Skip to main content
Journal of Applied Clinical Medical Physics logoLink to Journal of Applied Clinical Medical Physics
. 2019 Jul 31;20(8):65–77. doi: 10.1002/acm2.12679

Three IMRT advanced planning tools: A multi‐institutional side‐by‐side comparison

Lan Lu 1, Yang Sheng 2, Jeremy Donaghue 3, Zhilei Liu Shen 1, Matt Kolar 1, Q Jackie Wu 2, Ping Xia 1,
PMCID: PMC6698808  PMID: 31364798

Abstract

Purpose

To assess three advanced radiation therapy treatment planning tools on the intensity‐modulated radiation therapy (IMRT) quality and consistency when compared to the clinically approved plans, referred as manual plans, which were planned without using any of these advanced planning tools.

Materials and Methods

Three advanced radiation therapy treatment planning tools, including auto‐planning, knowledge‐based planning, and multiple criteria optimization, were assessed on 20 previously treated clinical cases. Three institutions participated in this study, each with expertise in one of these tools. The twenty cases were retrospectively selected from Cleveland Clinic, including five head‐and‐neck (HN) cases, five brain cases, five prostate with pelvic lymph nodes cases, and five spine cases. A set of general planning objectives and organs‐at‐risk (OAR) dose constraints for each disease site from Cleveland Clinic was shared with other two institutions. A total of 60 IMRT research plans (20 from each institution) were designed with the same beam configuration as in the respective manual plans. For each disease site, detailed isodoseline distributions and dose volume histograms for a randomly selected representative case were compared among the three research plans and manual plan. In addition, dosimetric endpoints of five cases for each site were compared.

Results

Compared to the manual plans, the research plans using advanced tools showed substantial improvement for the HN patient cases, including the maximum dose to the spinal cord and brainstem and mean dose to the parotid glands. For the brain, prostate, and spine cases, the four types of plans were comparable based on dosimetric endpoint comparisons.

Conclusion

With minimal planner interventions, advanced treatment planning tools are clinically useful, producing a plan quality similarly to or better than manual plans, improving plan consistency. For difficult cases such as HN cancer, advanced planning tools can further reduce radiation doses to numerous OARs while delivering adequate dose to the tumor targets.

Keywords: auto‐planning, knowledge‐based planning, multiple criteria optimization

1. INTRODUCTION

The key to intensity‐modulated radiation therapy (IMRT) IMRT planning is the use of a compute optimization to find an optimal balance between delivering adequate prescription dose coverage to the planning target volume (PTV), and sparing normal tissues. Even with compute optimization, finding a clinically optimal IMRT plan for a specific patient is still challenging and time‐consuming, often resulting in large variations in plan qualities among different institutions and planners.1, 2, 3 The large variations in plan quality are partly due to that the manual plan objectives are based on the simplified dose and volume, and not explicitly specified on spatial relationship between the organs‐at‐risk (OARs) and PTVs of the specific patient. Another reason is that most planning optimization algorithms find a solution with a local minimum rather than the global minimum.4, 5, 6 Because of these limitations, progressive adjustments in the planning objectives are often used by experienced planners to create clinical IMRT plans. However, extensive manual progressive adjustments increase the planning time and the outcomes are highly dependent on the planner's skills and experience.

Many research efforts have been devoted to reduce variations in plan quality and accelerate the IMRT planning process, henceforth, several advanced treatment planning tools have been developed and implemented clinically. One method, referred to as the automatic planning (AP), is developed by the Pinnacle treatment planning system (Philips Radiation Oncology Systems, Fitchburg, WI).7, 8, 9 The second method, referred to as the knowledge‐based planning (KBP), is available from the Eclipse treatment planning system (Varian, Palo Alto, CA).10, 11, 12, 13 The third approach, named multi‐criteria optimization (MCO) is available in the RaySearch System (Stockholm, Sweden).14, 15, 16

The purpose of this study is to assess advantages of these three advanced treatment planning tools when compared with the clinically accepted IMRT plans without using any of these advanced planning tools. For the remaining text, we refer to these clinical plans as the manual plans. Using the same planning datasets from brain tumor, head‐and‐neck (HN) cancer, advanced prostate cancer with pelvic lymph nodal involvement, and spinal tumors treated with SBRT, three participated institutions applied one of these three advanced planning tools to create IMRT plans with minimum human interventions. We evaluated plan quality of the advanced plans against the manual plans.

2. MATERIALS AND METHODS

2.1. Study dataset

Twenty clinical cases, including five brain cases, five HN cases, five advanced prostate cases with pelvic lymph nodal involvement, and five spinal tumor metastasis cases treated with SBRT, were retrospectively selected from Cleveland Clinic for this study. The patient identifications were anonymized using MIM software (MIM 6.4, Cleveland, OH) and subsequently sent to the other two participating institutions, Duke University and Akron General Hospital. Three advanced treatment planning tools, including AP, KBP, and MCO, were used in the study. For the KBP approach, the model was trained for each treatment site individually using the in‐house program proposed by Yuan et al.12 Each participating institution has clinically implemented one of these advanced planning tools.

For each cancer site, the general planning goals from Clevelan Clinic were sent to the other two institutions without case specific instruction. The manual plans were generated using the Pinnacle planning system according to the same clinical goals at Cleveland Clinic by experienced planners and were clinically approved. IMRT quality assurance measurements for manual plans were conducted and passed the local institution criteria. Table 1 lists the planning guidance for these four sites. Volumetric modulated arc (VMAT) plans were generated for head‐and‐neck, brain and prostate cases and step‐and‐shoot plans were generated for spine cases using these three advanced planning tools, resulting in a total of 60 plans. For each case, all plans used the same beam configurations (Table 2) as in the manual plans, such as the isocenter location, gantry angles for step and shoot plans, number of arcs and arc lengths for VMAT plans, and the collimator angles. All plans were normalized to ensure that 95% of all PTVs received the prescription doses.

Table 1.

Treatment planning goals for head‐and‐neck, brain, prostate with pelvic lymph nodes, and spin SBRT cases

Head‐and‐Neck Brain
Organ name Endpoint Goal Organ name Endpoint Goal
PTV_7000 V70 Gy >95% PTV_6000 V60 Gy >95%
PTV_5600 V56 Gy >95% PTV_5940 V59.4 Gy >95%
Brainstem D0.03cc <54 Gy PTV_5100 V51 Gy >95%
Brainstem V30 Gy <50% PTV_5040 V50.4 Gy >95%
Spinal_cord D0.03cc <45 Gy Brainstem D0.03 cc <60 Gy
Parotid_L Dmean <26 Gy OPTIC_NRV_L D0.03 cc <55 Gy
Parotid_R Dmean <26 Gy OPTIC_NRV_R D0.03 cc <55 Gy
Larynx Dmean <35 Gy GLOBE_L D0.03 cc <50 Gy
Mandible D0.03cc <75 or 65 Gy GLOBE_R D0.03 cc <50 Gy
Trachea Dmean <45 Gy LENS_L D0.03 cc <7 Gy
Esophagus Dmean <50 Gy LENS_R D0.03 cc <7 Gy
Lips Dmean <20 Gy CHIASM D0.03 cc <56 Gy
Oral cavity Dmean <35 Gy COCHLEA_L Dmean <45 Gy
Submadibular_L Dmean <39 Gy COCHLEA_R Dmean <45 Gy
Submadibular_R Dmean <39 Gy Spinal_cord D0.03 cc <56 Gy
Prostate +Pelvic Lymph nodes Spine SBRT
Organ name Endpoint Goal Organ name Endpoint Goal
PTV_7000 V70 Gy >95% Tumor_1800 V18 Gy >90%
PTV_6200 V62 Gy >95% Tumor_1600 V16 Gy >90%
PTV_4500 V45 Gy >95% Spinal_cord D0.03 cc <14 Gy
BLADDER V63 Gy <10% Spinal_cord V1000 <10%
BLADDER V54 Gy <20% Sacrum/Thecal sac) D0.03 cc <16 Gy
RECTUM V63 Gy <10% Sacrum (Thecal sac) V1200 <10%
RECTUM D10 cc <63 Gy Esophagus D0.03 cc <15.4 Gy
RECTUM V45 Gy <50% Esophagus V1190 <5cc
FEMORAL HEAD_L V50 Gy <5% Kidney_L V1060 <2/3
FEMORAL HEAD_R V50 Gy <5% Kidney_R V1060 <2/3
PENILE BULB Dmean <52.5 Gy      

Table 2.

The same Beam configurations used for the manual plan and advanced plans

  Head Neck Brain Prostate Spine
Gantry CCW: 182‐178
CW: 178‐182
CCW: 182‐178
CW: 178‐182
Arc1: 182‐172
Arc2: 178‐182
Arc3: 200‐270
Acr4: 160‐90
Step and shoot:
280, 255, 230, 205, 181, 155, 130, 105, 80
Collimator angle 10° or 20o 330° or 30o 90o 0o
MLC leave width 0.25 cm or 0.5 cm 0.4 cm or 0.5 cm 0.4 cm or 0.5 cm 0.25 cm or 0.5 cm
Energy 6 MV 6 MV 10 MV 6 MV

2.2. Three advanced planning tools

The automatic planning (AP) tool developed by the Pinnacle treatment planning system (version 9.10, Philips Radiation Oncology Systems, Fitchburg, WI), is designed to mimic the manual process by automatically adjusting and adding planning objectives progressively and iteratively.8 The research plans generated for this study did not have additional manual optimizations after AP planning. The AP planning tool has been clinically implemented at Cleveland Clinic .

The knowledge‐based planning (KBP) is implemented in the Eclipse treatment planning system dedicated for research (version 13.6, Varian, Palo Alto, CA), which hosts anonymous patient cases and is separate from clinic patient database. KPB planning uses the dose and volume endpoints achieved from previous IMRT plans as the planning objectives for the new patient with a similar anatomic site. More specifically, using machine learning algorithms, a cancer‐specific model can be built based on the training dataset which is composed of anatomic and dosimetric information of prior patients. In the current study, all the KBP models are trained and validated with one institution's datasets, outside the 20 cases used in this study. The research plans generated from KBP were not further tuned to improve plan quality after the plans were submitted for comparison. The KBP planning tool has been clinically implemented at Duke University.

The multi‐criteria optimization (MCO) is implemented in the RaySearch planning system (version 4.9.9, Stockholm, Sweden). Rather than generating a single plan, the Raysearch planning system using MCO creates a series of Pareto optimal plans for a specific case.14, 15, 17 The advantage of MCO planning is that the plan trade‐offs are visible to physicians and planners and allow them to evaluate these plans and then select an optimal plan for each patient‐specific case. The research plans generated from MCO were not further optimized after the plans were submitted for comparison. The MCO planning tool has been clinically implemented at Akron General Hospital.

2.3. Plan evaluation

The plan quality evaluation was performed based on quantitative dosimetric parameters extracted from the plans, including percentage of dose coverage on target volumes, the maximum or mean dose of the OARs, the conformity index (CI), and the homogeneity index (HI). The CI is defined as

CI=VRx/VPTV (1)

where VRx is the tissue volume covered by the prescription dose for the PTV (high dose PTV, if there are multiple PTVs) and VPTV is the volume of the corresponding PTV. For the ideal case, CI equals to 1. The HI is defined as

HI=Dmax/DRx (2)

where Dmax is the maximum dose to 0.03 cc and DRx is the prescription dose for the high dose PTV (HD‐PTV). In addition, the total monitor units (MUs) per plan were also used to assess the delivery efficiency.

For each site, we plotted box plots of the PTVs and OARs for the four types of plans. In addition, we randomly selected one case from each site to compare dose distributions and dose volume histograms among the four plans case by case.

3. RESULTS

3.1. HN cases

Figure 1 shows box plots of defined endpoints of the five head‐and‐neck cases. All advanced plans had equal or lower maximum doses to the spinal cord and brainstem, equal or lower mean dose of both parotid glands than those of manual plans [Figs. 1(a)‐1(d)]. Although the maximum doses of the brain stem and spinal cord in the manual plans were below the tolerance dose, the additional dose reductions from advanced plans could benefit if re‐irradiation is needed. For these OARs (spinal cord, brainstem, and both parotid glands), all three advanced plans showed narrow ranges on the defined dosimetric endpoints when compared to manual plans, indicating a higher consistency in plan quality. Furthermore, except for one KBP plan, all advanced plans had lower mean dose to the larynx when compared to those from the manual plans [Fig. 1(e)]. Except for one MCO plan, the mean dose to the oral cavity from the advanced plans was also lower than those from the manual plans, albeit the reductions were small [Fig. 1(f)]. Figure 1(m) and 1(n) compared plan conformal indices and homogeneity indices among advanced plans and five manual HN plans. The plan conformity indices from AP and KBP plans were equal to or better than those of manual plans, but conformity indices from five MCO plans were slightly worse than those of the manual plans [Fig. 1(m)]. Plan homogeneity indices from AP and KBP plans were equal to or better than the manual plans but homogeneity indices from MCO plans were worse than those of manual plans, indicating trade‐offs among the OAR sparing, plan conformity, and homogeneity [Fig. 1(n)]. The MUs from AP and MCO plans were slightly higher than those of manual plans and KBP plans had the lowest MUs when compared to the manual, AP, and MCO plans [Fig. 1(o)].

Figure 1.

Figure 1

Comparison of the dosimetric performance between the three advanced plans and manual plans in head‐and‐neck cases for maximum dose for brainstem (a) and spinal cord (b), mean dose for parotid glands (c and d), larynx (e) and oral cavity (f), trachea (g), lips (h), esophagus (j), and submandibular (k and l), and max dose to mandible (i) as well as conformity index (m), homogeneity index (n), and MU (o).

Figure 2 shows representative isodose distributions for a randomly selected HN case. Compared to the manual plan, the advanced plans showed more conformal dose distributions, especially in the middle to low dose ranges (for instance, the 45 and 30 Gy isodose lines in Fig. 2). In addition, noticeable hotspots were observed in the HD‐PTV with the MCO plan. Fig. 3 shows the DVHs for all four methods for the same HN case as in Fig. 2. The AP and the manual plans achieved similar PTV homogeneity, while KBP and MCO plans were less homogenous with long tails noted in the HD‐PTV DVHs. In addition to equal or better dose sparing to the spinal cord, brain stem, and parotid glands [Figs. 3(b) and 3(c)], advanced plans achieved better sparing for both left and right submandibular, esophagus, and trachea when compared to the manual plan [Figs. 3(d) and 3(e)]. The maximum dose to the mandible was similar from all plans [Fig. 3(f)]. However, for the larynx, KBP and MCO resulted in worse DVHs compared to AP and manual plans, indicating the trade‐off between larynx sparing and other OARs' sparing for KBP and MCO plans for this typical case [Fig. 3(f)].

Figure 2.

Figure 2

Representative isodose distributions from Manual, automatic planning (AP), knowledge‐based planning (KBP), and multi‐criteria optimization (MCO) plans for a head‐and‐neck patient.

Figure 3.

Figure 3

Representative DVHs of a head‐and‐neck patient for all four planning methods for tumor targets (a) and organs‐at‐risk (OARs), such as brainstem and spinal cord (b), parotid glands (c), submandibular (d), esophagus, oral cavity, trachea (e), and larynx, mandible (f).

3.2. Brain cases

Figure 4 shows box plots of defined endpoints of the five cases in this treatment site. For these cases, the maximum doses to the brainstem and chiasm are limiting doses. The maximum dose to the brain stem from the AP plans are comparable to those from the manual plans, exceeding or approaching the dose limit of 60 Gy while the KBP and MCO plans meet or lie below the dose limit [Fig. 4(a)]. The maximum doses to the chiasm are comparable among manual, AP, and KBP plans while the maximum doses to the chiasm from MCO plans are substantially lower than those from the manual plans [Fig. 4(b)]. The results of the maximum doses to optic nerves are mixed among four types of plans [Figs. 4(c) and 4(d)]. In general, the advanced plans had equal or lower maximum dose to the optic nerves than those of the manual plans. For the spinal cord, all three advanced plans achieved lower maximum dose with narrow ranges, indicating a higher consistency in plan quality [Fig. 4(e)]. For other normal structures including eyes, lens, and cochlea, the three advanced plans had equal or lower mean doses than those of manual plans [Figs. 4(g)–4(l)]. Plan conformal indices and homogeneity indices are important for these cases. Advanced plans are less conformal and less homogenous than the manual plans, indicating a trade‐off among the OAR sparing, plan conformity, and homogeneity [Figs. 4(m)–4(n)]. The MUs from KBP plans were the lowest and the MCO plans had the highest MUs in these plans [Fig. 4(f)].

Figure 4.

Figure 4

Comparison of the dosimetric performance between the three advanced plans and manual plans in brain cases for maximum dose to brainstem (a), chiasm (b), optical nerves (c and d), spinal cord (e), globes (g and h), and lens (i and j), mean dose to cochlea (k and l), as well as conformity index (m), homogeneity index (n), and MU (f).

Figure 5 shows representative isodose distributions for a randomly selected brain case. It showed that AP plan was slightly hotter than the manual plan with better sparing to the right eye. However, the 35 Gy isodose lines from the KBP and MCO plans were extended further to the contralateral brain. Figure 6 shows the DVHs for the same brain case as in Fig. 5. In this specific case, except for further decreasing doses to the eyes and lens using advanced plans, the quality of manual plan and advanced plans were comparable, which was reasonably balanced among plan conformity, homogeneity, and spring critical structures (i.e., brain stem and chiasm).

Figure 5.

Figure 5

Representative isodose distributions from Manual, automatic planning (AP), knowledge‐based planning (KBP), and multi‐criteria optimization (MCO) plans for a brain patient.

Figure 6.

Figure 6

Representative DVHs of a brain patient for all four planning methods for tumor targets (a) and organs‐at‐risk (OARs), such as brainstem, chiasm, spinal cord (b), cochlea, globes (c), lens, and optical nerves (d).

3.3. Prostate and pelvic lymph nodes cases

Figure 7 shows box plots of four planning methods based on defined dosimetric endpoints of the five cases in the prostate treatment site. These patients were treated with 70 Gy in 25 fractions to the prostate while concurrently treating 60 or 56 Gy to the seminal vesicle, and 50.4 or 45 Gy to the pelvic lymph nodes. The dose coverage to the prostate met the goal for all plans; the dose coverage to the seminal vesicle and pelvic lymph nodes for advanced plans met the goal while the manual plans were 1–2% below the goal for several cases [Fig. 7(a)], and this relaxed dose coverage allowed manual plan to achieve better mean dose to the penile bulb than the advanced plans [Fig. 7(e)]. For the rectum, advanced plans achieved decreased V63 Gy when compared to those of manual plans [Fig. 7(c)]. However, V45 Gy of the rectum in KBP plans was higher than those in the manual plans, which is likely due to the fact that this dosimetric endpoint was not included in the KBP model [Fig. 7(d)]. The conformity and homogeneity indices in advanced plans were comparable to those in manual plans with slightly worse conformity indices in MCO plans [Figs. 7(g) and 7(h)]. The total MUs were higher in MCO plans, indicating higher modulations in these plans [Fig. 7(f)].

Figure 7.

Figure 7

Comparison of the dosimetric performance between the three advanced plans and manual plans in prostate cases for percentage of volume for lymph node target (a), bladder V63Gy (b), rectum V63Gy (c), rectum V45Gy (d), mean dose to penile bulb (e), as well as conformity index (g), homogeneity index (h), and MU (f).

Figure 8 shows the representative isodose distributions for a randomly selected prostate case. For this particular case, AP and KBP were more conformal to the pelvic lymph nodal volume than the manual and MCO plans. At the low dose level of 35 Gy, AP plan showed the best control on low dose spillage. As see in Fig. 9, the DVHs of the rectum and bladder were improved in three advanced plans when compared to the manual plan.

Figure 8.

Figure 8

Representative isodose distributions from Manual, automatic planning (AP), knowledge‐based planning (KBP), and multi‐criteria optimization (MCO) plans for a prostate patient.

Figure 9.

Figure 9

Representative DVHs of a prostate patient for all four planning methods for PTV targets (a) and organs‐at‐risk (OARs), such as bladder, rectum, small bowel, and penile bulb (b).

3.4. Spine SBRT cases

The spine cases were treated with a prescription dose of 18 or 16 Gy, located at T11, L1, L3‐4, C4, and T3–T4. Figure 10 shows the box plots of four planning methods based on defined endpoints of the five spine cases. The MCO plans achieved the lowest maximum doses to the spinal cord and the KBP plans had the highest maximum dose to the spinal cord [Fig. 10(a)]. The advanced plans had worse V10 Gy to the spinal cord than the manual plans [Fig. 10(b)]. All KPB plans were most homogeneous with trade‐off of the least conformity, while all MCO were most conformal with trade‐off of least homogeneity [Figs. 10(c) and 10(d)]. MCO plans again, have the highest MUs among all plans [Fig. 10(e)]. Figure 11 and 12 show representative isodose distributions and DHVs for a randomly selected spine SBRT case. For this particular case, the MCO plan achieved the lowest maximum dose to the spinal cord and most conformity but with trade‐off of the least plan homogeneity.

Figure 10.

Figure 10

Comparison of the dosimetric performance between the three advanced plans and manual plans in spine cases for maximum dose to the spinal cord (a) and spinal cord V10Gy (b) as well as conformity index (c), homogeneity index (d) and MU (e).

Figure 11.

Figure 11

Representative DVHs of a spine patient for all four planning methods for PTV and spinal cord.

Figure 12.

Figure 12

Representative isodose distributions from Manual, automatic planning (AP), knowledge‐based planning (KBP), and multi‐criteria optimization (MCO) plans for a spine patient.

4. DISCUSSION

To our knowledge, this is the first side‐by‐side comparison of three advanced treatment planning tools with the manual plans. Three institutions participated the study, and each has implemented one of the advanced planning tools clinically. In the current study, the advanced plans from these institutions, however, did not undergo multiple iterative processes as the clinical plans often did. For the research purpose, these advanced plans were created by simply applying the advanced tools with one or two attempts. Therefore, the quality of these research plans did not undergo the clinical evaluation process, specifically physicians' evaluation as the manual plans.

Our results indicated that advanced planning methods are beneficial especially for the complex plans such as head‐and‐neck cancer. Due to the complexity of the tumor shape and location and more involved OARs, it is difficult to take all the planning objectives into account when manually optimizing a plan. Considering substantial variation between cases, the small sample size of five cases may not have enough power to evaluate the statistical difference between the advanced plans and manual plans, however, the trend of improvement for normal tissue sparing is clearly notable for each HN plan. For other disease sites, such as prostate and spine that contain fewer OAR, the advanced plans exhibit similar dose distribution and normal tissue sparing as compared to the manual plans. One possible reason is that both spine and prostate cases contain fewer OAR and tumor targets, and hence it is easier for human planners to achieve a clinically optimized plan with inverse planning tasks. Another possible reason is that there is no further optimization after one round of automatic optimization for the advanced planning methods.

Many previous studies have demonstrated promising results of improvement for OAR sparing using AP. For example, using ten challenging HN cases, Gintz et al. demonstrated that AP can provide lower OAR doses compared to human‐driven VMAT plans despite less homogeneous dose distributions in AP plans.18 Hansen et al. performed a prospective study, in which three senior oncologists blindly compared AP and manual plans and picked 29 out of 30 plans because of better homogenous dose distributions and significantly lower OAR doses without compromising the target coverage.7 In the current study, we did not further optimize the plan after initial automatic process. In contrast, all the previous work performed postoptimization after the automatic step, which included adding more objectives or adjusting weights of the objectives if violations of initial constraints for target or critical organs were found. Without postoptimization our AP plans may not reach the best dose sparing for OARs. As reported by Hazell et al., better OAR sparing was obtained with slightly compromising target coverage relative to the manual plans due to lack of the postoptimization.8

The concept of KBP can be broad, including early attempts of using class solutions,19, 20, 21 applying template‐based planning objectives,22, 23 and model‐based predictive planning objectives.10, 12, 13 The model‐based KBP builds a model that is trained by prior clinically optimal plans from a specific treatment site to predict patient‐specific achievable planning objectives. With the benefit of having patient‐specific achievable planning objectives, the dependence on planner to estimate the optimal planning objectives is largely reduced or eliminated, thus interplanner and interinstitution plan quality variations can be reduced.10 It is worthwhile noting that several aspects can affect the accuracy of the model and consequently the precision of the predicted objectives. For instance, the number of cases available for the model training, the variation range of the anatomy, and the quality of the plans, can all affect the model prediction accuracy.24, 25

MCO provides planners insight into the lowest achievable OAR dose, and navigated trade‐offs to allow physicians to be more involved in the treatment plan decision, which make interaction among treatment planning systems, planners, and physicians more efficient.16, 26, 27 Since the MCO created plans which are all located on the Pareto surface, obviously unacceptable plans were excluded from this dataset, and transiting from one plan to another relies on the priority of OAR sparing based on the patient‐specific situation. Hong et al. reported that MCO is suitable for patient with pancreatic cancer and helps physician select a favorite plan quickly.26 Craft et al. showed that MCO planning time was dramatically shorter than the clinical standard planning time.27

One advantage of these advanced plans is that the plan quality variation is much less than manually optimized plans and interplanner variation is reduced.8 For the manual plan, it is difficult to predict to what extent an OAR can be spared before executing the optimization, and the population‐based guideline was generally set either loose or tight initially. The amount that can be modified in the subsequent planning process depends on the experience of the planner or the clinical turnaround time available. Those advanced planning methods showed potential of improving the plan quality with better dosimetirc performance than manual methods with less variability, which can help transfer planning expertise to the clinical practice where large plan variation exists.28 In addition, the advanced planning methods are not limited to one modality of treatment and can be generalized to other delivery techniques such as TomoTherapy® (Accuray, Inc., Sunnyvale, CA) and Gamma Knife (Elekta, Stockholm, Sweden).29, 30

One limitation of the current study is the small sample size. Considering the complexity of this collaborative work involving three institutions with expertise in different advanced treatment planning systems, only 20 cases were included in this study. While the sample size is small, this multi‐institutional study provides valuable assessment of clinical usage for three important advanced treatment planning tools across four different disease sites. Another limitation of the study is that the manual plans were generated by experienced dosimetrists and, therefore, no substantial reduction of variation in the plans created by the advanced planning methods was observed compared to typical manual plans, which imply that the advanced plans are less dependent on the experience and skills of planners, and able to provide more consistent plans.

5. CONLUSION

Advanced treatment planning tools, including auto‐planning, knowledge based planning, and multi‐criteria optimization, can assist planners to create a better or equivalent plan quality compared to manual plans. These tools are more beneficial for complex cases such as HN patients with multiple PTVs, or cases where critical structures are spared. For less complex cases, the advanced plans showed comparable dosimetric endpoints with the manual plans.

CONFLICT OF INTERESTS

The authors received research support from Philip Medical Systems and Varian Master Research Agreement.

ACKNOWLEGMENTS

The authors would like to acknowledge financial support from NIH 1R01CA201212.

REFERENCES

  • 1. Chung HT, Lee B, Park E, Lu JJ, Xia P. Can all centers plan intensity‐modulated radiotherapy (IMRT) effectively? An external audit of dosimetric comparisons between three‐dimensional conformal radiotherapy and IMRT for adjuvant chemoradiation for gastric cancer. Int J Radiat Oncol Biol Phys. 2008;71:1167–1174. [DOI] [PubMed] [Google Scholar]
  • 2. Das IJ, Moskvin V, Johnstone PA. Analysis of treatment planning time among systems and planners for intensity‐modulated radiation therapy. J Am Coll Radiol. 2009;6:514–517. [DOI] [PubMed] [Google Scholar]
  • 3. Williams MJM, Bailey MJ, Forstner D, Metcalfe PE. Multicentre quality assurance of intensity‐modulated radiation therapy plans: a precursor to clinical trials. Australas Radiol. 2007;51:472–479. [DOI] [PubMed] [Google Scholar]
  • 4. Jeraj R, Wu C, Mackie TR. Optimizer convergence and local minima errors and their clinical importance. Phys Med Biol. 2003;48:2809–2827. [DOI] [PubMed] [Google Scholar]
  • 5. Wu Q, Djajaputra D, Wu Y, Zhou J, Liu HH, Mohan R. Intensity‐modulated radiotherapy optimization with gEUD‐guided dose‐volume objectives. Phys Med Biol. 2003;48:279–291. [DOI] [PubMed] [Google Scholar]
  • 6. Rowbottom CG, Webb S. Configuration space analysis of common cost functions in radiotherapy beam‐weight optimization algorithms. Phys Med Biol. 2002;47:65–77. [DOI] [PubMed] [Google Scholar]
  • 7. Hansen CR, Bertelsen A, Hazell I, et al. Automatic treatment planning improves the clinical quality of head and neck cancer treatment plans. Clin Transl Radiat Oncol. 2016;1:2–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Hazell I, Bzdusek K, Kumar P, et al. Automatic planning of head and neck treatment plans. J Appl Clin Med Phys. 2016;17:5901. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Purdie TG, Dinniwell RE, Letourneau D, Hill C, Sharpe MB. Automated planning of tangential breast intensity‐modulated radiotherapy using heuristic optimization. Int J Radiat Oncol Biol Phys. 2011;81:575–583. [DOI] [PubMed] [Google Scholar]
  • 10. Moore KL, Brame RS, Low DA, Mutic S. Experience‐based quality control of clinical intensity‐modulated radiotherapy planning. Int J Radiat Oncol Biol Phys. 2011;81:545–551. [DOI] [PubMed] [Google Scholar]
  • 11. Wu B, Ricchetti F, Sanguineti G, et al. Data‐driven approach to generating achievable dose‐volume histogram objectives in intensity‐modulated radiotherapy planning. Int J Radiat Oncol Biol Phys. 2011;79:1241–1247. [DOI] [PubMed] [Google Scholar]
  • 12. Yuan L, Ge Y, Lee WR, Yin FF, Kirkpatrick JP, Wu QJ. Quantitative analysis of the factors which affect the interpatient organ‐at‐risk dose sparing variation in IMRT plans. Med Phys. 2012;39:6868–6878. [DOI] [PubMed] [Google Scholar]
  • 13. Zhu X, Ge Y, Li T, Thongphiew D, Yin FF, Wu QJ. A planning quality evaluation tool for prostate adaptive IMRT based on machine learning. Med Phys. 2011;38:719–726. [DOI] [PubMed] [Google Scholar]
  • 14. Craft D, Halabi T, Shih HA, Bortfeld T. An approach for practical multiobjective IMRT treatment planning. Int J Radiat Oncol Biol Phys. 2007;69:1600–1607. [DOI] [PubMed] [Google Scholar]
  • 15. Craft D, Süss P, Bortfeld T. The tradeoff between treatment plan quality and required number of monitor units in intensity‐modulated radiotherapy. Int J Radiat Oncol Biol Phys. 2007;67:1596–1605. [DOI] [PubMed] [Google Scholar]
  • 16. Thieke C, Küfer KH, Monz M. A new concept for interactive radiotherapy planning with multicriteria optimization: first clinical evaluation. Radiother Oncol. 2007;85:292–298. [DOI] [PubMed] [Google Scholar]
  • 17. Craft DL, Halabi TF, Shih HA, Bortfeld TR. Approximating convex pareto surfaces in multiobjective radiotherapy planning. Med Phys. 2006;33:3399–3407. [DOI] [PubMed] [Google Scholar]
  • 18. Gintz D, Latifi K, Caudell J. et al. Initial evaluation of automated treatment planning software. J Appl Clin Med Phys. 2016;17:331–346. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Estall VJ, Eaton D, Burton KE, Jefferies SJ, Jena R, Burnet NG. Intensity‐modulated radiotherapy plan optimisation for skull base lesions: practical class solutions for dose escalation. Clin Oncol. 2010;22:313–320. [DOI] [PubMed] [Google Scholar]
  • 20. Osman SO, Jeevanandam P, Kanakavelu N. et al. Class solutions for SABR‐VMAT for high‐risk prostate cancer with and without elective nodal irradiation. Radiat Oncol. 2016;11:155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Schreibmann E, Xing L. Feasibility study of beam orientation class‐solutions for prostate IMRT. Med Phys. 2004;31:2863–2870. [DOI] [PubMed] [Google Scholar]
  • 22. Xia P, Lee N, Liu YM, et al. A study of planning dose constraints for treatment of nasopharyngeal carcinoma using a commercial inverse treatment planning system. Int J Radiat Oncol Biol Phys. 2004;59:886–896. [DOI] [PubMed] [Google Scholar]
  • 23. Hunt MA, Jackson A, Narayana A, Lee N. Geometric factors influencing dosimetric sparing of the parotid glands using IMRT. Int J Radiat Oncol Biol Phys. 2006;66:296–304. [DOI] [PubMed] [Google Scholar]
  • 24. Boutilier JJ, Craig T, Sharpe MB, Chan TCY. Sample size requirements for knowledge‐based treatment planning. Med Phys. 2016;43:1212–1221. [DOI] [PubMed] [Google Scholar]
  • 25. Sheng Y, Ge Y, Yuan L, Li T, Yin FF, Wu QJ. Outlier identification in radiation therapy knowledge‐based planning: a study of pelvic cases. Med Phys. 2017;44:5617–5626. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Hong TS, Craft DL, Carlsson F, Bortfeld TR. Multicriteria optimization in intensity‐modulated radiation therapy treatment planning for locally advanced cancer of the pancreatic head. Int J Radiat Oncol Biol Phys. 2008;72:1208–1214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Craft DL, Hong TS, Shih HA, Bortfeld TR. Improved planning time and plan quality through multicriteria optimization for intensity‐modulated radiotherapy. Int J Radiat Oncol Biol Phys. 2012;82:e83–e90. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Good D, Lo J, Lee WR, Wu QJ, Yin FF, Das SK. A knowledge‐based approach to improving and homogenizing intensity modulated radiation therapy planning quality among treatment centers: an example application to prostate cancer planning. Int J Radiat Oncol Biol Phys. 2013;87:176–181. [DOI] [PubMed] [Google Scholar]
  • 29. Lian J, Yuan L, Ge Y, et al. Modeling the dosimetry of organ‐at‐risk in head and neck IMRT planning: an intertechnique and interinstitutional study. Med Phys. 2013;40:121704. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Lee KJ, Barber DC, Walton L. Automated gamma knife radiosurgery treatment planning with image registration, data‐mining, and Nelder‐Mead simplex optimization. Med Phys. 2006;33:2532–2540. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley

RESOURCES