Table 1. Quantitative features of the proposed method for the selected analytes .
Analyte | LOD a | LOQ b | LR c | r 2 d | RSD % e | |
Intra-day | Inter-day | |||||
VPA | 0.057 | 0.20 | 0.20 – 100 | 0.998 | 8 | 9 |
3-Heptanone | 0.010 | 0.040 | 0.04 – 100 | 0.997 | 5 | 5 |
aLimit of detection (S/N=3) (mg L–1).
bLimit of quantification (S/N=10) (mg L–1).
cLinear range (mg L–1).
dCoefficient of determination
eRelative standard deviation (n=6, C= 2 mg L–1 of each analyte) for intra–day and (n=4, C= 2 mg L–1of each analyte) for inter-day precisions.