Skip to main content
Investigative and Clinical Urology logoLink to Investigative and Clinical Urology
. 2019 Jun 25;60(4):326–330. doi: 10.4111/icu.2019.60.4.326

Pure single-site robot-assisted pyeloplasty with the da Vinci SP surgical system: Initial experience

Ji Eun Heo 1, Sung Ku Kang 1, Dong Hoon Koh 2, Joon Chae Na 1, Yong Seung Lee 1, Woong Kyu Han 1, Young Deuk Choi 1, Won Sik Jang 1,
PMCID: PMC6607072  PMID: 31294143

Abstract

Purpose

Laparoendoscopic single-site surgery (LESS) reduces the limited invasiveness of conventional laparoscopy while providing superior cosmetic results. However, LESS remains a challenging surgical technique, even in robotic surgery, primarily due to the lack of triangulation and limited instrument movement. The da Vinci SP surgical system (Intuitive Surgical) was recently introduced to overcome these limitations. We describe our initial experience with pure single-site robot-assisted pyeloplasty (RAP) for ureteropelvic junction obstruction (UPJO) using the da Vinci SP surgical system.

Materials and Methods

Three consecutive patients who were diagnosed with UPJO underwent RAP with the da Vinci SP surgical system from December 2018 to February 2019 at our institution. The surgical technique involved reproducing the steps of multi-port RAP. A 30-mm umbilical incision was made and the GelPOINT was inserted. The multichannel robotic port and the assistant's port were placed through the GelSeal cap. In all patients, Anderson–Hynes dismembered pyeloplasty was performed. The ureteral double J stent was inserted antegrade, and the drain was not placed.

Results

The procedures were successfully completed using a pure single-site approach. There was no need for additional port placement or conversion to laparoscopic or open surgery. Total operative time in the three patients was 139, 180, and 213 minutes, respectively. No intraoperative complications occurred, and blood loss was minimal. The postoperative course of all patients was uneventful with no complications greater than Clavien–Dindo grade I surgical complications.

Conclusions

Pure single-site RAP using the da Vinci SP surgical system is feasible and safe.

Keywords: Laparoscopy, Robotic surgical procedures, Ureteral obstruction

INTRODUCTION

Laparoendoscopic single-site surgery (LESS) is a laparoscopic procedure performed through a single skin incision. This minimally invasive surgery was developed to minimize port-related complications, recovery time, and postoperative pain and to achieve superior cosmetic results [1,2]. Although LESS is feasible for several urologic procedures [3], the technique is highly challenging even for expert laparoscopic surgeons.

Since Kaouk et al. [4] reported the first series of single-incision surgeries for the upper urinary tract using the da Vinci surgical system (Intuitive Surgical, Sunnyvale, CA, USA), the use of robotic single-site surgery for urologic problems has been globally reported [5,6]. However, intracorporeal suturing and traction are difficult and technically time-consuming because of clashing of the external robotic instruments and the loss of triangulation [7,8]. These challenges are occasionally overcome by the additional placement of an assistant port [9].

The da Vinci SP surgical system was developed as a novel robotic platform for successfully performing “pure” robotic single-site surgery while overcoming the aforementioned difficulties [10]. The system recently received clearance from the Food and Drug Administration [11]. The da Vinci SP surgical system was introduced for the first time in Korea at the end of 2018.

Here, we describe our experience with our first three cases of robot-assisted pyeloplasty (RAP) performed using the da Vinci SP surgical system in Severance Hospital. We report on the technical feasibility and perioperative outcomes. To the best of our knowledge, this is the first case series on RAP performed using the da Vinci SP surgical system in Asia.

MATERIALS AND METHODS

1. Patients

Three consecutive patients diagnosed with ureteropelvic junction obstruction (UPJO) underwent RAP performed by use of the da Vinci SP surgical system from December 2018 to February 2019. The surgeries were done by a single surgeon (W.S.J) who had performed eight cases of conventional RAP in a recent year at our institution, which is a high-volume robotic center. All patients reported flank pain on the affected side due to UPJO and signed an informed consent after the novel robotic system had been explained to them. This study was approved by the Institutional Review Board (IRB) of Severance Hospital (IRB approval number: 1-2019-0003).

2. Surgical procedure

RAP was performed with the da Vinci SP surgical system by using the dismembered technique with a transperitoneal approach. Patients were positioned in a semilateral position. A 30-mm umbilical incision was made to reach the peritoneal cavity, and the GelPOINT advanced access platform (Applied Medical, Rancho Santa Margarita, CA, USA) was inserted (Fig. 1). A 25-mm multichannel port containing an articulating robotic camera and two double-jointed robotic instruments and a 12-mm port for an assistant's laparoscopic instrument were placed through the GelSeal cap (Fig. 2).

Fig. 1. Placement of the GelPOINT advanced access platform. A 25-mm multichannel port and a 12-mm assistant port were placed through the GelSeal cap.

Fig. 1

Fig. 2. Placement of the instruments and docked single-port platform.

Fig. 2

The surgical technique involved reproducing the steps of our own multi-port RAP. An incision was made along the white line of Toldt, and the colon was mobilized medially to expose Gerota's fascia; an incision was made in the Gerota's fascia to isolate the renal pelvis and ureter. Pyeloplasty was performed using the Anderson–Hynes technique. After excision of the obstructed ureteropelvic junction, the ureter was longitudinally spatulated. Anastomosis was performed on the posterior side with a running 4-0 absorbable, sterile suture and a double J stent was inserted antegrade. Once the stent was correctly placed, anastomosis was completed on the anterior side (Video clip, Supplementary material). A surgical drain was not inserted. Layered wound closure was performed, and the skin was closed with a skin bond.

Preoperative variables, intraoperative surgical outcomes, and immediate postoperative complications were investigated. The Clavien–Dindo classification was used to grade the severity of postoperative complications [12].

RESULTS

The procedures were successfully completed using a pure single-site approach. There was no need for additional port placement or conversion to laparoscopic or open surgery. The patients' demographic characteristics and perioperative data are shown in Table 1. Total operative time in the three patients was 139, 180, and 213 minutes, respectively. The third patient had a large parapelvic cyst and had undergone endopyelotomy with balloon dilation 3 years ago; marsupialization of the parapelvic cyst was performed at the same time, and there was a moderate adhesion near the uteropelvic joint. No intraoperative complications occurred, and blood loss was minimal. The postoperative course of all patients was uneventful with no complications greater than Clavien–Dindo grade I postoperative surgical complications. The cosmetic result was satisfactory (Fig. 3).

Table 1. Patients' characteristics and perioperative data.

Characteristic Case no. 1 Case no. 2 Case no. 3
Age (y) 42 30 71
Sex Male Male Female
Affected side Left Right Left
Body mass index (kg/m2) 22.49 27.08 18.82
Preoperative eGFR (mL/min/m2) 91 108 43
Grade of UPJO High Low High
Operative time (min) 139 180 213
Docking time (min) 5 10 10
Console time (min) 118 145 185
Suture time (min) 29 47 53
Intraoperative complications No No No
Conversion to other surgery No No No
Estimated blood loss (mL) Minimal Minimal 50
Discharge POD 4 POD 3 POD 3
Postoperative eGFR (mL/min/m2) 105 113 81
Postoperative complications No No No
Pain at discharge, NPIS score 2 2 3

eGFR, estimated glomerular filtration rate; UPJO, ureteropelvic junction obstruction; POD, postoperative day; NPIS, numerical pain intensity scale.

Fig. 3. Cosmetic result after skin closure.

Fig. 3

DISCUSSION

We have validated the feasibility of pure single-site RAP performed using the da Vinci SP surgical system. For pyeloplasty, an elongated skin incision is not required to collect the specimen and there are no oncologic risks associated with the procedure; thus, pyeloplasty is an option for suitable single-site surgery. Despite the technical difficulties associated with the procedures, single-site laparoscopic and robotic pyeloplasty have frequently been performed owing to the need for adequate cosmesis [6,7,13]. The main concern in single-site surgery is the loss of triangulation [7,8]; however, technologic advancements have aided in overcoming this problem.

The da Vinci SP surgical system was designed after development of the SP 999 and SP 1098 prototypes [14]. The da Vinci SP surgical system includes three double-jointed wristed instruments and a fully wristed three-dimensional camera. This double-jointed instrument, called the “EndoWrist,” performs an “elbow” movement to provide triangulation. The double-jointed instruments, the camera, and the assistant's 5-mm laparoscopic instrument are placed through a single multichannel port. The single arm enables 360° anatomic access.

Although the EndoWrist instrument provides elbow movement for achieving intracorporeal triangulation, the angle formed on curving the EndoWrist instrument is less than 90°, making proper spatulation of the ureter difficult to perform. Anastomotic suturing is not difficult. Moreover, the assistant experiences difficulty because of the limited working space, which is common with single-site surgery, and good coordination between the surgeon and the assistant is required. The surgeon should move the camera or instruments when the assistant's instrument enters into the operation field for traction or suction. Therefore, an additional assistant port could be inserted in cases of high risk of bleeding or when many assists are required. Although an additional port was not used in the three cases described in this report, we inserted an additional port to use an EndoStapler for robot-assisted radical nephroureterectomy using the da Vinci SP surgical system.

There were some limitations to this study: namely, the small sample size and short follow-up duration. When the cases of conventional RAP performed by the same surgeon (W.S.J.) in a recent 1-year period were reviewed, the mean operative time was 136±53.0 minutes and the mean suture time was 38±9.0 minutes. If the surgeon becomes accustomed to performing surgery using the da Vinci SP surgical system, the operative time or suture time could be shortened by as much as conventional RAP. The functional outcomes of RAP using da Vinci SP surgical system are not yet confirmed because the imaging studies will be performed after removal of the double J stent 3 months after surgery according to our hospital protocol. However, the results of this preliminary study are promising. There were no cases of conversion or complications, which showed the feasibility and safety of pure single-site RAP with the da Vinci SP surgical system. Studies analyzing more patients and long-term outcomes are needed. To confirm our results, the treatment outcomes of RAP performed using da Vinci SP surgical system should be compared with those of conventional RAP or LESS pyeloplasty.

CONCLUSIONS

Pure single-site RAP using the da Vinci SP surgical system is feasible and safe. Further studies involving more patients and long-term outcomes are needed to confirm our results.

Footnotes

CONFLICTS OF INTEREST: The authors have nothing to disclose.

SUPPLEMENTARY MATERIAL

The supplementary video clip can also be accessed by scanning a QR code, or will be available on YouTube: https://youtu.be/8WVgnxFLpAM.

graphic file with name icu-60-326-g004.jpg

Download video file (253.4MB, mp4)

References

  • 1.Gettman MT, Box G, Averch T, Cadeddu JA, Cherullo E, Clayman RV, et al. Consensus statement on natural orifice transluminal endoscopic surgery and single-incision laparoscopic surgery: heralding a new era in urology? Eur Urol. 2008;53:1117–1120. doi: 10.1016/j.eururo.2008.02.006. [DOI] [PubMed] [Google Scholar]
  • 2.Box G, Averch T, Cadeddu J, Cherullo E, Clayman R, Desai M, et al. Urologic NOTES Working Group. Nomenclature of natural orifice translumenal endoscopic surgery (NOTES) and laparoendoscopic single-site surgery (LESS) procedures in urology. J Endourol. 2008;22:2575–2581. doi: 10.1089/end.2008.0471. [DOI] [PubMed] [Google Scholar]
  • 3.Autorino R, Cadeddu JA, Desai MM, Gettman M, Gill IS, Kavoussi LR, et al. Laparoendoscopic single-site and natural orifice transluminal endoscopic surgery in urology: a critical analysis of the literature. Eur Urol. 2011;59:26–45. doi: 10.1016/j.eururo.2010.08.030. [DOI] [PubMed] [Google Scholar]
  • 4.Kaouk JH, Goel RK, Haber GP, Crouzet S, Stein RJ. Robotic single-port transumbilical surgery in humans: initial report. BJU Int. 2009;103:366–369. doi: 10.1111/j.1464-410X.2008.07949.x. [DOI] [PubMed] [Google Scholar]
  • 5.Han WK, Kim DS, Jeon HG, Jeong W, Oh CK, Choi KH, et al. Robot-assisted laparoendoscopic single-site surgery: partial nephrectomy for renal malignancy. Urology. 2011;77:612–616. doi: 10.1016/j.urology.2010.06.067. [DOI] [PubMed] [Google Scholar]
  • 6.Olweny EO, Park SK, Tan YK, Gurbuz C, Cadeddu JA, Best SL. Perioperative comparison of robotic assisted laparoendoscopic single-site (LESS) pyeloplasty versus conventional LESS pyeloplasty. Eur Urol. 2012;61:410–414. doi: 10.1016/j.eururo.2011.10.024. [DOI] [PubMed] [Google Scholar]
  • 7.Best SL, Donnally C, Mir SA, Tracy CR, Raman JD, Cadeddu JA. Complications during the initial experience with laparoendoscopic single-site pyeloplasty. BJU Int. 2011;108:1326–1329. doi: 10.1111/j.1464-410X.2011.10078.x. [DOI] [PubMed] [Google Scholar]
  • 8.Autorino R, Kaouk JH, Stolzenburg JU, Gill IS, Mottrie A, Tewari A, et al. Current status and future directions of robotic single-site surgery: a systematic review. Eur Urol. 2013;63:266–280. doi: 10.1016/j.eururo.2012.08.028. [DOI] [PubMed] [Google Scholar]
  • 9.Seideman CA, Tan YK, Faddegon S, Park SK, Best SL, Cadeddu JA, et al. Robot-assisted laparoendoscopic single-site pyeloplasty: technique using the da Vinci Si robotic platform. J Endourol. 2012;26:971–974. doi: 10.1089/end.2011.0573. [DOI] [PubMed] [Google Scholar]
  • 10.Kaouk J, Garisto J, Eltemamy M, Bertolo R. Pure single-site robot-assisted partial nephrectomy using the SP surgical system: initial clinical experience. Urology. 2019;124:282–285. doi: 10.1016/j.urology.2018.11.024. [DOI] [PubMed] [Google Scholar]
  • 11.Kaouk J, Bertolo R, Eltemamy M, Garisto J. Single-port robot-assisted radical prostatectomy: first clinical experience using the SP surgical system. Urology. 2019;124:309. doi: 10.1016/j.urology.2018.10.025. [DOI] [PubMed] [Google Scholar]
  • 12.Dindo D, Demartines N, Clavien PA. Classification of surgical complications: a new proposal with evaluation in a cohort of 6336 patients and results of a survey. Ann Surg. 2004;240:205–213. doi: 10.1097/01.sla.0000133083.54934.ae. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Sorokin I, Cadeddu JA. Robotic laparoendoscopic single site: pyeloplasty and nephrectomy chapter. J Endourol. 2018;32:S88–S92. doi: 10.1089/end.2017.0724. [DOI] [PubMed] [Google Scholar]
  • 14.Bertolo R, Garisto J, Gettman M, Kaouk J. Novel system for robotic single-port surgery: feasibility and state of the art in urology. Eur Urol Focus. 2018;4:669–673. doi: 10.1016/j.euf.2018.06.004. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Download video file (253.4MB, mp4)

Articles from Investigative and Clinical Urology are provided here courtesy of Korean Urological Association

RESOURCES