Skip to main content
Clinical Cardiology logoLink to Clinical Cardiology
. 2007 Dec 10;30(12):615–620. doi: 10.1002/clc.20164

Development and Validation of a Noninvasive Method to Estimate Cardiac Output Using Cuff Sphygmomanometry

Shiu‐Shin Chio 1, Jeffrey J Tsai 1, Yen‐Ming Hsu 1, Jeffrey C Lapointe 1, Thao Huynh‐Covey 2, Oi Ling B Kwan 2, Anthony N DeMaria 2,
PMCID: PMC6653276  PMID: 18069677

Abstract

Background

Obtaining cardiac output (CO) measurements noninvasively during routine blood pressure recording can improve hypertension management. A new method has been developed that estimates cardiac output using pulse‐waveform analysis (PWA) from a brachial cuff sphygmomanometer. This study evaluates the ability of PWA to track changes in CO as derived by Doppler ultrasound during dobutamine stimulation.

Hypothesis

This study aims to validate the PWA CO estimation over a wide CO range as would be obtained by dobutamine stimulation during Doppler ultrasound evaluation.

Method

A total of 48 patients undergoing standard dobutamine stress echocardiography testing for accepted clinical indications were enrolled. Among them, 44 patients (age 36–83, 18 females, 26 males) with good waveform data for analyses provided estimates of CO in this study. Noninvasive measurements of CO were performed using both Doppler ultrasound recordings and PWA techniques simultaneously at each stage of dobutamine infusion.

Results

A total of 207 simultaneous pulse‐waveform analyses and Doppler measurements were taken during dobutamine stress on 44 cardiac patients. Linear regression analysis revealed good intra‐patient correlation between pulse‐waveform analysis and Doppler at different dobutamine‐induced CO with coefficients from r = 0.69 to 0.98 (p < 0.05). Analysis of all patients yielded an overall correlation of r = 0.82 (p < 0.001, bias = 0.4 L/min, standard deviation = 1.8 L/min).

Conclusion

The CO measured noninvasively from a sphygmomanometer using this PWA method correlates well with those of Doppler through a range of dobutamine‐stimulated levels. The CO by PWA should be useful for monitoring hemodynamic changes in hypertensive and cardiac patients during routine blood pressure measurement. Copyright © 2007 Wiley Periodicals, Inc.

Keywords: cardiac output, pulse‐waveform analysis, thermo‐dilution, Doppler ultrasound, dobutamine stimulation, hypertension

Full Text

The Full Text of this article is available as a PDF (296.1 KB).

References

  • 1. Taler SJ, Textor SC, Augustine JE: Resistant hypertension: comparing hemodynamic management to specialist care. Hypertension 2002; 39: 982–988. [DOI] [PubMed] [Google Scholar]
  • 2. Messerli FH: Calcium antagonists in hypertension: from hemodynamics to outcomes. Am J Hypertens 2002; 15: 94S–97S. [DOI] [PubMed] [Google Scholar]
  • 3. Shoemaker WC, Wo CJ, Bishop MH, Appel PL, van de Water JM, et al.: Multicenter trial of a new thoracic electrical bioimpendance device for cardiac output estimation. Crit Care Med 1994; 22: 1907–1912. [PubMed] [Google Scholar]
  • 4. Hirschl MM, Kittler H, Woisetschager C, Siostrzonek P, Staudinger T, et al.: Simultaneous comparison of thoracic bioimpedance and arterial pulse waveform‐derived cardiac output with thermo‐dilution measurement. Crit Care Med 2000; 28: 1798–1802. [DOI] [PubMed] [Google Scholar]
  • 5. Hirschl MM, Binder M, Gwechenberger M, Herkner H, Bur A, et al.: Non‐invasive assessment of cardiac output in critically ill patients by analysis of the finger blood pressure waveform. Crit Care Med 1997; 25: 1909–1914. [DOI] [PubMed] [Google Scholar]
  • 6. Liu Z, Brin KP, Yin FC: Estimation of total arterial compliance: an improved method and evaluation of current methods. Am J Phisol 1986; 251: H588–H600. [DOI] [PubMed] [Google Scholar]
  • 7. McVeigh GE, Burns DE, Finkelstein SM, McDonald KM, Mock JE, et al.: Reduced vascular compliance as a marker for essential hypertension. Am J Hypertens 1991; 4: 245–251. [DOI] [PubMed] [Google Scholar]
  • 8. O'Rourke MF, Brunner HR: Introduction to arterial compliance and function. J Hypertens 1992; 10: S3–S5. [PubMed] [Google Scholar]
  • 9. Watt TB, Burrus CS: Arterial pressure contour analysis for estimating human vascular properties. J Appl Physiol 1976; 40: 171–176. [DOI] [PubMed] [Google Scholar]
  • 10. Simon AC, Safar ME, Levenson JA, London GM, Levy BI, et al.: An evaluation of large arteries compliance in man. Am J Physiol 1979; 237: H550–H554. [DOI] [PubMed] [Google Scholar]
  • 11. Brinton TJ, Kailasam MT, Wu RA, Cervenka JH, Chio SS, et al.: Arterial compliance by cuff sphygmomanometer. Application to hypertension and early changes in subjects at genetic risk. Hypertension 1996; 28: 599–603. [DOI] [PubMed] [Google Scholar]
  • 12. Brinton TJ, Cotter B, Kailasam MT, Brown DL, Chio SS, et al.: Development and validation of a noninvasive method to determine arterial pressure and vascular compliance. Am J Cardiol 1997; 80: 323–330. [DOI] [PubMed] [Google Scholar]
  • 13. Brinton TJ, Walls ED, Chio SS: Validation of pulse dynamic blood pressure measurement by auscultation. Blood Press Monit 1998; 3: 121–124. [PubMed] [Google Scholar]
  • 14. Armstrong WF, Zoghbi WA: Stress echocardiography current methodology and clinical applications. J Am Coll Cardiol 2005; 45: 1739–1747. [DOI] [PubMed] [Google Scholar]
  • 15. Chio SS. Method and apparatus for determining blood pressure and cardiovascular condition. U.S. Patent 4,880,013. November 14, 1989.
  • 16. Chio SS. Method for diagnosing, monitoring and treating hypertension and other cardiac problems. U.S. Patent 5,836,884. November 17, 1998.
  • 17. Chio SS. Method for diagnosing, monitoring and treating hypertension and other cardiac problems. U.S. Patent 6,165,130. December 26, 2000.
  • 18. Chio SS. Method for diagnosing, monitoring and treating hypertension and other cardiac problems. U.S. Patent 6,270,461, Aug. 7, 2001, and U.S. Patent 6,540,687 B2, Apr. 1, 2003.
  • 19. Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307–310. [PubMed] [Google Scholar]

Articles from Clinical Cardiology are provided here courtesy of Wiley

RESOURCES