Abstract
Objectives
To apply the MB (multiband) excitation and blipped-CAIPI (blipped-controlled aliasing in parallel imaging) techniques in a spin and gradient-echo (SAGE) EPI sequence to improve the slice coverage for vessel architecture imaging (VAI).
Materials and methods
Both MB excitation and blipped-CAIPI with in-plane parallel imaging were incorporated into a gradient-echo (GE)/spin-echo (SE) EPI sequence for simultaneous tracking of the dynamic MR signal changes in both GE and SE contrasts after the injection of contrast agent. MB and singleband (SB) excitation were compared using a 20-channel head coil at 3 Tesla, and high-resolution MB VAI could be performed in 32 glioma patients.
Results
Whole-brain covered high resolution VAI can be achieved after applying multiband excitation with a factor of 2 and in-plane parallel imaging with a factor of 3. The quality of the images resulting from MB acceleration was comparable to those from the SB method: images were reconstructed without any loss of spatial resolution or severe distortions. In addition, MB and SB signal-to-noise ratios (SNR) were similar. A relative low g-factor induced from the MB acceleration method was achieved after using a blipped-CAIPI technique (1.35 for GE and 1.33 for SE imaging). Performing quantitative VAI, we found that, among all VAI parametric maps, microvessel type indicator (MTI), distance map (I) and vascular-induced bolus peak-time shift (VIPS) were highly correlated. Likewise, VAI parametric maps of slope, slope length and short axis were highly correlated.
Conclusions
Multiband accelerated SAGE successfully doubles the number of readout slices in the same measurement time when compared to conventional readout sequences. The corresponding VAI parametric maps provide insights into the complexity and heterogeneity of vascular changes in glioma.
Introduction
Vessel architecture imaging (VAI) MRI is a recently coined term for a technique that noninvasively measures parameters who describe the structural heterogeneity of brain microvasculature [1–4]. The technique uses a temporal shift that occurs between the time-resolved curves of gradient-echo (GE) relaxation rate R2* and spin-echo (SE) relaxation rate R2 during contrast bolus administration. Depending on the structural properties of the microvessels, as well as their oxygen content and spatial arrangement, the different GE and SE images produce an apparent variation in the respective MRI signal readouts [5–9]. This effect is based on the varying susceptibility differences between blood vessels and surrounding tissue, as well as on their architecture in general, i.e. their structural alignment, density and the characterization of their vessel segment succession from arterial inflow via capillaries to venous outflow. Specifically, local differences in vessel size and bolus transit time between arterial inflow and venous outflow vessels can result in a relative shift in the shape and peak positions of the two relaxation rate curves, an effect first observed and described in [1]. When visualized in a time-parameterized plot, the pairwise GE and SE data points form a vortex or hysteresis curve of a certain shape that depends on the type of vascular architecture. Naturally, the loop direction depends on the axes that define the representation plane; in this study, we always plot GE against SE values to acquire vessel vortex curves. These curves therefore follow either a clockwise or a counterclockwise direction [1, 2]. Furthermore, the shape and enclosed volume of vessel vortex curves depend on the combination of vessel type and caliber. Typically, when the vascular system contains mostly venule-like vessels, the vessel vortices follow a counterclockwise direction. In contrast, if the vascular system contains mostly arteriole-type vessels, the vessel vortices transverse in a clockwise direction [3].
Previous studies found evidence for a dominance of the counterclockwise loop in normal tissue, and a possible alternation of the loop direction in a tumor, indicating a substantial change in tumor vasculature and/or permeability [1]. The size and direction of the hysteresis loop was first analyzed on a voxel basis for a small group in brain tumor (N = 4) and stroke (N = 13) patients, to show that the dominant counterclockwise loop direction found in [1] was a result of averaging of both directions present in the tissue [2]. An observed alternation of the loop direction in tumors was reported as significant, while the one noticeable in the stroke area was not [2]. The method was also applied to a patient cohort with recurrent glioblastoma, where it was found that the alternation of the loop direction in response to therapy predicts a longer patient survival [3]. The latter study used the term “vessel vortex curve” as opposed to “hysteresis loop” [1]; to avoid confusion, we only use the first term in our study. Likewise, the term VAI was first introduced in [3] as a mapping procedure of the vessel vortex curve, while the term vessel-size imaging, first used in [10], was renamed vessel caliber MRI.
VAI was also applied to examine differences in vasculature between different glioma types and to track angiogenesis in glioma growth to provide an imaging-based tumor grade prediction for low and high grade glioma [11].
To obtain VAI parameters, a dual GE/SE EPI sequence is needed to simultaneously track dynamic signal changes in both GE and SE images during contrast bolus administration. It is therefore important to acquire GE and SE signals with a high temporal sampling rate of 1-2s, usually implemented with a short repetition time (TR). Yet, for such short TR’s, it is not possible to cover the whole brain with a combined GE and SE readout. Whole-brain coverage, however, is mandatory for MR examinations in clinical routine. Specifically, brain tumors may appear in two different regions that cannot be covered with one image slab. In addition, reduced coverage has the problem of an inconsistent coverage in a longitudinal study of growing brain tumors. On the other hand, varying coverage becomes an issue in normalization of the perfusion maps to a reference region, as a consistent reference region is more difficult to find, resulting in less repeatable measurements [12].
In this work, we use a multiband (MB) EPI accelerated VAI sequence to double the readout slices for whole-brain coverage. With the combination of blipped-controlled aliasing in parallel imaging (blipped-CAIPI), a low geometry (g)-factor penalty and high SNR can be achieved. The phases imprinted by the first EPI readout are thereby canceled by the rephasing gradients just before the refocusing pulse to easily enable in-plane parallel imaging (Fig 1).
Methods
Data evaluation was approved by the local ethics committee of the University of Heidelberg (ethics approval number: S-320/2012), Heidelberg, Germany, and the requirement for informed consent was waived. All patients had consented to the scientific use of their data with admission to our hospital. They received scans using a 20-channel head receive RF coil on a 3T Prisma Siemens scanner (Siemens Healthcare, Erlangen, Germany), following a standardized tumor protocol at the department of neuroradiology at Heidelberg University Hospital, including FLAIR image acquisition, diffusion-weighted imaging, T2-weighted imaging, vessel architectural imaging during preload injection of intravenous gadoterate meglumine (Dotarem; Guerbet, France) and subsequent dynamic-susceptibility contrast imaging during contrast bolus injection, and pre- and postcontrast T1-weighted three-dimensional magnetization-prepared rapid acquisition gradient-echo imaging, see also [13, 14]. In total, 32 patients (11 women, 21 men; mean age ± standard deviation, 52.5±17.7) with low grade (14 patients) and high grade (18 patients) gliomas were included in this study, see Table 1. HGG patients with a time difference > 2.5 years between histopathologic diagnosis and MRI exam were excluded for the statistical analysis.
Table 1. Patient characteristics.
No. | Age | Sex | Glioma WHO | Location |
---|---|---|---|---|
1 | 37 | M | II | left hemisphere |
2 | 76 | F | IV | right hemisphere |
3 | 44 | F | II | left hemisphere |
4 | 51 | M | IV | right hemisphere |
5 | 52 | M | II | left hemisphere |
6 | 70 | M | IV | multifocal, basal ganglia |
7 | 67 | M | IV | left hemisphere |
8 | 78 | M | IV | right hemisphere |
9 | 49 | M | IV | left hemisphere |
10 | 36 | M | II | left hemisphere |
11 | 85 | F | IV | left hemisphere |
12 | 38 | F | II | left hemisphere |
13 | 58 | F | II | left hemisphere |
14 | 76 | M | IV | right hemisphere |
15 | 68 | M | IV | left hemisphere |
16 | 20 | F | IV | left cerebello-pontine angle |
17 | 43 | F | IV | right hemisphere |
18 | 34 | M | II | right hemisphere |
19 | 80 | M | IV | left hemisphere |
20 | 56 | M | IV | right hemisphere |
21 | 65 | F | IV | left hemisphere |
22 | 42 | M | II | both hemispheres |
23 | 76 | F | IV | left hemisphere |
24 | 57 | M | IV | right hemisphere |
25 | 62 | M | IV | multifocal, basal ganglia |
26 | 47 | M | IV | right hemisphere |
27 | 53 | M | II | left hemisphere |
28 | 37 | F | II | left hemisphere |
29 | 36 | M | II | right hemisphere |
30 | 24 | F | II | right hemisphere |
31 | 57 | M | II | left hemisphere |
32 | 25 | M | II | right hemisphere |
Vessel architecture imaging
The MEPI-SAGE sequence is based on a clean EPI sequence; by inserting the rephasing gradient after GE EPI, a higher in-plane resolution can be easily achieved by exploiting a factor of 3 for the parallel imaging technique. Without the rephasing gradient, the actual position in k-space is flipped to the diagonal position of the k-space after application of a 180° pulse. However, by using parallel imaging acceleration, the start point and end point in k-space are not point-symmetric to the k-space center, so that the k-space position before the second EPI readout train, which depends on the acceleration factor, needs to be corrected. To ensure correct data acquisition, and to allow the usage of a standard EPI module, the phases produced by the EPI readout were canceled just before the refocusing pulse.
By using readouts of dual GE/SE 2D EPI, 60 measurements including 10 baseline measurements were obtained in a time of 1.5 minutes. Explicit sequence parameters were as follows: TE (GE/SE) = 22/90 ms, MB factor = 2, parallel imaging factor = 3, dim = 120×120, resolution = 2×2 mm2, TR = 1.5s. The slice thickness is 4.5 mm, distance factor is 20% (0.9mm) and slice number is 24. The total coverage of MB EPI is 13 cm. VAI examinations were performed with administration of 0.1 mmol/kg-bodyweight gadoterate meglumine (Dotarem, Guerbet) at a rate of 4 ml/s using an MR-compatible injector. A 20 ml bolus of saline was injected subsequently at the same rate. Singleband (SB) acquisition at the same slice position without injection of contrast agent was performed after MB acquisition. The slice number is 12 and the total coverage of SB EPI is 6.5 cm.
Data processing
Before VAI analysis, motion correction including realignment and reslicing was achieved by using SPM (Wellcome Trust Center for Neuroimaging, UCL, London, UK). The relaxation rates ΔR(t)2,SE and ΔR(t)2,GE were calculated from the SE and GE data based on the following equation:
(1) |
where XE stands for SE or GE, respectively. TE is the echo time, S0 is the baseline signal from averaging of the first 10 signals before injection of preload contrast agent bolus, and S(t) represents the signal during the bolus passage of the VAI sequence. Only those voxels, whose peak signal, p, during preload contrast administration was larger than two times the sum of averaged baseline signal, μ, and baseline standard deviation, σ, or: p > 2[μ + σ], were included in further analysis. All other voxels were discarded. Contrast agent leakage effects were accounted for based on the method proposed by Boxerman et al. [15]. The corrected relaxation rate time curves were fitted in a gamma-variate function [16] and used for calculation of the ΔR(t)2,SE versus ΔR(t)2,GE diagram, producing a vascular hysteresis loop (VHL), see also [2]. The fitting procedure was based on the method of nonlinear least squares with a maximum iteration number of 4000; if fitting generated an error, the signal was set to zero. We excluded all fits with an r-value below 0.35. VAI quantification was performed using in-house developed MATLAB code (MathWorks, Natick, MA). Vessel size index and mean vessel density Q were calculated based on the procedure in [1, 2]. Since DTI was not included in our protocol, the ADC was set to the constant value of 0.8 μm2/ms [17].
To ensure readers of the robustness of the acquired data, one example of a voxel-wise uncorrected ΔR(t)2,XE curves in enhancing voxels is presented in S1 Fig. GE and SE voxels that exhibited T1 leakage effects were identified as the voxels that exhibit signals that increase above the pre-contrast baseline and overlapped (S2 Fig). These results demonstrate that leakage effects needed to be corrected.
All VAI parameters can be calculated from the VHL: the maximum distance between ascending and descending branches of the loop (I); the long axis (slope length), short axis and signed area of the vortex curve (microvessel type indicator MTI), the gradient of the long axis (slope), as well as the vascular-induced bolus peak-time shift (VIPS), which represents the temporal shift between the time-to-peak of the SE- and GE-EPI signal curves.
Statistics
For analysis of the VAI parameters, we determined the mean value and standard deviation at five different ROIs. The ROIs were manually drawn in HGG patients based on the FLAIR and the contrast-enhanced T1-weighted images, respectively, by two trained radiologists (FTK, CHZ). ROIs were placed in: (a) tumor, (b) peritumoral non-edematous tissue, (c) peritumoral edema, (d) contralateral normal appearing brain tissue (cNAB), and (e) necrosis. Examples of ROI placements are shown in S3 Fig. VAI parameters in tumor were compared with related values in peritumor, edema and cNAB using the Wilcoxon signed rank test (Table 2). The linear correlation coefficients (R) and P-values between two parameters in the tumor core (Table 3) and normal tissue (Table 4) were also determined. Bonferroni correction was performed for multiple comparisons.
Table 2. P-values for comparisons of VAI parameter distributions between HGG tumor core and peritumor, edema, and cNAB.
Tumor↔Peritumor | Tumor↔Edema | Tumor↔cNAB | |
---|---|---|---|
I | 0.010 | 0.007 | 0.001 |
MTI | 0.010 | 0.010 | 0.010 |
VIPS | 0.041 | 0.121 | 0.008 |
Slope | 0.188 | 0.421 | 0.330 |
Short Axis | 0.015 | 0.041 | 0.188 |
Slope Length | 0.022 | 0.064 | 0.421 |
rCBV | 0.599 | 0.002 | 0.048 |
Table 3. Correlation coefficients and p-values between pairs of VAI parameter distributions in the tumor core (N = 15).
Correlation Coefficient | |||||||
I | MTI | VIPS | Slope | Short Axis | Slope Length | rCBV | |
I | 1.000 | 0.864 | 0.768 | -0.878 | -0.642 | -0.826 | -0.545 |
MTI | 0.864 | 1.000 | 0.639 | -0.717 | -0.780 | -0.795 | -0.469 |
VIPS | 0.768 | 0.639 | 1.000 | -0.570 | -0.763 | -0.485 | -0.310 |
Slope | -0.878 | -0.717 | -0.570 | 1.000 | 0.563 | 0.919 | 0.730 |
Short Axis | -0.642 | -0.780 | -0.763 | 0.563 | 1.000 | 0.639 | 0.284 |
Slope Length | -0.826 | -0.795 | -0.485 | 0.919 | 0.639 | 1.000 | 0.633 |
rCBV | -0.545 | -0.469 | -0.310 | 0.730 | 0.284 | 0.633 | 1.000 |
P-value | |||||||
I | MTI | VIPS | Slope | Short Axis | Slope Length | rCBV | |
I | 0.000 | 0.000 | 0.001 | 0.000 | 0.010 | 0.000 | 0.035 |
MTI | 0.000 | 0.000 | 0.010 | 0.003 | 0.001 | 0.000 | 0.078 |
VIPS | 0.001 | 0.010 | 0.000 | 0.027 | 0.001 | 0.067 | 0.261 |
Slope | 0.000 | 0.003 | 0.027 | 0.000 | 0.029 | 0.000 | 0.002 |
Short Axis | 0.010 | 0.001 | 0.001 | 0.029 | 0.000 | 0.010 | 0.305 |
Slope Length | 0.000 | 0.000 | 0.067 | 0.000 | 0.010 | 0.000 | 0.011 |
rCBV | 0.035 | 0.078 | 0.261 | 0.002 | 0.305 | 0.011 | 0.000 |
Table 4. Correlation coefficients and p-values between pairs of VAI parameter distributions in the normal tissue (N = 15).
Correlation Coefficient | |||||||
I | MTI | VIPS | Slope | Short Axis | Slope Length | rCBV | |
I | 1.000 | 0.838 | 0.753 | -0.078 | 0.078 | 0.093 | 0.335 |
MTI | 0.838 | 1.000 | 0.455 | 0.148 | -0.182 | 0.194 | 0.491 |
VIPS | 0.753 | 0.455 | 1.000 | -0.200 | 0.281 | -0.029 | 0.047 |
Slope | -0.078 | 0.148 | -0.200 | 1.000 | -0.428 | 0.660 | 0.668 |
Short Axis | 0.078 | -0.182 | 0.281 | -0.428 | 1.000 | 0.083 | -0.190 |
Slope Length | 0.093 | 0.194 | -0.029 | 0.660 | 0.083 | 1.000 | 0.642 |
rCBV | 0.335 | 0.491 | 0.047 | 0.668 | -0.190 | 0.642 | 1.000 |
P-value | |||||||
I | MTI | VIPS | Slope | Short Axis | Slope Length | rCBV | |
I | 0.000 | 0.000 | 0.001 | 0.781 | 0.782 | 0.741 | 0.222 |
MTI | 0.000 | 0.000 | 0.089 | 0.597 | 0.517 | 0.488 | 0.063 |
VIPS | 0.001 | 0.089 | 0.000 | 0.476 | 0.311 | 0.918 | 0.869 |
Slope | 0.781 | 0.597 | 0.476 | 0.000 | 0.111 | 0.007 | 0.006 |
Short Axis | 0.782 | 0.517 | 0.311 | 0.111 | 0.000 | 0.770 | 0.499 |
Slope Length | 0.741 | 0.488 | 0.918 | 0.007 | 0.770 | 0.000 | 0.010 |
rCBV | 0.222 | 0.063 | 0.869 | 0.006 | 0.499 | 0.010 | 0.000 |
Results
Multiband and singleband EPI produce similar images
When compared to the SB technique, the MB produces multiple clones of the original SB profile, each of which has a symmetric frequency offset [18]. Fig 2 shows the EPI-based images acquired with the SB and MB techniques, respectively. A representative reconstructed volume is presented for each data set. All images were reconstructed after the unfolding procedure was applied to account for the 3-fold in-plane acceleration, which was introduced using parallel imaging. Two simultaneously excited slices are aliased to be contained in one slice with the use of the MB excitation. After the ‘Slice-GRAPPA’ reconstruction [19], the two simultaneously acquired slices were successfully separated from the MB-aliased images. Subtraction of MB EPI from SB EPI produces little differences at the same slice location (Fig 2). Presumably, this difference is mainly induced by motion during imaging.
Signal-to-noise ratios slightly lower in multiband imaging
The SNR values were calculated by 0.65×S/ σ, with S being the mean pixel intensity value in brain, and σ being the noise-associated standard deviation of background air, employing a correction factor of 0.65 for background noise (Rician distribution) [20]. The SNR of the reconstructed images from all subjects at the same slice position are listed in S1 Table. The mean SNR are 106.6 and 101.3 for SB and MB in GE EPI, and 38.2 and 36.8 for SB and MB in SE EPI, respectively.
G-factor maps and image histograms
The g-factor maps of GE and SE were calculated after the application of MB data reconstruction (Fig 3). The g-factor values induced from the MB acceleration of all subjects has an averaged value of 1.35 for GE and 1.33 for SE (S2 Table).
The histograms of the image slabs at the same location from 32 subjects show similar distributions between MB and SB techniques in GE and SE readout intensity (Fig 4). We determined the correlation coefficient between MB and SB histograms for each patient for both GE and SE readout to find an averaged correlation coefficient of 0.96 ± 0.03 for GE readout intensity and 0.96 ± 0.04 for SE.
Vascular hysteresis loops
A VHL from a voxel at the region of the tumor core is represented in Fig 5 for a patient with glioblastoma multiforme. To illustrate the variability of VHLs in tumor regions, we show VHL curves in three different tumor areas for the patient with glioblastoma (WHO IV; S4 Fig), another patient with glioblastoma (WHO IV; S5 Fig) and a patient with oligodendroglioma (WHO II; S6 Fig). After plotting the relaxation rate curves in a point-by-point time-parametrized plot, we found a clockwise loop that is typical for large arterial inflow, as shown previously [1, 2].
VAI parametric maps
Maps of VAI parameters from the same patient are represented in Fig 6. More cases of parametric maps can be found in S7 and S8 Figs. For instance, in S7 Fig, there is a strong decrease in distance map and MTI parameters in the tumor core area. The maps for slope length, slope and short axis are parameter maps based on absolute values, whereas parameter maps for I, MTI and VIPS are signed to account for the direction of the VHL. The VHL in a voxel transverses in the clockwise direction if the vascular system in this voxel contains mostly arterioles and capillaries, i.e. in case of relatively higher arterial blood volume content, whereas the VHL transverses in a counterclockwise direction if the vascular system in this voxel consists of mainly venule- and capillary-like vessel components, corresponding to a relatively higher venous blood volume content, see also [2, 3, 21]. According to the definition for the calculation of I and MTI, a clockwise VHL-direction was associated with a positive signed distance and signed area, i.e. positive I and MTI, and vice-versa for the counterclockwise VHL direction. In the maps of I and MTI, negative values were assigned to cool colors and positive to warm colors, respectively. A voxel with high arterial blood volume shows positive I and MTI values and an orange to red color in I and MTI maps; a capillary voxel shows low I and MTI values and green color in the maps; and a voxel with high venous blood volume shows negative I and MTI values and blue to purple color in the maps. In tissue areas that predominantly host venules and capillaries, the SE-EPI perfusion signal peaks earlier than the GE-EPI perfusion signal, corresponding to a negative VIPS value, and vice-versa for arteriole-dominated microvasculature. In fact, one can see that the distance maps display many dots of extreme colors, that typically show large arteries and veins, see also S9 Fig, where we demonstrate this effect, based on an additional SWI and TOF image of the same axial slice location as the distance map (see legend to S9 Fig for details).
In the VIPS maps, similarly to I and MTI, negative VIPS values were assigned to cool colors and positive to warm colors, respectively. In Fig 6G and 6H, we see an increase in negative I and MTI values in the larger contrast-enhancing tumor in the right frontal lobe, indicating a relatively venous/venule blood volume content than in the surrounding tissue. However, the smaller ring-enhancing tumor component with central necrosis, that is located posterior to the frontal bulk tumor, shows a more mixed microvascular architecture with partly venous and partly arterial-dominated blood volume content voxels, indicating a different state of angiogenesis. This distinction could not have been made based on the rCBV map alone that only shows an increased blood volume content in both bulk tumors.
In fact, among the parametric maps, the slope length map features the most similar contrast to rCBV: both maps exhibit increased value areas in the bulk tumor and decreased value areas in the surrounding edematous region, see Fig 6C and 6D. This is highly suggestive of a strong correlation (p = 0.011) between VHL slope length and rCBV, as confirmed for the bulk tumor region in Table 3.
Vessel size index (VSI) and parameter Q, a measure of microvessel density, are shown in Fig 6J and 6K for the same patient. The VSI is understood as a mean vessel radius averaged over the capillary population with the weight of its volumetric fraction; therefore, it can be used to monitor the dilation and contraction of microvessels [22]. The increase of VSI in the bulk tumor (Fig 6j) corresponds to a decrease of Q (Fig 6k). Values of VSI in the tumor core range between 90–140 μm, similar to results in previous study [23].
Statistical results
Within the HGG patient collective (N = 15), VAI parameters I and MTI were lower in the tumor core than in the peritumor, edema and cNAB regions (Fig 7, Table 2). VIPS in the tumor core for HGG patients was significantly decreased to the peritumor (p = 0.041) and cNAB regions (p = 0.008). Similarly, the short axis in the HGG tumor core was significantly decreased to the peritumor region (p = 0.015) and increased to edema (p = 0.041), the slope length in the tumor core was significantly increased to edema (p = 0.022), and, furthermore, tumor core rCBV was significantly increased to edema (p = 0.002) and cNAB (p = 0.048). A positive linear correlation between VAI parameters I and MTI were found both in the tumor core of HGG (Fig 8). Furthermore, parameter distributions of I, MTI and VIPS in the bulk tumor showed were highly correlated, see Table 3 (p<0.01). In fact, one can clearly see that the maps of I, MTI and VIPS are similar, see Fig 6, S7 and S8 Figs. Likewise, slope, short axis and slope length in the bulk tumor were correlated (p<0.03; see Table 3) and parametric maps of slope, short axis and slope length looked similar, see Fig 6, S7 and S8 Figs. In addition, CBV was found as the parameter to have no correlation with VAI parameters I, MIT, VIPS and short axis (p>0.08), but showed a high correlation with slope length. This can be explained by the slope length representing the root sum square of the peaks of the two relaxation rate curves, while rCBV is proportional to the area under the GE relaxation rate curve, which is determined by the peak of the GE relaxation rate curve. Similar to the tumor core, VAI parameters I, MTI and VIPS in normal tissue are in part correlated; however, MTI and VIPS are not correlated (p = 0.09). In addition, in normal tissue, only slope and slope length are highly correlated (p<0.01), whereas short axis is not.
Discussion
In this work, the MB technique was integrated in a simultaneous GE/SE EPI sequence for VAI imaging to increase the number of slices in order to achieve whole-brain coverage. The optimized GE/SE EPI sequence achieves twice the number of slices whilst maintaining the same repetition time (1.5s) as the SB technique. In previous studies with simultaneous GE/SE EPI based VAI, only a few slices could be acquired for the dynamic tracking of relaxation rates [1, 2, 24]. Generally, in order to capture the signal changes after the injection of contrast agent, there is a trade-off between temporal resolution and number of slices such that the temporal resolution needs to be reduced if more slices per readout are to be acquired. The use of simultaneous multi-slice imaging not only provides better tissue coverage, it also allows for modifications of the TR. In theory, a shorter TR should allow for a better sampling of the VAI vortex curve.
To achieve high SNR, high spatial resolution, and coverage of the whole brain, previous works utilized a dual contrast agent injection approach in combination with two separate SE and GE acquisitions to determine VAI parameters [4, 21, 23]. However, patient motion between two scans as well as differences in the time to peak for the two bolus injections may significantly affect data evaluation. Additionally, the dual contrast agent injection approach effectively doubles the dose of contrast agent together with a significant increase in measurement time, when compared to simultaneous GE/SE EPI acquisitions. Furthermore, leakage effects of contrast agent need to be corrected.
Applications of the MB technique have also been demonstrated previously for dynamic susceptibility contrast (DSC) imaging [12], where the authors applied MB excitation and blipped-CAIPI to a GE/SE EPI sequence with the same MB factor but lower in-plane iPAT factor of 2. However, by inserting the rephasing gradient after GE EPI, a higher in-plane resolution can easily be achieved by exploiting a higher iPAT factor of 3 for parallel imaging, as used in our study. Further optimization of MB RF to reduce peak power needs to be developed [25–27]. In this study, we employed a MB factor of 2 due to the relative low number of channels of our RF coil (20). A higher MB factor would reduce the TR for the same brain coverage and would also help to make the resolution more isotropic. Generally, the MB technique has been widely applied in the field of MR perfusion. MB with multi-echo EPI (3 and 7 echoes, respectively) has been applied to assess perfusion and permeability [28] and in hyperpolarized 13C spectroscopic imaging [29]. MB acquisition was also successfully applied in ASL [30].
As shown above, the quality of the images resulting from MB acceleration is comparable to those from the SB method, see Fig 2. The images were reconstructed without any loss of spatial resolution or severe distortions, and, in addition, MB and SB SNR were similar.
We have acquired VAI parameter maps from a patient collective with HGG tumors. In agreement with results from a previous study [4], we found that MTI in the tumor core of HGG was decreased when compared to peritumor, edema and cNAB (Table 2). Since MTI values are sensitive to the temporal shift between SE- and GE-perfusion, induced by changes in the vascular architecture, as well as the blood volume [2, 3], the observed decreased values in hyperperfused tumor tissue is likely dominated by the strong increase in blood volume. In contrast, VIPS in the tumor core of HGG was decreased when compared to peritumor and CNAB, but not to edema. Presumably, VIPS is less sensitive to blood volume, when compared to MTI, but more sensitive to early neovascularization [4], and, thus, provides additional information at the tumor periphery. A decreased VIPS in peritumoral edema may therefore indicate neovascular changes, e.g. vascular co-option through advancing tumor cells.
In addition, we found a strong correlation between VAI parameters I, MTI and VIPS in the tumor core, while MTI and VIPS in normal tissue were not significantly correlated. This again indicates that the strong increase in blood volume in the tumor core due to hyperperfused tumor tissue may have a more profound effect on MTI and VIPS values than changes in vascular architecture. Similarly, we also found a strong correlation in VAI parameters slope and slope length in both tumor core and normal tissue, as opposed to VAI parameter short axis, that does not correlate in normal tissue with either slope or slope length. This again indicates that the specific circumstances of increased blood volume and changes in vascular architecture of the tumor core are dominated by blood volume effects, thus allowing for a reduction in VAI parameters to characterize this phenomenon, e.g. MTI and slope length, while an extensive discrimination of vascular changes in healthy tissue and tumor periphery needs a more careful consideration of additional VAI parameters such as VIPS and short axis.
The increased temporal resolution as well as the improved brain coverage offered by a multiband dual echo GE/SE can also be a great asset for functional studies of the healthy brain. In fact, recent studies could show that GE and SE capture different features of the BOLD response respectively (e.g. a different linearity of the response or an increased benefit of SE for magnetic field inhomogeneities), see [31–33].
In conclusion, the high-resolution GE/SE EPI vessel architectural imaging sequence provides significantly improved slice coverage, and the corresponding VAI parametric maps offer more detailed insights into the complexity and heterogeneity of vascular changes in glioma.
Supporting information
Data Availability
Data are available upon request due to ethical and legal restrictions involving potentially identifying patient information, imposed by the ethics committee of the medical faculty of Heidelberg University (Alte Glockengießerei 11/1, 69115 Heidelberg, Germany). Data requests may be sent to the Department of Neuroradiology, Heidelberg University Hospital (Im Neuenheimer Feld 400, 69120 Heidelberg, Germany) or directly to the corresponding author.
Funding Statement
This work was supported by grants from the Deutsche Forschungsgemeinschaft (Contract Grant number: DFG ZI 1295/2-1 and DFG KU 3555/1-1). F. T. Kurz was supported by a postdoctoral fellowship from the medical faculty of Heidelberg University and the Hoffmann-Klose foundation of Heidelberg University.
References
- 1.Kiselev VG, Strecker R, Ziyeh S, Speck O, Hennig J. Vessel size imaging in humans. Magn Reson Med. 2005;53(3):553–63. 10.1002/mrm.20383 [DOI] [PubMed] [Google Scholar]
- 2.Xu C, Kiselev VG, Möller HE, Fiebach JB. Dynamic hysteresis between gradient echo and spin echo attenuations in dynamic susceptibility contrast imaging. Magn Reson Med. 2013;69(4):981–91. 10.1002/mrm.24326 [DOI] [PubMed] [Google Scholar]
- 3.Emblem KE, Mouridsen K, Bjornerud A, Farrar CT, Jennings D, Borra RJH, et al. Vessel architectural imaging identifies cancer patient responders to anti-angiogenic therapy. Nat Med. 2013;19(9):1178–83. 10.1038/nm.3289 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Stadlbauer A, Zimmermann M, Heinz G, Oberndorfer S, Doerfler A, Buchfelder M, et al. Magnetic resonance imaging biomarkers for clinical routine assessment of microvascular architecture in glioma. J Cereb Blood Flow Metab. 2017;37(2):632–43. 10.1177/0271678X16655549 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Boxerman JL, Hamberg LM, Rosen BR, Weisskoff RM. MR contrast due to intravascular magnetic susceptibility perturbations. Magn Reson Med. 1995;34:555–66. [DOI] [PubMed] [Google Scholar]
- 6.Kurz FT, Buschle LR, Kampf T, Zhang K, Schlemmer HP, Heiland S, et al. Spin dephasing in a magnetic dipole field around large capillaries: Approximative and exact results. J Magn Reson. 2016;273:83–97. 10.1016/j.jmr.2016.10.012 [DOI] [PubMed] [Google Scholar]
- 7.Zhang K, Sturm VJ, Buschle LR, Hahn A, Yun SD, Jon Shah N, et al. Dual-contrast pCASL using simultaneous gradient-echo/spin-echo multiband EPI. Magn Reson Imaging. 2019;57:359–67. 10.1016/j.mri.2018.11.018 [DOI] [PubMed] [Google Scholar]
- 8.Kurz FT, Kampf T, Buschle LR, Heiland S, Schlemmer H-P, Bendszus M, et al. CPMG relaxation rate dispersion in dipole fields around capillaries. Magn Reson Imaging. 2016;34(7):875–88. 10.1016/j.mri.2016.03.016 [DOI] [PubMed] [Google Scholar]
- 9.Buschle LR, Ziener CH, Zhang K, Sturm VJF, Kampf T, Hahn A, et al. Vessel radius mapping in an extended model of transverse relaxation. Magn Reson Mater Phys. 2018. August;31(4):531–51. [DOI] [PubMed] [Google Scholar]
- 10.Troprès I, Grimault S, Vaeth A, Grillon E, Julien C, Payen JF, et al. Vessel size imaging. Magn Reson Med. 2001;45(3):397–408. [DOI] [PubMed] [Google Scholar]
- 11.Lu H, Pollack E, Young R, Babb JS, Johnson G, Zagzag D, et al. Predicting Grade of Cerebral Glioma Using Vascular-Space Occupancy MR Imaging. Am J Neuroradiol. 2008;29(2):373–8. 10.3174/ajnr.A0794 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Eichner C, Jafari-Khouzani K, Cauley S, Bhat H, Polaskova P, Andronesi OC, et al. Slice accelerated gradient-echo spin-echo dynamic susceptibility contrast imaging with blipped CAIPI for increased slice coverage. Magn Reson Med. 2014;72(3):770–8. 10.1002/mrm.24960 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kickingereder P, Wiestler B, Sahm F, Heiland S, Roethke M, Schlemmer H-P, et al. Primary Central Nervous System Lymphoma and Atypical Glioblastoma: Multiparametric Differentiation by Using Diffusion-, Perfusion-, and Susceptibility-weighted MR Imaging. Radiology. 2014;272(3):843–50. 10.1148/radiol.14132740 [DOI] [PubMed] [Google Scholar]
- 14.Bonekamp D, Mouridsen K, Radbruch A, Kurz FT, Eidel O, Wick A, et al. Assessment of tumor oxygenation and its impact on treatment response in bevacizumab-treated recurrent glioblastoma. J Cereb Blood Flow Metab. 2017;37(2):485–94. 10.1177/0271678X16630322 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Boxerman JL, Schmainda KM, Weisskoff RM. Relative cerebral blood volume maps corrected for contrast agent extravasation significantly correlate with glioma tumor grade, whereas uncorrected maps do not. Am J Neuroradiol. 2006;27(4):859–67. [PMC free article] [PubMed] [Google Scholar]
- 16.Rausch M, Scheffler K, Rudin M, Radü EW. Analysis of input functions from different arterial branches with gamma variate functions and cluster analysis for quantitative blood volume measurements. Magn Reson Imaging. 2000;18(10):1235–43. [DOI] [PubMed] [Google Scholar]
- 17.Helenius J, Soinne L, Perkiö J, Salonen O, Kangasmäki A, Kaste M, et al. Diffusion-weighted MR imaging in normal human brains in various age groups. Am J Neuroradiol. 2002;23(2):194–9. [PMC free article] [PubMed] [Google Scholar]
- 18.Larkman DJ, Hajnal J V, Herlihy AH, Coutts GA, Young IR, Ehnholm G. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J Magn Reson Imaging. 2001;13(2):313–7. [DOI] [PubMed] [Google Scholar]
- 19.Setsompop K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn Reson Med. 2012;67(5):1210–24. 10.1002/mrm.23097 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Firbank MJ, Coulthard A, Harrison RM, Williams ED. A comparison of two methods for measuring the signal to noise ratio on MR images. Phys Med Biol [Internet]. 1999. December [cited 2019 Jun 7];44(12):N261–4. Available from: http://www.ncbi.nlm.nih.gov/pubmed/10616158 10.1088/0031-9155/44/12/403 [DOI] [PubMed] [Google Scholar]
- 21.Stadlbauer A, Zimmermann M, Kitzwögerer M, Oberndorfer S, Rössler K, Dörfler A, et al. MR Imaging–derived Oxygen Metabolism and Neovascularization Characterization for Grading and IDH Gene Mutation Detection of Gliomas. Radiology. 2017;283(3):799–809. 10.1148/radiol.2016161422 [DOI] [PubMed] [Google Scholar]
- 22.Xu C, Schmidt WU, Villringer K, Brunecker P, Kiselev V, Gall P, et al. Vessel Size Imaging Reveals Pathological Changes of Microvessel Density and Size in Acute Ischemia. J Cereb Blood Flow Metab. 2011;31(8):1687–95. 10.1038/jcbfm.2011.38 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Hsu Y-Y, Yang W-S, Lim K-E, Liu H-L. Vessel size imaging using dual contrast agent injections. J Magn Reson Imaging. 2009;30(5):1078–84. 10.1002/jmri.21960 [DOI] [PubMed] [Google Scholar]
- 24.Schmiedeskamp H, Straka M, Newbould RD, Zaharchuk G, Andre JB, Olivot J-M, et al. Combined spin- and gradient-echo perfusion-weighted imaging. Magn Reson Med. 2012;68(1):30–40. 10.1002/mrm.23195 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Eichner C, Wald LL, Setsompop K. A low power radiofrequency pulse for simultaneous multislice excitation and refocusing. Magn Reson Med. 2014;72(4):949–58. 10.1002/mrm.25389 [DOI] [PubMed] [Google Scholar]
- 26.Wong E. Optimized phase schedules for minimizing peak RF power in simultaneous multi-slice RF excitation pulses. Proc Intl Soc Mag Reson Med. 2012;(20):2209. [Google Scholar]
- 27.Auerbach EJ, Xu J, Yacoub E, Moeller S, Uğurbil K. Multiband accelerated spin-echo echo planar imaging with reduced peak RF power using time-shifted RF pulses. Magn Reson Med. 2013;69(5):1261–7. 10.1002/mrm.24719 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Wu J, Saindane AM, Zhong X, Qiu D. Simultaneous perfusion and permeability assessments using multiband multi-echo EPI (M2-EPI) in brain tumors. Magn Reson Med. 2019;81(3):1755–68. 10.1002/mrm.27532 [DOI] [PubMed] [Google Scholar]
- 29.Sigfridsson A, Weiss K, Wissmann L, Busch J, Krajewski M, Batel M, et al. Hybrid multiband excitation multiecho acquisition for hyperpolarized 13 C spectroscopic imaging. Magn Reson Med. 2015;73(5):1713–7. 10.1002/mrm.25294 [DOI] [PubMed] [Google Scholar]
- 30.Kim T, Shin W, Zhao T, Beall EB, Lowe MJ, Bae KT. Whole brain perfusion measurements using arterial spin labeling with multiband acquisition. Magn Reson Med. 2013;70(6):1653–61. 10.1002/mrm.24880 [DOI] [PubMed] [Google Scholar]
- 31.Chiacchiaretta P, Romani GL, Ferretti A. Sensitivity of BOLD response to increasing visual contrast: Spin echo versus gradient echo EPI. Neuroimage. 2013;82:35–43. 10.1016/j.neuroimage.2013.05.069 [DOI] [PubMed] [Google Scholar]
- 32.Khatamian YB, Golestani AM, Ragot DM, Chen JJ. Spin-Echo Resting-State Functional Connectivity in High-Susceptibility Regions: Accuracy, Reliability, and the Impact of Physiological Noise. Brain Connect. 2016;6(4). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Schwarzbauer C, Mildner T, Heinke W, Brett M, Deichmann R. Dual echo EPI—the method of choice for fMRI in the presence of magnetic field inhomogeneities? Neuroimage. 2010;49(1):316–26. 10.1016/j.neuroimage.2009.08.032 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data are available upon request due to ethical and legal restrictions involving potentially identifying patient information, imposed by the ethics committee of the medical faculty of Heidelberg University (Alte Glockengießerei 11/1, 69115 Heidelberg, Germany). Data requests may be sent to the Department of Neuroradiology, Heidelberg University Hospital (Im Neuenheimer Feld 400, 69120 Heidelberg, Germany) or directly to the corresponding author.