Treatment of type 1A endoleak using coil embolization: a case report

Sergio Quilici Belczak 1,2,, Guilherme Delicato Pedroso 1, Lara Cote Ogawa 1, Paula Thume Campos 1, Andre Lopes Padula 1, Glenna Paulain Machado 1, Matheus Zago Soares dos Santos 1, Beatriz Marques Abrão 1
PMCID: PMC6636908  PMID: 31360155

Abstract

In a type 1A endoleak, the endograft is unable to fully seal the proximal aneurysm neck and blood flow leaks between the wall of the aortic neck and the graft material. This article reports a case in which coil embolization was used and presents a literature review (PubMed, LILACS, and SciELO). Searches were run for articles published in the past 5 years using the descriptors “endoleak 1A”, “coil embolization,” and “treatment”. Type 1A endoleak occurs in 1.1% of patients within 30 days of graft placement. Treatment of an endoleak is obligatory and usually consists of sealing the proximal graft neck using stents and balloons to expand the landing zone or to increase the radial force of the graft. Some studies have suggested using embolization techniques with cyanoacrylate, fibrin glue, and Onyx, demonstrating success rates that exceed 97%. However, correction of type 1A endoleak using coil embolization has seldom been described.

Keywords: endoleak, therapeutic embolization, treatment

INTRODUCTION

Endovascular repair of aortic aneurysms (EVAR) has become a common alternative to open surgery because it is a less invasive procedure.1 However, certain complications can occur, including endoleaks. These are defined as persistent filling of the aneurysm sac after endovascular repair and they can increase the risk of rupture due to expansion of the aneurysm sac.

There are five main types of endoleak. Type 1 can be proximal (1A) or distal (1B) of the site of repair. In a type 1A leak, the endoprosthesis does not completely seal the neck of the aneurysm and there is arterial leakage between the wall of the proximal aortic neck and the graft material. In type 1B, leakage occurs between the wall of the distal aortic neck and the graft material. In a type 2 leak, the sac is filled by retrograde flow through vessels. A type 3 leak involves a tear or disconnection of the endoprosthesis. A type 4 leak is usually seen at the time of deployment of the implant in anticoagulated patients, caused by graft porosity. In type 5, the sac expands in the absence of a visible endoleak (endotension).2

Treatment is obligatory in cases of type 1 endoleaks,3 with the objective of improving the seal to the proximal implant, primarily by using stents and balloons to widen the landing zone or increase the radial force of the graft. This report describes a case in which coil embolization was used to repair a type 1A endoleak, which is a technique that has rarely been described according to a review of the literature.

CASE DESCRIPTION

A 72-year-old patient, assessed as in regular general condition, was admitted with abdominal pains. Tomography diagnosed a 6.4 cm abdominal aortic aneurysm with a 1.1 cm proximal neck that was conical, with a 60° angle. There was also an aneurysm involving the right internal iliac, with a diameter of 3.4 cm and no distal neck. This anatomy, with the internal iliac aneurysm extending deep into the pelvis, ruled out any possibility of open surgery. During planning, the decision was taken to use an endoprosthesis long enough to reach the external iliac, since treatment with embolization was necessary because of the large aneurysm of the internal iliac with no distal neck.

The patient was therefore treated using endovascular techniques to implant an Ovation® endoprosthesis (Endologix, Irvine, California). Control angiography showed significant leakage between the endoprosthesis and the proximal aorta neck (a type 1A endoleak). The decision was then taken to use controlled-release coils to seal the area of leakage. Control angiography showed that the endoleak had been sealed once the coils were deployed.

Embolization of the endoleak was accomplished by placing a 5 Fr Simmons® catheter in contact between the endoprosthesis and the aorta and then inserting a Maestro® microcatheter through the first catheter and advancing it up to the site requiring embolization, where six Complex Trufill 3D coils were released. The patient exhibited resolution of the endoleak and was discharged from hospital after 3 days (Figures 1 and 2).

Figure 1. Type 1A endoleak.

Figure 1

Figure 2. Type 1A endoleak after coil embolization.

Figure 2

DISCUSSION

Literature found by searching the PubMed, LILACS, and SciELO databases was reviewed. Articles published during the previous 5 years were identified using the descriptors “endoleak 1A”, “coil embolization”, and “treatment”.

The most common complication associated with endovascular treatment of aneurysms is internal leakage (endoleak), with incidence rates in the range of 6 to 57% reported in the literature. An endoleak is a failure of the implanted stent to exclude the aneurysm sac from the systemic circulation and can cause the aneurysm sac to expand and rupture. Management of endoleaks varies depending on type. Types 1 and 3 require treatment, whereas type 2 is increasingly managed expectantly, especially if there is no expansion of the aneurysm sac.4

The incidence of type 1 endoleaks may be attributable to the surgeon’s surgical skills, but it can also be related to skill at preoperatively sizing the endoprosthesis. This is a difficult task when the aneurysms to be repaired present complex anatomic characteristics, such as a short proximal aneurysm neck, reverse thinning of the neck, mural calcification or thrombus, and accentuated neck angles.5

Not all aortic aneurysms can be resolved satisfactorily. Studies have identified a direct association between short proximal aortic necks and EVAR failure, i.e., presence of endoleaks.6 Special techniques are therefore needed to solve these complex cases.

Although post-EVAR rupture of an aortic aneurysm involves a major risk of loss of life and demands immediate diagnosis and emergency intervention, there are no specific guidelines on the most appropriate management of this situation. Limited data from small case series indicate that these patients are most often treated with surgery. However, the risk of operating on abdominal aortic aneurysms that have ruptured post-EVAR is high and these patients generally have significant comorbidities. Endovascular repair is therefore the first-choice option if the patient’s conditions are favorable.7

Conversion to open surgery generally involves surgical exposure of the aneurysm, proximal and distal vascular control, complete removal of the endoprosthesis and substitution with an aortic prosthesis. Such a complex procedure is associated with significant morbidity and mortality .

Conventional methods for treatment of type 1 leakage include proximal aortic extensions, balloon angioplasty, and stents. Some cases are not suitable for endovascular reintervention and open surgical repair is the only option.4 EVAR with a chimney graft (Ch-EVAR) may be an option in cases with compromised cardiac function,8 while fenestrated stents (FEVAR) can be useful in patients with complex aortic anatomy.9 Successful cases of type 1A endoleak repair have been described in the literature using the Palmaz intraoperative stent.10 Other studies have also reported effective management of 1A endoleaks using Onyx, with or without coils, depending on the number of gutters involved.11

In conclusion, EVAR is increasingly the preferred method rather than open repair in cases with favorable anatomy because it is associated with less morbidity and mortality than open repair. However, there is a greater risk of reintervention because of complications, such as endoleaks. Studies show that complications have a robust relationship with short necks, graft migration, and anatomic variants. Treatments have therefore been proposed as alternatives to EVAR and new generations of endoprosthesis with improved designs for EVAR have also been developed. The resulting more satisfactory performance has reduced the need for reintervention and led to lower morbidity and mortality after aneurysm repair. Finally, despite the limited literature on the subject, studies report good short-term results. Since the techniques are recent, more time must elapse before long-term follow-up of these methods is possible.

Footnotes

How to cite: Belczak SQ, Pedroso GD, Ogawa LC, et al. Treatment of type 1A endoleak using coil embolization: a case report. J Vasc Bras. 2019;18:e20180130. https://doi.org/10.1590/1677-5449.180130

Financial support: None.

The study was carried out at Instituto de Aprimoramento e Pesquisa em Angiorradiologia e Cirurgia Endovascular (IAPACE), São Paulo, SP, Brazil.

REFERENCES

  • 1.Katada Y, Kondo S, Tsuboi E, Nakamura K, Rokkaku K, Irie Y. Type IA endoleak embolization after TEVAR via direct transthoracic puncture. Jpn J Radiol. 2015;33(3):169–172. doi: 10.1007/s11604-015-0392-7. [DOI] [PubMed] [Google Scholar]
  • 2.Ameli-Renani S, Morgan RA. Secondary interventions after endovascular aneurysm sac sealing: endoleak embolization and limb-related interventions. Semin Vasc Surg. 2016;29(1-2):61–67. doi: 10.1053/j.semvascsurg.2016.07.001. [DOI] [PubMed] [Google Scholar]
  • 3.Gandini R, Del Giudice C, Abrignani S, Vasili E, Pampana E, Simonetti G. Inexplicable late type ia endoleak associated with the low-profile ovation endograft in a patient with favorable neck anatomy: treatment with transcaval coil embolization. J Endovasc Ther. 2015;22(3):426–430. doi: 10.1177/1526602815579898. [DOI] [PubMed] [Google Scholar]
  • 4.Rusius VC, Davies R, Hopkins J, Duddy M, Smith S. Successful glue embolization of a late type 1A endoleak causing abdominal aortic aneurysm rupture. Vasc Endovascular Surg. 2011;45(2):198–201. doi: 10.1177/1538574410391821. [DOI] [PubMed] [Google Scholar]
  • 5.Buijs RV, Zeebregts CJ, Willems TP, Vainas T, Tielliu IF. Endograft sizing for endovascular aortic repair and incidence of endoleak type 1A. PLoS One. 2016;11(6):e0158042. doi: 10.1371/journal.pone.0158042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Galiñanes EL, Hernandez E, Krajcer Z. Preliminary results of adjunctive use of endoanchors in the treatment of short neck and pararenal abdominal aortic aneurysms. Catheter Cardiovasc Interv. 2016;87(4):E154–9. doi: 10.1002/ccd.26351. [DOI] [PubMed] [Google Scholar]
  • 7.Klonaris C, Georgopoulos S, Markatis F, Katsargyris A, Tsigris C, Bastounis E. Endovascular repair of late abdominal aortic aneurysm rupture owing to mixed-type endoleak following endovascular abdominal aortic aneurysm repair. Vascular. 2007;15(3):167–171. doi: 10.2310/6670.2007.00021. [DOI] [PubMed] [Google Scholar]
  • 8.Van den Eynde W, Van Breussegem A, Joos B, Keirse K, Verbist J, Peeters P. Endovascular repair of a type 1a endoleak after Ch-EVAR with a b-EVAR. Cardiovasc Intervent Radiol. 2016;39(9):1361–1363. doi: 10.1007/s00270-016-1388-0. [DOI] [PubMed] [Google Scholar]
  • 9.Marcelin C, Le Bras Y, Petitpierre F, et al. Embolization for persistent type IA endoleaks after chimney endovascular aneurysm repair with Onyx®. Diagn Interv Imaging. 2017;98(12):849–855. doi: 10.1016/j.diii.2017.04.005. [DOI] [PubMed] [Google Scholar]
  • 10.Law Y, Chan Y, Cheng S. Effectiveness of proximal intra-operative salvage Palmaz stent placement for endoleak during endovascular aneurysm repair. Hong Kong Med J. 2016;22(6):538–545. doi: 10.12809/hkmj154799. [DOI] [PubMed] [Google Scholar]
  • 11.Reddy NP, Ham S, Weaver F, Rowe V, Ziegler K, Han S. Repair of delayed type 1A endoleak using fenestrated and parallel endografts. Ann Vasc Surg. 2018;49:309.e7–15. doi: 10.1016/j.avsg.2018.01.070. [DOI] [PubMed] [Google Scholar]
J Vasc Bras. 2019 Jul 3;18:e20180130. [Article in Portuguese]

Tratamento de endoleak tipo 1A mediante embolização com molas: relato de caso

Sergio Quilici Belczak 1,2,, Guilherme Delicato Pedroso 1, Lara Cote Ogawa 1, Paula Thume Campos 1, Andre Lopes Padula 1, Glenna Paulain Machado 1, Matheus Zago Soares dos Santos 1, Beatriz Marques Abrão 1

Resumo

No endoleak tipo 1A, a endoprótese não sela completamente o colo do aneurisma proximal, e o fluxo arterial está presente entre a parede do colo aórtico e o material do implante. Este é um relato de um caso no qual foi utilizada embolização com molas, associado a uma revisão de literatura (PubMed, LILACS e SciELO). Foram pesquisados artigos publicados nos últimos 5 anos com os descritores “endoleak 1A”, “coil embolization” e “treatment”, combinados de formas aleatórias, sendo encontrados 25 artigos. O tipo 1A ocorre em 1,1% dos pacientes após 30 dias do implante. O tratamento consiste em aumentar a vedação do implante proximal, principalmente com o uso de stents e balões para alargar a zona de aterragem ou aumentar a força radial do implante. Alguns trabalhos sugerem técnicas de embolização com cianoacrilato, cola de fibrina e uso de Onyx, mostrando taxas de sucesso superiores a 97%. Contudo, a correção de endoleaks tipo 1A mediante embolização com molas é pouco descrita.

Palavras-chave: endoleak, embolização terapêutica, tratamento

INTRODUÇÃO

O reparo endovascular de aneurisma aórtico (endovascular aneurysm repair, EVAR) tornou-se uma alternativa comum à cirurgia aberta, por tratar-se de um procedimento menos invasivo1. Entretanto, algumas complicações podem acontecer, como o endoleak, definido como persistência de enchimento do saco aneurismático após a correção endovascular, aumentando o risco de ruptura por expansão do saco aneurismático.

Existem cinco tipos principais de endoleaks. O tipo 1 é o proximal (1A) ou distal (1B) ao local do reparo. No 1A, a endoprótese não sela completamente o colo do aneurisma, e o extravasamento arterial está presente entre a parede do colo aórtico proximal e o material de implante. No 1B, o extravasamento está presente entre a parede do colo aórtico distal e o material do implante. No tipo 2, há preenchimento do saco por fluxo retrógrado nos vasos. No tipo 3, há ocorrência de rasgo ou desconexão da endoprótese. Já o tipo 4 é usualmente visto no momento da implantação em pacientes anticoagulados devido à porosidade do implante. No tipo 5, há aumento do tamanho do saco na ausência de visibilidade do endoleak (endotensão)2.

O tratamento do endoleak tipo 1 é obrigatório3 e aumenta a vedação do implante proximal, principalmente com o uso de stents e balões para alargar a zona de aterragem ou aumentar a força radial do implante. O presente trabalho relata um caso no qual foi usada embolização com molas para resolução de endoleak tipo 1A, técnica pouco descrita na revisão de literatura.

DESCRIÇÃO DO CASO

Um paciente de 72 anos, avaliado com estado geral regular, foi internado com quadro de dor abdominal. Foi realizada uma tomografia, que diagnosticou um aneurisma de aorta abdominal de 6,4 cm e colo proximal de 1,1 cm, cônico, com angulação de 60°. Havia também a presença de um aneurisma de ilíaca interna direita de 3,4 cm de diâmetro sem colo distal. Essa anatomia, com o aneurisma de ilíaca interna se aprofundando na pelve, descartou a possibilidade de indicação de cirurgia aberta. Optou-se, no planejamento, por colocar endopótese com extensão permitindo alcançar a ilíaca externa, uma vez que o grande aneurisma de ilíaca interna sem colo distal impôs o tratamento mediante embolização.

O paciente foi submetido, então, ao tratamento endovascular com implante de endoprótese Ovation® (Endologix, Irvine, Califórnia). Na angiografia de controle, foi observado importante vazamento entre a endoprótese e o colo proximal na aorta (endoleak tipo 1A). Optou-se pelo implante de molas de liberação controlada para fechamento da área de extravasamento. Realizou-se uma angiografia de controle, que mostrou o fechamento do endoleak com as molas implantadas.

Para realização da embolização do endoleak, foi colocado um cateter Simmons® de 5 Fr encostado entre a endoprótese e a aorta. Assim, um microcateter Maestro® passou por dentro do outro cateter até alcançar o local onde se deveria embolizar, com posterior liberação de seis molas Complex Trufill 3D. O paciente apresentou resolução do endoleak e recebeu alta do hospital após 3 dias (Figuras 1 e 2).

Figura 1. Endoleak tipo 1A.

Figura 1

Figura 2. Endoleak tipo 1A após embolização com molas.

Figura 2

DISCUSSÃO

Para a revisão de literatura, foram utilizadas as bases de dados PubMed, LILACS e SciELO. Foram pesquisados artigos publicados nos últimos 5 anos com os seguintes descritores: “endoleak 1A”, “coil embolization” e “treatment”.

A complicação mais comum associada ao aneurisma endovascular é o vazamento interno, com taxas de incidência entre 6% e 57% relatadas na literatura. Um endoleak representa falha do implante de stent para excluir o saco aneurismático da circulação sistêmica e pode levar ao aumento e ruptura do saco do aneurisma. O manuseio de um vazamento interno varia de acordo com o tipo. Os tipos 1 e 3 requerem tratamento, enquanto o tipo 2 é cada vez mais gerenciado com expectativa, especialmente se não estiver associado ao alargamento do saco aneurismático4.

A incidência de vazamento interno tipo 1 pode ser atribuída à habilidade cirúrgica do cirurgião, mas também à habilidade de dimensionamento pré-operatório da endoprótese. Este é dificultado quando os aneurismas a serem reparados apresentam características anatômicas complexas, como um colo de aneurisma de curto comprimento proximal, afilamento reverso do colo, calcificação mural ou trombo e acentuada angulação do colo5.

Nem todos os aneurismas aórticos conseguem resoluções satisfatórias. Estudos identificaram uma associação direta entre colos aórticos proximais curtos e falha do EVAR, ou seja, presença de vazamento interno6. Diante disso, é preciso desenvolver técnicas especiais para resolver esses casos complexos.

Embora a ruptura de aneurisma aórtico pós-EVAR represente um grande risco de vida que requer diagnóstico imediato e intervenção emergencial, não há diretrizes específicas para a gestão mais adequada dessa situação. Dados limitados de pequenas séries de caso indicam que esses pacientes são tratados com cirurgia mais frequentemente. No entanto, o risco operatório de aneurismas da aorta abdominal rompidos pós-EVAR é alto e tais pacientes geralmente apresentam comorbidades significativas. Por essa razão, o reparo endovascular é a conduta de escolha quando favorável a cada paciente7.

A conversão para cirurgia aberta geralmente envolve exposição cirúrgica do aneurisma, controle vascular proximal e distal, remoção completa da endoprótese e substituição com prótese aórtica. Um procedimento tão complexo acarreta morbidade e mortalidade significativas.

Métodos de tratamento convencionais para vazamento tipo 1 incluem o uso de extensões aórticas proximais, angioplastia com balão e stents. Alguns casos não são passíveis de reintervenções endovasculares, e o reparo cirúrgico aberto é a única opção4. EVAR com implante de chaminé (Ch-EVAR) pode ser uma opção em casos de função cardíaca comprometida8, bem como endopróteses fenestradas (FEVAR), que podem ser úteis em pacientes com anatomia aórtica complexa9. Casos de sucesso de reparo de endoleak tipo 1A já foram relatados na literatura e incluem uso de stent intraoperatório Palmaz10. Outros estudos também relataram manejo eficaz de endoleak 1A com uso de Onyx com ou sem molas, dependendo da quantidade de calhas envolvidas11.

Em conclusão, o EVAR é um método cada vez mais preferível em relação à correção aberta nos casos com anatomia favorável, por apresentar menor morbimortalidade quando comparado à correção aberta. No entanto, foi demonstrado um maior risco de reabordagem por complicações, como os endoleaks. Estudos mostraram forte relação das complicações com colo curto, migração da prótese e variações anatômicas. Com isso, surgiram tratamentos alternativos ao EVAR, bem como novas gerações de endoprótese com melhorias no design do EVAR. Assim, performances mais satisfatórias diminuíram a necessidade de reabordagem e garantiram menor morbimortalidade na correção dos aneurismas. Por fim, apesar da literatura limitada, os estudos evidenciaram uma boa resposta em curto prazo; por ser uma técnica recente, existe a necessidade de um maior tempo para realização de acompanhamento em longo prazo desses métodos.

Footnotes

Como citar: Belczak SQ, Pedroso GD, Ogawa LC, et al. Tratamento de endoleak tipo 1A mediante embolização com molas: relato de caso. J Vasc Bras. 2019;18:e20180130. https://doi.org/10.1590/1677-5449.180130

Fonte de financiamento: Nenhuma.

O estudo foi realizado no Instituto de Aprimoramento e Pesquisa em Angiorradiologia e Cirurgia Endovascular (IAPACE), São Paulo, SP, Brasil.


Articles from Jornal Vascular Brasileiro are provided here courtesy of Sociedade Brasileira de Angiologia e Cirurgia Vascular

RESOURCES