Abstract
Left ventricular assist devices (LVADs) restore cardiovascular circulatory demand at rest with a spontaneous increase in pump flow to exercise. The relevant contribution of cardiac output provided by the LVAD and ejected through the aortic valve for exercises of different intensities has been barely investigated in patients. The hypothesis of this study was that different responses in continuous recorded pump parameters occur for maximal and submaximal intensity exercises and that the pump flow change has an impact on the oxygen uptake at peak exercise (pVO2). Cardiac and pump parameters such as LVAD flow rate (Q LVAD), heart rate (HR), and aortic valve (AV) opening were analyzed from continuously recorded LVAD data during physical exercises of maximal (bicycle ergometer test) and submaximal intensities (6‐min walk test and regular trainings). During all exercise sessions, the LVAD speed was kept constant. Cardiac and pump parameter responses of 16 patients for maximal and submaximal intensity exercises were similar for Q LVAD: +0.89 ± 0.52 versus +0.59 ± 0.38 L/min (P = 0.07) and different for HR: +20.4 ± 15.4 versus +7.7 ± 5.8 bpm (P < 0.0001) and AV‐opening with 71% versus 23% of patients (P < 0.0001). Multi‐regression analysis with pVO2 (R 2 = 0.77) showed relation to workload normalized by bodyweight (P = 0.0002), HR response (P = 0.001), AV‐opening (P = 0.02), and age (P = 0.06) whereas the change in Q LVAD was irrelevant. Constant speed LVADs provide inadequate support for maximum intensity exercises. AV‐opening and improvements in HR show an important role for higher exercise capacities and reflect exercise intensities. Changes in pump flow do not impact pVO2 and are independent of AV‐opening and response in HR. An LVAD speed control may lead to adequate left ventricular support during strenuous physical activities.
Keywords: Left ventricular assist device, Mechanical circulatory support, Exercise, Cardiac response, Cardiac rehabilitation
Increased support duration with newer generations left ventricular assist devices and availability of donor organs only for selected patients result in a widespread use of left ventricular assist devices (LVAD) 1. Smaller LVAD devices that allow less invasive surgical implantation techniques as well as enhanced strategies to manage blood pressure and anti‐coagulation have been important improvements of LVAD therapy 2.
LVADs increase perfusion, reduce pulmonary pressures, improve quality of life, and submaximal exercise capacity compared to pre LVAD condition 2, 3, 4, 5, 6. Significant improvements in NYHA functional class from NYHA class prior to LVAD implant of IIIb‐IV to I–II after 1 year were reported 7, 8. LVAD patients from our center perform cardiac rehabilitation to restore and improve aerobic work capacity, physical strength, and mobility after LVAD implantation 9. Exercise training for LVAD patients is safe and effective 10, 11, 12 with significant improvements in exercise capacity and quality of life scores after at least 6 weeks of training 11. However very heterogeneous inter‐individual exercise and functional capacities were reported, which are due to the different severities of heart failure and the presence of other organs’ dysfunction 10, 12.
Maximal physical capacity in patients during LVAD support remains very low. According to the literature 12, 13 oxygen uptake at peak exercise (pVO2) in LVAD recipients only reaches 40–50% of the predicted pVO2. The limited exercise capacities observed in most patients are related to the pathophysiological changes of the cardiovascular, neurologic, respiratory, and musculoskeletal system 12. These changes result in reduced cardiac output and pVO2 14, 15, 16, 17, 18, 19. Factors that could influence cardiac output and thus aerobic capacity are left and right ventricular function, chronotropic response, pulmonary functional capacity, endothelial function, and anemia 12, 13.
Most LVADs used nowadays are axial or centrifugal rotary blood pumps operating at a constant speed 20. These LVADs are sensitive to preload and afterload, because the intrinsic dependency of the pump flow from the pump’s head pressure (= aortic pressure–left ventricular pressure) 21. In previous studies, it has been shown that LVAD pump flow increases with exercise 15, 18. However, the level of LVAD flow increase to various submaximal exercises remained unclear.
Due to the complex interaction of the native left ventricle (LV), the LVAD, and the overall circulatory system, it is not yet well known if and to which extent the LVAD output contributes to maximal and submaximal exercises. In this study, a unique method to continuously monitor patients’ LVAD data from implant onward are used allowing observations of the LVAD output and other related factors during stationary as well as nonstationary exercises.
The aim of this study is to compare pump and cardiac response during physical exercises at maximal and submaximal levels as well as to explore associations of parameters with pVO2. Furthermore, exercise related increase in pump flow is investigated together with cardiac parameters during maximum exercise tests. We hypothesized that even at constant speed, the change in pump flow and cardiac response in LVAD patients is higher at maximal compared to submaximal exercise and that the increase in pump flow has an impact on the pVO2.
Patients and Methods
In order to investigate LVAD patients’ exercise responses, LVAD data were monitored during cardiac rehabilitation from the patients within the clinical study for continuous LVAD monitoring. LVAD data together with the obtained exercise parameters were analyzed retrospectively.
LVAD monitoring
A data recorder has been developed to record the continuous data stream provided by the HVAD (Medtronic Inc., Minneapolis, MN, USA) serial port. The recorder is battery‐powered and small in size (80 × 48 × 28 mm) to fit within the patient’s carriage bag for LVAD peripherals. Pump data were acquired with 50 samples/sec and stored onto a miniSD card. For safety reasons, the data link from the controller is galvanic isolated therefore the recorder cannot interfere with the controller under any circumstances. Previously developed algorithms to analyze the cardiac function were applied on the estimated pump flow 22. For each cardiac cycle, parameters such as mean LVAD output (Q LVAD), aortic valve (AV) opening 23, 24 and heart rate (HR) in terms of beat‐to‐beat interval 25 were calculated from the LVAD pump flow waveform. Pump data were retrospectively analyzed together with the exercise and testing protocols using Matlab (The Mathworks Inc., Natick, MA, USA).
Clinical study
In a continuing prospective observational study approved by the Institutional Review Board of the Medical University Vienna (ClinicalTrials.gov identifier: NCT01981642), patient data for LVAD diagnostic purposes were recorded. Patients with an HVAD implanted at the General Hospital in Vienna (Austria) with age ranging from 18 to 70 years, no coagulopathies and ability to cope with LVAD peripherals determined by VAD coordinators gave written consent and were enrolled into the clinical study. For this study, only the patients with recorded LVAD data during cardiac rehabilitation were analyzed.
Evaluation of LVAD data during exercises
Maximal bicycle exercise tests (MBET) consist of cardiopulmonary stress tests and bicycle stress echocardiography. Submaximal intensity exercises comprise the 6‐min walk test (6MWT) and four exercise trainings (interval bicycle ergometer training, walking, mobilization, and strength training for the lower limbs).
Cardiopulmonary stress tests were performed in the upright position on a sitting bicycle ergometer. Bicycle stress echocardiography was performed occasionally with the patient in the semi‐recumbent position. MBETs were performed until subjectively perceived maximum physical capacity symptoms. The incremental increase in workload for MBETs was applied individually based on the patient’s physical performances in the submaximal intensity exercises. The 6‐min walk test was performed indoor usually at the beginning and end of rehabilitation. Based on demographic parameters, expected pVO2 were determined by Cooper and expected 6‐min walk distance was calculated with the method by Enright 26.To improve physical capacities, regular exercise training sessions were performed regulated by Borg’s 27 subjectively perceived exertion (for further information on training therapy see 10). To determine LVAD exercise responses, LVAD data were analyzed and combined with the clinical evaluations of MBETs and submaximal intensity exercises.
Responses for MBETs and submaximal intensity exercises were calculated by the difference of baseline values prior to exercise and peak exercise. Data available for multiple medical training sessions and exercise tests were pooled for each patient before calculating the average responses of all patients. Aortic valve status was determined by the beat‐to‐beat calculated aortic valve opening prevalence 23, 24 with AV closed (<20% AV‐opening), intermittent AV‐opening (20–90% AV‐opening), and complete AV‐opening (≥90% AV‐opening).
Statistical analysis
Measurable responses to exercise were verified for normal distribution with a Kolmogorov–Smirnov test and means ± standard deviations (STD) were calculated. To test the differences of response in Q LVAD and HR for maximal and submaximal intensity exercises, the Bonferroni–Holm adjusted 2 sample t‐test for unequal variances was used. Bonferroni–Holm adjusted Fisher’s exact test was used to test differences in number of patients with complete AV‐opening. For MBETs, the relationship of the Q LVAD response with HR response was tested with Pearson correlation. The differences in ∆Q LVAD based on AV‐opening during bicycle ergometer stress tests were analyzed with the Bonferroni–Holm adjusted 2 sample t‐test for unequal variances. For statistical tests, the significance level of P < 0.05 was used. For the cardiopulmonary stress tests, a stepwise regression analysis was performed to estimate the relations of pVO2 with parameters associated with exercise capacity such as age, bodyweight, cardiovascular parameters, and muscular performance. The initial model was created with all variables and compared with the explanatory power of models with fewer variables. Predictors with insufficient evidence to reject the null hypothesis that the variable has a zero coefficient were excluded from the model. Backwards elimination of predictor variables was performed for P > 0.1.
Results
Patient population and cardiac rehabilitations
LVAD data of 16 patients were recorded during cardiac rehabilitation from February 2014 until December 2016. The patient demographics are shown in Table 1. Patients underwent cardiac rehabilitation immediately following hospitalization due to initial LVAD implantation and/or later. In particular 14 of the 16 patients underwent cardiac rehabilitation on postoperative day (POD) 47 ± 22, whereas seven patients performed an additional cardiac rehabilitation on POD 628 ± 327. Average duration for in‐patient cardiac rehabilitation was 35 ± 8 days during which LVAD data were recorded for 84% of days. The average LVAD impeller speed among all patients and during all rehabilitation periods was 2884 ± 221 rpm.
Table 1.
n | Mean ± STD | |
---|---|---|
Patients | 16 | |
Gender (male/female) | 14/2 (87/13%) | |
Age (years) | 57.4 ± 12.8 | |
BMI (kg/m2) | 27.9 ± 5.1 | |
Intermacs level | 1:5 (31%) | |
2:2 (13%) | ||
3:5 (31%) | ||
4:4 (25%) | ||
Etiology: CMP (isc. / non‐isc.) | 7/9 (44/56%) | |
LVAD indication | BTT: 6 (38%) | |
BTC: 5 (31%) | ||
BTR: 1 (6%) | ||
DT: 4 (25%) | ||
Co‐morbidities | ||
Diabetes mellitus | 5 (31%) | |
Pulmonary hypertension | 4 (25%) | |
Arterial hypertension | 7 (44%) | |
Atrial fibrillation | 3 (19%) | |
ICD | 11 (69%) | |
Renal Insufficiency | 4 (25%) | |
COPD | 3 (19%) |
BTT, bridge to transplantation, BTC, bridge to candidacy, BTR, bridge to recovery, DT, destination therapy.
Comparison of LVAD data during exercise
LVAD data during MBETs (n = 24) and submaximal intensity exercises comprising 6MWT (n = 16), bicycle ergometer training (n = 100), walking training (n = 137), strength training (n = 71), and mobilization training (n = 134) were analyzed. Figure 1 shows the comparison of the average responses for maximal and submaximal intensity exercises. LVAD output increased by +0.89 ± 0.52 L/min during MBETs, +0.69 ± 0.48 L/min during 6MWT and on average by +0.55 ± 0.33 L/min during all training sessions. The increase in HR was +20.4 ± 15.4 bpm during MBETs, +12.9 ± 6.4 bpm during 6MWT and +5.3 ± 4.3 bpm during training sessions. Complete AV‐opening occurred in 71% of patients during MBETs, in 55% during 6MWT and in 15 ± 4.6% during training sessions. The achieved 6‐min walk distance was 316 ± 80 m and corresponds to 52 ± 9% of the expected distance.
For MBETs an average peak workload of 0.61 ± 0.34 W/kg was performed within 7.8 ± 1.9 min. Complete AV‐opening at peak exercise occurred in 10 patients, in two patients the AV remained closed and the remaining two patients showed intermittent AV‐opening. No correlation (r = 0.11, P = 0.7) between response in Q LVAD and HR existed. Similar Q LVAD response for complete AV‐opening and intermittent AV‐opening together with AV closed (+0.87 ± 0.52 vs. +0.94 ± 0.58 L/min, P = 0.8) were observed.
Cardiopulmonary bicycle ergometer stress tests
Twelve of the 16 patients underwent a total of 18 cardiopulmonary bicycle stress tests. The predictors of the regression model were peak workload normalized by bodyweight (P = 0.0002), ∆HR (P = 0.001), peak AV‐opening prevalence (P = 0.02), and age (P = 0.06). From the initial set of predictors considered in the regression model, ∆Q LVAD (P = 0.8) was not predictive enough. The multilinear regression model explained (R 2) 77% of the variation in pVO2 (P = 0.0001) and is expressed in Eq. 1. Clinical data combined with LVAD data from the cardiopulmonary stress tests are shown in Table 2.
(1) |
Table 2.
Mean ± STD | |
---|---|
Exercise duration (min) | 8.0 ± 1.6 |
pVO2 (mL/kg/min) | 9.9 ± 2.3 |
Expected pVO2 (%) | 35 ± 2.5% |
Respiratory exchange ratio (RER) | 1.1 ± 0.1 |
VE/VCO2 | 47.8 ± 6.7 |
Maximum workload (W/kg) | 0.55 ± 0.28 |
∆Q LVAD (L/min) | +1.0 ± 0.6 (P < 0.0001) |
∆Heart rate (bpm) | +16.7 ± 15.1 (P = 0.0002) |
Aortic valve at peak exercise (% of patients, n) | AV closed: 17% (2 of 12) |
Intermittent AV‐opening: 17% (2 of 12) | |
Complete AV‐opening: 66% (8 of 12) | |
(P = 0.04) |
Discussion
The main mechanism that leads to an increase in LVAD output at constant speed is a decrease in the pump’s head pressure. The key determinates for a decreased pump head are increasing venous return 28 and LV contractility or reducing afterload, for example, due to vasodilation. These effects occur as a cardiovascular response during exercise depending on exercise intensity and duration 15, 18.
The results from the current study suggest that this increase in pump flow (when an LVAD is driven at constant speed) is moderate and similar for both maximal and submaximal exercises. While the increase in LVAD output in response to low‐intensity exercise seem sufficient, additional improvements to maximal exercise was expected. Indeed, the response of the supported heart (reflected by the ∆HR and AV‐opening) was proportional to the intensity (shown in Fig. 1), whereas the ∆Q LVAD remained unchanged. The additional circulatory demand during MBETs therefore must be compensated by the remaining cardiovascular function rather than supported by an increase in LVAD output.
This result underlines the importance of the remaining LV contractility to generate ejection through the AV additional to the pump output. AV‐opening prevalence at peak exercise and not ∆Q LVAD was a sufficient predictor of exercise capacity (see Eq. 1). This confirms that the necessary additional cardiac output during peak exercise is mainly generated by the ejection through the AV, as it was also reported in Refs. 15, 18, 29.
Additionally, when the AV opens, a competition between the supported ventricle and the LVAD can occur. The increasing ejection through the AV leads to a reduction of the concomitant increase expected in pump output due to the LVADs’ pump head sensitivity. Increased ejection through the AV results in increased arterial pulse pressure and afterload 15, 18, 30. At peak exercise, preload and afterload together with the ejection through the AV are maximized 15, 16, 17, 18, 19. Depending on the remaining LV contractility during exercise, the increase in ejection through the AV results in increased afterload and leads to attenuation of the increase in LVAD output.
This competing mechanism explained in the previous paragraph, where the ejection through the AV has a relevant concomitant effect on LVAD output is evident in additional results. During MBETs, similar ∆Q LVAD for closed and open AVs were obtained despite the existing relationship of AV‐opening and exercise capacity (see Eq. 1). Furthermore, higher exercise capacities were obtained by patients with a better LV function, in this study analyzed by ∆HR and AV‐opening, as reported in Ref. 31. Additionally, heart rate is a potent facilitator of the competing mechanism and no correlation between ∆HR and ∆Q LVAD existed, as mentioned in reference 28.
During submaximal trainings, the limited cardiovascular contribution resulted in more patients remaining in full‐support (see Fig. 1) with the whole cardiac output being provided by the LVAD. With increasing exercise intensity increase in preload, inotropic, chronotropic, and vascular response leads to ejection through the AV and reduction of the concomitant increase in Q LVAD. Such behavior can be seen in Ref. 32 where ∆Q LVAD was +1.3 L/min from rest to light exercise, +0.2 L/min from light to moderate exercise, and +0.3 L/min from moderate to heavy exercise. Such dependency of ∆Q LVAD on exercise intensity additionally contributes to the significantly higher (P = 0.003) expected values for submaximal 6‐min walk distance compared with expected pVO2s in this study and mentioned in Ref. 32.
This study shows that LVAD pump output at peak exercise is rather limited and patients with better exercise capacity rely heavily on their remaining cardiovascular function. LVAD speed increase during exercise results in potential benefits independent on the patient’s ability to increase cardiac output through the AV. Patients in full‐support at peak exercise with constant LVAD speed would increase cardiac output and reduce LV preload whereas in partially supported patients a better unloading would allow the supported ventricle to still keep some of its functional reserve. It seems important that the strategy of support depends on the underlying remaining function of the assisted left ventricle and the right ventricle. Therefore, strategies to change pump speed in response to exercise will be different depending on the individual patients’ condition.
Furthermore, the question if current miniaturized LVADs are powerful enough to overtake the whole cardiac output at peak exercise in “less‐sick” patients has to be addressed, since these patients would most likely be in a condition of partial support with their left ventricle engaging in “competition” with the LVAD.
Among the factors contributing to the pVO2, those representing the remaining cardiovascular function (∆HR and AV‐opening) and the skeletal muscle contractile and metabolic system (performed workload) dominated rather than pump response in terms of LVAD output (see Eq. 1). For patients remaining in full‐support during exercise, Q LVAD reflects cardiac output. In partial‐supported patients, ∆HR and AV‐opening seem to be sufficient parameters representing the cardiac contribution to an increase in cardiac output (see Fig. 1). Thus monitoring such as used in this study allows determining cardiac responses as sensed by the LVAD. The measurement about the contribution of the LV to the overall cardiac output would, of course, provide a more comprehensive picture. Continuous LVAD monitoring could be used to guide training sessions which would require additional investigations. Furthermore, improved monitoring could also reveal pattern of daily activity and thus lead to a better understanding of activity efforts and compliance to outpatient training protocols.
This study has some limitations. As for most observational studies, this work presents results about the associations of LVAD parameters during physical exercise rather than causations. The sample size for this study is too small to draw broader conclusions and only patients within the clinical study were analyzed. The quantification of the flow redistribution between the LV and the LVAD would be very interesting to assess the cooperation of the two pumping systems limiting factors at peak exercise. Cardiac output estimation, for example through rebreathing methods 33 were not available during MBETs which would provide important information. Assumptions on the change in CO were based on surrogates such as AV‐opening and ∆HR. Blood pressure measurements would provide valuable information on the observed ∆Q LVAD resulting from the decrease in the pump’s head pressure. Due to the low arterial pulse pressure, pulsatility of LVAD patients most blood pressure measurements were unreliable, especially at rest 34.
Conclusion
For near‐maximal exercise intensities, patients rely heavily on their remaining left ventricular function with left ventricular assist devices providing support sufficiently, at best, for only low‐intensity exercises. LVAD output response is indeed similar for high‐ and low‐intensity exercises. Parameters representing the cardiovascular function such as ∆HR and AV‐opening contribute to pVO2s instead of ∆Q LVAD. Thus indicating greater contribution of the heart than the LVAD to cardiac output at peak exercise compared to rest. An appropriate LVAD speed control may provide optimized pump output at peak exercise and increase exercise capacity of LVAD patients.
Authors Contributions
FM and HS developed the concept and design, CG, FM, and TS performed the statistical analysis, all authors contributed to the interpretation of data, funding was secured by FM, HS, DZ, CM, JM and JA, CG drafted the article, CG and TS collected the data, all authors performed critical revision of the article and approved the final version.
Conflict of interest
The authors have declared no conflicts of interest for this article.
Funding information
This study was supported by the Austrian Science Fund (FWF): KLI357.
Acknowledgments
The authors would like to thank Dr. Erwin Kosanyi (Rehabilitation Center Felbring, Austria) and Dr. Richard Steinacher (Rehabilitation Center Großgmain, Austria) for their assistance with the performed exercise tests.
References
- 1. Kirklin JK, Naftel DC, Pagani FD, et al. Sixth INTERMACS annual report: a 10,000‐patient database. J Heart Lung Transplant 2014;33:555–64. [DOI] [PubMed] [Google Scholar]
- 2. Pinney SP, Anyanwu AC, Lala A, Teuteberg JJ, Uriel N, Mehra MR. Left ventricular assist devices for lifelong support. J Am Coll Cardiol 2017;69:2845–61. [DOI] [PubMed] [Google Scholar]
- 3. Salzberg SP, Lachat ML, vonHarbou K, Zünd G, Turina MI. Normalization of high pulmonary vascular resistance with LVAD support in heart transplantation candidates? Eur J Cardiothorac Surg 2005;27:222–5. [DOI] [PubMed] [Google Scholar]
- 4. Dunlay SM, Allison TG, Pereira NL. Changes in cardiopulmonary exercise testing parameters following continuous flow left ventricular assist device implantation and heart transplantation. J Card Fail 2014;20:548–54. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Rogers JG, Aaronson KD, Boyle AJ, et al. Continuous flow left ventricular assist device improves functional capacity and quality of life of advanced heart failure patients. J Am Coll Cardiol 2010;55(17):1826–34. [DOI] [PubMed] [Google Scholar]
- 6. Rosenbaum AN, Dunlay SM, Pereira NL, et al. Determinants of improvement in cardiopulmonary exercise testing after left ventricular assist device implantation. ASAIO J 2018;64:1–10. [DOI] [PubMed] [Google Scholar]
- 7. Krabatsch T, Netuka I, Schmitto JD, et al. Heartmate 3 fully magnetically levitated left ventricular assist device for the treatment of advanced heart failure–1 year results from the Ce mark trial. J Card Surg [Internet]. 2017;12 Available at: https://cardiothoracicsurgery.biomedcentral.com/articles/10.1186/s13019-017-0587-3. Accessed July 3, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Benton CR, Sayer G, Nair AP, et al. Left ventricular assist devices improve functional class without normalizing peak oxygen consumption. ASAIO J 2015;61:237–43. [DOI] [PubMed] [Google Scholar]
- 9. Marko C, Xhelili E, Lackner T, Zimpfer D, Schima H, Moscato F. Exercise performance during the first two years after left ventricular assist device implantation. ASAIO J 2017;63:408–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Marko C, Danzinger G, Käferbäck M, et al. Safety and efficacy of cardiac rehabilitation for patients with continuous flow left ventricular assist devices. Eur J Prev Cardiol 2014;7 10.1177/2047487314558772. [DOI] [PubMed] [Google Scholar]
- 11. Ganga HV, Leung A, Jantz J, et al. Supervised exercise training versus usual care in ambulatory patients with left ventricular assist devices: a systematic review. PLOS One . 12:e0174323. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Hayward CS, Fresiello L, Meyns B. Exercise physiology in chronic mechanical circulatory support patients: vascular function and beyond. Curr Opin Cardiol . 2016;31:292–8. [DOI] [PubMed] [Google Scholar]
- 13. Jung MH, Gustafsson F. Exercise in heart failure patients supported with a left ventricular assist device. J Heart Lung Transplant 2015;34:489–96. [DOI] [PubMed] [Google Scholar]
- 14. Loyaga‐Rendon RY, Plaisance EP, Arena R, Shah K. Exercise physiology, testing, and training in patients supported by a left ventricular assist device. J Heart Lung Transplant 2015;34:1005–16. [DOI] [PubMed] [Google Scholar]
- 15. Jacquet L, Vancaenegem O, Pasquet A, et al. Exercise capacity in patients supported with rotary blood pumps is improved by a spontaneous increase of pump flow at constant pump speed and by a rise in native cardiac output: exercise capacity with rotary pumps. Artif Organs 2011;35:682–90. [DOI] [PubMed] [Google Scholar]
- 16. Muthiah K, Robson D, Prichard R, et al. Effect of exercise and pump speed modulation on invasive hemodynamics in patients with centrifugal continuous‐flow left ventricular assist devices. J Heart Lung Transplant 2015;34:522–9. [DOI] [PubMed] [Google Scholar]
- 17. Burrell A, Hayward C, Mariani J, Leet A, Kaye DM. Clinical utility of invasive exercise hemodynamic evaluation in LVAD patients. J Heart Lung Transplant 2015;34(12):1635–7. [DOI] [PubMed] [Google Scholar]
- 18. Martina J, Jonge N, Rutten M, et al. Exercise hemodynamics during extended continuous flow left ventricular assist device support: the response of systemic cardiovascular parameters and pump performance: exercise hemodynamics during CF‐LVAD support. Artif Organs 2013;37:754–62. [DOI] [PubMed] [Google Scholar]
- 19. Haft J, Armstrong W, Dyke DB, et al. Hemodynamic and exercise performance with pulsatile and continuous‐flow left ventricular assist devices. Circulation . 2007;116(11_suppl):I‐8–I‐15. [DOI] [PubMed] [Google Scholar]
- 20. Kirklin JK, Naftel DC, Pagani FD, et al. Seventh INTERMACS annual report: 15,000 patients and counting. J Heart Lung Transplant 2015;34:1495–504. [DOI] [PubMed] [Google Scholar]
- 21. Moazami N, Fukamachi K, Kobayashi M, et al. Axial and centrifugal continuous‐flow rotary pumps: a translation from pump mechanics to clinical practice. J Heart Lung Transplant 2013;32:1–11. [DOI] [PubMed] [Google Scholar]
- 22. Granegger M, Moscato F, Casas F, Wieselthaler G, Schima H. Development of a pump flow estimator for rotary blood pumps to enhance monitoring of ventricular function: pump flow estimator for rotary blood pumps. Artif Organs 2012;36:691–9. [DOI] [PubMed] [Google Scholar]
- 23. Granegger M, Schima H, Zimpfer D, Moscato F. Assessment of aortic valve opening during rotary blood pump support using pump signals: aortic valve opening during RBP support. Artif Organs 2014;38:290–7. [DOI] [PubMed] [Google Scholar]
- 24. Granegger M, Masetti M, Laohasurayodhin R, et al. Continuous monitoring of aortic valve opening in rotary blood pump patients. IEEE Trans Biomed Eng 2016;63:1201–7. [DOI] [PubMed] [Google Scholar]
- 25. Moscato F, Granegger M, Edelmayer M, Zimpfer D, Schima H. Continuous monitoring of cardiac rhythms in left ventricular assist device patients. Artif Organs 2014;38:191–8. [DOI] [PubMed] [Google Scholar]
- 26. Enright PL, Sherrill DL. Reference equations for the six‐minute walk in healthy adults. Am J Respir Crit Care Med 1998;158:1384–7. [DOI] [PubMed] [Google Scholar]
- 27. Borg G. Borg’s perceived exertion and pain scales. Hum Kinet Champaign, IL, USA, 1998. [Google Scholar]
- 28. Muthiah K, Gupta S, Otton J, et al. Position and activity, but not heart rate, affect pump flows in patients with continuous‐flow left ventricular assist devices. JACC: Heart Fail 2014;2(4):323–30. [DOI] [PubMed] [Google Scholar]
- 29. Fresiello L, Rademakers F, Claus P, Ferrari G, DiMolfetta A, Meyns B. Exercise physiology with a left ventricular assist device: Analysis of heart‐pump interaction with a computational simulator. PLOS One. 2017;12:e0181879. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Myers TJ, Bolmers M, Gregoric ID, Kar B, Frazier OHH. Assessment of arterial blood pressure during support with an axial flow left ventricular assist device. J Heart Lung Transplant 2009;28:423–7. [DOI] [PubMed] [Google Scholar]
- 31. Noor MR, Bowles C, Banner NR. Relationship between pump speed and exercise capacity during HeartMate II left ventricular assist device support: influence of residual left ventricular function. Eur J Heart Fail 2012;14:613–20. [DOI] [PubMed] [Google Scholar]
- 32. Schmidt T, Bjarnason‐Wehrens B, Mommertz S, et al. Changes in total cardiac output and oxygen extraction during exercise in patients supported with an HVAD left ventricular assist device: changes in total cardiac output and oxygen extraction. Artif Organs 2018. Available at: https://doi.wiley.com/10.1111/aor.13102. Accessed May 16, 2018. [DOI] [PubMed] [Google Scholar]
- 33. Reiss N, Schmidt T, Workowski A, et al. Physical capacity in LVAD patients: hemodynamic principles, diagnostic tools and training control. Int J Artif Organs 2016;39:451–9. [DOI] [PubMed] [Google Scholar]
- 34. Castagna F, Stöhr EJ, Pinsino A, et al. The unique blood pressures and pulsatility of LVAD patients: current challenges and future opportunities. Curr Hypertension Rep 2017;19:85. [DOI] [PMC free article] [PubMed] [Google Scholar]