Contrast Enhanced Ultrasound Imaging by Nature-Inspired Ultrastable Echogenic Nanobubbles

Al de Leon 1, Reshani Perera 2, Christopher Hernandez 2, Michaela Cooley 2, Olive Jung 2, Selva Jeganathan 2, Eric Abenojar 1, Grace Fishbein 3, Amin Jafari Sojahrood 3, Corey C Emerson 4, Phoebe L Stewart 4, Michael C Kolios 3, Agata A Exner 1,2
PMCID: PMC6716144  NIHMSID: NIHMS1047462  PMID: 31408083

Abstract

Advancement of ultrasound molecular imaging applications requires not only a reduction in size of the ultrasound contrast agents (UCAs) but also a significant improvement in the in vivo stability of the shell-stabilized gas bubble. The transition from first generation to second generation UCAs was marked by an advancement in stability as air was replaced by a hydrophobic gas, such as perfluoropropane and sulfur hexafluoride. Further improvement can be realized by focusing on how well the UCAs shell can retain the encapsulated gas under extreme mechanical deformations. Here we report the next generation of UCAs for which we engineered the shell structure to impart much better stability under repeated prolonged oscillation due to ultrasound, and large changes in shear and turbulence as it circulates within the body. By adapting an architecture with two layers of contrasting elastic properties similar to bacterial cell envelopes, our ultrastable nanobubbles (NBs) withstand continuous in vitro exposure to ultrasound with minimal signal decay and have a significant delay on the onset of in vivo signal decay in kidney, liver, and tumor. Development of ultrastable NBs can potentially expand the role of ultrasound in molecular imaging, theranostics, and drug delivery.

Graphical Abstract:

graphic file with name nihms-1047462-f0001.jpg

We report an ultrasound contrast agent for which we engineered the shell structure to impart much better stability under intense stress and deformation.

Introduction

With an estimated 2.8 billion imaging procedures done in 2011 worldwide, ultrasound is among the most widely used and safest modalities for visualizing tissues and organs in real-time.1 Therefore, a slight advancement on ultrasound technology can have a potentially large impact in the medical field. As an alternative to commercial UCAs, nanodroplets, nanovesicles, and NBs have been studied.29 Of growing interest is the use of NBs as UCAs for cancer detection, tumor characterization, image-guided surgeries and biopsies, as well as theranostics.1018 Clinically-used UCAs (1 −10 μm microbubbles) are limited to intravascular applications due to their larger size.12 For both bubbles, one of the main limitations is the rapid in vivo signal decay; commercially available microbubble formulations have a signal half-life of under 2 min.1922 Development of stable UCAs has been focused primarily on reducing surface tension, varying membrane components, counteracting Laplace pressure, preventing bubble coalescence, and utilizing hydrophobic gases.15,2327 Still, most of the UCAs that show extended stability in vitro do not perform as well when continuously insonated or used in vivo because their shell is not designed to endure large deformations. Insonated UCAs undergo fast and extreme oscillations for which gas leaks out during expansion and shell materials shed off during compression.28,29 Furthermore, UCAs experience strong shear and turbulence as they circulate within the body.30,31 Ultrastable NBs can be designed by engineering the bubble membrane shell to be able to withstand deformations due to ultrasound and flow in vivo.

One particular membrane architecture that has been shown to redistribute stress, dissipate excess energy, and deform without irreversible damage is a bilayer with contrasting elastic properties. Membranes comprised of contrasting elastic properties have been shown to provide synergistic improvement in toughness and resilience against defects, stress, and deformations.3234 In nature, this architecture can be found not only in the bacterial cell envelope, but also apparent in the design of interfaces in mammalian blood vessels, skin, and red blood cells.35,36 We hypothesize that adapting this architecture can impart NBs an enhanced capacity to minimize gas leakage and membrane shedding during insonation and therefore provide better persistence as it circulates within the body.

Development of ultrastable NBs can facilitate diagnostic ultrasound imaging of various malignancies by providing more accurate, less transient visualization of tumor vasculature and with appropriate molecular targets, detection of cellular biomarkers. In addition, for molecular imaging, stable echogenic NBs can enable a more reliable extravasation beyond the tumor vasculature. Likewise, for drug delivery, a more robust shell can prevent pre-mature release of loaded drug far from the target sites brought about by the collapse of NBs in circulation. Here, we engineered the membrane of ultrastable NBs to be composed of a compliant phospholipid layer by incorporating propylene glycol (PG), an edge-activator used in ultra-deformable liposomal formulations, and glycerol (Gly), a membrane stiffener known to increase buckling of lipid monolayers (Figure 1a).3742 The unique interfacial design reduces in vitro signal decay of NBs under continuous exposure to ultrasound of varying strength and pulse repetition frequencies, enabling a substantial improvement in both in vivo half-life and in vivo signal decay onset as compared to a common commercially available UCAs. The oscillation behavior of single NB with different membrane composition and stiffness is studied. In addition, a rigorous comprehensive comparison of peak negative pressure and transmit frequency on the effect of shell components in vitro and in vivo, showing significantly distinct behavior between the four formulations, was done. A comprehensive physical, chemical, structural, and acoustic characterization of NBs, and conclusively demonstrate the advantage of having both PG and glycerol in the bubble membrane in terms of NB stability was also provided.

Figure 1. Schematic, size, concentration, and structural property of NBs.

Figure 1.

(a) Schematic representation of the membrane of the ultrastable NB showing a bilayer architecture with elastic contrast. (b). Cryo-electron microscopy (cryo-EM) images of PG-Gly-PL showing the nanobubble membrane and the dense C3F8 gas core. Diameter (c) and concentration (d) measurement of NB via a resonant mass measurement capable of differentiating buoyant (NBs) and non-buoyant particles (micelles, liposomes, etc.). Mean ± SE (n = 5) (e) X-ray diffractogram of the membrane of the freeze-dried NBs. The NBs were isolated and freeze-dried overnight. Peak corresponding to d = 4.5 Å is observed for all NB but only PG-PL and PG-Gly-PL have a peak corresponding to d = 10.3 Å implying the incorporation of PG into the phospholipid membrane.

Results and Discussion

NBs containing a perfluoropropane (C3F8) gas core were prepared by dissolution of a phospholipid mixture in PBS solution containing PG and Gly (PG-Gly-PL), followed by activation via amalgamation and isolation by differential centrifugation.26,27 To directly compare the effects of the glycerol and propylene glycol we also prepared phospholipid stabilized NBs without glycerol (propylene glycol only (PG-PL)), without propylene glycol (glycerol only (Gly-PL)), or using phospholipids alone (PL) without additives. Cryo-EM images shown in Fig 1b shows a dense C3F8 gas core enclosed by a phospholipid membrane.43 Particle size distribution and concentration for each NB type were determined through resonant mass measurement capable of distinguishing buoyant and non-buoyant particles (liposomes, micelles, etc).4448 The average diameter of NB was measured to be 274 ± 8 nm, at a concentration of 4.07 × 1011 ± 3.15 × 1010 NBs/mL, with no particles larger than 1 μm observed in the population (Figure 1c, 1d, S1). Cryo-electron microscopy confirms an average diameter of ~200 nm and reveals the nanobubble membrane and dense C3F8 gas core (Figures 1b). Aside from NBs, non-buoyant particles were also measured to be 194 ± 3 nm in average diameter and concentration of 6.32 × 1010 ± 1.58 × 1010 particles/mL. These non-buoyant particles, which are invisible under ultrasound, were reported to sterically hinder coalescence of UCAs, thereby contributing to their overall stability.49 NBs in PBS showed a slightly negative charge as measured via zeta potential, with no significant differences among formulations (Table S2). Although the isolated NBs have similar size, charge, and concentration, analysis of the shell material by X-ray diffraction (XRD) showed an additional broad amorphous peak centered at 2θ = ~10° for PG-PL and PG-Gly-PL (Figure 1e) suggesting that PG expands the distance between phospholipids from ~4.2 Å to ~10.3 Å at some locations.50,51

This is consistent with the reported interaction of PG with phospholipids. Warren et al., showed through molecular simulation that PG incorporates between lipids with increasing degree as the concentration of PG in water increases.52 Harvey et al., used SANS and SAXS to study the interdigitation of phospholipid due to the increased head group solvation by PG, partitioning of PG into the hydrophobic core of the bilayers, or to a combination of these two phenomena. In addition, wide-angle X-ray scattering shows the existence of a gel phase at all concentrations of PG at 25°C due to the association of PG with the lipid headgroups, causing an increase in molecular area.53 Boncheva, et al. examined the effect of PG on the molecular organization of lipid matrix in intact Stratum corneum (SC) using Fourier Transform infrared/attenuated total reflection (FT-IR/ATR).54 By investigating the increase in the degree of CH2-scissoring of the lipid, they concluded that incorporation of PG to the lipid membrane changes its lateral organization. With this regard, PG has been used as an edge-activator into membranes of ultradeformable liposomes.51,5557 These ultradeformable liposomes have been reported to deform as they pass through a narrow channel (e.g. skin pore) and reform without disruption of their vesicular structure.37,38,5861 Of note, the broad peak centered at 2θ = ~10° is not present for pure components (Figure S2a) and no preferred crystalline orientation is observed in 2D XRD (Figure S2bd).62

The ability of a NB to function as an UCA depends on the acoustic impedance difference between the gas and its surrounding and the nonlinearity of the NB oscillation.6365 Tissues do not exhibit nonlinear scattering of the ultrasound wave, enabling high contrast, which is ideally suited for molecular imaging. Therefore, understanding the nonlinear behavior of NB oscillation can facilitate better parameter optimization for molecular imaging and theranostics.66 Subharmonic, ultraharmonic, second harmonic, as well as fundamental harmonic oscillations, were observed when single NBs were insonated by a 100% bandwidth 25 MHz ultrasound transducer and pressure amplitude of 300 kPa as shown in Figure 2a. Dilute suspensions of NBs were introduced in PBS, and the scattering of single NBs were recorded following a procedure described elsewhere.67,68 PG-PL, with the most flexible shell, shows the largest oscillation amplitude for both fundamental (f) and second harmonic (2f) oscillations. This is consistent with the Marmottant model, which predicts that the backscatter intensity is directly proportional to shell flexibility.6971 In contrast, Gly-PL shows the lowest oscillation amplitude for fundamental and second harmonic, suggesting that it has a stiffer shell than PG-PL. The interaction between Gly and phospholipids has been well-studied experimentally, numerically, and via molecular dynamics.39,40,7275 Recently, Terakosolphan, et al. found via molecular dynamics simulation that glycerol preferentially interacts with the phosphate and choline in the headgroup of phospholipids. Furthermore, these are experimentally confirmed via Langmuir pressure-area isotherms, FT-IR spectroscopy, fluorescence-based membrane transition temperature determination, and small angle neutron scattering.39 Pocivavsek et al., in particular, reported the formation of a 10 Å-thick Gly-enriched vitrified adlayer at the phospholipid/solvent interface as measured by neutron and X-ray reflectivity.41 In the same study, it was shown that the Gly adlayer stiffens the phospholipid monolayer at the air-water interface enabling its folding/buckling normal to the interface when compressed. This stiffening, caused by inter- and intra-hydrogen bonding with the phospholipid, limits the oscillation of Gly-PL NBs when exposed to ultrasound field. PG-Gly-PL, which has a membrane stiffness between that of PG-PL and Gly-PL, shows intermediate amplitude for both fundamental and second harmonic. Furthermore, the PG-Gly-PL NBs undergo a highly stable oscillation with minimal change in amplitude with time. Aside from fundamental and second harmonic, subharmonic (f/2) and ultraharmonic (3f/2) signals can also be detected for all samples further confirming the non-linear nature of oscillation of NB.

Figure 2. Nonlinear oscillation of a single NB and a solution of NBs.

Figure 2.

(a) Representative nonlinear backscatter from a single NB from a narrowband 30 cycle ultrasound pulse and the corresponding frequency spectrum (25 MHz transducer at 300kPa). Presence of subharmonic (f/2), ultraharmonic (3f/2), and second harmonic (2f), in addition to its fundamental (f) harmonic signal in the frequency spectra confirms the nonlinear oscillation of NBs. Minor changes in backscattered amplitude during the 30 cycle pulse shows how stably PG-Gly-PL NB oscillates when insonated by 25 MHz ultrasound. The magnitude of amplitude is consistent with the predicted stiffness of the NB shell having PG-PL as the most flexible, Gly-PL as the stiffest, and PG-Gly-PL with intermediate stiffness. (b) Contrast harmonic images (2nd harmonic) of PG-Gly-PL at different receiving frequency showing improvement in both intensity and resolution of the backscattered signal. NB with concentration of 4 × 109 NBs/mL was placed in a tissue-mimicking agarose phantom. Only the second harmonic backscattered signal is visible in contrast harmonic imaging mode which implies that the backscattered signal from the agarose phantom, similar to tissues, does not have non-linear component. Enhancement of nonlinear backscattered signal of NB (concentration of 4 × 109 NB/mL) relative to tissue-mimicking agarose phantom at different (c) receiving frequency and (d) peak negative pressure.

In vitro echogenicity of a solution of NBs (concentration of 4 × 109 NB/mL) relative to a tissue-mimicking agarose phantom (Figure S3a) was studied by analyzing the nonlinear backscatter at different ultrasound frequencies using a clinical scanner in contrast harmonic imaging mode (Figure S4). The enhancement, as well as the resolution, increased as the receiving frequency is changed from 7 MHz to 12 MHz (Figure 2b and 2c). As expected, the agarose phantom did not show any nonlinear acoustic activity, similar to tissues and organs, as shown by dark areas in Figure 2b.76 Consistent with what has been observed when single NB were insonated (Figure 2a), the flexible PG-PL bubbles showed the greatest enhancement while, the stiff Gly-PL showed a lower enhancement suggesting lower magnitude of oscillation as compared to PG-PL. PG-Gly-PL has intermediate enhancement because it has both a compliant and a stiff adlayer and thus intermediate enhancement. A similar trend is observed as the peak negative pressure of the transmitted ultrasound (Figure 2d and S5) is increased with PG-PL showing the greatest enhancement.

As stated above, one of the main challenges with current UCAs is rapid signal decay due to dissipation of the gas from the bubble core, which is immensely amplified when the UCAs are continuously oscillating due to ultrasound. To assess bubble in vitro stability under constant insonation, we placed the NB solution in a tissue-mimicking phantom with a narrow channel (Figure S3b) and exposed it to ultrasound at varying frame rates and pressures (Figure S6 and S7). A narrow channel was specifically used to ensure that the NBs were continuously insonated, in contrast to the typical stability test wherein the NB can move in and out of the ultrasound field (Figure S3b).22,77 In these experiments, PG-Gly-PL showed the slowest signal decay (Figure 3a, b) and the longest in vitro half-life for all frame rates and pressure (Figure 3c, d). The synergistic effect of having a bilayer with elastic contrast designed similar to natural membranes enabled the PG-Gly-PL to resist failure under continuous oscillation. We hypothesize that while PG introduces elastic and compliant defects in the shell that absorbs excess energy during compression as demonstrated with XRD in Figure 1d, Gly forms a vitrified adlayer that provides stiffness, limits strain during expansion, and functions as an additional layer resisting gas diffusion.41,78,79 Furthermore, PG-PL has a longer in vitro half-life than Gly-PL suggesting that having a compliant shell imparts better stability under ultrasound than having a stiff shell. For all cases, PL always showed the lowest enhancement and shortest in vitro half-life, likely brought about by its inability to endure oscillating deformations. The stabilizing effect of glycerol has been previously demonstrated by Yin, et al. where they showed that the NBs with glycerol under ultrasound have longer in vivo and in vitro half-life relative to microbubbles without glycerol.80 Interestingly, and similar to prior observations by our group, the equilibrium surface tension appeared to have a lesser influence on the bubble stability in this system as Gly-PL or PG-PL showed the lowest surface tension (but not the highest stability) when measured via pendant drop or rising bubble with perfluoropropane gas, respectively (Figure S8).26

Figure 3. In vitro characterization of NBs in tissue mimicking agarose phantom.

Figure 3.

(a) Contrast harmonic images of PG-Gly-PL (concentration of 4 × 109 NB/mL) before and after application of 500 frames of ultrasound at different peak negative pressure (P = 245 kPa to 465 kPa) at 1 frame per second. A dilute solution of PG-Gly-PL NB is placed in a tissue-mimicking agarose phantom with a thin channel (1 mm) to ensure minimal diffusion of NB in and out of the ultrasound field. (b) Representative signal decay curve of NBs continuously exposed to 245 kPa pressure at 1 frame per second. Mean ± SE (n = 3). (c) In vitro half-life of nonlinear backscattered signal of dilute NBs placed in a tissue mimicking agarose phantom and at 245 kPa pressure and varying frame rate. Mean ± SE (n = 3). (d) In vitro half-life (t1/2) of nonlinear backscattered signal dilute NBs placed in a tissue mimicking agarose phantom and exposed to ultrasound with varying peak negative pressure at 15 frames per second. Mean ± SE (n = 3).

Development of long-circulating NBs can have a significant impact in advancing the role of ultrasound in applications such as molecular imaging, and can open the door to more effective targeted theranostic or drug-delivery applications in cancer and a number of other diseases where nanomedicine already plays a critical role.80,81 Clinical practice with current UCAs is limited to circulation times of ~5 mins, which is the reported in vivo lifetime of Definity® and Sonovue®, as UCAs quickly dissipate and are consumed by macrophages.22 The in vivo lifetime of current UCAs, even with a smaller, sub-micron footprint, is thus insufficient to enable reliable extravasation and retention in tumors. To confirm if the observed in vitro stability of PG-Gly-PL NBs is translatable to in vivo, we injected our NB solution into the tail vein of a mouse and continuously acquired images of the kidney and liver for 30 mins with a clinical ultrasound in contrast harmonic imaging mode (Figure 4a, b). For comparison, we performed an identical experiment with a commercially-available and FDA-approved UCA (Lumason®). All samples show improvement in kidney contrast within 1 min after injection, but the signal from Lumason® dissipated within 5 min with in vivo half-life (t1/2) of 1.56 ± 0.27 min. The in vivo t1/2 for Gly-PL was determined to be 14.24 ± 1.08 min, PG-PL to be 12.19 ± 1.2 min, and PG-Gly-PL to be 18.32 ± 0.40 min. Despite having similar maximum non-linear image enhancement, PG-Gly-PL showed a 12x longer in vivo half-life than Lumason® (Figure 4c and d). The enhancement of PG-Gly-PL was similar to PG-PL and higher than Gly-PL, but PG-Gly-PL bubbles showed a longer in vivo half-life than both PG-PL and Gly-PL. This indicates that PG-Gly-PL is not only the most stable under continuous ultrasound exposure in vitro, but also at higher temperature, under shear and turbulence in physiological blood flow, and in the presence of macrophages in vivo. It is interesting to note that the circulation time of bubbles without glycerol was shorter than both bubbles formulated with glycerol. This could also be indicative of reduced RES clearance due to the presence of glycerol on the bubble surface, an effect previously explored using liposomes.82 Furthermore, while Lumason® followed the two-compartment pharmacokinetic model typical for UCAs (Figure 4b), the NBs followed a logistic behavior wherein the intensity remained at its maximum for some time before it started to decay (in vivo decay onset).83 The in vivo decay onset (Figure 4e) of PG-Gly-PL was significantly longer than Lumason® (20x longer), PG-PL, and Gly-PL. This implies that continuous acquisition of high quality images can be performed in preclinical animal models for over 15 min following only a single bolus injection, which is not possible with Lumason® as its signal immediately decay after reaching the peak.84,85 Similar behavior was also observed in the liver despite the presence of stellate macrophages known to efficiently eliminate foreign objects (Figure S9).

Figure 4. In vivo characterization of NBs in mouse kidney stability relative to commercially available UCA in mice.

Figure 4.

(a) NBs and commercially available UCA (Lumason) were injected via tail-vein and a cross-section of the kidney and liver was imaged at 12 MHz, 245 kPa pressure, and 0.2 frames per second. Left: Representative B-mode images of the kidney and liver before the injection of UCAs. 0 min-20 mins: Series of images showing the signal onset and signal decay of various UCAs at different time points. Minimal to no contrast can be observed before injection (t= 0) which is expected because backscatter signals from kidney and liver do not have any nonlinear properties. (b) Representative signal decay curve of NBs and Lumason® showing the delayed signal decay onset and longer in vivo half-life of PG-Gly-PL. (c) In vivo maximum enhancement, (d) in vivo washout half-life, and (e) in vivo decay onset of NBs and Lumason extracted from the enhancement vs time curve.

To further demonstrate the unique capabilities of the PG-Gly-PL for practical applications, we performed a similar imaging protocol in a mouse with flank colorectal tumor (Figure 5). For comparison, a parallel experiment was performed with Lumason®. In addition, as an internal control, we also monitored the contrast dynamics in the kidney for both agents. Consistent with data shown in Figure 4, Lumason® enhances both tumor and kidney a minute after injection, but rapidly dissipates within 5 minutes (Figure 5ac). In contrast, PG-Gly-PL enhances the contrast of the kidney immediately, with in vivo decay onset at 19.67 ± 2.56 min and in vivo half-life of 24.38 ± 3.39 min (Figure S10). It is also worth noting that although the in vivo max enhancement of Lumason® (16.94 ± 0.27 dB) in kidney is not significantly different than PG-Gly-PL (16.17 ± 0.86 dB), PG-Gly-PL (11.59 ± 1.11 dB) shows a 2-fold increase in in vivo max enhancement in tumor as compared to Lumason® (5.14 ± 1.10 dB) (Figure S11).

Figure 5. In vivo characterization of NBs in mouse kidney and flank colorectal tumor stability relative to commercially available UCA (Lumason®) in mice.

Figure 5.

(a) NBs or Lumason® were injected via tail-vein and a cross-section of the kidney and flank tumor was imaged at 12 MHz, 245 kPa pressure, and 0.2 frames per second. Left: Representative B-mode images of the kidney and tumor before the injection of UCAs. 0 min-20 mins: Series of images showing the signal onset and signal decay of various UCAs at different time points. Minimal to no contrast can be observed before injection (t= 0) which is expected because backscatter signals from kidney and tumor do not have any nonlinear properties. Representative signal decay curves of PG-Gly-PL and Lumason® in kidney (b) and in tumor (c) showing the delayed signal decay onset and longer in vivo half-life of PG-Gly-PL.

Although the use of clinical ultrasound equipment demonstrates the potential for clinical translation of ultrastable NB, the equipment is not ideal for imaging small mouse tumors. We therefore examined performance of PG-Gly-PL compared to MicroMarker® research microbubble (FujiFilm VisualSonics) optimized for use with a research ultrasound scanner (VisualSonics Vevo 3100). For this experiment, we injected PG-Gly-PL or Micromarker® into a mouse with colorectal tumor and captured 2D contrast images for 5 mins before performing a 3D scan of the tumor. For both experiments, the transducer positon was held constant for a more direct comparison. Injection of PG-Gly-PL led to significantly higher enhancement at 1.5 min that highlights the entire tumor volume (outer shell and inner core), in comparison to MicroMarker® that highlights mainly the periphery of the tumor. This difference could be brought about by the fact that tumors have smaller vessels at the center, and the concentration of the MicroMarker® there is not enough to be visualized by ultrasound.86 3D volume analysis of the presence of contrast signal in the entire tumor volume shows >4-fold increase in coverage with PG-Gly-PL nanobubbles. Expanding the use of ultrasound in these applications could make crucial diagnostic imaging tools significantly more accessible to a broader global population as compared to modalities such as magnetic resonance imaging (MRI) and computed tomography (CT). Colorectal cancer, for example, is the second leading cause of cancer-related deaths for both men and women in the United States,87,88 yet the use of ultrasound for colorectal cancer diagnosis has been essentially abandoned by current guidelines in favor of computed tomography (CT).89 However, there has been recent interest in assessing colorectal tumor angiogenesis by quantitative contrast-enhanced endoscopic ultrasound.90 New contrast agents could provide complementary data in this area and eventually lead to a more effective diagnosis.

Conclusions

We have shown that our ultrastable NB (PG-Gly-PL) with membrane structure similar to bacterial cell envelope has minimal signal decay when insonated continuously in vitro, has longer in vivo half-life, and has a delayed onset of in vivo signal decay. Despite having a bilayer structure with contrasting elastic property, the backscattered ultrasound from the ultrastable NB still has nonlinear properties when insonated as a single NB or as population. The enhanced in vivo stability of the ultrastable NB can advance the use of ultrasound in molecular imaging, theranostics, and drug delivery. Compared to Lumason® (FDA-approved UCA) and MicroMarker®, our ultrastable nanobubble is able to more effectively enhance the entire tumor vasculature for an extended duration, making it the ideal potential agent for enabling future molecular imaging applications of ultrasound.

Experimental Section

Preparation and Isolation of NB solution.

PG-GLy-PL solution (10 mg/mL) was prepared by first dissolving 1,2-dibehenoyl-sn-glycero-3-phosphocholine (C22, Avanti Polar Lipids Inc., Pelham, AL), 1,2 Dipalmitoyl-sn-Glycero-3-Phosphate (DPPA, Corden Pharma, Switzerland), 1,2-dipalmitoyl-sn-glycero-3-phosphoethanolamine (DPPE, Corden Pharma, Switzerland), and 1,2-distearoyl-snglycero-3-phosphoethanolamine-N-[methoxy(polyethylene glycol)-2000] (ammonium salt) (DSPE-mPEG 2000, Laysan Lipids, Arab, AL) into propylene glycol (PG, Sigma Aldrich, Milwaukee, WI) by heating and sonicating at 80 °C until all the lipids were dissolved. Mixture of glycerol (Gly, Acros Organics) and phosphate buffer solution (0.8 mL, Gibco, pH 7.4) preheated to 80 °C was added to the lipid solution. The resulting solution was sonicated for 10 mins at room temperature. Note that we typically prepared batches of 5 or 10 samples at a time and the amounts were adjusted as such. The solution (1 mL) was transferred to a 3 mL headspace vial, capped with a rubber septum and aluminum seal, and sealed with a vial crimper. Air was manually removed with a 30 mL syringe and was replaced by injecting octafluoropropane (C3F8, Electronic Fluorocarbons, LLC, PA) gas. PG-PL and Gly-PL solutions were prepared similar to PG-Gly-PL but with 1 mL of PG and 1 mL of Gly, respectively, instead of 0.5 mL of PG and 0.5 mL of Gly. PL solution was prepared by directly dissolving the phospholipid mixture into 1 mL PBS. After air was replaced by C3F8, the phospholipid solution was activated by mechanical shaking with a VialMix shaker (Bristol-Myers Squibb Medical Imaging Inc., N. Billerica, MA) for 45s. Nanobubbles were isolated from the mixture of foam and microbubbles by centrifugation at 50 rcf for 5 mins with the headspace vial inverted, and the 100 μL NB solution withdrawn from a fixed distance of 5 mm from the bottom with a 21G needle.

Cryo-EM of NBs.

Samples of the PG-Gly-PL NBs washed with PBS after isolation at concentrations around 1010 particles were applied to Quantifoil holey carbon EM grids (R2/2, 400 mesh:EMS). The EM grids were glow discharged for 30 sec at 15 mA before the sample was applied. 4 μL aliquots of the NB solution were applied to the EM grids and blotted for 6 sec. Following application, the grids were plunge frozen using a Vitrobot. The prepared cryoEM grids were imaged using a ThemoFisher/FEI TecnaiSpirit cryo-transmission electron microscope (operated at 80kV) equipped with a Gatan US4000 CCD camera. Micrographs were collected at 30kX magnification with a defocus range of −2.5 to −3 μm.

Size, concentration, and surface charge of NBs and non-buoyant particles.

The size distribution and concentration of NBs were characterized with a Resonant Mass Measurement (Archimedes®, Malvern Panalytical) equipped with a nanosensor capable of measuring particle size between 50 nm and 2000 nm. The nanosensor was pre-calibrated with 565 nm polystyrene beads (Thermo Scientific™ Nanospehere Size Standards 3560A). The NB solution was diluted with PBS (500x) to obtain an acceptable limit of detection (< 0.01 Hz) and coincidence (< 5%). Prior to any measurement, a 5-min PBS blank was run to ensure that the system fluidics and sensor were free of particles. Also, between measurements the sensor and microfluidic tubing were rinsed for 30 seconds with PBS followed by 2 “sneezes” for at least 3 cycles. During the sample measurement, NB solution was loaded for 120 seconds and analyzed at 2 and 5 psi, respectively. Samples measurement was finalized after 1000 particles were measured. Data was exported from the Archimedes software (version 1.2) and analyzed for positive and negative counts. Dilution was accounted for in calculating the NB concentration. Surface charge of the diluted NB solution (500X) was measure with an Anton Paar Litesizer™ 500.

Crystallinity analysis of NB membrane by x-ray diffractometer (XRD).

1D and 2D X-ray diffraction experiments were performed using a Bruker Discover D8 X-ray diffractometer, with a monochromated X-ray source (used with a Co Kα X-ray tube), configured in point focus mode. Isolated NB solution was transferred to a 1.5 mL Eppendorf microcentrifuge tube and was flashed frozen with liquid nitrogen before being loaded to a freeze-dryer (Labconco Freezone 2.5) set at – 50 °C. The NB solution was freeze-dried for at least 48 hours before placing on a zero diffraction plate sample holder for XRD.

Nonlinear oscillation of a single NB.

The backscatter signal from single NB events were captured through insonation of a dilute solution of NBs, with continuous pulse trains of 30 cycles at 25 MHz frequency for pressures value of 300 kPa. The ultrasonic signals were transmitted and received by a Vevo 770 ultrasound instrument (VisualSonics Inc. Toronto, Ontario). The transducer has a 20 MHz center frequency and 100% bandwidth. The dilute solution enabled the recording of single scattering events from individual NBs within the focus of the ultrasound transducer, as has been shown previously with cells and microspheres

Echogenicity of NB solution.

Nonlinear ultrasound imaging was done on AplioXG SSA-790A clinical ultrasound imaging system (Toshiba Medical Imaging Systems, Otawara-Shi, Japan) with a 12 MHz center frequency linear array transducer (PLT-1204BT). Images were acquired in contrast harmonic imaging (CHI) mode with parameters set as: 65 dB dynamic range, 70 dB gain, receiving frequency (7, 8, 10, and 12 MHz) and peak negative pressure (245, 318, 382, 465 kPa) depending on the experiment. NB solution was diluted (1mL, 100x diluted, concentration of 4 × 109 NBs/mL) to avoid signal attenuation and was placed in a tissue-mimicking agarose phantom. The agarose phantom was composed of 1.5 wt% agarose in Milli-Q water (resistivity of ~18 MΩ·cm) heated in a microwave until the agarose is dissolved. The hot agarose solution was then poured into a mold avoiding any trapped bubbles and cooled down to obtain phantom with the desired channel dimension (Figure S3a). Intensity of the backscattered nonlinear ultrasound signal was determined using a pre-loaded quantification software (CHI-Q) setting the ROI to be around inside the channel as shown in Figure 2b. The experiments were replicated three times. Enhancement was calculated by normalizing the measured backscattered ultrasound intensity of the NB solution with respect to the backscattered ultrasound intensity of the tissue-mimicking agarose phantom. The mean Enhancement was analyzed statistically with one-way analysis of variance (ANOVA), with the differences among the means evaluated by the Tukey–Kramer multiple comparison test. A p-value of < 0.05 was considered significant.

Stability of NB under continuous insonation.

NB solution was diluted (400 μL, 100x, concentration of 4 × 109 NBs/mL) with PBS and was placed in a tissue-mimicking agarose phantom with a thin channel (L × W × H = 22 × 1 × 10 mm) as shown in Figure S3b. Thin channel was chosen to ensure that the diffusion of NB in and out of the ultrasound field is minimized and NBs were continuously insonated. Nonlinear ultrasound imaging was done on AplioXG SSA-790A clinical ultrasound imaging system (Toshiba Medical Imaging Systems, Otawara-Shi, Japan) with a 12 MHz center frequency linear array transducer (PLT-1204BT). Images were acquired in contrast harmonic imaging (CHI) mode with parameters set as: 65 dB dynamic range, 70 dB gain, imaging frame rate (1, 5, 10, 15 fps) and peak negative pressure (245, 318, 382, 465 kPa) depending on the experiment. 500 frames of ultrasound images were acquired and the intensity per frame is analyzed with a built-in software (CHI-Q). Enhancement was calculated by normalizing the measured backscattered ultrasound intensity of the NB solution with respect to the the tissue-mimicking agarose phantom. In vitro half-life (t1/2) was calculated by using the calculated enhancement and with the formula for first order decay. To calculate for the half-life, we used the 1st order exponential decay model and linearized the data. The half-life is then calculated using t1/2 = (ln 2)/(-slope). The mean half-life was analyzed statistically with one-way analysis of variance (ANOVA), with the differences among the means evaluated by the Tukey–Kramer multiple comparison test. A p-value of < 0.05 was considered significant.

Surface tension measurement of NB solution.

Surface tension via pendant drop method was performed by dispensing a drop of the lipid solution at the tip of a microsyringe (SY20, Kruss Scientific, Germany) with a 0.5 mm outer diameter needle (Krüss, model: NE31). The needle used is made with stainless steel with PTFE coating and with stainless steel luer-lock connector. Images of the drops (n=3) was acquired with a KSV CAM 200 Optical Contact Angle Meter. Surface tension was calculated using the built-in software by analyzing the dimension of the drop and fitting into the Young-Laplace equation. Surface tension via rising drop method with air is done by dispensing an air bubble at the tip of a J-shaped needle (Krüss, model: NE71, 0.484 mm diameter) immersed in a lipid solution. Images of the bubbles (n = 3) was acquired and analyzed similar to the previous technique. Surface tension using rising bubble method with C3F8 was done similarly but replacing air with C3F8.

NB and Lumason® stability in vivo studies.

Animal experiments were performed in accordance with Public Health Service (PHS) Policy on Humane Care and Use of Laboratory Animals, the Guide for the Care and Use of Laboratory Animals: Eighth Edition (2011, The National Academies Press. https://doi.org/10.17226/12910), and in compliance with the Institutional Animal Care and Use Committee (IACUC) of Case Western Reserve University. In all procedures, the animals were anesthetized with 3% isoflurane with 1 L/min oxygen. Both NB solutions and Lumason (sulfur hexafluoride lipid-type A microspheres, Bracco Diagnostics Inc.) were tested in vivo. Lumason was prepared according to the protocol provided by the manufacturer. NB (200 μL of 4 × 1011 NBs/mL) or Lumason (1:4 dilution) was injected via the mouse tail vein (n=3). Nonlinear ultrasound imaging was done on the AplioXG SSA-790A clinical ultrasound imaging system with the same 12 MHz transducers above. The transducer was oriented along the sagittal plane to simultaneously image the kidney and liver of the mouse. Images were acquired in contrast harmonic imaging (CHI) mode with parameters set as: 65 dB dynamic range, 70 dB gain, 0.2 imaging frame rate per second, and peak negative pressure of 245 kPa (equivalent to mechanical index of 0.1). 500 frames (~30 mins) of ultrasound images were acquired and the intensity per frame is analyzed with a built-in software (CHI-Q). Enhancement was calculated by normalizing the measured intensity of the NB solution with respect to the intensity before the injection of NB with ROI shown in Figure 4a. In vivo half-life and in vivo decay onset were calculated by using calculated enhancement and with the formula for a 5-parameter logistic function using nonlinear curve fitting function of Origin. The 5-parameter logistic function is as follows y=Amin+AmaxAmin(1+(xx0)h)s.

NB and MicroMarker® stability in vivo studies.

Animal experiments were performed in accordance with Public Health Service (PHS) Policy on Humane Care and Use of Laboratory Animals, the Guide for the Care and Use of Laboratory Animals: Eighth Edition (2011, The National Academies Press. https://doi.org/10.17226/12910), and in compliance with the Institutional Animal Care and Use Committee (IACUC) of Case Western Reserve University. HCT-15 cells (0.1 mL) in sterile 1X PBS were subcutaneously injected into the dorsal side of BALB/c nude mice at a concentration of 5 ×106 cells/mL. Animals were observed every other day until tumors reached at least 8 mm in diameter. NB (200 μL) or Micromarker® (50 μL) was injected via the mouse tail vein. Nonlinear ultrasound imaging was done in VisualSonics Vevo 3100 using MX250 transducer set at 18 MHz, 4% power, 30 dB contrast gain, 28 dB gain, 40 dB dynamic range, 22 frame per second). The entire tumor was then scanned via a 3d motor with respiratory gating at 0.051 mm step size. The 2d slices were reconstructed and analyzed using VevoCQ. The percent agent in the tumor is calculated based on the voxels with signal from a contrast agent relative to the total number of voxels. The ROI of the volume analyzed is shown in blue-green color in the last panel in Figure 6.

Figure 6.

Figure 6.

In vivo characterization of PG-Gly-PL in flank colorectal tumor stability relative to MicroMarker® in mice showing comparison on the extent of UCA filling of tumor both in 2D and 3D. Tumor volume: 229.975 mm3.

Supplementary Material

ESI

Acknowledgments

This work was funded by the National Institutes of Health (1R01EB025741–01) and the Office of the Assistant Secretary of Defense for Health Affairs, through the Prostate Cancer Research Program under Award No. W81XWH-16-1-037I and W81XWH-16-1-0372. We also acknowledge additional support from the Case Comprehensive Cancer Center P30CA043703 in the form of a pilot grant and T32 GM008803 training grant support to CCE. Views and opinions of, and endorsements by the author(s) do not reflect those of the National Institutes of Health or of the Department of Defense.

Footnotes

Supporting Information

Supporting Information is available from the Nanoscale or from the author.

Conflict of Interest

The authors declare no conflict of interest.

Notes and References

  • 1.Szabo T, No TitleDiagnostic Ultrasound Imaging: Inside Out, Academic Press, Second Edi., 2013. [Google Scholar]
  • 2.Paproski RJ, Forbrich A, Huynh E, Chen J, Lewis JD, Zheng G and Zemp RJ, Small, 2016, 12, 371–380. [DOI] [PubMed] [Google Scholar]
  • 3.Wang X, Chen H, Zhang K, Ma M, Li F, Zeng D, Zheng S, Chen Y, Jiang L, Xu H and Shi J, Small, 2014, 10, 1403–1411. [DOI] [PubMed] [Google Scholar]
  • 4.Hadinger KP, Marshalek JP, Sheeran PS, Dayton PA and Matsunaga TO, Ultrasound Med. Biol, 2018, 44, 2728–2738. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Rojas JD and Dayton PA, Ultrasound Med. Biol, 2019, 45, 177–191. [DOI] [PubMed] [Google Scholar]
  • 6.Nyankima AG, Rojas JD, Cianciolo R, Johnson KA and Dayton PA, Ultrasound Med. Biol, 2018, 44, 368–376. [DOI] [PubMed] [Google Scholar]
  • 7.Zha Z, Wang S, Zhang S, Qu E, Ke H, Wang J and Dai Z, Nanoscale, 2013, 5, 3216. [DOI] [PubMed] [Google Scholar]
  • 8.Dong F, Zhang J, Wang K, Liu Z, Guo J and Zhang J, Nanoscale, 2019, 11, 1123–1130. [DOI] [PubMed] [Google Scholar]
  • 9.Xie Y, Wang J, Wang Z, Krug KA and Rinehart JD, Nanoscale, 2018, 10, 12813–12819. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kaufmann BA, Lindner JR, Wu JC and Narula J, Curr. Opin. Biotechnol, 2007, 18, 11–16. [DOI] [PubMed] [Google Scholar]
  • 11.Borden MA, Zhang H, Gillies RJ, Dayton PA and Ferrara KW, Biomaterials, 2008, 29, 597–606. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Goldberg BB, Liu JB and Forsberg F, Ultrasound Med. Biol, 1994, 20, 319–333. [DOI] [PubMed] [Google Scholar]
  • 13.Schutt EG, Pelura TJ and Hopkins RM, Acad. Radiol, 1996, 3 Suppl 2, S188–S190. [DOI] [PubMed] [Google Scholar]
  • 14.Yang F, Hu S, Zhang Y, Cai X, Huang Y, Wang F, Wen S, Teng G and Gu N, Adv. Mater, 2012, 24, 5205–5211. [DOI] [PubMed] [Google Scholar]
  • 15.Lin PL, Eckersley RJ and Hall EAH, Adv. Mater, 2009, 21, 3949–3952. [Google Scholar]
  • 16.Huang HY, Liu HL, Hsu PH, Chiang CS, Tsai CH, Chi HS, Chen SY and Chen YY, Adv. Mater, 2015, 27, 655–661. [DOI] [PubMed] [Google Scholar]
  • 17.Zhou Y, Wang Z, Chen Y, Shen H, Luo Z, Li A, Wang Q, Ran H, Li P, Song W, Yang Z, Chen H, Wang Z, Lu G and Zheng Y, Adv. Mater, 2013, 25, 4123–4130. [DOI] [PubMed] [Google Scholar]
  • 18.Lukianova-Hleb EY, Ren X, Zasadzinski JA, Wu X and Lapotko DO, Adv. Mater, 2012, 24, 3831–3837. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.McDannold N, Vykhodtseva N and Hynynen K, AIP Conf. Proc, 2007, 911, 547–553. [Google Scholar]
  • 20.Unger EC, Porter T, Culp W, Labell R, Matsunaga T and Zutshi R, Adv. Drug Deliv. Rev, 2004, 56, 1291–1314. [DOI] [PubMed] [Google Scholar]
  • 21.Paefgen V, Doleschel D and Kiessling F, Front. Pharmacol, 2015, 6, 1–16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Garg S, Thomas AA and Borden MA, Biomaterials, 2013, 34, 6862–6870. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Segers T, De Rond L, De Jong N, Borden M and Versluis M, Langmuir, 2016, 32, 3937–3944. [DOI] [PubMed] [Google Scholar]
  • 24.Katiyar A and Sarkar K, J. Colloid Interface Sci, 2010, 343, 42–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Mulvana H, Stride E, Tang MX, Hajnal JV and Eckersley RJ, Ultrasound Med. Biol, 2012, 38, 1097–1100. [DOI] [PubMed] [Google Scholar]
  • 26.Hernandez C, Nieves L, De Leon AC, Advincula R and Exner AA, ACS Appl. Mater. Interfaces, 2018, 10, 9949–9956. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Perera RH, Wu H, Peiris P, Hernandez C, Burke A, Zhang H and Exner AA, Nanomedicine Nanotechnology, Biol. Med, 2017, 13, 59–67. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Uhlendorf V, IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 1994, 41, 70–79. [DOI] [PubMed] [Google Scholar]
  • 29.Ferrara K, Pollard R and Borden M, Annu. Rev. Biomed. Eng, 2007, 9, 415–447. [DOI] [PubMed] [Google Scholar]
  • 30.Murai Y, Oishi Y, Takeda Y and Yamamoto F, Exp. Fluids, 2006, 41, 343–352. [Google Scholar]
  • 31.Kanai A and Miyata H, Int. J. Numer. Methods Fluids, 2001, 35, 593–615. [Google Scholar]
  • 32.Leguillon D and Martin E, Int. J. Fract, 2013, 179, 157–167. [Google Scholar]
  • 33.Leguillon D and Martin E, Int. J. Fract, 2013, 179, 169–178. [Google Scholar]
  • 34.Discher BM, Won Y, Ege DS, Lee JC, Bates FS, Discher DE and Hammer DA, Adv. Sci, 2012, 284, 1143–1146. [DOI] [PubMed] [Google Scholar]
  • 35.Mohandas N and Gallagher PG, Blood, 2009, 112, 3939–3948. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Huck WTS, Nat. Mater, 2005, 4, 271–272. [DOI] [PubMed] [Google Scholar]
  • 37.Maheshwari RGS, Tekade RK, Sharma PA, Darwhekar G, Tyagi A, Patel RP and Jain DK, Saudi Pharm. J, 2012, 20, 161–170. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.El Maghraby GM, Barry BW and Williams AC, Eur. J. Pharm. Sci, 2008, 34, 203–222. [DOI] [PubMed] [Google Scholar]
  • 39.Terakosolphan W, Trick JL, Royall PG, Rogers SE, Lamberti O, Lorenz CD, Forbes B and Harvey RD, Langmuir, 2018, 34, 6941–6954. [DOI] [PubMed] [Google Scholar]
  • 40.Biondi AC and Disalvo EA, Biochim. Biophys. Acta - Biomembr, 1990, 1028, 43–48. [DOI] [PubMed] [Google Scholar]
  • 41.Pocivavsek L, Gavrilov K, Cao KD, Chi EY, Li D, Lin B, Meron M, Majewski J and Lee KYC, Biophys. J, 2011, 101, 118–127. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Abenojar EC, Nittayacharn P, de Leon AC, Perera R, Wang Y, Bederman I and Exner AA, Langmuir, 2019, acs.langmuir.9b00462. [DOI] [PubMed] [Google Scholar]
  • 43.Hernandez C, Gulati S, Fioravanti G, Stewart PL and Exner AA, Sci. Rep, 2017, 7, 13517. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Lee J, Shen W, Payer K, Burg TP and Manalis SR, Nano Lett, 2010, 10, 2537–2542. [DOI] [PubMed] [Google Scholar]
  • 45.Godin M, Bryan AK, Burg TP, Babcock K and Manalis SR, Appl. Phys. Lett, 2007, 91, 1–4. [Google Scholar]
  • 46.Burg TP, Godin M, Knudsen SM, Shen W, Carlson G, Foster JS, Babcock K and Manalis SR, Nature, 2007, 446, 1066–1069. [DOI] [PubMed] [Google Scholar]
  • 47.Olcum S, Cermak N, Wasserman SC, Christine KS, Atsumi H, Payer KR, Shen W, Lee J, Belcher AM, Bhatia SN and Manalis SR, Proc. Natl. Acad. Sci, 2014, 111, 1310–1315. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Hernandez C, Abenojar EC, Hadley J, de Leon AC, Coyne R, Perera R, Gopalakrishnan R, Basilion JP, Kolios MC and Exner AA, Nanoscale, 2019, 11, 851–855. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Segers T, Lohse D, Versluis M and Frinking P, Langmuir, 2017, 33, 10329–10339. [DOI] [PubMed] [Google Scholar]
  • 50.TAKEUCHI Y, YASUKAWA H, YAMAOKA Y, KATO Y, MORIMOTO Y, FUKUMORI Y and FUKUDA T, Chem. Pharm. Bull. (Tokyo), 1992, 40, 1887–1892. [DOI] [PubMed] [Google Scholar]
  • 51.Elmoslemany RM, Abdallah OY, El-Khordagui LK and Khalafallah NM, AAPS PharmSciTech, 2012, 13, 723–731. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Birru WA, Warren DB, Han S, Benameur H, Porter CJH, Pouton CW and Chalmers DK, Mol. Pharm, 2017, 14, 580–592. [DOI] [PubMed] [Google Scholar]
  • 53.Harvey RD, Ara N, Heenan RK, Barlow DJ, Quinn PJ and Lawrence MJ, Mol. Pharm, 2013, 10, 4408–4417. [DOI] [PubMed] [Google Scholar]
  • 54.Boncheva M, Damien F and Normand V, Biochim. Biophys. Acta - Biomembr, 2008, 1778, 1344–1355. [DOI] [PubMed] [Google Scholar]
  • 55.Vanić Ž, Hurler J, Ferderber K, Golja Gašparović P, Škalko-Basnet N and Filipović-Grčić J, J. Liposome Res, 2014, 24, 27–36. [DOI] [PubMed] [Google Scholar]
  • 56.Elsayed MMA, Abdallah OY, Naggar VF and Khalafallah NM, J. Pharm. Pharmacol, 2007, 59, 1447–1450. [DOI] [PubMed] [Google Scholar]
  • 57.Željka Vanić J Drug Discov Dev. Deliv, 2014, 1, 8. [Google Scholar]
  • 58.Lee EH, Kim A, Oh YK and Kim CK, Biomaterials, 2005, 26, 205–210. [DOI] [PubMed] [Google Scholar]
  • 59.Alvarez-Román R, Naik A, Kalia YN, Guy RH and Fessi H, Pharm. Res, 2004, 21, 1818–1825. [DOI] [PubMed] [Google Scholar]
  • 60.Trotta M, Peira E, Carlotti ME and Gallarate M, Int. J. Pharm, 2004, 270, 119–125. [DOI] [PubMed] [Google Scholar]
  • 61.Elmoslemany RM, Abdallah OY, El-Khordagui LK and Khalafallah NM, AAPS PharmSciTech, 2012, 13, 723–731. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Sato K and Ueno S, Curr. Opin. Colloid Interface Sci, 2011, 16, 384–390. [Google Scholar]
  • 63.Hoche S, Hussein MA and Becker T, Ultrasonics, 2015, 57, 65–71. [DOI] [PubMed] [Google Scholar]
  • 64.Zell K, Sperl JI, Vogel MW, Niessner R and Haisch C, Phys. Med. Biol, , DOI: 10.1088/0031-9155/52/20/N02. [DOI] [PubMed] [Google Scholar]
  • 65.Persson HW and Hertz CH, Ultrasonics, 1985, 23, 83–89. [DOI] [PubMed] [Google Scholar]
  • 66.Falou O, Jafari Sojahrood A, Carl Kumaradas J and Kolios MC, J. Acoust. Soc. Am, 2012, 132, 1820–1829. [DOI] [PubMed] [Google Scholar]
  • 67.Falou O, Church C, Baddour RE, Nathanael G, Church C, Czarnota GJ, Church C, Kumaradas JC, Kolios MC and Church C, J. Acoust. Soc. Am, 2008, 124, EL278–EL283. [DOI] [PubMed] [Google Scholar]
  • 68.Baddour RE, Sherar MD, Hunt JW, Czarnota GJ and Kolios MC, J. Acoust. Soc. Am, 2005, 117, 934–943. [DOI] [PubMed] [Google Scholar]
  • 69.van Rooij T, Luan Y, Renaud G, van der Steen AFW, Versluis M, de Jong N and Kooiman K, Ultrasound Med. Biol, 2015, 41, 1432–1445. [DOI] [PubMed] [Google Scholar]
  • 70.Marmottant P, van der Meer S, Emmer M, Versluis M, de Jong N, Hilgenfeldt S and Lohse D, J. Acoust. Soc. Am, 2005, 118, 3499–3505. [Google Scholar]
  • 71.Lum JS, Dove JD, Murray TW and Borden MA, Langmuir, 2016, 32, 9410–9417. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72.Alonso-Romanowski S, Biondi AC and Disalvo EA, J. Membr. Biol, 1989, 108, 1–11. [Google Scholar]
  • 73.CROWE JH, CROWE LM and CHAPMAN D, Science (80-. ), 1984, 223, 701–703. [DOI] [PubMed] [Google Scholar]
  • 74.Boggs JM and Rangaraj G, Biochim. Biophys. Acta - Biomembr, 1985, 816, 221–233. [DOI] [PubMed] [Google Scholar]
  • 75.Abou-Saleh RH, McLaughlan JR, Bushby RJ, Johnson BR, Freear S, Evans SD and Thomson NH, Langmuir, 2019, acs.langmuir.8b04130. [DOI] [PubMed] [Google Scholar]
  • 76.Huynh E, Leung BYC, Helfield BL, Shakiba M, Gandier JA, Jin CS, Master ER, Wilson BC, Goertz DE and Zheng G, Nat. Nanotechnol, 2015, 10, 325–332. [DOI] [PubMed] [Google Scholar]
  • 77.Sboros V, Adv. Drug Deliv. Rev, 2008, 60, 1117–1136. [DOI] [PubMed] [Google Scholar]
  • 78.López-Montero I, Arriaga LR, Monroy F, Rivas G, Tarazona P and Vélez M, Langmuir, 2008, 24, 4065–4076. [DOI] [PubMed] [Google Scholar]
  • 79.Wegst UGK, Bai H, Saiz E, Tomsia AP and Ritchie RO, Nat. Mater, 2015, 14, 23–36. [DOI] [PubMed] [Google Scholar]
  • 80.Yin T, Wang P, Zheng R, Zheng B, Cheng D, Zhang X and Shuai X, Int. J. Nanomedicine, 2012, 7, 895–904. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 81.Gao Y, Hernandez C, Yuan HX, Lilly J, Kota P, Zhou H, Wu H and Exner AA, Nanomedicine Nanotechnology, Biol. Med, 2017, 13, 2159–2168. [DOI] [PubMed] [Google Scholar]
  • 82.MARUYAMA K, OKUIZUMI S, ISHIDA O, YAMAUCHI H, KIKUCHI H and IWATSURU M, Int. J. Pharm, 1994, 111, 103–107. [Google Scholar]
  • 83.Chen CC, Sirsi SR, Homma S and Borden MA, Ultrasound Med. Biol, 2012, 38, 492–503. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Perera R, DeLeon A, Wang X, Ramamurtri G, Peiris P, Basilion J and Exner AA, in 2018 IEEE International Ultrasonics Symposium (IUS), IEEE, 2018, pp. 1–4. [Google Scholar]
  • 85.Perera R, bioRxiv, , DOI: 10.1101/634444. [DOI] [Google Scholar]
  • 86.Wilhelm S, Tavares AJ, Dai Q, Ohta S, Audet J, Dvorak HF and Chan WCW, Nat. Rev. Mater, , DOI: 10.1038/natrevmats.2016.14. [DOI] [Google Scholar]
  • 87.Zhang Z, Xu J, Liu B, Chen F, Li J, Liu Y, Zhu J and Shen C, Inflammopharmacology, , DOI: 10.1007/s10787-018-0534-5. [DOI] [Google Scholar]
  • 88.Sartore-Bianchi A, Marsoni S and Siena S, JAMA Oncol, 2017, 4, 2017–2018. [DOI] [PubMed] [Google Scholar]
  • 89.Nasseri Y and Langenfeld SJ, Surg. Clin. North Am, 2017, 97, 503–513. [DOI] [PubMed] [Google Scholar]
  • 90.Cartana E-T, Gheonea D, Cherciu I, Streaţa I, Uscatu C-D, Nicoli E-R, Ioana M, Pirici D, Georgescu C-V, Alexandru D-O, Şurlin V, Gruionu G and Săftoiu A, Endosc. Ultrasound, 2018, 7, 175. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ESI

RESOURCES