.1399-3011.1988.tb00900.x. [DOI] [PubMed] [Google Scholar]
- 35.Sogami M, Shunji N, Seiichi E, Makoto H, Kohji N. Resolution of human mercapt- and nonmercaptalbumin by high-performance liquid chromatography. Int.J.Peptide Protein Res. 1984;24:96–103. doi: 10.1111/j.1399-3011.1984.tb00933.x. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The authors confirm that the data supporting the findings of this study are available within the article and its supplementary materials.