Skip to main content
Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry logoLink to Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry
. 1973 Jul-Aug;77A(4):395–405. doi: 10.6028/jres.077A.028

Heat Capacities of Polyethylene from 2 to 360 K. I. Standard Samples of Linear and Branched Polyethylene Whole Polymer

S S Chang 1, A B Bestul 1,*
PMCID: PMC6728472  PMID: 32189750

Abstract

Heat capacities of two well characterized polyethylene samples have been measured from 2 to 360 K in a precision vacuum adiabatic calorimeter. The two samples are derived from the same stocks from which NBS standard reference materials (SRM) 1475 and 1476 for linear and branched polyethylene whole polymers, respectively, were established. Both samples have been studied in the conditions as received. The branched polyethylene sample has also been studied following various thermal treatments in the calorimeter. The effect of thermal history on the behavior of branched polyethylene has also been studied by differential scanning calorimetry.

Keywords: Branched polyethylene, calorimetry, glass transition, heat capacity, linear polyethylene, polyethylene, thermal analysis, thermodynamic properties

1. Introduction

Polyethylene is the simplest hydrocarbon chain polymer composed of methylene groups. This basic polymer has been the most widely used and studied polymer. There are no less than twenty papers [121]1 concerning the experimental heat capacity behavior of polyethylene of various origins and treatments. However most of the research papers present their results mainly or entirely in graphs [1, 2, 5, 6, 7, 10, 11, 13, 16, 18], in simple (linear or quadratic) analytical equations [1, 8, 12, 21] or in tables of smoothed values [3, 4, 9, 14, 15, 19, 20]. Hence even if the data is of high precision some fine features will be lost in the smoothing procedures and representations. Only one paper [17], concerning heat capacities of polyethylene samples below 30 K, listed the actual heat capacity data. Partial listing of the heat capacity data above 320 K appeared in another paper [21].

The present paper reports detailed investigations by precision adiabatic calorimetry from 2 to 360 K on two well characterized polyethylene samples in their as-received conditions. Over most of the temperature range investigated, the precision of the calorimetric measurement is better than 0.1 percent. These two samples are available as the National Bureau of Standards (NBS) standard reference materials (SRM) 1475 and 1476 for linear and branched polyethylene whole polymer, respectively. They are intended primarily as characterization standards for molecular weights and rheological properties.

Subsequent papers in this series will deal with the heat capacity behavior of densified linear polyethylene samples, the detection of the glass transformation in partially crystalline polyethylene, the deduction of heat capacity of 100 percent crystalline linear polyethylene, and a comparison of the heat capacity behavior of polyethylene from various origins and investigations.

2. Experimental Detail

2.1. Calorimetric Technique

Heat capacity measurements on the two polyethylene samples were performed in the vacuum adiabatic calorimeter described previously [22] with major modifications noted elsewhere [23]. The measuring procedures and methods of data treatment were discussed in more detail in another paper [24].

A calibrated platinum resistance thermometer was used to interpolate the temperature according to the International Practical Temperature Scale of 1968 [25, 26] above 13.81 K. Below 13.81 K the platinum thermometer was compared against a germanium resistance thermometer which has been calibrated in accordance with the NBS 1965 (2–20 K) provisional scale [27]. From this comparison, a fifth-degree polynomial [28] was generated to interpolate the temperature of the platinum thermometer below 13.81 K.

2.2. Material

Both calorimetric samples of the linear and of the branched polyethylene were taken from the stocks from which NBS standard reference materials (SRM) 1475 and 1476, respectively, were established.

a. Linear Polyethylene

The Certificate for SRM 1475, Linear Polyethylene (Whole Polymer) gives the following information. The weight-averaged molecular weight, Mw, is 52,000 as determined by light-scattering in 1-chloronaphthalene at 135 °C. Gel permeation chromatography gave a value of Mw as 53,070 with the ratio Mz:Mw:Mn = 7.54:2.90:1. The methyl group content is 0.15 methyl groups per 100 carbon atoms. The limiting viscosity numbers (dl/g) are 0.890, 1.010, and 1.180 in 1-chloronaphthalene, 1,2,4-trichlorobenzene and decalin, respectively, at 130 °C with a pellet to pellet coefficient of variation of 3 percent. This sample has an ash content of 0.002 percent. No volatiles were detected by a gas-chromatographic procedure capable of detecting 0.5 percent volatiles. The manufacturer of this sample added to the polymer 111 ppm of the anti-oxidant, tetrakis[methylene-3-(3′,5′-di-t-butyl-4′-hydroxy-phenyl)propionate] methane. The density reported in the certificate is 0.97844 g cm−3 at 23 °C, as determined by ASTM Method D1505–67 on a sample prepared by Procedure A, ASTM Method D1928–68. This density is obtained after a pretreatment of the sample and therefore differs from the density of the sample as received. Detailed reports on the investigations required for the certificate are described in a collection of papers [29].

The calorimetric sample for the linear polyethylene (SRM 1475) in the as received condition was composed of ellipsoidal pellets about 2–3 mm in diameter and about 3–5 mm in length. No treatment was applied to the sample before the heat capacity measurement. The density of the sample before the calorimetric measurement was 0.954 g cm−3 at 23 °C with a variation of ±0.001 g cm−3 as determined by a flotation method in an ethanol-water mixture. After the heat capacity measurement, the density of the calorimetric sample was determined again. No change, greater than the initial variation, was found in the density from that of the sample before the calorimetric measurement. The variation in the density of the sample after the calorimetric measurement is however reduced to about 0.0005 g cm−3. The density measurements were performed on no less than ten randomly chosen pellets from each of the samples.

The mass of the calorimetric sample was 78.572 g in vacuo, which corresponds to 5.6015 base mole or gram formula weight of methylene group, − CH2 −. The density values of 0.954 and 0.00117 g cm−3 for the sample and for the air, respectively, were used to estimate the buoyancy correction. The top of the sample container was soldered on with In-Sn solder while the container was surrounded with a water jacket, so that the sample was not heated above the ambient temperature during the soldering process. Helium gas at a pressure of 10 cm Hg was sealed in to facilitate the thermal conduction within the sample container.

b. Branched Polyethylene

The certificate for SRM 1476, Branched Polyethylene (Whole Polymer) gives the following information. The limiting viscosity numbers (dl/g) are 0.8132, 0.9024, and 1.042 in 1-chloronaphthalene, 1,2,4-trichlorobenzene and decalin, respectively, at 130 °C. No pellet to pellet variation in limiting viscosity number was found. The melt index is 1.19 g/10 min. The density is 0.9312 g cm−3 at 23 °C as determnied by previously mentioned ASTM procedures. The manufacturer of this sample has added to the polymer 50 ppm of the antioxidant, 4,4′-thio-bis(6-t-butyl-3-cresol).

The calorimetric sample for the branched polyethylene was taken from the same stock from which SRM 1476 was established. The material was in the shape of cylindrical pellets, about 3 mm in diameter and 2–3 mm in length. The initial density of the sample before the calorimetric measurement was 0.9247 g cm−3, at 23 °C with a variation less than 0.001 g cm−3 by flotation method. The final density of the sample after the heat capacity measurement was 0.9272 g cm−3 with a variation of less than 0.0005 g cm−3.

The mass of the calorimetric sample was 66.258 g in vacuo which corresponds to 4.7236 base mole or gram formula weight of methylene group, − CH2 −. The density values of 0.925 and 0.00117 g cm−3 for the sample and for air, respectively, were used to estimate the buoyancy correction. Precautions were taken so as not to alter the condition of the sample from that as received. The top of the container was soldered on with In-Sn solder while the container was surrounded with a water jacket to prevent heating of the sample. Helium gas at a pressure of 10 cm Hg was sealed in to aid the thermal conduction within the container.

3. Results

The results of the heat capacity measurements are tabulated in table 1 and shown graphically in figure 1. The table is arranged in the order of increasing initial temperature of a series of heat capacity determinations. The series are numbered in chronological sequence in order to facilitate the tracing of thermal history of the sample. The temperature increment for a heat capacity determination may be inferred from the differences in the mean temperatures of the adjacent determinations within the series. Curvature corrections have been added to correct for the effect of the finite temperature rise of a determination [30].

Table 1.

Heat capacities of polyethylene (base mole [− CH2 − ] = 14.027)

T,K Cp, J/K mol
I. Linear Polyethylene (SRM 1475 As Received)
SERIES VI
2.32 .0020
3.12 .0055
4.20 .0143
5.29 .0291
6.33 .0492
7.40 .0770
8.53 .1136
9.69 .1599
10.86 .2141
12.02 .2767
13.18 .3479
14.41 .4285
15.81 .5349
17.37 .6630
19.05 .8139
20.83 .9866
22.75 1.186
24.96 1.430
27.47 1.722
SERIES VII
4.68 .0200
5.56 .0335
6.35 .0497
7.18 .0709
8.20 .1017
9.17 .1380
10.16 .1804
11.21 .2321
12.30 .2931
13.40 .3621
14.58 .4410
15.95 .5458
17.48 .6722
19.21 .8294
21.18 1.019
23.33 1.247
25.73 1.516
28.38 1.828
31.21 2.178
34.29 2.575
37.73 3.024
41.45 3.513
45.55 4.049
50.06 4.630
SERIES V
47.15 4.259
51.16 4.774
55.89 5.359
61.44 6.008
67.41 6.662
SERIES II
73.12 7.252
79.78 7.890
87.35 8.563
96.19 9.286
105.82 10.01
115.56 10.70
125.36 11.38
135.25 12.08
145.15 12.81
SERIES III
147.08 12.97
157.14 13.71
167.25 14.42
177.34 15.12
187.31 15.80
197.20 16.47
207.01 17.14
SERIES I
212.27 17.52
222.16 18.26
231.95 19.08
241.73 19.99
251.40 20.89
260.88 21.83
SERIES VIII
253.16 21.07
SERIES IX
262.66 22.01
272.24 23.03
SERIES IV
265.49 22.33
SERIES X
282.30 24.07
SERIES XI
291.59 24.99
301.15 25.98
310.78 27.03
SERIES XII
320.21 28.15
329.93 29.47
SERIES XIII
339.34 30.73
348.93 32.43
357.00 33.97
II. Branched Polyethyelene (SRM 1476)
II.a As Received
SERIES VII
2.29 .0031
3.06 .0074
3.95 .0172
4.80 .0315
5.72 .0532
6.64 .0820
7.56 .1160
8.47 .1568
9.37 .2024
10.31 .2558
11.36 .3220
12.56 .4064
13.91 .5080
15.38 .6318
17.04 .7822
18.90 .9639
20.91 1.173
23.03 1.406
25.24 1.661
27.57 1.939
30.13 2.253
32.98 2.614
36.25 3.033
40.01 3.516
44.30 4.060
49.08 4.659
54.31 5.294
60.11 5.961
66.73 6.672
SERIES III
74.66 7.469
81.87 8.144
90.03 8.855
99.41 9.616
109.38 10.38
SERIES IV
114.88 10.79
124.43 11.49
134.11 12.19
143.95 12.92
153.84 13.67
163.81 14.43
SERIES V
172.38 15.10
182.13 15.86
192.01 16.65
201.89 17.47
211.68 18.35
SERIES VI
211.68 18.38
221.73 19.44
231.54 20.52
SERIES II
233.46 20.68
243.62 22.12
253.36 23.63
262.87 25.10
SERIES VIII
265.72 25.53
275.32 26.92
SERIES IX
284.78 28.29
294.45 29.48
304.10 30.97
313.59 33.30
SERIES I
300.30 30.07
309.79 32.50
SERIES X
317.85 33.79
328.02 36.71
337.93 39.08
347.52 43.35
355.68 47.76
II.b. Stabilized at 360 K and then Quenched
SERIES XI
25.46 1.680
28.04 1.990
30.93 2.345
34.17 2.758
37.74 3.217
42.03 3.765
46.94 4.395
52.16 5.038
57.77 5.690
63.77 6.356
70.55 7.059
78.18 7.797
SERIES XIII
87.07 8.607
96.56 9.385
106.47 10.16
116.25 10.89
SERIES XVI
154.85 13.75
164.67 14.51
174.37 15.26
183.97 16.03
193.58 16.80
SERIES XVII
203.28 17.67
213.11 18.54
223.06 19.51
232.98 20.58
242.74 21.92
252.30 23.35
261.76 24.78
271.05 26.14
280.37 27.36
289.70 29.06
SERIES XIV
206.63 17.92
216.56 18.86
226.49 19.84
233.45 20.60
237.44 21.13
241.41 21.68
245.38 22.30
249.32 22.85
253.24 23.47
260.14 24.52
SERIES XVIII
298.63 30.66
308.09 32.16
317.88 33.97
327.73 36.18
337.54 37.90
346.43 38.71
352.88 40.09
357.72 42.51
SERIES XII
304.59 31.66
314.17 33.36
323.86 35.53
333.48 37.35
343.21 38.48
SERIES XV
316.39 33.91
326.30 35.92
336.17 37.66
346.02 38.63
355.35 41.43
II.c. Annealed at 230 K
SERIES XXI
2.17 .0024
2.79 .0057
3.65 .0129
4.66 .0282
5.66 .0505
6.63 .0794
7.57 .1140
8.49 .1545
9.45 .2025
10.53 .2637
11.72 .3394
12.98 .4308
14.37 .5363
15.91 .6685
17.54 .8194
19.30 .9929
21.24 1.195
23.38 1.434
25.81 1.716
28.59 2.048
31.68 2.434
34.55 2.802
37.31 3.157
40.63 3.584
44.68 4.100
49.43 4.695
54.78 5.344
60.68 6.021
SERIES XXII
65.55 6.551
71.45 7.155
78.60 7.843
87.31 8.619
96.91 9.411
SERIES XX
121.97 11.29
131.77 12.01
141.71 12.74.
151.71 13.47
161.77 14.21
SERIES XXIII
142.99 12.82
152.83 13.54
162.63 14.28
172.41 15.03
182.19 15.78
192.09 16.55
SERIES XXIV
201.74 17.34
211.65 18.27
221.54 19.22
231.40 20.34
241.17 21.63
SERIES XIX
214.04 18.42
223.87 19.45
233.54 20.58
243.06 22.26
252.62 23.41
262.23 24.81
SERIES XXV
250.22 22.99
260.04 24.62
269.86 26.06
279.72 27.27
289.55 28.48
299.26 30.31
SERIES XXVI
307.21 32.21
317.04 34.18
326.92 36.12
336.76 37.87
346.58 38.75
355.85 41.33
II.d. Slow Cooled at 1 K/h
SERIES XXVII
196.16 16.85
206.05 17.69
215.95 18.56
225.81 19.61
235.62 20.78
245.32 22.19
255.02 23.66
264.77 25.19
SERIES XXVIII
273.98 26.49
283.87 27.98
293.74 29.57
303.60 31.21
313.40 33.11
323.21 35.16
SERIES XXIX
331.81 36.78
341.68 38.26
350.47 39.66
357.08 41.70

Figure 1. Heat capacity of polyethylene.

Figure 1.

Linear, (SRM 1475 as received): ●. Branched, (SRM 1476): ◯ as received, △ stabilized at 360 K and then quenched, □annealed at 230 K, ▼ slow cooled.

The precision of the measurement above 25 K is in the order of 0.05 percent. Below 25 K, the precision gradually changes to 1 percent at 5 K and about 5 percent at 2 K. The accuracy over most of the temperature range of the measurement is believed to be comparable to the precision as seen from the result of the heat capacity measurement on a Calorimetry Conference standard sample of sapphire [22].

Analytically smoothed heat capacity values at rounded temperatures along with values of other derived thermodynamic functions are listed in table 2. H0 and S0 refers to the zero point enthalpies and entropies of the individual samples. Since these samples are expected to have undetermined residual entropies at 0 K, Gibbs free energies for these samples are not given in table 2.

Table 2.

Thermodynamic functions of polyethylene (units in J, K and base mole [− CH2 − ] = 14.027)

Linear polyethylene (SRM 1475) As Received
T Cp H-H0 S-S0
5 0.024 0.029 0.0076
10 .173 .46 .062
15 .473 2.01 .184
20 .904 5.41 .376
25 1.433 11.22 .633
30 2.027 19.85 .946
35 2.664 31.57 1.306
40 3.322 46.53 1.705
45 3.981 64.79 2.134
50 4.626 86.31 2.587
60 5.841 138.7 3.540
70 6.935 202.7 4.524
80 7.911 277.1 5.515
90 8.786 360.6 6.498
100 9.579 452.5 7.465
110 10.31 552.0 8.413
120 11.01 658.6 9.341
130 11.70 772.2 10.25
140 12.44 892.8 11.14
150 13.18 1021 12.03
160 13.91 1156 12.90
170 14.61 1299 13.77
180 15.30 1449 14.62
190 15.98 1605 15.47
200 16.66 1768 16.30
210 17.36 1938 17.13
220 18.10 2116 17.96
230 18.91 2301 18.78
240 19.80 2494 19.60
250 20.76 2697 20.43
260 21.76 2909 21.26
270 22.78 3132 22.10
280 23.80 3365 22.95
290 24.83 3608 23.80
300 25.87 3862 24.66
310 26.95 4126 25.53
320 28.11 4401 26.40
330 29.39 4688 27.29
340 30.86 4989 28.19
350 32.59 5306 29.10
360 34.65 5642 30.05
273.15 23.10 3204 22.37
298.15 25.68 3814 24.50
Branched Polyethylene (SRM 1476) As Received
T Cp H-H0 S-S0
5 0.036 0.043 0.011
10 .238 .651 .088
15 .599 2.686 .248
20 1.077 6.835 .483
25 1.631 13.58 .782
30 2.233 23.22 1.132
35 2.871 35.97 1.524
40 3.511 51.93 1.949
45 4.151 71.09 2.400
50 4.773 93.40 2.869
60 5.945 147.1 3.845
70 7.008 211.9 4.842
80 7.971 286.9 5.842
90 8.849 371.1 6.832
100 9.662 463.7 7.807
110 10.43 564.2 8.764
120 11.16 672.1 9.703
130 11.89 787.4 10.63
140 12.63 910.0 11.53
150 13.38 1040 12.43
160 14.14 1178 13.32
170 14.91 1323 14.20
180 15.69 1476 15.07
190 16.49 1637 15.94
200 17.32 1806 16.81
210 18.22 1983 17.68
220 19.21 2170 18.55
230 20.32 2368 19.42
240 21.66 2577 20.32
250 23.14 2801 21.23
260 24.63 3040 22.17
270 26.08 3294 23.12
280 27.51 3562 24.10
290 28.96 3844 25.09
300 30.50 4141 26.09
310 32.23 4455 27.12
320 34.31 4787 28.18
330 36.92 5143 29.27
340 40.27 5528 30.42
350 44.62 5952 31.65
360 50.25 6425 32.98
273.15 26.53 3377 23.48
298.15 30.20 4085 25.91
Branched Polyethylene (SRM 1476) Annealed
T Cp H-H0 S-S0
5 0.035 0.043 0.01l
10 .233 .637 .086
15 .590 2.637 .244
20 1.065 6.732 .476
25 1.620 13.42 .772
30 2.223 23.01 1.120
35 2.859 35.71 1.510
40 3.503 51.61 1.934
45 4.142 70.73 2.384
50 4.766 93.01 2.853
60 5.943 146.6 3.827
70 7.012 2ll.5 4.825
80 7.974 286.5 5.825
90 8.846 370.7 6.816
100 9.652 463.2 7.790
110 10.41 563.6 8.746
120 11.15 671.4 9.684
130 11.88 786.6 10.61
140 12.60 909.0 11.51
150 13.34 1039 12.41
160 14.09 1176 13.29
170 14.84 1320 14.17
180 15.60 1473 15.04
190 16.38 1632 15.90
200 17.19 1800 16.76
210 18.06 1976 17.62
220 19.03 2162 18.48
230 20.13 2357 19.35
240 21.27 2564 20.23
250 22.90 2785 21.13
260 24.50 3022 22.05
270 25.92 3274 23.00
280 27.30 3541 23.96
290 28.84 3821 24.95
300 30.60 4119 25.95
310 32.59 4434 26.98
320 34.55 4770 28.04
330 36.33 5125 29.14
340 37.95 5497 30.24
350 39.74 5883 31.37
360 42.82 6295 32.52
273.15 26.35 3359 23.28
298.15 30.26 4064 25.74

During the measurements from Series VIII to XIII for the linear polyethylene sample, relatively long negative temperature drifts were observed. These long drifts were probably due mainly to the slow responding adiabatic shield rather than due to thermal effects generated by the sample. Upon unloading the cryostat, it was found that the heater wires for the top and bottom of the adiabatic shield had partially been peeled off from the shield. Hence the shield was responding sluggishly to the controller. Since the density of the sample after the heat capacity measurement did not show significant change from that of the original sample, it may be concluded that little or no change in crystallinity had occurred during the course of heat capacity measurement up to 360 K.

The low temperature heat capacity of the linear polyethylene sample, SRM 1475 as received, agrees within 1 percent of that reported for Marlex 2,2 one of the two linear polyethylene samples studied by Tucker and Reese [17]. This agreement may be expected, since the densities and hence the crystallinities of these two samples are very close to each other. The densities at room temperatures are 0.954 and 0.958 g cm−3 for SRM 1475 as received and Marlex 2, respectively. Below 30 K the heat capacity behavior of the branched polyethylene sample, SRM 1476, is similar to that of the Low Density (L.D.) polyethylene sample also reported in reference [17]. The heat capacity of SRM 1476 is in general about 1–2 percent lower than that of L.D. This is in accord with the density differences of the two samples. The room temperature density of SRM 1476 at 0.925–0.927 g cm−3 is slightly higher than that of L.D. at 0.915. The lowering of the heat capacity due to the increase in density or due to annealing can also be detected when the results for SRM 1476 in the as received condition (ρ = 0.925) is compared with that in the annealed condition (ρ = 0.927). From 20 to 360 K, heat capacities of linear polyethylene samples from several previous works [3, 9, 21, 31] are within 5 percent of the values for the linear polyethylene sample reported here, irrespective of their origins and densities. Even the values for various branched polyethylene [3, 9, 14, 15, 31] do not differ more than 5 percent from the value of the linear polyethylene sample of this work in the temperature range from 40 to 230 K.

Figures 2a and 2b show the heat capacity differences between the branched polyethylene sample subjected to various thermal treatments and the linear polyethylene sample in the condition as received. These two figures also show approximately the degree of precision of the measurements reported here. A low temperature maximum in the heat capacity difference may be seen in figure 2a centering around 30 K. Similar features may also be observed if low temperature heat capacities of branched polyethylene from other sources [14, 17, 31] are compared against that of linear polyethylene. Maxima in heat capacity differences between the glass and crystal of the same substance have been observed to occur in the temperature range around 15 to 40 K in many substances. Such a feature has been attributed mainly to the volumetric differences between the two forms [32] and has also been correlated with Schottky functions [31].

Figure 2. Comparison of heat capacity of polyethylene.

Figure 2.

Baseline: smoothed values for linear polyethylene, SRM 1475 as received. All legends same as in figure 1. (a) Below 200 K. (b) Above 160 K.

A small irregularity in the order of 1 percent of the heat capacity at about 140 K is probably caused by the heat capacity behavior of the linear polyethylene sample in that region. In the temperature region of 230 to 260 K the heat capacity difference seems to increase more steeply than that in the regions immediately above or below. The abrupt increase in the heat capacity of branched polyethylene in the temperature region around 240 K is more pronounced in a calorimetric study [31] on a sample having lower density, 0.91 g cm−3, than that of SRM 1476, and in thin film calorimetric results [6] on an even lower density, 0.89 g cm−3, sample. The magnitude of this heat capacity irregularity, characteristic of a glass transition, increases as the density of the sample is decreased or as the amorphous content of the sample is increased. Hence the glass transition temperature of branched polyethylene is located around 235 to 240 K.

Below 250 K the heat capacity of the annealed or slow-cooled branched polyethylene samples is lower than that of the as received or the quenched sample. Above 250 K, however, the apparent heat capacity which may include any crystallization or premelting processes seems to be partially a function of where the sample had been held previously at some higher temperature. When the sample is stored at temperatures above 250 K, the heat capacity is usually lowered in the vicinity of the storage temperature.

In order to assess the effect of the thermal history on the heat capacity of branched polyethylene and on the spontaneous temperature drift behavior (fig. 3) the following procedure was carried out. During the loading of the sample, a water-cooled jacket was used to surround the sample container, so that the sample would not be heated above room temperature during the process of soldering the container top. The sample was cooled in the cryostat by radiation and conduction at a rate of about 5 K h−1. The heat capacity of the as-received sample was measured below room temperature before the sample was heated above room temperature. No large spontaneous temperature drift was observed until the sample was heated above 330 K. Positive temperature drifts as high as 0.1 K min−1 were observed in the region 350 to 360 K, indicating that additional crystallization may have taken place. Figure 3 shows the spontaneous temperature drifts observed at about 30 min after the termination of the electrical energy input to the calorimeter. The sample was then “stabilized” by maintaining the sample temperature at 360 K for 5 days. The sample was then quenched to temperatures below 100 K at a rate of about 5 K min−1. Upon heating, a spontaneous warming trend starting around 210 K and reaching a peak around 230 K was observed. However large warming drifts persisted to about 300 K. The sample was again quenched from high temperature and then “annealled” at 230 K for 4 days. During the course of heat capacity measurement on the annealed sample, a spontaneous cooling peak was observed around 240 K. Upon further heating a large warming drift again showed up. Finally, heat capacity measurements were performed on the sample after it had been “slowly cooled” from 360 to 200 K at a rate of about 1 K h−1. A rather broad region of spontaneous cooling drifts with a peak around 240 K was observed.

Figure 3. Spontaneous calorimetric temperature drift in branched polyethylene.

Figure 3.

Data within a series are linked by straight lines. All legends same as in figure 1.

The first large warming drift peaks at about 230 K and the first cooling peaks at about 240 K due to quenching and annealing (or slow cooling), respectively, are probably associated with the glass transition phenomena. The large warming drifts seen above 250 K in both quenched samples and quenched samples followed by annealing are however due to crystallization of the polymer above the glass transition temperature.

A commercial differential scanning calorimeter (DSC) was also used to demonstrate qualitatively the behavior of the branched polyethylene sample. A sample of SRM 1476 weighing approximately 10 mg was subjected to various thermal treatments in the DSC and was then observed under identical operational conditions, such as the scanning rate, slope adjustment, temperature calibration adjustments, etc. The recorded traces of the DSC observations are reproduced in figure 4. Each trace was shifted vertically by a certain amount except trace (g) which was shifted a little more than the others. The vertical axis denotes the difference in the power inputs to the sample and to the reference, and hence is related to the difference in the heat capacity between the sample and the reference. The values on the abscissa are the readings of the temperature indicator of the DSC. The annealing temperatures for various traces were also read directly from the indicator. Neither the indicated programming temperature during a scan nor the isothermal temperature during annealing were corrected to the sample temperature. Under the operational conditions of the present observations, these corrections were less than 3 K.

Figure 4. DSC melting curves of branched polyethylene.

Figure 4.

All curves obtained with a scanning rate of 10 K min−1 after various thermal treatments: (a) cooled at − 2.5 K min−1. All other curves, cooled rapidly from the melt and then held at: (b) 350 K, (c) 360 K, (d) 370 K, (e) 360 and then at 370 K, (f) 370 and then at 360 K, (g) 350 and then at 5 K intervals to 375 K.

All the curves are obtained with a scanning speed of 10 K min−1 from 310 to 420 K after the sample has been cooled from the melt and received various thermal treatments. Curve (a) is a smooth trace obtained after cooling the sample continuously at a rate of 2.5 K min−1 from the melt. Curves (b), (c), and (d) show the effect of annealing or holding the sample in the DSC for 15 min at 350, 360, and 370 K, respectively. Annealing lowers the apparent heat capacity, including any premelting and recrystallization phenomena, near the annealing temperature. At 5 to 10 K above the annealing temperature a hump may be observed. The magnitude of these effects is increased with increasing annealing temperature. Curves (e) and (g) show the effect on the sample subjected to a multiple annealing process by successively raising the annealing temperature: (e) first at 360 and then at 370 K and (g) from 350 to 375 K at 5 K intervals. Both the dip near the highest annealing temperature, and the hump above that in these two curves are very pronounced. The apparent heat capacity below 370 K in curve (g) seems to be noticeably higher than the smooth curve (a). Curve (f) indicates the effect produced by annealing first at 370 K and then at 360 K. Two dips and humps may be seen on the same curve. The hump above 370 K is similar in magnitude to that in curve (d), however the hump between 360 and 370 K is somewhat smaller than that in curve (c). Similar phenomena have also been observed to occur in branched polyethylene by differential thermal analysis [33], and in high density polyethylene annealed within 15 K of its melting point by DSC [34].

Branched polyethylene, due to its geometrical irregularities, may form crystallites having a very wide range of melting points. As long as the cooling from the melt is continuous, the melting phenomena is also continuous. The smooth distribution of the crystallites, however, may be modified by discrete thermal treatments. By annealing or holding the sample at a particular temperature, the crystallites which melt near that temperature may either re-crystallize or grow into less defective or larger crystallites. Thus the annealing process may deplete part of crystallites that melt near the annealing temperature and increase the amount of crystallites that melt at higher temperatures. Crystallites melting at much lower temperatures may not be affected by this process as they may not be incorporated into the high melting crystallites and may be formed again upon subsequent cooling. Therefore, in the DSC traces for the annealed samples, the apparent background heat capacity below the annealing temperature is similar to that of a smooth distribution. By correlating the more sensitive plot of heat capacity difference in figure 2b with the plot of observed calorimetric temperature drift in figure 3, one finds that the dips in the heat capacity curves at 290, 300, and 350 K are not associated with any large positive drifts. Therefore the dip in heat capacity curve near the annealing temperature is probably due to a depletion of crystallites which melt in that region.

Acknowledgments

The authors wish to thank C. H. Pearson in assisting with the calorimetric measurement.

Footnotes

1 Figures in brackets indicate the literature references at the end of this paper.

2 Certain commercial materials are identified in this paper in order to adequately specify the experimental procedure. Such identification does not imply recommendation or endorsement by the National Bureau of Standards.

4. References

  • [1].Atkinson C. M. L., and Richardson M. J., Trans. Faraday Soc. 61, 1764 (1969). [Google Scholar]
  • [2].Aukward J. A., Warfield R. W., Petree M. C., and Donovan P., Rev. Sci. Instr. 30, 597 (1959). [Google Scholar]
  • [3].Dainton F. S., Evans D. M., Hoare F. E., and Melia T. P., Polymer 3, 277 (1962). [Google Scholar]
  • [4].Dole M., Hettinger W. P. Jr., Larson N. R., and Wethington J. A. Jr., J. Chem. Phys. 20, 781 (1952). [Google Scholar]
  • [5].Gray A. P., and Brenner N., ACS Polymer Reprint 6, 956 (1965). [Google Scholar]
  • [6].Hager N. E. Jr., Rev. Sci. Instr. 35, 618 (1964). [Google Scholar]
  • [7].Hellewege K. H., Knappe W., and Wetzel W., Kolloid Z. 180, 126 (1962).
  • [8].Issacs L. L., and Garland C. W., J. Phys. Chem. Solids 23, 311 (1962). [Google Scholar]
  • [9].Passaglia E., and Kevorkian H. K., J. Appl. Polymer Sci. 7, 119 (1963). [Google Scholar]
  • [10].Peterlin A., and Meinel G., Polymer 3, 783 (1965). [Google Scholar]
  • [11].Raine H. C., Richards R. B., and Ryder H., Trans Faraday Soc. 41, 56 (1945). [Google Scholar]
  • [12].Reese W., and Tucker J. E., J. Chem. Phys. 43, 105 (1965). [Google Scholar]
  • [13].Richardson M. J., Trans. Faraday Soc. 61, 1876 (1965). [Google Scholar]
  • [14].Sochava I. V., Doklady Akad. Nauk SSSR 130, 126 (1960). [Google Scholar]
  • [15].Sochava I. V., and Trapeznikova O. N., Soviet Phys. Doklady 2, 164 (1957). [Google Scholar]
  • [16].Tautz H., Glück M., Hartmann G., and Leuteritz R., Plaste u. Kautschuk 10, 648 (1963); ibid 11, 657 (1964). [Google Scholar]
  • [17].Tucker J. E., and Reese W., J. Chem. Phys. 46, 1388 (1967). [Google Scholar]
  • [18].Warfield R. W., Petree M. C., Donovan P., SPE Journal 15, 1055 (1959). [Google Scholar]
  • [19].Wilski H., Kunstoffe 50, 281 (1960). [Google Scholar]
  • [20].Wunderlich B., J. Phys. Chem. 69, 2078 (1965). [Google Scholar]
  • [21].Wunderlich B., and Dole M., J. Polymer Sci. 24, 201 (1957). [Google Scholar]
  • [22].Sterrett K. F., Blackburn D. H., Bestul A. B., Chang S. S., and Horman J. A., J. Res. Nat. Bur. Stand. (U.S.), 69C (Eng. and Instr.), No. 1, 19–29 (Jan.–Mar. 1965). [Google Scholar]
  • [23].Chang S. S., and Bestul A. B., J. Res. Nat. Bur. Stand. (U.S.), 75A (Phys. and Chem.), No. 2, 113–120 (Mar.–Apr. 1971). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Chang S. S., Horman J. A., and Bestul A. B., J. Res. Nat. Bur. Stand. (U.S.), 71A (Phys. and Chem.), No. 41, 293–305 (July–Aug. 1967). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Comité International des Poids et Mesures, Metrologia 5, 35 (1969). [Google Scholar]
  • [26].Bedford R. E., Durieux M., Muijlwijk R., and Barber C. R., Metrologia 5, 47 (1969). [Google Scholar]
  • [27].Cataland G., and Plumb H. H., J. Res. Nat. Bur. Stand. (U.S.), 70A (Phys. and Chem.), No. 3, 243–252 (May–June 1966). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Berry R. J., Can J. Phys. 45, 1963 (1967). [Google Scholar]
  • [29].Wagner H. L., and Verdier P. H., The Characterization of Linear Polyethylene SRM 1475, NBS Spec. Publ. 260–42 (1972). Reprinted from J. Res. Nat. Bur. Stand. (U.S.), 76A (Phys. and Chem.), No. 2, 137–170 (Mar.–Apr. 1972). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Osborne N. S., Stimson H. F., Sligh T. S. Jr. and Cragoe C. S., Sci. Papers NBS 20, 65 (1925). [Google Scholar]
  • [31].Westrum E. F. Jr., unpublished. [Google Scholar]
  • [32].Guttman C. M., J. Chem. Phys. 56, 627 (1972). [Google Scholar]
  • [33].Holden H. W., J. Polymer Sci. C6, 53 (1964). [Google Scholar]
  • [34].Harland W. G., Khadr M. M., and Peters R. H., Polymer 13, 13 (1972). [Google Scholar]

Articles from Journal of Research of the National Bureau of Standards. Section A, Physics and Chemistry are provided here courtesy of National Institute of Standards and Technology

RESOURCES