Skip to main content
Journal of Foot and Ankle Research logoLink to Journal of Foot and Ankle Research
. 2019 Aug 2;12:39. doi: 10.1186/s13047-019-0350-1

The reliability of the ankle brachial index: a systematic review

Sarah Casey 1,, Sean Lanting 1, Christopher Oldmeadow 2, Vivienne Chuter 1
PMCID: PMC6679535  PMID: 31388357

Abstract

Background

The ankle brachial index (ABI) is widely used in clinical practice as a non-invasive method to detect the presence and severity of peripheral arterial disease (PAD). Current guidelines suggest that it should be used to monitor potential progression of PAD in affected individuals. As such, it is important that the test is reliable when used for repeated measurements, by the same or different health practitioners. This systematic review aims to examine the literature to evaluate the inter- and intra-rater reliability of the ABI.

Methods

A systematic search of MEDLINE, EMBASE and CINAHL Complete was conducted to 20 January 2019. Two authors independently reviewed and selected relevant studies and extracted the data. Methodological quality was determined using the Quality Appraisal of Reliability (QAREL) Checklist.

Results

Fifteen studies of ABI reliability in a range of patient populations were identified as suitable for inclusion in the review: seven considered inter-rater reliability, four intra-rater reliability, and four studies evaluated both inter- and intra-rater reliability. Inter-rater reliability was found to be highly variable, with intraclass correlation coefficients (ICC’s) ranging from poor to excellent (ICC 0.42–1.00), while intra-rater also demonstrated considerable variation, with ICCs from 0.42–0.98. Meta-analysis was not possible due to the lack of statistical information reported.

Conclusions

Results of included studies suggest the inter- and intra-tester reliability of the ABI is acceptable. However, inconsistencies in obtaining systolic pressure measurements, calculating ABI values, and incomplete reporting of methodologies and statistical analysis make it difficult to determine the validity of the results of included studies. Further research, with more consistent reliability methodology, statistical analysis and reporting conducted in populations at risk of PAD is needed to conclusively determine the ABI reliability.

Keywords: Ankle brachial index, Peripheral arterial disease, Lower extremity artery disease, Repeatability, Reproducibility, Reliability

Introduction/background

Peripheral arterial disease (PAD) describes the process of progressive atherosclerosis affecting arteries, most frequently in the lower limb. The prevalence in the general population has been estimated at up to 19% in people over the age of 55 years [1], with incidence increasing with advancing age and in the presence of smoking, inactivity and obesity [1, 2]. The presence of PAD is associated with increased risk of mortality and morbidity from cardiac atherosclerosis [2], and, in its advanced stages, can result in lower extremity ulceration and amputation [3]. Diabetes mellitus is an independent risk factor for the development of PAD [4], and in people with diabetes, atherosclerotic plaques tend to have a more distal and diffuse distribution and there is a more aggressive disease presentation [5].

Due to the high risk of concurrent cardiovascular morbidity, mortality and lower limb complications associated with PAD, accurate and reliable diagnostic testing methods are required for screening and ongoing monitoring [2, 6]. Early detection of PAD allows for intervention and management to reduce the risk of mortality and morbidity related to atherosclerosis (lifestyle modification, pharmacotherapy, e.g. statins, antiplatelets, and measures to address systemic risk factors such as hypertension or diabetes) [7]. Current recommendations for non-invasive lower limb vascular assessment include using the ankle-brachial index (ABI) as an objective measurement of peripheral blood flow [7, 8]. The ABI represents the ratio of ankle to brachial systolic pressure and is recommended to be calculated by dividing the higher systolic pressure of the dorsalis pedis and tibialis posterior vessels at the ankle with the higher of the systolic pressures measured in the brachial artery in both arms [7, 8].

The ABI is widely used to screen for PAD in different clinical settings and by different health professionals, from general medical practitioners to specialist vascular technicians [9, 10]. Reliability of the test for accurate ongoing monitoring of lower limb vascular status has the potential to be affected by a number of factors. As an operator-dependent test, this includes the experience and skills of the clinician, particularly as multiple clinicians are frequently involved in ongoing monitoring measurements [11, 12]. There are also a number of types of equipment (e.g. automated versus manual) and methods used to measure ankle and arm blood pressures (e.g. stethoscope, Doppler, photoplethysmography probe), with variable findings as to whether the results are interchangeable [1316]. The pre-test protocol and test environment have also been demonstrated to affect the resting ABI at measurement, with variations in body position [17], recency of tobacco smoking, caffeine intake [18, 19] and exercise [20, 21], and pre-measurement rest time [22] all likely to introduce error to the measurement and affect the test-retest reliability.

Objectives

Given that the ABI is the recommended method for screening for the presence and progression of PAD, it is important that it is reliable. Therefore, the aim of this review was to systematically evaluate the literature to determine the inter- and intra-rater reliability of the ABI in adults.

Methods

Search strategy

A search of relevant biomedical journal databases from the University of Newcastle library website was performed to identify studies that consider the reliability of ABI measurement from database inception to January 2019 using MEDLINE (1946+), EMBASE (1947+), and CINAHL Complete. Truncated versions of some search terms were used to ensure that relevant studies were included (Table 1).

Table 1.

Search terms: searches were limited to human studies

Databases: MEDLINE (1946+), EMBASE (1947+), and CINAHL Complete
1 Ankle brachial pressure
2 Ankle arm pressure
3 Ankle brachial ind*
4 Reliab*
5 Consistenc*
6 Accura*
7 Reproduc*
8 Repeat*
9 Agreement
10 Precision
11 1 or 2 or 3 AND 4 or 5 or 6 or 7 or 8 or 9 or 10

Inclusion and exclusion criteria

The review was conducted with reference to the Preferred Reporting Items for Systematic Review and Meta-Analysis (PRISMA) statement [23]. The following criteria had to be satisfied for inclusion in the review: published original research evaluating the reliability of the ABI in adults. Studies were excluded if the test-retest time frame made it likely that results may be affected by disease progression e.g. > 12 months. No language restrictions were applied to the database searches.

Other sources

Hand searching of the reference list of appropriate articles was also conducted.

Data collection and analysis

All abstracts obtained were assessed independently by SC and SL for inclusion. There were no instances of disagreement between reviewers, so arbitration by a third person (VC) was not necessary. Data extraction was performed by SC and SL. It was pre-determined that a meta-analysis of reliability outcomes for inter- and intra-rater reliability would be conducted provided there were sufficient studies that report the estimator of interest, and a measure of uncertainty for this estimator (e.g. standard error, 95% confidence interval, non-truncated p-value). Given the expectation for a high degree of study heterogeneity, we believed a fixed effect meta-analysis would generally not be appropriate so we aimed to only pool estimates using a random effects approach provided there were at least 5 studies [24].

Methodological quality assessment

The studies that met the inclusion criteria were appraised for risk of bias using the Quality Appraisal of Reliability (QAREL) Checklist and qualitative methodological assessment [25]. All full-text papers were assessed for methodological quality independently by two reviewers (SC/SL), and as there were no disagreements arbitration by a third reviewer (VC) was not necessary.

Results

A total of 1703 articles were retrieved, of which 36 were identified as suitable for full-text review. Twenty-one papers were excluded based on the exclusion criteria (Fig. 1): 10 papers reported comparison of methods [2635], five studies did not report measures of reliability [3640], two studies compared raters’ experience [41, 42] and one reported a novel trial design, for which the reliability results were duplicated in another included paper [43]. One paper used measures repeated at up to 365 days apart, with a mean time between measures of 228 days, which is long enough to encompass changes attributable to progression of PAD [44]. Two papers were conference abstracts, for which the full text could not be obtained as the authors did not respond to a request for further information [40, 45].

Fig. 1.

Fig. 1

PRISMA flow chart

Of the included papers, seven measured inter-tester reliability [12, 4651], four assessed intra-tester reliability [5255], and four considered both inter- and intra-tester reliability [13, 16, 56, 57].

Characteristics and overview of included studies

The 15 studies in this review included a total of 916 participants, with data collected from a combination of one and both lower limbs (1396 limbs in total). Two studies did not state the number of limbs included [52, 53]. Eleven studies assessed inter-rater reliability [12, 13, 16, 4650, 56, 57], and eight studies reported intra-rater reliability [13, 16, 5257]. The characteristics of included studies are described in Table 2. Eleven studies reported participants’ gender, with more men (n = 416, 56.4%) overall than women, whilst gender was unreported in four studies [12, 46, 49, 50]. Most of the studies included predominantly older participants (age range (41–92 years) [12, 13, 16, 4749, 51, 5355, 57], however two studies recruited only younger adults (age range 22–30 years) [46, 56], one study included 18–80 year olds [52] and one study did not report participants’ ages [50]. The majority of studies [12, 4751, 55, 57] included only participants with suspected PAD, or risk factors for atherosclerosis; three studied a mixed population including those without risk factors or clinical indicators of PAD [13, 16, 52]; two studies included only participants with diabetes [53, 54], and two studies included only healthy individuals [46, 56].

Table 2.

Participant characteristics and reliability measure

Reference Number (n) Gender (M,F) Age (years) Height (cm) Weight (kg) DM & control DM duration Medical history PAD Reliability measure
Inter-rater reliability (n = 11)
 Aboyans et al. (2008) [13] 54 28, 26 52.8 ± 17.1 NR NR DM 35.2% control NR NR HT 38.9%, DL 38.9%, CAD 33.3% CVD 11.1% TS 22.2%, 19 IC, 25 RF, 10 healthy

ICC

Doppler: Inter: 0.79 (0.70–0.85)

Pulse: inter: 0.40 (0.5–0.57)

Auto ABI inter: 0.44 (0.27–0.58)

 Alvaro-Afonso et al. (2018) [51] 21 15, 6 67 ± 8.7 NR NR All DM2: control NR NR HT 19, DL 17, NEU 18, NEP 1, TS 6 NR

Kappa coefficient

Normal: 0.4 (p < 0.001), PAD 0.7 (p < 0.001), MAC 0.43 (p < 0.001)

 Chesbro et al. (2011) [46] 20 NR 22–30 NR NR NR NR Healthy volunteers only None

ICC

Doppler: R: 1.00 (95%CI 0.999–1.00, p < .001); L: 0.99 (0.997–1.00, p < .001)

 Chesbro et al. (2013) [56] 10 5, 5 NR NR NR NR NR ‘Healthy young adults’ None

ICC

Vascular Cuff

Right: Trial 1: 0.853; Trial 2: 0.898

Left: Trial 1: 0.448; Trial 2: 0.938

Standard Cuff

Right: Trial 1: 0.902; Trial 2:0.817

Left: Trial 1: 0.826; Trial 2: 0.867

 de Graff et al. (2001) [57] 54 31, 23 66 ± 12 NR NR DM 36% control NR NR HT 43%, Dl 35%, CAD 36%, CVD 20%, Suspected

Repeatability Coefficient/ICC

Inter: day 20 / 0.92

Inter: week 27 / 0.87

 Georgakarakos et al. (2013) [47] 18 12, 6 54–74 NR NR 9 DM: 3 oral, 6 insulin NR HT 15, DL 9, TS 4 All PAD

Mean, Standard Error, t-test

PAD: 0.77 ± 0.19, p = 0.95

No PAD: 1.37 ± 0.12, p < 0.0001

Severe PAD: 0.23 ± 0.07, p = 0.0002

 Holland-Letz et al. (2007) [16] 108 50, 58 68.1 ± 1.5 NR BMI 29 ± 4.3 DM 15.7% control NR NR HT 58.1% DL 54.8%, TS 9.2% current, 40.7% ex-, Hx vasc surg 6 subjects 68.1 ± 1.5

ICC for inter-observer: 0.423

Inter-ob SD 0.103

 Jaffer et al. (2008) [50] 25 NR NR NR NR NR NR NR All suspected PAD Pearson CC r = 0.516 (p < 0.001)
 Langen et al. (2009) [48] 20 11, 9 41–75 NR NR NR NR NR All IC Symptoms Interobserver variability: 10% (SD 0.8)
 Mätzke et al. (2003) [12] 30 18, 15 (no of limbs) 26 > 65 yrs., 7 ≤ 65 yrs NR NR 8 DM, control NR NR NR Ischaemic pressure lesion or rest pain Coefficient of variation: 3.2
 Span et al. (2016) [49] 136 NR 64 ± 7.8 NR NR 19 (14%) control not reported NR HT66, DL 58, TS 22 current, 39 ex- RF or IC

Coefficient of variation

Doppler: 5.9% R & L legs

Auto: Right 3.2% Left 3.5%

Intra-rater reliability (n = 8)
 Aboyans et al. (2008) [13] 54 28, 26 52.8 ± 17.1 NR NR DM 35.2% control NR NR HT 38.9%, DL 38.9%, CAD 33.3% CVD 11.1% TS 22.2%, 19 IC, 25 RF, 10 healthy

ICC

Doppler: Intra: 0.89 (0.84–0.92)

Pulse: Intra: 0.60 (0.44–0.73),

 Chesbro et al. (2013) [56] 10 5, 5 NR NR NR NR NR ‘Healthy young adults’ None

ICC- Intra-rater

Vascular Cuff

Rater 1 R: 0.750; Rater 1 R: 0.696

Rater 2 R: 0.551; Rater 2 L: 0.869

Standard Cuff

Rater 1 R: 0.628; Rater 1 L: 0.420

Rater 2 R: 0.620; Rater 2 L:0.585

 de Graff et al. (2001) [57] 54 31, 23 66 ± 12 NR NR DM 36% control NR NR HT 43%, Dl 35%, CAD 36%, CVD 20%, Suspected

Repeatability Coefficient/ICC

Intra: day 9 / 0.98

Intra: week 22 / 0.89

 Demir et al. (2016) [52] 161 87, 74 52.03 ± 18.99 165.12 ± 8.88 75.61 ± 13.4 DM NR NR HT 62.7%, DL 46.6%, TS 29.8%, Mixed population

ICC

Single measurement 0.808

Mean: 0.927

 Faccenda et al. (1988) [53] 36 28,8 56 ± 11 NR NR All DM1 NR No other hx reported NR Coefficient of variation 8%
 Holland-Letz et al. (2007) [16] 108 50, 58 68.1 ± 1.5 NR BMI 29 ± 4.3 DM 15.7% control NR NR HT 58.1% DL 54.8%, TS 9.2% current, 40.7% ex-, Hx vasc surg 6 subjects 68.1 ± 1.5 Intra-observer Variance: 0.008, SD 0.87 [0.081; 0.095]
 Millen et al. (2018) [54] 66 51, 15 69.5 ± 12 yrs. (range 35–92) NR NR 4 DM1, 14 DM2 HT 79%, DL 68%, CAD 44%, CVD 17%, TS 15% current, 59% ex- 36 IC, 4 rest pain

Coefficient of variation: Dopplex Ability: 9.65 ± 12%

Parks Flo-lab: 4.95 ± 3%

 Rosenbaum et al. (2012) [55] 157 80, 77 59.1 ± 13.2 NR NR 35 DM, control NR NR HT88, DL 103, CAD 14, CVD 2, TS 27 current, 49 ex- 11 PAD, all RF or IC

Coefficient of variation: Dopplex Ability: 9.65 ± 12%

Parks Flo-lab: 4.95 ± 3%

NR Not reported, DM Diabetes, HT Hypertension, DL Dyslipidaemia, CAD Coronary artery disease, CVD Cerebrovascular disease, NEU Neuropathy, NEP Nephropathy, TS smoking, IC Intermittent claudication, RF risk factors, PAD Peripheral arterial disease

There was little consistency in the training and qualifications of the raters used, with experience ranging from students [46, 47] to experienced vascular technicians and/or vascular specialist doctors [12, 13, 48, 54, 57]. Six studies did not state the background of the personnel performing the test [49, 5153, 55, 56]. The majority of the studies used Doppler and manual sphygmomanometer to measure systolic blood pressures; [12, 13, 16, 4649, 5153, 56] however three studies used an automated device to obtain some or all of the pressure readings [54, 55, 57] and one study did not report the method used [50]. The reported pre-measurement rest time varied from five minutes [55] to 15 min [48], with seven studies not reporting a period of rest before testing commenced [44, 47, 49, 50, 5254]. The time between repeat testing varied from five minutes [46, 56] to 4 weeks [52]; six studies did not report time between repeated measures [12, 4951, 54, 55]. Several different methods were used to calculate the ABI. The majority of studies [4749, 51, 52, 56, 57] divided the highest ankle pressure by the higher brachial pressure measurement, two [13, 16] used the highest ankle pressure and the mean brachial value, and one used the lowest ankle pressure and the highest brachial pressure [55]. One study used a fully automated device that calculated the ABI value [54], and four did not state how the ABI was calculated [12, 46, 50, 53].

Methodological quality

The quality of studies was variable with regard to reported blinding of raters, order of examination and the time between repeated measurements, with no study clearly addressing all of these variables. While most studies used appropriate statistical measures of agreement, reporting of results was frequently incomplete and the true extent of reliability could not be determined (Table 3).

Table 3.

QAREL Checklist

Item Aboyans et al. (2008) [13] Alvaro-Afonso et al. (2018) [51] Chesbro et al. (2011) [46] Chesbro et al. (2013) [56] de Graaff et al. 2001) [57] Demir et al. (2016) [52] Faccenda et al. (1988) [53] Georgakarakos et al. (2013) [47] Holland-Letz et al. (2007) [16] Jaffer et al. (2008) [50] Langen et al (2009) [48] Matzke et al. (2003) [12] Millen et al. (2018) [54] Rosenbaum et al. (2012) [55] Span et al. (2016) [49]
 1. Was the test evaluated in a sample of subjects who were representative of those to whom the authors intended the results to be applied? Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y
 2. Was the test performed by raters who were representative of those to whom the authors intended the results to be applied? Y U Y Y Y Y U Y Y Y Y Y Y Y Y
 3. Were raters blinded to the findings of other raters during the study? Y U Y Y Y NA NA Y U U Y Y Y NA Y
 4. Were raters blinded to their own prior findings of the test under evaluation? Y NA U U U U Y NA Y NA NA NA NA U Y
 5. Were raters blinded to the results of the reference standard for the target disorder (or variable) being evaluated? NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA
 6. Were raters blinded to clinical information that was not intended to be provided as part of the testing procedure or study design? U U U U U U U U U U U Y N U U
 7. Were raters blinded to additional cues that were not part of the test? U U U U U U U U U U U U N U U
 8. Was the order of examination varied? U U Y Y U N U U U U N U Y U U
 9. Was the time interval between repeated measurements compatible with the stability (or theoretical stability) of the variable being measured? U Y Y Y Y Y Y Y Y U Y U Y U U
 10. Was the test applied correctly and interpreted appropriately? N Y Y Y Y Y N Y N U Y Y Y N Y
 11. Were appropriate statistical measures of agreement used? Y Y Y Y Y Y P Y Y Y U N Y Y Y

Y Yes, N No, U Unclear, P Partly, NA Not applicable

Meta-analysis

A number of the eligible papers identified lacked sufficient data relating to the main outcomes to allow for inclusion in a meta-analysis. For example, the paper by Chesbro et al., [46] provided no details on the intra-rater reliability of measurements taken using a Doppler, which was the main outcome being assessed in this review. Similarly, papers by De Graaff et al., [57] and Demir et al. [52] detailed no measure of variability for the intraclass correlation coefficients (ICC) reported, which is required when pooling results in a meta-analysis. It is not clear whether Chesbro et al. [46, 56] used data from the same population in both studies, and the authors did not respond to a request for clarification. Finally, for the paper by Aboyans et al., [13] the type of ICC calculated was not reported, and while pooling of this data would be possible, understanding which ICC was used is preferred to allow for accurate and appropriate calculation of the standard error. As there were only a small number of eligible papers identified we would require data from all articles to allow for appropriate pooling of ICCs. Thus, as a consequence of the small number of papers reviewed and insufficient data reported by several of the papers it was not possible to conduct a meta-analysis as part of this review. None of the authors responded to requests for missing data. A narrative review of results is presented instead.

Inter-rater reliability

Inter-rater reliability results are included in Table 2. Statistical methods for calculating reliability were inconsistent. Of the eleven included studies, five reported levels of agreement with ICCs [13, 16, 46, 56, 57]. Of these, only three [13, 46, 56] reported 95% confidence intervals, which limits the interpretation of reliability in the context of clinically meaningful results. Based on ICC values alone, inter-rater reliability was highly variable, ranging from poor (ICC: 0.42) [16], to excellent (ICC: 1.0) [46].

Other estimates of reliability reported in included studies were coefficient of variation between raters [12, 49] (ranging from 3.2 to 5.9%), inter-observer reliability of 10% for raters [48], and a moderate Pearson’s correlation coefficient of 0.52 in a population with suspected PAD [50].

Of the remaining studies, one demonstrated statistically significant differences in ABI between raters in a population with severe PAD and in those with no disease, which did not occur in those participants with mild to moderate PAD [47], suggesting increased reliability with this disease state. In contrast, another paper reported Kappa coefficients of 0.4 (low agreement) for healthy limbs, 0.7 (good agreement) for limbs with PAD, and 0.43 (moderate agreement) for limbs with medial arterial calcification (MAC) (p < 0.001 for all values) [51].

Intra-rater reliability

Intra-rater reliability results are included in Table 2. Various methods of calculating reliability were used. Of the eight included studies, four reported ICCs [13, 52, 56, 57], with ICC values ranging from poor (ICC: 0.42) [56] to excellent (0.98) [57]. Interpretation of the results was limited again by the fact that not all studies reported 95% confidence intervals, with only two articles having done so. [13, 56]. Other estimates of reliability included coefficient of variation [5355] (range 4.95% [54] – 15.8% [55]), and an intra-observer variance of 8% [16].

Discussion

The findings of this review are that the inter- and intra- tester reliability of the ABI across a number of mixed populations appears to be acceptable, however statistical tests of reliability in included papers were heterogeneous and levels of statistical reporting were inconsistent and incomplete. This makes interpretation of the reliability of the ABI in the context of clinical detection, evaluation and ongoing monitoring of peripheral arterial supply challenging, and prevented meta-analysis. For example, where studies lack 95% confidence intervals for ICCs, the validity of interpretation of the value is reduced as it fails to provide the lowest level of reliability that it represents. Similarly for coefficient or estimate of variation, values between 3.2 and 15.8% were reported.. Whilst this is considered an acceptable level of variation for many clinical tests, for the ABI it can represent a range of values that may indicate both normal and pathological results; which could reduce the ability of ABI to reliably determine the presence and extent of PAD. For example, assuming a variation of 15%, an ABI of 1.0 (which is considered ‘borderline’ when ABI is used as a screening tool [6]) could represent a true value between 0.85 (indicative of PAD) and 1.15 (‘normal’).

Further complicating the interpretation and generalisability of the inter- and intra-rater reliability results of included studies was the heterogeneity of participant populations. Whilst the majority of studies included older people with PAD risk factors or suspected PAD, three studies also included healthy participants [13, 16, 52], and two used an exclusively young and healthy population [46, 56]. In clinical practice, ABI is used to evaluate peripheral arterial supply in people with risk factors for atherosclerosis, and in those with clinical signs and symptoms of PAD. The variation in the disease status of participants across the studies included in this review provides some difficulty in evaluating how the studies’ findings apply to the people in whom the ABI would clinically be used. The study that reported near-perfect inter- and intra- tester reliability included only healthy individuals under the age of 30 [56]. This population would not typically undergo vascular screening, and the results obtained do not indicate the ability of the ABI to perform reliably in the presence of pathology where the result is likely to be lower and therefore change in result indicative of worsening pathology is likely to be small. In contrast, inter-tester and intra-tester reliability was found to be poor in several populations in which this test is recommended including people with diabetes and without MAC, [51] and older people with risk factors for PAD [16].

Methodological differences between studies is also likely to have contributed to variable reliability outcomes, with automated oscillometric devices demonstrating marginally better reliability than manual assessment using Doppler [49, 55], while Doppler evaluation was found to be more reliable than the use of pulse palpation [13] or stethoscope [46]. Higher ABI reliability was found in more experienced raters [47]. Whilst most of the studies reported that participants rested for 5–15 min prior to testing [12, 13, 16, 46, 48, 51, 5557], six studies did not describe any pre-test preparation [47, 49, 50, 5254], and only one paper took steps to ensure that participants did not consume alcohol, caffeine or tobacco (which are known to affect blood pressure) in the two hours prior to testing which may have affected measurements, particularly when taken across two different testing sessions [55]. This lack of reporting of the methodology used to obtain systolic blood pressure measurements makes it difficult to compare results across the included studies as it is unknown how much external factors are likely to contribute measurement variability.

Two papers identified the presence of diabetes mellitus as a factor that may affect reliability of the ABI [12, 51], however only one study included a large enough sample of this cohort to perform statistical tests [51]. This study, which used only participants with diabetes, reported the Kappa coefficient for inter-tester measures for participants classed as having PAD or not, rather than performing ICCs on the measures obtained. The authors reported ‘good’ reproducibility of the ABI (Κ 0.7) in people classified by their ABI measurement as having PAD, but low reproducibility in those without PAD and in those with MAC. Previous research has also shown that people with diabetes demonstrate a different response to pre-measurement rest [22], and that brachial blood pressure measurement is also less reliable in these individuals [58]. Diabetes-related autonomic neuropathy has been shown to affect blood pressure regulation, with a lack of vasoconstriction arising from reduced sympathetic input, particularly in response to changes in temperature and position [59, 60].

Limitations

While the search methods employed in this study were designed to be robust, there may be some evidence that was not captured, for example unpublished data. Further limitations to this study are the inability to perform meta-analysis in order to obtain a quantitative analysis of the available reliability data for the ABI, and the inability perform any sub-analyses relating to individual populations such as those with diabetes, or methods of measurement such as automated or manual methods. Furthermore, there has been some disagreement in the literature about which pressure measurement should be used to calculate the ABI [61, 62], with no studies exploring the effect of calculation method on reliability. However, the method of calculation cannot be excluded as a factor affecting reliability that has not been considered by this review.

Conclusion

Results of included studies suggest the inter- and intra-tester reliability of the ABI is acceptable. However, inconsistencies in obtaining systolic pressure measurements, calculating ABI values, and incomplete reporting of methodologies and statistical analysis make it difficult to determine the validity of the results of included studies. Further research of ABI reliability using a more consistent approach to study design and implementation and more detailed reporting of results in populations with vascular pathology and at risk of PAD is required. Based on current available data clinicians should ensure they interpret ABI results in the context of other vascular assessment findings, and patient management is not based upon this measurement alone.

Acknowledgements

Not applicable.

Availability of data and material

All data generated or analysed during this study are available in this article.

Abbreviations

ABI

ankle brachial index

IC

intermittent claudication

ICC

intraclass correlation coefficient

MAC

medial arterial calcification

PAD

peripheral arterial disease

QAREL

Quality Appraisal of Reliability

Authors’ contributions

VC and SC conceived the study. Database searches and data extraction was conducted by SC and SL. Statistical evaluation was performed by CO. All authors contributed to the manuscript and approved the final manuscript.

Funding

Funding sources include the national Research Training Scheme (Australia).

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Contributor Information

Sarah Casey, Email: Sarah.Casey@newcastle.edu.au.

Sean Lanting, Email: Sean.Lanting@newcastle.edu.au.

Christopher Oldmeadow, Email: Christopher.Oldmeadow@hmri.org.au.

Vivienne Chuter, Email: Vivienne.Chuter@newcastle.edu.au.

References

  • 1.Meijer WT, Hoes AW, Rutgers D, Bots ML, Hofman A, Grobbee DE. Peripheral arterial disease in the elderly. Rotterdam Stud. 1998;18(2):185–192. doi: 10.1161/01.atv.18.2.185. [DOI] [PubMed] [Google Scholar]
  • 2.Norman PE, Eikelboom JW, Hankey GJ. Peripheral arterial disease: prognostic significance and prevention of atherothrombotic complications. Med J Aust. 2004;181(3):150–154. doi: 10.5694/j.1326-5377.2004.tb06206.x. [DOI] [PubMed] [Google Scholar]
  • 3.Bonham PA. Get the LEAD out: noninvasive assessment for lower extremity arterial disease using ankle brachial index and toe brachial index measurements. J Wound Ostomy and Continence Nurs. 2006;33(1):30–41. doi: 10.1097/00152192-200601000-00004. [DOI] [PubMed] [Google Scholar]
  • 4.American Diabetes Association Peripheral arterial disease in people with diabetes. Diabetes Care. 2003;26(12):3333–3341. doi: 10.2337/diacare.26.12.3333. [DOI] [PubMed] [Google Scholar]
  • 5.Jude EB, Oyibo SO, Chalmers N, Boulton AJM. Peripheral arterial disease in diabetic and nondiabetic patients: a comparison of severity and outcome. Diabetes Care. 2001;24(8):1433–1437. doi: 10.2337/diacare.24.8.1433. [DOI] [PubMed] [Google Scholar]
  • 6.Aboyans V, Criqui MH, Abraham P, Allison MA, Creager MA, Diehm C, et al. Measurement and interpretation of the ankle-brachial index. A Scientific Statement From the American Heart Association. Circulation. 2012;126(24):2890–2909. doi: 10.1161/CIR.0b013e318276fbcb. [DOI] [PubMed] [Google Scholar]
  • 7.Gerhard-Herman MD, Gornik HL, Barrett C, Barshes NR, Corriere MA, Drachman DE, et al. 2016 AHA/ACC guideline on the management of patients with lower extremity peripheral artery disease: executive summary. A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. 2016. [Google Scholar]
  • 8.Norgren L, Hiatt WR, Dormandy JA, Nehler MR, Harris KA, Fowkes FGR. Inter-society consensus for the Management of Peripheral Arterial Disease (TASC II) J Vasc Surg. 2007;45(1):S5–S67. doi: 10.1016/j.jvs.2006.12.037. [DOI] [PubMed] [Google Scholar]
  • 9.Haigh K, Bingley J, Golledge J, Walker P. Peripheral arterial disease -screening in general practice. Aust Fam Physician. 2013;42:391–395. [PubMed] [Google Scholar]
  • 10.Al-Qaisi M, Nott DM, King DH, Kaddoura S. Ankle brachial pressure index (ABPI): an update for practitioners. Vasc Health Risk Manag. 2009;5:833–841. doi: 10.2147/vhrm.s6759. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Nicolai SP, Kruidenier LM, Rouwet EV, Bartelink ML, Prins MH, Teijink JA. Ankle brachial index measurement in primary care: are we doing it right? Br J Gen Pract. 2009;59(563):422–427. doi: 10.3399/bjgp09X420932. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Matzke S, Franckena M, Alback A, Railo M, Lepantalo M. Ankle brachial index measurements in critical leg ischaemia - the influence of experience on reproducibility. Scand J Surg. 2003;92:144–147. doi: 10.1177/145749690309200206. [DOI] [PubMed] [Google Scholar]
  • 13.Aboyans V, Lacroix P, Doucet S, Preux PM, Criqui MH, Laskar M. Diagnosis of peripheral arterial disease in general practice: can the ankle-brachial index be measured either by pulse palpation or an automatic blood pressure device? Int J Clin Pract. 2008;62. [DOI] [PubMed]
  • 14.Staessen JA, Li Y, Wang J-G. Comparison of three measures of the ankle-brachial blood pressure index in a general population. Hypertens Res. 2007;30:555. doi: 10.1291/hypres.30.555. [DOI] [PubMed] [Google Scholar]
  • 15.Ramanathan A, Conaghan PJ, Jenkinson AD, Bishop CR. Comparison of ankle-brachial pressure index measurements using an automated oscillometric device with the standard doppler ultrasound technique. ANZ J Surg. 2003;73(3):105–108. doi: 10.1046/j.1445-2197.2003.02582.x. [DOI] [PubMed] [Google Scholar]
  • 16.Holland-Letz T, Endres HG, Biedermann S, Mahn M, Kunert J, Groh S, et al. Reproducibility and reliability of the ankle—brachial index as assessed by vascular experts, family physicians and nurses. Vasc Med. 2007;12(2):105–112. doi: 10.1177/1358863X07077281. [DOI] [PubMed] [Google Scholar]
  • 17.Pollak EW, Chavis P, Wolfman EF. The effect of postural changes upon the ankle arterial perfusion pressure. Vasc Surg. 1976;10(4):219–222. doi: 10.1177/153857447601000404. [DOI] [PubMed] [Google Scholar]
  • 18.Pilli R, Naidu M, Pingali U, Takallapally RK. Study of cardiovascular effects of caffeine in healthy human subjects, with special reference to pulse wave velocity using photoplethysmography. Int J Nutr Pharmacol Neurol Dis. 2012;2(3):243–250. [Google Scholar]
  • 19.Waring WS, Goudsmit J, Marwick J, Webb DJ, Maxwell SR. Acute caffeine intake influences central more than peripheral blood pressure in young adults. Am J Hypertens. 2003;16(11 Pt 1):919–924. doi: 10.1016/s0895-7061(03)01014-8. [DOI] [PubMed] [Google Scholar]
  • 20.Yataco AR, Gardner AW. Acute reduction in ankle/brachial index following smoking in chronic smokers with peripheral arterial occlusive disease. Angiology. 1999;50(5):355–360. doi: 10.1177/000331979905000501. [DOI] [PubMed] [Google Scholar]
  • 21.Carter SA. Response of ankle systolic pressure to leg exercise in mild or questionable arterial disease. N Engl J Med. 1972;287(12):578–582. doi: 10.1056/NEJM197209212871202. [DOI] [PubMed] [Google Scholar]
  • 22.Chuter VH, Casey SL. Effect of premeasurement rest time on systolic ankle pressure. J Am Heart Assoc. 2013;2(4):e000203. [DOI] [PMC free article] [PubMed]
  • 23.Liberati Alessandro, Altman Douglas G., Tetzlaff Jennifer, Mulrow Cynthia, Gøtzsche Peter C., Ioannidis John P. A., Clarke Mike, Devereaux P. J., Kleijnen Jos, Moher David. The PRISMA Statement for Reporting Systematic Reviews and Meta-Analyses of Studies That Evaluate Health Care Interventions: Explanation and Elaboration. PLoS Medicine. 2009;6(7):e1000100. doi: 10.1371/journal.pmed.1000100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Borenstein M, Hedges LV, Higgins JP, Rothstein HR. A basic introduction to fixed-effect and random-effects models for meta-analysis. Res Synth Methods. 2010;1(2):97–111. doi: 10.1002/jrsm.12. [DOI] [PubMed] [Google Scholar]
  • 25.Lucas NP, Macaskill P, Irwig I, Bogduk N. The development of a quality appraisal tool for studies of diagnostic reliability. J Clin Epidemiol. 2010;63:854–861. doi: 10.1016/j.jclinepi.2009.10.002. [DOI] [PubMed] [Google Scholar]
  • 26.Jvd S, Verbogt NP, Mulder PG, Steunenberg SL, Steunenberg BE, Lvd L. The clinical applicability of an automated plethysmographic determination of the ankle-brachial index after vascular surgery. Vascular. 2016;24(5):545–548. doi: 10.1177/1708538115625379. [DOI] [PubMed] [Google Scholar]
  • 27.Harrison ML, Lin H-F, Blakely DW, Tanaka H. Preliminary assessment of an automatic screening device for peripheral arterial disease using ankle–brachial and toe–brachial indices. Blood Press Monit. 2011;16(3):138–141. doi: 10.1097/MBP.0b013e328346a839. [DOI] [PubMed] [Google Scholar]
  • 28.Khandanpour N, Armon MP, Jennings B, Clark A, Meyer FJ. Photoplethysmography, an easy and accurate method for measuring ankle brachial pressure index: can photoplethysmography replace Doppler? Vasc Endovasc Surg. 2009;43(6):578–582. doi: 10.1177/1538574409334829. [DOI] [PubMed] [Google Scholar]
  • 29.Bonham PA, Cappuccio M, Hulsey T, Michel Y, Kelechi T, Jenkins C, et al. Are ankle and toe brachial indices (ABI-TBI) obtained by a pocket Doppler interchangeable with those obtained by standard laboratory equipment? J Wound Ostomy Continence Nurs. 2007;34(1):35–44. doi: 10.1097/00152192-200701000-00007. [DOI] [PubMed] [Google Scholar]
  • 30.Raines JK, Farrar J, Noicely K, Pena J, Davis WW, Willens HJ, et al. Ankle/brachial index in the primary care setting. Vasc Endovasc Surg. 2004;38(2):131–136. doi: 10.1177/153857440403800204. [DOI] [PubMed] [Google Scholar]
  • 31.Gestin S, Delluc A, Saliou AH, Colas A, Guéguen F, Gladu G, et al. Ankle brachial pressure index (ABPI): color-Doppler versus ultrasound Doppler correlation study in 98 patients after analysis of interobserver reproducibility. J Mal Vasc. 2012;37(4):186–194. doi: 10.1016/j.jmv.2012.05.007. [DOI] [PubMed] [Google Scholar]
  • 32.Kollias A, Xilomenos A, Protogerou A, Dimakakos E, Stergiou GS. Automated determination of the ankle-brachial index using an oscillometric blood pressure monitor: validation vs. Doppler measurement and cardiovascular risk factor profile. Hypertens Res. 2011;34(7):825–830. doi: 10.1038/hr.2011.53. [DOI] [PubMed] [Google Scholar]
  • 33.Hamel J-F, Foucaud D, Fanello S. Comparison of the automated oscillometric method with the gold standard Doppler ultrasound method to access the ankle-brachial pressure index. Angiology. 2010;61(5):487–491. doi: 10.1177/0003319709360522. [DOI] [PubMed] [Google Scholar]
  • 34.Davies JH, Williams EM. Automated plethysmographic measurement of the ankle-brachial index: a comparison with the doppler ultrasound method. Hypertens Res. 2016;39(2):100–106. doi: 10.1038/hr.2015.124. [DOI] [PubMed] [Google Scholar]
  • 35.Bundó M, Urrea M, Muñoz-Ortíz L, Pérez C, Llussà J, Forés R, et al. Measurement of the ankle brachial index with a non-mercury sphygmomanometer in diabetic patients: a concordance study. BMC Cardiovasc Disord. 2013;13(1):1–7. doi: 10.1186/1471-2261-13-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Fowkes FG, Housley E, Macintyre CC, Prescott RJ, Ruckley CV. Variability of ankle and brachial systolic pressures in the measurement of atherosclerotic peripheral arterial disease. J Epidemiol Community Health. 1988;42(2):128–133. doi: 10.1136/jech.42.2.128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Ena J, Lozano T, Verdú G, Argente CR, González VL. Accuracy of ankle–brachial index obtained by automated blood pressure measuring devices in patients with diabetes mellitus. Diabetes Res Clin Pract. 2011;92(3):329–336. doi: 10.1016/j.diabres.2011.02.015. [DOI] [PubMed] [Google Scholar]
  • 38.Chongthawonsatid S, Dutsadeevettakul S. Validity and reliability of the ankle-brachial index by oscillometric blood pressure and automated ankle-brachial index. J Res Med Sci. 2017;22:44. doi: 10.4103/jrms.JRMS_728_16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Alzamora MT, Baena-Díez JM, Sorribes M, Forés R, Toran P, Vicheto M. Peripheral Arterial Disease study (PERART): prevalence and predictive values of asymptomatic peripheral arterial occlusive disease related to cardiovascular morbidity and mortality. BMC Publ Health. 2007;7. 10.1186/1471-2458-7-348. [DOI] [PMC free article] [PubMed]
  • 40.Bakalakou K, Marinakos A, Nouli A, Taxiarchou E, Margariti C, Papanikitas K, et al. Efficacy of automatic blood pressure device to determine reliably the ankle brachial index (ABI) Eur J Intern Med. 2011;22:S9. [Google Scholar]
  • 41.Monti M, Calanca L, Alatri A, Mazzolai L. Accuracy of in-patients ankle-brachial index measurement by medical students. Vasa. 2016;45(1):43–48. doi: 10.1024/0301-1526/a000494. [DOI] [PubMed] [Google Scholar]
  • 42.Nexøe J, Damsbo B, Lund JO, Munck A. Measurement of blood pressure, ankle blood pressure and calculation of ankle brachial index in general practice. Fam Pract. 2012;29(3):345–351. doi: 10.1093/fampra/cmr094. [DOI] [PubMed] [Google Scholar]
  • 43.Endres H, Hucke C, Holland-Letz T, Trampisch H-J. A new efficient trial design for assessing reliability of ankle-brachial index measures by three different observer groups. BMC Cardiovasc Disord. 2006;6(1):33. doi: 10.1186/1471-2261-6-33. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Weatherley BD, Chambless LE, Heiss G, Catellier DJ, Ellison CR. The reliability of the ankle-brachial index in the atherosclerosis risk in communities (ARIC) study and the NHLBI family heart study (FHS) BMC Cardiovasc Disord. 2006;6(1):1–11. doi: 10.1186/1471-2261-6-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Balkanay OO, Arapi B, Arslan C, Gurel SA. Reliability of ankle brachial index measurement by oscillometric blood pressure monitoring device. Heart Surg Forum. 2012;15:S66. [Google Scholar]
  • 46.Chesbro SB, Asongwed ET, Brown J, John EB. Reliability of Doppler and stethoscope methods of determining systolic blood pressures: considerations for calculating an ankle-brachial index. J Natl Med Assoc. 2011;103(9):863–869. doi: 10.1016/s0027-9684(15)30441-7. [DOI] [PubMed] [Google Scholar]
  • 47.Georgakarakos E, Papadaki E, Vamvakerou V, Lytras D, Tsiokani A, Tsolakaki O, et al. Training to measure ankle–brachial index at the undergraduate level: can it be successful? Int J Low Extrem Wounds. 2013;12(2):167–171. doi: 10.1177/1534734613483768. [DOI] [PubMed] [Google Scholar]
  • 48.Hv L, Jv G, Rubbens L. Interobserver variability of ankle–brachial index measurements at rest and post exercise in patients with intermittent claudication. Vasc Med. 2009;14(3):221–226. doi: 10.1177/1358863X08101017. [DOI] [PubMed] [Google Scholar]
  • 49.Špan M, Geršak G, Millasseau SC, Meža M, Košir A. Detection of peripheral arterial disease with an improved automated device: comparison of a new oscillometric device and the standard Doppler method. Vasc Health Risk Manag. 2016;12:305–311. doi: 10.2147/VHRM.S106534. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Jaffer U, Elmagrabi AW, Cameron A, Osman I. Agreement of community-performed ankle brachial pressure indices (ABPI) with vascular laboratory performed assessment. Vasc Dis Manag. 2008;5(2):71–72. [Google Scholar]
  • 51.Alvaro-Afonso FJ, Garcia-Morales E, Molines-Barroso RJ, Garcia-Alvarez Y, Sanz-Corbalan I, Lazaro-Martinez JL. Interobserver reliability of the ankle-brachial index, toe-brachial index and distal pulse palpation in patients with diabetes. Diab Vasc Dis Res. 2018;15(4):344–347. doi: 10.1177/1479164118767599. [DOI] [PubMed] [Google Scholar]
  • 52.Demir O, Tasci I, Acikel C, Saglam K, Gezer M, Acar R, et al. Individual variations in ankle brachial index measurement among Turkish adults. Vascular. 2016;24(1):53–58. doi: 10.1177/1708538115584506. [DOI] [PubMed] [Google Scholar]
  • 53.Faccenda F, Rubba P, Vaccaro O, Carbone L, Pauciullo P, Riccardi G. Reproducibility of non-invasive vascular diagnosis in patients with diabetes mellitus. Artery. 1989;16(4):189–207. [PubMed] [Google Scholar]
  • 54.Millen Rebecca N., Thomas Kate N., Majumder Arunesh, Hill Brigid G., Van Rij Andre M., Krysa Jo. Accuracy and repeatability of the Dopplex Ability. Expert Review of Medical Devices. 2018;15(3):247–251. doi: 10.1080/17434440.2018.1431121. [DOI] [PubMed] [Google Scholar]
  • 55.Rosenbaum D, Rodriguez-Carranza S, Laroche P, Bruckert E, Giral P, Girerd X. Accuracy of the ankle-brachial index using the SCVL(®), an arm and ankle automated device with synchronized cuffs, in a population with increased cardiovascular risk. Vasc Health Risk Manag. 2012;8:239–246. doi: 10.2147/VHRM.S29405. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 56.Chesbro SB, Asongwed ET, John EB, Haile N. Reliability of ankle-brachial index measurements: a comparison of standard and vascular blood pressure cuffs. Top Geriatr Rehabil. 2013;29(3):195–202. [Google Scholar]
  • 57.de Graaff JC, Ubbink DT, Legemate DA, de Haan RJ, Jacobs MJHM. Interobserver and intraobserver reproducibility of peripheral blood and oxygen pressure measurements in the assessment of lower extremity arterial disease. J Vasc Surg. 2001;33(5):1033–1040. doi: 10.1067/mva.2001.108011. [DOI] [PubMed] [Google Scholar]
  • 58.Sonter JA, Chuter V, Casey S. Intratester and intertester reliability of toe pressure measurements in people with and without diabetes performed by podiatric physicians. J Am Podiatr Med Assoc. 2015;105(3):201–208. doi: 10.7547/0003-0538-105.3.201. [DOI] [PubMed] [Google Scholar]
  • 59.Christensen NJ. Spontaneous variations in resting blood flow, postischaemic peak flow and vibratory perception in the feet of diabetics. Diabetalogia. 1969;5(3):171–178. doi: 10.1007/BF01213675. [DOI] [PubMed] [Google Scholar]
  • 60.Hilsted J, Parving HH, Christensen NJ, Benn J, Galbo H. Hemodynamics in diabetic orthostatic hypotension. J Clin Invest. 1981;68(6):1427–1434. doi: 10.1172/JCI110394. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Schröder F, Diehm N, Kareem S, Ames M, Pira A, Zwettler U, et al. A modified calculation of ankle-brachial pressure index is far more sensitive in the detection of peripheral arterial disease. J Vasc Surg. 2006;44(3):531–536. doi: 10.1016/j.jvs.2006.05.016. [DOI] [PubMed] [Google Scholar]
  • 62.Niazi K, Khan TH, Easley KA. Diagnostic utility of the two methods of ankle brachial index in the detection of peripheral arterial disease of lower extremities. Catheter Cardiovasc Interv. 2006;68(5):788–792. doi: 10.1002/ccd.20906. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

All data generated or analysed during this study are available in this article.


Articles from Journal of Foot and Ankle Research are provided here courtesy of Wiley

RESOURCES