Abstract
Cephalopelvic disproportion (CPD)-related obstructed labour requires delivery via Caesarean section (C/S); however, in low-resource settings around the world, facilities with C/S capabilities are often far away. This paper reports three low-cost tools to assess the risk of CPD, well before labour, to provide adequate time for referral and planning for delivery. We performed tape measurement- and three-dimensional (3D) camera-based anthropometry, using two 3D cameras (Kinect and Structure) on primigravida, gestational age ≥ 36 weeks, from Addis Ababa, Ethiopia. Novel risk scores were developed and tested to identify models with the highest predicted area under the receiver-operator characteristic curve (AUC), detection rate (true positive rate at a 5% false-positive rate, FPR) and triage rate (true negative rate at a 0% false-negative rate). For tape measure, Kinect and Structure, the detection rates were 53%, 61% and 64% (at 5% FPR), the triage rates were 30%, 56% and 63%, and the AUCs were 0.871, 0.908 and 0.918, respectively. Detection rates were 77%, 80% and 84% at the maximum J-statistic, which corresponded to FPRs of 10%, 15% and 11%, respectively, for tape measure, Kinect and Structure. Thus, tape measurement anthropometry was a very good predictor and Kinect and Structure anthropometry were excellent predictors of CPD risk.
Keywords: obstructed labour, pregnancy risk, low-cost diagnostic, developing world, anthropometry
1. Introduction
Ethiopia had the fourth highest number of maternal deaths worldwide in 2015 [1]. One out of every 64 Ethiopian women die due to complications with pregnancy and 17% of deaths of 15- to 49-year-old Ethiopian women are due to maternal causes [1,2]. The maternal mortality ratios (MMRs), in Ethiopia and across the world, have significantly improved over the past 25 years; however, in 2015, the MMR of 353 deaths per 100 000 live births in Ethiopia is five times higher than Sustainable Development Goal 3.1; namely, the goal to reduce the global MMR to less than 70 per 100 000 live births by 2030. The percentage of deliveries attended by skilled health professionals in Ethiopia is far below the sub-Saharan Africa and worldwide percentages. In 2005, only 6% of deliveries were attended by a skilled health provider; although this percentage increased in the 2016 survey to 28%, Ethiopia remains far below the sub-Saharan Africa and worldwide percentages of 52% and 71%, respectively, and maintains a large gap between rural and urban areas. The cultural aspect of at-home deliveries likely plays an important role in Ethiopia's subpar percentage of attended deliveries. These numbers parallel the low uptake in antenatal care (ANC). In 2015, only 62% of pregnant Ethiopian women sought at least one ANC visit and less than 32% sought four or more ANC visits, compared to 49% of women in sub-Saharan Africa who sought four or more ANC visits [1,2].
Cephalopelvic disproportion (CPD) is an inadequate size of the maternal pelvis, compared to the fetal head, which prevents the fetus from passing through the pelvic cavity during delivery, causing obstructed labour [3]. Obstructed labour accounts for 3–8% of the maternal deaths worldwide and 11% of maternal deaths in Ethiopia [4,5]; however, in some regions of Ethiopia, up to 22% of maternal deaths may be attributed to obstructed labour [6–8]. CPD prevalence increases in regions where girls are small in stature, grow up malnourished, marry at a young age and become pregnant before the pelvis is fully grown [9,10], all of which are common in Ethiopia. For example, in a cross-sectional study of obstetric fistula patients in Ethiopia, the mean ages at the first marriage and at the delivery that caused the fistula were 14.7 and 17.8 years, respectively [11]. Indeed, 13% of the girls surveyed in the study in 2016 between 15 and 19 years of age had begun childbearing, including 1.6% of 15-year-olds, 4.4% of 16-year-olds and 13% of 17-year-olds [2].
CPD-related obstructed labour requires delivery via Caesarean section (C/S). In rural Ethiopia and many low-resource settings around the world, it may require hours or days of travel for mothers to reach facilities with the infrastructure and expertise to perform a C/S. After the onset of labour, it is too late for the mother to travel to these facilities for a C/S and the consequence of CPD-related obstructive labour (in the absence of C/S) is often maternal and/or perinatal mortality or long-term morbidity. Maternal mortality occurs via haemorrhage, puerperal sepsis or rupture of the uterus; perinatal conditions include fetal asphyxia, fetal distress, intracranial haemorrhage and hypoxic isachemic encephalopathy. Long-term maternal morbidity from obstructed labour includes trauma to the bladder that can lead to stress incontinence and obstetric fistulas. An obstetric fistula is a hole that forms in the vaginal wall communicating into the bladder (vesico-vaginal fistula) or the rectum (recto-vaginal fistula), or both [9,12], resulting in urine and/or bowel contents passing uncontrollably through the vagina. In Ethiopia, 1 out of every 250 women report having an obstetric fistula from obstructed labour [2]. Women who develop fistulas are often socially rejected. In one Ethiopian study, more than 50% of the women were rejected and left by their husbands after the fistulae developed [13]. Thus, the consequences of undiagnosed CPD-related obstructed labour go far beyond maternal and perinatal mortality, affecting a significant fraction of the population.
Over the last century, much effort has been dedicated to developing methods to accurately assess the risk of CPD. Clinical pelvimetry requires highly skilled personnel, is subjective and shows low predictive capability. MRI- and CT-pelvimetry are cost prohibitive in low-resource settings and X-ray-pelvimetry poses an unwanted radiation exposure to the fetus. Published anthropometry-based CPD risk scores show only fair predictability. Recently, we demonstrated the feasibility of using a low-cost KinectV2 three-dimensional (3D) camera (Microsoft, Inc.) to generate 3D models of pregnant women, from which anthropometric features are automatically extracted via novel image processing algorithms and used to calculate a novel CPD risk score to identify at-risk mothers [14]. Our novel 3D camera-based CPD risk score performed on-par with our novel MRI-based CPD risk score and out-performed published CPD risk scores based on radiological pelvimetry, clinical pelvimetry and anthropometry. Further, our novel tape measure-based anthropometry risk score showed very good predictive capabilities, albeit not as predictive as the 3D camera-based score.
The goal of this paper is to develop and test three novel tools to assess the risk of CPD-related obstructed labour in primigravida; namely, (i) a Kinect V2 sensor (Microsoft, Inc., developed for the XBox gaming system), combined with a laptop computer, (ii) a Structure 3D camera (Occipital, Inc.), combined with an iOS or Android-based smartphone, and (iii) a tape measure-based anthropometry tool (figure 1). The Kinect and Structure 3D cameras generate digital 3D geometric models of pregnant women, from which anthropometric measurements are automatically extracted via novel image processing algorithms and are used to calculate novel CPD risk scores to identify at-risk mothers. We extend our previous work in several important ways. First, we employ a new 3D camera (Structure), which is smartphone-compatible, offers higher resolution images, and easier image-to-3D model acquisition compared to the Kinect. Second, we studied a larger sample set, consisting of only first-time mothers who tried labour. First-time mothers represent the most important group to assess risk, since they lack previous obstetric history and are at the highest risk for CPD-related obstructed labour. Our central hypothesis is that risk scores based on measurements taken with 3D camera and tape measurements from Ethiopian primigravidae between 36 and 42 weeks of gestation can predict the risk of CPD-related obstructed labour, with a detection rate above 60% at a 5% false-positive rate, a triage rate above 60% at a 0% false-negative rate and an area under the receiver operator curve above 0.90.
2. Material and methods
2.1. Participant enrolment and data collection procedure
This study was approved by the Institutional Review Board at Addis Ababa University, College of Health Sciences (Protocol number: 054/15/gyn, approved on 3 January 2017). Primigravida, 18–40 years of age, at gestational age 36 + 0 weeks or above with vertex presentation and singleton pregnancy were recruited from Tikur Anbessa Specialized Referral Hospital, Girar Health Center and Worreda 08 Health Center in Addis Ababa, Ethiopia. We recruited 685 subjects to participate and combined these data with the data from 148 primigravida from Gleason et al. [14]; in [14], anthropometric measurements and Kinect 3D images were collected, but Structure 3D images were not collected.
Participants entered the exam room and a nurse fully informed the participant of the study and obtained written informed consent. The nurse recorded the participant information, including hospital card number, age and gestational age. The nurse confirmed that the subject satisfies the inclusion/exclusion criteria (i.e. primigravida, singleton pregnancy, vertex presentation, plan for trial of labour, plan to deliver in a health facility). Height and weight were measured with a stadiometer and calibrated scale. Tape measurements of shoulder height, waist height, hip height, head circumference, shoulder diameter, waist circumference, hip circumference and foot length were collected following the published methods [14]. The participant was asked to disrobe (except for tight-fitting undergarments) and stand facing the Kinect camera (Kinect V2 sensor, Microsoft, Inc.), with their arms approximately 45° from the ground and legs spread approximately 50 cm apart. Kinect 3D images were collected from the anterior, posterior, left and right sides, with the Kinect camera remaining stationary while the participant adjusted their position. A Structure 3D camera (Occipital, Inc.) was used to collect a 3D point-cloud image of the subject in the same position; the participant remained stationary, as the nurse moved the Structure camera around the subject to collect the 3D image.
After delivery, the pregnancy outcome was collected both via interview of the participant and review of the participant's hospital card; if necessary, the attending health professional was interviewed regarding the indication for C/S. The nurse recorded the date of birth, the method of delivery (vaginal or C/S), whether C/S was scheduled or performed after trial of labour, the reason the C/S was performed, the infant weight, gender and Apgar scores. Gestational age at delivery was determined and any complications for the mother or the child during labour and delivery were noted. Based on the outcome of the current pregnancy, subjects were categorized into three groups: CPD-related C/S, non-CPD-related C/S or vaginal delivery (VD). We excluded non-CPD-related C/S from analysis; subjects were grouped as CPD (those delivering via a CPD-related C/S) or VD (those having a vaginal delivery).
2.2. Three-dimensional image analysis
2.2.1. Kinect scan analysis
The Kinect software algorithms (Kinect SDK for Window, 2.0, Microsoft, Inc.) determined the location of 25 key joints from the 3D model, which are tracked in real time in three dimensions to quantify movement. From the skeletal maps, we extracted the following features: head height, face height, shoulder height, hip height, shoulder width and hip width. From the 3D point cloud surface maps, an automated MATLAB© code was developed to identify key landmarks (figure 2), from which the following anthropometric measurements were determined: height, shoulder height, waist height, iliac height, symphysis height, hip height, shoulder diameter (measured as the distance between the left and right arm pit), waist diameter (measured as the smallest diameter of the torso), iliac diameter, hip diameter (measured as the widest diameter in the pelvic area), five diameters (Torso Diameter 1, 2, 3, 4 and 5) equally spaced between the hip and shoulder, fundal height, fundal volume and belly height. In addition to assessing measurements from individual views, we also calculated several ratios of measurements.
2.2.2. Structure scan analysis
The Structure scan produces a 3D model of the participant, from which key landmarks were identified and the following anthropometric measurements were determined, using algorithms similar to those used with the Kinect 3D model: height, shoulder height, waist height, iliac height, S1 height, hip height, symphysis height, shoulder diameter, waist diameter, iliac diameter, hip diameter, five diameters (Torso Diameter 1, 2, 3, 4 and 5) equally spaced between the hip and shoulder, waist circumference, hip circumference, external conjugate, fundal height and fundal volume. Several ratios of measurements were also calculated.
2.3. Cephalopelvic disproportion risk score development
Following Gleason et al. [14], the risk scores were defined as
2.1 |
where denotes the risk assessment modality (tape measure, Kinect, Structure), denotes z-score for the ith feature (except age, for which the raw values were used), and are model parameters, xi are the values of the features fi (e.g. fi = height, xi = 155 cm) measured with the respective tool, and are the mean and standard deviation of the feature values across all subjects. For all data, a Grubb's test for outliers was performed and outlier data points were removed from the dataset. Equation (2.1) represents possible permutations of feature sets and parameters that may be calculated for each value of k for all subjects, where is the total number of features considered for inclusion in the model and k is the actual number of features included in the model. The best fit parameters ( and ) for all np possible permutations of feature sets were determined using the glmfit MATLAB© subroutine. For each of the np parameter sets, for each value of k, the perfcurve MATLAB© subroutine was performed to generate a receiver-operator characteristic (ROC) curve. Note that, for Kinect, because np was too large to evaluate all permutation sets, the feature elimination routine of Gleason et al. [14] was employed.
Let us define the following metrics:
2.2 |
where TPR is the true-positive rate (also known as sensitivity), TP is the number of true-positive cases (i.e. CPD cases that score as at risk), is the total number of CPD cases, FNR is the false-negative rate, FN is the number of false-negative cases (i.e. CPD cases that do not score as at risk), FPR is the false-positive rate, FP is the number of false-positive cases (i.e. vaginal deliveries that score as at risk), nVD is the total number vaginal deliveries, TNR is the true-negative rate (also known as specificity) and TN is the number of true-negative cases (i.e. vaginal deliveries that do not score as at risk). From equation (2.2), it can be seen that and . The likelihood ratio for CPD is defined as .
We developed two risk models. The first risk model (model 1) identified a single set of features and parameter values that maximized the area under the ROC, denoted AUC. The second risk model employed three models to maximize the detection rate and triage rate of the CPD risk assessment for each modality; each model had a different set of characteristics (denoted generally as C). For model 2, we identified the model (denoted model 2.1) that maximized TNR at a false FNR of 0%; i.e. (figure 3). With model 2.1, we sought to ‘triage’ as many low-risk subjects as possible, without incorrectly classifying any CPD cases as low risk. Next, we identified model 2.2 that maximized the TPR at an FPR of 5%; i.e. we sought a model that detected as many high-risk subjects, while classifying less than 5% of vaginal deliveries as high risk. On the remaining subjects that were not triaged by model 2.1 or scored as high risk using model 2.2, we applied model 1.
2.4. Model training, validation and testing
CPD cases and VD controls were randomized, using the default randperm random number generator command in MATLAB©; the first 20% of subjects in the randomized sample were assigned as the testing set and the remaining 80% were assigned as the training set. Training and validation were performed on the training set. After the most predictive model was identified, based on the training set, these features and parameters were applied to the testing set to assess the model's predictive capability. For each of the np parameter sets, for each value of k, the characteristic C was calculated with the perfcurve MATLAB© subroutine and the best model was defined as the feature set and model parameters with the greatest C, denoted Cbest, based on the training dataset. To identify the models expected to have the highest predictive capabilities, we evaluated the optimism, O, following the approach of Harrell et al. [14,15]. Briefly, nB = 1000 non-parametric bootstrap samples were generated from the original training set. For each bootstrap sample set, the top feature sets were selected and the model parameters ( and ), for models 1, 2.1 and 2.2, were identified and the value of C for each bootstrap sample, denoted , was calculated. The feature sets and fitted model parameters ( and ) were then applied to the original dataset and the C, denoted , was calculated. The optimism was calculated as . The adjusted C, denoted , which represents the expected value of C if the model was applied to a new dataset (i.e. the predictive capability), was calculated as . For each number of features, k, O and were calculated for the top 1000 feature sets that yielded the highest Cbest. The most predictive model, based on the training set, was defined as the model with the highest Cadj. This model was then applied to the testing set to calculate Ctest. In summary, the training step determined the feature sets and parameters with the highest Cbest, the validation step identified the most predictive model as that with the highest Cadj and the testing set calculated the Ctest using the feature set and model (determined from training set) applied to the testing set.
2.5. Assessing predictive capabilities
To assess the predictive capability, we used a four-point risk score contingency table (very high, high, moderate and mild risk) based on the risk scores from the final model applied to the complete dataset with the features and parameters. The very high–high cut-off was the highest risk value where the FPR was less than or equal to 5%. We define our detection rate as the TPR of subjects scored as very high risk. The high–mild cut-off was defined as the risk score associated with the highest Youden's J-statistic, , which represents distance between the ROC and a diagonal line from [0,0] to [1,1] on the ROC curve. The mild–low cut-off was the highest risk value where TPR = 1. We define our triage rate as subject that score as low risk.
2.6. Statistical analysis
For continuous variables, a two-sample t-test was performed (p < 0.05). Statistical analysis was performed in MATLAB© (MathWorks).
3. Results
3.1. Pregnancy outcomes
Of the 810 primigravida recruited for anthropometry and Kinect imaging, 656 of these subjects were also imaged with the Structure 3D camera. Of the 810 women, 34 (4.2%) women delivered via C/S due to CPD, 655 (80.9%) delivered vaginally (VD) and 121 (14.9%) delivered via C/S for reasons other than CPD. Kinect 3D images were collected for 31 CPD cases and 610 vaginal deliveries; the missed subjects arose due to the loss of power in the exam room, technical difficulties in operation or low-quality images. Structure 3D images were collected for 26 CPD cases and 532 vaginal deliveries. The Structure 3D camera was introduced partway through the study, after the first 148 subjects were recruited. A majority of the anthropometric measurements, made with a tape measure and the Kinect and Structure 3D cameras, were significantly different between the CPD and VD groups (tables 1–3; electronic supplementary material, tables S1.csv and S2.csv).
Table 1.
feature | CPD | VD | model 1 | model 2.1 | model 2.2 | |
---|---|---|---|---|---|---|
(n = 34) | (n = 655) | p-value | −5.975 | −5.800 | −3.710 | |
age (years) | 25.0 ± 4.0 | 23.0 ± 3.3 | 0.001 | 0.08529 | 0.08634 | — |
weight (kg) | 65.8 ± 10.9 | 61.4 ± 8.6 | 0.005 | — | — | — |
tape measurements | ||||||
height (cm) | 154.5 ± 5.0 | 158.6 ± 5.6 | 0.000 | — | — | 1.733 |
shoulder height (cm) | 128.4 ± 5.0 | 131.5 ± 5.2 | 0.001 | — | — | — |
waist height (cm) | 91.3 ± 4.7 | 95.9 ± 4.8 | 0.000 | −0.9011 | −0.9250 | −0.9682 |
hip height (cm) | 86.0 ± 4.7 | 88.9 ± 4.7 | 0.000 | — | — | — |
head circumference (cm) | 56.4 ± 1.6 | 56.3 ± 2.4 | 0.69 | 5.205 | 0.2628 | — |
shoulder diameter (cm) | 38.5 ± 2.7 | 38.7 ± 2.5 | 0.57 | — | — | −5.846 |
waist circumference (cm) | 102.7 ± 6.1 | 97.8 ± 7.1 | 0.000 | −8.508 | −6.257 | — |
hip circumference (cm) | 99.2 ± 9.3 | 96.2 ± 6.5 | 0.009 | — | 6.641 | 2.891 |
foot length (cm) | 23.4 ± 1.2 | 24.0 ± 1.2 | 0.003 | — | — | — |
ratios of measurements | ||||||
BMI | 27.5 ± 4.3 | 24.4 ± 3.1 | 0.000 | 0.6824 | — | 0.9406 |
waist-to-hip ratio | 1.05 ± 0.09 | 1.02 ± 0.05 | 0.001 | 6.453 | 4.960 | — |
head circ.-to-height ratio | 0.37 ± 0.01 | 0.36 ± 0.02 | 0.002 | −6.008 | — | 2.254 |
hip circ.-to-hip height ratio | 1.16 ± 0.12 | 1.08 ± 0.08 | 0.000 | — | — | — |
hip circ.-to-height ratio | 0.64 ± 0.06 | 0.61 ± 0.04 | 0.000 | 7.722 | — | — |
shoulder diameter-to-hip circ. | 0.39 ± 0.04 | 0.40 ± 0.03 | 0.013 | −0.4359 | — | 6.726 |
head circ.-to-hip circ. | 0.58 ± 0.06 | 0.59 ± 0.04 | 0.14 | — | — | −2.953 |
(hip-to-shoulder length)-to-hip height | 0.49 ± 0.04 | 0.48 ± 0.06 | 0.43 | — | — | — |
head circ.-to-foot length ratio | 2.41 ± 0.14 | 2.35 ± 0.14 | 0.009 | — | — | — |
Table 2.
feature | CPD | VD | model 1 | model 2.1 | model 2.2 | |
---|---|---|---|---|---|---|
(n = 31) | (n = 610) | p-value | −6.979 | −3.679 | −5.654 | |
age | 25 ± 4 | 23 ± 3 | 0.000 | 0.1347 | — | 0.0739 |
weight (kg) | 65.8 ± 10.2 | 61.4 ± 8.7 | 0.006 | — | 0.3158 | — |
measurements from skeletal mapping | ||||||
SM: head height (cm) | 146.6 ± 4.9 | 150.2 ± 6.6 | 0.003 | — | — | — |
SM: face height (cm) | 133.1 ± 4.8 | 136.8 ± 6.2 | 0.001 | — | — | −1.264 |
SM: shoulder height (cm) | 125.8 ± 4.9 | 129.4 ± 6.1 | 0.002 | — | — | — |
SM: hip height (cm) | 77.7 ± 5.1 | 83.0 ± 5.2 | 0.000 | — | −1.148 | — |
SM: shoulder width (cm) | 28.6 ± 2.5 | 27.9 ± 2.0 | 0.07 | — | — | 2.569 |
SM: hip width (cm) | 12.5 ± 1.3 | 12.6 ± 1.0 | 0.79 | — | — | — |
measurements from 3D point-cloud | ||||||
3D: height (cm) | 155.6 ± 5.3 | 159.0 ± 7.3 | 0.012 | — | — | — |
3D: shoulder height (cm) | 121.4 ± 5.6 | 125.4 ± 6.0 | 0.000 | — | 2.730 | — |
3D: waist height (cm) | 102.0 ± 6.4 | 106.1 ± 6.2 | 0.000 | — | — | 0.4868 |
3D: illiac height (cm) | 82.8 ± 5.4 | 87.2 ± 5.0 | 0.000 | — | −0.4469 | — |
3D: symphysis height (cm) | 79.7 ± 5.8 | 85.3 ± 5.7 | 0.000 | −0.7362 | — | — |
3D: hip height (cm) | 69.8 ± 6.9 | 74.0 ± 4.9 | 0.000 | — | 0.3995 | — |
3D: shoulder diameter (cm) | 30.5 ± 1.9 | 30.1 ± 2.3 | 0.28 | — | −3.810 | −1.604 |
3D: waist diameter (cm) | 24.4 ± 2.5 | 23.5 ± 2.1 | 0.016 | — | 0.5221 | −0.1861 |
3D: illiac diameter (cm) | 33.5 ± 3.0 | 32.4 ± 2.3 | 0.011 | — | — | — |
3D: hip diameter (cm) | 36.0 ± 3.0 | 35.1 ± 2.6 | 0.08 | — | — | — |
3D: torso diameter 1 (cm) | 33.2 ± 3.0 | 32.6 ± 3.2 | 0.27 | — | — | — |
3D: torso diameter 2 (cm) | 35.1 ± 2.4 | 34.3 ± 2.6 | 0.09 | — | — | −0.1761 |
3D: torso diameter 3 (cm) | 30.1 ± 3.5 | 29.7 ± 3.0 | 0.51 | — | — | — |
3D: torso diameter 4 (cm) | 25.0 ± 2.6 | 24.1 ± 2.2 | 0.029 | — | −0.4033 | — |
3D: torso diameter 5 (cm) | 30.1 ± 4.6 | 28.9 ± 2.8 | 0.034 | — | — | — |
3D: belly height (cm) | 24.5 ± 5.3 | 22.4 ± 5.2 | 0.026 | — | — | — |
3D: fundal height (cm) | 35.8 ± 5.7 | 32.1 ± 4.6 | 0.000 | — | — | — |
3D: fundal volume (cm) | 0.4 ± 0.2 | 0.3 ± 0.1 | 0.000 | 0.3503 | — | 0.4779 |
ratios of measurements | ||||||
hip height-to-hip diameter (3D) | 2.0 ± 0.2 | 2.1 ± 0.2 | 0.001 | — | — | −0.4378 |
hip height-to-hip width (SM) | 6.2 ± 0.7 | 6.6 ± 0.7 | 0.000 | −0.2881 | 0.2918 | — |
belly height-to-hip diameter (3D) | 0.7 ± 0.1 | 0.6 ± 0.2 | 0.036 | — | 0.09400 | — |
belly height (3D)-to-hip width (SM) | 2.0 ± 0.5 | 1.8 ± 0.5 | 0.006 | — | — | 0.8422 |
hip height-to-(hip-to-shoulder) (SM) | 1.6 ± 0.2 | 1.7 ± 0.1 | 0.000 | — | — | — |
head height-to-(hip-to-shoulder) (SM) | 3.0 ± 0.2 | 3.2 ± 0.2 | 0.000 | −0.6134 | — | −0.7463 |
height-to-belly height (3D) | 6.5 ± 1.5 | 7.4 ± 1.6 | 0.004 | 0.4485 | 0.1040 | 1.121 |
head height-to-shoulder width (SM) | 5.1 ± 0.4 | 5.4 ± 0.4 | 0.000 | — | — | — |
shoulder height-to-shoulder width (SM) | 4.4 ± 0.4 | 4.6 ± 0.3 | 0.000 | — | — | 1.970 |
shoulder height-to-shoulder diameter (3D) | 4.0 ± 0.2 | 4.2 ± 0.3 | 0.001 | — | −4.568 | −1.786 |
Table 3.
feature | CPD | VD | model 1 | model 2.1 | model 2.3 | |
---|---|---|---|---|---|---|
(n = 26) | (n = 532) | p-value | −7.4417 | −4.237 | −6.4754 | |
age | 25.0 ± 4.5 | 22.8 ± 3.3 | 0.001 | 0.1334 | — | 0.09969 |
weight (kg) | 64.9 ± 10.1 | 60.4 ± 8.1 | 0.007 | — | — | — |
anthropometry measures | ||||||
height (cm) | 154.2 ± 6.0 | 160.4 ± 6.5 | 0.000 | −2.534 | −2.021 | — |
S1 height (cm) | 79.7 ± 5.2 | 83.6 ± 4.6 | 0.000 | 2.247 | 2.091 | — |
hip height (cm) | 77.0 ± 4.4 | 80.8 ± 4.3 | 0.000 | — | — | — |
symphysis height (cm) | 75.0 ± 5.5 | 81.1 ± 5.5 | 0.000 | −0.9096 | −1.058 | — |
shoulder diameter (cm) | 38.7 ± 3.3 | 37.5 ± 2.6 | 0.026 | — | — | −1.861 |
waist diameter (cm) | 28.5 ± 3.0 | 27.3 ± 1.9 | 0.004 | — | — | — |
iliac diameter (cm) | 33.3 ± 3.0 | 31.8 ± 2.4 | 0.002 | 2.520 | 2.626 | 0.9101 |
hip diameter (cm) | 35.9 ± 3.2 | 34.7 ± 2.4 | 0.020 | — | — | — |
torso diameter 1 (cm) | 35.4 ± 3.2 | 34.2 ± 3.4 | 0.08 | — | — | 0.2823 |
torso diameter 2 (cm) | 36.7 ± 3.3 | 35.3 ± 2.6 | 0.010 | — | — | 0.8183 |
torso diameter 3 (cm) | 33.0 ± 3.3 | 31.4 ± 2.5 | 0.001 | −1.360 | −1.579 | — |
torso diameter 4 (cm) | 29.9 ± 4.3 | 27.9 ± 2.3 | 0.000 | — | — | 0.4089 |
torso diameter 5 (cm) | 36.3 ± 6.0 | 35.1 ± 5.0 | 0.25 | — | — | −0.8137 |
hip circumference (cm) | 116 ± 13 | 115 ± 13 | 0.72 | — | — | — |
waist circumference (cm) | 121 ± 23 | 114 ± 19 | 0.07 | −0.5341 | — | −0.5301 |
external conjugate (cm) | 27.8 ± 3.2 | 27.0 ± 2.5 | 0.14 | — | 0.1232 | — |
fundal height (cm) | 37.5 ± 4.1 | 34.8 ± 4.9 | 0.008 | — | — | — |
fundal volume (cm3) | 0.430 ± 0.129 | 0.322 ± 0.120 | 0.000 | — | −2.756 | −1.972 |
ratios of measurements | ||||||
WHR | 1.02 ± 0.23 | 1.00 ± 0.19 | 0.54 | 0.5478 | — | 0.5157 |
hip height-to-hip width | 2.17 ± 0.24 | 2.34 ± 0.18 | 0.000 | — | — | — |
height-to-hip width | 4.33 ± 0.39 | 4.64 ± 0.31 | 0.000 | — | — | — |
fundal height to hip width | 1.05 ± 0.15 | 1.01 ± 0.15 | 0.13 | — | — | — |
fundal volume to hip width | 12.5 ± 4.23 | 9.33 ± 3.51 | 0.000 | 0.7511 | 3.565 | 2.752 |
height-to-hip height | 2.00 ± 0.06 | 1.99 ± 0.06 | 0.18 | 0.7590 | 0.5347 | — |
shoulder height-to-diameter | 3.28 ± 0.35 | 3.56 ± 0.28 | 0.000 | — | −0.5660 | −2.093 |
fundal height-to-external conjugate | 1.37 ± 0.23 | 1.30 ± 0.22 | 0.12 | — | — | — |
3.2. Cephalopelvic disproportion risk assessment
The AUC for tape measure, Kinect and Structure were 0.868, 0.846 and 0.889 for model 1, respectively, and 0.871, 0.908 and 0.918 for model 2, respectively (figures 4 and 5 and table 4). In general, model 2 outperformed model 1 across the indicators (AUC, detection rate and triage rate). The detection rates at 5% FPR for tape measure, Kinect and Structure were 53%, 47% and 48% for model 1, respectively, and 53%, 61% and 64% for model 2, respectively. The triage rates at 0% FNR for tape measure, Kinect and Structure were 24%, 17% and 53% for model 1, respectively, and 30%, 56% and 63% for model 2, respectively. When applied to the testing set, these models showed good consistency with the predictions from the training set; these results were consistent with the optimism predicted using our validation strategy.
Table 4.
AUC |
detection rate (very high risk) |
triage rate (low risk) |
|||||||
---|---|---|---|---|---|---|---|---|---|
training | testing | combined | training (%) | testing (%) | combined (%) | training (%) | testing (%) | combined (%) | |
anthropometry | |||||||||
model 1 | 0.869 | 0.870 | 0.868 | 59 | 43 | 53 | 23 | 37 | 24 |
model 2 | 0.879 | 0.832 | 0.871 | 67 | 43 | 53 | 42 | 29 | 30 |
Kinect | |||||||||
model 1 | 0.839 | 0.907 | 0.846 | 46 | 17 | 47 | 22 | 79 | 17 |
model 2 | 0.921 | 0.862 | 0.908 | 68 | 50 | 61 | 63 | 57 | 56 |
Structure | |||||||||
model 1 | 0.896 | 0.855 | 0.889 | 57 | 25 | 48 | 53 | 69 | 53 |
model 2 | 0.914 | 0.940 | 0.918 | 62 | 75 | 64 | 64 | 81 | 63 |
The four-point contingency table for model 2 (table 5) shows that 77%, 80% and 84% of CPD cases scored as either very high or high risk, at FPR of 10%, 15% and 11%, respectively, for tape measure, Kinect and Structure. Note that, after removing outliers, the number of subjects included in the risk calculations reduced slightly; cf. tables 1–3 with 5. Perhaps the most relevant difference between the tape measure tool and the 3D camera tools is in the number of subjects in the low-risk (triage) group, with more subjects being scored with the less definitive, more ambiguous diagnosis of mild risk for the tape measure tool. As a result, 57% of the subjects (calculated as (8 + 372)/(34 + 632) from table 5) were scored as mild risk with the tape measure tool, compared to 29 and 25% with the Kinect and Structure tools.
Table 5.
four-point risk score contingency table |
||||
---|---|---|---|---|
very high | high | mild | low | |
tape measure risk score | ||||
CPD (n = 34) | 18 (53%) | 8 (24%) | 8 (24%) | 0 (0%) |
VD (n = 632) | 34 (5%) | 34 (5%) | 372 (59%) | 192 (30%) |
LRCPD = | 9.84 | 4.37 | 0.40 | 0.00 |
Kinect risk score | ||||
CPD (n = 31) | 19 (61%) | 6 (19%) | 6 (19%) | 0 (0%) |
VD (n = 601) | 32 (5%) | 58 (10%) | 176 (29%) | 335 (56%) |
LRCPD = | 11.51 | 2.01 | 0.66 | 0.00 |
Structure risk score | ||||
CPD (n = 25) | 16 (64%) | 5 (20%) | 4 (16%) | 0 (0%) |
VD (n = 508) | 27 (5%) | 30 (6%) | 132 (26%) | 319 (63%) |
LRCPD = | 12.07 | 3.39 | 0.62 | 0.00 |
4. Discussion
CPD-related obstructed labour poses a major global health burden that contributes significantly to maternal and perinatal mortality and morbidity. Ethiopia has a high incidence of CPD-related obstructed labour, low ANC coverage, low rate of attended deliveries and a large portion of the population without nearby access to emergency C/S. Current strategies to assess CPD risk have limited scale-up potential in the developing world.
Over the last century, much effort has been dedicated to developing methods to accurately assess the risk of CPD, prior to labour, including clinical and radiological pelvimetry and anthropometry. Clinical pelvimetry assesses the size of the pelvic cavity by means of the systematic vaginal palpation of specific bony landmarks in the pelvis. This technique requires highly skilled personnel, is subjective and shows low predictive capability [16–18]. Further, given that Ethiopia's densities of doctors (0.025/1000 people) and nurse/midwifery personnel (0.236/1000 people) are among the lowest in the world [19], the number of qualified personnel in Ethiopia is too low to effectively scale this strategy. Radiological pelvimetry uses X-ray, magnetic resonance imaging (MRI) and computed tomography (CT) to provide accurate quantification of the pelvic passage and fetal size, from which a number of CPD risk scores have been proposed [20–27]. MRI and CT are cost prohibitive in low-resource settings and a majority of mothers in the developing world do not have access to MRI or CT. Further, X-ray radiation exposure to the fetus poses an unwanted health risk and may represent a significant barrier to widespread scale-up in the developing world. In the developed world, assessment of fetal size is routinely done via ultrasound imaging; ultrasound does not provide information on pelvic dimensions. Ultrasound technology and expertise are also limited in most low-resource settings. Further, the accuracy of radiological pelvimetry at predicting CPD-related obstructive labour is controversial and lacks testing via rigorous randomized trials [20,28].
Height has long been recognized as a risk indicator for CPD [29–31] across multiple nationalities and regions of the world [29,32–42], including sub-Saharan Africa [31,43–53]; however, the sensitivity and specificity of height as a predictor of CPD are generally only fair at best [31,34,38,45,47,51,53–59]. Other anthropometric measurements or combinations of measurements have also been identified as potential risk indicators for CPD, including foot length [32,42,49,52,60], shoulder diameter [42], lower limb length [49], BMI [39,61–64], maternal head circumference-to-height ratio [39] and external pelvimetry (e.g. external intercrestal, interspinous, intertrochanteric and intertuberous transverse pelvic diameters, anteroposterior external conjugate, and transverse and vertical diagonals of the Michaelis sacral rhomboid area) [35,41,42,52,58]. Published anthropometric-based CPD risk scores only show fair predictability and have not been rigorously employed in clinical practice. Measurements are prone to measurement errors and inter- and intra-user variability [65–68].
Taken together, there is a pressing clinical need to identify novel, low-cost, safe, easy-to-use, scalable methods for timely and accurate assessment of risk of obstructed labour from CPD in low-resource settings. Accurate diagnosis of CPD, well before the onset of labour, using resource-appropriate tools, would allow at-risk mothers to be referred for delivery in a setting where emergency C/S is an option. This paper describes novel innovations designed to address this clinical need.
4.1. Innovation 1: three-dimensional camera-based anthropometry
Our first innovation employs low-cost 3D cameras to generate 3D models of pregnant women, from which anthropometric measurements are automatically extracted via our novel image processing algorithms and used to calculate our novel CPD risk score to identify at-risk mothers. A Kinect V2 sensor, combined with a laptop, was used to generate digital 3D geometric models of pregnant women. Scans that take less than 60 s to collect (30 frames s−1) pose no risk for the mother or fetus, are non-invasive and contact-free and do not require skilled medical personnel to operate or interpret results. The second-generation Kinect V2 sensor, released in 2014 as part of the Xbox One gaming system, includes a high definition (1920 × 1080 pixel) colour camera and a 512 × 424-pixel depth sensor. The depth acquisition is based on the time-of-flight principle where the distance to points on the surface is measured by computing the phase-shift distance of modulated infrared light. Since the depth map is generated with infrared light, it is relatively insensitive to changes in ambient illumination, allowing for gaming in dark and well lit rooms. Skeletal mapping algorithms detect 25 Kinect joint locations [69]; while these joint locations are not necessarily accurate locations of the corresponding anatomical joints, they offer repeatable approximations of these locations [70]. One disadvantage of the Kinect sensor platform is that it currently requires external power to operate. Also, the acquisition of a true three-dimensional model of a subject requires the ‘stitching’ together of 3D images from multiple single-view point clouds [14].
The Structure 3D camera uses structured light technology to generate a fully 3D model of objects and is compatible with iOS- and Android-based smartphones. This device has the advantages of improved resolution of 3D surfaces and easier data-acquisition-to-model-generation capability, compared to the Kinect V2 platform. Further, the smartphone compatibility of the Structure platform does not require a constant power supply, which lends well to scale-up in the developing world. One disadvantage is that the Structure device requires that the participant remain still while being scanned (approx. 60 s); motion during scanning can result in reduced resolution or produce erroneous 3D models. Also, the skeletal mapping capabilities of the Kinect platform are not available for the Structure device. Our results showed that both the Kinect and Structure 3D camera platforms yielded CPD risk scores that outperformed traditional anthropometry; namely, model 2 from the Kinect and Structure showed higher detection rates, triage rates and AUCs, compared to model 2 from the tape measure tool. The Structure 3D platform yielded slightly higher triage rates and AUCs, compared to Kinect, perhaps due to the higher resolution camera and easier data-acquisition-to-model-generation.
While the four-point contingency table (table 5) should be interpreted with caution, since 80% of the data in this table was used to train our CPD risk models, this table may be helpful in illustrating how the CPD risk scores may ultimately be used clinically. The Structure 3D platform showed the highest detection and triage rates across modalities, scoring 64% of the CPD cases as very high risk and another 20% of the cases as high risk and 63% of the vaginal deliveries as low risk. Therefore, with the Structure tool, 43 (8%) of all women (CPD and VD) were scored as very high risk and another 35 (7%) of all women were scored as high risk. Thus, this contingency table suggests that if the 8% of all primigravida that scored as very high risk were referred, 64% of all CPD cases would be captured; if the 15% that scored very high or high were referred, 84% of all CPD cases would be captured, using the Structure tool. Similarly, with the Kinect tool, referring the top 8% (very high risk) would capture 61% of the cases and referring 18% would capture 80% of the cases. Note that the clinically acceptable FPR and how one might recommend to health systems, clinicians and patients on how to use the diagnostic information will certainly require additional research with regard to cost–benefit analysis, availability of resources and many other factors for a given health system.
The Kinect risk score, in the current paper, has a 61% detection rate at a 5% FPR (i.e. very high-risk group) and an 80% detection rate at a 15% FPR (i.e. high + very high group), which is a significant improvement from our previous work [14]. Our previous paper had a 57% detection rate at a 10% FPR. Therefore, the current work showed a higher detection rate (61 versus 57%) at a lower FPR (5 versus 10%). Further, in our previous paper, the low-risk group was defined at an FNR of 10%, while in the current paper, the low-risk group is defined at an FNR of 0%. The demonstration that a risk score can detect 63% of low-risk subjects, before making a false detection is a significant improvement from the models in our previous publication.
4.2. Innovation 2: combining the tape measure-based tool with a three-dimensional camera-based tool
Our results suggest that our novel anthropometry-based risk score that combines multiple anthropometric measurements shows very good predictability. We demonstrated that the tape measure-based CPD risk score (model 2) was able to accurately detect 53% of the CPD cases as very high risk and another 24% as high risk and accurately triage 30% of low-risk subjects. Our tape measure-based risk score, in the current paper, has a 53% detection rate at a 5% FPR (i.e. very high-risk group) and a 77% detection rate at a 10% FPR (i.e. high + very high group). Our previous paper had a 53% detection rate at a 10% FPR. Therefore, the current work improved the detection rate from 53 to 77% at a 10% FPR.
Given the ultra-low cost and ease of use of the tape measure-based tool, this tool may have tremendous scale-up potential; it simply requires a smartphone application (for entry of measurements and calculation of the risk score) and a tape measure. Although our tape measure-based risk score is not as accurate as our 3D camera-based risk scores, particularly showing a much lower triage rate, and is likely more prone to measurement errors, this tool may provide an important ultra-low-cost tool for preliminary screening to ‘detect’ and ‘triage’ a significant number of patients and refer remaining subjects for further diagnosis with the more accurate 3D camera-based tool (figure 6). For example, our preliminary data suggest that we may be able to accurately triage 30% of pregnant mothers at low risk and accurately detect 53% of high-risk cases. This leaves approximately two-thirds of the mothers with an ambiguous diagnosis. On the other hand, our data suggest that the Structure 3D camera-based tools can accurately classify 63% of mothers a low risk and accurately detect 60% of the CPD cases. In the latter case, 25% of mothers would be considered as mild risk, compared to 60% of mothers who would score as mild risk with the tape measurement-based tool. We submit that this multi-stage approach could accelerate the access to CPD risk assessment and would require fewer 3D camera-based devices for a given population, while still providing superior risk assessment, compared to the tape measure-based tool alone.
5. Limitations
This study has several limitations. First, while the sample size was fairly large (n = 810), given the low prevalence of CPD-related C/S (4.2%), the dataset only includes 34 CPD cases. Learning curves developed based on the current dataset suggest that significant improvements in model predictivity could be achieved with a larger sample size. Second, this dataset is based in one hospital and two health centres in Addis Ababa, Ethiopia. Anthropometric measurements and body features differ across different regions and people groups [71]. Even within the nine regional states of Ethiopia, which represent different ethnic and tribal people groups, we expect to see different correlations between anthropometric measurements and CPD. It is also noteworthy that women in the Tigray region report a nearly three times higher prevalence of obstetric fistula, compared to the rest of Ethiopia, suggesting a heightened CPD risk in Tegrayian women [2]. Thus, even within Ethiopia, data from different regions are required, not to mention differences that may arise from cultures around the world. Third, all of our data were collected between 36 and 42 weeks of gestation; however, validating CPD risk assessment tools across a wider range of gestational ages will expand the tools' utility. For example, healthcare providers could assess risk at multiple gestational ages, which could help identify risk earlier in pregnancy and provide patients adequate preparation time. Further, given that a majority of women in Ethiopia have three or fewer ANC visits [2], expanding the validated gestational age range would increase the number of pregnant women who access the tool. Fourth, we have not assessed the robustness of the 3D camera-based anthropometry; there is a need to perform inter- and intra-user variability studies to quantify measurement precision.
6. Conclusion
This work demonstrates the potential of using 3D camera-based anthropometry and traditional anthropometry to assess the risk of CPD-related obstructed labour in Ethiopia. While these tools show significant potential, broader clinical studies across multiple regions and people groups are required, to increase the widespread applicability and the predictive capabilities, prior to translation to use in a clinical setting.
Supplementary Material
Supplementary Material
Acknowledgements
We gratefully acknowledge the support from Grand Challenges Canada and the Saving Lives at Birth Partners, United States Agency for International Development (USAID), Norwegian Agency for Development Cooperation (Norad), Bill & Melinda Gates Foundation, Grand Challenges Canada, UKaid and the Korea International Cooperation Agency (KOICA).
Ethics
This study was approved by the Institutional Review Board at Addis Ababa University, College of Health Sciences (Protocol no. 054/15/gyn, approved on 3 January 2017).
Data accessibility
Traditional anthropometric measurements, Kinect measurements, Structure measurements and pregnancy outcomes are provided as electronic supplementary material files labelled, tables S1.csv and S2.csv.
Authors' contributions
All authors contributed to writing and editing the manuscript. L.T., M.Y., S.T. and R.L.G. contributed to data collection activities. L.T., S.A., M.W., N.F., J.B.D. and R.L.G. contributed to data analysis activities.
Funding
This work was supported by Grand Challenges Canada (grant no. 0748-03).
Competing interests
M.W. and N.F. are employed by LymphaTech, Inc. and J.B.D. owns equity in LymphaTech, Inc. R.L.G., J.B.D. and S.A. were paid consultants to Because of Kennedy to conduct this research. A utility patent (16/143020) is pending for this innovation.
References
- 1.WHO. 2015. Trends in maternal mortality: 1990 to 2015. Geneva, Switzerland: WHO.
- 2.Ethiopia Demographic and Health Survey 2016. 2016. Key indicators report. Addis Ababa, Ethiopia and Rockville, MD: Ethiopian Central Statistical Agency & ICF International. [Google Scholar]
- 3.El-Hamamy E, Arulkumaran S. 2005. Poor progress of labour. Curr. Obstet. Gynaecol. 15, 1–8. ( 10.1016/j.curobgyn.2004.09.001) [DOI] [Google Scholar]
- 4.WHO. 2005. The world health report 2005 make every mother and child count. Geneva, Switzerland: WHO. [Google Scholar]
- 5.Say L, Chou D, Gemmill A, Tunçalp Ö, Moller AB, Daniels J, Gülmezoglu AM, Temmerman M, Alkema L. 2014. Global causes of maternal death: a WHO systematic analysis. Lancet Glob. Health 2, e323–e333. ( 10.1016/S2214-109X(14)70227-X) [DOI] [PubMed] [Google Scholar]
- 6.Warren C. 2010. Care seeking for maternal health: challenges remain for poor women. Ethiop. J. Health Dev. 24, 100–104. ( 10.4314/ejhd.v24i1.62950) [DOI] [Google Scholar]
- 7.Victora CG, et al. 2016. Countdown to 2015: a decade of tracking progress for maternal, newborn, and child survival. Lancet 387, 2049–2059. ( 10.1016/S0140-6736(15)00519-X) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Tessema GA, et al. 2017. Trends and causes of maternal mortality in Ethiopia during 1990–2013: findings from the Global Burden of Diseases study 2013. BMC Public Health 17, 160 ( 10.1186/s12889-017-4071-8) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Wall LL. 2006. Obstetric vesicovaginal fistula as an international public-health problem. Lancet 368, 1201–1209. ( 10.1016/S0140-6736(06)69476-2) [DOI] [PubMed] [Google Scholar]
- 10.Konje JC, Ladipo OA. 2000. Nutrition and obstructed labor. Am. J. Clin. Nutr. 72, 291S–297S. ( 10.1093/ajcn/72.1.291S) [DOI] [PubMed] [Google Scholar]
- 11.Muleta M. 2004. Socio-demographic profile and obstetric experience of fistula patients managed at the Addis Ababa Fistula Hospital. Ethiop. Med. J. 42, 9–16. [PubMed] [Google Scholar]
- 12.Dolea C, AbouZahr C.. 2003. Global burden of obstructed labour in the year 2000. Evidence and Information for Policy (EIP).
- 13.Kelly J. 1995. Ethiopia: an epidemiological study of vesico-vaginal fistula in Addis Ababa. World Health Stat. Q. 48, 15–17. [PubMed] [Google Scholar]
- 14.Gleason RL, et al. 2018. A low-cost, easy-to-use platform to assess risk of obstructed labor due to cephalopelvic disproportion in Ethiopia. PLoS ONE 13, e0203865 ( 10.1371/journal.pone.0203865) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Harrell FE, Lee KL, Mark DB. 1996. Multivariable prognostic models: issues in developing models, evaluating assumptions and adequacy, and measuring and reducing errors. Stat. Med. 15, 361–387. () [DOI] [PubMed] [Google Scholar]
- 16.Anderson DM, Anderson LE, Glanze WD.. 2002. Mosby's medical dictionary. St Louis, MO: Mosby. [Google Scholar]
- 17.Adinma JI, Agbai AO, Anolue FC. 1997. Relevance of clinical pelvimetry to obstetric practice in developing countries. West Afr. J. Med. 16, 40–43. [PubMed] [Google Scholar]
- 18.Maharaj D. 2010. Assessing cephalopelvic disproportion: back to the basics. Obstet. Gynecol. Surv. 65, 387–395. ( 10.1097/OGX.0b013e3181ecdf0c) [DOI] [PubMed] [Google Scholar]
- 19.World Health Organization's Global Health Workforce Statistics. 2017. OECD, supplemented by country data. See https://data.worldbank.org/indicator/SH.MED.NUMW.P3?view=map&year=2010.
- 20.Caldwell WE, Moloy HC. 1933. Anatomical variations in the female pelvis and their effect in labor with a suggested classification. Am. J. Obstet. Gynecol. 26, 479–505. ( 10.1016/S0002-9378(33)90194-5) [DOI] [PubMed] [Google Scholar]
- 21.Ball RP. 1936. Roentgen pelvimetry and fetal cephalometry. Surg. Gynecol. Obstet. 62, 798–810. [Google Scholar]
- 22.Mengert WF. 1948. Estimation of pelvic capacity. J. Am. Med. Assoc. 138, 169–174. ( 10.1001/jama.1948.02900030001001) [DOI] [PubMed] [Google Scholar]
- 23.Abitbol MM, Taylor UB, Castillo I, Rochelson BL. 1991. The cephalopelvic disproportion index. Combined fetal sonography and x-ray pelvimetry for early detection of cephalopelvic disproportion. J. Reprod. Med. 36, 369–373. [PubMed] [Google Scholar]
- 24.Friedman EA, Taylor MB. 1969. A modified nomographic aid for x-ray cephalopelvimetry. Am. J. Obstet. Gynecol. 105, 1110–1115. ( 10.1016/0002-9378(69)90134-3) [DOI] [PubMed] [Google Scholar]
- 25.Spörri S, Thoeny HC, Raio L, Lachat R, Vock P, Schneider H. 2002. MR imaging pelvimetry: a useful adjunct in the treatment of women at risk for dystocia? AJR Am. J. Roentgenol. 179, 137–144. ( 10.2214/ajr.179.1.1790137) [DOI] [PubMed] [Google Scholar]
- 26.Morgan MA, Thurnau GR, Fishburne JI. 1986. The fetal-pelvic index as an indicator of fetal-pelvic disproportion: a preliminary report. Am. J. Obstet. Gynecol. 155, 608–613. ( 10.1016/0002-9378(86)90288-7) [DOI] [PubMed] [Google Scholar]
- 27.Morgan MA, Thurnau GR. 1992. Efficacy of the fetal-pelvic index in nulliparous women at high risk for fetal-pelvic disproportion. Am. J. Obstet. Gynecol. 166, 810–814. ( 10.1016/0002-9378(92)91338-B) [DOI] [PubMed] [Google Scholar]
- 28.Rozenberg P. 2007. Is there a role for X-ray pelvimetry in the twenty-first century? Gynécol. obstét. fertil. 35, 6–12. ( 10.1016/j.gyobfe.2006.09.028) [DOI] [PubMed] [Google Scholar]
- 29.Thomson AM. 1959. Maternal stature and reproductive efficiency. Eugenics Rev. 51, 157–162. [PMC free article] [PubMed] [Google Scholar]
- 30.Baird D. 1945. The influence of social and economic factors on stillbirths and neonatal deaths. J. Obst. Gynaecol. British Empire 52, 339–366. ( 10.1111/j.1471-0528.1945.tb07636.x) [DOI] [Google Scholar]
- 31.Everette VJ. 1975. The relationship between maternal height and cephalopelvic disproportion in Dar es Salaam. East Afr. Med. J. 52, 251–256. [PubMed] [Google Scholar]
- 32.Kennedy JL, Greenwald E. 1981. Correlation of shoe size and obstetric outcome: an anthropometric study. Am. J. Obstet. Gynecol. 140, 466–467. ( 10.1016/0002-9378(81)90048-X) [DOI] [PubMed] [Google Scholar]
- 33.Chen HY, Chen YP, Lee LS, Huang SC. 1982. Pelvimetry of Chinese females with special reference to pelvic type and maternal height. Int. Surg. 67, 57–62. [PubMed] [Google Scholar]
- 34.Kappel B, Eriksen G, Hansen KB, Hvidman L, Krag-Olsen B, Nielsen J, Videbech P, Wohlert M. 1987. Short stature in Scandinavian women—an obstetrical risk factor. Acta Obstet. Gynecol. Scand. 66, 153–158. ( 10.3109/00016348709083038) [DOI] [PubMed] [Google Scholar]
- 35.Hanzal E, Kainz C, Hoffmann G, Deutinger J. 1993. An analysis of the prediction of cephalopelvic disproportion. Arch. Gynecol. Obstet. 253, 161–166. ( 10.1007/BF02766641) [DOI] [PubMed] [Google Scholar]
- 36.Hin LY, Khairuddin Y, Ng KB. 1994. The predictive value of extremes of birth weight, mother's height and ethnic origin on the likelihood of emergency caesarean section. Asia Ocean. J. Obstet. Gynaecol. 20, 389–394. ( 10.1111/j.1447-0756.1994.tb00486.x) [DOI] [PubMed] [Google Scholar]
- 37.McGuinness BJ, Trivedi AN. 1999. Maternal height as a risk factor for caesarean section due to failure to progress in labour. Aust. N. Zeal. J. Obstet. Gynaecol. 39, 152–154. ( 10.1111/j.1479-828X.1999.tb03360.x) [DOI] [PubMed] [Google Scholar]
- 38.Merchant KM, Villar J, Kestler E. 2001. Maternal height and newborn size relative to risk of intrapartum caesarean delivery and perinatal distress. Br. J. Obstet. Gynaecol. 108, 689–696. ( 10.1016/s0306-5456(00)00181-9) [DOI] [PubMed] [Google Scholar]
- 39.Connolly G, Naidoo C, Conroy RM, Byrne P, McKenna P. 2003. A new predictor of cephalopelvic disproportion? J. Obstet. Gynaecol. 23, 27–29. ( 10.1080/0144361021000043173) [DOI] [PubMed] [Google Scholar]
- 40.Khunpradit S, Patumanond J, Tawichasri C. 2005. Risk indicators for cesarean section due to cephalopelvic disproportion in Lamphun hospital. J. Med. Assoc. Thai. 88(Suppl. 2), S63–S68. ( 10.1111/j.1447-0756.2007.00548.x) [DOI] [PubMed] [Google Scholar]
- 41.Bansal S, Guleria K, Agarwal N. 2011. Evaluation of sacral rhomboid dimensions to predict contracted pelvis: a pilot study of Indian primigravidae. J. Obstet. Gynaecol. India 61, 523–527. ( 10.1007/s13224-011-0078-8) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Benjamin SJ, Daniel AB, Kamath A, Ramkumar V. 2012. Anthropometric measurements as predictors of cephalopelvic disproportion. Acta Obstet. Gynecol. Scand. 91, 122–127. ( 10.1111/j.1600-0412.2011.01267.x) [DOI] [PubMed] [Google Scholar]
- 43.Kasongo Project Team. 1984. Antenatal screening for fetopelvic dystocias. A cost-effectiveness approach to the choice of simple indicators for use by auxiliary personnel. J. Trop. Med. Hyg. 87, 173–183. [PubMed] [Google Scholar]
- 44.Adadevoh SWK, Hobbs C, Elkins TE. 1989. The relation of the true conjugate to maternal height and pbstetric performance in Ghanaians. Int. J. Gynaecol. Obstet. 28, 243–251. ( 10.1016/0020-7292(89)90725-X) [DOI] [PubMed] [Google Scholar]
- 45.Sokal D, Sawadogo L, Adjibade A. 1991. Short stature and cephalopelvic disproportion in Burkina Faso, West Africa. Operations Research Team. Int. J. Gynaecol. Obstet. 35, 347–350. ( 10.1016/0020-7292(91)90671-Q) [DOI] [PubMed] [Google Scholar]
- 46.van Roosmalen J, Brand R.. 1992. Maternal height and the outcome of labor in rural Tanzania. Int. J. Gynaecol. Obstet. 37, 169–177. ( 10.1016/0020-7292(92)90377-U) [DOI] [PubMed] [Google Scholar]
- 47.Kwawukume EY, Ghosh TS, Wilson JB. 1993. Maternal height as a predictor of vaginal delivery. Int. J. Gynaecol. Obstet. 41, 27–30. ( 10.1016/0020-7292(93)90150-U) [DOI] [PubMed] [Google Scholar]
- 48.Tsu VD. 1994. Antenatal screening: its use in assessing obstetric risk factors in Zimbabwe. J. Epidemiol. Community Health 48, 297–305. ( 10.1136/jech.48.3.297) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Van Bogaert LJ. 1999. The relation between height, foot length, pelvic adequacy and mode of delivery. Eur. J. Obstet. Gynecol. Reprod. Biol. 82, 195–199. ( 10.1016/S0301-2115(98)00232-2) [DOI] [PubMed] [Google Scholar]
- 50.Ould El Joud D, Bouvier-Colle MH, Group M. 2001. Dystocia: a study of its frequency and risk factors in seven cities of West Africa. Int. J. Gynaecol. Obstet. 74, 171–178. ( 10.1016/S0020-7292(01)00407-6) [DOI] [PubMed] [Google Scholar]
- 51.Brabin L, Verhoeff F, Brabin BJ. 2002. Maternal height, birthweight and cephalo pelvic disproportion in urban Nigeria and rural Malawi. Acta Obstet. Gynecol. Scand. 81, 502–507. ( 10.1034/j.1600-0412.2002.810605.x) [DOI] [PubMed] [Google Scholar]
- 52.Rozenholc AT, Ako SN, Leke RJ, Boulvain M. 2007. The diagnostic accuracy of external pelvimetry and maternal height to predict dystocia in nulliparous women: a study in Cameroon. BJOG: Int. J. Obstet. Gynaecol. 114, 630–635. ( 10.1111/j.1471-0528.2007.01294.x) [DOI] [PubMed] [Google Scholar]
- 53.Aitken IW, Walls B. 1986. Maternal height and cephalopelvic disproportion in Sierra-Leone. Trop. Doct. 16, 132–134. ( 10.1177/004947558601600313) [DOI] [PubMed] [Google Scholar]
- 54.Dujardin B, Van Cutsem R, Lambrechts T.. 1996. The value of maternal height as a risk factor of dystocia: a meta-analysis. Trop. Med. Int. Health 1, 510–521. ( 10.1046/j.1365-3156.1996.d01-83.x) [DOI] [PubMed] [Google Scholar]
- 55.Scott RT, Strickland DM, Hankins GDV, Gilstrap LC. 1989. Maternal height and weight-gain during pregnancy as risk-factors for cesarean-section. Mil. Med. 154, 365–367. ( 10.1093/milmed/154.7.365) [DOI] [PubMed] [Google Scholar]
- 56.Tsu VD. 1992. Maternal height and age: risk factors for cephalopelvic disproportion in Zimbabwe. Int. J. Epidemiol. 21, 941–946. ( 10.1093/ije/21.5.941) [DOI] [PubMed] [Google Scholar]
- 57.Burgess HA. 1997. Anthropometric measures as a predictor of cephalopelvic disproportion. Trop. Doct. 27, 135–138. ( 10.1177/004947559702700305) [DOI] [PubMed] [Google Scholar]
- 58.Liselele HB, Boulvain M, Tshibangu KC, Meuris S. 2000. Maternal height and external pelvimetry to predict cephalopelvic disproportion in nulliparous African women: a cohort study. BJOG: Int. J. Obstet. Gynaecol. 107, 947–952. ( 10.1111/j.1471-0528.2000.tb10394.x) [DOI] [PubMed] [Google Scholar]
- 59.Sheiner E, Levy A, Katz M, Mazor M. 2005. Short stature—an independent risk factor for Cesarean delivery. Eur. J. Obstet. Gynecol. Reprod. Biol. 120, 175–178. ( 10.1016/j.ejogrb.2004.09.013) [DOI] [PubMed] [Google Scholar]
- 60.Frame S, Moore J, Peters A, Hall D. 1985. Maternal height and shoe size as predictors of pelvic disproportion: an assessment. Br. J. Obstet. Gynaecol. 92, 1239–1245. ( 10.1111/j.1471-0528.1985.tb04869.x) [DOI] [PubMed] [Google Scholar]
- 61.Witter FR, Caulfield LE, Stoltzfus RJ. 1995. Influence of maternal anthropometric status and birth-weight on the risk of Cesarean delivery. Obstet. Gynecol. 85, 947–951. ( 10.1016/0029-7844(95)00082-3) [DOI] [PubMed] [Google Scholar]
- 62.Jensen H, Agger AO, Rasmussen KL. 1999. The influence of prepregnancy body mass index on labor complications. Acta Obstet. Gynecol. Scand. 78, 799–802. ( 10.1080/j.1600-0412.1999.780911.x) [DOI] [PubMed] [Google Scholar]
- 63.Young TK, Woodmansee B. 2002. Factors that are associated with cesarean delivery in a large private practice: the importance of prepregnancy body mass index and weight gain. Am. J. Obstet. Gynecol. 187, 312–318. ( 10.1067/mob.2002.126200) [DOI] [PubMed] [Google Scholar]
- 64.Chen G, Uryasev S, Young TK. 2004. On prediction of the cesarean delivery risk in a large private practice. Am. J. Obstet. Gynecol. 191, 616–624. ( 10.1016/j.ajog.2004.07.004) [DOI] [PubMed] [Google Scholar]
- 65.Ayele B, et al. 2012. Reliability of measurements performed by community-drawn anthropometrists from rural Ethiopia. PLoS ONE 7, e30345 ( 10.1371/journal.pone.0030345) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Ulijaszek SJ, Kerr DA. 1999. Anthropometric measurement error and the assessment of nutritional status. Br. J. Nutr. 82, 165–177. ( 10.1017/S0007114599001348) [DOI] [PubMed] [Google Scholar]
- 67.Johnson TS, Engstrom JL, Warda JA, Kabat M, Peters B. 1998. Reliability of length measurements in full-term neonates. J. Obstet. Gynecol. Neonatal Nurs.: JOGNN/NAACOG 27, 270–276. ( 10.1111/j.1552-6909.1998.tb02649.x) [DOI] [PubMed] [Google Scholar]
- 68.Sicotte M, Ledoux M, Zunzunegui M-V, Ag Aboubacrine S, Nguyen V-K, Group A. 2010. Reliability of anthropometric measures in a longitudinal cohort of patients initiating ART in West Africa. BMC Med. Res. Methodol. 10, 102 ( 10.1186/1471-2288-10-102) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Shotton J, et al. 2013. Real-time human pose recognition in parts from single depth images. Commun. ACM 56, 116–124. ( 10.1145/2398356.2398381) [DOI] [Google Scholar]
- 70.Wang Q, Kurillo G, Ofli F, Bajcsy R. 2015. Evaluation of pose tracking accuracy in the first and second generations of Microsoft Kinect. In 2015 Int. Conf. on Healthcare Informatics (ICHI), Dallas, TX, USA, 21–23 October 2015, pp. 380–389. ( 10.1109/ICHI.2015.54) [DOI] [Google Scholar]
- 71.Otieno AO, Mehtre A, Fera O, Lema O, Gebeyehu S. 2016. Developing standard size charts for Ethiopian men between the ages of 18–26 through anthropometric survey. J. Text. Apparel Technol. Manage. 10, 1–10. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Traditional anthropometric measurements, Kinect measurements, Structure measurements and pregnancy outcomes are provided as electronic supplementary material files labelled, tables S1.csv and S2.csv.