Abstract
The splitting of the 57Fe Mössbauer spectrum in TiFe has been carefully measured in a magnetic field and compared with the zero-field quadrupolar splitting of sodium nitroprusside. The comparison gives a value of 1.699 mm/s at 295 K for the sodium nitroprusside splitting. This determination makes use of the known Knight shift of Fe in TiFe and is independent of any velocity measurement. Comparison is made with the splitting of the inner two lines of a pure Fe spectrum.
Keywords: Calibration, chemical shift, iron, Knight shift, Mössbauer effect, sodium nitroprusside, TiFe
1. Introduction
The recent discovery [1]1 of the 57Fe nuclear magnetic resonance in the non-ferromagnetic intermetallic compound TiFe makes this material useful as a reference material for use in Mössbauer effect spectroscopy.
The Mössbauer effect of TiFe in the absence of an external field gives a single resonance line. Applying a known external field splits this line and the separation of the resulting lines can be used to calibrate the Mössbauer effect spectrometer in magnetic field units, energy units, and velocity units. Conversely, if the velocity calibration is known, the TiFe spectrum can be used to determine the applied magnetic field. This is possible because the Knight shift of 57Fe in TiFe is accurately known [1], and because TiFe has an ordered CsCl structure which places each Fe in a cubic site, giving a Mössbauer resonance with no resultant distortion due to the 57Fe excited state electric quadrupole moment. Also the Knight shift of TiFe has the advantage that it is only weakly temperature dependent. By varying the applied field, the TiFe line splitting can be varied. Thus this material is useful for calibration over a wide range of velocities and for checking spectrometer linearity.
In this paper we present data comparing Mössbauer effect spectra of TiFe in a magnetic field of about2 50 kG and spectra of sodium nitroprusside (SNP) in zero field. Using this comparison we obtain a calibration for the quadrupole split lines of sodium nitroprusside. Although intrinsically no more precise than a direct velocity determination, this calibration is of interest because it is independent of any velocity measurement. Values quoted in the literature [2–7] for the room-temperature quadrupole splitting of sodium nitroprusside determined by direct velocity measurement vary from 1.65 to 1.85 mm/s. Several of the more precise of these determinations are compared in figure 1. The results obtained by the method used here are in close agreement with the direct velocity determinations of references [2] and [7].
Figure 1. Comparison of various results on the quadrupole splitting of sodium nitroprusside.
(a) this work.
(b) reference [2].(c) reference [4].
(d) reference [3].
(e) reference [7].
We have been informed that with new direct velocity measurements, Drs. J. R. DeVoe, J. J. Spijkermann, and J. C. Travis have obtained a value above 1.700 mm/s but within our error bar shown in (a).
The sodium nitroprusside calibration is of considerable interest in Mössbauer spectroscopy with 57Fe. It has been issued [3] for use as a standard reference material for calibrating Mössbauer spectrometers and for comparing isomer shifts in iron-containing compounds. As previously noted by a number of authors [4, 8] sodium nitroprusside has several desirable characteristics for use as a standard. The fairly large quadrupole splitting is only weakly dependent on temperature (on the order [5] of 10−4 mm s−1 K−1 at room temperature). This is combined with an easily available high chemical purity and stability in addition to near natural resonance linewidths.
Using the results presented here, the Knight shifts and chemical shifts3 of 57Fe in Fe compounds can be measured to better than 0.5 percent by the Mössbauer effect. This is poor accuracy when compared with normal NMR results. However, it still offers the possibility of valuable results, since the NMR of 57Fe in most compounds is much too weak to be measurable.
2. Experimental Procedure
The Mössbauer effect spectrometer-superconducting magnet system has been described in detail elsewhere [9]. A schematic of the experimental arrangement is shown in figure 2. The velocity drive unit was used in the constant acceleration mode and spectra were stored in a multichannel analyzer.
Figure 2. Schematic drawing of the experimental arrangement.
The source temperature was controlled at 210 K, the TiFe absorber at 120 or 60 K. The sodium nitroprusside temperature was 295 K. The source-to-detector distance was 0.33 m. The γ-ray beam was collimated to a 12-mm diameter at the TiFe absorber position.
The source was about 100 mCi of 57Co in a Pd matrix. It was prepared4 by plating the requisite amount of carrier-free 57Co in a 6-mm-diam circle on one side of a 25-μm foil of pure Pd. This was diffused in a hydrogen atmosphere at about 1270 K for 1 hr followed by a water quench.
The TiFe alloys were prepared by arc melting weighed quantities in an argon atmosphere. The ingots were turned and remelted three times during arc melting. The purity of the starting materials was Ti 99.97 percent and Fe 99.999 percent (no 57Fe enrichment was used). An excess of one atomic percent Ti above the stoichiometric quantity was used. This gives a more perfectly ordered alloy, as evidenced by NMR signal strength [10] and ME line shape [11]. The arc-melted ingot was homogenized at 1470 K for 72 hr and then crushed to a powder. The powder was then annealed by sealing in an evacuated quartz ampoule, heating at 1170 K for 1 hr, then cooling to 870 K by reducing the temperature 30 K/hr. Two samples were prepared from separate melts and the measurements to be reported later detected no significant difference in their ME. (The same annealing procedure was used by Swartz et al. [1] for the TiFe samples in which 57Fe Knight shift measurements were made.) From the annealed powders −200 mesh particles were obtained. These particles were embedded in a lucite disk about 12 mm in diameter and 2 mm thick with enough powder to give 0.02 μg/m2 of 57Fe.
The sodium nitroprusside sample5 was in the form of a 10 mm square cut from a single crystal with the surface parallel to the ac plane of the crystal [2]. It was approximately 0.8 mm thick yielding 0.055 μg/m2 of 57Fe.
The magnetic field was calibrated by measuring the NMR frequency of Cu at 4.2 K with a super-regenerative detector. Details of this measurement are described more fully in the section on systematic errors.
Runs were made alternately on sodium-nitroprusside and TiFe. A total of eleven sodium nitroprusside spectra and nine TiFe spectra were obtained over a period of about 12 days.
3. Results
A summary of the measurement conditions for each run is given in table 1. The spectrum, w(x), taken with no sample was used to determine the small perturbation introduced by the presence of Fe in the beryllium windows of the magnet dewar and proportional counter. This spectrum is a doublet and was least squares fitted to obtain its position, splitting, and width. Since the sodium nitroprusside is in a small fringing field of several hundred gauss, two runs were taken with the magnet coil de-energized. No significant difference was detected when compared to the spectra taken with the coil energized.
Table 1. Summary of measurement conditions and results obtained.
All runs taken with source at 210 K, magnetic field at 50.131 kG, SNP at 295 K, and TiFe at 120 K, except as noted. The Δ1 ratios are obtained by averaging the SNP results before and after each TiFe run.
| Run No. | Absorber | Δ1 | |
|---|---|---|---|
| 010970.1 | SNP | 175.97 | |
| 010970.2 | TiFe | 169.20 | 1.0403 |
| 011070.1 | SNP | 176.06 | |
| 011070.2 | TiFea | 169.26 | 1.0396 |
| 011170.1 | SNP | 175.88 | |
| 011270.1 | SNPb | 175.73 | |
| 011270.2 | TiFe | 169.03 | 1.0398 |
| 011370.1 | SNP | 175.80 | |
| 011470.1 | Nonec | ||
| 011470.2 | TiFea | 168.48 | 1.0446 |
| 011570.1 | SNPb | 176.18 | |
| 011570.2 | TiFe | 169.10 | 1.0409 |
| 011670.1 | SNP | 175.84 | |
| 011670.2 | TiFe | 168.59 | 1.0426 |
| 011770.1 | SNP | 175.72 | |
| 011770.2 | TiFed | 168.79 | 1.0403 |
| 011970.1 | SNP | 175.47 | |
| 011970.2 | TiFed | 168.85 | 1.0409 |
| 012070.1 | SNP | 176.03 | |
| 012070.2 | TiFed | 168.30 | 1.0457 |
| 012170.1 | SNP | 175.94 |
Second of two samples of TiFe.
No magnetic field.
This run used to obtain window doublet parameters.
TiFe at 60 K.
All spectra were least squares fitted using an equation of the form
where
| (1) |
Here y(x) refers to the amount of resonant absorption as a function of channel No. x. The parameters a, b, and c represent a background parabola arising from the fact that the source-detector distance is varied in the experiment. The doublet due to the presence of Fe in the beryllium windows, determined in the run with no absorber, is represented by w(x) and d is the height of this doublet. For fitting the sodium nitroprusside spectra one doublet is used (see fig. 3) and the summation index α runs over only α = l. The TiFe spectra are fitted to two doublets, an outer and a narrowly spaced inner (see fig. 4) and the summation index α runs over α =1 and α = 2. The parameter Δα represents the doublet splitting, Cα the center of the doublet, Γα the full width at half maximum (FWHM) of each doublet line, ϵhα any difference in the height of the two doublet lines, and ϵWα any difference in the width of the two doublet lines.
Figure 3. Typical spectrum of sodium nitroprusside.
Circles are experimental points and the solid line is the least squares fit. The direction of gamma-ray propagation is along the b axis of the crystal. Distortion of background due to Fe in beryllium windows is evident. Velocity zero is taken as the center of the two prominent peaks.
Figure 4. Typical spectrum of TiFe in a magnetic field.
Circles are experimental points and the solid line is the least squares fit. The sample was a polycrystalline powder. Velocity zero the same as for figure 2.
The least squares fit was accomplished by an iterative procedure. The function y(x) was expanded in a Taylor series with only the linear terms retained. Partial derivatives with respect to the various parameters were determined numerically. Initial estimates of a, hα, Cα, and Γα could be made from the raw data to within 5 percent of their final value. The initial estimates for b, c, ϵhα, and ϵWα were taken as zero. The initial estimate for d was determined from a spectrum taken with no absorber. On two of the spectra several least squares fitting procedures were used to determine the effect of deleting various combinations of the background parabola, window doublet, and assymmetry parameters. This is discussed in the next section on systematic errors.
Determination of the splitting of the outer doublet of the TiFe spectrum Δ1 (TiFe) in a known field calibrates the spectrometer. From the splitting of the sodium nitroprusside doublet in channels Δ1(SNP), the splitting in velocity, v(SNP) is given by
| (2) |
where λ is the wavelength of the Mössbauer effect gamma ray, g1/g0 is the ratio of the 57Fe excited state level splitting to ground state splitting, K(T) is the Knight shift of Fe in TiFe, γ/2π is the gyromagnetic ratio of the 57Fe ground state, and H is the applied field.
In table 1 are listed the 9 values obtained for the ratio Δ1(SNP)/ Δ1(TiFe). For the parameters we adopt
The selection of these values and their contribution to the systematic error are discussed in the next section. Using these values, and the numbers listed in table 1, we obtain
as a 90 percent confidence interval8 for the sodium nitroprusside splitting at 295 K determined as described above. The contribution of the systematic errors are greater than the uncertainty in this value. They are discussed and estimated in the next section.
4. Systematic Errors
Besides the obvious sources of error, the determination of the position of a resonance line to within a few percent of its linewidth presents numerous difficulties. In this section we evaluate the known systematic errors in our determination of the sodium nitroprusside line splitting.
4.1. Magnetic Field Measurement
The magnetic field measurement utilized the 63Cu nuclear magnetic resonance at 4.2 K. The Knight shift of Cu is accurately known [13] (0.232 %) and is nearly temperature independent. A super-regenerative detector of the type described by Rubinstein and Shtrikman [14] was used with a quench frequency of 20 kHz. The 63Cu line was broadened to about 30 kHz by the field inhomogeneity. The accuracy of this method is better than one-half the resonance linewidth and is thus better than about 0.02 percent. No change in resonance frequency was detected when the superconducting magnet was switched to persistent mode. However, there was a field drift of 0.25 percent per hour in the persistent mode. To correct for this, the voltage drop across an 0.001Ω resistor required to give a field of 50.196 kG was measured. During the TiFe runs the field was readjusted every 0.5 hr to this value, giving an average field of 50.131 kG. Measurements of the field were made both before and after taking the ME spectra with no significant change detected. By moving the NMR probe in the magnetic field, the homogeneity of the field was estimated to be better than 0.1 percent over a 12.5-mm sphere. Considering all these factors, we estimate that the magnetic field error could contribute no more than 0.0015 mm/s to the systematic error.
4.2. Gamma-Ray Wavelength
The uncertainty in the value of λ[12] is less than 100 ppm. This is an order of magnitude smaller than the uncertainty of our measurement and will contribute less than 0.0002 mm/s to the systematic error in the sodium nitroprusside splitting.
4.3. Knight Shift Values
The uncertainty in the product [1 + K(T) ] γ/2π is likewise small and is given [1] as about 300 ppm. Since we use only the product and not K(T) alone, any uncertainty in the ground state nuclear moment of 57Fe does not affect our result. Also, there is no reason to expect a field dependence of the Knight shift of TiFe up to fields much larger than those used here. This view is supported by the fact that there is no detectable change in the TiFe spectra upon lowering the temperature from 120 K to 60 K. The contribution of the Knight shift value to the systematic error is less than 0.0005 mm/s.
4.4. Velocity Drive Linearity
A schematic diagram of the velocity feedback circuit is shown in figure 5. In this experiment we depend critically on the linearity of the reference signal, and the ability of the feedback circuit to reduce the error signal to zero. Throughout the experiment, the error signal was continuously monitored at the output of operational amplifier A2. Knowing the parameters of the velocity transducer and the values of R1, R2, R3, and C1, a correction could be made for the feedback error. However, in no case would this error have exceeded 0.03 channels (out of 175 channels peak separation) and hence it was neglected.
Figure 5. Velocity spectrometer feedback circuit.
Values of the parameters used were: R1 —250 kΩ; R2 —825 kΩ; R3 —1 kΩ; C1 —0.047 μF; velocity transducer—19 kΩ, 2.4 Hr, with 20 vs/m sensitivity.
The channel address voltage of the multichannel analyzer was used to control the reference signal (as described by Ruegg et al. [15]). Operation can be linear only if the reference signal voltage is proportional to the analyzer channel number. This was checked by measuring the reference signal voltage with a precision voltmeter as a function of channel number. A small nonlinearity of ±0.15 channels over the 332 channels utilized was detected. However, since the TiFe peaks and the sodium nitroprusside peaks fall within 18 channels of each other, the change in Δ1(SNP)/ Δ1(TiFe) by correcting for this nonlinearity would be only 0.0003 mm/s. This same reasoning applies for any nonlinearity in the velocity transducer. The transducer9 was a magnetically shielded coil about 180 mm long with an O.D. of 15 mm and with a bar magnet about 110 mm long moving through the inside of the coil. The difference in position of the bar magnet between corresponding peaks of the TiFe and the sodium nitroprusside spectra was only 80 μm. We thus expect the nonlinearity in the velocity transducer to shift the peaks of both TiFe and sodium nitroprusside equally and thus not affect the ratio Δ1(SNP)/Δ1(TiFe) significantly. Our estimate of the total systematic error due to velocity drive nonlinearity is 0.0008 mm/s.
4.5. Nuclear Moment Ratio
The ratio of the excited state to the ground state 57Fe moment, 3 g1/g0, appears in eq (2) and affects the accuracy of the results. (This ratio is actually negative but here we use 3 g1/g0 to refer to its absolute value.) It cannot be determined accurately from our own spectra because of the reduced intensity and very close spacing of the inner two doublet lines. We therefore use the value of 3 g1/g0 determined for pure Fe. Several precision measurements of pure Fe Mössbauer spectra have been made. Using an ultrasonic technique, Cranshaw and Reivari [16] obtained 3 g1/g0 = 1.715 ± 0.003. Using an interferometer to calibrate their spectrometer, Fritz and Schulze [17] obtained 3 g1/g0 = 1.711 ± 0.002. Using Moiré fringes for velocity measurement, DeWaard [18] obtained 3 g1/g0 =1.7l3 ±0.004, and using the lead screw of a lathe as a velocity drive, Preston et al. [19] obtained 3 g1/g0 = 1.715±0.004. Considering these values we adopt 3 g1/g0 = 1.714± 0.003. The uncertainty in this value for g1/g0 yields an uncertainty of 0.0019 mm/s in the sodium nitroprusside splitting.
We mention the possibility of an anomalous hyper-fine field effect [20] which could cause the apparent value of 3 g1/g0 to be different for 57Fe in TiFe and 57Fe in pure iron. The anomaly arises from the fact that the distribution of magnetic dipole density over the finite nuclear volume may be different for different isotopes or for the ground and excited nuclear states of the same isotope. The effect has been observed, for example, for the Knight shifts of 85Rb and 87Rb in rubidium metal [21] which are, respectively, 0.6527 percent and 0.6502 percent. This anomaly would lead to an error in magnetic field of 0.002 percent using Rb. The anomalous effect is thus expected to be negligible for 57Fe in different materials because it generally decreases with decreasing nuclear size. It has not been detected within the precision of any Fe Mössbauer experiments to date. We therefore feel confident in neglecting it for this experiment.
4.6. Background, Asymmetry, and Vibration
The background for the spectra obtained here were perturbed from an ideal straight line due to the presence of Fe in the Be windows of the dewar and due to source motion. To correct for these perturbations they were included as the parabola and window doublet of eq (1). Also, as indicated in eq (1), the heights and the widths of the doublets were allowed to be asymmetric. The window doublet fraction, d, averaged about 5 percent and the background parabola about 0.5 percent of the total absorption area. The height asymmetry, ϵh, averaged about 0.01 and width asymmetry, ϵw, about 0.02.
To test the effect of these factors on the results, various fits were made to the data of one TiFe spectrum and one sodium nitroprusside spectrum under various constraints on the parameters of eq (1). The results are shown in table 2.
Table 2.
Effect of constraints on lease squares fitting results for run No. 012070.2 (TiFe) and run No. 012170.1 (SNP).
| Parameters of fit | Δ1(TiFe) | Δ1 (SNP) | Percent change froma | Sfitf | ||
|---|---|---|---|---|---|---|
| TiFe | SNP | |||||
| (a) | 168.30 | 175.94 | 1.0454 | 0.013 | 0.012 | |
| (b) | 168.07 | 175.77 | 1.0458 | + 0.04 | .016 | .015 |
| (c) | 167.88 | 175.64 | 1.0462 | + .08 | .014 | .015 |
| (d) | 167.90 | 175.67 | 1.0463 | + .09 | .017 | .016 |
| (e) | 168.19 | 175.79 | 1.0452 | -.02 | .012 | .010 |
Fit to eq (1), no constraints, all parameters allowed to vary independently.
Fit to eq (1), background parabola terms b and c and asymmetry terms ϵhα and ϵwα constrained to be zero.
Fit to eq (1), height of window doublet d constrained to be zero.
Fit to eq (1), height of window doublet d, background parabola terms b and c, and asymmetry terms ϵhα and ϵwα constrained to be zero.
Fit to eq (1), with hα given a Gaussian distribution as a function of Cα. Widths Γα constrained to 28.0 channels (corresponding to 0.27 mm/s). No other constraints.
f Standard deviation of fit, , where N is the number of data points, P is the number of independent parameters used in the fit, and the data points γobs have been normalized to lie in the range of zero to unity. Both runs had an average of about 5 × 105 counts in each of 330 channels.
The Γα’s obtained for all runs varied from 34 to 36 channels, corresponding to a FWHM of 0.34 ±0.01 mm/s. This is about 0.07 mm/s larger than the width of 0.27 mm/s exhibited by both the TiFe and the sodium nitroprusside samples under ideal conditions. The excess width is due to vibrations arising from nearby vacuum pumps, building vibrations, etc. The effect of such vibrations is to give the center of the doublets a Gaussian probability distribution. When this assumption is put into eq (1) and a least squares fit made with the parameters Γα constrained to correspond to 0.27 mm/s, the results shown in table 2, line (e) are obtained.
A statistical analysis of the various fits shows that the constraints introduced in (b), (c), and (d) of table 2 all result in a statistically significant (5% level or better) increase in the standard deviation of fit. The analysis also shows that the assumptions made in (e) give the best fit to the data. As seen from table 2, the value of Δ1(SNP)/Δ1(TiFe) is in all cases quite insensitive to the method of fitting the data. The values of Δ1, though still relatively insensitive, vary considerably more than the ratio Δ1(SNP)/Δ1(TiFe). When vibrations may be a problem the results shown in table 2 suggest that the most appropriate method would be to fit with a Gaussian distribution of natural-width Lorentzians. The results presented here are quite insensitive to the choice of fitting method since they depend only on the ratio Δ1(SNP)/Δ1(TiFe).
As seen in table 2, no correction due to background, asymmetry or vibration greater than 0.1 percent is used. We estimate the systematic error due to these factors to be no greater than one-half this correction, 0.0008 mm/s.
The systematic error contributions are summarized in table 3. The result for the sodium nitroprusside splitting is then 1.6993 mm/s with a precision of ± 0.0024 mm/s based on a 90 percent confidence limit for the mean value of all determinations, and with a maximum systematic error of 0.0021 mm/s, not including possible errors in 3 g1/g0, λ, and K(T).
Table 3. Contributions to systematic error in this determination of the sodium nitroprusside splitting.
The first four are independently determined constants. The last three are unique with this measurement.
| Parameter | Estimated error contribution |
|---|---|
| mm/s | |
| γ | 0.0002 |
| K(T) | .0005 |
| 3g 1/g 0 | .0020 |
| Hyperfine field anamoly | .0000 |
| H | .0015 |
| Linearity | .0008 |
| Background, etc | .0008 |
5. Discussion
The sodium nitroprusside splitting obtained here is compared with other results in figure 1. Within the limit of error, the results compare favorably with those of Grant et al. [2] and Fritz [7]. It is also interesting to compare the splitting obtained for the sodium nitroprusside with a pure iron spectrum. From the results of four runs we estimate
where (g0-g1) is the splitting of the inner two lines of a pure iron spectra. Using this, and the previously quoted value for 3 g1/g0 and assuming Δ1(SNP)= 1.699 mm/s we obtain g0 = 3.918 ± 0.004 mm/s and g1=2.238 ± 0.004 mm/s for pure iron at 295 K. These values are compared with previous ME results and with NMR results [22] in table 4. The Mössbauer effect results are in good agreement and give confidence in the correctness of the SNP value. The NMR results are listed for reference. Though inherently more accurate than the ME results, they cannot be directly compared since NMR in ferromagnetic iron measures the splitting for 57Fe atoms in domain walls, whereas the ME measures the splitting for 57Fe atoms within the domains. As described by Schechter et al. [23], the values of g0 and g1 are useful for calibrating and checking linearity of spectrometers using pure Fe.
Table 4.
Comparison of g0 and g1 values obtained for 57Fe in pure Fe at 295 ± 1 K.
| Technique and Ref. | g0 | g1 |
|---|---|---|
| mm/s | mm/s | |
| ME, this work | 3.918 ± 0.004 | 2.238 ± 0.004 |
| MEa [16] | 3.911 ± 0.003 | 2.236 ± 0.003 |
| ME [17] | 3.931 ± 0.003 | 2.242 ± 0.002 |
| ME [18] | 3.916 ± 0.005 | 2.236 ± 0.006 |
| MEb [19] | 3.920 ± 0.008 | 2.240 ± 0.005 |
| NMRa [22] | 3.909 ± 0.001 |
Results given in frequency units, which we have converted to velocity units.
Interpolated between values given for 293 and 298 K.
The results presented for TiFe and sodium nitroprusside make it possible to measure Knight shifts and chemical shifts of Fe in suitable alloys and compounds to about 0.5 percent or better.
To obtain more accurate results, several modifications to the equipment within the state of the art would be useful. This would include an interferometer or Moiré fringe counter for direct velocity measurement while the data are being accumulated. This would allow a check on the internal consistency of the results and, once calibrated, an accurate determination of applied field using TiFe spectra. Higher fields than used here are desirable in order better to resolve the inner two lines of the TiFe applied field spectrum. This would provide a sensitive test for the possible presence of a small hyperfine field anomaly and would allow calibrations of materials such as pure Fe with larger internal field splittings. Also, a more homogeneous and stable magnet would be useful, since the magnet field drift and inhomogeneity gave the largest contribution to the systematic error of this experiment.
Acknowledgments
The authors wish to thank Dr. J. J. Spijkerman and Dr. J. C. Travis for helpful discussions, Dr. J. Mandel for advice on statistical analysis, and Dr. R. W. Grant for useful comments. We also thank D. P. Fickle for aid with sample preparation and R. D. Robbins and R. L. Parke for technical assistance.
Supported in part by the Office of Saline Water, Department of the Interior.
Footnotes
Figures in brackets indicate the literature references at the end of this paper.
The field value was carefully measured using NMR, as explained in detail later.
By chemical shift we mean the NMR frequency shifts often observed in nonmetals. The term chemical shift has sometimes been used to refer to the isomer shift, which is measureable only by ME.
The source preparation was performed by the New England Nuclear Corp., Boston, Mass.
Standard reference material 725; disodium pentacyanonitrosoferrate dihydrate (Na2Fe(CN)5NO · 2H2O). Otained from the Office of Standard Reference Materials, National Bureau of Standards, Washington, D.C. in October 1966.
This value is based on comparison with λWKα = 0.2090100 Å* [12]. Professor J. A. Bearden has kindly informed us that this is his most accurate value at present.
Taken from ref. [1]. These values are based on an effective nuclear moment of 39Kin the reference salt (aqueous KI) of 0.39090 μN.
Calculated as described by M. G. Natrella in NBS Handbook 91 (U.S. Government Printing Office, 1963) p. 2–2.
L. V. Syn Model 6LV2, Sanborn Electronics, Waltham, Massachusetts.
6. References
- [1].Swartz J. C., Swartzendruber L. J., Bennett L. H., and Watson R. E., Phys. Rev. B 1,146 (1970). [Google Scholar]
- [2].Grant R. W., Housley R. M., and Gonser U., Phys. Rev. 178, 523 (1969). [Google Scholar]
- [3].Spijkerman J. J., Snediker D. K., Ruegg F. C., and DeVoe J. R., Nat. Bur. Stand. (U.S.), Misc. Publ. 260–13 (1967). [Google Scholar]
- [4].Spijkerman J. J., Ruegg F. C., and DeVoe J. R., in Applications of the Mössbauer Effect in Chemistry and Solid-State Physics (International Atomic Energy Agency, Vienna, 1966), p. 254; and [Google Scholar]; Mössbauer Effect Methodology, Gruverman I. J. ed. (Plenum Press Inc., New York, 1965) Vol. 1, p. 115. [Google Scholar]
- [5].Kerler W., Physik Z. 167, 194 (1962); W. Kerler and W. Neu-wirth, Z. Physik 167,176 (1962). [Google Scholar]
- [6].Danon J., J. Chem. Phys. 41,3378 (1964). [Google Scholar]
- [7].Fritz R., Phys. Letters 31A,226 (1970). [Google Scholar]
- [8].Herber R. H., in Mössbauer Effect Methodology, ed. Gruverman I. J. (Plenum Press Inc., New York, 1965), Vol. 1, p. 3. [Google Scholar]
- [9].Swartzendruber L. J., Nucl. Instr. and Meth. 69, 101 (1969). [Google Scholar]
- [10].West G. W., J. Appl. Phys. 39,2213(1968). [Google Scholar]
- [11].Swartzendruber L. J., and Bennett L. H., J. Appl. Phys. 39,2215 (1968). [Google Scholar]
- [12].Bearden J. A., Phys. Rev. 137,B455(1965). [Google Scholar]
- [13].Carter G. C., NBS Alloy Data Center, Gaithersburg, Md., private communication. [Google Scholar]
- [14].Rubinstein M., and Sktrikman S., Am. J. Phys. 35, 945 (1967). [Google Scholar]
- [15].Ruegg F. C., Spijkerman J. J., and DeVoe J. R., Rev. Sci. Instr. 36, 356 (1965). [Google Scholar]
- [16].Cranshaw T. E., and Reivari P., Proc. Phys. Soc. 90, 1059 (1967). [Google Scholar]
- [17].Fritz R., and Schulze D., Nucl. Instr. and Meth. 62, 317 (1968). [Google Scholar]
- [18].DeWaard H., Rev. Sci. Instr. 36, 1728 (1968). [Google Scholar]
- [19].Preston R. S., Hanna S. S., and Heberle J., Phys. Rev. 128, 2207 (1962). [Google Scholar]
- [20].Bohr A., and Weisskopf V. F., Phys. Rev. 77, 94 (1950). [Google Scholar]
- [21].Blumberg W. E., Eisinger J., and Klein M. P., Phys. Rev. 124 , 206 (1961). [Google Scholar]
- [22].Budnick J. I., Bruner L. J., Blume R. J., and Boyd E. L., J. Appl. Phys. 32, 120S (1961). [Google Scholar]
- [23].Shechter H., Ron M., Niedzwiedz S., and Herber R. H., Nucl. Instr. and Meth. 44, 268 (1966). [Google Scholar]