Skip to main content
Wiley Open Access Collection logoLink to Wiley Open Access Collection
. 2018 Oct 3;42(12):1164–1173. doi: 10.1111/aor.13278

Requirements and Pitfalls of Dialyzer Sieving Coefficients Comparisons

Michael Hulko 1,, Ulrike Haug 1, Julia Gauss 1, Adriana Boschetti‐de‐Fierro 1, Werner Beck 2, Bernd Krause 1
PMCID: PMC6585607  PMID: 30281162

Abstract

Sieving coefficients reported in dialyzer data sheets and instructions for use (IFUs) indicate the potential of different solutes to pass across a particular membrane. Despite being measured in vitro, sieving coefficient data are often used as a predictor of the clinical performance of dialyzers. Although standards for the measurement of sieving coefficients exist, the stated methodologies do not offer sufficient guidance to ensure comparability of test results between different dialyzers. The aim of this work was to investigate the relationship between sieving coefficients and published clinical performance indicators for two solutes, albumin loss and beta‐2 microglobulin (β2M) reduction ratio (RR), and to assess the impact of different in vitro test parameters on sieving coefficient values for albumin, β2M, and myoglobin. Clinical albumin loss and β2M RR for commercially available dialyzers used in hemodialysis (HD) and post‐dilution hemodiafiltration (HDF) were extracted from the literature and plotted against sieving coefficients reported in data sheets and IFUs. Albumin, β2M, and myoglobin sieving coefficients of a selection of dialyzers were measured per the ISO 8637 standard. The impact of in vitro testing conditions was assessed by changing blood flow rate, ultrafiltration (UF) rate, sampling time, and origin of test plasma. Results showed variation in albumin loss and β2M RR for the same sieving coefficient across different dialyzers in HD and HDF. Changes in blood flow rates, UF rates, sampling time, and test plasma (bovine vs. human) caused marked differences in sieving coefficient values for all investigated solutes. When identical testing conditions were used, sieving coefficient values for the same dialyzer were reproducible. Testing conditions have a marked impact on the measurement of sieving coefficients, and values should not be compared unless identical conditions are used. Further, variability in observed clinical data in part reflects the lack of definition of test conditions.

Keywords: Albumin, Beta‐2 microglobulin, Dialyzer, Hemodialysis, Membrane, Myoglobin, Renal replacement therapy, Sieving coefficient, Hemodiafiltration, In vitro testing


Sieving coefficients indicate the potential of different solutes (e.g. albumin, beta‐2 microglobulin [β2M], or myoglobin) to pass across a particular dialyzer membrane 1. Two international standards (ISO 8637/EN 1283) 2, 3 are commonly referenced for the measurement of in vitro sieving coefficients.

Sieving coefficients for different dialyzers are often compared with each other, as a result of the perception that citing the above standards is sufficient to assure data comparability. Although referencing standards implies that clear and unambiguous guidance on test procedures is provided to achieve consistent testing across laboratories, they still allow for a certain variability of testing conditions. In the specific case of sieving coefficients, it is questionable whether the standards listed above provide sufficient guidance to enable comparability.

Furthermore, in vitro sieving coefficient values are often used as surrogates of solute removal performance, and to inform decisions that may impact clinical practice, despite not being measured using the same test conditions. In fact, a sieving coefficient is only one of many factors, including blood composition, dialyzer design, and treatment modality, that can impact solute loss and removal during dialysis 4, 5, 6.

The aims of this study were: (i) to assess the impact of different test parameters on sieving coefficient values within the framework provided and (ii) to investigate the relationship between sieving coefficient values and clinical albumin loss/β2M reduction ratio (RR), as a predictor of solute removal performance.

METHODS

Literature review analysis of published solute removal and reported sieving coefficient values

Clinical data published between 2002 and April 2017 for albumin loss or β2M RR were identified using two searches: (i) OVID searches of Embase and MEDLINE databases and (ii) searches of abstract archives for ERA‐EDTA, ASN, and ASAIO congresses. The search was limited to human, clinical data only. Publications were only included if an identifiable dialyzer was noted (and had publicly available sieving coefficient data, e.g., as mentioned in a data sheet, instructions for use [IFU], or manufacturer's brochure) and if the dialyzer was high‐flux (high‐cut‐off dialyzers were excluded). Data on albumin loss (including both mean or median values per treatment) or β2M RR in patients during hemodialysis (HD) or hemodiafiltration (HDF) sessions were also required. For studies with the HDF treatment modality, only post‐dilution was included, with convective volumes >15L or infusion volumes >14L (assuming weight loss of ≥1kg per treatment session); publications with no HDF volumes reported were excluded. Duplicates of study data published in abstract form at different congresses and publications were removed.

Data on albumin loss and β2M RR during dialysis were extracted from the identified publications and plotted against sieving coefficients as reported on data sheets and IFUs for the dialyzers studied. For these analyses, sieving coefficient values listed as “less than” were considered equal to the stated value, but are clearly marked to aid comparison.

Analysis of the impact of test parameters on sieving coefficient values

Dialyzers used in these analyses were: Revaclear 400, Polyflux 170H (Baxter International Inc., Hechingen, Germany), Xevonta Hi18 (B. Braun Medical Ltd., Melsungen, Germany), Elisio 17H (Nipro Medical Corporation, Bridgewater, NJ, USA), FX CorDiax 80, and FX 100 classix (Fresenius Medical Care, Bad Homburg, Germany).

For the first series of experiments, only the Revaclear dialyzer was used to assess the impact of testing conditions on the sieving coefficients of albumin, β2M, and myoglobin; tests were repeated six times to ensure robust data. For the second series of experiments, albumin, β2M, and myoglobin sieving coefficients of all the dialyzers listed above were determined.

For both sets of experiments, sieving coefficients were measured per the methodology stated in ISO 8637, with a recirculating fluid apparatus that provides a test setup comparable to hemofiltration mode (Supporting Information Fig. 1). Prior to testing, filters were primed with 0.9% NaCl solution. Both human plasma (Octapharma, Langenfeld, Germany) and bovine plasma (Kraeber & Co., Ellerbek, Germany) were used in separate experiments and had a specified total protein content of 60 ± 5g/L. For the first series of experiments, blood flow rate (Q B) and ultrafiltration rate (UF) were set according to predefined parameters (human plasma with Q B of 300mL/min, UF 60mL/min; human plasma with Q B/UF 500/100mL/min and bovine plasma with Q B/UF 300/60mL/min). For the second series of experiments, sieving coefficients were measured using human plasma with Q B/UF of 300/60mL/min. For both series of experiments, Q B/UF were set (at the aforementioned levels) and kept at a constant and stable pressure; temperature and flow rates were verified. Samples were drawn 60 min after setting blood flow. Stability of test conditions was verified twice, 10 min after experimental setup and prior to drawing the samples.

To investigate the influence of the time of sampling on sieving coefficient, samples were taken after 15, 30, 45, and 60 min (n = 3). Sieving coefficients were calculated as solute concentration on the filtrate side, divided by the average solute concentration as measured in the blood inlet and blood outlet sides. Concentrations of albumin, β2M, or myoglobin were analyzed using turbidimetric assays on a BN Prospec clinical analyzer (Siemens, Marburg, Germany) with the respective reagent kits used according to the manufacturer's instructions. Myoglobin concentration in bovine plasma was analyzed photometrically on a GE Ultrospec 8000 photometer (GE Healthcare, Frankfurt, Germany) via light absorbance at 405nm. Bovine serum albumin was quantified using a Cobas Mira clinical autoanalyzer (Roche Diagnostics, Penzberg, Germany), system with an albumin reagent kit (LT SYS, Berlin, Germany), based on the bromocresol green method and using protein standards as a reference.

Statistical methodology

The relationship between published solute loss/RRs and the reported sieving coefficient values was analyzed by linear regression, and correlation coefficients were tested for statistical significance. Measured sieving coefficient data are displayed as mean values of six independent replicates plus/minus standard deviation. Normal distribution and equal variance of the data were verified using Shapiro–Wilk and equal variance tests. Data that were normally distributed with equal variance were analyzed using a t‐test; P values less than 0.05 were considered significant. Data that were not normally distributed were analyzed with a Mann–Whitney rank sum test. Statistical calculations were conducted using SigmaPlot (Version 12.5, Systat Software Inc., San Jose, CA, USA).

RESULTS

Analysis of published solute loss and reported sieving coefficient values

For albumin loss, 31 publications were identified that met the search criteria, providing data from treatments in HD mode or post‐dilution HDF mode. Albumin loss data were available for 24 dialyzers used in HDF mode and 13 dialyzers in HD mode. In HD mode, treatment times ranged from 199 to 295 min. In post‐dilution HDF mode, treatment times ranged from 224 to 294 min, infusion volumes ranged from 14 to 27.9L, and UF volumes ranged from 15 to 29.9L. A significant correlation could be seen between albumin loss and reported sieving coefficients (as listed in data sheets and IFUs) for HD treatment modalities (Supporting Information Table; n = 21; P = 0.0316; R 2 = 0.2208). For HDF treatment modalities, reported albumin sieving coefficients did not show a significant correlation with clinical data, that is, albumin loss as reported in the literature (Supporting Information Table; n = 63; P = 0.2318; R 2 = 0.0234). The majority of the reported sieving coefficient values in data sheets and IFUs were reported as “less than” values (e.g., “<0.01”), making comparison between dialyzers and their clinical performance more difficult as specific sieving coefficients were not provided.

For β2M removal, 12 publications were identified that met the search criteria (including treatments in HD mode or post‐dilution HDF mode). In post‐dilution HDF mode, treatment times ranged from 239 to 264 min, infusion volumes ranged from 14 to 28.7L, and UF volumes ranged from 15 to 30.6L. In HD mode, treatment times ranged from 180 to 264 min. Sieving coefficient data were available for eight dialyzers used in HDF mode and 11 dialyzers in HD mode. No correlation was identified between β2M RR and the reported sieving coefficient in HD mode (n = 21; P = 0.3455; R 2 = 0.04693), but a weak, statistically significant correlation was evident in HDF mode (P = 0.0002; R 2 = 0.42915; Supporting Information Table).

Analysis of the impact of test parameters on sieving coefficient values

Changes in Q B/UF flow rate and test plasma (human vs. bovine) had a marked impact on the sieving coefficient values for β2M, myoglobin, and albumin observed for the Revaclear dialyzer (Table 1). Human plasma at Q B/UF 300/60mL/min resulted in the highest β2M and myoglobin sieving coefficient measurements of the three sets of testing conditions assessed. Human plasma with Q B/UF of 500/100mL/min resulted in the lowest (0.16%) sieving coefficient for albumin, while bovine plasma with a Q B/UF of 300/60mL/min produced the highest (0.69%; Table 1).

Table 1.

The impact of different testing conditions on Revaclear dialyzer sieving coefficients

Albumin β2M Myoglobin
Testing conditions Mean SC (%) SD Mean SC (%) SD Mean SC (%) SD
Human plasma Q B/UF 300/60mL/min 0.27[1,2] 0.04 95[3,4] 4 68[1] 1
Bovine plasma Q B/UF 300/60mL/min 0.69[1] 0.04 78[3] 4 58[5] 4
Human plasma Q B/UF 500/100mL/min 0.16[2] 0.01 85[4] 3 54[6] 3

β2M, beta‐2 microglobulin; Q B, blood flow rate; SC, sieving coefficient; SD, standard deviation; UF, ultrafiltration rate. Superscript numbers indicate specific comparisons and statistical results.

[1,3,4] t‐test: significant difference; P < 0.001.

[2,5,6]Mann–Whitney rank sum test; P = 0.002.

Sample number = 6.

The influence of sampling time on albumin, myoglobin, and β2M sieving coefficients was analyzed with Revaclear dialyzers using human plasma and testing conditions of Q B/UF 300/60mL/min (sample n = 3). Results are shown in Table 2. Sampling time had the most obvious impact on the albumin sieving coefficient; after 60 min, the sieving coefficient was only 46% of the measurement taken at 15 min. For myoglobin, the effect of sampling time was still visible but less pronounced than for albumin. No effect of sampling time was demonstrated for β2M, and the β2M sieving coefficient remained stable throughout the experiment (Table 2).

Table 2.

Revaclear sieving coefficients for albumin, myoglobin, and β2M after different times of sampling (human plasma, QB 300mL/min, UF 60mL/min)

Albumin β2M Myoglobin
Time (min) Mean SC (%) SD Mean SC (%) SD Mean SC (%) SD
15 0.59 0.09 93 1 79 1
30 0.40 0.07 92 3 71 2
45 0.31 0.06 92 4 68 1
60 0.27 0.07 91 4 67 1

β2M, beta‐2 microglobulin; Q B, blood flow rate; SC, sieving coefficient; SD, standard deviation; UF, ultrafiltration rate.

Sample number n = 3.

Comparison of published and measured sieving coefficient values

Albumin, β2M, and myoglobin sieving coefficients were measured for all dialyzers, using identical testing conditions (Q B/UF 300/60mL/min; human plasma). Of the six dialyzers investigated, a specific sieving coefficient for albumin was only available for one dialyzer, with the other five listed as having sieving coefficients less than a certain value, for example, <1%, which would allow the actual value to be as high as 0.99%. Specific sieving coefficients for albumin generated by measurements using human plasma were lower than the maximum values that comply with the “less than” values reported in data sheets and IFUs (Table 3).

Table 3.

Sieving coefficients of commercial dialyzers generated using identical testing conditions (human plasma, QB 300mL/min, UF 60mL/min)

Albumin β2M Myoglobin
Dialyzer Mean SC (%) SD Mean SC (%) SD Mean SC (%) SD
Revaclear 400 * 0.27 (DS: <1) 0.04 95 (DS: 70) 4 68 (DS: NR) 1
Polyflux 170H * 0.22 (DS: <1) 0.10 82 (DS: 70) 6 37 (DS: NR) 7
Xevonta Hi18 0.04 (DS: <0.1) 0.01 84 (DS: >80) 6 36 (DS: NR) 3
Elisio 17H 0.16 (DS: 0.2) 0.03 96 (DS: NR) 2 72 (DS: 22) 2
FX CorDiax 80 * 0.04 (DS: <0.1) 0.01 92 (DS: 90) 4 50 (DS: 50) 4
FX 100 classix * 0.07 (DS: <0.1) 0.1 70 (DS: 70) 2 10 (DS: 10) 0.2

β2M, beta‐2 microglobulin; DS, sieving coefficient value as displayed in the product data sheet; NR, not reported; Q B, blood flow rate; SC, sieving coefficient; SD, standard deviation; UF, ultrafiltration rate. A value preceded by “<” indicates that a specific sieving coefficient value is not included on the data sheet; rather it is less than the value indicated. A value preceded by “>” indicates that a specific sieving coefficient value is not included on the data sheet; rather it is greater than the value indicated.

*Q B/UF and test solution not identified in the data sheet.

Data sheet values taken from product documentation 79, 1619.

Q B/UF listed as 300/60mL/min in the data sheet; test solution not identified.

Sample number = 6.

Specific β2M sieving coefficients were available for four of the dialyzers, while a “greater than” value was provided for one dialyzer and no value was provided for one other. The β2M sieving coefficients reported here were higher than those listed in data sheets and IFUs, where provided, except for FX 100 classix, which was equivalent to that reported in the FX classix product brochure 7. Sieving coefficient values for β2M ranged from 70 to 96%, with Revaclear 400 and Elisio 17H showing the highest values (95% and 96%, respectively), despite Revaclear 400 having the lowest reported sieving coefficient in its data sheet/IFU (70%; Table 3).

Sieving coefficient data for myoglobin are not always listed in data sheets; therefore, no comparison with reported values can be made. Data ranged from 10 to 72%, with Revaclear 400 and Elisio 17H having the highest values (68% and 72%, respectively; Table 3).

Comparison of measured coefficient values vs. reported clinical albumin loss and β2M RR

For both treatment modalities, large differences were seen between many of the measured albumin sieving coefficient values and those reported in the IFUs/data sheets (Fig. 1). For example, in HD mode, two of the dialyzers with reported albumin sieving coefficient values of <1% had sieving coefficient values of 0.27% and 0.22%, respectively, when measured under identical conditions. The same was true for dialyzers tested in HDF mode. The differences were particularly evident for sieving coefficient values that were reported as “less than” figures (Supporting Information Table).

Figure 1.

Figure 1

Correlation of clinical albumin loss in hemodialysis (HD) mode and sieving coefficient, measured with human plasma 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19. Albumin sieving coefficients are plotted for a range of dialyzers used in clinical studies that provided albumin removal data, as reported in the literature. Sieving coefficients were measured by Baxter using identical testing conditions, per the methodology stated in ISO 8637, and using human plasma with Q B/UF of 300/60mL/min.

Comparison of the sieving coefficients with published albumin loss data showed significant correlations in HD mode, with an improvement in the correlation when sieving coefficients were measured with the same test parameters (n = 12 published: R 2 = 0.4216, P = 0.032 vs. measured R 2 = 0.9655, P ≤ 0.001; Fig. 1). However, in HDF mode no significant correlation was seen (n = 35 published: R 2 = 0.0504, P = 0.195 vs. measured R 2 = 0.036; P = 0.275; Supporting Information Table).

When β2M sieving coefficients were compared with β2M RR in HD mode, the correlation was not significant, although it was close to the significance threshold when sieving coefficients were measured with the same test parameters (n = 16, reported: R 2 = 0.0011, P = 0.905 vs. measured: R 2 = 0.2315, P = 0.0592; Fig. 2A). However, for β2M RR in HDF mode, a significant correlation could be seen, but only when sieving coefficients were measured with the same test parameters (n = 14, reported: R 2 = 0.1231, P = 0.218 vs. measured: R 2 = 0.5318, P = 0.0031; Fig. 2B).

Figure 2.

Figure 2

Correlation of clinical beta‐2 microglobulin reduction ratio (β2M RR) data and sieving coefficient using (A) values reported in data sheets and instructions for use and (B) values measured under identical testing conditions (per the methodology stated in ISO 8637, using human plasma with Q B/UF of 300/60mL/min) 7, 8, 9, 10, 11, 12, 15, 19, 20, 21, 22, 23, 24, 25, 26. HD, hemodialysis; HDF, hemodiafiltration.

DISCUSSION

Sieving coefficients reported in dialyzer data sheets and IFUs, and hence those used in this analysis, indicate the potential of different solutes (e.g., albumin, β2M, or myoglobin) to pass across a particular membrane 1. Despite this definition and the fact that sieving coefficient data are generated in vitro, they are often used as a predictor of clinical solute removal performance. To our knowledge, there are no publications that have analyzed a possible correlation between sieving coefficient and clinical performance data, that is, highlighting the problem of extrapolating clinical performance from sieving coefficient data generated in vitro, given the weak relationship with clinical performance markers (i.e., albumin loss and β2M RR).

Our analyses of the published literature on solute loss and the sieving coefficients reported in manufacturers’ materials demonstrated that although correlations are sometimes identified, sieving coefficients do not always show a strong correlation to clinical performance indicators. Other factors such as the patient, treatment conditions, and the design of the dialyzer need to be considered as well 1, 4, 6, 27, 28, 29, 30. In vitro sieving coefficients have limitations as parameters on which to base decisions in clinical practice and clinicians and other key decision makers should consider all available clinical information, which may provide a more accurate representation of clinical performance of a dialyzer than technical data generated in vitro alone.

Variation in reported sieving coefficients values may not be the sole reason for the discrepancies seen in the correlation with solute loss. For example, even when measured using identical conditions, we did not observe any improvement in the correlation for albumin loss in HDF mode. This suggests that other parameters are likely to be more important for impacting the amount of albumin lost during treatment, such as treatment factors (e.g., UF volumes in relation to dialyzer surface area or blood flow rate), patient factors (e.g., blood viscosity), or the method used for assessing albumin loss. Furthermore, the borderline significance of the correlation between β2M RR in HDF mode and sieving coefficient values measured under the same conditions could be explained by factors relating to the patient. For example, based on the finding that β2M removal with HDF treatment was limited by the rate of intercompartmental transfer of β2M within the body, Ward et al. 31 recommended that altering the duration and frequency of treatment sessions, rather than increasing extracorporeal clearance in a given dialysis session, would be the most effective way of returning serum β2M levels to normal 31.

That there was no association between β2M sieving coefficient and RR in HD mode while there was a correlation in HDF mode that can be explained, in part, by the different mechanisms of transport across the membrane in these modalities. During HD mode, diffusion is the primary driving factor for mass transport and the other impacting factors, such as membrane permeability, will become less important, resulting in the lack of observed correlation. The opposite is true in HDF mode, where convection due to a transmembrane pressure gradient provides a strong driving force that makes the impact of other limiting factors, such as membrane permeability, more relevant; this results in a correlation between the β2M sieving coefficient and observed RR. However, it is important to note that—as observed here with albumin, a large molecule possessing a low sieving coefficient—multiple factors, such as membrane fouling over time, result in complex effects that limit the impact of small changes in sieving coefficient despite the strong driving force.

Our data also demonstrate the impact of different methodologies on generating in vitro sieving coefficient data and have important implications regarding the comparability of data from different sources. The recommended methods to measure sieving coefficients allow for variability in testing conditions in terms of plasma source (human vs. bovine), flow rates, and testing times. As we have observed, and building on the observations of other groups, all these factors can have a significant impact on the measured sieving coefficient, such that quite different values can be obtained for the same membrane and solute if conditions are altered. As such, decision makers should be aware of the limitations of comparing sieving coefficients reported in product documentation that have been obtained using different methods. For example, a change in flow rate caused a marked difference in results; when flow rates (Q B/UF) were lowered, sieving coefficient values increased. This finding is supported by work by Bresler et al. and Opong and Zydney, who reported that sieving coefficient values are filtrate flux and, therefore, flow dependent 32, 33.

Our results also highlight the impact that the origin of plasma can have on sieving coefficient values. Given this knowledge, it could be argued that bovine plasma is not the most appropriate test medium to use for studies of devices intended for use in humans. The exception to this would be in situations where comparability studies have been conducted to ensure results generated using bovine plasma are equivalent to those generated using human plasma. Work by Windberger et al. supports this conclusion, in that the authors analyzed blood and plasma properties and found that differences in viscosity and shear rate were apparent between different species of animals 34. Although the data published by Windberger et al. were not generated in the context of sieving coefficients, it is known that both viscosity and shear rates can affect solute removal in dialysis 4, 6. Besides this, it was also found that plasma protein composition can be different between human and bovine plasma. Human plasma tends to have higher concentrations of serum albumin and lower concentrations of globulins, particularly in the alpha 1‐globulin and beta‐globulin fractions, when compared with bovine plasma 35, 36. Differences in plasma composition can be considered as an explanation for the observed differences in plasma viscosity. Beyond that, difference in plasma composition can also have a direct impact on sieving coefficients by creating different layers of adsorbed proteins upon contact with membrane material. This evidence also brings into question the comparability of data generated using plasma from different species.

In this work, an impact of sampling time on sieving coefficients was shown. Reduction of dextran sieving coefficients upon blood contact was also reported in work by Langsdorf and Zydney, in which sieving coefficients stabilized after 40 min of contact with blood 37. This effect can be explained by the formation of a protein layer which has an impact on sieving properties of the membrane 38. Sampling time is typically not mentioned in conjunction with published sieving coefficient data. This lack of information adds to the difficulty of sieving coefficient comparability.

Our analyses also highlight that sieving coefficients generated using identical testing methods stated here were similar for some dialyzers, despite reported values in data sheets and IFUs being markedly different. Based on results reported here, it can be presumed that such differences can be attributed to variations in testing methodology. There are inconsistences between sieving coefficient values cited in product brochures and those reported in data sheets and IFUs 8, 9; to the authors’ knowledge, this is the first time that this finding has been formally reported specifically for commonly used dialyzers. Indeed, these findings reinforce the theory that sieving coefficients are subjected to variability (depending on testing conditions used) and should not be compared unless identical testing methods have been used.

There are limitations to the statistical analyses presented here. First, as with any analysis of data obtained from multiple studies, the differences in the trial methodologies will limit how comparable the data are, and this will in turn impact the statistical analyses that can be used. In addition, as these data were obtained from varying patient populations and treatment conditions (e.g., flow rates, treatment times, and substitution volumes in HDF mode), and as removal during dialysis does not have a linear correlation with many of these factors, it was not possible to normalize for these potential differences. This analysis is limited by which data are available, and an analysis of correlation can be skewed by limited data, especially at the extreme ends of the curve. Finally, in the case of albumin, exact sieving coefficient values are often not reported; instead, a range of values are provided in manufacturers’ materials (e.g., <0.01). Therefore, although a positive correlation is formally indicated in some cases, the clinical relevance of this could be questioned due to the large spread of albumin loss reported for the same sieving coefficient values. Differences between the methods used to measure clinical albumin loss also make it difficult to draw comparisons and can add to the variation seen. For example, albumin loss reported for the FX 1000 dialyzer ranges from 1.2 to 3.611g per session 39, 40, 41. It is possible that differences in the methods used to measure albumin in spent dialysate may also contribute to the variation in results seen 13, 39. Regardless of the potential limitations, there is a value to the statistical analyses; potential correlations were observed, which—in the absence of a direct comparative study—may help to support filter performance evaluations.

CONCLUSIONS

In summary, a lack of standardization of testing parameters makes it difficult to accurately compare sieving coefficients of different dialyzers; test parameters such as origin of plasma, Q B/UF, and time will all impact the final measured sieving coefficient. The results presented here, using standardized conditions, allow comparability of sieving coefficients of some commonly used commercial dialyzers and highlight the need for stricter standardization of experimental conditions for the measurements of sieving coefficients. Unless measured under identical conditions, sieving coefficients should not be compared. The limitations and influencing factors shown here must be taken into consideration when considering the impact of sieving coefficients on clinical solute removal performance.

Author Contributions

All authors contributed to the conception/design and the interpretation of the data and provided critical review of the manuscript draft.

Supporting information

Additional supporting information may be found online in the Supporting Information section at the end of the article.

FIG. S1. Setup for measuring sieving coefficients.

TABLE S1. Analyses of A) β2M reduction ratios and B) albumin loss vs. sieving coefficient values in HD and HDF treatment modes (sieving coefficient values reported in datasheets and IFUs vs. identical testing conditions).

Acknowledgments

Medical writing assistance was provided by Melissa Purves, Eleanor LeFeuvre, and Siobhán Ahern of SciMentum, UK, and funded by Baxter International Inc.

Conflict of Interest

All authors are employees of Gambro Dialysatoren GmbH (part of Baxter International Inc.).

REFERENCES

  • 1. Cheung AK, Ward RA. Hemodialyzers. ASN Dialysis Advisory Group. Available at: https://www.asn-online.org/education/distancelearning/curricula/dialysis/hemodialyzerscheung.pdf. Accessed October 10, 2017.
  • 2. ISO . BS EN ISO 8637:2014. Cardiovascular Implants and Extracorporeal Systems. Haemodialysers, Haemodiafilters, Haemofilters and Haemoconcentrators, 2014. Available at: https://shop.bsigroup.com/ProductDetail/?pid=000000000030280719. Accessed October 10, 2017.
  • 3. British Standards Institution . Haemodialysers, Haemodiafilters, Haemofilters, Haemoconcentrators and their Extracorporeal Circuits (BS EN 1283:1996), 1996. Available at: http://shop.bsigroup.com/ProductDetail/?pid=000000000000875837. Accessed October 10, 2017.
  • 4. Ronco C, Crepaldi C, Brendolan A, et al. Evolution of synthetic membranes for blood purification: the case of the Polyflux family. Nephrol Dial Transplant 2003;18:vii10–20. [DOI] [PubMed] [Google Scholar]
  • 5. Boschetti‐de‐Fierro A, Voigt M, Storr M, Krause B. MCO membranes: enhanced selectivity in high‐flux class. Sci Rep 2016;5:18448. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Ronco C, Technology of dialysis and associated methods In: Hörl WH, Koch K‐M, Lindsay RM, Ronco C, Winchester JF, eds. Replacement of Renal Function by Dialysis. New York: Kluwer Academic Publishers, 2013;201–725. [Google Scholar]
  • 7. Fresenius Medical Care . FX Classix. High‐Flux Dialysis For Improved Survival, 2012. Available at: https://www.freseniusmedicalcare.com.ar/fileadmin/data/masterContent/pdf/Healthcare_Professionals/Haemodialysis/Dialysers/FX_classix/FX-classix_Broschuere_08_14_MT-EN_w__RGB_.pdf. Accessed October 10, 2017.
  • 8.B. Braun. Xevonta Brochure. 7080148A. Available between 2012 and May 2017. Available at:http://www.bbraun-dialysis.com. Accessed 2017.
  • 9. Baxter POLYFLUX H Datasheet . EUMP/MG135/16‐0004 May 2016. Accessed October 10, 2017.
  • 10. Gayrard N, Ficheux A, Duranton F, et al. Consequences of increasing convection onto patient care and protein removal in hemodialysis. PLoS One 2017;12:e0171179. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Kirsch AH, Lyko R, Nilsson LG, Beck W, Amdahl M, Lechner P. Performance of hemodialysis with novel medium cut‐off dialyzers. Nephrol Dial Transplant 2017;32:165–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Gayrard N, Ficheux A, Duranton F, et al. Influence of high convection volumes in removal performances of on‐line haemodiafiltration. Nephrol Dial Transplant 2013;28:30–2. [Google Scholar]
  • 13. Ficheux A, Gayrard N, Szwarc I, et al. The use of SDS‐PAGE scanning of spent dialysate to assess uraemic toxin removal by dialysis. Nephrol Dial Transplant 2011;26:2281–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Pellicano R, Polkinghorne KR, Tuominen O, Woods HF, Kerr PG. Crossover trial of porous dialysis membrane versus conventional high‐flux membrane: assessment of middle molecule clearance and nutritional parameters. Nephrology 2007;12:1071. [Google Scholar]
  • 15. Ward RA, Ouseph R, Modification of membrane characteristics allows a reduction in dialyzer membrane area without loss of performance. Proceedings of the 40th American Society of Nephrology Congress San Francisco, CA, USA, November 2–5, 2007.
  • 16. Fresenius Medical Care . FX CorDiax Brochure, 2012. Available at: https://www.freseniusmedicalcare.ae/fileadmin/data/masterContent/pdf/Healthcare_Professionals/Haemodialysis/Dialysers/FX_Cordiax/FX_CorDiax_Broch_08_14_MT-EN_w__RGB_.pdf. Accessed October 10, 2017.
  • 17. Baxter REVACLEAR Datasheet . EUMP/MG135/16‐0001b May 2016. Accessed October 10, 2017.
  • 18. Nipro Medical Corp . Elisio POLYNEPHRON: Research Program Overview, 2014. Available at: http://www.nipro.com/wp-content/uploads/2014/11/ELISIO-Clinical-Studies.pdf. Accessed October 10, 2017.
  • 19. Nipro Medical Corp . Elisio Product Brochure. Available at: www.renalfarma.lt/download.php/fileid/71. Accessed October 10, 2017
  • 20. Ouseph R, Hutchison CA, Ward RA. Differences in solute removal by two high‐flux membranes of nominally similar synthetic polymers. Nephrol Dial Transplant 2008;23:1704–12. [DOI] [PubMed] [Google Scholar]
  • 21. Eloot S, Van Biesen W, Dhondt A, et al. Impact of hemodialysis duration on the removal of uremic retention solutes. Kidney Int 2008;73:765–70. [DOI] [PubMed] [Google Scholar]
  • 22. Mandolfo S, Malberti F, Imbasciati E, Cogliati P, Gauly A. Impact of blood and dialysate flow and surface on performance of new polysulfone hemodialysis dialyzers. Int J Artif Organs 2003;26:113–20. [DOI] [PubMed] [Google Scholar]
  • 23. Uhlin F, Holmar J, Yngman‐Uhlin P, Fernström A, Fridolin I. Optical estimation of beta 2 microglobulin during hemodiafiltration—does it work? Blood Purif 2015;40:113–9. [DOI] [PubMed] [Google Scholar]
  • 24. Potier J, Queffeulou G, Bouet J. Are all dialyzers compatible with the convective volumes suggested for postdilution online hemodiafiltration? Int J Artif Organs 2016;39:460–70. [DOI] [PubMed] [Google Scholar]
  • 25. Ahrenholz PG, Winkler RE, Michelsen A, Lang DA, Bowry SK. Dialysis membrane‐dependent removal of middle molecules during hemodiafiltration: the beta2‐microglobulin/albumin relationship. Clin Nephrol 2004;62:21–8. [DOI] [PubMed] [Google Scholar]
  • 26. Macias N, Santos García A, Vega Martínez A, Abad Estébanez S, Goicoechea Diezhandino M, López Gómez JM. Importance of body water in the efficacy of convective solute transport in online hemodiafiltration. Ther Apher Dial 2017;21:88–95. [DOI] [PubMed] [Google Scholar]
  • 27. Maduell F, Navarro V, Cruz MC, et al. Osteocalcin and myoglobin removal in on‐line hemodiafiltration versus low‐ and high‐flux hemodialysis. Am J Kidney Dis 2002;40:582–9. [DOI] [PubMed] [Google Scholar]
  • 28. Maduell F, Arias M, Durán CE, et al. Nocturnal, every‐other‐day, online haemodiafiltration: an effective therapeutic alternative. Nephrol Dial Transplant 2012;27:1619–31. [DOI] [PubMed] [Google Scholar]
  • 29. Sargent JA, Gotch FA, Principles and biophysics of dialysis In: Jacobs C, Kjellstrand CM, Koch KM, Winchester JF, eds. Replacement of Renal Function by Dialysis. Dordrecht, The Netherlands: Springer, 1996;34–102. [Google Scholar]
  • 30. Teixeira Nunes F, de Campos G, Xavier de Paula SM, et al. Dialysis adequacy and nutritional status of hemodialysis patients. Hemodial Int 2008;12:45–51. [DOI] [PubMed] [Google Scholar]
  • 31. Ward RA, Greene T, Hartmann B, Samtleben W. Resistance to intercompartmental mass transfer limits beta2‐microglobulin removal by post‐dilution hemodiafiltration. Kidney Int 2006;69:1431–7. [DOI] [PubMed] [Google Scholar]
  • 32. Bresler EH, Mason EA, Wendt RP. Appraisal of equations for neutral solute flux across porous sieving membranes. Biophys Chem 1976;4:229–36. [DOI] [PubMed] [Google Scholar]
  • 33. Opong WS, Zydney AL. Diffusive and convective protein transport through asymmetric membranes. AIChE J 1991;37:1497–510. [Google Scholar]
  • 34. Windberger U, Bartholovitsch A, Plasenzotti R, Korak KJ, Heinze G. Whole blood viscosity, plasma viscosity and erythrocyte aggregation in nine mammalian species: reference values and comparison of data. Exp Physiol 2003;88:431–40. [DOI] [PubMed] [Google Scholar]
  • 35. Longo DL, Fauci AS, Kasper DL, Hauser SL, Jameson JL, Loscalzo J, eds. Harrison's Principles of Internal Medicine, 18th Edition. New York: McGraw‐Hill Education, 2012;A1–A15. [Google Scholar]
  • 36. Fielder S . Serum biochemical reference ranges [MSD Veterinary Manual Appendix], 2016. Available at: http://www.msdvetmanual.com/appendixes/reference-guides/serum-biochemical-reference-ranges#v5934790. Accessed October 10, 2017.
  • 37. Langsdorf LJ, Zydney AL. Effect of blood contact on the transport properties of hemodialysis membranes: a two‐layer membrane model. Blood Purif 1994;12:292–307. [DOI] [PubMed] [Google Scholar]
  • 38. Boyd RF, Langsdorf LJ, Zydney AL. A two layer model for the effects of blood contact on membrane transport in artificial organs. ASAIO J 1994;40:M864–9. [DOI] [PubMed] [Google Scholar]
  • 39. Fournier A, Birmelé B, François M, Prat L, Halimi JM. Factors associated with albumin loss in post‐dilution hemodiafiltration and nutritional consequences. Int J Artif Organs 2015;38:76–82. [DOI] [PubMed] [Google Scholar]
  • 40. Nicoud P, Delobel S, Gavard M, Gmerek A, Wagner J, Dialytic performance of a new high flux membrane in on line HDF: Experience of a French dialysis center. Proceedings of the 47th European Dialysis and Transplant Association, and European Renal Association (EDTA‐ERA) Conference Munich, Germany. June 25–28, 2010.
  • 41. Le Roy F, Hanoy M, Claeyssens S, Bertrand D, Freguin C, Godin M. Beta2microglobulin removal and albumin losses in post‐dilution hemodiafiltration: membrane effect. Nephrol Dial Transplant Plus 2009;2:Issue suppl_2,1150. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Additional supporting information may be found online in the Supporting Information section at the end of the article.

FIG. S1. Setup for measuring sieving coefficients.

TABLE S1. Analyses of A) β2M reduction ratios and B) albumin loss vs. sieving coefficient values in HD and HDF treatment modes (sieving coefficient values reported in datasheets and IFUs vs. identical testing conditions).


Articles from Artificial Organs are provided here courtesy of Wiley

RESOURCES