Skip to main content
The Journal of International Medical Research logoLink to The Journal of International Medical Research
. 2019 Apr 24;47(6):2666–2673. doi: 10.1177/0300060519844399

Evaluation of a manual identification system for detection of Mycobacterium tuberculosis in a primary tuberculosis laboratory in China

Ping Zhao 1,, Qin Yu 2, Yu Zhang 3
PMCID: PMC6567736  PMID: 31014140

Short abstract

Objective

To compare the diagnostic performance of the manual BACTEC™ Mycobacteria Growth Indicator Tube (MGIT™) system (M-MGIT) with the automated BACTEC™ MGIT™ 960 system (A-MGIT) and Löwenstein-Jensen (L-J) culture method in detecting mycobacteria in sputum specimens from patients with suspected pulmonary tuberculosis (TB).

Methods

For this cross-sectional study, sputum samples were taken from patients aged ≥18 years attending a TB clinic in Beijing, China between July 2015 and October 2016. Processed sputum samples were inoculated into the MGIT systems and L-J medium for up to 6 and 8 weeks, respectively.

Results

The M-MGIT and A-MGIT methods detected significantly more Mycobacterium tuberculosis complex (MTC) isolates than L-J culture from the 565 sputum samples (39%, 40% and 32%, respectively). Using a positive result from any of the three culture systems as reference, the sensitivity of M-MGIT, A-MGIT and L-J methods were 92%, 94%, and 74%, respectively. The time-to-detection of mycobacteria was 12.9±4.2 days for M-MGIT, 11.8±5.2 days for A-MGIT and 24.2±8.7 days for L-J.

Conclusions

M-MGIT has a similar diagnostic performance to A-MGIT, and is a fast and reliable alternative to conventional culture methods in the diagnosis of pulmonary TB in a developing country.

Keywords: Mycobacterium tuberculosis, BACTEC MGIT, liquid culture, tuberculosis, Lowenstein–Jensen, mycobacteria

Introduction

Worldwide, tuberculosis (TB) causes ill-health in millions of people each year and in 2015 was one of the top 10 causes of death.1 In 2015, there were an estimated 918,000 incident new cases of TB in China and 35,000 related deaths. 1 Therefore, addressing gaps in detection and treatment is of paramount importance. 1 Sputum smear microscopy remains the most commonly used method for diagnosing TB worldwide.2 Developments in diagnostics over the past few years has resulted in the use of rapid molecular tests such as the Xpertw Mycobacterium tuberculosis (MTB)/rifampicicn (RIF) assay and the real-time fluorescent based quantitative polymerase chain reaction (PCR) detection method. 2However, these molecular tests are relatively expensive and their implementation is difficult in resource-limited settings.2 Non-radiometric liquid culture methods, such as the Mycobacteria Growth Indicator Tube (MGIT ™), have been developed to speed up the isolation of slow-growing mycobacteria and have been reported to be faster and a more reliable alternative to conventional culture on Lowenstein-Jensen egg (LJ) solid medium.3,4 Unfortunately, liquid culture systems are expensive which has precluded their widespread use, particularly in resource-limited settings.5 However, while the BACTEC ™ MGIT ™ 960 system (Becton-Dickinson, Franklin Lakes, NJ, USA) is automated for continuous monitoring, the older manual MGIT system is more affordable for laboratories with a small budget. 4

The present study was designed to compare the performance of the manual BACTEC MGIT system (M-MGIT) with the automated BACTEC MGIT 960 system (A-MGIT) and a conventional L-J culture method in detecting mycobacteria in sputum specimens from patients with suspected pulmonary TB.

Methods

For this cross-sectional study, sputum samples were taken from patients attending a TB clinic in Chaoyang District Centre for Disease Control and Prevention, Chaoyang District, Beijing, China between July 2015 and October 2016. Patients were ≥18 years of age and had a high clinical suspicion of TB according to the Chinese National Diagnostic Guidelines.6The study protocol was approved by the Ethics Committee of Chaoyang District Centre for Disease Control and all patients provided written informed consent.

One sputum specimen (3–5ml) was collected from each patient before treatment. Prior to processing, smears were prepared and stained with Ziehl-Neelsen stain and examined under a light microscope to confirm the presence of acid-fast bacilli (AFB). The remaining sputum was decontaminated and digested according to the N-acetyl-L-cysteine (NALC) sodium hydroxide method.7 A portion (0.5ml) of the processed specimen was inoculated into a tube of the M-MGIT system. The tubes were incubated at 37°C and examined daily in a 365nm wavelength UV light source fluorescence detector (BACTEC ™ MicroMGIT ™ device). For the A-MGIT system, 0.5ml of the processed specimen was inoculated into a tube which was incubated at 37°C and monitored automatically every 60 minutes for increased fluorescence. All culture tubes from both methods were incubated for 6 weeks or until they were found to be positive. A portion (0.1ml) of the remaining processed specimen was inoculated into the L-J medium and incubated at 37°C with daily examinations for eight weeks until mycobacterial colonies were detected. Typical colonies were tested for Mycobacterium tuberculosis complex (MTC) organisms using para nitro benzoic acid (PNB) and thiophene-2-carboxylic acid hydrazide (TCH) medium growth tests.6

Statistical analyses

Data were analysed using the Statistical Package for Social Sciences (SPSS®) for Windows® release 13.0 (SPSS Inc., Chicago, IL, USA) and a P-value <0.05 was considered to indicate statistical significance. The recovery and contamination rates of the three systems were compared using χ2 test. Concordance between tests was evaluated using the kappa statistic and 95% confidence intervals (CI). Student t-test was used to compare the time-to-detection (TTD) in different media.

Results

Of the 565 sputum samples available from patients with presumptive pulmonary TB, 237 (42%) grew mycobacteria by the M-MGIT method, 243 (43.0%) by the A-MGIT method, and 190 (34%) by the L-J method (Table 1). There was no statistically significant difference in yield observed between the MGIT systems, but there was a statistically significant difference in yield between each MGIT system and the L-J method (χ2=8.32 and 10.52; P <0.01).

Table 1.

Comparison of results from the three different culture systems in the analysis of sputum samples from 565 patients with suspected pulmonary tuberculosis


Sputum samples
M-MGIT A-MGIT L-J All systems
Total processed 565 565 565 565
Positive growth 237(41.9) 243 (43.0) 190 (33.6) 259 (45.8)
MTC 222 (39.3) 227 (40.2) 179 (31.7) 241 (42.6)
NTM 15 (2.6) 16 (2.8) 11 (1.9) 18 (3.2)
Contaminated 33 (5.8) 35 (6.2) 15 (2.7) 8 (1.4)
Negative 295 (52.3) 287 (50.8) 360 (63.7) 298 (52.8)

Data are presented as n or n (%).

Contaminated or negative on all three media

M-MGIT™, manual Mycobacterial Growth Indicator Tube; A-MGIT, automated Mycobacterial Growth Indicator Tube (BACTEC™ MGIT™ 960); L-J, Löwenstein-Jensen culture method, MTC, Mycobacterium tuberculosis complex; NTM, non-tuberculous mycobacteria

Of the total number of isolates positive for mycobacteria, 241 (43%) were positive for MTC and 18 (3%) had non-tuberculous mycobacteria (NTM) (Table 1). To calculate the sensitivity of each system, we defined the ‘gold standard’ as being a specimen positive for MTC on at least one of the culture systems. Therefore, the sensitivity of each culture system for MTC isolation was as follows: M-MGIT, 92% (222/241); A-MGIT, 94% (227/241); L-J, 74% (179/241). There was no significant difference between the MGIT systems, but there was a statistically significant difference between each MGIT system and the L-J method (χ2=27.44 and 35.99; P <0.01).

From a total of 565 sputum samples that were processed using the M-MGIT, A-MGIT and L-J methods, MTC organisms were detected in 222 (39%), 227 (40%) and 179 (32%) isolates, respectively. (Table 1). There was no significant difference between the MGIT systems but the difference between each MGIT system and L-J method was statistically significant (χ2=7.15 and 8.86; P <0.05).

Contamination rates were statistically significantly higher for both MGIT systems compared with the L-J culture method (5.8%, 6.2% and 2.7%, respectively; χ2=7.05 and 8.37; P  < 0.01) but there was no significant difference between the two MGIT systems (Table 1). Of the 565 sputum samples, 77 smears were AFB positive and 488 were AFB negative (Table 2). Statistically significantly more AFB-negative samples were found to be MTC-positive in the two MGIT systems compared with the L-J method (34%, 34%, and 25%, respectively; χ2=8.30 and 9.50; P <0.01). There was no difference between the two MGIT systems. The sensitivities of the M-MGIT, A-MGIT and L-J methods were 92% (164/178), 94% (167/178) and 69% (123/178), respectively. Although there was no significant difference between the two MGIT methods, the difference between each MGIT method and the L-J method was statistically significant (χ2=30.22 and 36.01; P <0.01).

Table 2.

Overview of results from the three different culture systems in the analysis of sputum samples from 565 patients with suspected pulmonary tuberculosis

Smear Sample n
M-MGIT

A-MGIT

L-J

All systems

Positive growth
−ve or Con
Positive growth
−ve or Con
Positive growth
−ve or Con
Positive growth
−ve or Con
MTC NTM MTC NTM MTC NTM MTC NTM
+ve AFB 77 58 (75.3) 13 (16.9) 6 (7.8) 60 (77.9) 12 (15.6) 5 (6.5) 56 (72.7) 10 (13.0) 11 (14.3) 63 (81.8) 12 (15.6) 2 (2.6)
−ve AFB 488 164 (33.6) 2 (0.4) 322 (66.0) 167 (34.2) 4 (0.8) 317 (65.0) 123 (25.2) 1 (0.2) 364 (74.6) 178 (36.5) 6 (1.2) 304 (62.3)
Total 565 222 (39.3) 15 (2.6) 328 (58.1) 227 (40.2) 16 (2.8) 322 (57.0) 179 (31.7) 11 (1.9) 375 (66.4) 241 (42.6) 18 (3.2) 306 (54.2)

Data are presented as n or n (%).

M-MGIT™, manual Mycobacterial Growth Indicator Tube; A-MGIT, automated Mycobacterial Growth Indicator Tube (BACTEC™ MGIT™ 960); L-J, Löwenstein-Jensen culture method; AFB, acid-fast bacilli; +ve, positive; −ve, negative; Con, contaminated

With regard to AFB-positive samples, there was no significant difference between MTC-positive samples found by the two MGIT methods or the L-J method (i.e., 75%, 78% and 73%) (Table 2). In addition, there was no significant difference in the sensitivities of the three systems (92% [58/63] for M-MGIT, 95% [60/63] for A-MGIT and 89% [56/63] for L-J).

The agreement between M-MGIT and A-MGIT was 93% (i.e., [228 + 273+22]/565; kappa=0.87; 95% CI 0.79-0.94), and between M-MGIT and L-J was 84% ([177 + 285+10]/565; kappa=0.69; 95% CI 0.61-0.76) (Table 3).

Table 3.

Agreement between the three different culture systems in the analysis of sputum samples from 565 patients with suspected pulmonary tuberculosis


A-MGIT

L-J
Positive Negative Con Total Positive Negative Con Total
M-MGIT
 Positive, n 228 7 2 237 177 58 2 237
 Negative, n 11 273 11 295 7 285 3 295
 Contaminated, n 4 7 22 33 6 17 10 33
Total, n 243 287 35 565 190 360 15 565
Kappa, 95% CI 0.87 (0.79, 0.94) 0.69 (0.61, 0.76)

M-MGIT ™, manual Mycobacterial Growth Indicator Tube; A-MGIT, automated Mycobacterial Growth Indicator Tube (BACTEC™ MGIT™ 960); L-J, Löwenstein-Jensen culture method; Con, contaminated; CI, confidence interval

The mean ± standard deviation TTD was 12.9±4.2 (range 3-35) days for M-MGIT, 11.8±5.2 (range 4–33) days for A-MGIT and 24.2±8.7 (range 11–69) days for the L-J method. There was no significant difference between the two MGIT systems but the difference between each MGIT system and the L-J method was statistically significant (P <0.01).

Discussion

The MGIT system is based on fluorescence detection of mycobacterial growth in a tube containing a modified Middlebrook 7H9 medium together with a fluorescence quenching-based oxygen sensor.5 Consumption of the dissolved oxygen by the growing mycobacteria produces an orange fluorescence and its intensity is proportional to the number of bacteria present. 5While the automated system (BACTEC MGIT 960) is a fully automated, continuously monitoring, high-capacity instrument, the older manual system requires tubes to be examined for fluorescence manually under a Wood’s lamp or with some other long wave UV light source. 5 However, although the manual system requires more technical time than the automated system, it is less of a financial burden for low resource countries. 5

The results of this present study showed that while the M-MGIT and A-MGIT systems had a similar yield for mycobacteria and MTC, both methods had a significantly better yield than the conventional L-J culture medium. To avoid biased estimates of the test characteristics the presence of M. tuberculosis was defined using a composite reference standard8 whereby a positive culture in any medium was used as the ‘gold standard’. This practice has been used in several other studies.911 Using the derived ‘gold standard’, the sensitivity for MTC isolation was similar for the M-MGIT and A-MGIT systems (92% and 94%, respectively) and both systems were significantly higher than that of L-J culture method (74%). Although the ‘gold standard’ used in this present study may have been different from other studies, our findings are similar to results from several different countries. For example, in two studies where the ‘gold standard’ was positive L-J cultures, the sensitivity of M-MGIT was 92% in a study from Peru12 and 90% in a study from Malaysia.4 However, in a study where the ‘gold standard’ was consistent with ours, sensitivities of M-MGIT, A-MGIT and L-J were found to be 82%, 80%, and 47% respectively.13

The diagnostic performance of M-MGIT was good even for smear-negative sputum samples. Indeed, among the smear-negative samples, the M-MGIT and A-MGIT methods yielded 34% MTC growth, which was significantly higher than L-J culture medium (25%). The sensitivities of the MGIT systems for MTC in smear-negative samples were similar (M-MGIT, 92% and A-MGIT, 94%) and significantly greater than L-J medium (69%). However, a study from Peru that evaluated the performance of M-MGIT in the diagnosis of 542 smear-negative samples found a yield of 24% and a sensitivity of 85%. It is important to note that sensitivities of methods may be affected by the decontamination protocol.13 For instance, if specimens are decontaminated using excess sodium hydroxide or the decontamination time is too long, MTB may be killed, and so the positive rate will be low. By contrast, if specimens are decontaminated using an insufficient amount of sodium hydroxide or the decontamination time is short, other bacteria besides MTB may grow, and the rate of contamination will increase and the positive rate of MTB will decrease.

As a general rule, a contamination rate of 1–4% is acceptable in laboratories that receive fresh specimens,15 and the Chinese Antituberculosis Association suggests the contamination rate should be controlled at 2–5%.16 In the present study, the MGIT systems had similar contamination rates (6% ) but they were significantly higher than for the L-J culture method (3%). A possible explanation for the higher contamination with the MGIT systems compared with the L-J culture method, is that the medium in the MGIT systems is more enriched than the L-J medium.17

The TTD is an important feature since early identification of the bacteria allows timely treatment and can prevent the disease from spreading. In this study, the TTD for the MGIT systems was similar (approximately 12–13 days) and significantly shorter than for the L-J culture (approximately 24 days). Our findings are similar to those from other studies (i.e. 11–13 days for M-MGIT system and 21–33 days for LJ culture medium).4,7,13

In the diagnosis of TB, the M-MGIT system had a similar performance, turnaround time and percentage of contaminated cultures compared with the A-MGIT system. Although in our study the costs of the M-MGIT method (70 Chinese Yuan per tube and the additional costs for a manual fluorescence reader) were higher than those of the L-J culture method (30 Chinese Yuan per tube), they were much lower than those of the A-MGIT method (70 Chinese Yuan per tube and a million- Chinese Yuan for the equipment). Therefore, in spite of the higher cost compared with the L-J culture method, the M-MGIT system appears to be more cost-effective because of its higher efficiency.

The present study was limited by the fact that the potential agreement between A-MGIT system and L-J method was not evaluated. However, this has been studied elsewhere, 18and the focus of this study was the evaluation of the M-MGIT system. In addition, our results were obtained from one centre. Therefore, further multicentre studies are required.

In summary, the M-MGIT system has a similar diagnostic performance to the A-MGIT system, with a relatively low contamination rate and short TTD compared with the LJ method. Therefore, the M-MGIT culture system offers a fast and reliable alternative to the conventional solid medium culture system for the diagnosis of pulmonary TB in a developing country.

Acknowledgments

The authors would like to thank Dr Yan Zhu and Dr Fang He (Tuberculosis Clinic, Chaoyang District Centre for Disease Control and Prevention) for data and specimen collection.

Declaration of conflicting interest

The authors declare that there are no conflicts of interest.

Funding

This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.

References

  • 1.World Health Organization. Global tuberculosis report 2016. Geneva: WHO, WHO/HTM/TB/2016.13. [Google Scholar]
  • 2.Lawn SD, Mwaba P, Bates M, et al. Advances in tuberculosis diagnostics: the Xpert MTB/RIF assay and future prospects for a point-of-care test. Lancet Infect Dis 2013; 13: 349–361. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Ryu YJ. Diagnosis of pulmonary tuberculosis: recent advances and diagnostic algorithms. Tuberc Respir Dis (Seoul) 2015; 78: 64–71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Fadzilah MN, Ng KP, Ngeow YF. The manual MGIT system for the detection of M. tuberculosis in respiratory specimens: an experience in the University Malaya Medical Centre. Malays J Pathol 2009; 31: 93–97. [PubMed] [Google Scholar]
  • 5.Palomino JC, Martin A, Von Groll A, et al. Rapid culturebased methods for drug-resistance detection in Mycobacterium tuberculosis. J Microbiol Methods 2008; 75: 161–166. [DOI] [PubMed] [Google Scholar]
  • 6.Ministry of Health of P.R. China. WS288-2008, Diagnostic criteria for pulmonary tuberculosis. Beijing, China: People’s Medical Publishing House; 2008. http://www.codeofchina.com/standard/WS288-2008.html
  • 7.Al-Mutairi N, Ahmad S, Mokaddas E. Performance comparison of four methods for rapid detection of multidrug-resistant Mycobacterium tuberculosis strains. Int J Tuberc Lung Dis 2011; 15: 110–115. [PubMed] [Google Scholar]
  • 8.Banoo S, Bell D, Bossuyt P, et al. Evaluation of diagnostic tests for infectious diseases: general principles. Nat Rev Microbiol 2006; 4: S20–S32. [DOI] [PubMed] [Google Scholar]
  • 9.Chew WK, Lasaitis RM, Schio FA, et al. Clinical evaluation of the Mycobacteria Growth Indicator Tube (MGIT) compared with radiometric (Bactec) and solid media for isolation of Mycobacterium species. J Med Microbiol 1998; 47: 821–827. [DOI] [PubMed] [Google Scholar]
  • 10.Somoskövi A, Magyar P. Comparison of the mycobacteria growth indicator tube with MB redox, Löwenstein-Jensen, and Middlebrook 7H11 media for recovery of mycobacteria in clinical specimens. J Clin Microbiol 1999; 37: 1366–1369. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Boehme CC, Nabeta P, Hillemann D, et al. Rapid molecular detection of tuberculosis and rifampin resistance. N Engl J Med 2010; 363: 1005–1015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Quispe R, Valle GA, Huapaya JA, et al. Manual MGIT™ system for the detection of Mycobacterium tuberculosis: insights from a high TB burden setting. Int J Tuberc Lung Dis 2016; 20: 605–610. [DOI] [PubMed] [Google Scholar]
  • 13.Muyoyeta M, Schaap JA, De Haas P, et al. Comparison of four culture systems for Mycobacterium tuberculosis in the Zambian National Reference Laboratory. Int J Tuberc Lung Dis 2009; 13: 460–465. [PubMed] [Google Scholar]
  • 14.Battaglioli T, Soto A, Agapito J, et al. Manual liquid culture on simple Middlebrook 7H9 or MGIT for the diagnosis of smear-negative pulmonary tuberculosis. Trop Med Int Health 2014; 19: 1500–1503. [DOI] [PubMed] [Google Scholar]
  • 15.Somoskövi A, Ködmön C, Lantos A, et al. Comparison of recoveries of mycobacterium tuberculosis using the automated BACTEC MGIT 960 system, the BACTEC 460 TB system, and Löwenstein-Jensen medium. J Clin Microbiol 2000; 38: 2395–2397. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Song YY, Zheng HW, Zhao YL. The development of TB laboratory in China. Chin J Antituberc 2014; 36: 764–768 [in Chinese, English Abstract]. [Google Scholar]
  • 17.Zhao P, Yu Q, Chen L, et al. Evaluation of a liquid culture system in the detection of mycobacteria at an antituberculosis institution in China; A retrospective study. J Int Med Res 2016; 44: 1055–1060. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Diriba G, Kebede A, Yaregal Z, et al. Performance of Mycobacterium Growth Indicator Tube BACTEC 960 with Lowenstein-Jensen method for diagnosis of Mycobacterium tuberculosis at Ethiopian National Tuberculosis Reference Laboratory, Addis Ababa, Ethiopia. BMC Res Notes 2017; 10:181. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from The Journal of International Medical Research are provided here courtesy of SAGE Publications

RESOURCES