Abstract
Background
Whole-heart magnetic resonance angiography (MRA) requires sophisticated methods accounting for respiratory motion. Our purpose was to evaluate the image quality of compressed sensing-based respiratory motion-resolved three-dimensional (3D) whole-heart MRA compared with self-navigated motion-corrected whole-heart MRA in patients with known thoracic aorta dilation.
Methods
Twenty-five patients were prospectively enrolled in this ethically approved study. Whole-heart 1.5-T MRA was acquired using a prototype 3D radial steady-state free-precession free-breathing sequence. The same data were reconstructed with a one-dimensional motion-correction algorithm (1D-MCA) and an extradimensional golden-angle radial sparse parallel reconstruction (XD-GRASP). Subjective image quality was scored and objective image quality was quantified (signal intensity ratio, SIR; vessel sharpness). Wilcoxon, McNemar, and paired t tests were used.
Results
Subjective image quality was significantly higher using XD-GRASP compared to 1D-MCA (median 4.5, interquartile range 4.5–5.0 versus 4.0 [2.25–4.75]; p < 0.001), as well as signal homogeneity (3.0 [3.0–3.0] versus 2.0 [2.0–3.0]; p = 0.003), and image sharpness (3.0 [2.0–3.0] vs 2.0 [1.25–3.0]; p < 0.001). SIR with the 1D-MCA and XD-GRASP was 6.1 ± 3.9 versus 7.4 ± 2.5, respectively (p < 0.001); while signal homogeneity was 274.2 ± 265.0 versus 199.8 ± 67.2 (p = 0.129). XD-GRASP provided a higher vessel sharpness (45.3 ± 10.7 versus 40.6 ± 101, p = 0.025).
Conclusions
XD-GRASP-based motion-resolved reconstruction of free-breathing 3D whole-heart MRA datasets provides improved image contrast, sharpness, and signal homogeneity and seems to be a promising technique that overcomes some of the limitations of motion correction or respiratory navigator gating.
Keywords: Aorta, Dilatation, Image processing (computer–assisted), Magnetic resonance angiography, Motion
Key points
Prototype compressed sensing whole-heart magnetic resonance angiography (MRA) provides improved signal homogeneity and sharpness.
Motion-resolved three-dimensional whole-heart MRA reconstruction outperformed one-dimensional image-based motion correction.
Motion-resolved MRA allows for image acquisition without the need for respiratory navigator.
Background
Whole-heart magnetic resonance angiography (MRA) has been extensively used as a clinical tool to visualize three-dimensional (3D) cardiac anatomy [1]. While 3D whole-heart MRA can be obtained using contrast-enhanced techniques [2], the most common application remains the unenhanced steady-state free-precession (SSFP) acquisition [3–6]. As the heart is relatively still only for a short mid-diastolic or late systolic period of the cardiac cycle, collection of such MRA data requires hundreds of cardiac cycles and a protocol designed for free-breathing imaging.
Current 3D SSFP MRA techniques depend on sophisticated respiratory motion compensation or gating. Diaphragmatic navigators, the only tool clinically available, have several limitations including unpredictable acquisition time and low scan efficiency, which can render the length of the acquisition excessively long, up to 28 min [7–10]. Alternatively, a variety of self-navigated techniques have been proposed, which allow the extraction of a respiratory motion signal directly from the image data that can be used for respiratory motion correction during post-processing [11–16]. Self-navigation, employing a radial trajectory combined with one-dimensional correction, has been one of the most widely used investigational approaches which can provide 100% scan efficiency and predictable acquisition times, enabling the application of this technique even in children with limited compliance [17, 18]. However, self-navigation also suffers from certain limitations that are related to the 1D motion model used for correction [19], which may not be accurate when a wide range of respiratory motion is present [20].
Recent developments have shifted towards more sophisticated reconstruction techniques that are able to extract a respiratory signal from the image data and use it to sort images into different respiratory motion states. The extradimensional golden-angle radial sparse parallel (XD-GRASP) method, a novel compressed sensing image reconstruction framework, combines the benefits of reduced k-space sampling and sparse reconstruction [20, 21]. In particular, XD-GRASP enables the reconstruction of 3D radial golden-angle free-breathing coronary artery MRA data at multiple respiratory phases by exploiting the sparsity along the respiratory dimension [20]. This is a paradigm shift for motion compensation, as with this technique, respiratory motion is no longer corrected using an approximation of the displacement (e.g., reducing the motion to a 1D correction). Instead, the strong similarity between different respiratory phases is used to improve the image quality of the reconstruction.
The purpose of this study was to evaluate the subjective and objective image quality of XD-GRASP-based respiratory motion-resolved 3D whole-heart MRA in comparison with respiratory motion-corrected whole-heart MRA in patients with known thoracic aorta dilation.
Methods
Patient selection
Our study protocol was approved by the local Institutional Review Board and written informed consent was obtained from all patients. The study was conducted in compliance with the Health Insurance Portability and Accountability Act guidelines. Patients (n = 25) with known thoracic aorta dilation were prospectively enrolled for a research study between July 2017 and September 2018. Further inclusion criteria were: (1) > 18 years of age; (2) previous clinically indicated aorta, chest, triple-rule-out, or pulmonary embolism computed tomography examination; and (3) willing to comply with all study procedures and provide written informed consent. General magnetic resonance exclusion criteria were applied to patient selection. Patient’s demographics were obtained by medical record chart review.
Acquisition protocol
Image acquisition was performed on a 1.5-T system (Magnetom Avanto DOT, Siemens Healthcare, Erlangen, Germany). Patients were scanned head-first in a supine position. A multi-channel spine phased-array radiofrequency coil with 24 elements integrated into the patient table and a 6 element, 6-channel phased-array body coil were used for signal reception. Acquisitions were electrocardiographically gated. The entire protocol was performed in a free-breathing fashion. Following the initial scout images, a free-breathing two-dimensional balanced SSFP cine image set in a parasagittal long-axis view of the left ventricle was acquired using the following typical parameters: repetition time/echo time, 2.3/1.1 ms; field of view 220–340 mm2; matrix 1922; number of segments 15; reconstructed phases 25; temporal resolution 45 ms; flip angle 77°; number of excitations 3; and parallel acquisition acceleration factor 2. Cine image data were used to determine the optimal mid-diastolic timing for the whole-heart MRA.
Whole-heart MRA was performed using a prototype fat-saturated and T2-prepared pulse sequence that employs the 3D radial trajectory [7, 22]. A coronal saturation slab was placed at the level of the anterior chest wall and the following imaging parameters were used: repetition time/echo time 3.1/1.5 ms; field of view 320 × 320 × 320 mm3; reconstructed voxel size, 1.7 × 1.7 × 1.7 mm3; matrix 1923; flip angle 115°; and bandwidth, 898 Hz/pixel. An acceleration factor of 5 with respect to the Nyquist sampling for 3D radial imaging was applied [21]. The acquisition was performed in free-breathing and the following two approaches were used to address respiratory motion within the reconstruction.
Post-processing
Respiratory motion-corrected approach
For the self-navigation approach, respiratory motion was extracted by cross-correlating the automatically segmented blood pool of the 1D Fourier transform of a readout along the superior-inferior direction acquired consistently at the beginning of each heartbeat. The detected 1D superior-inferior respiratory displacement was then used for correcting each readout before the gridding operation. The correction was performed by applying a phase shift to all k-space radial readouts and was adapted for the polar orientation of each readout according to the spiral phyllotaxis pattern [22]. This reconstruction has been implemented inline at the scanner and takes about 1–2 min. Further details about the algorithm employed for motion correction were described by Piccini et al. [12].
Respiratory motion-resolved approach
For the respiratory motion-resolved approach, the same raw data of the whole-heart MRA acquisition were exported and processed on a dedicated workstation using an adaptation of the previously described framework [20] implemented in MATLAB 2015a (MathWorks, Natick, Massachusetts, USA). Using a respiratory signal extracted directly from the image data, individual readouts of the 3D radial acquisition were binned according to their respiratory phase [23]. The resultant series of motion-resolved undersampled images were then reconstructed using an XD-GRASP algorithm [21], which aims at exploiting the intrinsic similarities between distinct respiratory phases (or motion states) of the whole-heart acquisition to perform a compressed sensing reconstruction along the respiratory motion dimension. To achieve this, first, the acquired readouts are separated and grouped according to the respiratory phase they belong to. Subsequently, a k-t sparse SENSE iterative reconstruction is performed, where the temporal domain is represented by the different respiratory phases. Out of the four reconstructed respiratory phases, the end-expiratory phase was selected in all datasets for the subsequent analyses. The motion resolved reconstruction has not been implemented inline and takes between 15–30 min using high-end computers.
Image analysis
Respiratory motion-corrected and motion-resolved MRA reconstructions were randomized and independently reviewed by two readers (with 1 and 11 years of experience in cardiovascular imaging, respectively) on a dedicated workstation (Aquarius iNtuition Edition v4.4.12, TeraRecon, Inc., Foster City, CA, USA). Standard axial, coronal, and sagittal planes were used to generate multi-planar reformats (MPR), but readers were allowed to use curved MPR or maximum intensity projection (MIP) series according to their preference. The presence of artifacts was noted.
Qualitative analysis
The overall image quality was subjectively rated independently by each reader on a 5-point Likert-scale: (1) vascular anatomy not assessable due to severe image artifacts and/or poor contrast, (2) vascular anatomy assessable despite severe image artifacts and/or poor contrast, (3) acceptable image quality with artifacts and/or limited contrast, (4) good image quality with minor artifacts and/or good contrast, and (5) excellent image quality without artifacts and excellent contrast. Signal homogeneity in the intra-aortic blood pool was rated using a 3-point Likert-scale as (1) inhomogeneity affecting diagnosis, (2) subtle inhomogeneity with no effect on diagnosis, and (3) excellent homogeneity. Image sharpness was evaluated on a 3-point scale as follows: (1) motion affecting diagnosis, (2) motion with no effect on diagnosis, and (3) no significant motion. Finally, diagnostic confidence was also rated by each reader using a 3-point scale as (1) low reader confidence, (2) marginal reader confidence, and (3) high reader confidence.
Quantitative analysis
Readers visualized seven standard anatomical levels of the thoracic aorta using a double oblique technique as follows: sinuses of Valsalva, sinotubular junction, mid ascending aorta, proximal aortic arch, mid aortic arch, proximal descending aorta, and mid descending aorta [24]. At each level, the signal intensity ratio (SIR) between the intravascular signal and the surrounding lung tissue was calculated. Regions of interest were placed in the center of the aorta. Aorta blood pool signal homogeneity was quantified by measuring the standard deviation of blood signal as a function of distance along the thoracic aorta on centerline reconstructions. Finally, the sharpness of the right coronary artery was quantitatively evaluated using a dedicated prototype application (Soap-Bubble, John’s Hopkins University, Baltimore, MD, USA) [25]. This software examines a user-assisted definition of a curved subvolume enclosed in the 3D MRA dataset and measures the magnitude of local change in signal intensity at the vessel borders in the reformatted image. The resulting vessel edge value indicates quantitative sharpness, whereas a value of 100% refers to an abrupt signal intensity change at the vessel border, and lower values are associated with lower vessel sharpness [25]. Measurements were taken in the proximal segment of the right coronary artery, being the most sensitive to motion [26].
Statistical analysis
Statistical analysis was performed using SPSS v213 (IBM Corporation, Armonk, NY, USA). Categorical variables are represented as total number and percentages, and continuous variables as mean ± standard deviation or median (interquartile range), depending on their distribution (tested with Shapiro Wilkes). Subjective image quality scores were compared between the respiratory motion-corrected and motion-resolved techniques using the Wilcoxon signed-rank test, and the McNemar test was used to compare the presence of image artifacts. Interclass correlations (ICC) were used to assess the absolute agreement between readers and were interpreted as follows: < 0.2, poor; 0.2–0.4, acceptable; 0.41–0.6, moderate; 0.61–0.8, good; and > 0.8, excellent. Objective image quality measures were compared using a paired Student’s t test. A p value < 0.05 was considered significant.
Results
A total of 25 patients with thoracic aorta dilation were enrolled, aged 70 ± 9 years (mean ± standard deviation, 16 men and 9 women). The mean patient body weight and body mass index were 88.5 ± 14.6 kg and 27.8 ± 3.0 kg/m2, respectively. All patients were successfully scanned using the free-breathing whole-heart MRA protocol. Representative high-quality images of respiratory motion-corrected and motion-resolved reconstructions of corresponding datasets are shown in Fig. 1.
Overall subjective image quality was rated significantly higher using the motion-resolved reconstruction compared to motion correction (4.5 versus 4.0, p < 0.001), including signal homogeneity and image sharpness (Table 1). Representative images highlighting the signal homogeneity and sharpness differences between the two techniques are shown in Fig. 2. Image artifacts mostly arising from motion were noted on 7 motion-corrected and 3 motion-resolved datasets (p = 0.219). The improved image quality of the motion-resolved reconstruction also resulted in higher diagnostic confidence scores (3.0 versus 2.0, p = 0.016). Inter-reader assessment showed moderate to excellent agreement between the readers in motion-corrected datasets (ICCs between 0.577 and 0.841) and good to excellent agreement in the motion-resolved image sets (ICCs between 0.648 and 0.860) (Table 2).
Table 1.
Respiratory self-navigated | Respiratory motion-resolved | p value | |
---|---|---|---|
Overall image quality | 4.0 [2.25–4.75] | 4.5 [4.5–5.0] | < 0.0001* |
Signal homogeneity | 2.0 [2.0–3.0] | 3.0 [3.0–3.0] | 0.003* |
Image sharpness | 2.0 [1.25–3.0] | 3.0 [2.0–3.0] | 0.0001* |
Presence of artifacts | 7 (25%) | 3 (10.7%) | 0.219 |
Diagnostic confidence | 2.0 [2.0–3.0] | 3.0 [2.0–3.0] | 0.016* |
*Indicating significant difference
Table 2.
Respiratory self-navigated | Respiratory motion-resolved | |
---|---|---|
Overall image quality | 0.841 | 0.860 |
Signal homogeneity | 0.577 | 0.648 |
Image sharpness | 0.784 | 0.772 |
Presence of artifacts | 0.825 | 0.680 |
Diagnostic confidence | 0.726 | 0.780 |
Objective image quality parameters showed improvement when the motion-resolved technique was used. Overall SIR values with the motion-corrected and motion-resolved techniques were 6.1 ± 3.9 versus 7.4 ± 2.5 (p < 0.001). Individual SIR measurements taken at the various standard anatomical levels of the thoracic aorta showed significant improvement at the mid arch, proximal descending aorta, and mid descending aorta (Table 3). No statistically significant difference was observed in signal homogeneity (motion-corrected versus motion-resolved 274.2 ± 265.0 vs 199.8 ± 67.2, p = 0.129); however, the standard deviations for the motion-corrected acquisitions were higher when compared with the motion-resolved technique, indicating a more uniform signal measurement in the latter. Soap-Bubble-based image sharpness analysis revealed that motion-resolved reconstruction provides higher coronary vessel sharpness (45.3 ± 10.7 versus 50.6 ± 10.1, p = 0.025), as shown in the image example in Fig. 3.
Table 3.
Respiratory self-navigated | Respiratory motion-resolved | p value | |
---|---|---|---|
Signal intensity ratio | |||
Sinus | 6.9 ± 4.5 | 7.7 ± 2.2 | 0.325 |
Sinotubular junction | 7.2 ± 5.4 | 7.8 ± 2.2 | 0.601 |
Ascending aorta | 7.1 ± 5.2 | 7.3 ± 2.5 | 0.829 |
Proximal arch | 5.6 ± 3.4 | 6.1 ± 1.8 | 0.475 |
Mid arch | 4.7 ± 1.9 | 5.6 ± 1.7 | 0.035* |
Proximal descending aorta | 5.0 ± 1.3 | 8.0 ± 2.7 | < 0.0001* |
Mid-descending aorta | 6.0 ± 2.8 | 9.2 ± 2.9 | 0.0001* |
Signal homogeneity | 274.2 ± 265.0 | 199.8 ± 67.2 | 0.129 |
Right coronary artery sharpness | 45.3 ± 10.7 | 50.6 ± 10.1 | 0.025* |
*Indicating significant difference
Discussion
This study aimed to evaluate if respiratory-resolved compressed sensing reconstruction, specifically XD-GRASP, provides improved image quality of free-breathing 3D whole-heart MRA, when compared to self-navigated motion-corrected reconstructions of the same sets of data in a patient population. Subjective and objective image quality measures were assessed by two readers. Overall, we found improved image quality and better diagnostic confidence using the respiratory motion-resolved reconstruction, indicating that the technique has the potential to overcome the limitations of other approaches used for respiratory motion compensation or correction.
In this study, we reported improvements in both subjective and objective image quality parameters when the motion-resolved technique was used. Subjective ratings were higher in the overall assessment, but also showed a significant increase when signal homogeneity and image sharpness were rated separately. The improvement in image quality also resulted in an increase in the readers’ diagnostic confidence. Inter-reader assessment showed similar agreement between the readers for both motion-corrected and motion-resolved.
Vascular signal in the motion-corrected data sets proved to be less uniform, especially in the center of the image, most likely because the compressed sensing reconstruction technique reduced the standard deviation of the noise. Such inhomogeneity was not observed on motion-resolved reconstructions, despite the utilization of the same raw data for both algorithms. Although the quantitative signal homogeneity assessment did not show any significant difference, a substantially higher standard deviation can be observed in the motion-corrected datasets, which potentially contributed to the statistical outcome.
Coronary vessel sharpness, on the other hand, was also found to be significantly improved when the respiratory motion-resolved reconstruction was used. The Soap-Bubble analysis confirmed the increase of sharpness, which otherwise can also be visually observed. The quantitative vessel edge sharpness values measured in the motion-resolved datasets in this study population were comparable to those obtained using respiratory navigator-gated MRA [25] and higher than those reported with self-navigated motion-corrected whole-heart MRA with or without contrast administration [7, 17]. A previous study performed in a limited number of healthy volunteers reported similar improvement in vessel edge sharpness using the motion-resolved reconstruction when compared to motion correction [20]. Our study, however, was different in multiple aspects. We not only set out to use the technique in a cohort of clinical patients, creating a more relevant study design, but we also imaged the entire thoracic aorta. The latter required the use of an increased field of view (in the 300–350-mm range), which subsequently reduced spatial resolution. Despite the decrease in spatial resolution, we were able to demonstrate the superiority of the motion-resolved technique over the motion-corrected reconstruction in objective, quantifiable vessel edge sharpness. While vessel sharpness does not necessarily represent the quality of respiratory motion compensation or correction, this measure may also be influenced by other factors including cardiac motion (e.g., patients with high heart rate variability) and patient movements. However, the very same datasets were used and compared in our study, meaning that the cardiac motion and patient movements were exactly the same in both reconstructions.
Finally, SIR measurements indicated significant improvement in image contrast using the respiratory motion-resolved reconstruction. The difference in SIR was the most apparent at the mid arch, proximal descending aorta, and mid descending aorta levels. One possible explanation is that the XD-GRASP reconstruction provides both a better suppression of the motion artifacts (resolving motion versus correcting) and a reduced amount of noise (regularization term in the reconstruction) not only in the heart and vessels, but also in the lung tissue. Since the region of interest selected within the pulmonary tissue averages pixels with relatively very low intensities, even small changes in the noise levels will produce improved SIRs.
The XD-GRASP-based reconstruction is an algorithm which outputs a motion resolved 3D whole-heart dataset without the need for any navigator or motion correction. As the reconstruction can be performed offline after the acquisition, theoretically any free-breathing radial 3D whole-heart dataset can be reprocessed without the need to acquire new data. The motion-resolved algorithm overcomes most of the limitations that diaphragmatic navigator gating or self-navigation combined with motion correction is subject to [20], such as long and unpredictable image acquisition times [27–31] or artifacts and residual motion issues with the 1D self-navigation [6, 20]. While several other groups have proposed continuous data acquisition regardless of the phase of the respiratory cycle [11, 12, 14–16], most of the techniques apply retrospective motion correction using registration algorithms or motion models [13, 14, 32, 33], which, in contrast, are not used in the XD-GRASP algorithm. The novelty in applying a respiratory motion-resolved algorithm is that the image data can be continuously acquired in a free-breathing fashion over a certain number of predetermined heartbeats, without the need for any real-time navigation or motion correction [20]. The reconstruction algorithm considers respiratory motion as an additional dimension without imposing a specific motion model for the reconstruction and allows the reader to choose the most optimal phase from the respiratory domain during the post-processing steps. Such XD-GRASP-based motion-resolved reconstruction approaches are now moving towards five-dimensional continuous imaging, where the same concept is applied at the same time to respiratory and cardiac motion [34].
Our study has some limitations to consider. While our sample size is limited, such image quality assessment does not require a large patient cohort. Our study did not demonstrate how image quality improvement affects diagnostic value; however, going forward, we are planning to expand the assessment to other aspects of the evaluation (diagnostic accuracy, coronary artery visualization, etc.) in a wider range of patients (including various age, as children with low compliance) and for different disease groups (e.g., congenital heart disease). Another limitation that highly depends on the available infrastructure is the non-negligible computational power that the motion-resolved reconstruction requires. Currently, the processing time is in the 15–30-min range using high-end computers, which is expected to decrease with the continuous improvement in computer technologies. Finally, MPRs and MIPs generated individually by the two readers may have influenced visualization and consequently image quality assessment.
In conclusion, XD-GRASP-based motion resolved reconstruction of free-breathing 3D whole-heart MRA datasets provides improved image contrast, sharpness, and signal homogeneity and seems to be a promising technique that overcomes some of the limitations of motion correction or respiratory navigator gating.
Availability of data and materials
The relevant data have been included in the manuscript.
Funding
This study was supported by a research grant from Siemens.
Abbreviations
- 1D
One-dimensional
- 2D
Two-dimensional
- 3D
Three-dimensional
- ICC
Interclass correlation
- MIP
Maximum intensity projection
- MPR
Multi-planar reformat
- MRA
Magnetic resonance angiography
- SIR
Signal intensity ratio
- SSFP
Steady-state free-precession
- XD-GRASP
Extradimensional golden-angle radial sparse parallel
Authors’ contributions
RES contributed to the data analysis and manuscript drafting. DP contributed to the data analysis and manuscript drafting. UJS contributed to the study design and manuscript revision. JH contributed to the data analysis and manuscript revision. JY contributed to the data analysis and manuscript revision. LDS contributed to the data analysis and manuscript revision. JDR contributed to the data analysis and manuscript revision. AMF contributed to the statistical analysis and manuscript revision. PS contributed to the study design, MRI studies, and manuscript revision. AVS contributed to the study design, MRI studies, and manuscript drafting and revision. All authors read and approved the final manuscript.
Ethics approval and consent to participate
The study protocol was approved by the Institutional Review Board at the Medical University of South Carolina and complied with the HIPAA guidelines.
Consent for publication
All patients provided written informed consent.
Competing interests
U. Joseph Schoepf is a consultant for and/or receives research support from Astellas, Bayer, Elucid Bioimaging, GE, Guerbet, HeartFlow Inc., and Siemens. Davide Piccini is an employee of Siemens. John Heerfordt’s doctoral studies are financially supported by Siemens. Akos Varga-Szemes receives institutional research support and travel support from Siemens and is a consultant for Elucid Bioimaging. The other authors declare that they have no competing interests.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Robert E. Stroud, Phone: +1-843-876-0097
U. Joseph Schoepf, Phone: +1-843-876-0097.
Jonathan D. Rollins, Phone: +1-843-876-0097
Andreas M. Fischer, Phone: +1-843-876-0097
Pal Suranyi, Phone: +1-843-876-0097.
Akos Varga-Szemes, Phone: +1-843-876-0097, Email: vargaasz@musc.edu.
References
- 1.Weber OM, Martin AJ, Higgins CB. Whole-heart steady-state free precession coronary artery magnetic resonance angiography. Magn Reson Med. 2003;50:1223–1228. doi: 10.1002/mrm.10653. [DOI] [PubMed] [Google Scholar]
- 2.Liu X, Bi X, Huang J, Jerecic R, Carr J, Li D. Contrast-enhanced whole-heart coronary magnetic resonance angiography at 3.0 T: comparison with steady-state free precession technique at 1.5 T. Invest Radiol. 2008;43:663–668. doi: 10.1097/RLI.0b013e31817ed1ff. [DOI] [PubMed] [Google Scholar]
- 3.Srichai MB, Kim S, Axel L, Babb J, Hecht EM. Non-gadolinium-enhanced 3-dimensional magnetic resonance angiography for the evaluation of thoracic aortic disease: a preliminary experience. Tex Heart Inst J. 2010;37:58–65. [PMC free article] [PubMed] [Google Scholar]
- 4.Krinsky GA, Rofsky NM, DeCorato DR, et al. Thoracic aorta: comparison of gadolinium-enhanced three-dimensional MR angiography with conventional MR imaging. Radiology. 1997;202:183–193. doi: 10.1148/radiology.202.1.8988210. [DOI] [PubMed] [Google Scholar]
- 5.Neimatallah MA, Ho VB, Dong Q, et al. Gadolinium-enhanced 3D magnetic resonance angiography of the thoracic vessels. J Magn Reson Imaging. 1999;10:758–770. doi: 10.1002/(SICI)1522-2586(199911)10:5<758::AID-JMRI21>3.0.CO;2-A. [DOI] [PubMed] [Google Scholar]
- 6.Renker M, Varga-Szemes A, Schoepf UJ, et al. A non-contrast self-navigated 3-dimensional MR technique for aortic root and vascular access route assessment in the context of transcatheter aortic valve replacement: proof of concept. Eur Radiol. 2016;26:951–958. doi: 10.1007/s00330-015-3906-x. [DOI] [PubMed] [Google Scholar]
- 7.Piccini D, Monney P, Sierro C, et al. Respiratory self-navigated postcontrast whole-heart coronary MR angiography: initial experience in patients. Radiology. 2014;270:378–386. doi: 10.1148/radiol.13132045. [DOI] [PubMed] [Google Scholar]
- 8.Stuber M, Weiss RG. Coronary magnetic resonance angiography. J Magn Reson Imaging. 2007;26:219–234. doi: 10.1002/jmri.20949. [DOI] [PubMed] [Google Scholar]
- 9.Abdalla EK, Bauer TW, Chun YS, D'Angelica M, Kooby DA, Jarnagin WR. Locoregional surgical and interventional therapies for advanced colorectal cancer liver metastases: expert consensus statements. HPB (Oxford) 2013;15:119–130. doi: 10.1111/j.1477-2574.2012.00597.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Sakuma H, Ichikawa Y, Chino S, Hirano T, Makino K, Takeda K. Detection of coronary artery stenosis with whole-heart coronary magnetic resonance angiography. J Am Coll Cardiol. 2006;48:1946–1950. doi: 10.1016/j.jacc.2006.07.055. [DOI] [PubMed] [Google Scholar]
- 11.Stehning C., Börnert P., Nehrke K., Eggers H., Stuber M. Free-breathing whole-heart coronary MRA with 3D radial SSFP and self-navigated image reconstruction. Magnetic Resonance in Medicine. 2005;54(2):476–480. doi: 10.1002/mrm.20557. [DOI] [PubMed] [Google Scholar]
- 12.Piccini D, Littmann A, Nielles-Vallespin S, Zenge MO. Respiratory self-navigation for whole-heart bright-blood coronary MRI: methods for robust isolation and automatic segmentation of the blood pool. Magn Reson Med. 2012;68:571–579. doi: 10.1002/mrm.23247. [DOI] [PubMed] [Google Scholar]
- 13.Pang J, Bhat H, Sharif B, et al. Whole-heart coronary MRA with 100% respiratory gating efficiency: self-navigated three-dimensional retrospective image-based motion correction (TRIM) Magn Reson Med. 2014;71:67–74. doi: 10.1002/mrm.24628. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Henningsson M, Koken P, Stehning C, Razavi R, Prieto C, Botnar RM. Whole-heart coronary MR angiography with 2D self-navigated image reconstruction. Magn Reson Med. 2012;67:437–445. doi: 10.1002/mrm.23027. [DOI] [PubMed] [Google Scholar]
- 15.Prieto C, Doneva M, Usman M, et al. Highly efficient respiratory motion compensated free-breathing coronary MRA using golden-step Cartesian acquisition. J Magn Reson Imaging. 2015;41:738–746. doi: 10.1002/jmri.24602. [DOI] [PubMed] [Google Scholar]
- 16.Ingle RR, Wu HH, Addy NO, et al. Nonrigid autofocus motion correction for coronary MR angiography with a 3D cones trajectory. Magn Reson Med. 2014;72:347–361. doi: 10.1002/mrm.24924. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Albrecht MH, Varga-Szemes A, Schoepf UJ, et al. Coronary artery assessment using self-navigated free-breathing radial whole-heart magnetic resonance angiography in patients with congenital heart disease. Eur Radiol. 2018;28:1267–1275. doi: 10.1007/s00330-017-5035-1. [DOI] [PubMed] [Google Scholar]
- 18.Albrecht MH, Varga-Szemes A, Schoepf UJ et al (2019) Diagnostic accuracy of noncontrast self-navigated free-breathing MR angiography versus CT angiography: a prospective study in pediatric patients with suspected anomalous coronary arteries. Acad Radiol. 10.1016/j.acra.2018.12.010 [DOI] [PubMed]
- 19.Manke Dirk, Nehrke Kay, Börnert Peter, Rösch Peter, Dössel Olaf. Respiratory motion in coronary magnetic resonance angiography: A comparison of different motion models. Journal of Magnetic Resonance Imaging. 2002;15(6):661–671. doi: 10.1002/jmri.10112. [DOI] [PubMed] [Google Scholar]
- 20.Piccini D, Feng L, Bonanno G, et al. Four-dimensional respiratory motion-resolved whole heart coronary MR angiography. Magn Reson Med. 2017;77:1473–1484. doi: 10.1002/mrm.26221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Feng L, Axel L, Chandarana H, Block KT, Sodickson DK, Otazo R. XD-GRASP: Golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing. Magn Reson Med. 2016;75:775–788. doi: 10.1002/mrm.25665. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Piccini D, Littmann A, Nielles-Vallespin S, Zenge MO. Spiral phyllotaxis: the natural way to construct a 3D radial trajectory in MRI. Magn Reson Med. 2011;66:1049–1056. doi: 10.1002/mrm.22898. [DOI] [PubMed] [Google Scholar]
- 23.Di Sopra L, Piccini D, Coppo S, Bastiaansen J, Stuber M, Yerly J. Motion-resolved 5D imaging of the heart: time to get rid of the ECG? Proc Intl Soc Mag Reson Med. 2017;25:3148. [Google Scholar]
- 24.Hiratzka LF, Bakris GL, Beckman JA, et al. 2010 ACCF/AHA/AATS/ACR/ASA/SCA/SCAI/SIR/STS/SVM guidelines for the diagnosis and management of patients with Thoracic Aortic Disease: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, American Association for Thoracic Surgery, American College of Radiology, American Stroke Association, Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Interventions, Society of Interventional Radiology, Society of Thoracic Surgeons, and Society for Vascular Medicine. Circulation. 2010;121:e266–e369. doi: 10.1161/CIR.0b013e3181d4739e. [DOI] [PubMed] [Google Scholar]
- 25.Etienne A, Botnar RM, Van Muiswinkel AM, Boesiger P, Manning WJ, Stuber M. “Soap-Bubble” visualization and quantitative analysis of 3D coronary magnetic resonance angiograms. Magn Reson Med. 2002;48:658–666. doi: 10.1002/mrm.10253. [DOI] [PubMed] [Google Scholar]
- 26.Pang J, Chen Y, Fan Z, et al. High efficiency coronary MR angiography with nonrigid cardiac motion correction. Magn Reson Med. 2016;76:1345–1353. doi: 10.1002/mrm.26332. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Francois CJ, Tuite D, Deshpande V, Jerecic R, Weale P, Carr JC. Unenhanced MR angiography of the thoracic aorta: initial clinical evaluation. AJR Am J Roentgenol. 2008;190:902–906. doi: 10.2214/AJR.07.2997. [DOI] [PubMed] [Google Scholar]
- 28.Krishnam MS, Tomasian A, Malik S, Desphande V, Laub G, Ruehm SG. Image quality and diagnostic accuracy of unenhanced SSFP MR angiography compared with conventional contrast-enhanced MR angiography for the assessment of thoracic aortic diseases. Eur Radiol. 2010;20:1311–1320. doi: 10.1007/s00330-009-1672-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Potthast S, Mitsumori L, Stanescu LA, et al. Measuring aortic diameter with different MR techniques: comparison of three-dimensional (3D) navigated steady-state free-precession (SSFP), 3D contrast-enhanced magnetic resonance angiography (CE-MRA), 2D T2 black blood, and 2D cine SSFP. J Magn Reson Imaging. 2010;31:177–184. doi: 10.1002/jmri.22016. [DOI] [PubMed] [Google Scholar]
- 30.von Knobelsdorff-Brenkenhoff F, Gruettner H, Trauzeddel RF, Greiser A, Schulz-Menger J. Comparison of native high-resolution 3D and contrast-enhanced MR angiography for assessing the thoracic aorta. Eur Heart J Cardiovasc Imaging. 2014;15:651–658. doi: 10.1093/ehjci/jet263. [DOI] [PubMed] [Google Scholar]
- 31.Ruile P, Blanke P, Krauss T, et al. Pre-procedural assessment of aortic annulus dimensions for transcatheter aortic valve replacement: comparison of a non-contrast 3D MRA protocol with contrast-enhanced cardiac dual-source CT angiography. Eur Heart J Cardiovasc Imaging. 2016;17:458–466. doi: 10.1093/ehjci/jev188. [DOI] [PubMed] [Google Scholar]
- 32.Bhat H, Ge L, Nielles-Vallespin S, Zuehlsdorff S, Li D. 3D radial sampling and 3D affine transform-based respiratory motion correction technique for free-breathing whole-heart coronary MRA with 100% imaging efficiency. Magn Reson Med. 2011;65:1269–1277. doi: 10.1002/mrm.22717. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Aitken AP, Henningsson M, Botnar RM, Schaeffter T, Prieto C. 100% efficient three-dimensional coronary MR angiography with two-dimensional beat-to-beat translational and bin-to-bin affine motion correction. Magn Reson Med. 2015;74:756–764. doi: 10.1002/mrm.25460. [DOI] [PubMed] [Google Scholar]
- 34.Feng L, Coppo S, Piccini D, et al. 5D whole-heart sparse MRI. Magn Reson Med. 2018;79:826–838. doi: 10.1002/mrm.26745. [DOI] [PMC free article] [PubMed] [Google Scholar]