. 1975 Nov-Dec;79A(6):713–725. doi: 10.6028/jres.079A.028

Absolute Isotopic Abundance Ratios and the Atomic Weight of a Reference Sample of Potassium

E L Garner 1, T J Murphy 1, J W Gramlich 1, P J Paulsen 1, I L Barnes 1
PMCID: PMC6589417  PMID: 32184525

Abstract

Solid sample, thermal ionization, mass spectrometry has been used to obtain absolute values for the isotopic abundance ratios of a reference sample of potassium. Standards of known isotopic composition, prepared by gravimetrically mixing nearly isotopically and chemically pure separated isotopes of, 39K and 41K, were used for calibration. The absolute isotopic abundance ratios are : 39K/41K = 13.8566 ± 0.0063 and 40K/41K = 0.0017343 ± 0.0000061 which yield atom percent compositions of 39K=93.2581 ± 0.0029. 40K = 0.01167 ± 0.00004, and 41K = 6.7302 ± 0.0029. The calculated atomic weight for potassium is 39.098304 ± 0.000058. The indicated uncertainties are overall limits of error which are the sum of the uncertainty components for ratio determinations and the components covering the effects of known sources of possible systematic error.

Keywords: Absolute ratios, atomic weight, isotopic abundance, potassium, reference standard

1. Introduction

The Analytical Spectrometry Section of the National Bureau of Standards is conducting a long term program of absolute isotopic abundance ratio and atomic weight determinations using the mass spectrometric method. Previous atomic weight determinations include silver [1],1 chlorine [2], copper [3], bromine [4], chromium [5], magnesium [6], lead [7], boron [8], rubidium [9], rhenium [10], and silicon [11]. The present work extends the study to potassium.

A mass spectrometric determination of the atomic weight of an element requires a determination of the absolute isotopic abundance of a reference standard and also the establishment of limits for natural variations. The essential criteria for obtaining absolute isotopic abundances from relative or observed isotopic ratios is calibration of the instrument for the effects of bias using calibration standards of the analyte element with accurately known isotopic ratios. Accurately known calibration standards are prepared by gravimetrically mixing nearly isotopically and chemically pure separated isotopes of the analyte element. The uncertainty in the absolute ratio of the reference standard due to the assay and mixing of the separated isotopes must be within ±0.02 percent to obtain the desired high accuracy atomic weight results. Mass spectrometric analysis of the reference standard and the calibration standard under nearly identical analytical conditions provides the means of calculating correction factors (calculated isotopic ratio of calibration standard/observed isotopic ratio of calibration standard) which when applied to the observed ratios of the reference standard being calibrated yield absolute isotopic abundance ratios for the standard. The atomic weight of the reference standard can then be calculated from the absolute abundances and the nuclidic masses reported by Wapstra and Gove [12]. Establishing the limits of natural variability requires a sampling of minerals from different geographic and geological origins on a world-wide basis. The uncertainty components due to possible natural variations, the ratio determination and other known sources of possible systematic error are summed to obtain an overall limit of error for the atomic weight of the element. A more detailed discussion of the mass spectrometric determination of the atomic weights of the elements is given by Cameron [13].

Naturally occurring potassium consists of stable isotopes 39K and 41K, and a radioactive isotope 40K. With a half-life of approximately 1.4 × 109 years, 40K is changing by beta emission to 40Ca and by orbital electron capture to 40Ar. Since the stable isotopes are 99.988 percent of the total number of potassium atoms present, the effect of the total amount of 40K on the atomic weight is only 2.6 parts-per-million. During the course of a 109 year half-life the decay of 40K will decrease the atomic weight by 1.5 parts-per-million (ppm).

Prior to 1951 the International Commission on Atomic Weights used data from the chemical combining weight method as the source for establishing the potassium atomic weight. All of the mass spectro-metric isotopic abundance data available at that time were relative measurements, which could not be corrected for bias effects because of insufficient information. Secondly, there were conflicting reports which indicated both large natural variations [14, 15, 16, 17, 18, 19, 20, 21], and a high degree of constancy [22, 23] in the isotopic abundances of potassium. These two factors weighed heavily against the use of mass spectrometric data as a source of information for the atomic weight of potassium.

In what was considered a benchmark in analytical mass spectrometry at the time of the measurement in 1950, Nier [24] calibrated for the effects of bias with standards of known isotopic composition prepared from high purity separated isotopes of argon. Reuterswärd [25], in an extensive effort to identify and correct for bias in the mass spectrometric measurement, discussed the risk of accepting a calibration which did not use potassium separated isotopes. Although extensive in scope, the calibration of Reuterswärd did not include calibration with standards of known isotopic composition. The atomic weight calculated from his abundance data was 39.101 and did not provoke any serious misgivings about the 1951 report of the International Commission on Atomic Weights [26] which accepted the 1950 isotopic abundance measurements of Nier as the basis for calculating the atomic weight of potassium.

Since 1951, the values recommended for the atomic weight of potassium by the International Commission on Atomic Weights have fluctuated between 39.098, obtained from the chemical method, and 39.102, obtained from the mass spectrometric method. Much of the indecision was due to an apparent difference of 0.01 percent between the methods and the lack of new sources of data or a high accuracy measurement to resolve the discrepancy. Tables 1 and 2 give a detailed summary of atomic weight (12C = 12) data by these methods. The mass spectrometric data is not inclusive of all published isotopic data but is representative of the measurements published since 1930. Where investigators reported natural variations, the 39K/41K value for the standard or mineral most representative of terrestrial potassium was used for the atomic weight calculation. Examining the data from a point of view that none of the measurements satisfy the criteria for an absolute abundance determination, the average and range obtained is similar to the chemical data. The new source of information which eventually produced the revision to 39.098 ± 0.003 as the accepted value by the International Commission on Atomic Weights was Marinenko’s [36] discussion of the data of Bates and Wichers [37], obtained from potentiometric titration of acids.

Table 1.

Recalculation of atomic weight of potassium (12C = 12)

Year Investigator Method Atomic weight
1906 Richards and Staehler [27] KClAg=0.691072 39.092
KClAgCl=0.0520118 39.091
1907 Richards and Müeller [28] KClAg=0.69107 39.091
KClAgCl=0.52012 39.091
1928 Hönigschmid and Gaubeau [29] KBrAg=1.103270 39.104
KBrAgCl=0.63374 39.101
1929 Hönigschmid and Gaubeau [30] KClAg=0.691148 39.100
KClAgCl=0.520180 39.100
1933 Baxter and MacNevin[31] KClAg=0.691057 39.090
1933 Hönigschmid and Sachtleben [32] KClAg=0.69069 39.091
KClAgCl=0.520132 39.093
KBrAg=1.103197 39.096
KBrAgBr=0.633720 39.091
1935 Johnson [33] KClAg=0.691108 39.095
1940 Baxter and Harrington [34] KClAg=0.691085 39.093
1941 McAlpine and Bird [35] KBrO3KBr=1.403320 39.104
KBrAg=1.103193 39.095
1957 Marinenko [36] (Recalculated data of Bates and Wickers [37]) C8H5O4KC7H6O2=1.673374 39.099
Average (overall) = 39.095
Average (since 1935) = 39.097

Table 2.

Calculation of atomic weight of potassium from relative mass spectrometric measurements

Year Investigator 39K/41K Atomic weighta (12C = 12)
1934 Brewer and Kueck [38] 13.88 ± 0.4 39.098
1935 Nier [39] 13.96 ± 0.1 39.097
1936 Manley [40] 13.4 ± 0.5 39.102
1936 Bondy and Vanicek [41] 14.1 ± 0.1 39.096
1943 Cook [22] 14.12 ± 0.28 39.096
1944 Paul[42] 13.96 ± 0.6 39.097
1948 White and Cameron [43] 13.66 ± 0.1 39.100
1948 Hibbs and Redmond [44] 13.81 ± 0.15 39.099
1950 Nier [24] 13.48 ± 0.07 39.102
1951 Reuterswärd [45] 14.32 ± 0.04 39.094
1956 Reuterswärd [25] 13.57 ± 0.09 39.101
1956 White et al. [46] 13.79 ± 0.05 39.099
1958 Omura and Morito [47] 13.96 ± 0.05 39.097
1960 Kendall [23] 13.769 ± 0.029 39.099
1962 Harms et al. [15] 14.33 ± 0.06 39.094
1962 Letolle [16] 13.85 ± 0.06 39.098
1962 Stauffer and Honda [48] 14.23 ± 0.01 39.095
1963 Shukolyukov et al. [49] 13.96 ± 0.09 39.097
1965 Schreiner and Verbeek [50] 14.19 ± 0.01 39.095
1973 Barnes et al. [51] 14.015 ± 0.014 39.097
Average 39.098
a

Calculated with a 40K/41K value of 0.0017343 for all investigators.

The primary purpose of the research reported here was a high accuracy determination of the atomic weight of potassium which would resolve the apparent discrepancy between the chemical and mass spectro-metric data. To accomplish this goal, six calibration standards of known isotopic composition were prepared by gravimetrically blending isotopically and chemically pure separated isotopes of 39K and 41K. A secondary purpose was to provide an isotopic standard of accurately known composition which can serve as a stabilizer in geochemistry where the search for potassium variations has produced a collection of diverse and conflicting relative isotopic measurements. The material chosen for the isotopic standard was a potassium chloride (Standard Reference Material 985) obtained from the J. T. Baker Chemical Company2, Phillipsburg, New Jersey.

2. Experimental Procedure

2.1. Mass Spectrometry

Isotopic abundance ratios of the reference standard (Standard Reference Material 985), calibration mixes and separated isotopes were determined by solid sample, thermal ionization mass spectrometry. Two single stage, solid sample mass spectrometers (a 15 cm radius of curvature, 60° analyzer tube instrument and a 30 cm radius of curvature, 90° analyzer tube instrument) were used to obtain two independent sets of isotopic ratio measurements. Both mass spectrometers were equipped with interchangeable and nearly identical electronic components, identical thin lens “Z” focusing ion sources [52], and the same basic design of a multielement, deep-bucket faraday cage collector [52, 53, 54].

The collectors, especially designed for in-depth suppression of secondary electrons, were equipped with a tungsten ribbon (0.025 ×0.76 mm) transmission grid shadowing a series of 0.10 mm suppressor grids. The other components for each measuring circuit were a pair of vibrating reed electrometers, an expanded-scale strip-chart recorder, a voltage to frequency converter and a scaler-timer. Data acquisition was computer controlled with the 15 cm instrument interfaced to a central, time-shared computer and the 30 cm instrument interfaced to its own minicomputer. Under computer controlled operation, all phases of the ratio measurement and data reduction were programed and automatic with the exception of ranging of the vibrating reed electrometers. This was accomplished manually by programing a magnet switching delay at the end of each peak top observation which was of sufficient duration to permit manual ranging. Operators retained at all times the options for digital, expanded scale recorder or simultaneous digital-recorder measurements.

Absolute isotopic abundance studies of uranium [54] established that the basic expanded-scale measuring circuit, including the multielement faraday cage collector, is linear over a ratio range 20:1 to 1:20. It was also established that for this range of isotopic ratios, a single mass dependent correction factor could be experimentally determined and applied to correct for bias effects. Corroborative support for this is provided by previous work in the atomic weight program where the isotopic ratios of a wide range of synthetic calibration mixes of the elements boron [8], chromium [5], magnesium [6], rubidium [9], and rhenium [10], were measured and compared to known values. Since most of the absolute isotopic abundance data obtained at NBS was with the expanded scale recorder, a considerable effort was made to prove the linearity of the digital measuring circuit. This was accomplished by a comparison of the digital and expanded scale recorder data obtained under simultaneous measurement conditions. Linearity of the digital circuits used in this experiment were found to be ≤ 0.02 percent.

Potassium was thermally ionized from a tantalum triple filament (0.025 × 0.76 mm) source. This filament material was selected in preference to other materials because of its poor ionization efficiency for calcium and the relative ease of reducing the potassium background ion current of a tantalum filament to 1 × 10−15A. Prior to filament fabrication, the tantalum ribbons were boiled in ultra-high-purity water to remove all water soluble surface contaminants, rinsed in a second wash of ultra-high-purity water, and dried with a heat lamp in a clean air environment. After fabrication, the filaments were degassed in a vacuum and under a potential field. All filaments were degassed for 1 h at 4 Å and then at an increased current of 41/2 Å for one-half hour. After minimum cooling of 30 min, the final cleaning was accomplished by a flashing procedure. This procedure consisted of a series of l-min cycles in which the filaments were alternately pulsed with a 4 Å current and then cooled for periods of 5 to 15 s. Bare filaments cleaned in this manner exhibited a potassium background ion current of ≤ 1 × 10−15 A.

An evaluation of borosilicate glass, polyethylene, quartz, and Teflon containers over a 3-month period resulted in the selection of Teflon FEP3 as the most acceptable for long-term storage of potassium solutions. Polyethylene containers were the least acceptable because of the relatively large amounts of material leached from some of the containers after stringent cleaning. Microgram amounts of organic or inorganic impurities on the sample filaments with 10–15 μg of potassium, changed the isotopic fractionation pattern of a high purity potassium chloride standard from a significantly decreasing 39K/41K ratio to a nearly constant or increasing ratio with time [55]. Thus, a “dirty” standard, independent of the source of the impurities, showed a bias when compared to the same high purity standard analyzed under identical conditions. The deleterious effects of impurities on the accuracy of potassium isotopic measurements is well documented [22, 23, 25, 50, 55, 56]. The Teflon FEP containers were cleaned for a prolonged period of 1 week in hot, ultra-high purity (1 + 1) nitric acid [57], rinsed with ultra-high purity water, leached for 1 week in hot, ultra-high purity (1+1) hydrochloric acid, rinsed with ultra-high purity water and dried in a Class 100 clean air environment. The containers were then filled with ultra-high purity (1 + 1) hydrochloric acid, capped and stored in a clean air environment until needed. Prior to adding potassium solutions, the containers were thoroughly rinsed with ultra-high purity water.

Quartz tubing was used to make the pipets which were used to transfer solution from sample storage containers to the filaments. The quartz tubing was cleaned by heating in nitric acid (1 + 1) for 48 h, rinsed several times in warm high purity water, boiled in ultra-high purity nitric acid (1 +1), rinsed in warm ultra-high purity water, and dried in a clean air environment by heat lamps. Immediately before transferring solution for analysis, additional cleaning was accomplished by flushing with high purity hydrochloric acid (1 +49), rinsing with ultra-high purity water, and conditioning with an expendable aliquot of the sample to be analyzed. The effects of environmental contamination on cleaned filaments and pipets were minimized by careful handling and storage in a Class 100 clean air environment.

Potassium was deposited on the sample filaments as dilute hydrochloric acid solutions (1 +49) of potassium chloride at a concentration of 2.5 mg of potassium per ml. One drop, 5 μl for MS #1 and 10 μl for MS #5, were evaporated to dryness on each sample filament with heat lamps and electrical currents which were adjusted as follows: 1.0 Å for 10 min: 1.2 Å for 5 min: and 1.5 Å for 5 min. The heat lamps and electrical currents were turned off and the filaments allowed to cool 5 min before removal from the clean air environment. Sample filament alinement and spacing criteria required positioning these filaments the maximum distance of approximately 0.5 mm from the edges of the ionizing filament. This type of open spacing was used to reduce the effects of radiant heat by the ionizing filament.

Potassium ions were measured at a constant accelerating voltage of 3.8 kV by varying the magnetic field via a temperature compensated. Hall effect controller. For the mass spectrometric analysis, the ionizing filament was maintained at 1250 °C (direct pyrometer reading) and the sample filament currents were initially adjusted to 0.6 A;. At specified times of 5 and 10 min into the analysis, the K+ ion current was adjusted to 5 × 10−11 A and 1.0 × 10−10 A, respectively. No further adjustments were made until the ion current exhibited a growth pattern. Failure to produce an ion current growth pattern within 25 min after starting the run was sufficient cause for termination and rejection of the analysis. Under ion current growth conditions, the sample filament currents were adjusted to yield intensities of 1.5 × 10−10 A and 2.0 × 10−10 A at specified times of 15 and 20 min into the analysis. No further adjustments were normally required. The ratio measurement which was started after 40 min of heating was made in the sequence of 39K/41K, 41K/40K, 39k/41k. While the two operators did not select or use the same parameters, such as sample size, sample mounting, heating pattern, and signal intensity, rigid standardization was maintained on each mass spectrometer. Isotopic fractionation was observed and, under normal conditions, the 39k/41k decreased ≤ 0.2 percent during the 45 min ratio measuring period. The effects of secondary electrons were observed in the vicinity of the 40K base and were due primarily to the 10−10 A 39K ion current. Reducing the x-axis opening into the collector with a baffle removed the large 39k ion beam from the collector plates when the 40K ion beam was focused on the faraday cage. This procedure was successfully used on both instruments to minimize the effects of secondary electrons. However, as collector plates were coated with a thin layer of potassium, suppression of secondary electrons became uncontrollable. It was necessary to remove the collector, thoroughly clean all surfaces and replace the tungsten transmission grids with virgin ribbon. This cleaning procedure was necessary and essential after approximately 50 analyses. Efforts to clean the tungsten transmission grids were unsuccessful and, under such conditions, deterioration of secondary electron suppression was rapid and unacceptable within 20 analyses. Contributions from the 39K tail at the 40K mass position were not detectable on the 30 cm instrument but were ≤ 1 × 10−15 A on the 15 cm instrument. This tailing effect limited the accuracy of 40K/41K measurements with the digital system.

Although memory was not detected when the 39k and 41k separated isotopes were analyzed in an alternate pattern on the same source, or when analyzed in a similar pattern with natural or isotope dilution samples, it was known that significant amounts of potassium were deposited on the ion source during an analysis. Since K ions contributed from residual potassium can be a source of significant error, the ion source was removed from the mass spectrometer, disassembled, and thoroughly cleaned before analysis of each separated isotope, the equal atom calibration mix, and the reference standard and remaining calibration mixes. All tools and associated sample mounting apparatus were also cleaned with each source cleaning. The clean ion source was first degassed for at least 1 h with a bare set of filaments and then conditioned for 90 min with a sample of the approximately isotopic abundance to be analyzed.

2.2. Purification of the Separated Isotopes

Electromagnetically separated 39K and 41K isotopes in the form of potassium chloride were obtained from the Isotopes Division. Oak Ridge National Laboratory of the Union Carbide Nuclear Company. The 39KC1 and the 41KC1 were designated series 107401 and 149401, respectively. The certificates of analysis which accompanied each sample included a semiquantitative spectrographic analysis which showed that zinc could be present at the 0.2 percent level and several other elements, including aluminum, cadmium, cobalt, chromium, and nickel could be present up to the 0.5 percent level. To reduce these impurities to a level low enough so that they could not cause a significant error in the determination of the potassium, the separated isotope samples were further purified. The purification method used which was based on the crystallization of potassium perchlorate had been previously found by this laboratory to be effective for removing cationic and anionic impurities from rubidium chloride with the exception of potassium and cesium which cocrystallized with the rubidium [9]. The effectiveness of this purification procedure for potassium, in which potassium perchlorate is crystallized, was first tested by the purification of natural potassium chloride to which 1,000 ppm of 21 common elemental impurities had been added. The results of the analysis of the purified material by spark source — isotope dilution analysis is shown in table 3. Only sodium and iron were detected at the 10-ppm level and most of the other elements were at the low ppm level. The purification method does not separate rubidium or cesium.

Table 3.

Spark source-isotope dilution mass spectrographic analysis of purified natural potassium chloride

Element Concentration
μg/g
Element Concentration
μg/g
Al 0.6 Mo 5
Ba 1 Na 10
Bi 0.3 Ni 4
Ca 1 Pb 1
Cd 2 Se 6
Co 3 Sn 4
Cr 2 Sr 0.1
Cu 2 Te 2
Fe 11 Tl 0.4
Mg 1 Zn 0.4
Mn 1

Each separated isotope sample was purified as follows: The potassium chloride (about 3.5 g) was dissolved in 50 ml of 0.02N HC1 in a Teflon-FEP beaker. The solution was filtered through a close textured filter paper that had been previously washed with 200 ml of 5N HC1 into a Teflon-FEP beaker. The filtrate was diluted to 100 ml, and 6 ml of high-purity perchloric acid were added. The precipitated KCl04 was dissolved by covering the beaker with a Teflon cover and heating the solution to near boiling. Potassium perchlorate was crystallized by removing the cover and allowing the solution to concentrate to about 35 ml. The solution was allowed to cool to room temperature and then to about 5 °C by placing it in the refrigerator overnight. The crystallized potassium perchlorate was recovered from the solution by filtering it through a fine porosity polyethylene filter. After the KCIO4 was washed with a cold 95 percent ethanol solution, it was returned to the Teflon beaker and the crystallization process was repeated except that only 2 ml of perchloric acid were added. The recrystallized KCIO4 was again caught on the polyethylene filter and washed with cold 95 percent ethanol. The purified material was dried by passing clean air over the KClO4.

The purified KCIO4 was transferred back to the Teflon beaker and dissolved in 150 ml of hot water. This solution, while still hot, was passed through a cation exchange column containing 50 ml of AG50X8 resin, 100–200 mesh which is a strongly acidic cation exchange resin. (The polyethylene and resin were previously cleaned with 800 ml of 5NHCl.)The column was washed with 100 ml of water to completely remove perchloric acid from the column and the potassium was eluted with 200 ml of 5N HC1. The eluate containing the potassium was caught in the original Teflon beaker and evaporated to dryness.

The residue of potassium chloride was transferred to a 15 cm× 1 cm × l cm platinum boat. The boat was transferred to a quartz tube furnace and the boat and contents were ignited under oxygen at 550 °C for 2 h to oxidize the organic residue from the column. (To avoid any possibility of cross isotopic contamination, the ignitions were performed in new separate quartz tubes which had never been exposed to potassium.)

Calculations based on the starting weights of 39KCl and 41KC1 and the weights of the corresponding purified salts showed about 95 percent of the potassium was recovered in each case.

The acids and water used during these purifications were produced at NBS by subboiling distillation [57] and have been shown to be extremely low in trace cation contamination. Apparatus such as beakers, columns, and filters were vigorously cleaned with high-purity acid.

2.3. Preparation and Analysis of the Separated Isotope Solutions

The 3.7 g of purified 39KC1 were dissolved in about 50 ml of (1 + 49) HC1, transferred to a tared 250 ml Teflon-FEP screw cap bottle and diluted to about 190 ml with (1 +49) HC1. The 3.3 g of purified 41KC1 were dissolved in about 50 ml of (1+49) HC1, transferred to a tared 500 ml Teflon-FEP screw cap bottle and diluted to about 350 ml with (1 +49) HC1. The bottles and contents were weighed to ±0.2 mg and the preliminary potassium concentration were calculated. The solution of 39KC1 was designated “K39” and the solution of 41KC1 was designated “K41”.

Samples of the “K39” and “K41” solutions were analyzed for impurity elements by isotope dilution-spark source mass spectrometry [57]. Samples equivalent to about 100 mg of K were spiked with 10−7 g of 109Ag, 137Ba, 44Ca, 111Cd, 53Cr, 65Cu, 54Fe, 113In, 26Mg, 62Ni, 206Pb, 82Se, 117Sn, 86Sr, 125Te, 203Tl, and 67Zn. Most of the potassium matrix was removed by crystallization of potassium perchlorate and concentration of the solution to about 0.2 ml. The liquid was removed from the crystallized KClO4 by decantations and concentrated to about 0.05 ml by evaporation. Each solution was then evaporated to dryness on high purity gold wires and analyzed by spark-source mass spectrometry. In addition to these elements, A1, Bi, Co, Mo, and Mn were estimated by comparison to other nuclides. In addition, Rb and Cs, which are not removed by the purification procedure were determined by thermal ionization mass spectrometry by spiking with 87Rb, measuring the 85Rb/87Rb ratio and comparing the 132Cs to87Rb. The results of the analysis for impurities are shown in table 4. No element was detected at a level high enough to cause a significant error in the assay procedure, that is greater than 100 μg/g.

Table 4.

Impurities in potassium separated isotopes

Element “K39”
μg/ga
“K41”
μg/ga
Alb 1.5 0.5
Ba 0.6 .09
Ca .4 .3
Cd .3 .09
Cr .4 .5
Cs .1 .1
Cu .3 .1
Fe .8 1.2
Mg .2 0.2
Nab 1 1.4
Ni 0.1 0.06
Pb .09 .03
Rb .3 .4
Sn 1 .01
Sr 0.04 .01
Te .06 .05
Tl .1 .05
Zn .5 .09
a

Essentially all of the values represent upper limit values.

b

Compared to 26Mg.

2.4. Assay of Separated Isotope Solutions

The bottles containing the potassium separated isotope solutions were shaken to thoroughly mix the solution and the screw cap closures were replaced with prepunctured polyethylene stoppers. Four weighed portions of about 15 g for the “K39” solution and about 30 g for the “K41” solution, equivalent to about 4 mmol of K, were withdrawn from each solution in the following manner. A 16-cm platinum needle was inserted through the prepunctured polyethylene stopper. A 20-ml polyethylene hypodermic syringe was attached to the Kel-F hub of the needle and the desired amount of solution was withdrawn. The syringe was then disconnected from the hub and the tip was capped with a Kel-F hub. Any static charge that might be present on the plastic syringe was dissipated by wiping it with a damp lintless towel and the syringe and contents were weighed on a semimicrobalance to ±0.02 mg. The solution was then delivered into a Teflon-FEP beaker and the syringe was again capped, wiped and weighed. The weight of the sample was determined from the weights of the syringe before and after delivery of the sample. Since about 30 ml of the “K41” solution were required to produce a 4-mmol sample, two loadings of the 20-ml syringe were weighed for each assay sample. Two assay samples were withdrawn from each solution before the calibration samples were withdrawn and two samples were withdrawn after the calibration samples to insure that no change in concentration occurred during this time interval (about 6 h).

Each sample was then assayed as follows: Two ml of (1 + 1) HClO4 was added to the weighed portion of the separated isotope solution. The solution was heated to dissolve the precipitated potassium perchlorate,KClO4. and evaporated to constant volume at about 85 °C. The temperature of the hot plate was then raised to about 110 °C to volatilize the excess perchloric acid. The beaker and contents were cooled to room temperature, 15 ml of water were added and the solution was heated until the crystallized KClO4 dissolved. The solution was again evaporated and any excess perchloric acid was removed by heating at about 110 °C. These evaporations freed the KClO4 from chloride and occluded perchloric acid.

The recrystallized potassium perchlorate was again dissolved in 15 ml of water by heating and the solution was evaporated until only about 1 ml of solution remained with the crystallized KC104. (This was estimated by weighing the beaker and contents.) After cooling to room temperature, 4 ml of perchloric acid were added and the beaker and contents were cooled to about 5 °C in a refrigerator for 16 h (overnight).

The crystallized potassium perchlorate was transferred with cold (5 °C) 95 percent ethanol to a tared 15 ml fine porosity borosilicate glass filtering crucible. As much of the salt as possible was transferred by scraping the sides of the beaker with a Teflon “policeman” and washing the salt into the crucible with cold ethanol. The filtrate and original beaker were reserved for the determination of dissolved and untransferred potassium.

The filtering crucible and contents were dried at 110 °C for 2 h and at 300 °C for 4 h, cooled in a desiccator, and then transferred to the case of a micro-balance and allowed to stand for several hours. The crucible and contents were then weighed to ±0.002 mg. A buoyancy correction for the glass crucible was made by averaging three empty tare crucibles. The air weight of the KClO4 was then determined and converted to vacuum weight using 2.522 for the density of 39 KClO4 and 2.557 for the density of 41KClO4. These densities were calculated by assuming that they are proportional to the density of natural KClO4 in the same relationship as their molecular weights. For simplicity of calculations, the vacuum weight of the potassium perchlorate was converted to millimoles of potassium using the calculated atomic weight for potassium and 1973 atomic weight values for chlorine and oxygens. The formula weights used were 138.4148 for 39KClO4 and 140.3959 for 41KC1O4.

The beaker in which the precipitation was performed was thoroughly washed with hot water and the washings were combined with the filtrate. The “K39” solutions were spiked with about 15 μmol of 41K and the “K41” solutions were spiked with about 15 μmol of 39K for determining soluble and untransferred potassium by isotope dilution mass spectrometry. The spiked solution was mixed, 1 ml of nitric acid was added and the solution was evaporated to constant volume at about 85 °C. A few ml of water was added and the solution was evaporated to dryness at a higher temperature. The residue was taken up in 5 ml of water and passed through a column containing 3 ml of AG50X8 cation exchange resin. After washing the column with 20 ml of H2O, the potassium was eluted with 20 ml of 5N HC1. The eluate was evaporated to dryness and taken up in 0.3 ml of (1+49) HC1 and the 39K/41K ratio was determined by thermal ionization mass spectrometry. The potassium found by isotope dilution was added to the potassium from the gravimetric determination to yield the total potassium in the sample. The concentration was then calculated by dividing this value by the weight of the sample. Table 5 shows the results of these determinations.

Table 5.

Assay—potassium separated isotopes solutions

Solution Sample Weight
KClO4
g
KClO4
K
mmol
Filtrate
K
mmol
Total
K
mmol
Weight
Sample
g
Conc.
Sample
mmol
1 0.517358 3.737737 0.007600 3.745337 14.52021 0.257939
2 .543999 3.930210 .005786 3.935996 15.25967 .257934
“K39” 3 .531849 3.842430 .004935 3.847365 14.91381 .257973
4 .522707 3.776382 .003977 3.780359 14.65616 .257936
Average = .257946
s.d. = .000018
1 0.514726 3.666247 0.006539 3.672786 30.03932 0.122266
2 .492306 3.506555 .005527 3.512082 28.72888 .122249
“K41” 3 .487376 3.471440 .006114 3.477554 28.44437 .122258
4 .498044 3.547426 .005187 3.552613 29.05879 .122256
Average = .122257
s.d. = .000008

This method of determining the concentration of potassium solutions was previously tested on solutions containing known amounts of potassium. Six solutions were prepared from SRM 999 which is certified to contain 52.435±0.004 percent potassium by weight in the approximate concentration of the separated isotope solution. Four samples containing from 3.5 to 5.0 mmol of potassium were withdrawn from each solution and the potassium concentrations were determined as described.

Comparison of the calculated and measured concentrations of the six solutions showed that: (a) the concentration of potassium determined by this method agreed to within 0.01 percent of the calculated potassium concentration, (b) systematic errors were negligible, and (c) the analyses were of equal precision.

From these analyses and the analyses of the separated isotope solutions, the standard deviation of an individual measurement of the concentration of a potassium solution was computed to be0.000017 mmol/g (24 degrees of freedom). The uncertainty of the concentration at the 95 percent confidence level is computed to be 0.000018 mmol/g for the “K39” solution and 0.000008 mmol/g for the “K41” solution.

2.5. Isotopic Analysis of Separated Isotopes

Two aliquots of each separated isotope were utilized for mass spectrometric analysis. One aliquot was withdrawn from each separated isotope solution before blending of the first calibration mix and the other immediately after finishing the last calibration mix. Each of the premixing separated isotope solutions was analyzed eight times and each of the postmixing solutions was analyzed two times for a total of 10 analyses per instrument. A comparison of the observed isotopic ratios of these solutions revealed no significant contamination during preparation of the calibration mixes.

Exclusive of variable isotopic fractionation between analyses, the major sources of error in determining the isotopic composition of highly enriched separated isotopes are memory, contamination, filament background and nonlinearity of the measuring circuit. Memory effects were minimized by thorough cleaning and conditioning of the ion source for analysis of each separated isotope. The 39K background contribution of the bare degassed filaments was confirmed to be 1 × 10−15 A by mass spectrometric analysis. A selected number of these filaments were removed from the mass spectrometer and a known quantity, either 10 μg or 25 μg, of potassium separated isotope was immediately deposited on each sample filament. When analyzed, the effect of natural potassium was much larger than could be accounted for by a nominal 1 × 10−15 A filament background. This “threshold effect,” where a small amount of sample on the filament surface is necessary to obtain an accurate measure of filament blank, was observed when the ionizing filament was used for a single analysis or when used for as many as four different analyses with four different sets of sample filaments. The effect of filament blank on the 39K/41K ratios of the separated isotopes was approximately 0.002 percent and 0.05 percent for 39K and 41K, respectively.

Bias due to RC response of the measuring circuit was minimized by delaying a minimum of 30 s to allow the system to stabilize after magnetically switching peak tops. Continuous monitoring of 39K on the faraday cage produced a condition in which the 39K/41K ratio decreased more than could be accounted for by the isotopic fractionation associated with the heating pattern. Removing the 39K ion current from the collector for less than 5 min produced a recovery to the initially observed 39K/41K ratio. This change was in the same direction and, in most cases, of the same magnitude as that obtained from isotopic fractionation. The maximum change observed, including isotopic fractionation, for a normal analysis in which 39K was continuously focused on the faraday cage was 0.5 percent. A most likely cause of this condition, secondary electrons, was investigated with negative results. The number and position of collector suppressor magnets, increased suppressor voltage on collector plates and two different thoroughly cleaned collectors on two different mass spectrometers did not have any significant effect. A nonlinear response of a component of the measuring circuit because of continuous and prolonged exposure to 10−10 A ion currents was also a likely cause but could not be proven. Although this effect was not detected for any other sample, precautions were taken to avoid prolonged or continuous monitoring of 10−10 A ion currents, to limit ratio measurements to less than 10 min, and to defocus the ions from the collector for 5 min between each measurement cycle within an analysis.

Corrected isotopic compositions of the separated isotopes are reported in table 6. The uncertainties are limits of error which are larger than the calculated 95 percent confidence limits and which include additional components for nonlinearity and the effects of a natural potassium filament blank.

Table 6.

Isotopic composition of separated potassium isotopes

Separated isotope Isotopic composition (atom percent)
“K39” 39K 99.9765 ± 0.0004
40K 0 0014 ± 0001
41K .0221 ± .0004
“K41” 39K 0.8186 ± 0.0020
40K .0004 ± .0001
41K 99.1810 ± .0020

The uncertainties are based on minimum errors for the ratio determinations. The calculated 95 percent confidence limits are well below the minimum error statement.

2.6. Preparation of the Calibration Samples

Six calibration samples were prepared by mixing weighed portions of the “K39” and “K41” solutions to produce five calibration samples with 39K/41K ratios that approximated natural potassium and varied over a range of 4 percent. A sixth calibration sample was prepared with a 39K/41K ratio of 1. The portions were withdrawn from the bottles and weighed in the manner previously described. To eliminate any possibility of change in concentration of the isotope solutions with time, the portions for the calibration samples were withdrawn from the bottles at the same time the samples for determining the potassium concentrations were withdrawn.

Table 7 shows the calculated isotopic composition of the calibration samples. The isotopic ratio of each calibration sample was calculated from the isotopic analysis of the separated isotopes and the mmol of potassium from each separated isotope solution.

Table 7.

Composition of potassium calibration samples

Sample
No.
Isotope
solution
Weight
Solution
g
39K
mmol
41K
mmol
39K/41K
ratio
1 “K39”
“K41”
6.74034
1.01137
1.738235
0.001012
0.0003842
.1226344
14.13808
2 “K39”
“K41”
7.49182
1.14244
1.932031
0.001143
.0004271
.1385274
13.91228
3 “K39”
“K41”
7.07292
1.07620
1.824002
0.001077
.0004032
.1304954
13.94269
4 “K39”
“K41”
7.19697
1.11621
1.855994
0.001117
.0004103
.1353469
13.67965
5 “K39”
“K41”
7.25150
1.13500
1.870055
0.001136
.0004134
.1376252
13.55556
6 “K39”
“K41”
2.45747
5.32666
.6337456
.0053309
.0001401
.6458880
0.989239

Each calibration sample was thoroughly mixed by stirring and evaporated to dryness on a hot plate. The calibration samples were taken up with (1 + 49) HCI so that 1 ml of solution contained 2.5 mg of K.

2.7. Isotopic Analysis of the Calibration Mixes and the Reference Standard

Two complete and independent sets of analyses of the calibration mixes and the reference standard were made, one on MS #1 by Operator I and the other on MS #5 by Operator II. Each set consisted of 18 analyses of the calibration mixes and 18 analyses of the standard. Exclusive of the equal atom calibration mix which was analyzed on a clean source, a simple alternating pattern of standard and calibration mix was maintained until three analyses per calibration mix were obtained. Four solutions of the high purity potassium chloride standard were prepared and stored in specially cleaned Teflon bottles. Three of these samples were representative of different positions of the bulk material and the fourth was a composite of the first three. The composite sample was processed through an ion exchange column to duplicate the treatment of the separated isotopes during the purification procedure.

Operator I utilized the digital system to determine the potassium isotopic ratios of all samples and also simultaneously recorded the data by expanded scale recorder. Of the strip chart recorder data by Operator I, only the 41K/40K ratio was calculated with the intent of providing a rigorous evaluation of the linearity of the digital system. Operator II utilized the digital and expanded scale recorder in a simultaneous mode of operation to measure the isotopic ratios of all samples. Two sets of measurement by Operator II, the set of digital measurements by Operator I, and the 41K/40K recorder measurements by Operator I were submitted to a statistician for analysis.

3. Results and Discussion

Table 8 summarizes the analysis of the calibration mixes and the calculation of correction factors. Calibration mixes one through five formed the point calibration for determining the correction factors. Analysis of equal atom mix (No. 6) yielded correction factors slightly different from the averages on both instruments. This calibration mix was excluded in calculating the average correction factors but was included in determining the uncertainty in the ratio measurement.

Table 8.

Determination of correction factors

Calibration
sample No.
39K/41K Ratio Correction factor a
Operator I
(digital)
Operator II
(digital)
Operator II
(recorder)
Operator I
(digital)
Operator II
(digital)
Operator II
(recorder)
1 14.2879 14.2723 14.2709 0.989590 0.990583 0.990684
2 14.0525 14.0432 14.0378 .990097 .990666 .991050
3 14.0871 14.0766 14.0749 .989825 .990475 .990598
4 13.8221 13.8088 13.8024 .989769 .990636 .991098
5 13.6940 13.6838 13.6797 .989966 .990617 .990917
6 0.999965 0.998973 0.998859 .989391 .990250 .990367
Mean value of correction factors = b 0.989849 b 0.990595 b 0.990869
a

These correction factors are calculated using the individually determined compositions of the separated isotopes to compute the isotopic ratio of each calibration sample.

b

Values of calibration sample No. 6 were not used in calculating the averages, but were included in the computation of the uncertainty in the ratio determination.

The average values for the observed isotopic ratios of four subsamples of the reference standard are given in table 9. The agreement between subsamples indicates that the Teflon sample containers were adequately cleaned or that any impurities leached from the containers did not affect the isotopic analysis. Organic contributions from the ion exchange column purification were also too small to significantly affect the isotopic ratios. The 41K/40K ratio as determined by Operator II on the digital system is biased high with respect to the recorder data on the same instrument. This bias is due to a slight amount of 39K tail at the 40K mass position on the 15 cm instrument. The digital system takes the linear average of two points on opposite sides of the 40K peak to determine the baseline, but the tail contribution is nonlinear. This type correction is more accurately determined from the recorder spectra. The precision of the 39K/41K ratio measurements are approximately the same for both operators using either digital or expanded scale recorder measurements. The two complete sets of measurements by Operator II do not satisfy the criteria for two independent determinations. Satisfaction of this criteria requires each set of analyses to be different in all respects from sample loading through the ratio measurement. Therefore, the digital measurements of Operator I and the recorder measurement of Operator II were used as two independent determinations from which the absolute isotopic ratios, the isotopic compositions, and the atomic weight were calculated. The absolute isotopic ratios are summarized in table 10.

Table 9.

Observed isotopic ratios of reference standard

Bottle No. Operator I Operator II
39K/41K Digital 39K/41K Recorder 39K/41K Digital 39K/41 K Recorder
B-2 13.99743 13.98520 13.98286
B-6 13.99798 13.98952 13.98477
B-11 13.99840 13.98805 13.98590
Composite 13.99826 13.98951 13.98653
Average 13.99802 13.98807 13.98502
41K/40K Digital 41K/40k Recorder 41k/40K Digital 41K/40K Recorder
B-2 573.363 574.204 575.301 573.987
B-6 573.307 573.289 576.027 574.090
B-11 573.921 573.653 575.135 573.571
Composite 573.852 573.851 575.648 574.453
Average 573.611 573.749 575.528 574.025

NOTE. — The value for each bottle is the average of five analyses.

Table 10.

Determination of corrected isotopic ratios

39K/41K 41K/40K
Observed
39K/41K
Correction
factor
Corrected
39K/41K
Observed
41K/40K
Correction
factor
Corrected
41K/40K
Digital I 13.99802 0.989849 13.85593 573.611 1.005128 576.552
s.d. = 0.00022 .000114 0.00141 (8 d.f.) 0.160 0.000051 0.163 (8 d.f.)
Digital II 13.98807 .990595 13.85651 575.528 1.004747 578.260
s.d. = 0.00102 .000070 0.00195 (8 d.f.) 0.198 0.000035 0.199 (8 d.f.)
Recorder II 13.98502 .990869 13.85732 574.025 1.004608 576.670
s.d. = 0.00081 .000114 0.00161 (8 d.f.) 0.182 0.000057 0.163 (8 d.f.)
Mean = 13.85662 ± a0.00627 Mean = 576.6 ± a2.0
Uncertainty Components:
 95 percent confidence limits on ratio determination ± 0.00267 ± 1.88
 Bounds due to possible systematic error in composition of separated isotopes ± 0.00119 ± 0.05
 Bounds due to possible systematic error in chemical analysis ± 0.00241 ± 1.10
a

Digital I and Recorder II were used as the two independent determinations to calculate the means and uncertainty component due to ratio determination.

The calculation of the atomic weight of the reference standard is summarized in table 11. Each uncertainty component is at least as large as the 95 percent confidence limit. For the 40K/41K ratio measurement of the standard and the isotopic ratios of the separated isotopes, the statistically calculated 95 percent confidence limits are smaller than the limiting error of the instrument. For this condition an uncertainty component larger than the calculated 95 percent confidence limit, but equivalent to the limiting error of instrument for the analytical conditions of the measurement, is used. The overall limit of error is the sum of the uncertainty components due to the isotopic composition of the separated isotopes, the chemical assay and mixing of the separated isotopes, and the mass spectrometric ratio measurement.

Table 11.

Summary calculations of the atomic weight of potassium in Standard Reference Material 985

Value Uncertainty Components
Overall limit of error a Mass spectrometric analytical error Possible systematic error in composition of separated isotopes Possible systematic error in chemical analysis
Atomic weight = 39.098304 ± 0.000058 ± 0.000025 ± 0.000011 ± 0.000022
Nuclidic masses (12C = 12)
39K = 38.9637089 ± 0.0000008
40K= 39.9640001 ± 0.0000008
41K = 40.9618270 ± 0.0000011
Atomic percent
39K = 93.25811 ± 0.00292 ± 0.00130 ± 0.00053 ± 0.00109
40K= 0.011672 ± 0.000041 ±0.000038 0.000001 0.000002
41K= 6.73022 ± 0.00292 ±0.00130 ± 0.00053 ± 0.00109
Isotopic ratios
39K/41K = 13.85662 ± 0.00627 0.00267 ± 0.00119 ± 0.00241
40K/41K = 0.0017343 ± 0.0000061 ± 0.00000564 ± 0.00000015 ± 0.0000003
a

The overall limit of error is the sum of the 95 percent confidence limits and the terms covering effects of known sources of possible systematic error.

An extensive survey of terrestrial potassium by Garner et al. [56] found no major isotopic abundance variations among approximately 70 mineral samples of different geological origins and geographic locations. Particular attention was given to include samples which might be expected to show isotopic variations or which had been previously reported to show significant isotopic variations. The average value of the minerals is biased approximately 0.03 percent higher than the 39K/41K of the reference standard, SRM 985. This difference is attributed to an altered fractionation pattern rather than to real differences in isotopic abundance between the minerals and reference standard. The alteration in fractionation pattern of the minerals is produced by the relative small amounts of organic and/or inorganic impurities which are not removed by the chemical separation and purification procedure.

Based on this survey, the atomic weight of terrestrial potassium is calculated to be 39.09830±0.00019. The error limits include an additional component of ±0.000136 to cover uncertainties in the terrestrial mineral survey data resulting from possible natural isotopic variations and bias introduced by the chemical purification procedure. The reported isotopic variations of lunar potassium [51, 58] were treated as a special case and are not covered by the error limits for terrestrial potassium.

Acknowledgments

We are indebted to: Hsien H. Ku for statistical analysis; W. R. Shields, former Section Chief, for his advice and efforts in initiating the project; W. A. Bowman, III, for maintenance support; Carrie A. Garfink for filament fabrication and assistance in reduction of strip chart recorder data; Joy J. Shoemaker and Josephine Ledford for skillful manuscript preparation.

Footnotes

1

Figures in brackets indicate the literature references at the end of this paper.

2

Company names are identified in this paper. Such identification does not imply recommendation or endorsement by the National Bureau of Standards.

3

A commercial material is identified in this paper in order to adequately specify experimental procedure. Such identification does not imply recommendation or endorsement by the National Bureau of Standards.

4. References

  • [1].Shields W. R., Garner E. L., and Dibeler V. H., J. Res. Nat. Bur. Stand. (U.S.) 66A (Phys. and Chem.) 1–3 (Jan-Feb 1962). [Google Scholar]
  • [2].Shields W. R., Murphy T. J., Garner E. L., and Dibeler V. H., J. Am. Chem. Soc. 84, 1519–1522 (1962). [Google Scholar]
  • [3].Shields W. R., Murphy T. J., and Garner E. L., J. Res. Nat. Bur. Stand. (U.S.) 68A (Phys. and Chem.) 589–592 (Nov-Dec 1964). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Catanzaro E. J., Murphy T. J., Garner E. L., and Shields W. R., j. Res. Nat. Bur. Stand. (U.S.) 68A (Phys. and Chem.) 593–599 (Nov-Dec 1964). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Shields W. R., Murphy T. J., Catanzaro E. j., and Garner E. L., J. Res. Nat. Bur. Stand. (U.S.) 70A (Phys. and Chem.) 193–197 (Mar-Apr 1966). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Catanzaro E. J., Murphy T. J., Garner E. L., and Shields W. R., J. Res. Nat. Bur. Stand. (U.S.) 70A (Phys. and Chem.) 453–458 (Nov-Dec 1966). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Catanzaro E. J., Murphy T. J., Shields W. R., and Garner E. L., J. Res. Nat. Bur. Stand. (U.S.) 72A (Phys. and 261–267 (May-Jun 1968). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [8].Catanzaro E. J., Champion C. E., Garner E. L., Marinenko G., Sappenfield K. M., and Shields W. R., Nat. Bur. stand (U.S.), Spec. Publ. 260–17, 70 pages (Feb. 1969). [Google Scholar]
  • [9].Catanzaro E. J., Murphy T. J., Garner E. L., and Shields W. R., J. Res. Nat. Bur. Stand. (U.S.) 73A (Phys. and Chem.) 511–516 (Sep-Oct 1969). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [10].Gramlich J W., Murphy T. j., Garner E. L., and Shields W. R., J. Res. Nat. Bur. Stand. (U.S.) 77A (Phys. and Chem.) 691–698 (Nov-Dec 1973). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Barnes I. L., Moore L. J., Machlan L. A., Murphy T. J., and Shields W. R., Absolute Isotopic Abundance Ratios and the Atomic Weight of a Reference Sample of Silicon, J. Res. Nat. Bur. Stand. (U.S.) 79A (Phys. and Chem.) No.6 (Nov-Dec 1975) [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Wapstra A. H., and Gove N. B., Nuclear Data Tables 9, 265–301 (1971). [Google Scholar]
  • [13].Cameron A. E., Anal. Chem. 36, 23A–35A (1963). [Google Scholar]
  • [14].Brewer A. K., J. Amer. Chem. Soc. 58, 365–370 (1936). [Google Scholar]
  • [15].Harms G., Blanc G., Espagno L., and Blanc D., Comptes Rendus 255, 3203–3204 (1962). [Google Scholar]
  • [16].Letolle R., Comptes Rendus 254, 2205–2206 (1962). [Google Scholar]
  • [17].Letolle R., Comptes Rendus 257, 3996–3998 (1963). [Google Scholar]
  • [18].Schreiner G. D. L., and Verbeek A. A., Proc. Roy. Soc. A285, 423–429 (1964). [Google Scholar]
  • [19].Lebadev V. I., Prokofyev L. M., Kirillov A. S., Tarasov A. V., Geokhimiya 10, 1255–1257 (1966). [Google Scholar]
  • [20].Verbeek A. A., and Schreiner G. D. L., Geochim. Cosmochim. Acta 31, 2125–2133 (1967). [Google Scholar]
  • [21].Schreiner G. D. L., and Welke H. J. H. F. D., Geochim. Cosmochim. Acta 35, 719–726 (1971). [Google Scholar]
  • [22].Cook K. L., Phys. Rev. 64, 278–293 (1943). [Google Scholar]
  • [23].Kendall B. R. F., Nature 186, 255–256 (1963). [Google Scholar]
  • [24].Nier A. O., Phys. Rev. 77, 789–793 (1950). [Google Scholar]
  • [25].Reuterswärd C., Arkiv Fysik 11, 1–54 (1956). [Google Scholar]
  • [26].Wichers E., J. Am. Chem. Soc. 74, 2447–2450 (1952). [Google Scholar]
  • [27].Richards T. W., and Staehler, J. Am. Chem. Soc. 29, 623–639 (1907). [Google Scholar]
  • [28].Richards T. W., and Müeller E., .J. Am. Chem. Soc. 29, 639–656 (1907). [Google Scholar]
  • [29].Hünigschmid O., and Gaubeau J., Z. Anorg. Allgem. Chem. 163, 93–164 (1927). [Google Scholar]
  • [30].Höigschmid O., and Gaubeau J., Z. Anorg. Allgem. Chem. 177, 102–108 (1928). [Google Scholar]
  • [31].Baxter G. P., and MacNevin W. M., J. Am. Chem. Soc. 55, 3185–3190 (1933). [Google Scholar]
  • [32].Hönigschmid O., and Sachtleben R., Z. Anorg. Allegem. Chem. 213, 365–376 (1933). [Google Scholar]
  • [33].Johnson C. R., J. Phys. Chem. 39, 781–789 (1935). [Google Scholar]
  • [34].Baxter G. P., and Harrington C. D., J. Am. Chem. Soc. 62, 1836–1837(1940). [Google Scholar]
  • [35].McAlpine R. K., and Bird E. J., J. Àm. Chem. Soc. 63, 2960–2965 (1941). [Google Scholar]
  • [36].Marinenko G.,Talanta 16,1339–1340(1969). [DOI] [PubMed] [Google Scholar]
  • [37].Bates R. G., and Wichers E., J. Res. Nat. Bur. Stand. (U.S.) 59, 9–17 (July 1957). [Google Scholar]
  • [38].Brewer A. K., and Kueck P. D., Phys. Rev. 46, 894–897 (1934). [Google Scholar]
  • [39].Nier A. O., Phys. Rev. 48, 283–284 (1935). [Google Scholar]
  • [40].Manley J. H., Phys. Rev. 49, 921–924 (1936). [Google Scholar]
  • [41].Bondy H., and Vanicek V., Z. Physik 101, 186–192(1936). [Google Scholar]
  • [42].Paul W., Z. Physik 124, 244–257(1948). [Google Scholar]
  • [43].White J. R., and Cameron A. E., Phys. Rev. 74, 991–1000 (1948). [Google Scholar]
  • [44].Hibbs R. F., and Redmond J. W., Y-290, AECI-24 (1949). [Google Scholar]
  • [45].Reuterswärd C., Ark. Fysik 4, 203–205 (1952). [Google Scholar]
  • [46].White F. A., Collins T. H., and Rourke F. M., Phys. Rev. 101, 1786–1791 (1956). [Google Scholar]
  • [47].Omura I., and Morito N., J. Phys. Soc. Japan 13,659 (1958). [Google Scholar]
  • [48].Stauffer H., and Honda M., J. Geophys. Res. 67, 3503–3512 (1962). [Google Scholar]
  • [49].Shukolyukov Yu A., Tolstikhin I. N., and Varshavskaya E. S., Geokhimiya 4, 373–375 (1964). [Google Scholar]
  • [50].Schreiner G. D. L., and Verbeek A. A., Proc. Roy. Soc. A285, 423–429 (1965). [Google Scholar]
  • [51].Barnes I. L., Garner E. L., Gramlich J. W., Machlan L. A., Moody J. R., Moore L. J., Murphy T. J., and Shields W. R., Proceedings of the Fourth Lunar Science Conference, Geochim. Cosmochim. Acta Suppl. 4, 2, 1197–1207 (1973). [Google Scholar]
  • [52].Shields W. R., (Ed.), Nat. Bur. Stand. (U.S.) Tech. Note; 426, 53 pages (Sept. 1967). [Google Scholar]
  • [53].Shields W. R., (Ed.), Nat. Bur. Stand. (U.S.) Tech. Note; 277, 99 pages (July 1966). [Google Scholar]
  • [54].Garner E. L., Machlan L. A., and Shields W. R., Nat. Bur. Stand. (U.S.) Special Publication; 260–27, 150 pages (April 1971). [Google Scholar]
  • [55].Garner E. L., Machlan L. A., Gramlich J. W., Moore L. A., Murphy T. J., and Barnes I. L., 7th IMR Research Symp., Gaithersburg, MD, Oct. 7–11, 1974. NBS Spec. Publ. 422, U.S. Government Printing Office, Washington. D.C. 20402 (In press). [Google Scholar]
  • [56].Garner E. L., Goldich S. S., Gramlich J. W., Machlan L. A., and Murphy T. J., Absolute Isotopic Composition of Terres-trial Potassium (In preparation).
  • [57].Kuehner E. C., Alvarez R., Paulsen P. J., and Murphy T. J., Anal. Chem. 44, 2050–2056(1972). [Google Scholar]
  • [58].Garner E. L., Machlan L. A., and Barnes I. L., Proceedings of the Sixth Lunar Science Conference (In press).

Articles from Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry are provided here courtesy of National Institute of Standards and Technology

RESOURCES