Dual-Axis π-Pulse Magnetometer with Suppressed Spin-Exchange Relaxation

Elena Zhivun 1,*, Michael Bulatowicz 1, Alexander Hryciuk 1, Thad Walker 1
PMCID: PMC6604831  NIHMSID: NIHMS1034551  PMID: 31276011

Abstract

We present a spin-exchange relaxation-free vector magnetometer with suppressed 1/f probe noise, achieved by applying a small dc bias field and a comb of magnetic dc π pulses along the pump direction. This results in a synchronous orthogonal ac response for each of its two sensitive axes. The magnetometer is particularly well suited to applications such as biomagnetism in which the signal to be measured carries a dominant component of its power at low frequencies. The magnetometer reaches a technical noise floor of 8.4 fT Hz-1/2 (x^) and 11 fT Hz-1/2 (y^) at 0.01 Hz. A single-axis dc spin-exchange relaxation-free (SERF) magnetometer sharing the same experimental apparatus attains 61 fT Hz−1/2 at the same frequency. A noise minimum of 1.1 fT Hz-1/2 (x^) and 2.0 fT Hz-1/2 (y^) is reached by the magnetometer at10 Hz, compared to 0.7 fT Hz−1/2 at 25 Hz for a dc SERF magnetometer.

I. INTRODUCTION

Precision measurement of weak magnetic fields can yield important information that is not obtainable by other methods. The residual magnetization of geological samples reveals the Earth’s magnetic field history and the formation and movement of the continents, and provides means to verify geophysical theories [1,2]. The magnetic fields generated by electrical signals in the human body are used in both research and clinical diagnosis. Fetal magnetocardiography (fMCG), for example, is an important tool for diagnosing arrhythmia in a developing fetus in utero [3]. Electric fetal heart signals are attenuated and distorted by the surrounding tissue and vernix caseosa (a waxy substance covering the fetus), making electrocardiography (ECG) challenging [4]. Similarly, magnetoencephalography (MEG), used to detect and localize brain responses to external stimuli or to diagnose and localize pathological activity [5,6], offers better source localization and complementary information to electroencephalography (EEG).

While superconducting quantum-interference devices (SQUIDs) have been an established state-of-the-art tool for these applications, optical atomic magnetometers are becoming a viable alternative. These magnetometers are compact, reach similar magnetic sensitivity levels (approximately 1 fT Hz−1/2) [7,8], and do not require liquid helium or a large magnetically shielded room, substantially reducing the cost of operation and potentially making high-sensitivity magnetometers more accessible in the future. Optical atomic magnetometers have been employed in high-sensitivity measurements of remnant rock magnetization as a function of temperature [9], brain auditory response [1012], multichannel MEG [13,14], and fMCG signal measurements that are competitive with SQUIDs [15,16].

Spin-exchange relaxation-free (SERF) magnetometers [17] are a subtype of optical atomic magnetometers that exhibit exceptional sensitivity (record of 0.16 fT Hz−1/2) [9], making them particularly attractive for fMCG. While traditional dc SERF magnetometers suffer from 1/f noise that can dominate the fMCG signal, this can be mitigated by adding an external magnetic field modulation, which facilitates signal detection at a higher frequency [18]. Typically, SERF magnetometers measure a single field vector component orthogonal to the optical pumping (z^) and probing (x^) axes [Fig. 1(a)]. Other field components can be measured by adiabatically modulating the magnetic field at the cost of drastically reducing the bandwidth [19]. An alternate approach (the “Z mode”) is to modulate the field along z^ at a frequency fmod outside of the magnetometer’s bandwidth [20]. Demodulation of the signal at fmod provides an independent measurement of the field along x^, while the y^ component is detected either at dc with gain comparable to the x^ component or at 2fmod with a substantially reduced gain. In our system, better fMCG measurements [21] are achieved through both diffusive suppression of the ac Stark shifts [22] and detection of the y^ component at dc. However, this renders the y^ field measurement prone to 1/f technical noise, degrading the y^ sensitivity at low frequencies. While it is possible to circumvent this issue by introducing another probe beam along the y^ axis [23,24], this requires three orthogonal optical axes and increases the complexity of each individual sensor.

FIG. 1.

FIG. 1.

(a) The experimental setup: LP, linear polarizer; WP, Wollaston prism; CL, condenser lens; PD, differential polarimeter; DAQ, data acquisition system. The heaters, field coils, and magnetic shielding are not shown. (b) The π-pulse magnetometer PD response to constant positive x^ and y^ fields and the corresponding demodulation waveforms (illustration).

We present here a method for measuring both x^ and y^ field components using synchronous detection with a single probe, while retaining high sensitivity and spin-exchange relaxation suppression. This is achieved by applying a superposition of a dc offset field B0 and a comb of π-pulses Bπ to the sensor in the z^ direction. The signals produced by both the x^ and y^ magnetic field components are periodic at the π-pulse frequency fπ and orthogonal to each other. The signal demodulation for each axis is performed in real time via multiplying the probe polarization rotation signal by an appropriately phased square wave at the π-pulse frequency, followed by low-pass filtering. This approach suppresses 1/f technical noise along the y^ axis in the π-pulse magnetometer, as compared to the dc SERF and the Z-mode magnetometers. The technical noise limit reached by the π-pulse magnetometer is comparable to that of the Z-mode magnetometer, while the dc SERF magnetometer attains a lower noise limit due to the higher Faraday rotation gain.

II. THEORY

Consider a spin ensemble in the magnetic field B = B0+ B + Bπ, where B=Bxx^+Byy^ is the field to be measured and B0 is the offset field in the z^ direction, which is parallel to the pump wave vector kpmp. A comb of short π-pulses Bπ, also parallel to kpmp, has repetition rate fπ. Here, π-pulse is defined as a magnetic field pulse causing the atomic spin vector S to undergo Larmor precession by the angle π around z^. Let Ω+ = γB+ and Ω0 = γB0 be the corresponding precession rates in the constant magnetic field, where γ is the gyromagnetic ratio and B+ = Bx + iBy. The Bloch equation in the spherical basis for the S+ = Sx + iSy component is

dS+dt=[Γ+i(πfπ+dϕπdt+Ω0)]S+iΩ+Sz, (1)

where Γ is the spin relaxation rate, γ Bπ = π/dt, and ϕπ is defined as ϕπ ≡ −π [fπt (mod 1)]. After the transformation S+ = A+eiϕπ, Eq. (1) is simplified:

dA+dt=[Γ+i(πfπ+Ω0)]A+iΩ+Szeiϕπ. (2)

Since eiϕπ is a periodic function, the resonance condition can be found by substituting A+=pA(+,p)eiptωπ, eiϕπ=pjpeiptωπ in the steady state:

A(+,p)=iΩ+jpSzΓi(πfπpωπ+Ω0), (3)

where

jp=1/(2fπ)1/(2fπ)exp{iπ[fπt( mod 1)]2πipfπt}fπdt. (4)

When B0 is chosen such that Ω0 = πfπ, the p = 1 term dominates, the B+ field response is maximized and S+ becomes

S+=ΩSz|j1|Γexp(iωπt+iϕπ+iαiπ2), (5)

where α = arg(j1Ω+) and Ω = |Ωx + iΩy|. Calculating j1 = −2i/π, we can find the spin projection on the direction of the probe propagation x^:

Sx=2ΩxSzπΓcos(ωπt+ϕπ)+2ΩySzπΓsin(ωπt+ϕπ). (6)

The probe polarization rotation signal can thus be synchronously detected at Ωπ and the components corresponding to Bx and By are orthogonal. Note that the By signal has a nonzero average and that the square-wave demodulation shown in Fig. 1(b) discards the dc Fourier component of the signal. This suppresses the additive 1/f technical noise but degrades the gain in the By channel by a factor of approximately 2.4 compared to the Bx channel.

Qualitatively, the shapes of the x^ and y^ signals [Fig. 1(b)] can understood as follows. Consider an ensemble of spins initially polarized along z^ in a zero net magnetic field. A small applied field Bx or By generates components of spin polarization along y^ or x^ respectively, as described in Ref. [17]. Now superimpose an additional static field Bz; the spins begin to precess about z^. The addition of an infinitely short π-pulse parallel to Bz after π radians of precession effectively eliminates half of each precession cycle. Thus, a Bx field generates static time-average polarization along y^, which sweeps out an arc about z^ by ±π/2 radians, at the π-pulse repetition rate; likewise, a By field generates a time-average static polarization along x^, which sweeps out an arc about z^ by ±π/2 radians.

In order In order to determine how small variations ΔB0 in the leading field affect the magnetometer performance, we find the modified A(+,p) expression by substituting Ω0 = πfπ + δ, δ = γ B0, δπfπ into Eq. (3):

A(+,p)=iΩ+SzjpΓi(Ω0+πfπ+δpωπ)A(+,p)(1+iδΓ)A(+,p)eiδ/Γ. (7)

Similarly, small variations in the π-pulse area Aπ = π + δ, δπ result in the modified A(+,p) expression:

A(+,p)=iΩ+SzjpΓi[Ω0+(π+δ)fπpωπ]A(+,p)(1+iδfπΓ)A(+,p)eiδfπ/Γ. (8)

Equations (7) and (8) suggest that when the offset field B0 deviates from the resonance condition [Eq. (3)] or the pulse area Aπ deviates from π, the magnetometer’s sensitive axes rotate by the angles Δϕ = γ ΔB0/ Γor Δϕ = γ ΔAπfπ/Γ, correspondingly.

III. THE EXPERIMENTAL SETUP

The experimental setup [Fig. 1(a)] allows for direct comparison between the performance of the dc SERF, the Z-mode SERF, and the π-pulse SERF magnetometers. We independently optimize each magnetometer’s parameters and find that all three have the largest optical gain at the same laser tuning and power.

The core of the setup is a rectangular vapor cell (10 × 10 × 30 mm, 87Rb + 165 Torr N2) with two clear optical axes for the pump and probe beams, which propagate through the short dimensions of the cell. The cell is enclosed by a set of high-resistance ceramic heaters with counterpropagating wire traces in order to minimize stray magnetic fields, similar to Ref. [25]. Heat-insulating padding made from 10-mm-thick aerogel sheets is placed between the heater assembly and a 3D-printed plastic housing. The housing also provides frames for rectangular Bx and By field coils (37 mm × 32 mm, Δx = 38 mm). The ac fields for the Z-mode SERF, B0, and π-pulses are created with a larger auxiliary coil system to improve the field uniformity (see the Appendix). The vapor cell is heated to approximately 175 °C with ac at 401.5 kHz, chosen to minimize aliasing of the interference from the current in the heating elements into the demodulated signal [26].

The linearly polarized probe beam (780 nm, 800 μW, x^) is delivered into a four-layer μ-metal shield via a polarizing single-mode fiber (IXfiber λ = 780 nm, Ø125 μmcore). The optical frequency is tuned to the blue side of the D2 line and adjusted to maximize the magnetometer response. After the fiber, the probe polarization is additionally cleaned up with an absorptive linear polarizer (LP) and the residual birefringence in the vapor cell walls is compensated with a λ/4 wave plate. The probe polarization rotation is measured with a balanced differential polarimeter consisting of a Wollaston prism (WP), a condenser lens (CL), and a matched photodiode pair (PD). A differential current amplifier [27] converts the PD difference current into voltage, which is then acquired and demodulated by the data-acquisition system (DAQ) in real time.

The pump beam (23 mW, 795 nm, z^) is delivered into the magnetic shield via the same type of fiber as the probe beam and is circularly polarized before entering the vapor cell. The optical frequency is locked on the red side of the D1 line and fine-tuned to maximize the magnetometer response. Light-shift gradients are minimized by operating the magnetometer in the diffusive SERF regime [22], with high light intensity within the pump beam (w0 = 0.3 cm) to ensure that the pumped atoms remain primarily polarized along z^. The suppression of ac-Stark-shift gradients is particularly important in the π-pulse magnetometer setup, as they cause nonuniformity of Ω0 across the cell volume. This broadens the magnetic resonance [Eq. (3)], reduces the response amplitude, and introduces transients into the signal as the atoms precess out of phase with each other. Atoms diffusing outside of the pump beam dominate the magnetic signal, unaffected by the light shifts and broadening. With a relaxation rate Γ = 435 s−1 limited by Rb-Rb spin-destruction collisions and a diffusion-coefficient estimate of D = 7 mm2/s, the atoms traverse Λ=2πD/Γ=3 mm before being depolarized.

The pump laser power and frequency are stabilized using two proportional-integral-derivative (PID) controllers implemented in the magnetometer field-programmable gate array (FPGA) code, thus ensuring that the feedback is synchronous with the magnetic data acquisition. The uncoated front surface of the magnetometer cell serves as a power pickoff, enabling monitoring of the pump-power noise immediately before the cell. The pump power is measured via a ceramic photodiode placed inside the magnetic shields, with the error signal fed back to a liquid-crystal modulator (Meadowlark Optics D3060HV), stabilizing the pickoff light power. The pump laser frequency is locked to a feature in the transmission signal of an auxiliary vacuum saturated-absorption spectroscopy cell, which contains natural-abundance rubidium. The photocurrents of both the saturated absorption system and the pump-power pickoff are amplified with SRS 570 current-to-voltage converters.

The polarimeter signal is digitized by a 16-bit ADC at 500 kilosamples per second synchronously with the π-pulse control signal produced by the FPGA (NI-7851R). The demodulation is performed in real time by multiplying the ADC data with a square wave [Fig. 1(b)]. The demodulated raw Bx and By data are streamed at the rates of 1 kilosample per second (Z mode) or 500 Hz (π-pulse) per channel to the host computer, where they are converted into magnetic field units.

IV. NOISE ANALYSIS

We determine the sensitivity of the π-pulse magnetometer to B0 variations by applying a low-frequency sinusoidal magnetic field along Bz and measuring the response in Bx and By channels of the π-pulse magnetometer. The corresponding cross-talk coefficients are Bx/Bz = 6 × 10−3 and By/Bz = 27 ×10−3. The discrepancy between the responses is caused by nonorthogonality between By, Bx, and B0, as the coils producing these fields are located on different frames.

The dc magnetic fields in the setup are generated by custom-made current supplies [26]. Based on the noise-density measurements at 0.1 Hz, 1 Hz, and 30 Hz, we extrapolate the 1/f noise to lower frequencies. We estimate the added noise due to B0 current drifts at 0.01 Hz in B0, Bx, and By to be 292 fT Hz−1/2, 1.8 fT Hz−1/2, and 7.9 fT Hz−1/2, correspondingly. Similarly, the estimated 1/f noise in the Bx and By supplies is 117 fT Hz−1/2 at 0.01 Hz. Although the Bx and By drifts exceed the technical noise floor at frequencies below 10 Hz, it is still possible to achieve high magnetic field sensitivity either by reducing the dynamic range of an individual sensor or by employing an array of sensors sharing the bias field compensation, along with a low-current gradient compensation field.

The π-pulses are generated with a home-made half-H-bridge circuit, described in the Appendix. To the leading order, the pulse area is proportional to the square of the pulse time and to the coil power supply voltage. The fractional noise of the coil supply voltage measures 3 × 10−9 Hz−1/2 at 1 Hz. The timing jitter tj ≈ 88 ps root-mean-square (rms) (250 ps peak to peak, f0 = 0 40 MHz) on the π-pulse duration tπ = 4.675 μs (nominal) generates a fractional noise Ntπ:

Ntπ=(tπ+tj)2tπfπ2tjtπtπfπ=1.7×106Hz1/2. (9)

Over the time scale of the measurements (100 – 400 s), Ntπ exceeds the typical center-frequency drift in a quartz oscillator (approximately 10−8) [28]. The π-pulse area noise is thus dominated by the short-term phase noise of the FPGA clock. Noting that the π-pulses generate the same total precession as B0, we estimate the equivalent Bz noise density induced by the π-pulse duration instability as Ntπ×B0=75fT Hz1/2. Based on the cross-talk coefficients, the induced Bx and By noise is 0.4 fT Hz−1/2 and, correspondingly, 2.0 fT Hz−1/2.

V. RESULTS

We begin by implementing a dc SERF magnetometer in order to provide a performance baseline for our experimental setup. The noise spectral density of a dc SERF magnetometer optimized for the best technical noise performance is presented in Fig. 2. Each noise trace is created by averaging the spectra of several 100-s-long samples. The technical noise (blue) is calculated by adding the noise contributions from the probe, pump-power, and pump-frequency fluctuations in quadrature. The photon-shot noise (green dashed line) is calculated theoretically and the magnetic noise (red) is the measured magnetic field noise in the setup. The lowest magnetic noise measured in this setup is 10 fT Hz−1/2, limited by the Johnson noise of the magnetic shield. The technical noise limit of the setup approaches the photon-shot noise limit at frequencies above 25 Hz, attaining the minimum of 0.7 fT Hz−1/2. Although the photon-shot noise decreases further below this frequency, the magnetometer sensitivity still degrades due to the technical noise increase.

FIG. 2.

FIG. 2.

The dc SERF magnetic (red), technical (blue), and calculated photon-shot noise (green dashed line).

The optical gains of the dc SERF, Z-mode SERF, and π-pulse magnetometers are presented in Fig. 3. The gain is measured by applying a known magnetic field and measuring the optical rotation signal as a function of the frequency [21]. In the π-pulse magnetometer, B0 = 44 nT and the π-pulses jointly generate full precession cycles at fπ = 500 Hz. The gyromagnetic ratio is therefore γ = 5.56 Hz/nT, which corresponds to a polarization of p = 0.29 in the spin-temperature limit [29]. In the Z-mode SERF magnetometer, the modulation amplitude (25 nT) is selected to maximize the Bx sensitivity, while the modulation frequency matches the π-pulse repetition rate.

FIG. 3.

FIG. 3.

The dc SERF (red), Z-mode SERF (green), and π-pulse (blue) magnetometer gains.

The technical noise comparison for the Z-mode and π-pulse-mode magnetometers is presented in Fig. 4. In contrast to the dc SERF magnetometer, the Z-mode and π-pulse magnetometers have an improved low-frequency noise performance, except for the y^ direction in Z mode [Fig. 4(b)], which does not benefit from the added modulation. In addition, we assess the stability of each magnetometer using the Allan deviation (Fig. 5), computed from a set of 400-s-long data samples. This provides an estimate of the magnetometers’ performance when they are operated as a sensor array in a closed feedback loop. The π-pulse magnetometer readout can be averaged for up to 10 s to attain a technical noise floor of 2.2 fT (x^) and 4.0 fT (y^), which is within the stability requirements of operating an fMCG array. The Z-mode y^ and dc SERF signals exhibit drifts at time scales above 0.1 s, while the Z-mode x^ signal has the lowest drift and is dominated by the sensor noise up to at least 40 s of integration time.

FIG. 4.

FIG. 4.

A comparison of the π-pulse (blue) and Z-mode (green) magnetometers. The dashed lines represent corresponding photon-shot noise limits. The magnetic noise (red) is measured with the Z-mode magnetometer. (a) x^ axis; (b) y^ axis.

FIG. 5.

FIG. 5.

The Allan deviation of the dc SERF (red), Z-mode (green), and π-pulse (blue) magnetometers.

The detailed technical noise composition of each magnetometer is presented in Figs. 6 (dc SERF), 7 (Z-mode SERF), and 8 (π-pulse). On these plots, the magnetic noise trace (red) shows fluctuations of the real magnetic field during the measurements. The probe noise trace (blue) is the magnetic sensitivity limit imposed by the optical detection scheme. It is measured by recording the magnetometer signal while blocking the pump laser beam. The probe photon-shot noise (dashed green line) is calculated as ρ = 4eIpd, where e is the electron charge and Ipd is the current through a single photodiode of the balanced polarimeter. The electronic noise trace (black) is the sensitivity limit due to the electronic noise in the front-end amplifier, combined with the data-acquisition and demodulation process. The electronic noise is measured by recording the magnetometer signal with both the pump and probe laser beams blocked. The pump-power (yellow) and frequency (pink) noise traces are calibrated by sequentially applying a sinusoidal (f = 23 Hz) modulation to each corresponding PID set point. The power and frequency monitor readouts are captured simultaneously with the magnetometer signals. By comparing the peak amplitudes at the calibration frequency in the magnetic signal and the readouts, we calibrate the readout signals into the magnetic field units.

FIG. 6.

FIG. 6.

The dc SERF magnetometer noise density.

FIG. 7.

FIG. 7.

The Z-mode magnetometer noise density: (a) y^ axis; (b) x^ axis.

FIG. 8.

FIG. 8.

The π-pulse magnetometer noise density: (a) y^ axis; (b) x^ axis.

Although the peak π-pulse field amplitude (16 μT) is multiple orders of magnitude larger than the intended device sensitivity, it is still reachable, since the magnetic field only determines the instantaneous precession frequency of the atoms, while the magnetometer measures their phase. The π-pulse magnetometer can operate in the SERF regime provided that the π-pulse duration is short compared to the time between spin-exchange collisions and the Larmor precession rate in B0 is less than the spin-exchange collision rate (approximately 0.3 × 106 1/s).

The effect of spin-exchange collisions occurring during the π-pulse manifests as an additional spin-relaxation mechanism, resulting in a decreased magnetic-response magnitude and an increased bandwidth. In Fig. 9, we measure the magnetometer response while varying the π-pulse duty cycle between 0.25% and 50%. During the measurements, both the pulse duration and the amplitude are adjusted to maintain the pulse-induced precession at π, while keeping B0 constant. Although the π-pulse duration increase adversely affects the magnetometer response, it only degrades by a factor of 2 as the duty cycle increases by an order of magnitude. Even in the extreme case of 50% duty cycle, the Bx and By responses are still measurable. In this experiment, the minimum duration of the π-pulse is limited by the largest π-pulse coil voltage that can be applied without damaging the electronics.

FIG. 9.

FIG. 9.

The effects of the π-pulse duty cycle on the magnetometer’s response: (a) the magnetic response amplitude; (b) the magnetic response bandwidth.

VI. CONCLUSION

In this paper, we demonstrate a dual-axis spin-exchange relaxation-free magnetometer and measure its technical noise. While, overall, it has lower gain compared to the traditional dc SERF magnetometer, it offers the advantage of enabling synchronous detection in two directions simultaneously. The x^ and y^ magnetic field signals are generated the chosen π-pulse frequency and are orthogonal, minimizing the technical 1/f noise contribution from the laser and temperature drifts in both signals simultaneously. This is especially important in biomagnetic applications, since a significant fraction of the signal power is contained in the 0.1–100 Hz frequency range, where the measurement sensitivity is often limited by the 1/f noise in the detection system [26]. This also opens up a possibility of precise gradiometry measurements with several independent sensors, as the improved long-term stability can be used for field stabilization and better cancellation of the environmental noise. Our next goal is to implement a π-pulse magnetometer array for fMCG and perform a direct low-frequency low-noise magnetic-field-gradient measurement.

ACKNOWLEDGMENTS

This work was supported by Grant No. R01HD057965–05 awarded by the National Institutes of Health, as well as Grant No. PHY-1607439 by the National Science Foundation.

APPENDIX

The offset field and the π-pulse coils are wrapped on top of each other on the auxiliary coil frame (Fig. 10) and consist of two square coils (L = 119 mm), N = 10 wraps each, separated by Δx = 149 mm. The π-pulse coils have an additional compensation coil set (L = 79 mm, Δx = 101 mm, N = 3), wrapped in the opposite direction to the primary coils. The π-pulse coil geometry is designed to minimize the dBz/dz field gradients along the pump axis. The calculated current-to-field conversion coefficients are β = 31 nT/mA for the π-pulse and β = 55 nT/mA for the offset field coil.

FIG. 10.

FIG. 10.

The auxiliary coil frame housing the π-pulse and the offset field coils.

The π-pulse control signals are generated by the NI-7851R FPGA and converted to current pulses via a custom half-H-bridge circuit (Fig. 11). The power rail V_pulse controls the pulse amplitude; it is connected to a low-noise HP6205C dc power supply. During the π-pulse, transistors Q1 and Q2 are switched “on,” connecting V_pulse and ground (GND) to the coil leads. The current through the π-pulse coils increases approximately linearly, with a slope determined by V_pulse. Capacitors C1 and C2 provide an additional low-impedance source for the pulse current and a sink for the return current, helping to minimize the dynamic loading of the power supply. Schottky diodes with zero reverse-recovery time and minimal parasitic capacitance are chosen for this application, in order to reduce undesirable current oscillations through the π-pulse coil. Further suppression of the current oscillation is achieved by activating a supplementary ringing suppression gate Q3 shortly after completion of the main pulse, when the current through the π-pulse coil reaches zero. This shunts any residual energy stored in the parasitic capacitances of the circuit and the magnetic field to ground through resistor R1. This prevents the oscillations of the residual energy between the π-pulse coil and the parasitic capacitances that would otherwise have occurred. The value of resistor R1=L/CQ3 is chosen to balance the energy-dissipation rate against oscillations of the current during the π-pulse. The optimal timing and duration of the Q3 control signal are determined by connecting high-impedance scope probes to the coil leads and verifying that the ringing after the pulse is minimized. The π-pulse control signals are buffered with metal-oxide semiconductor field-effect transistor (MOSFET) drivers (MIC4420, not shown), which are powered using the HPE3620A dc power supply. Optionally, the control signals may be buffered with a bridge driver to increase the maximum allowed V_pulse.

FIG. 11.

FIG. 11.

A schematic of the pulse circuit (the actual pulse circuit is a full H-bridge; in this experiment, only one half is used). The MOSFET drivers (MIC4420) and the corresponding power rail are not shown. πin, Main pulse control signal; Rin, ringing suppression gate control signal; L+ and L−, coil leads; V_pulse, pulse power rail; GND, ground.

To assess the magnetic field pulse shape, we connect a 1 Ω resistor in series with the π-pulse coil and measure the voltage drop across the resistor throughout the pulse. With V_pulse = 14 V, the current rises linearly from 0 A to 0.5A over 4.7 μs during the pulse active phase and drops back to 0 A over 2.7 μs after the pulse completion. This provides an estimate for the coil inductance L = 130 μH, and the peak magnetic field B = 16 μT. With γ = 5.6 Hz/nT, we can calculate the pulse area A = 2.1 rad, which is within the order of magnitude of π.

References

  • [1].Courtillot Vincent and Le Mouël Jean-Louis, The study of Earth’s magnetism (1269–1950): A foundation by Peregrinus and subsequent development of geomagnetism and paleomagnetism, Rev. Geophys 45, 11 (2007). [Google Scholar]
  • [2].Verosub Kenneth L. and Roberts Andrew P., Environmental magnetism: Past, present, and future, J. Geophys. Res 100, 2175 (1995). [Google Scholar]
  • [3].Strasburger Janette F., Cheulkar Bageshree, and Wakai Ronald T., Magnetocardiography for fetal arrhythmias, Heart Rhythm 5, 1073 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Wakai RT, Lengle JM, and Leuthold AC, Transmission of electric and magnetic foetal cardiac signals in a case of ectopia cordis: The dominant role of the vernix caseosa, Phys. Med. Biol 45, 1989 (2000). [DOI] [PubMed] [Google Scholar]
  • [5].Malmivuo Jaakko and Plonsey Robert, Bioelectromagnetism: Principles and Applications of Bioelectric and Biomagnetic Fields (Oxford University Press, New York, 1995). [Google Scholar]
  • [6].Yoshinaga Harumi, Nakahori Tomoyuki, Ohtsuka Yoko, Oka Eiji, Kitamura Yoshihiro, Kiriyama Hideki, Kinugasa Kazumasa, Miyamoto Keiichi, and Hoshida Toru, Benefit of simultaneous recording of EEG and MEG in dipole localization, Epilepsia 43, 924 (2002). [DOI] [PubMed] [Google Scholar]
  • [7].Clarke J and Braginski AI, The SQUID Handbook (Wiley-VCH, Weinheim, 2004). [Google Scholar]
  • [8].Budker Dmitry and Jackson Kimball Derek F., Optical Magnetometry (Cambridge University Press, New York, 2013). [Google Scholar]
  • [9].Dang HB, Maloof AC, and Romalis MV, Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer, Appl. Phys. Lett 97, 151110 (2010). [Google Scholar]
  • [10].Xia H, Ben-Amar Baranga A, Hoffman D, and Romalis MV, Magnetoencephalography with an atomic magnetometer, Appl. Phys. Lett 89, 211104 (2006). [Google Scholar]
  • [11].Sander TH, Preusser J, Mhaskar R, Kitching J, Trahms L, and Knappe S, Magnetoencephalography with a chip-scale atomic magnetometer, Biomed. Opt. Express 3, 981 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Sheng Jingwei, Wan Shuangai, Sun Yifan, Dou Rongshe, Guo Yuhao, Wei Kequan, He Kaiyan, Qin Jie, and Gao Jia-Hong, Magnetoencephalography with a Cs-based high-sensitivity compact atomic magnetometer, Rev. Sci. Instrum 88, 094304 (2017). [DOI] [PubMed] [Google Scholar]
  • [13].Borna Amir, Carter Tony R., Goldberg Josh D., Colombo Anthony P., Jau Yuan-Yu, Berry Christopher, McKay Jim, Stephen Julia, Weisend Michael, and Schwindt Peter D. D., A 20-channel magnetoencephalography system based on optically pumped magnetometers, Phys. Med. Biol 62, 8909 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [14].Johnson Cort N., Schwindt PDD, and Weisend M, Multi-sensor magnetoencephalography with atomic magnetometers, Phys. Med. Biol 58, 6065 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15].Wyllie Robert, Kauer Matthew, Wakai Ronald T., and Walker Thad G., Optical magnetometer array for fetal magneto-cardiography, Opt. Lett 37, 2247 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Alem Orang, Sander Tilmann H., Mhaskar Rahul, John LeBlanc Hari Eswaran, Steinhoff Uwe, Okada Yoshio, Kitching John, Trahms Lutz, and Knappe Svenja, Fetal magnetocardiography measurements with an array of microfabricated optically pumped magnetometers, Phys. Med. Biol 60, 4797 (2015). [DOI] [PubMed] [Google Scholar]
  • [17].Allred JC, Lyman RN, Kornack TW, and Romalis MV, High-Sensitivity Atomic Magnetometer Unaffected by Spin-Exchange Relaxation, Phys. Rev. Lett 89, 130801 (2002). [DOI] [PubMed] [Google Scholar]
  • [18].Shah V and Romalis MV, Spin-exchange relaxation-free magnetometry using elliptically polarized light, Phys. Rev. A 80, 013416 (2009). [Google Scholar]
  • [19].Seltzer SJ and Romalis MV, Unshielded three-axis vector operation of a spin-exchange-relaxation-free atomic magnetometer, Appl. Phys. Lett 85, 4804 (2004). [Google Scholar]
  • [20].Li Zhimin, Wakai Ronald T., and Walker Thad G., Parametric modulation of an atomic magnetometer, Appl. Phys. Lett 89, 134105 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].DeLand Zachary J., Ph.D. thesis, University of Wisconsin, 2017. [Google Scholar]
  • [22].Sulai IA, Wyllie R, Kauer M, Smetana GS, Wakai RT, and Walker TG, Diffusive suppression of ac-Stark shifts in atomic magnetometers, Opt. Lett 38, 974 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [23].Li Rujie, Quan Wei, Fan Wenfeng, Xing Li, Wang Zhuo, Zhai Yueyang, and Fang Jiancheng, A dual-axis, high-sensitivity atomic magnetometer, Chin. Phys. B 26, 120702 (2017). [Google Scholar]
  • [24].Seltzer Scott Jeffrey, Ph.D. thesis, Princeton University, 2008. [Google Scholar]
  • [25].Bulatowicz MD, Temperature system with magnetic field suppression, U.S. Patent No. 8,138,760 (2012).
  • [26].Wyllie Robert IV, Ph.D. thesis, University of Wisconsin, 2012. [Google Scholar]
  • [27].Hobbs Philip C. D., Building Electro-Optical Systems: Making It All Work, 2nd ed (Wiley, Hoboken, N.J, 2009). [Google Scholar]
  • [28].Lombardi M, The accuracy and stability of quartz watches, Horological J. 150, 57 (2008). [Google Scholar]
  • [29].Happer W and Tang H, Spin-Exchange Shift and Narrowing of Magnetic Resonance Lines in Optically Pumped Alkali Vapors, Phys. Rev. Lett 31, 273 (1973). [Google Scholar]

RESOURCES