Skip to main content
Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry logoLink to Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry
. 1970 Sep-Oct;74A(5):647–653. doi: 10.6028/jres.074A.052

Vapor Pressure and Heat of Sublimation of Platinum

E R Plante 1, A B Sessoms 1, K R Fitch 1
PMCID: PMC6696569  PMID: 32523215

Abstract

The vapor pressure of platinum was measured by the Langmuir method in the temperature range 1700–2000 K using a vacuum microbalance. Eight series of data gave concordant results and an average third-law heat of sublimation of 564.49 kJ mol−1 with an estimated overall uncertainty of 2.1 kJ mol−1 (134.92 ±0.5 kcal mol−1). Three out of eight second-law heats agreed with the third-law heats within one standard error but there was a tendency for second-law heats to be low. This was attributed to small systematic errors in the measurements. A vapor-pressure equation representing the data is log P(atm) = —29020/T+7.502, based on our third-law heat and tabulated entropies evaluated at 1800 K. Our data agree well with several previous Langmuir determinations but significantly decrease the error in the heat previously accepted.

Inability to obtain saturation pressures at lower temperatures, previously reported in the literature, was confirmed. It was shown that microgram quantities of carbon are capable of blocking the sublimation reaction.

Keywords: Heat of sublimation, Langmuir vaporization, platinum, rate of vaporization, vapor pressure

1. Introduction

This study was undertaken as part of an NBS contribution to a program involving the measurement of vapor pressures of selected standard materials in various cooperating laboratories. The object of these measurements is to determine reliable standard vapor pressure data and to reveal, if possible, any systematic differences in vapor pressures which might be attributable to different methods of measurement. Data leading to vapor pressures or heats of sublimation of platinum have been reported by a number of investigators [110]1 using several methods of measurement over a wide temperature range. Results from some of these data are listed in table 2 and comments on these studies are reserved for section 4 of this paper.

TABLE 2.

Derived quantities for platinuma

Source Second-Law Values Third-Law Values
ΔH298° b S.E. cΔS298° S.E. ΔHT°/R S.E. ΔST°/R S.E. bS.D.(log P) ΔH298° S.E.
kJ mol1 kJ mol1 J mol1K1 J mol1K1 K K kJ mol‒1 kJ mol‒1
SRM 680 I 555.20 6.16 144.85 3.38 28520 321 7.189 0.176 0.042 565.85 0.31
SRM 680 II 553.82 4.77 144.66 2.63 28457 254 7.184 .140 .018 564.78 .26
SRM 680 III 552.48 5.09 144.07 2.79 28393 264 7.156 .145 .018 564.56 .27
SRM 680 IV 564.76 8.62 151.11 4.76 29034 454 7.524 .250 .031 563.97 .32
SRM 747–15 I 558.56 6.41 147.56 3.53 28710 335 7.338 .185 .025 564.24 .27
SRM 747–17 II 568.25 6.39 153.06 3.50 29208 334 7.621 .183 .027 563.91 .29
SRM 747–17 III 557.23 6.11 147.09 3.35 28629 316 7.307 .174 .025 563.79 .26
SRM 747–15 IV 554.95 4.67 145.30 2.55 28514 247 7.216 .135 .017 564.82 .22
Mean, This Work Jones et al. [2] 537.51 26.37 135.99 14.12 27564 1385 6.708 .742 .131 564.49 .24
565.13 1.71
D & M [3] 606.45 32.31 173.95 18.81 31285 1683 8.759 .980 .118 566.38 1.36
H & W [4] 539.54 23.40 137.60 11.71 27571 1220 6.742 .611 .051 565.71 0.46
NSB [6] 536.97 17.20 27518 896
P & A [7] (s) 536.29 25.15 133.53 12.69 27411 1311 6.534 .662 .040 570.30 .45
P & A [7] (1) 582.93 14.56 156.14 6.56 28684 760 7.147 .342 .039 570.79 .44
Koch et al., 1 [9] 617.04 26.95 176.74 11.97 30433 1406 8.208 .624 .094 558.11 1.64
Koch et al., 1A [9] 601.94 23.32 169.01 10.46 29653 1219 7.808 .547 .119 560,87 1.43
a

All temperatures were converted to the 1968 IPTS.

b

S.E.=standard error; S.D.= standard deviation.

c

Based on tabulated data, △S°298 = 150.68 J mol−1K‒1.

2. Experimental Method

Data were obtained by Langmuir rate of sublimation measurements using a vacuum microbalance. Techniques and procedures were similar to those used previously [11]. Vacuum in the range of 10−7 to 10−9 torr was maintained and indicated by a commercial sputter-ion pump. Preliminary data were obtained using Standard Reference Material, SRM 680, but this material is not generally available in a convenient form for Langmuir measurements. Later measurements were made on SRM 747 stock which will be certified as a vapor-pressure standard. Samples were machined from the stock materials, both of about five 9’s purity, into right circular cylinders having nominal diameters of 0.25 or 0.20 cm and lengths of 1.9 cm. A hole 0.1 cm in diameter and 1.5 cm long, assumed to represent blackbody conditions, was drilled along the cylinder axis and a suspension hole 0.025 cm in diameter was drilled along a diameter about 0.20 cm from the other end. Machining was accomplished by ordinary machining methods using oil cutting fluid on the 0.25-cm-diam samples except that designated SRM 747–17, which was machined using oil-free tools and trichloro-ethylene cutting fluid. The blackbody hole of the 0.20cm sample was machined by arc erosion. Samples machined using oil cutting fluids were rinsed in acetone or ethanol to remove any residual traces of oil. However, this procedure was not effective, as two of the three samples treated in this way gave evidence of contamination as discussed later.

The sample was suspended from one arm of an equal-arm quartz beam microbalance by a chain of 0.025-cm-diam sapphire or quartz rods connected together by V-shaped hooks made by heating and bending the rods. The lower 10 cm of the suspension was 0.0075-cm platinum wire, which passed through the suspension hole in the samples and over the hook on the lowest suspension rod.

The appendage of the vacuum chamber in which the sample hung was a 22-mm o.d. Vycor2 tube made with a fused silica window at the bottom. The window could be protected during sublimation experiments with a magnetically actuated shutter. The shutter was used to protect the window during the first three experiments but was not used during the remaining five. For our experimental arrangement it was found that the increase in the window correction factor was not significantly greater when the shutter was not used than when it was. This is because the shutter must be kept open for a large fraction of the time during short experiments when the rate of sublimation is highest. Data were designated as belonging to a new experimental series when new window and/or mirror correction values were determined, when a different observer made temperature measurements, when the sample was changed, or when a combination of these changes was made.

Prior to experiments, a thin platinum coating which did not heat inductively was deposited on the interior surface of the Vycor tube and a grounding device consisting of a split circular stainless steel ring, about 1 cm wide and having a magnetically actuated wire hinge, was inserted into the Vycor tube so that it made contact with the platinum coating. The Vycor tube was connected to the system by means of a standard taper joint using Apiezon W sealant. A wire connected the grounding device to an electrical ground. With the sample in place, the hinge of the grounding device could be magnetically deflected until it made contact with the wire supporting the sample. This allowed for removal of any static charge generated during the high-temperature heating. Heating was accomplished by induction at 450 kHz. The metal sample served as its own susceptor.

Temperatures were measured with an NBS-cali-brated optical pyrometer through a calibrated window and mirror. Calibration corrections for the window and mirror were determined in separate experiments using a band lamp. Corrections were determined in terms of “A” values where A= (1/T) — (1/Tw); T is the brightness temperature in Kelvins of the source; and Tw is the brightness temperature of the source with the window or mirror in the optical path. Window corrections were determined before and after each series of measurements and the average value accepted. Mirror corrections were determined less frequently; an average of two independent sets of determinations was used.

In obtaining each datum point the following sequence of operations was followed: (1) the rest point of the balance was determined, (2) the sample was heated to a temperature about 100 K below the lowest temperature where vapor pressure measurements were practicable, (3) the power of the oscillator was adjusted to a predetermined setting and held constant, (4) the power was turned off, (5) the sample and platinum coating on the Vycor sleeve were grounded, and (6) the rest point of the balance was redetermined.

The mass change of the sample was determined from the displacement of the beam of the microbalance and the previously determined sensitivity, which was about 0.5 μg/μm. The change in sensitivity with load is sufficiently small so that the weight change during a series of experiments (about 2 mg) has negligible effect on it. Data during these experiments were obtained with two different but similar balances. One was gold plated and had been used for previous experiments. The second was unplated and had slightly higher sensitivity. Both balances exhibited excellent zero-point stability in contrast to some previous drift problems [11]. This is attributed to a more nearly constant ambient temperature at our new NBS facility, use of a narrower slot on the kinematic table which supported the gold-plated balance, and the grounding procedure which eliminates static charge on the sample and its surroundings.

Initial time for an experiment was taken when the brightness of the blackbody hole matched the preset brightness of the pyrometer filament set for a temperature 50 K or so below the expected temperature for a particular power setting. Final time was taken as the time the power was turned off. The first temperature measurement was usually obtained within the first minute of the experiment, at which time the sample had attained its ultimate temperature. This method of determining the duration of the experiment represents a slight change from our previous practice and was made because the rate of heating is slower than the rate of cooling. Qualitatively, the time required to heat from the base temperature to the operating temperature is inversely proportional to the temperature difference between them. For low-temperature runs the length of time to go from the base temperature to 50 K below the operating temperature would be about 15 s, while for high-temperature runs about 3 s would be required. We believe that systematic error resulting from this approximation is small.

Some difficulties were encountered in achieving linear log P versus 1/T relationships in two of the platinum samples which had been exposed to oil during the machining process. These samples exhibited the same type of behavior as described by Hampson and Walker [4] with platinum and by Carrera et al. with ruthenium [12]; that is, below about 1800 K the apparent vapor pressure deviated negatively from the normal curve. This behavior is apparently due to carbon contamination as a result of pyrolysis of oil residue introduced during machining, since a sample of the same reference material did not exhibit this behavior when machined using trichloroethylene cutting fluid. In addition, after the sample showing abnormal behavior was heated in air it gave normal vapor-pres-sure measurements. Finally, after the vapor-pressure measurements were completed, the 0.25-cm sample 747–15, which weighed 1.74 g, was heated in contact with graphite powder at 1735 Kfor half an hour. During this process it gained 72 μg. Microscopic examination of the sample at 75× showed a uniform darkening of the surface. Attempts at measuring the vapor pressure of this sample showed complete blocking of the vaporization process at temperatures up to 1860 K, while at 1900 K and above, the vaporization rate was normal. Measurements in the transition range were not very satisfactory because the sample temperature tended to increase or decrease at constant power setting. This behavior differs greatly from that of a pure sample, whose temperature would remain constant within a few degrees. These observations would seem to indicate that the carbon contamination stays close to the surface of the sample and that the extent of diffusion into the sample during the measurements is negligible. We can speculate that above the transition temperature the rate of sublimation is normal because the carbon is dissolved by the platinum and the rate of platinum diffusion through the solution is high, while below the transition temperature, the carbon would precipitate, and the rate of platinum diffusion through the carbon would be low.

3. Data and Thermodynamic Treatment

Vapor pressures were calculated using the equation3

P=mαat(2πRTM)1/2 (1)

where m is the mass of material sublimed, t is the duration of the experiment, a is the projected surface area of the sample, T is the absolute temperature on the 1968 IPTS [13], R is the gas constant, M is the atomic weight of the vaporizing species, monatomic platinum, and a is the vaporization coefficient which we assumed is equal to unity. The value of the sample area at temperature, AT, was calculated using the equation

AT=AR[1+2β(T300)]

where AR is the area calculated from measurements made at room temperature and β is the linear thermal expansion coefficient. For platinum, β was taken to be 11.3 × 10−6 K−1. This correction amounts to an increase in the surface area at temperature of 2 to 4 percent.

Sample areas at room temperature for the various samples were 1.60 cm2, 1.64 cm2, 1.64 cm2, and 1.27 cm2 for the platinum SRM 747 I through IV series, respectively, and 1.64 cm2 for the platinum SRM 680 data.

A linear equation was fitted to the data by least squares solution of the approximate integrated form of the Clausius-Clapeyron equation

logP(atm)=ΔH°RT+ΔS°R (2)

where △ and △S° are the heat and entropy changes at an average temperature and R’ is R In 10.

In addition, third-law heats of sublimation were calculated using the equation

ΔH298°=T[Δ(GT°H298°T)RlnP(atm)] (3)

where Δ(GT°H298°T) is the difference in free-energy functions of product and reactant. Free-energy function data for platinum were from Hultgren et al. [14]. Finally, accurate second-law heats and entropies were obtained using a method suggested by Horton [15] which is similar to Cubicciotti’s method [16]. This consists of fitting by least squares Δ(GT°H298°T)R In P versus 1/T and allowing for non-zero intercept. The slope corresponds to an accurately adjusted second-law heat, while the intercept corresponds to △S°298 (3d law) — △S°298 (2d law). To show this, add and subtract each of the quantities Δ(HT°H298°)/T and Δ(ST°S298°) from the negative of eq (2). Combining appropriate terms, the result is

RlnP=ΔH298°(2)T+Δ(HT°H298°T)ΔS298°(2)Δ(ST°S298°), (4)

which is equivalent to Cubicciotti’s [16] eq (4) or Horton’s [15] eq (22). Here, the (2)’s and (3)’s following heats or entropies are used to designate second-law or third-law values. The quantity Δ(ST°S298°)Δ(HT°H2983)/T equals Δ(GT°H298°)/TΔS298° (3). Therefore, collecting temperature-dependent terms, we can write

RlnP+Δ(GT°H298°T)=ΔS298°(3)ΔS298°(2)+ΔH298°(2)T. (5)

By subtracting the intercept from Δ(G298°H2980)/T=ΔS2980 (3d law), we can also obtain ΔS298° (2d law). The convenience of this method results from the fact that the Δ(GT°H298°)/TR logP terms are used in evaluating the third-law heats so that the data necessary for calculating an accurate second-law heat are already available. This method gives the same results as that described by Cubicciotti [16]. We should note also that the difference between the third-law entropy and the second-law entropy is a constant and is independent of temperature. The second-law entropy can, therefore, be determined at any temperature by subtracting the intercept from △ S (3d law). We have, however, chosen to evaluate the second-law entropy at 298 K for convenience.

4. Results

Basic data used in the calculation, the vapor pressures, and individual third-law heats are listed in table 1. Table 2 lists the average second-law heat and entropy change at 298.15 K and their standard errors, the coefficients of eq (2) and their standard errors, the standard deviation in the pressure in log units, and the average third-law heat and its standard error for each series of data. The mean third-law heat, calculated as the average of the means for each run, is 564.49 kJ mol−1 (134.92 kcal mol−1).

TABLE 1.

Compilation of platinum dataa

VAPOR PRESSURE OF PLATINUM SRM 680 .25 CM. SAMPLE SERIES I
Temp. Time Wgt. loss Pressure ΔH298° ΔH298°
Kelvins seconds micrograms atmospheres cal/mol J/mol
1809 1500 112 2.98 E–9 134796 563987
1781 2100 74.2 1.40 E–9 135416 566580
1866 900 168 7.57 E–9 135521 567018
1735 6600 95 5.64 E–10 135106 565285
1835 1200 133 4.46 E–9 135234 565817
1777 3000 102 1.35 E–9 135245 565865
1852 900 130 5.84 E–9 135475 566826
1821 1200 97.6 3.26 E–9 135352 566312
1704 7200 54.6 2.95 E–10 134922 564515
1675 10800 38.5 1.37 E–10 135210 565718
1909 300 135 1.84 E–8 135221 565767
1902 240 83.8 1.43 E–8 135688 567719
1807 1500 93.6 2.49 E–9 135294 566072
1978 180 210 4.86 E–8 136192 569826
1978 180 284 6.57 E–8 135007 564868
1759 3600 73.7 8.08 E–10 135690 567729
1945 210 190 3.74 E–8 134979 564751
1930 210 148 2.90 E–8 134935 564566
1752 4800 104 8.53 E–10 134970 564714
1802 1800 113 2.50 E–9 134911 564469
1697 9000 68.1 2.94 E–10 134387 562276
1875 600 136 9.21 E–9 135433 566652
1748 4800 81.3 6.66 E–10 135526 567041
1951 210 194 3.82 E–8 135305 566114
VAPOR PRESSURE OF PLATINUM SRM 680 .25 CM. SAMPLE SERIES II
1823 1200 111 3.71 E–9 135030 564964
1949 150 182 5.02 E–8 134111 561119
1755 2400 57.1 9.38 E–10 134866 564281
1879 600 162 1.10 E–8 135054 565067
1700 9000 65.3 2.82 E–10 134762 563846
1790 1800 89.3 1.97 E–9 134875 564315
1838 900 116 5.19 E–9 134898 564411
1727 7200 88.4 4.80 E–10 135046 565033
1908 240 115 1.96 E–8 134912 564474
VAPOR PRESSURE OF PLATINUM SRM 680 .25 CM. SAMPLE SERIES III
1833 1200 139 4.66 E–9 134929 564541
1882 600 163 1.11 E–8 135232 565813
1906 240 105 1.79 E–8 135117 565331
1939 210 169 3.32 E–8 135030 564965
1841 900 128 5.73 E–9 134752 563803
1827 1200 105 3.51 E–9 135523 567027
1692 9000 56.8 2.45 E–10 134610 563207
1781 2100 83.6 1.58 E–9 134988 564789
1732 7200 102 5.55 E–10 134932 564554
VAPOR PRESSURE OF PLATINUM SRM 680 .25 CM. SAMPLE SERIES IV
1933 180 130 2.98 E-8 135036 564989
1833 1200 132 4.42 E-9 135121 565347
1903 300 132 1.80 E-8 134888 564371
1822 2400 246 4.11 E-9 134586 563108
1753 3600 86.5 9.46 E-10 134685 563524
1699 10800 70.5 2.53 E-10 135051 565052
1873 900 256 1.16 E-8 134432 562464
1817 1500 130 3.47 E-9 134834 564143
1797 2400 153 2.54 E-9 134486 562690
1733 7200 108 5.88 E-10 134810 564043
VAPOR PRESSURE OF PLATINUM SRM 747–15 .25 CM. SAMPLE SERIES I
1864 600 114 7.89 E-9 135224 565778
1912 240 114 2.00 E–8 135113 565313
1806 1800 121 2.75 E–9 134864 564272
1698 10800 70.9 2.61 E–10 134867 564284
1750 4800 96.8 8.14 E–10 134981 564760
1932 180 130 3.05 E–8 134878 564329
1837 1200 158 5.43 E–9 134660 563419
1781 2700 116 1.75 E–9 134626 563276
1726 7200 94.4 5.26 E–10 134655 563398
1890 300 107 1.49 E–8 134692 563552
VAPOR PRESSURE OF PLATINUM SRM 747–17 .25 CM SAMPLE SERIES II
1936 180 164 3.76 E–8 134346 562105
1854 1200 216 7.28 E–9 134807 564031
1823 1800 176 3.92 E–9 134830 564130
1794 2700 150 2.21 E–9 134761 563841
1889 480 174 1.48 E–8 134647 563365
1726 7200 88.5 4.81 E–10 134962 564681
1780 3600 150 1.65 E–9 134760 563836
1748 5400 113 8.23 E–10 134791 563965
1932 180 120 2.75 E–8 135275 565993
1706 10800 89.4 3.22 E–10 134781 563926
1930 180 139 3.18 E–8 134581 563087
VAPOR PRESSURE OF PLATINUM SRM 747–17 .25 CM. SAMPLE SERIES III
1869 600 143 9.67 E–9 134826 564110
1836 1200 159 5.33 E–9 134656 563400
1731 5400 78.7 5.71 E–10 134757 563824
1924 180 117 2.67 E–8 134839 564168
1781 2700 113 1.66 E–9 134813 564058
1807 1800 123 2.73 E–9 134964 564689
1701 10800 93.4 3.36 E–10 134248 561695
1903 300 139 1.90 E–8 134683 563516
1963 120 156 5.39 E–8 134777 563905
1753 3600 80.7 8.83 E–10 134925 564528
VAPOR PRESSURE OF PLATINUM SRM 747–15 .20 CM. SAMPLE SERIES IV
1925 240 111 2.46 E–8 135221 565767
1830 1800 150 4.32 E–9 134987 564784
1876 600 115 1.01 E–8 135160 565510
1780 3000 93.6 1.60 E–9 134869 564291
1945 180 119 3.52 E–8 135202 565686
1800 1800 87.8 2.51 E–9 134749 563792
1894 360 104 1.52 E–8 134898 564411
1708 10800 67.1 3.12 E–10 135044 565025
1746 7200 112 7.90 E–10 134781 563924
1858 900 127 7.37 E–9 135048 565039
a

Data are listed in experimental sequence

The overall uncertainty in this value is estimated to be ±2100 J mol−1 (±0.5 kcal mol−1). This is calculated using an uncertainty in the pyrometer calibration of ±5 K, an uncertainty in the window and prism correction of ±4 K, and three standard errors in the mean third-law heat. This estimation neglects two possible sources of systematic error about which little is known; namely, deviation of the blackbody hole from blackbody conditions, and temperature inhomogeneity of the sample. Error in the third-law heat resulting from errors in the free-energy function data is considered negligible.

The average third-law heats for each series of measurements are reasonably consistent within the expected error limits. There is a tendency for second-law heats and entropies to be lower than the respective third-law values, but the significance of this is questionable.

This could be understood in terms of a non-unit evaporation coefficient or error in the free energy function data for Pt(s), but it is more likely that small systematic errors in our measurements are responsible. Specifically, the method of establishing the duration of each experiment could lead to larger negative deviations in the observed pressure as the time of an experiment is shortened. However, such systematic error, if it occurred, cannot be seen in the present data, possibly because of insufficient precision.

One set of data, series I on SRM 747, shows a trend in the third-law heats and residuals with order of experiment which could be indicative of a changing “A” value of the window during the series. However, measurements showed a “normal” increase of about 10 percent in the “A” value, which corresponds to a change in temperature of about 1° at 1800 K. To remove the trend in residuals, the “A” value would have had to change systematically by about 70 percent during the experiments. Several other possibilities to account for trend could be put forward but discussion of this point seems futile. The data for this series are normal in that they give heats and entropies in good agreement with the other series but are abnormal because of the trend in third-law heats and residuals.

Below the melting point, the vapor pressure of platinum can be adequately represented by the equation

logP(atm)=29020T+7.502,

which is based on our mean third-law heat and tabulated entropies evaluated at 1800 K. At 1500 K, an accurately calculated vapor pressure will deviate from the pressure predicted by this equation by only 3 percent. The poorest agreement with this equation occurs in the SRM 680 III data. At 1600 and 2000 K, the approximate temperature extremes for this series, the least squares equation representing the SRM 680 III data gives pressures deviating by +10 percent and − 10 percent, respectively. In terms of temperature this corresponds to an error of +5° at 1600 and −5° at 2000 K.

Table 2 also contains some derived quantities obtained from other studies. These data were reevaluated after conversion of temperature scales to the 1968 IPTS. This increased the average third-law heats by approximately 800 J and had negligible effect on second-law heats. Data rejected by the original authors were also rejected in our calculations. Second-law heats based on the data of Dreger and Margrave [3] and Norman, Staley, and Bell [6] differed significantly from those reported. However, it was not clear that their reported second-law heats were evaluated by least squares.

Data obtained by Langmuir sublimation studies include the original work of Langmuir and Mackay [1] corrected by Jones, Langmuir, and Mackay [2]; Dreger and Margrave [3]; and Hampson and Walker [4]. A preliminary analysis of Hampson and Walker’s data was also published by Diamond et al. [5], These three studies give average third-law heats in substantial agreement with our data. Some mass spec-trometric Knudsen effusion data were presented by Norman, Staley, and Bell [6], who reported a second-law heat in fair agreement with the Langmuir data. Peleg and Alcock [7] carried out some torsion Langmuir and torsion Knudsen experiments. Their data were published only in graphical form but evaluation of the graphical data indicates pressures lower by roughly 40 percent at 2000 K than those obtained in the classical Langmuir experiments. This difference was attributed by Peleg and Alcock to errors arising in determining the duration of the experiment in the classical Langmuir experiments. They argue that weight losses occurring during the preheating and cooling cycles, as well as outgassing of the sample during the experiment, lead to high pressures. However, errors of this magnitude resulting from measurement of the duration of the experiment or outgassing | are unlikely. A systematic difference in temperature measurement of about 20 K would account for the difference in measured pressures. Errors of this magnitude are frequently encountered and the arrangement used by Peleg and Alcock, which consisted of sighting an optical pyrometer on the surface of a specimen heated by radiation, can lead to significant errors in either direction depending on the geometric relationship of heater to specimen and the emittance of the specimen. Very high temperature measurements in the range 3400 to 3900 K were reported by Falk [8], who used a shock tube technique. These measurements yielded pressures agreeing within a factor of two with extrapolated pressures based on the classical Langmuir experiments. This agreement is excellent considering the length of the extrapolation, the extremely high temperatures, and the complexity of the experimental method. However, these data are subject to greater interpretive error than those of the simpler experiments and no attempt has been made to derive thermodynamic quantities from them. Recent Langmuir measurements carried out on molten platinum by Koch et al. [9] by a novel technique give average third-law heats somewhat below our average third-law heat. However, lack of knowledge of the emittance of molten platinum as a function of temperatures makes their result relatively unreliable. Finally, some rate of vaporization data by Rytvin and Ulybysheva [10] were noted but were too fragmentary to be considered further.

In summary, our data give good agreement with those of three previous Langmuir determinations. The major contribution of our data is that they significantly decrease the error in the heat previously accepted, primarily because of improved precision. Unanswered questions remain concerning the effect and importance of temperature gradients in our sample, extent of deviation from blackbody conditions in the blackbody sight hole, and whether or not apparent systematic errors are experimental artifacts or are properties of platinum. Questions of this sort can be answered only by systematic studies which yield improved precision.

Footnotes

1

Figures in brackets indicate the literature references at the end of this paper.

2

Reference to trade names is made only for completeness of description and does not imply in any way the endorsement of the product by the National Bureau of Standards

3

Values of constants used in the equation were: R = 8.3143 J mol−1 K−1 or 1.98717 cal mol−1 K−1, atomic weight of platinum = 195.09; one standard atmosphere = 101.325 Nm−2

5. References

  • [1].Langmuir I., and Mackay G. M. J., Phys. Rev. 4,377 (1914). [Google Scholar]
  • [2].Jones H. A., Langmuir I., and Mackay G. M. J., Phys. Rev. 30, 201 (1927). [Google Scholar]
  • [3].Dreger L. H., and Margrave J. L., J. Phys. Chem. 64, 1323(1960). [Google Scholar]
  • [4].Hampson R. F. Jr., and Walker R. F., J. Res. Nat. Bur. Stand.(U.S.), 65A (Phys. and Chem.), 289 (1961). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Diamond J. J., Efimenko J., Hampson R. F., and Walker R. F., Reactivity of Solids, Proc. Fourth Intern. Symp. Reactivity of Solids, Amsterdam, 1960, deBoer J. H. et al. , eds. (Elsevier Publ. Co., Amsterdam, The Netherlands; ). [Google Scholar]
  • [6].Norman J. H., Staley H. G., and Bell W. E., J. Phys. Chem. 71,3686 (1967). [Google Scholar]
  • [7].Peleg M., and Alcock C. B., J. Sci. Inst. 43, 558 (1966). For a larger graph of the Pt data see also Bulletin de la Societe Franc.aise de Ceramique Nr. 72 (1966). [Google Scholar]
  • [8].Falk T. J., J. Chem. Phys. 49, 3727 (1968); also, J. Chem. Phys.48,3305 (1968). [Google Scholar]
  • [9].Koch R. K., Calvert E. D., Thomas C. R., and Beall R. A.,U.S. Bur. Mines RI 7271 (1969).
  • [10].Rytvin E. I., and Ulybysheva L. P., Izv. Akad. Nauk SSSR,Metal [1], 247(1969). [Google Scholar]
  • [11].Plante E. R., and Szwarc R., J. Res. Nat. Bur. Stand. (U.S.),70A (Phys. and Chem.), 175 (1966). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Carrera N. J., Walker R. F., and Plante E. R., J. Res. Nat.Bur. Stand. (U.S.), 68A (Phys. and Chem.), 325 (1964). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].The International Temperature Scale of 1968, Metrologia 5, 35 (1969). [Google Scholar]
  • [14].Hultgren R., Orr R. L., Anderson P. D., and Kelley K. K., Selected Values of Thermodynamic Properties of Metals and Alloys, p. 221 (John Wiley & Sons, Inc., New York, N.Y., 1963). [Google Scholar]
  • [15].Horton W. S., J. Res. Nat. Bur. Stand. (U.S.), 70A (Phys. And Chem.), 533 (1966). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Cubicciotti D., J. Phys. Chem. 70, 2410 (1966). [Google Scholar]

Articles from Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry are provided here courtesy of National Institute of Standards and Technology

RESOURCES