Abstract
Background
No data exist on comparisons of efficacy, safety, and recurrence risk factors of paroxysmal and persistent atrial fibrillation (AF) ablation using robotic magnetic navigation system (MNS), respectively.
Methods
About 151 AF patients were prospectively enrolled and divided into paroxysmal AF group (n = 102) and persistent AF group (n = 49). Circumferential pulmonary vein antrum isolation (CPVI) was performed in all patients. Linear ablation at the left atrial roof and mitral isthmus was performed in patients with persistent AF in addition to CPVI. The procedural time, X‐ray exposure time, acute and long‐term success rates of CPVI, and procedure‐related complications were analyzed. The AF recurrence rates in the two groups were compared during 1 year, and Cox regression was used to analyze the recurrence risk factors.
Results
The acute success rates of CPVI in the two groups were 98.04% and 97.96%, respectively. There were no significant differences in the procedural time, X‐ray exposure time, and ablation time between the two groups (P > 0.05). No serious complications appeared in either group. The AF ablation success rates were 70.6% and 57.1% for the paroxysmal and persistent groups respectively at 12‐month follow‐up (P = 0.102). AF duration and coronary heart disease prior to ablation were associated with the higher AF recurrence in patients with persistent AF.
Conclusion
Ablation using MNS is effective and safe both in patients with paroxysmal and persistent AF. AF duration and coronary heart disease prior to ablation are two independent risk factors of AF recurrence in patients with persistent AF postoperatively.
Keywords: atrial fibrillation, catheter ablation, magnetic navigation system, recurrence, risk factor
1. INTRODUCTION
Atrial fibrillation (AF) is the most common sustained arrhythmia, and it is associated with higher mortality and morbidity,1 with similar outcomes amongst current treatment strategies. Currently, catheter ablation has become one of the preferred methods for AF treatment2, 3; however, manual AF ablation has some disadvantages that include a complex procedure, long procedural times, long learning curve, risk of serious complications, and high recurrence rate. Consequently, finding new techniques for AF ablation that provide clinical benefit for the treatment of AF patients is important.
In the recent decade, ablation using the Niobe II robotic magnetic navigation system (MNS) (Stereotaxis, Inc. St. Louis, Missouri) has been one of the novel techniques used to address the challenges of AF ablation procedures.4, 5, 6 MNS procedures have fewer complications and reduced fluoroscopy exposure time to patients, physicians and Ep lab staff. Many studies have demonstrated that ablation using MNS is noninferior to manual ablation in efficacy and safety.7, 8 On the other hand, as we expected based on numerous prior studies, the efficacy of manual ablation was lower for persistent than for paroxysmal AF. However, there are still no data on comparisons of efficacy and safety of paroxysmal and persistent AF ablation using MNS respectively. Therefore, the aim of this study was to investigate the efficacy, safety, and recurrence risk factors of catheter ablation using MNS for paroxysmal and persistent AF, respectively.
2. METHODS
2.1. Study population
In this prospective study, a total of 151 consecutive AF patients, including 102 paroxysmal and 49 persistent AF patients, with mean age of 63.9 ± 7.8 (18‐77) years, were enrolled for AF ablation using MNS from October 2012 to February 2017. These patients were all first time ablation procedures, and no redo patients were included.
2.2. Criteria of inclusion and exclusion
Inclusion criteria were patients with drug‐refractory symptomatic AF, more than one risk factor of thrombosis, and no thrombus found by transoesophageal echocardiography examination in left atrium prior to ablation. Exclusion criteria were patients with ages >80 or < 18 years, severely decompensated heart failure (ejection fraction (EF) ≤40%), left atrium diameter ≥ 55 mm, contraindications to anticoagulation, severe hepatic, renal or lung dysfunction, hyperthyroidism, and thrombus in the left atrium illustrated on either echocardiography or computed tomography (CT) scan. The potential risks of the procedure were explained, and written informed consent was obtained from all patients. Owing to good efficacy and better safety, AF ablation was performed in all cases using MNS in our center. This study was reviewed and approved by the Institutional Review Board of Wuxi People's Hospital affiliated to Nanjing Medical University, China.
2.3. Definitions of paroxysmal AF and persistent AF
In this study, AF was categorized as paroxysmal and persistent. Paroxysmal AF was defined as self‐terminating within 7 days or terminated with pharmacological or electrical cardioversion; persistent AF was defined as lasting >7 days, requiring cardioversion or other intervention or failed cardioversion, or cardioversion was no longer attempted.
2.4. Electrophysiological study and catheter ablation
The MNS Niobe II system (Stereotaxis Inc.) is composed of two focused‐field permanent magnets, irrigated magnetic catheter and Cardiodrive catheter advancing system (QuikCAS; Stereotaxis, Inc.). The two large magnets with 0.08 to 0.1 T magnetic field are located on either side of the patients' body. The magnetic catheter can be actively deflected by changing the magnetic field orientations. The third‐generation irrigated magnet catheter (Navistar Thermocool RMT; Biosense Webster, Inc. Diamond Bar, California) can be steered leftwards or rightwards following the direction of the vectors by magnet field‐controlled system, and can be advanced or retracted with minimal 1 mm step by mechanical device. Three‐dimensional electroanatomic mapping along side an X‐ray image (CARTO RMT; Biosense Webster Inc.) can be displayed on MNS monitor. Ablation was performed remotely in the control room, away from radiation exposure, by clicking the computer mouse and keyboard of MNS.
The decapolar catheter (Inquiry, St. Jude Medical, Inc. St. Paul, Minnesota) was routinely placed in coronary sinus via the right femoral vein. Transseptal puncture was done under the fluoroscopic guidance, and the angiographies of left and right pulmonary veins were performed after atrial septal puncture. The MNS catheter, used both for mapping and ablation, was moved to left atrium through the sheaths (Swartz SR0, St. Jude Medical Inc.). Using the preset navigation feature, the catheter can virtually move automatically to the target pulmonary veins after clicking the short‐cut menu. Heparin was given to maintain ACT of 250 to 300 seconds during the whole ablation procedure. Circumferential pulmonary vein antrum isolation (CPVI) only was performed in patients with paroxysmal AF. Linear ablation at left atrial roof and mitral isthmus was done in patients with persistent AF in addition to CPVI (Figure 1). Electrical cardioversion was carried out with failure to maintain sinus rhythm after ablation. The duration of radiofrequency ablation on each point was set at 30 to 60 seconds, or the potential amplitude declined by 80%. The ablation was performed with the temperature setting at 43°C, power of 30 to 35 W for the anterior wall and 25 to 30 W for the posterior wall, with a flow rate of 17 mL/min during ablation and 2 mL/min during ablation free interval. Finally, CPVI was confirmed by Lasso catheter at each pulmonary antrum. Procedure‐related parameters such as acute success rate of CPVI, procedure time, mapping and CPVI isolation time, radiofrequency ablation time, exposure to X‐ray time, and ablation complications were recorded.
Figure 1.

Catheter ablation of paroxysmal (left) and persistent (right) atrial fibrillation using magnetic navigation system. Left panel is paroxysmal atrial fibrillation ablation using magnetic navigation system, and the ablation protocol is only circumferential pulmonary vein antrum isolation. Right panel is persistent atrial fibrillation ablation using magnetic navigation system, and the ablation protocol is the linear ablation of mitral isthmus and left atrial roof in addition to circumferential pulmonary vein antrum isolation. The upper panel and the lower panel in left and right are the pictures at the same time, but the upper panel can simultaneously show three‐dimensional model of left atrium and ablation lines on fluoroscopy monitor, and the lower panel only shows ablation lines on fluoroscopy monitor
2.5. Follow‐ups
Follow‐ups were performed for all patients in outpatient clinic or by telephone interviews at 1, 3, 6, 9, and 12 months post procedure. Oral anti‐coagulation therapy with a target international normalized ratio between 1.8 and 2.5 and anti‐arrhythmic drugs were administered for 2 to 3 months after the procedure. In addition to the clinic follow‐up and telephone interview, 24‐hour Holter monitoring was routinely performed every 2 months post‐ablation in order to obtain the patient clinical manifestations and the occurrence of arrhythmias. The efficacy endpoints of AF ablation using MNS evaluated in this study were acute success rate of CPVI for the paroxysmal group, and CPVI plus lines for the persistent group, and the AF recurrence at 1, 3, 6, 9, and 12 months postoperatively. AF recurrence was defined as a documented episode of AF or atrial tachycardias lasting more than 30 seconds within 1 year from ablation date. The safety of AF ablation using MNS was evaluated by assessing the rate of procedure‐related complications perioperatively and the time of X‐ray exposure.
2.6. Statistical analysis
Continuous variables were expressed as mean values ± SD (−x ± s) or as median with inter‐quartile range. Categorical variables were expressed as ratios and percentages. SPSS version 22.0 (SPSS Inc, Chicago, Illinois) was used for statistical analysis. Normally distributed data were compared using the independent Student's t test. Non‐normally distributed data between two groups were compared using the Mann‐Whitney U test. The χ2 test or the Fisher exact test was used to measure the association for categorical variables. The Pearson or the Spearman correlation analysis was used to identify the related risk factors of AF recurrence after ablation. A Kaplan‐Meier analysis was used to compare the probability of AF recurrence in two groups. Univariate analysis and multivariate analysis were performed using Cox‐regression test to identify the risk factors of AF recurrence after ablation. A P value <0.05 was considered statistically significant.
3. RESULTS
3.1. Patient characteristics
The clinical baseline characteristics of patients in the two groups were nearly identical. There were no statistically significant differences between groups in terms of sex, age, weight, AF duration, heart function class, left ventricular ejection fraction, number of anti‐arrhythmic drugs, underlying diseases and lab results, but significant differences in diameters and volumes of left atrium were noted. (Table 1).
Table 1.
Baseline characteristics of patients prior to ablation
| Variable | All patients (n = 151) | Paroxysmal AF (n = 102) | Persistent AF (n = 49) | P value |
|---|---|---|---|---|
| Sex (M/F) | 93/58 | 59/43 | 34/15 | 0.172 |
| Age (years) | 58.62 ± 10.32 | 59.25 ± 9.86 | 57.33 ± 11.21 | 0.290 |
| Weight (kg) | 68.57 ± 10.10 | 68.77 ± 11.26 | 70.68 ± 11.06 | 0.403 |
| AF duration (month) | 36 | 36 | 24 | 0.174 |
| LAD (mm) | 39.01 ± 5.12 | 37.57 ± 4.36 | 42 ± 5.34 | 0.000 |
| LAV (mL) | 105.41 ± 36.66 | 93.82 ± 26.84 | 129.54 ± 42.48 | 0.000 |
| EF (%) | 63.48 ± 4.45 | 63.92 ± 4.26 | 62.55 ± 4.73 | 0.076 |
| Heart Function Class (NYHA) | 1.25 ± 0.47 | 1.21 ± 0.41 | 1.34 ± 0.57 | 0.086 |
| No. of antiarrhythmic drugs (n) | 1.81 ± 0.44 | 1.79 ± 0.45 | 1.84 ± 0.43 | 0.573 |
| Underlying diseases (n) | ||||
| HTN | 73 | 49 | 24 | 0.914 |
| CAD | 23 | 13 | 10 | 0.220 |
| DM | 3 | 3 | 0 | 0.551 |
| Hyperthyroidism | 2 | 2 | 0 | 1.000 |
| ACI | 2 | 0 | 2 | 0.104 |
| Others | 2 | 1 | 1 | 0.545 |
| Absence of underlying diseases (n) | 46 | 34 | 12 | 0.269 |
| Lab examinations | ||||
| BUN (mmol/L) | 4.81 ± 1.48 | 4.75 ± 1.56 | 4.94 ± 1.32 | 0.450 |
| Cr (mmol/L) | 78.15 ± 16.83 | 76.65 ± 16.08 | 81.42 ± 18.11 | 1.110 |
| FBG (mmol/L) | 5.10 ± 1.37 | 4.97 ± 0.81 | 5.37 ± 2.12 | 0.100 |
| TC (mmol/L) | 4.15 ± 0.82 | 4.15 ± 0.79 | 4.15 ± 0.87 | 0.950 |
| LDL (mmol/L) | 2.24 ± 0.74 | 2.21 ± 0.75 | 2.29 ± 0.73 | 0.560 |
| TG (mmol/L) | 2.10 ± 1.14 | 1.90 ± 1.47 | 2.08 ± 1.32 | 0.288 |
Abbreviations: LAD, left atrial diameter; LAV, left atrial volume; HTN, hypertension; CAD, coronary artery disease; DM, diabetes mellitus; ACI, acute cerebral infarction; BUN, blood urea nitrogen; Cr, creatinine; FBG, fasting blood glucose; TC, total cholesterol; LDL, low‐density lipoprotein; TG, triglyceride.
3.2. Procedural parameters and acute success rates of CPVI
There were no remarkable differences of procedure duration, time from left atrium electroanatomic mapping to pulmonary vein isolation, time of radiofrequency catheter ablation, fluoroscopy exposure time, and the incidence of complications during the procedures between the two groups. The acute success rates of CPVI in the paroxysmal and persistent groups were 98.04% and 97.96% (P = 0.974), respectively (Table 2).
Table 2.
Procedure parameters and follow‐up in two groups
| Variable | Paroxysmal AF (n = 102) | Persistent AF (n = 49) | P value |
|---|---|---|---|
| Procedure time (min) | 146.73 ± 23.05 | 152.45 ± 28.78 | 0.230 |
| Mapping and CPVI time (min) | 101.86 ± 21.17 | 110.31 ± 28.24 | 1.857 |
| Ablation time (min) | 38.75 ± 8.37 | 39.63 ± 7.54 | 0.532 |
| X‐ray exposure time (min) | 9.70 ± 5.90 | 10.15 ± 5.01 | 0.640 |
| Sheath (n) | |||
| SR0 | 99 | 49 | 0.551 |
| SR1 | 3 | 0 | 0.551 |
| Dosage of heparin (u) | 5619.09 ± 1500.87 | 5864.58 ± 1511.44 | 0.353 |
| Cases of complications (n) | |||
| Sudden death | 0 | 0 | — |
| Cardiac perforation | 0 | 0 | — |
| Esophageal fistula | 0 | 0 | — |
| PVS | 0 | 0 | — |
| Thrombus | 0 | 0 | — |
| Hemothorax | 2 | 0 | 1.000 |
| Hematoma | 0 | 1 | 0.327 |
| Hemopneumothorax | 0 | 0 | — |
| Acute success rates of CPVI (%) | 98.04 | 97.96 | 0.974 |
| Percents of free from AF (%) | |||
| 1 mo | 88.20 | 81.60 | 0.272 |
| 3 mo | 83.30 | 75.50 | 0.253 |
| 6 mo | 80.40 | 67.30 | 0.078 |
| 9 mo | 72.50 | 61.20 | 0.159 |
| 12 mo | 70.60 | 57.10 | 0.102 |
Abbreviations: AF, atrial fibrillation; CPVI, pulmonary vein isolation; PVS, pulmonary vein stenosis.
3.3. Recurrence rates of AF
The recurrence rates of paroxysmal AF were 11.8%, 16.7%, 19.6%, 27.5%, and 29.4% and those of persistent AF were 18.4%, 20.5%, 32.7%, 38.8%, and 42.9% during the follow‐ups of 1, 3, 6, 9, and 12 months after ablation. There were no statistical differences in AF recurrence rates between the two groups though AF recurrence rates in patients with persistent AF tended to be higher than those in patients with paroxysmal AF during each follow up time period (Table 2). The Kaplan‐Meier analysis showed that the incidences of freedom from AF were 70.6% and 57.1% in patients with paroxysmal AF and persistent AF respectively at 12‐month follow‐up, and there were no statistical significance (Log rank X2 = 1.23, P = 0.267) (Figure 2).
Figure 2.

Kaplan‐Meier curve of atrial fibrillation recurrence after ablation using magnetic navigation system. A Kaplan‐Meier analysis was used to compare the probability of atrial fibrillation recurrence between the two groups. The Kaplan‐Meier analysis showed that the incidences of freedom from atrial fibrillation were 70.6% and 57.1% in patients with paroxysmal atrial fibrillation and paroxysmal atrial fibrillation respectively after 12‐month follow‐up
3.4. Predictors of AF recurrence
Cox regression of univariate analysis demonstrated that the recurrence of paroxysmal AF was not associated with sex, age, weight, duration of AF, diameter of left atrium, heart function class, left ventricular ejection fraction, and some common underlying diseases (r < 0.3, P > 0.05). However, Cox regression of univariate and multivariate analysis showed that the duration of AF and coronary heart disease prior to ablation were two independent risk factors of predicting the recurrence of persistent AF postoperatively (Table 3).
Table 3.
Risk factors of atrial fibrillation recurrence after ablation guided by magnetic navigation system
| Variable | Paroxysmal AF | Persistent AF | ||||||
|---|---|---|---|---|---|---|---|---|
| Univariate analysis | Univariate analysis | Multivariate analysis | ||||||
| Score | P value | Score | P value | Wald | Risk ratio | 95% CI | P value | |
| Sex | 2.120 | 0.145 | 2.968 | 0.085 | ||||
| Age | 0.979 | 0.322 | 0.115 | 0.735 | ||||
| Weight | 2.798 | 0.094 | 0.060 | 0.806 | ||||
| AF duration | 0.000 | 0.989 | 6.142 | 0.013 | 4.458 | 1.938 | 1.049‐3.583 | 0.035 |
| LAD | 0.284 | 0.594 | 2.123 | 0.145 | ||||
| Heart functional class | 0.006 | 0.939 | 4.670 | 0.031 | ||||
| EF | 0.103 | 0.749 | 4.071 | 0.044 | ||||
| Underlying diseases | ||||||||
| HTN | 1.109 | 0.292 | 0.471 | 0.492 | ||||
| CAD | 0.027 | 0.870 | 6.463 | 0.011 | 4.323 | 2.641 | 1.057‐6.599 | 0.038 |
| DM | 0.231 | 0.631 | 3.704 | 0.054 | ||||
| Hyperthyroidism | 0.028 | 0.867 | 0.474 | 0.491 | ||||
| ACI | 0.103 | 0.749 | 0.474 | 0.491 | ||||
Abbreviations: LAD, left atrial diameter; HTN, hypertension; CAD, coronary artery disease; DM, diabetes mellitus; ACI, acute cerebral infarction.
3.5. Procedural complications
No serious complications such as sudden death, cardiac perforation, cardiac tamponade, pulmonary vein stenosis, and esophageal fistula appeared in two groups. Hemothorax occurred in two patients with paroxysmal AF and groin hematoma occurred in one patient with persistent AF.
4. DISCUSSION
4.1. Main findings
AF ablation using MNS is performed routinely; however, no study has focused on efficacy, safety, and recurrence risk factors of paroxysmal and persistent AF ablation using MNS. This study is the first, to the best of our knowledge, to investigate and compare the efficacy, safety and recurrence risk factors of catheter ablation using MNS for paroxysmal and persistent AF. The main findings of this study were as follows: (a) AF ablation using MNS was effective and safe both in patients with paroxysmal AF and patients with persistent AF; (b) AF duration and coronary heart disease prior to ablation were two independent risk factors of AF recurrence in patients with persistent AF postoperatively.
4.2. Development of AF ablation guided by MNS
It has been proven that catheter ablation is one of the effective treatments for AF patients. However, the conventional manual ablation technique for AF ablation has potential disadvantages of complex procedures, prolonged radiation exposure to patients and operators, increased risk of complications, and lengthened learning curve and so on. Therefore, advancements to remedy these issues will have great clinical implications especially those that simplify the ablation procedure, minimize the risk of complications, reduce the fluoroscopic exposure time, and decrease the learning curve.
In the last decade, it has demonstrated that AF ablation using MNS may be one of the promising ablation techniques that may address issues surrounding conventional manual ablation techniques.9, 10 In 2006, Pappone et al11 first reported the MNS‐guided AF ablation. In 2007, Di Biase et al12 conducted a further study on feasibility and effectiveness of AF ablation using MNS and this study found that it was difficult to achieve CPVI in most of cases with non‐irrigated catheter ablation due to the formation of the char in heart. However, with the application of saline‐irrigated magnetic catheters in 2008, AF ablation using MNS came into clinical practice again.13, 14 In the last decade, many clinical trials have confirmed that there are the same acute and long‐term success rates of AF ablation, less procedural complications, and better safety in MNS‐guided AF ablation compared with the conventional manual ablation.13, 14, 15 Currently, MNS‐guided AF ablation is widely used in many developed countries.16 However, there is still no comparative study on paroxysmal and persistent AF ablation using MNS. Therefore, this study is mainly focused on efficacy, safety, and risk factors of AF recurrence for paroxysmal and persistent AF after ablation guided by MNS.
4.3. Efficacy of AF ablation guided by MNS
Despite different ablation protocols for patients with paroxysmal and persistent AF, there were no significant statistical differences in acute success rates of CPVI, mapping time, ablation duration, total procedure time, X‐ray exposure, and incidence of procedure‐related complications or recurrence rates in follow‐up. In our study, the recurrence rates were statistically the same between the two groups, however, trended higher in the persistent AF cohort. These results suggested that efficacy and safety were the same for paroxysmal and persistent AF ablation using MNS. The reason may be related to the following factors: (a) The MNS ablation catheter has better stability and remains in constant contact with heart tissue regardless of heart rhythm / rate and respiration, which makes a higher efficacy of ablation lesions.17 (b) Irrigated magnetic catheters can reach the most difficult anatomic regions of left atrium due to its tip flexibility, which improves the success rate of CPVI. In addition, computer controlled catheter advancement system allows the operators to control the catheters even with 1 mm forward or backward step, and at 1° angle to leftward or rightward, which increases the precision when ablation in the target sites. (c) These advantages may be more prominent during persistent AF ablation using MNS. In general, the sizes of left atrium in persistent AF are usually larger than those in paroxysmal AF, and it is difficult for maneuver manual catheters during the ablation procedure. However, in contrast to manual control ablation, magnetic ablation catheters in large left atrium target areas are easy to reach allowing full contact with the left or right pulmonary vein antrum, which improves the efficacy of persistent AF ablation.18
Owing to better stability and maneuverability of magnetic ablation catheters, many single‐center studies and meta analyses have illustrated that the acute success rates of CPVI in MNS‐guided catheter ablation is as good as manual navigation ablation,14, 16, 19, 20 and the long‐term success rate may be better during the follow‐up.21 Weiss JP et al22 performed a comparison of remote magnetic irrigated tip ablation vs manual catheter irrigated tip catheter ablation with and without force sensing feedback and found MNS resulted in outcomes similar to manual navigation. The addition of contact force sensing catheters did not improve relative procedural outcome in comparison to MNS‐guided AF ablation.
4.4. Safety of AF ablation using MNS
Studies have demonstrated that the serious complications such as cardiac perforation, pulmonary vein stenosis, and left atrial esophageal fistula can be significantly decreased when AF ablation is performed using MNS,23, 24 and furthermore the incidence of complications in MNS‐guided ablation was much lower, compared with conventional manual ablation.25, 26
There is significant decrease in the risk of cardiac perforation and cardiac tamponade in MNS‐guided AF ablation due to its special soft catheter. This is achieved by stable catheter‐to‐tissue contact even on the setting of constant changes of patients' heart rate and breath. In contrast, it is difficult for the operators to precisely control the contact force to the catheter lesion during manual ablation.27 Shurrab et al28 compared the safety of MNS ablation catheter and SmartTouch ablation catheter (Biosense Webster Inc.). The study showed that the incidence of complications with MNS ablation catheter was only 0.03%, while it was as high as 2.3% to 4% with SmartTouch ablation catheter. In our study, no serious complications occurred both in patients with paroxysmal AF and patients with persistent AF.
4.5. Risk factors of AF recurrence after ablation using MNS
AF recurrence rates after single catheter ablation may range from 30% to 50%, often requiring repeated ablation and leading to increased treatment costs. Therefore, the accurate prediction of negative events in patients undergoing AF ablation has important clinical significance. 29
There are few data on the recurrence risk factors of paroxysmal and persistent AF ablation using MNS. Sohns et al30 reported that pulmonary vein anatomy assessed by multidetector CT is a good predictor of AF recurrence after PVA using MNS. To explore the risk factors of AF recurrence after ablation using MNS in patients with paroxysmal and persistent AF, univariate analysis and multivariate analysis were performed using Cox‐regression test to identify the risk factors of AF recurrence after ablation. The results showed that the recurrence of paroxysmal AF following ablation using MNS was not correlated with sex, age, weight, AF duration, inner diameter of left atrium, heart function class, ejection fraction, and some common underlying diseases; therefore, the risk factors for the recurrence of paroxysmal AF still need to be further studied. Nevertheless, we found that AF duration and coronary heart disease were two independent risk factors for the recurrence of persistent AF, which means atrial fibrosis due to long‐term AF duration and myocardial ischemia may the key causes of higher recurrence in patients with persistent AF after ablation using MNS.
Recently, more attention has been paid on predictors of improvement in symptoms and health‐related quality of life after AF ablation in addition to predictors of AF recurrence.31 If predictors of AF recurrence are combined with predictors of improvement in symptoms and health‐related quality of life, they may better evaluate the effects of AF ablation using MNS for paroxysmal AF and persistent AF, respectively.
4.6. Study limitations
There are at least two limitations in this study. First, the study is a single‐center study, and the patients enrolled in this study are most of those with almost normal heart function. Second, the number of enrolled cases is not large, especially for persistent AF, only 49 cases. A multiple center study is needed to further explore efficacy, safety, and recurrence risk factors of catheter ablation using MNS for paroxysmal AF and persistent AF respectively in future.
5. CONCLUSIONS
Ablation using MNS is effective and safe both in patients with paroxysmal and persistent AF. AF duration and coronary heart disease prior to ablation are two independent risk factors of AF recurrence in patients with persistent AF postoperatively.
CONFLICTS OF INTEREST
The authors have no conflicts of interest to report.
ACKNOWLEDGMENTS
This study was supported in part by grants from the National Natural Science Foundation of China (81770331) and Natural Science Foundation of Jiangsu Province (BK20151110) to Dr. Ru‐Xing Wang.
Zhao X‐X, Li K‐L, Wang R‐X, et al. Comparisons of efficacy, safety, and recurrence risk factors of paroxysmal and persistent atrial fibrillation catheter ablation using robotic magnetic navigation system. Clin Cardiol. 2019;42:418–424. 10.1002/clc.23156
Xiao‐Xi Zhao and Ku‐Lin Li contributed equally to the work.
Funding information Dr. Ru‐Xing Wang, Grant/Award Number: National Natural Science Foundation of China (81770331) and Natural Science Foundation of Jiangsu Province (BK20151110)
Contributor Information
Ru‐Xing Wang, Email: ruxingw@aliyun.com.
Xiang‐Jun Yang, Email: xiangjun_yang218@163.com.
REFERENCES
- 1. Magnani JW, Rienstra M, Lin H, et al. Atrial fibrillation: current knowledge and future directions in epidemiology and genomics. Circulation. 2011;124:1982‐1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. January CT, Wann LS, Alpert JS, et al. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2014;64:e1‐e76. [DOI] [PubMed] [Google Scholar]
- 3. Andrade JG, Macle L, Nattel S, Verma A, Cairns J. Contemporary atrial fibrillation management: a comparison of the current AHA/ACC/HRS, CCS, and ESC guidelines. Can J Cardiol. 2017;33:965‐976. [DOI] [PubMed] [Google Scholar]
- 4. Maurer T, Sohns C, Deiss S, et al. Significant reduction in procedure duration in remote magnetic‐guided catheter ablation of atrial fibrillation using the third‐generation magnetic navigation system. J Interv Card Electrophysiol. 2017;49:219‐226. [DOI] [PubMed] [Google Scholar]
- 5. Errahmouni A, Latcu DG, Bun SS, Rijo N, Dugourd C, Saoudi N. Remotely controlled steerable sheath improves result and procedural parameters of atrial fibrillation ablation with magnetic navigation. Europace. 2015;17:1045‐1050. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Sohns C, Bergau L, Seegers J, Lüthje L, Vollmann D, Zabel M. Single‐ring ablation compared with standard circumferential pulmonary vein isolation using remote magnetic catheter navigation. J Interv Card Electrophysiol. 2014;41:75‐82. [DOI] [PubMed] [Google Scholar]
- 7. Adragão PP, Cavaco D, Ferreira AM, et al. Safety and long‐term outcomes of catheter ablation of atrial fibrillation using magnetic navigation versus manual conventional ablation: A propensity‐score analysis. J Cardiovasc Electrophysiol. 2016;27(Suppl 1):S11‐S16. [DOI] [PubMed] [Google Scholar]
- 8. Lim PCY, Toh JJH, Loh JKXY, et al. Remote magnetic catheter navigation versus conventional ablation in atrial fibrillation ablation: fluoroscopy reduction. J Arrhythm. 2017;33:167‐171. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Szili‐Torok T, Akca F. Remote magnetic navigation in atrial fibrillation. Expert Rev Med Devices. 2012;9:249‐255. [DOI] [PubMed] [Google Scholar]
- 10. Bradfield J, Tung R, Mandapati R, et al. Catheter ablation utilizing remote magnetic navigation: a review of applications and outcomes. Pacing Clin Electrophysiol. 2012;35:1021‐1034. [DOI] [PubMed] [Google Scholar]
- 11. Pappone C, Vicedomini G, Manguso F, et al. Robotic magnetic navigation for atrial fibrillation ablation. J Am Coll Cardiol. 2006;47:1390‐1400. [DOI] [PubMed] [Google Scholar]
- 12. Di Biase L, Fahmy TS, Patel D, et al. Remote magnetic navigation: human experience in pulmonary vein ablation. J Am Coll Cardiol. 2007;50:868‐874. [DOI] [PubMed] [Google Scholar]
- 13. Luthje L, Vollmann D, Seeger J, et al. Remote magnetic versus manual catheter navigation for circumferential pulmonary vein ablation in patients with atrial fibrillation. Clin Res Cardiol. 2011;100:1003‐1011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Choi MS, Oh YS, Jang SW, et al. Comparison of magnetic navigation system and conventional method in catheter ablation of atrial fibrillation: is magnetic navigation system is more effective and safer than conventional method? Korean Circ J. 2011;41:248‐252. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Pappone C, Vicedomini G, Frigoli E, et al. Irrigated‐tip magnetic catheter ablation of AF: a long‐term prospective study in 130 patients. Heart Rhythm. 2011;8:8‐15. [DOI] [PubMed] [Google Scholar]
- 16. Proietti R, Pecoraro V, Di Biase L, et al. Remote magnetic with open‐irrigated catheter vs. manual navigation for ablation of atrial fibrillation: a systematic review and meta‐analysis. Europace. 2013;15:1241‐1248. [DOI] [PubMed] [Google Scholar]
- 17. Pezawas T, Ristl R, Bilinski M, Schukro C, Schmidinger H. Single remote‐magnetic catheter approach for pulmonary vein isolation in patients with paroxysmal and non‐paroxysmal atrial fibrillation. Int J Cardiol. 2014;174:18‐24. [DOI] [PubMed] [Google Scholar]
- 18. Bun SS, Ayari A, Latcu DG, Errahmouni A, Saoudi N. Radiofrequency catheter ablation of atrial fibrillation: Electrical modification suggesting transmurality is faster achieved with remote magnetic catheter in comparison with contact force use. J Cardiovasc Electrophysiol. 2017;28:745‐753. [DOI] [PubMed] [Google Scholar]
- 19. Kataria V, Berte B, Vandekerckhove Y, et al. Remote magnetic versus manual navigation for radiofrequency ablation of paroxysmal atrial fibrillation: long‐term, controlled data in a large cohort. Biomed Res Int. 2017;2017:6323729. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Aksu T, Bozyel S, Golcuk E, et al. Atrial fibrillation ablation using magnetic navigation comparison with conventional approach during long‐term follow‐up. J Atr Fibrillation. 2015;8:1276‐1283. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Yuan S, Holmqvist F, Kongstad O, et al. Long‐term outcomes of the current remote magnetic catheter navigation technique for ablation of atrial fibrillation. Scand Cardiovasc J. 2017;51:308‐315. [DOI] [PubMed] [Google Scholar]
- 22. Weiss JP, May HT, Bair TL, et al. A comparison of remote magnetic irrigated tip ablation versus manual catheter irrigated tip catheter ablation with and without force sensing feedback. J Cardiovasc Electrophysiol. 2016;27(Suppl 1):S5‐S10. [DOI] [PubMed] [Google Scholar]
- 23. CappatoR CH, Chen SA, et al. Worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation. Circulation. 2005;111:1100‐1105. [DOI] [PubMed] [Google Scholar]
- 24. Jin Q, Pehrson S, Jacobsen PK, et al. Efficacy and safety of atrial fibrillation ablation using remote magnetic navigation: experience from 1,006 procedures. J Cardiovasc Electrophysiol. 2016;27(Suppl 1):S23‐S38. [DOI] [PubMed] [Google Scholar]
- 25. Cappato R, Calkins H, Chen S, et al. Updated worldwide survey on the methods, efficacy and safety of catheter. Circ Arrhythm Electrophysiol. 2010;3:32‐38. [DOI] [PubMed] [Google Scholar]
- 26. Danon A, Shurrab M, Nair KM, et al. Atrial fibrillation ablation using remote magnetic navigation and the risk of atrial‐esophageal fistula: international multicenter experience. J Interv Card Electrophysiol. 2015;43:169‐174. [DOI] [PubMed] [Google Scholar]
- 27. Pappone C, Augello G, Sala S, et al. A randomized trial of circumferential pulmonary vein ablation versus antiarrhythmic drug therapy in paroxysmal atrial fibrillation: the APAF Study. J Am Coll Cardio1. 2006;48:2340‐2347. [DOI] [PubMed] [Google Scholar]
- 28. Shurrab M, Danon A, Lashevsky I, et al. Robotically assisted ablation of atrial fibrillation: a systematic review and meta‐analysis. Int J Cardiol. 2013;169:157‐165. [DOI] [PubMed] [Google Scholar]
- 29. Kosich F, Schumacher K, Potpara T, Lip GY, Hindricks G, Kornej J. Clinical scores used for the prediction of negative events in patients undergoing catheter ablation for atrial fibrillation. Clin Cardiol. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Sohns C, Sohns JM, Bergau L, et al. Pulmonary vein anatomy predicts freedom from atrial fibrillation using remote magnetic navigation for circumferential pulmonary vein ablation. Europace. 2013;15:1136‐1142. [DOI] [PubMed] [Google Scholar]
- 31. Barmano N, Charitakis E, Karlsson JE, Nystrom FH, Walfridsson H, Walfridsson U. Predictors of improvement in arrhythmia‐specific symptoms and health‐related quality of life after catheter ablation of atrial fibrillation. Clin Cardiol. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]