. 2019 Jun 9;11(3):510–515. doi: 10.1111/os.12467

Technical Note: Pedicle Cement Augmentation with Proximal Screw Toggle and Loosening

Wen Jie Choy 1,2, William R Walsh 3, Kevin Phan 1,2,4, Ralph J Mobbs 1,2,4,
PMCID: PMC6595100  PMID: 31179643

Abstract

Background

Cement augmentation is a technique used to increase the stability and purchase of pedicle screws in poor quality bone. Various methods can be applied for cement delivery, such as cement injection before screw placement and the use of fenestrated screws. However, potential problems can arise with the use of cement augmentation.

Case Presentation

A 66‐year‐old man with a lower trunk deformity, severe kyphosis, and sagittal imbalance following fusion (L2‐5), with minimal comorbidities, was referred to our unit 9 months after surgery. Pain and progressive kyphosis were investigated clinically and radiographically with computed tomography (CT) scans to assess the status of the hardware and fusion. CT imaging revealed that cement was present only at the distal tip of the fenestrated screws at the L4 vertebral level. A non‐union was present along with loosening and a halo around the body of the pedicle screws, and there was evidence of pullout of inferior screws.

Conclusion

Single‐level cement augmentation of pedicle screw in a posterior construct and distal tip cement augmentation of the screw results in a fixed pivot point. Micromotion in cranio‐caudal loading during flexion and extension may result in screw toggling with the single‐level cement‐augmented tip as a fulcrum. This may cause screw loosening, which can lead to pullout and loss of construct stability. The halo around the screw suggests bone loss and/or a fibrous tissue interface, which further complicates revision surgery. Stress shielding and polymethylmethacrylate cement present additional difficulties.

The findings of this technical note question the risks and benefits of cement‐augmented fenestrated pedicle screw fixation for spinal fusion. Although incidences of such cases are uncommon, surgeons should perform this technique with caution. Accurate restoration of lumbar lordosis during index procedures is important to minimize the risk of construct failure.

Keywords: Cement augmentation, Fenestrated pedicle screws, Lumbar spine, Peri‐screw halo, Screw loosening, Screw toggling

Introduction

The use of vertebral screws for spinal fixation dates back to the 1940s and has since evolved into pedicle screw fixation for stabilization to encourage intervertebral arthrodesis1 and for the treatment of various spinal pathologies such as stabilizing traumatic injuries, correction of deformities and spinal fusion. Various methods of pedicle screw fixation have been used including minimally invasive percutaneous screw fixation, open pedicle screw insertion2, 3, 4, and more recent trajectories including the cortical screw have gained popularity5, 6. To achieve improved stability and purchase, vertebral body and pedicle screw augmentation (kyphoplasty via the central screw cannulation and distal holes, and transpedicle injection) utilizing different bone cements such as calcium apatite cement (CAC), injectable calcium phosphate (CaP) and polymethylmethacrylate (PMMA) have been trialled7, 8, 9, 10. Multiple techniques of cement delivery are currently in practice. Formerly, bone cement was injected into the drilled lumen before insertion of the pedicle screw; recently, fenestrated pedicle screws (Fig. 1B) have been utilised which allows cement, such as PMMA, to be injected through the screw into the vertebral body after the screw is in place11, 12, 13, 14. In the setting of decreased bone mineral density (BMD) such as osteoporosis, cement augmentation is proven to improve pedicle screw purchase and outcomes15, 16.

Figure 1.

Figure 1

(A) Standing full‐body EOS scan; red box showing position of L5 vertebral body; blue arrow showing pull‐out of L5 pedicle screws. The acute kyphotic angle at L3/4 due to the construct failure was approximately 45o. (B) Sagittal CT with L2‐L5 pedicle screw construct with L4 cement‐augmented pedicle screws in vertebral body (green arrow). (C) L4 cement‐augmented pedicle screws (green arrow) on coronal CT view. Inset: L5 Proximal screw toggle with halo around pedicle screw and loosening of construct. Orange arrows indicate peri‐screw haloing due to cranial‐caudal toggling.

However, potential problems can arise with cement augmentation. Intraoperative mis‐positioning of screw holes17 and cement leakage;18 cement leakage into surrounding structures and post‐operative screw migration;19 which all can result in significant neurological deficit, morbidities or mortality such as pulmonary embolism14, 18, 19, 20, 21. Incomplete curing of PMMA upon injection can also cause thermonecrosis of surrounding neural structures22. Moreover, micromovements, loading, and twisting/ rotational movements of the relevant spinal motion segments before complete osseointegration of the fusion construct can predispose to non‐union.

Although clinical studies discussing the potential toggling of pedicle screws have been carried out, to the authors’ knowledge there is no formal publication reporting the outcomes of screw toggling in patients. This technical note will discuss the outcomes of distal cement augmented pedicle screw which serves as a pivot point for proximal screw toggling and loosening.

Case Presentation

We report a case of a 66‐year‐old male with minimal co‐morbidities who presents with lower trunk flat back deformity, severe iatrogenic kyphosis and sagittal imbalance following 9 months postoperative lumbar decompression and fusion (L2‐L5) with postero‐lateral grafting, without inter‐body implants from a different institute. He experienced significant pain due to L4 and L5 motor / sensory radiculopathy. Reflexes were absent in both lower limbs with a bilateral foot drop. Oswestry Disability Index was 74%, with Visual Analog Scale pain score of 10 in the standing position, 8 on lying flat.

Computed tomography (CT) scan revealed loosening and non‐union of the inferior aspect of the fusion construct. There is pull‐out of the inferior screws (Fig. 1A) with haloing around the body and distal aspect of the pedicle screws. (Fig. 1). Standing EOS scan reveals gross sagittal imbalance. The kyphotic angle due to the construct failure at L3/4 level was approximately 45°. A flat back deformity can be seen along the vertebral levels above the level of screw pull‐out. Bone mineral density was normal for his age. Figure 2A shows the pre‐operative presentation, with significant sagittal plane deformity and bent‐knees in order to maintain a gaze along the horizon while requiring a walking aid for ambulatory.

Figure 2.

Figure 2

(A) Pre‐operative image of patient with gross sagittal deformity and bent‐knees to accommodate for the kyphosis in the lumbar spine in order to maintain a gaze along the horizon. (B) Post‐operative image of patient with a corrected sagittal posture and requiring no walking aid for ambulatory.

The haloing effect demonstrated by the pedicle screws is consistent with hardware failure, and also seen in the “PEEK‐Halo” effect when PolyEther‐Ether‐Ketone (PEEK) is being used as an intervertebral implant which results in poor osseointegration23. However, in this case, the poor osseointegration was a result of repetitive screw cranial‐caudal micromotions and toggling which prevent consistent screw – bone contact but not due to hardware material.

A revision procedure utilizing both anterior and posterior approach was carried out (Fig. 3). Lordosis was restored utilizing 2 anterior lumbar interbody fusion (ALIF) cages (L3/4 and L4/5 levels) and Posterior Smith‐Peterson osteotomy at L3/4. Focal lordosis of approximately 22o was corrected resulting in a total correction of over 65 degrees at L3/4. The patient was able to stand up‐right which enabled him to maintain his view on horizon when his spine was in a neutral position. Figure 2B shows patient post‐op with a corrected posture requiring no walking aids to ambulate.

Figure 3.

Figure 3

(A) Post‐operative standing X‐Ray of patient. Lordosis was restored utilizing 2 ALIF cages in L3/4 and L4/5 level post‐removal of previous construct. Revision construct with up‐sized L2 and L3 pedicle screws combined with S1 and S2AI screws were used to provide a distal construct foundation. A focal lordotic correction at L3/4 of approximately 22o was restored. (B) Removed right L4 cement‐augmented pedicle screw with cement surrounding the screw tip. (C) Revision with Anterior construct (ALIF) and Posterior Smith‐Peterson osteotomy to restore lordosis of the lumbar spine.

Discussion

To the authors’ knowledge, there is currently no literature reporting such an adverse event, however the use of fenestrated cement fixation is relatively recent. A cadaveric study by Bostelmann et al. studying the effect of cranio‐caudal cyclic loading on non‐augmented and cement‐augmented pedicle screws concluded the superiority of augmentation compared to standard pedicle screws24. A separate study comparing sequence of cement augmentation before or after reduction manoeuvre of pedicle screws in a cadaveric study showed the superiority of cement augmentation post‐reduction to reinforce screw purchase25. However, the end effect of toggling due to cranio‐caudal loading was not studied.

In the case of fenestrated screws were only the tip is strongly fixated by cement in the vertebral body, the tip can act as a fulcrum in which micro‐movements and torsion can cause toggling of the screw body within the pedicle, resulting in proximal loosening, and non‐union of the fusion construct. As time passes, cranio‐caudal loading during lumbar flexion and extension, accumulates into a significant amount of force which eventually results in minute motions turning into significant toggling upon the cemented fulcrum. Further loading result in more proximal toggling. The loss of bone structure around the screw appears as a haloing effect seen on imaging (Fig. 4). The toggling effect can be further exacerbated if there is loss of bone mineral density which allows the screws to pivot easily with movements and torsion of the spine. These motions can also result in loss of cortical bone which weaken the screw fixations26. Eventually, this can lead to proximal and distal screw loosening and affect the cemented region as in this case.

Figure 4.

Figure 4

Failure of Distal Cement Augmentation of Pedicle Screw. (A) Pedicle screw position with distal fenestration for cement augmentation. (B) Cement delivery via distal fenestration. Non‐union may lead to toggle of screw and proximal loosening and failure. (C) Further movement will increase halo effect around proximal screw insertion point. (D) Halo and fracture of cement with pull‐out of screw.

The reduced forces withstood by the spine may be a potential risk factor for stress shielding to occur. As seen in the hip, stress reduction in an implanted bone can lead to bone loss27. Weight‐bearing is an important factor that promotes bone growth and deposition28. Inter‐body constructs serve as a secondary anterior column support for the spine29 and may result in reduced load being transferred through the pedicle screw construct, as in the current case. With a patient fixated with poor sagittal balance and no anterior column support, coupled with toggling and micro‐movements of the posterior pedicel screw, the chances of non‐union further increases.

The end result of such an effect can be disastrous to the patient. Revision of the loose pedicle screws presents a significant problem, as removal of these screws may lead to further damage of the surrounding bone or fracture of the vertebral body and the pedicles. The patient may require further levels of fusion as well as other invasive surgeries (e.g. anterior or lateral approaches) to correct the kyphosis. Blattert et al. argued that removal of cement‐augmented fenestrated screws was possible even in severe osteoporosis. They concluded that the cement screw connection was fragile enough to break off during extraction30. However, Bullmann et al. showed the axial pull‐out strength and torque is significantly higher in cement‐augmented fenestrated screws compared to non‐augmented screws. These were associated with pedicle and vertebral body fractures. Additionally, there is an increased chance of cement leakage if cement augmentation is used during revision31. Moreover, there are limitations associated with in‐vitro lab studies. Osseointegration and bone remodelling is not applicable in such settings as cadaveric bones do not heal. The end result of bone remodelling may affect the stability of the construct with time.

Different methods have been proposed and used for the revision of non‐augmented pedicle screws. These include the use of a larger and longer pedicle screws32, cement augmentation of the revised screws33 and pre‐operative planning of different trajectory for revision screw placements34. However, there are currently no promising techniques for the revision of cement‐augmented fenestrated screws. Removal of the screws may risk fracturing the vertebrae and neurological injury with screw/cement removal. Mesfin et al. reported a revision for a failed cement‐augmented fenestration screws fixation in a osteoporotic patient by extending the posterior construct, cement‐augmentation, and addition of a titanium hook and woven polyester band to increase the points of fixation35. If, however, the cement was broken off from the screws during revision as described by Blattert et al.,30 the residual cement can potentially act as a barrier for proper osseointegration between bone and the newly inserted screw. A cement‐in‐cement revision technique which is commonly used in hip arthroplasty36, 37 may be required to aid stability for the revised construct.

In addition to extending the posterior construct for re‐establishing sagittal balance, the authors agree that the use of interbody fusion techniques may further benefit the patient for anterior column support. In terms of correcting sagittal alignment, the use of ALIF as in the current case has been shown to compliment other realignment techniques38, 39. A minimally invasive, lateral transpsoas approach which incorporates lateral interbody fusion and anterior longitudinal ligament release reported by Murray et al. and Pimenta et al.40, 41 is another promising technique in this scenario.

Further studies and data are required to study the long‐term efficacy of cement‐augmented pedicle screw fixation, associated risks and complications. Long‐term outcomes of pedicle screw fixation comparing different screws should be carried out to compare the pros and cons of different screw designs. Recently, expandable pedicle screws have gained an interest amongst spine surgeons42, 43. Standalone comparison of expandable pedicle screws with standard pedicle screws demonstrated the former has a greater peak pullout force44. Aycan et al. suggested due to toggling and the brittleness of PMMA, expandable pedicle screws with PEEK shell might be a better alternative in dynamic loading45. Nevertheless, major complications such as the one reported here will further question the safety and efficacy of this technique. Better alternatives should be sought out to reduce the chance of such adverse outcomes in addition to a surgical focus on alignment correction with the initial surgery46.

Conclusion

The use of single‐level cement augmentation has resulted in a fulcrum for screw toggling when load was applied along the construct. Accurate restoration of sagittal alignment with ideal lumbar lordosis at the index procedure while using cement augmentation techniques would have reduced the risk of such a complication. Separately, the use of cement‐augmented fenestrated pedicle screws is a recent technique. Although there is currently an adequate body of literature on the pull‐out strength and stability of cement‐augmented fenestrated pedicle screws, there are no reports on non‐union biomechanics and the effect of cranio‐caudal toggling. Haloing around pedicle screw due to toggling is a major complication especially if cement‐augmentation was used. This phenomenon adds significant difficulty and potential morbidity to the revision procedure. Anterior or lateral interbody cages may aid with sagittal realignment. The appearance of such phenomenon potentially questions the role of such fixation technique. The use of intra‐operative on table CT scan may be beneficial to access the accuracy of screw placement in order to limit post‐operative toggling due to abnormal stress loading.

Declaration

All authors contributed significantly and met the authorship criteria according to the latest guidelines of the International Committee of Medical Journal Editors. All authors agree to the final submitted manuscript.

Disclosure: The authors declare no conflict of interest.

References

  • 1. Kabins MB, Weinstein JN. The history of vertebral screw and pedicle screw fixation. Iowa Orthop J, 1991, 11: 127–136. [Google Scholar]
  • 2. Phan K, Rao PJ, Mobbs RJ. Percutaneous versus open pedicle screw fixation for treatment of thoracolumbar fractures: Systematic review and meta‐analysis of comparative studies. Clin Neurol Neurosurg, 2015, 135: 85–92. [DOI] [PubMed] [Google Scholar]
  • 3. Smith JS, Ogden AT, Fessler RG. Minimally invasive posterior thoracic fusion. Neurosurg Focus, 2008, 25: E9. [DOI] [PubMed] [Google Scholar]
  • 4. Hsieh PC, Koski TR, Sciubba DM, et al Maximizing the potential of minimally invasive spine surgery in complex spinal disorders. Neurosurg Focus, 2008, 25: E19. [DOI] [PubMed] [Google Scholar]
  • 5. MR J. The “medio‐latero‐superior trajectory technique”: an alternative cortical trajectory for pedicle fixation. Orthop Surg, 2013, 5: 56–59. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Santoni BG, Hynes RA, McGilvray KC, et al Cortical bone trajectory for lumbar pedicle screws. Spine J, 2009, 9: 366–373. [DOI] [PubMed] [Google Scholar]
  • 7. Sun H, Liu C, Liu H, et al A novel injectable calcium phosphate‐based nanocomposite for the augmentation of cannulated pedicle‐screw fixation. Int J Nanomedicine, 2017, 12: 3395–3406. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Wuisman PI, Van Dijk M, Staal H, Van Royen BJ. Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique. European spine journal: official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical. Eur Spine J, 2000, 9: 528–533. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Alvarez L, Gomez S, Vlad MD, Pinera AR, Tome‐Bermejo F, Fernandez E. Analysis of PMMA distribution around cannulated and fenestrated cement‐augmented pedicle screws: clinical study. Spine J, 2015, 15 ( Supplement 1: 189S–190S. [Google Scholar]
  • 10. Burval DJ, McLain RF, Milks R, Inceoglu S. Primary pedicle screw augmentation in osteoporotic lumbar vertebrae: biomechanical analysis of pedicle fixation strength. Spine, 2007, 32: 1077–1083. [DOI] [PubMed] [Google Scholar]
  • 11. Klingler JH, Scholz C, Kogias E, et al Minimally invasive technique for PMMA augmentation of fenestrated screws. Scientific World Journal, 2015, 2015: 1–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Chandra VVR, Prasad BCM, Jagadeesh MA, Jayachandar V, Kumar SA, Kumar R. Segmental polymethylmethacrylateaugmented fenestrated pedicle screw fixation for lumbar spondylolisthesis in patients with osteoporosis ‐ a case series and review of literature. Neurol India, 2017, 65: 89–95. [DOI] [PubMed] [Google Scholar]
  • 13. Hu M‐H, Wu HTH, Chang M‐C, Yu W‐K, Wang S‐T, Liu C‐L. Polymethylmethacrylate augmentation of the pedicle screw: the cement distribution in the vertebral body. Eur Spine J, 2011, 20: 1281–1288. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Amendola L, Gasbarrini A, Fosco M, et al Fenestrated pedicle screws for cement‐augmented purchase in patients with bone softening: a review of 21 cases. J Orthop Traumatol, 2011, 12: 193–199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Wang W, Baran GR, Garg H, Betz RR, Moumene M, Cahill PJ. The benefits of cement augmentation of pedicle screw fixation are increased in osteoporotic bone: a finite element analysis. Spine Deform 2014;4:248–59. [DOI] [PubMed] [Google Scholar]
  • 16. Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS 3rd, Cook SD. Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976), 1994, 19: 2415–2420. [DOI] [PubMed] [Google Scholar]
  • 17. Tang J, Zhu Z, Sui T, Kong D, Cao X. Position and complications of pedicle screw insertion with or without image‐navigation techniques in the thoracolumbar spine: a meta‐analysis of comparative studies. J Biomed Res 2014;3:228–39. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Dai F, Liu Y, Zhang F, Sun D, Luo F, Zhang Z, Xu J Surgical treatment of the osteoporotic spine with bone cement‐injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients. Clinics 2015;70:114–149. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Chang MC, Kao HC, Ying SH, Liu CL. Polymethylmethacrylate augmentation of cannulated pedicle screws for fixation in osteoporotic spines and comparison of its clinical results and biomechanical characteristics with the needle injection method. J Spinal Disord Tech 2013;26:305–315. [DOI] [PubMed] [Google Scholar]
  • 20. Pesenti S, Blondel B, Peltier E, Adetchessi T, Dufour H, Fuentes S. Percutaneous cement‐augmented screws fixation in the fractures of the aging spine: is it the solution? Biomed Res Int, 2014, 2014: 1–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Janssen I, Ryang YM, Gempt J, et al Risk of cement leakage and pulmonary embolism by bone cement‐augmented pedicle screw fixation of the thoracolumbar spine. Spine J 2017;17:837–844. [DOI] [PubMed] [Google Scholar]
  • 22. Turner AW, Gillies RM, Svehla MJ, Saito M, Walsh WR. Hydroxyapatite composite resin cement augmentation of pedicle screw fixation. Clin Orthop Relat Res, 2003, 1: 253–261. [DOI] [PubMed] [Google Scholar]
  • 23. Phan K, Hogan JA, Assem Y, Mobbs RJ. PEEK‐Halo effect in interbody fusion. J Clin Neurosci, 2016, 24: 138–140. [DOI] [PubMed] [Google Scholar]
  • 24. Bostelmann R, Keiler A, Steiger HJ, Scholz A, Cornelius JF, Schmoelz W. Effect of augmentation techniques on the failure of pedicle screws under cranio‐caudal cyclic loading. Eur Spine J, 2017, 26: 181–188. [DOI] [PubMed] [Google Scholar]
  • 25. Schmoelz W, Heinrichs CH, Schmidt S, et al Timing of PMMA cement application for pedicle screw augmentation affects screw anchorage. Eur Spine J, 2017, 26: 2883–2890. [DOI] [PubMed] [Google Scholar]
  • 26. Pelletier MH, Bertollo N, Al‐Khawaja D, Walsh WR. The contribution of the cortical shell to pedicle screw fixation. J Spine Surg, 2017, 3: 184–192. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Niinimaki T, Junila J, Jalovaara P. A proximal fixed anatomic femoral stem reduces stress shielding. Int Orthop, 2001, 25: 85–88. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Parfitt AM. The two faces of growth: Benefits and risks to bone integrity. Osteoporos Int, 1994, 4: 382–398. [DOI] [PubMed] [Google Scholar]
  • 29. Mobbs RJ, Phan K, Malham G, Seex K, Rao PJ. Lumbar interbody fusion: techniques, indications and comparison of interbody fusion options including PLIF, TLIF, MI‐TLIF, OLIF/ATP, LLIF and ALIF. Journal of Spine. Surgery, 2015, 1: 2–18. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Blattert TR, Glasmacher S, Riesner HJ, Josten C. Revision characteristics of cement‐augmented, cannulated‐fenestrated pedicle screws in the osteoporotic vertebral body: a biomechanical in vitro investigation. Technical note. J Neurosurg Spine, 2009, 11: 23–27. [DOI] [PubMed] [Google Scholar]
  • 31. Bullmann V, Schmoelz W, Richter M, Grathwohl C, Schulte TL. Revision of cannulated and perforated cement‐augmented pedicle screws: a biomechanical study in human cadavers. Spine (Phila Pa 1976), 2010, 35: E932–E939. [DOI] [PubMed] [Google Scholar]
  • 32. Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims‐‐what is best? Spine (Phila Pa 1976), 1998, 23: 1374–1379. [DOI] [PubMed] [Google Scholar]
  • 33. Kiner DW, Wybo CD, Sterba W, Yeni YN, Bartol SW, Vaidya R. Biomechanical analysis of different techniques in revision spinal instrumentation: larger diameter screws versus cement augmentation. Spine (Phila Pa 1976), 2008, 33: 2618–2622. [DOI] [PubMed] [Google Scholar]
  • 34. Fujibayashi S, Takemoto M, Neo M, Matsuda S. Strategy for salvage pedicle screw placement: A technical note. Int J Spine Surg, 2013, 7: e67–e71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Mesfin A, Komanski CB, Khanna AJ. Failure of cement‐augmented pedicle screws in the osteoporotic spine: a case report. Geriatr Orthop Surg Rehabil, 2013, 4: 84–88. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Sandiford NA, Jameson SS, Wilson MJ, Hubble MJW, Timperley AJ, Howell JR. Cement‐in‐cement femoral component revision in the multiply revised total hip arthroplasty. Bone Joint J, 2017, 99‐B: 199–203. [DOI] [PubMed] [Google Scholar]
  • 37. Duncan WW, Hubble MJW, Howell JR, Whitehouse SL, Timperley AJ, Gie GA. Revision of the cemented femoral stem using a cement‐in‐cement technique. J Bone Joint Surg, 2009, 91‐B: 577–582. [DOI] [PubMed] [Google Scholar]
  • 38. Saville PA, Kadam AB, Smith HE, Arlet V. Anterior hyperlordotic cages: early experience and radiographic results. J Neurosurg Spine, 2016, 25: 713–719. [DOI] [PubMed] [Google Scholar]
  • 39. Hosseini P, Mundis GM Jr, Eastlack RK, et al Preliminary results of anterior lumbar interbody fusion, anterior column realignment for the treatment of sagittal malalignment. Neurosurg Focus, 2017, 43: E6. [DOI] [PubMed] [Google Scholar]
  • 40. Murray G, Beckman J, Bach K, Smith DA, Dakwar E, Uribe JS. Complications and neurological deficits following minimally invasive anterior column release for adult spinal deformity: a retrospective study. Eur Spine J, 2015, 24: 397–404. [DOI] [PubMed] [Google Scholar]
  • 41. Pimenta L, Fortti F, Oliveira L, et al Anterior column realignment following lateral interbody fusion for sagittal deformity correction. Eur J Orthop Surg Traumatol, 2015, 25: S29–S33. [DOI] [PubMed] [Google Scholar]
  • 42. Wu ZX, Gong FT, Liu L, et al A comparative study on screw loosening in osteoporotic lumbar spine fusion between expandable and conventional pedicle screws. Arch Orthop Trauma Surg, 2012, 132: 471–476. [DOI] [PubMed] [Google Scholar]
  • 43. Vishnubhotla S, McGarry WB, Mahar AT, Gelb DE. A titanium expandable pedicle screw improves initial pullout strength as compared with standard pedicle screws. Spine J, 2011, 11: 777–781. [DOI] [PubMed] [Google Scholar]
  • 44. Kiyak G, Balikci T, Heydar AM, Bezer M. Comparison of the pullout strength of different pedicle screw designs and augmentation techniques in an osteoporotic bone model. Asian Spine J, 2018, 12: 3–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Aycan MF, Yaman ME, Usta Y, Demir T, Tolunay T. Investigation of toggling effect on pullout performance of pedicle screws. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2018, 232: 395–402. [DOI] [PubMed] [Google Scholar]
  • 46. Phan K, Nazareth A, Hussain AK, et al Relationship between sagittal balance and adjacent segment disease in surgical treatment of degenerative lumbar spine disease: meta‐analysis and implications for choice of fusion technique. Eur Spine J, 2018, 27: 1981–1991. [DOI] [PubMed] [Google Scholar]

Articles from Orthopaedic Surgery are provided here courtesy of Wiley

RESOURCES