. 2017 Mar 30;9(1):129–132. doi: 10.1111/os.12317

Rod Migration to the Thoracic Subarachnoid Space after C1–2 Instrumentation: A Case Report and Literature Review

Seref Dogan 1,, Elif Basaran Gundogdu 1, Mevlüt Özgür Taşkapılıoğlu 1, Ahmet Karaoglu 1
PMCID: PMC6584166  PMID: 28371499

Abstract

Posterior instrumented fusion of the cervical spine is a common surgical procedure in the treatment of cervical subluxation, fractures, and stenosis. Although malpositions are commonly seen, it is rare to observe the malposition of the rod or interconnection because of hardware failure. A 62‐year‐old woman with spastic tetraparesis as a sequel to pediatric meningitis with C1–C2 cervical subluxation and myelomalacia had undergone laminectomy of C1 and C1 lateral mass and C2 bilateral pedicular screw fixation. Three years after the stabilization, she presented with complaints of headache, neck pain, and difficulty walking. There was no history of trauma during that period. A previously unrecorded and unusual migration of a rod through the thoracic subarachnoid space was detected. In this study, we report a case of atlantoaxial stabilization using the screw‐rod technique that was followed by rod migration to the thoracic subarachnoid space, and outline the subsequent management of the case. Failure of bony fusion can result in micromotion and subsequent migration of fixation device components. Routine radiographic follow‐up could be used to identify migration events.

Keywords: Atlantoaxial stabilization, Cervical stabilization, Complication, Migration, Posterior cervical stabilization

Introduction

Posterior cervical stabilization is a method commonly applied in the treatment of several pathologic conditions (e.g. stenotic cervical canal, cervical fracture, or cervical mass) accompanied with cervical instability. Wiring, interlaminar clamps, lateral mass screws, and cervical pedicle screws are among the techniques and instrumentation options that can be used for posterior stabilization of the subaxial cervical spine1, 2, 3. Wire fixation is an inexpensive and ancient method that requires little expertise, but is not effective if the patient have osteoporotic bones, or the posterior elements are not secure4. In many cases, absent or broken posterior elements preclude the use of wires; thus, rigid stabilization devices were developed for better fusion rates.

Lateral mass screws were first introduced by Roy‐Camille et al. in 19795. The lateral mass screws made it possible to place longer screws due to their angulated trajectories toward the superior‐lateral‐ventral corner of the lateral mass3. In a human cadaveric study, Coe et al. studied the stability of the lateral mass plate and confirmed the results of Roy‐Camille et al. that lateral mass fixation increases segmental stability in flexion by 92% and in extension by 60%6, 7. However, various complications and malpositioning may be encountered postoperatively. Although malpositions (e.g. screw malpositions) are commonly seen, it is rare to observe the malposition of the rod or interconnection because of hardware failure. Here we present a case of atlantoaxial stabilization using the screw‐rod technique that was followed by rod migration to the thoracic subarachnoid space in the late postoperative period.

Case Report

A 62‐year‐old woman with mental retardation was admitted to our outpatient department with spastic tetraparesis as a sequel to pediatric meningitis; she had no history of additional systemic diseases. After having been diagnosed with C1–C2 cervical subluxation and myelomalacia using cervical magnetic resonance imaging (MRI) (Fig. 1), the patient underwent laminectomy of C1, and the C1 lateral mass and C2 were fixed using bilateral pedicular screws (Oasys, Stryker, New Jersey, USA) on 20 June 2011 (Fig. 2A). The final tightening was performed with a torque screw driver. After drilling the C1–2 joint, posterior arthrodesis was performed with bone chips by placing the chips into the joint space for fusion. The rod was not spread outside the screws to avoid touching the C1 arc, but it was an adequate fit for tightening. Our follow‐up policy after posterior cervical fusion was: 10 days, 1 month, 3 months, 1 year after the surgery and then annually. A control X‐ray was performed as part our routine at every follow‐up visit.

Figure 1.

Figure 1

Preoperative cervical T2WI MRI showing cervical subluxation and myelomalasic changes at C1–C2.

Figure 2.

Figure 2

(A) Postoperative radiography after C1 laminectomy and posterior stabilization at the level of C1–C2. (B) Control radiography demonstrating missing rod 3 years after the surgery.

Postoperatively, the patient's complaints were resolved and no radiological or additional neurological findings were determined at her follow‐up period, until she presented with complaints of headache, neck pain and walking difficulty, all of which started at the end of the third postoperative year (Fig. 2B). Physical examination revealed tetraparesis with 4/5 muscle strength. There was no history of trauma during that period. On radiological evaluation, the left rod was not detected, but there was no apparent sign of compression on MRI. Thoracal X‐ray and computed tomography demonstrated an intradurally‐placed rod of 2 cm in length at the level of T7 vertebral body level (Fig. 3). On cervical exploration, the left rod was found to be displaced, the set screws were within the paravertebral muscles, and a dural defect was observed just above C1. Although fusion was achieved, the instruments were removed and the defect was repaired; we did not extend the fusion and did not recommend the use of a cervical collar in the postoperative period. A T7 laminectomy was performed as a second‐look surgery and the migrated rod was removed, which was situated intradurally at that level (Fig. 4). No additional complications were encountered after the operation. At the third year follow‐up after the second surgery the patient had no additional neurological deficit (Fig. 5).

Figure 3.

Figure 3

(A) Thoracal X‐ray and (B, C) computed tomography demonstrate a rod of 2‐cm length located at the level of T7.

Figure 4.

Figure 4

Intraoperative photograph showing intradurally migrated rod at T7.

Figure 5.

Figure 5

Cervical T2WI MRI showing cervical fusion with no additional neurological deficit at postoperative third year.

Discussion

The instruments used for posterior cervical stabilization have recently shown a rapid evolution and the lateral mass C1–2 screw‐rod technique is relatively new among them. Ease of rod manipulation in all three dimensions, more precise placement due to the flexibility during screw placement, and the resulting lower failure (malposition) rate are the most important advantages of this technique8. The malposition rate is reported between 2 and 5% and is seen mostly as screw malposition9, 10. Plant and Ruff describe a migrated rod through the skull and into the cerebellum following C1–C2 instrumentation11. In this case, the rod was migrated through the thoracal region subdurally, and we speculated that the rod was migrated to the thoracic subdural region because of gravity.

The complication rate has been reported to decrease after the first postoperative year10. After atlantoaxial fixation, some authors report a minimum follow‐up of 2 years11. However, the hardware failure presented here manifested in the third postoperative year. Our follow‐up policy after posterior cervical fusion was: 10 days, 1 month, 3 months, 1 year after the surgery, and then annually. The control X‐ray is routine at every follow‐up visit. The patient quit the follow‐up after the first year postoperatively and was admitted to hospital at the third year postoperatively with neck pain, headache, and walking difficulty. There was no indication of hardware failure 1 year after surgery. The screw heads were not broken, so we speculate that the rod migration was due to an absence of nuts. This case is likely to be a result of hardware failure, rather than an error by the surgeon, because a surgeon's error would have manifested itself earlier. It is important to remember that, even after the 1‐year regular postoperative follow‐up, the possibility of a late hardware failure (and its potentially life‐threatening effects) persists. The experience of the surgeon and knowledge of the 3D anatomy of the region are key factors in decreasing complication rates. In the presented case, early diagnosis and intervention prevented further morbidity.

Conclusion

Although the lateral mass screw‐rod technique has high success and low complication rates in the hands of an experienced surgeon, care should be taken for prevention and early diagnosis of probable hardware failure.

Disclosure: The authors declare no conflict of interest with respect to this study.

References

  • 1. Benzel EC, Kesterson L. Posterior cervical interspinous compression wiring and fusion for mid to low cervical spinal injuries. J Neurosurg, 1989, 70: 893–899. [DOI] [PubMed] [Google Scholar]
  • 2. Abumi K, Kaneda K, Shono Y, Fujiya M. One‐stage posterior decompression and reconstruction of the cervical spine by using pedicle screw fixation systems. J Neurosurg, 1999, 90: 19–26. [DOI] [PubMed] [Google Scholar]
  • 3. Yoshihara H, Passias PG, Errico TJ. Screw‐related complications in the subaxial cervical spine with the use of lateral mass versus cervical pedicle screws: a systematic review. J Neurosurg Spine, 2013, 19: 614–623. [DOI] [PubMed] [Google Scholar]
  • 4. Wellman BJ, Follett KA, Traynelis VC. Complications of posterior articular mass plate fixation of the subaxial cervical spine in 43 consecutive patients. Spine (Phila Pa 1976), 1998, 23: 193–200. [DOI] [PubMed] [Google Scholar]
  • 5. Roy‐Camille R, Gaillant G, Bertreaux D. Early management of spinal injuries In: McKibben B, ed. Recent Advances in Orthopedics. Edinburgh: Churchill‐Livingstone, 1979; 57–87. [Google Scholar]
  • 6. Coe JD, Warden KE, Sutterlin CE 3rd, McAfee PC. Biomechanical evaluation of cervical spinal stabilization methods in a human cadaveric model. Spine (Phila Pa 1976), 1989, 14: 1122–1131. [DOI] [PubMed] [Google Scholar]
  • 7. Roy‐Camille R, Saillant G, Mazel C. Internal fixation of the unstable cervical spine by a posterior osteosynthesis with plates and screws In: Sherk HH, Dunn HJ, Eismont FJ, et al., eds. The Cervical Spine, 2nd edn. Philadelphia: JB Lippincott Inc, 1989; 390–404. [Google Scholar]
  • 8. Deen HG, Birch BD, Wharen RE, Reimer R. Lateral mass screw‐rod fixation of the cervical spine: a prospective clinical series with 1‐year follow‐up. Spine J, 2003, 3: 489–495. [PubMed] [Google Scholar]
  • 9. Deen HG, Nottmeier EW, Reimer R. Early complications of posterior rod‐screw fixation of the cervical and upper thoracic spine. Neurosurgery, 2006, 59: 1062–1067. [DOI] [PubMed] [Google Scholar]
  • 10. Elliott RE, Tanweer O, Boah A, et al. Atlantoaxial fusion with screw‐rod constructs: meta‐analysis and review of literature. World Neurosurg, 2014, 81: 411–421. [DOI] [PubMed] [Google Scholar]
  • 11. Plant JG, Ruff SJ. Migration of rod through skull, into brain following C1‐C2 instrumental fusion for os odontoideum: a case report. Spine (Phila Pa 1976), 2010, 35: 90–92. [DOI] [PubMed] [Google Scholar]