Abstract
Controlling light propagation through a step-index multimode optical fiber (MMF) has several important applications, including for biological imaging. However, little consideration has been given to the coupling of fiber and tissue optics. Here, we characterized the effects of tissue-induced light distortions, in particular those arising from a mismatch in the refractive index of the pre-imaging calibration and biological media. By performing the calibration in a medium matching the refractive index of brain, optimal focusing ability was achieved, as well as a gain in focus uniformity within the field-of-view. These changes in illumination resulted in a 30% improvement in spatial resolution and intensity in fluorescence images of beads and live brain tissue. Beyond refractive index matching, our results demonstrate that sample-induced aberrations can severely deteriorate images from MMF-based systems.
Focusing laser light some distance away from the distal end of a MMF (away from the source) is of particular interest in biological sensing and imaging applications where the small diameter (< 0.2 mm) of the MMF minimizes invasiveness and enables multiplexing of measurements [1–3]. As light propagates in a MMF, unpredictable phase delays and coupling occur between modes causing the output intensity profile from a Gaussian input to have a speckle-like appearance. Several methods have been developed to evaluate this deterministic transformation such that the complex light field propagating through MMF could be controlled using active wavefront shaping [4–6]. In each case, a calibration procedure involving distal optical components is necessary; however the impact of distal optical aberrations present during calibration versus imaging is not known. In this Letter, we characterise the effects of a refractive index mismatch between the biological and calibration media on focus formation in the context of brain imaging using one-photon fluorescence point-scanning microscopy through MMF. This type of mismatch is commonly encountered as calibrations are routinely performed in air, while imaging is performed in brain tissue [1–3].
The system comprised 3 units [1, 7]: a source, a wavefront shaping, and a calibration unit (Fig. 1). In the source unit, laser light (CrystaLaser, DL488-020-S, CW, 488 nm) was coupled into two single-mode optical fibers (Thorlabs, P1-488PM-FC-2 and P1-405B-FC-5) for delivery to the wavefront shaping and calibration unit, respectively. The wavefront shaping unit was composed of a liquid-crystal spatial light modulator (LC-SLM, Meadowlark Optics, HSPDM512), the MMF (Thorlabs, FG050UGA, core diameter 50 μm, NA 0.22), and relay optics. The LC-SLM was conjugated to an iris letting through the on-axis first-order diffraction beam for wavefront shaping [8]. The modulated wavefront was then given a circular polarization with a quarter-wave plate (Thorlabs, WPQ05M-488) before it reached the MMF because this polarization state is power-invariant in a MMF [9]. Finally, the calibration unit imaged the MMF output at the distal facet with a microscope objective lens (Olympus, Plan N 20X, NA 0.4) and an achromatic doublet lens (f = 150 mm) on a CCD camera (Basler pilot, piA640-210gm), where it interfered with the reference beam. A second quarter-wave plate was positioned behind the objective to make the polarization linear and parallel to that of the reference beam. The calibration camera was also used to capture the illumination focal volume, but without the reference beam being present.
All fiber segments were cut to a length of 2.0 cm and glued into a ferrule (Thorlabs, CF128-10) before being positioned into the optical system. Subsequently, the wavefront shaping module was aligned with the calibration unit by imaging the distal facet of the MMF on the calibration camera. Then, the coupling of light into the MMF was optimized by translating axially and laterally the lens located immediate before (Thorlabs, C240TME-A, NA 0.5). To focus a known distance away from the MMF distal facet, the wavefront shaping unit was moved by such a distance from the calibration unit before performing the calibration (to select an axial focusing plane by imaging it onto the camera). Data were acquired 50 μm away from the distal facet unless otherwise mentioned. The calibration consisted of the acquisition of the transmission matrix (TM), which describes the field propagation between two planes within the optical system [1]. The TM enabled the determination of the linear superposition of input modes (2401 modes) across the proximal fiber facet, or equivalently the LC-SLM pattern generating the shaped wavefront, required to produce a single focus at a specific location of the distal fiber facet. The fiber could support 2504 modes.
First, we assessed the effect of a refractive index mismatch between the calibration and imaging media on the illumination focus. For a single measurement, the illumination focus was captured with 4 different illumination powers (optical densities from 0 to 3) in order to reconstruct high dynamic range images [3]. Three metrics were used to characterize focus quality: the peak intensity, the intensity enhancement factor (IEF), and the lateral full width at half maximum (FWHM). The IEF was defined as the peak intensity divided by the average intensity from all pixels within the central 50 μm diameter of the field imaged by the fiber. The FWHM was quantified by fitting a Gaussian profile to a line profile drawn through the focus.
The TM is frequently evaluated with the distal facet in air (with an air objective lens corrected for some thickness of glass). For several applications, fibers might be inserted in other milieus than air such as an aqueous solution or biological tissue. The primary optical property affecting focusing that will differ in such cases is the index of refraction (nair = 1.00, nwater = 1.33, and nbrain = 1.37 [10]). In order to quantify the effect of an index of refraction mismatch on focusing, we evaluated the FWHM and IEF of the illumination focal volume when the distal end of MMF were submerged in a solution of 22% glucose to mimic the rodent brain (nglu,22% = 1.37 [11]). A coverglass was located just below the desired focal plane. Calibrations were acquired with a media present between the MMF distal facet and the coverglass, either air or a 22% glucose solution, which will be referred to as the air and glucose calibration for the remainder of this Letter.
At the center of the imaged field, the peak intensity and IEF were 20 ± 11% and 11 ± 10% larger, respectively, when using the calibration acquired in a 22% glucose solution over the one acquired in air (n = 6, one-sided paired Wilcoxon signed rank (opWSR) test, pvalue = 0.016, Fig. 2(a)). Similarly, the FWHM decreased from 1.28 ± 0.05 μm with the air calibration to 1.24 ± 0.05 μm with the 22% glucose one (n = 6, opWSR rank test, pvalue = 0.03, Fig. 2(b)). These results indicate that matching the refractive index of the calibration and imaging media will significantly improve the signal-to-background ratio and the lateral width of the illumination focus. By analyzing the variation in peak intensity and FWHM as a function of the distance from the center of the fiber, we also observed that the peak intensity and FWHM were constant over a larger radius in the index-matched condition, thus providing a more uniform field-of-view (Fig. 2(c,d)). Of note, we also compared focus quality achieved with calibrations acquired in air using an air objective lens corrected for 170 μm of glass with and without a coverglass in place. The illumination focal volume was then evaluated in air immediately at the fiber output in its center. The IEF indicates that significantly better focusing was achieved when a coverglass was used during the calibration (Table 1; 4 segments per group, 5 trials per segments, one-sided paired t-test, ΔIEF = 7%, pvalue = 0.019). This results indicates that aberrations in the calibration unit can also degrade the quality of the illumination focus.
Table 1. Effect of a coverglass during calibration in air.
Coverglass thickness [μm] | FWHM [μm] | IEF [1] |
---|---|---|
0 | 1.30 ± 0.02 | 343 ± 7 |
170 | 1.30 ± 0.02 | 366 ± 5 |
Having demonstrated the improved illumination focus achieved when determining the TM in a medium having the same refractive index as the sample to be imaged, we characterized how calibrations performed in different media impacted the imaging of fluorescent beads. Beads (FluoSpheres carboxylate-modified microspheres, 1.0 μm, yellow-green fluorescent (505/515), 2% solids) were placed on a glass slide and immersed, together with the distal MMF facet, in a 22% glucose solution. Images were acquired with both an air (Fig. 3(a)) and a glucose calibration (Fig. 3(b)) at their respective optimal focal plane. Comparing the two conditions revealed that individual beads had a circular profile in images acquired with the glucose calibration (0.91 ± 0.05), whereas the bead shape was less symmetric for the air calibration (0.79 ± 0.11, circularity, n = 36, opWSR rank test, pvalue = 2.1 × 10−6). Circularity was calculated after fitting an ellipse to individual beads and calculating the ratio of the minor to major axis. We also measured the FWHM of single beads and found that the ratio between the glucose and air calibration was of 0.69 ± 0.17 (n = 51, Δ35%, op-WSR test, pvalue = 3 × 10−9). Most of the variability originated from the FWHM measured with the air calibration (2.1 ± 0.5 μm, Fig. 3(c)), while the FWHM with the glucose calibration was more consistent between beads (1.37 ± 0.13 μm). This observation is in line with a more uniform illumination being achieved with the glucose calibration. However, the significant gain in peak intensity, a factor of 1.3 ± 0.2 (n = 75, opWSR test, pvalue = 1.0 × 10−13, Fig. 3(d)), was larger than that of the illumination peak intensity. This difference likely results from the detected fluorescence signal being a lateral and axial integration within the illumination focus. Thus, the latter result corresponds more closely to the actual improvement that will be achieved in images of biological samples [12].
Deep-brain imaging is an important application for MMF-based microscopes because of the modest damage imparted to surrounding tissue as compared to alternative methods [1–3]. Representative images of a live neuron in organotypic hippocampal slices are shown in Fig. 4(a,b) (Wistar rats, thickness 350 μm). For visualization, CA1 pyramidal cells were patched with Alexa Fluor 488 fluorescent dye (0.5-1 mM, Life Technologies) for 3-5 minutes. To maintain slice health, imaging was performed in physiological Tyrode’s solution (in mM: 120 NaCl, 2.5 KCl, 30 glucose, 2 CaCl2, 1 MgCl2, and 20 HEPES; Sigma Aldrich). Two calibrations were performed: in air and in Tyrode’s solution. With the air calibration, no synaptic specialisations (dendritic spines) were visible (Fig. 4(a)). In contrast, dendritic spines were clearly visible with the Tyrode’s calibration (Fig. 4(b)). This improvement in spatial resolution and signal-to-background ratio was also apparent in the Fourier transform images (Fig. 4(c,d)), revealing the larger spatial frequency bandwidth accessible with the Tyrode’s calibration. In addition, the dendrite appeared tilted with respect to the focal plane with the air calibration. This observation has previously been assumed to reflect the actual position of the dendrite [1], however this is clearly erroneous as the dendrite lay within the focal plane of the Tyrode’s image. In fact, it was the reduced uniformity in focusing ability with the air calibration that generated less resolved structures, appearing out-of-focus, at the circumference of the image.
Fiber-specific calibration is required to shape light through MMF for imaging. It is well known that such calibration will compensate not only distortions occurring inside the fibre, but proximal aberrations as well. We showed that this is not true for distal aberrations. Indeed, the ability to resolve subcellular, synaptic structure, essential when addressing many questions in neuroscience, is substantially improved when one takes account of the refractive index of brain during the calibration. Even when focusing light at 488 nm through only 50 μm of tissue a significant difference was measured, both in terms of focus quality and uniformity across the field-of-view.
The aberrations present when performing the calibration in air are akin to spherical aberrations. They nevertheless differ in that the light field at the distal facet, the interface where the refractive index changes, is not uniform and varies between fibers. Accordingly, the aberrated focal profile will be fiber-dependence, depending on the spatial intensity and phase variations at the interface. It is also interesting to note that the lateral extent of the focus deteriorated when a mismatch was present, while spherical aberrations typically have a limited effect on this parameter [12]. We expect that as the numerical aperture (NA) is increased, so will the aberrations originating from a mismatch in refractive index between the calibration medium and the sample. A fiber with a relatively low NA, 0.22, was used in this study. Using fibers with higher NA is desirable because they would provide an increased spatial resolution, but also because high-NA MMF have been linked to a decreased sensitivity to bend translation [13, 14]. In addition, two-photon fluorescence (2PF) MMF-systems are desirable because of the intrinsic optical sectioning they offer, and high-NA fibers would again be needed as 2PF efficiency increases nonlinearly with the NA [6, 15].
Our results point to the importance of assessing the performance of MMF-based imaging systems, in the context of optical aberrations or in any other context, using the signal of interest. Indeed, the effects of refractive-index matching on fluorescence imaging were substantial, while the improvements in illumination focus were moderate. This suggests that it might not be sufficient to look exclusively at the illumination, as is commonly done for MMF-based imaging system, to characterize the impact on imaging performance. While we have emphasized the significance of these observations for biological imaging, our conclusion relating to the importance of considering distal aberrations, applies to all applications requiring focusing light at the distal end of a MMF, including microfabrication [16] as well as other types of optical manipulation.
Acknowledgments
Biotechnology and Biological Sciences Research Council (BB-SRC) (BB/P0273OX/1); European Research Council Advanced Grant AdOMIS (695140).
The authors thank Rudi Tong for technical assistance.
References
- 1.Vasquez-Lopez SA, Turcotte R, Koren V, Plöschner M, Padamsey Z, Booth MJ, Čižmár T, Emptage NJ. Subcellular spatial resolution achieved for deep-brain imaging in vivo using a minimally invasive multimode fiber. Light Sci & Appl. 2018;7:110. doi: 10.1038/s41377-018-0111-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Turtaev S, Leite IT, Altwegg-Boussac T, Pakan JMP, Rochefort NL, Čižmár T. High-fidelity multimode fibre-based endoscopy for deep brain in vivo imaging. Light Sci & Appl. 2018;7:92. doi: 10.1038/s41377-018-0094-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Ohayon S, Caravaca-Aguirre A, Piestun R, DiCarlo JJ. Minimally invasive multimode optical fiber microendoscope for deep brain fluorescence imaging. Biomed Opt Express. 2018;9:1492. doi: 10.1364/BOE.9.001492. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Di Leonardo R, Bianchi S. Hologram transmission through multimode optical fibers. Opt Express. 2011;19:247–254. doi: 10.1364/OE.19.000247. [DOI] [PubMed] [Google Scholar]
- 5.Čižmár T, Dholakia K. Shaping the light transmission through a multimode optical fibre: complex transformation analysis and applications in biophotonics. Opt Express. 2011;19:18871. doi: 10.1364/OE.19.018871. [DOI] [PubMed] [Google Scholar]
- 6.Morales-Delgado EE, Psaltis D, Moser C. Two-photon imaging through a multimode fiber. Opt Express. 2015;23:32158. doi: 10.1364/OE.23.032158. [DOI] [PubMed] [Google Scholar]
- 7.Plöschner M, Čižmár T. Compact multimode fiber beam-shaping system based on GPU accelerated digital holography. Opt Lett. 2015;40:197. doi: 10.1364/OL.40.000197. [DOI] [PubMed] [Google Scholar]
- 8.Neil MA, Booth MJ, Wilson T. Dynamic wave-front generation for the characterization and testing of optical systems. Opt Lett. 1998;23:1849–1851. doi: 10.1364/ol.23.001849. [DOI] [PubMed] [Google Scholar]
- 9.Plöschner M, Tyc T, Čižmár T. Seeing through chaos in multimode fibres. Nat Photonics. 2015;9:529–535. [Google Scholar]
- 10.Binding J, Ben Arous J, Leger J-F, Gigan S, Boccara C, Bourdieu L. Brain refractive index measured in vivo with high-NA defocus-corrected full-field OCT and consequences for two-photon microscopy. Opt Express. 2011;19:4833–4847. doi: 10.1364/OE.19.004833. [DOI] [PubMed] [Google Scholar]
- 11.Lide DR. CRC Handbook of Chemistry and Physics. 82nd Edition. Taylor & Francis; Cleveland, OH: 2001. [Google Scholar]
- 12.Turcotte R, Liang Y, Ji N. Adaptive optical versus spherical aberration corrections for in vivo brain imaging. Biomed Opt Express. 2017;8:3891–3902. doi: 10.1364/BOE.8.003891. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Loterie D, Psaltis D, Moser C. Bend translation in multimode fiber imaging. Opt Express. 2017;25:6263. doi: 10.1364/OE.25.006263. [DOI] [PubMed] [Google Scholar]
- 14.Bianchi S, Rajamanickam VP, Ferrara L, Di Fabrizio E, Liberale C, Di Leonardo R. Focusing and imaging with increased numerical apertures through multimode fibers with micro-fabricated optics. Opt Lett. 2013;38:4935–4938. doi: 10.1364/OL.38.004935. [DOI] [PubMed] [Google Scholar]
- 15.Morales-Delgado EE, Farahi S, Papadopoulos IN, Psaltis D, Moser C. Delivery of focused short pulses through a multimode fiber. Opt Express. 2017;23:9109–9120. doi: 10.1364/OE.23.009109. [DOI] [PubMed] [Google Scholar]
- 16.Morales-Delgado EE, Urio L, Conkey DB, Stasio N, Psaltis D, Moser C. Three-dimensional microfabrication through a multimode optical fiber. Opt Express. 2017;25:7031. doi: 10.1364/OE.25.007031. [DOI] [PubMed] [Google Scholar]