Technetium‐99m‐sestamibi Redistribution after Exercise Stress Test Identified by a Novel Cardiac Gamma Camera: Two Case Reports

Yuri Sheikine 1,, Daniel S Berman 2, Marcelo F Di Carli 1,
PMCID: PMC6652908  PMID: 20358533

Abstract

Single photon emission computer tomography (SPECT) myocardial perfusion imaging (MPI) employing technetium‐99m (Tc‐99m)‐based imaging tracers is the mainstay of nuclear cardiology for the detection of myocardial ischemia. Current guidelines for same day rest/stress Tc‐99m‐sestamibi SPECT MPI recommend image acquisition 15–60 minutes after the stress testing. A novel sensitive SPECT imaging technique, D‐SPECT, allows fast acquisition of images and captures rapid changes in radiotracer distribution. Here we report 2 cases of SPECT MPI in patients with angiographically confirmed coronary artery disease (CAD) where Tc‐99m‐sestamibi exhibited marked redistribution between early (6–8 min) and late (60–70 min) post‐stress imaging leading to an underestimation of the extent and severity of ischemia on late images. These observations suggest that early imaging maybe more sensitive for CAD detection. Fast SPECT imaging techniques, such as D‐SPECT, will facilitate similar studies in the future as they will allow fast image acquisition at several time points after the stress test. Copyright © 2010 Wiley Periodicals, Inc.

Full Text

The Full Text of this article is available as a PDF (3.2 MB).

References

  • 1. Klocke FJ, Baird MG, Lorell BH, et al. ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging—executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ ASNC Committee to Revise the 1995 Guidelines for the Clinical Use of Cardiac Radionuclide Imaging). Circulation 2003; 108: 1404–1418. [DOI] [PubMed] [Google Scholar]
  • 2. Patton JA, Slomka PJ, Germano G, Berman DS. Recent technologic advances in nuclear cardiology. J Nucl Cardiol 2007; 14: 501–513. [DOI] [PubMed] [Google Scholar]
  • 3. Taillefer R, Lambert R, Bisson G, Benjamin C, Phaneuf DC. Myocardial technetium 99m‐labeled sestamibi single‐photon emission computed tomographic imaging in the detection of coronary artery disease: comparison between early (15 minutes) and delayed (60 minutes) imaging. J Nucl Cardiol 1994; 1: 441–448. [DOI] [PubMed] [Google Scholar]
  • 4. Taillefer R, Primeau M, Costi P, et al. Technetium‐99m‐sestamibi myocardial perfusion imaging in detection of coronary artery disease: comparison between initial (1‐hour) and delayed (3‐hour) postexercise images. J Nucl Med 1991; 32: 1961–1965. [PubMed] [Google Scholar]
  • 5. Franceschi M, Guimond J, Zimmerman RE, et al. Myocardial clearance of Tc‐99m hexakis‐2‐methoxy‐2‐methylpropyl isonitrile (MIBI) in patients with coronary artery disease. Clin Nucl Med 1990; 15: 307–312. [DOI] [PubMed] [Google Scholar]
  • 6. Toba M, Kumita S, Cho K, et al. Usefulness of gated myocardial perfusion SPECT imaging soon after exercise to identify postexercise stunning in patients with single‐vessel coronary artery disease. J Nucl Cardiol 2004; 11: 697–703. [DOI] [PubMed] [Google Scholar]
  • 7. Sciagra R, Sotgia B, Dona M, Pupi A. Influence of the postexercise acquisition delay on the detection of functional abnormalities in sestamibi‐gated SPECT. J Nucl Cardiol 2007; 14: 334–340. [DOI] [PubMed] [Google Scholar]