Abstract
Background
Electrical devices generate electromagnetic fields that may interfere with pacemakers. Media players cause telemetry interference with pacemakers, but it is not known whether they cause direct interference with pacemakers. The purpose of this study was to examine the interaction between pacemakers and 3 different media players.
Methods
In this prospective, randomized study, 54 patients with dual chamber pacemakers who were in sinus rhythm underwent baseline observation, followed by observation under telemetry communication. These patients were then randomly evaluated with 3 media players (iPod 3G, iPod Photo, and iPod Touch Apple, Cupertino, CA) with and without telemetry communication for 1 minute each. Patients were monitored for pacemaker malfunction using a single‐channel ECG during exposure to media players. The pacemaker was interrogated after each exposure and an interrogation report was printed for evaluation. Pacemaker interference was categorized as type I, II, or III. Types I and II interference described telemetry interference and type III interference was defined as any direct interference with pacemaker function or programmed parameters.
Results
A total of 54 patients (29 men and 25 women; mean age 77.2 ± 9.3 y) were evaluated. In total, of the 162 tests (for telemetry interference) 36.4% were positive (Type I and II). Type III interference was also evaluated in 162 tests and none showed any evidence of direct interference.
Conclusion
Media players cause telemetry interference with pacemakers, but they do not directly interfere with pacemaker function. Copyright © 2009 Wiley Periodicals, Inc.
Full Text
The Full Text of this article is available as a PDF (369.2 KB).
References
- 1. Webster G, Ligia J, Martuscello M, et al. Digital music players cause interference with interrogation telemetry for pacemakers Titcolor and implantable cardioverter‐defibrillators without affecting device function. Heart Rhythm 2008; 5: 545–550. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Thaker JP, Patel MB, Jongnarangsin K, Liepa VV, Thakur RK. Electromagnetic interference with pacemakers caused by portable media players. Heart Rhythm 2008; 5: 538–544. [DOI] [PubMed] [Google Scholar]
- 3. Thaker JP, Patel MB, Liepa VV, Jongnarangsin J, Thakur RK. Media players cause interference in implantable loop recorders. Heart Rhythm 2008; S278. [DOI] [PubMed] [Google Scholar]
- 4.Applications for FDA approval to market a new drug or an antibiotic drug. 21 CFR 314.126 1992.
- 5.Medtronic, Inc., http://wwwp.medtronic.com/Newsroom/ NewsReleaseDetails.do?itemId = 1165849256153&lang = en_US.
- 6.St Jude, Inc., http://www.medicalnewstoday.com/articles/107312. php.
- 7.Boston Scientific, Inc., http://www.aboutlatitude.com/how‐latitude‐works/index.html.
- 8.Biotronik, Inc., http://www.biotronik.com/en/us/5153.
- 9. Savci H, Sula A, Wang Z, Dogan NS. MICS transceivers: regulatory standards and applications (Medical Implant Communications Service). Proc IEEE Southeast Con; 2005; 179–182. [Google Scholar]
- 10. Bassen H. Low frequency magnetic emissions and resulting induced voltage in a pacemaker by iPod portable media players. BioMedical Engineering OnLine 2008; 7: 7 http://www.biomedical‐engineering‐online.com/content/7/1/7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Seidman SJ, Rugggera PS, Brockman RG, Lewis B, Shein MJ. Electromagnetic compatibility of pacemakers and implantable cardiac defibrillators exposed to RFID readers. IJRFITA 2007; 1: 237–246. [Google Scholar]
- 12. van der Togt R, van Lieshout EJ, Hensbrock R, Beinat E, Binnekade JM, Bakker PJM. Electromagnetic interference from radio frequency identification inducing potentially hazardous incidents in critical care medical equipment. JAMA 2008; 299(24): 2884–2890. [DOI] [PubMed] [Google Scholar]
- 13. Berwick DM. Taming the technology beast. JAMA 2008; 299(24): 2898–2899. [DOI] [PubMed] [Google Scholar]
- 14. Halperin D, Clark SS, Fu K, et al. Pacemakers and implantable cardiac defibrillators: software radio attacks and zero‐power defense. Proceedings of IEEE Symposium on Security and Privacy, May 2008. DOI: 10.1109/sp.2008.31. [Google Scholar]