Skip to main content
PLOS One logoLink to PLOS One
. 2019 Sep 27;14(9):e0219276. doi: 10.1371/journal.pone.0219276

Inhibition of polymerase chain reaction: Pathogen-specific controls are better than human gene amplification

Guillaume Roux 1,¤, Christophe Ravel 2, Emmanuelle Varlet-Marie 2, Rachel Jendrowiak 2, Patrick Bastien 2, Yvon Sterkers 2,*
Editor: Antonella Marangoni3
PMCID: PMC6764677  PMID: 31560697

Abstract

PCR inhibition is frequent in medical microbiology routine practice and may lead to false-negative results; however there is no consensus on how to detect it. Pathogen-specific and human gene amplifications are widely used to detect PCR inhibition. We aimed at comparing the value of PCR inhibitor detection using these two methods. We analysed Cp shifts (ΔCp) obtained from qPCRs targeting either the albumin gene or the pathogen-specific sequence used in two laboratory-developed microbiological qPCR assays. 3152 samples including various matrixes were included. Pathogen-specific amplification and albumin qPCR identified 62/3152 samples (2.0%), and 409/3152 (13.0%) samples, respectively, as inhibited. Only 16 samples were detected using both methods. In addition, the use of the Youden's index failed to determine adequate Cp thresholds for albumin qPCR, even when we distinguished among the different sample matrixes. qPCR targeting the albumin gene therefore appears not adequate to identify the presence of PCR inhibitors in microbiological PCR assays. Our data may be extrapolated to other heterologous targets and should discourage their use to assess the presence of PCR inhibition in microbiological PCR assays.

Introduction

Molecular biology and particularly real-time PCR (qPCR) has revolutionized the biological diagnosis of infectious diseases. Nonetheless, PCR may fail because it is based on an enzymatic reaction susceptible to various mechanisms of inhibition [1]. Inhibition of PCR reaction is frequent in clinical microbiology and exposes to the risk of false negative results, hence PCR inhibition screening is recommended [2, 3]. PCR inhibition appears as a hardly predictable event and data about its actual frequency in routine practice of clinical biology laboratories are scarce [4, 5]. Differential susceptibility of each type of qPCR to different inhibitors and heterogeneity of sample matrixes make its detection non trivial. Many methods to detect PCR inhibition have been reported and some guidelines have been published. The use of PCR controls with a defined quantity of DNA molecules to check for the presence of a Cp switch due to the presence of inhibitors in the sample extract is relevant; however it is risky since technicians may have to manipulate and distribute target DNA, thus increasing the risk of inter-well contamination. 'Internal' amplification controls, based on alien DNA added at a low concentration in the specimen before DNA extraction, and of which the presence must be checked in a duplex PCR together with the target, is also highly relevant. Most of commercial inhibition controls are based on this principle. The rationale of using a human gene as an extraction or inhibition control is less acceptable because the human DNA target is present in high quantity in the sample as compared to the target DNA [6]. Importantly, there is presently no consensus for PCR inhibition detection in routine practices. For example, in the single field of molecular diagnosis of toxoplasmosis in France, seven different methods are used to detect PCR inhibition in 30 laboratories [7]. Most of them use either a pathogen pathogen-specific PCR or a human gene amplification method. The former is an internal control made of either genomic T. gondii DNA or a plasmid containing the targeted pathogen DNA sequence. It has one major drawback which is the increased risk of false positive results due to increased amounts of amplicons and airborne contamination of reaction wells. Amplification controls may be prepared from genomic DNA (as in our Toxoplasma-PCR assay) or from the target DNA sequence cloned in a plasmid (as in our Pneumocystis-PCR assay). One development of the plasmid strategy is to clone a chimeric sequence to be amplified by the same primers than the PCR target but detected by specific probe(s) [811]. In the second type of method, albumin, beta-globin or human RNase P genes are targeted. This method is found attractive since (i) they can be implemented in all qPCR assays involving human samples; and (ii) they constitute a complete process control for DNA extraction and amplification. Yet, human gene qPCR cycle of positivity (Cp) depends also on the initial human DNA content or cellularity in the clinical sample, which is highly variable. Indeed, DNA content or cellularity depends on the size/volume of the sample, the matrix, i.e. the nature of the sample, such as blood or cerebrospinal fluid, and on the pathophysiological state of the patients. The control of DNA extraction is another critical step of molecular diagnosis, but was not explored in our study. In addition, it is reported in the literature that PCR methods differing by their primers and/or amplified sequence have variable susceptibility to inhibitors [1214]. These studies tend to invalidate the assumption that absence of inhibition in a qPCR targeting any human gene has a good predictive value for assessing inhibition in a qPCR targeting a pathogen. Consequently, it is critical to assess the performances of human gene-based PCR inhibition screening methods in clinical samples in clinical microbiology routine practice.

In this study, (i) we analyse the frequency of PCR inhibition in a large range of clinical samples and (ii) we show that human gene-based qPCR methods are not efficient to affirm the presence/absence of PCR inhibitors.

Methods

We retrospectively analysed Toxoplasma and Pneumocystis qPCR tests for all the clinical samples analysed in 2016 in the Department of Parasitology-Mycology at the academic hospital of Montpellier (Montpellier, France).

DNA extraction

The DNA extraction method was dependent on the PCR assay and the sample matrix. For Pneumocystis-PCR, all types of respiratory samples were analysed, including bronchoalveolar lavage fluid (BALF), sputum, bronchoaspiration, nasal aspiration and pleural fluid; DNA was extracted using QIAamp DNA minikit (Qiagen) according to the manufacturer specifications. For Toxoplasma-PCR, the DNA extraction method depended on the matrix. For low cellular samples such as cerebrospinal fluids (CSF), aqueous humor, amniotic fluids and crystal clear BALF, DNA was extracted using the Tween-Nonidet-NaOH method[15]. When Pneumocystis and Toxoplasma qPCRs were performed on the same sample, QIAamp DNA minikit (Qiagen) was used. Other samples were processed using protein precipitation solution (A795A; Promega, France)[16]. As recommended, DNA extracts were then stored at– 20°C prior to PCR for best preservation[17].

Toxoplasma and Pneumocystis qPCR assays

The Toxoplasma qPCR targeted the ‘rep529’ sequence (AF146527) and was performed as described by Reischl et al. [18]. The Pneumocystis qPCR targeted the major surface glycoprotein gene (AF372980) and was performed as described by Fillaux et al.[19]. We used LC480® Probe Master 2X (Roche®) for both qPCR and Uracil-N-Glycosylase (Roche®) for Toxoplasma qPCR. Each PCR well contained 15 μL of mix and 5 μL of DNA extract for Toxoplasma qPCR and 18 μL of mix and 2 μL of DNA extract for Pneumocystis qPCR. Amplification and detection occurred in LightCycler® 480 (Roche) and raw fluorescence data were analysed using the LightCycler® 480 software release 1.5.0 (Roche). Cycles of positivity were determined by the second derivative method.

PCR controls and definition of inhibition

Each PCR plate contained one well for negative control (sterile water) and one for positive control (calibrated positive sample). Cp values of positive plate controls were plotted into Levey-Jennings charts and interpreted according to the Westgard rules[20]. DNA extraction control was performed in routine practice by qPCR targeting the human albumin gene[21]. Amplification controls were made of reaction wells containing pathogen DNA on top of the patient DNA extract. For the Toxoplasma qPCR assay, this pathogen-specific amplification positive control was prepared from a 105 T. gondii tachyzoites/mL freeze-dried standard as described in Varlet-Marie et al. [22]; 2 μL of T. gondii DNA extract were added to obtain a final concentration of 1.5 T. gondii genome/tube for which the expected Cp value is 35.2 ±1.5. The control was performed in duplicate. A sample was considered inhibited if both amplification control wells showed a Cp>38. If only one well showed a Cp>38, we considered this as irrelevant and due to stochastic events. For Pneumocystis-qPCR, the amplification control was made of the pathogen target DNA sequence, which had been cloned into pGEM-T easy®. Plasmid DNA maxipreparation (Qiagen Maxiprep®) was then diluted in order to reach a Cp of 30.1 ±1.2, and 1 μL of this dilution was added to the reaction tube. A Pneumocystis-qPCR reaction was considered inhibited if the Cp difference between amplification control and the positive plate control was ≥3. In this study, albumin-qPCR was considered as inhibited if the difference between the Cp of albumin-qPCR (Cpalb) obtained for a sample and the mean Cpalb for the matrix considered was ≥3.

Data management

PCR results and bio-clinical data were exported from LightCycler® 480 release 1.5.0 software (Roche®) and DxLab® software (MedaSys®) respectively, to spreadsheets (S1 and S2 Figs).

Determination of the PCR efficiencies

PCR efficiencies of the techniques were determined using a standard curve generated by performing a logarithmic serial dilution of DNA extracts. The standard curves demonstrating a linear relationship between the logarithm of the copy number and the Cp value, allows to determine the PCR efficiency using: Efficiency = -1+10exp(-1/slope) [23].

Statistics

Data were analyzed using R 3.4.0 and RStudio 1.0.143 softwares [24, 25]. We calculated the Youden’s index considering the pathogen-specific amplification control as the reference method to detect PCR inhibition. The Youden’s index combines sensitivity and specificity into a single measure (Sensitivity + Specificity– 1) and its value ranges between -1 and 1. A value of 1 indicates that there are neither false positive nor false negative results. The best Cpalb cut-off values of albumin-qPCR to detect inhibition were defined using the Youden's index. Sensitivities and specificities of these Cpalb cut-offs were next compared by binomial tests to the 95% theoretical value in order to assess their inferiority. Correlation of Cpalb to leucocyte counts was assessed using Spearman’s rank correlation. Holm correction was performed for all the tests. Samples without Cpalb value due to a flat amplification curve were excluded from the graph and the correlation. A p value <0.05 was considered as significant.

Ethical approval and informed consent

This work was carried out in accordance with the relevant guidelines and regulations; it does not include potentially identifying patient/participant information. The study corresponds to a non-interventional retrospective study and according to the French Health Public Law (CSP Art L1121-1.1), such studies are exempt from informed consent requirement and do not require approval by an ethics committee.

Results and discussion

General description

The Dept. of Parasitology-Mycology is a regional and national reference centre which coordinates the "Molecular biology pole" of the National Reference Center for Toxoplasmosis (http://cnrtoxoplasmose.chu-reims.fr). The laboratory is accredited according to ISO 15189:2012 for the Toxoplasma and Pneumocystis PCR assays. The Toxoplasma-qPCR assay is being used in routine since July 2009, >17000 clinical samples have been routinely tested, and PCR efficiency is regularly controlled and found at ≥97.5% [18]. For the Pneumocystis-qPCR assay, the corresponding features are March 2013, >2500, and 90 ±9%. PCR efficiency of the albumin qPCR was 98.5% (Fig 1). These laboratory-developed qPCR assays may therefore be considered as robust and finely optimized.

Fig 1. PCR efficiencies.

Fig 1

A. LightCycler real-time PCR amplification curves in a serial dilution assay. B-D. PCR efficiencies were determined in triplicate in two/three independent experiments. PCRs were performed in triplicate on five/six samples representing a 10-fold serial dilution. All standard curves demonstrated a linear relationship between the logarithm of the copy number and the Cp value, allowing determining the PCR Efficiency using: Efficiency = -1+10exp(-1/slope).

During the year 2016, 3152 samples were referred to the laboratory as part of the routine activity (Fig 2): 2225 samples were analysed by Toxoplasma-qPCR, 964 by Pneumocystis-qPCR, and 37 samples were analysed by both methods (S1 Table). Two samples analysed by Toxoplasma-qPCR were excluded from this study, one due to cancellation of the analysis and another due to identity problem (S1 and S2 Figs). Most Toxoplasma-qPCR analysed samples were blood samples (1874 out of 2225), followed by 95 CSF and 114 postnatal samples (Table 1). No Pneumocystis-qPCR samples had to be excluded from the analysis; the 964 Pneumocystis-qPCR samples comprised 787 BALF and 141 sputa samples (Table 1). The percentage of positive samples for the Toxoplasma-qPCR was 1.7%, ranging from 0.3% for the blood to 12% for the amniotic fluid samples. For Pneumocystis-qPCR, the percentage of positive sample was 7.1% (Table 1).

Fig 2. Venn diagram of number of analyzed samples.

Fig 2

Inhibited samples: Samples were considered inhibited if the Cp of the pathogen-specific control was >38 for both wells for Toxoplasma-PCR (N = 45), and if ΔCp ≥3 for Pneumocystis-PCR (N = 17). Cp(albumin)>threshold: For albumin-PCR, the inhibition threshold was the mean Cpalb + 3 for a given matrix (see Methods and Table 1).

Table 1. Frequency of PCR inhibition as a function of sample matrix for Toxoplasma PCR and Pneumocystis PCR.

        Toxoplasma PCR     Pneumocystis PCR
    Positive samples Inhibited samples Cpalb≥ threshold* Positive samples Inhibited samples Cpalb≥ threshold*
    Nber % Nber % Threshold Nber % Nber % Nber % Threshold Nber %
Blood   4/1309 0.3 31/1309 2,3 23,1 104/1309 7,9 - - - - - - -
Blood from leucopenic patient 7/565 1.2 2/565 0,4 27,6 123/565 21,8 - - - - - - -
CSF   1/95 1.1 3/95 3.2 34,9 16/95 16,8 - - - - - - -
Placenta   3/58 5.2 4/58 6.9 21,7 7/58 12,1 - - - - - - -
Cord blood   2/56 3.6 2/56 3.6 23,8 10/56 17,9 - - - - - - -
AF   15/44 34.1 0/44 0 29,3 5/44 11,4 - - - - - - -
BALF**   1/41 2.4 0/41 0 29.9 5/41 12.2 51/787 6,5 12/787 1.5 24.7 100/787 12.7
Aqueous humor 2/28 7.1 0/28 0 33,6 8/28 28,6 - - - - - - -
Sputum   - - - - - - - 16/141 11.4 4/141 2.8 23,9 23/141 16,3
Bronchoaspiration - - - - - - - 1/35 2.9 1/35 2.8 22,1 4/35 11,4
Miscellaneous** 1/29 3.4 3/29 10,3 32 7/29 24,1 0/1 0 0/1 0 24.7 0/1 0
Cerebral biopsy 1/14 7.1 2/14 14.3 28,4 1/14 7.1 - - - - - - -
  Lymph node 0/6 0 0/6 0 28,4 1/6 16.7 - - - - - - -
  Liver biopsy 0/2 0 1/2 50 28,4 1/2 50 - - - - - - -
  Ascites 0/2 0 0/2 0 28,4 1/2 50 - - - - - - -
  Pericardia fluid 0/2 0 0/2 0 28,4 1/2 50 - - - - - - -
  Pleural fluid 0/1 0 0/1 0 28,4 1/1 100 - - - - - - -
  Bone marrow 0/1 0 0/1 0 28,4 0/1 0 - - - - - - -
  Retina 0/1 0 0/1 0 28,4 1/1 100 - - - - - - -
  Nasal aspiration - - - - - - - 0/1 0 0/1 - 24,7 0/1 0
Total   37/2225 1,7 45/2225 2   285/2225 12,8 68/964 7,1 17/964 1.8   127/964 13,2

CSF: cerebrospinal fluid; AF: amniotic fluid (including external quality assessment samples); BALF: bronchoalveolar lavage fluid. Samples were considered inhibited if Cp of amplification controls was >38 for both amplification control wells for Toxoplasma PCR, and if ΔCp was ≥3 for Pneumocystis PCR

*threshold = mean Cp + 3

** for Toxoplasma and Pneumocystis PCR respectively, 37 BALF samples were analysed by both methods (for details, see Methods).

Inhibition rates in clinical samples as assessed from the pathogen-specific amplification control

Of the 3152 analysed samples, 62 (2.0%) showed evidence of PCR inhibition as assessed from the pathogen-specific amplification control: 45 samples for Toxoplasma-qPCR and 17 for Pneumocystis-qPCR (Fig 2). However, this percentage varied from 1.8–15% depending on the matrix (Table 1). In 61/62 inhibited samples, tenfold dilution of the DNA extract released inhibition, confirming that the unexpected Cp values were due to PCR inhibitors. The remaining sample was a cord blood tested for the presence of Toxoplasma for which PCR remained inhibited despite the dilution, showing flat amplification curves for both the target- and the albumin-specific inhibition controls before and after tenfold dilution. No additional dilution was applied to this sample due to the risk of reduced sensitivity. Interestingly, blood samples, which are reported elsewhere [26] to be prone to PCR inhibition did not exhibit the highest inhibition rate in our study. Samples like sputum or placenta were also prone to PCR inhibition, probably because it is difficult to optimize a DNA extraction protocol for heterogeneous matrixes with highly different structures and properties from one sample to another. Using a similar pathogen-specific control method to detect PCR inhibitors in BALF samples, Döskaya et al. found a 23.8% inhibition rate for Pneumocystis PCR [5] but a large study of Buckwalter et al. reported inhibition rates under 1% on 386,706 samples [4]. These discrepant results illustrate that PCR inhibition measurement is a problem of variable importance in routine practices, probably depending on the ability of extraction methods to remove inhibitors, on the susceptibility of PCR methods to inhibitors and on the performance of PCR inhibition detection.

Performances of PCR inhibition detection using albumin-qPCR

409/3152 samples (12.9%) exhibited a Cpalb above the threshold, i.e. above the mean Cp for the matrix concerned + 3 (Table 1). Thus, (i) the albumin-qPCR identified many more samples as containing inhibitors than the pathogen-specific controls, and this was so in both qPCR assays, i.e. 285 vs. 45 for the Toxoplasma assay, and 127 vs. 17 for the Pneumocystis assay. In addition, (ii) it is noteworthy that only 13 and 3 samples were detected as inhibited using both inhibitor-detection methods, for the Toxoplasma and Pneumocystis assays, respectively (Fig 2). We therefore conclude that (i) the human gene-based qPCR detected inhibitors much more often than the pathogen-specific qPCR and (ii) it most often did not detect them correctly. Considering this poor correlation between pathogen-specific controls and albumin qPCR to detect PCR inhibitors, we wished to determine Cpalb thresholds more adapted to our assays. To do so, we calculated the sensitivity, specificity of the albumin qPCR to detect PCR-inhibiting samples using the pathogen-specific control as reference and the Youden's index (Sensitivity + Specificity– 1) and we varied the Cpalb threshold (see Methods and Fig 3A). We could not determine a Cpalb threshold allowing to obtain 95% sensitivity and 95% specificity (Fig 3B), whether this strategy was used for all matrixes together or for each type of matrix (S3 and S4 Figs). We therefore concluded that no Cpalb cut-off values can be determined to efficiently detect pathogen-specific inhibitions for both pathogen-specific PCR assays. To check the influence of the cellularity of clinical samples on Cpalb values, we compared white blood cell counts to albumin qPCR results for blood samples. We analysed 1690/1874 blood samples; the remaining 184 samples were excluded from the analysis due to the absence of whole blood cell count on the same day as Toxoplasma-PCR. Cpalb values were correlated to leucocyte counts (p<0.0001; Spearman’s rank correlation), but for a given white blood cell count a large range of Cpalb was observed (Fig 4). The lack of sensitivity and specificity of albumin PCR to detect target specific inhibitions may be explained by a differential susceptibility of each qPCR assay to inhibitors and by the variable quantity of human DNA in clinical samples. These results prevent using human gene-based PCR, e.g. albumin, betaglobin or human RNase P genes, as PCR inhibition detection method. We based our demonstration on two models, i.e. Toxoplasma and Pneumocystis but these results should be expanded to the detection of other pathogens in human samples.

Fig 3. Youden’s index, sensitivity and specificity of albumin-PCR cut-off values.

Fig 3

A and C. Black curves: Youden’s index = sensitivity + specificity– 1; B and D. Red curves: sensitivity; blue curves: specificity. For Toxoplasma qPCR (A), the maximal Youden’s index was very low. The maximum value was 0.19 and was reached for a Cpalb cut-off value of 36. At this albumin qPCR cut-off value sensitivity is 20% (p< 0.0001; Binomial test for 95% theoretical sensitivity) and specificity 98.8% (p = 1, Binomial test for 95% theoretical specificity). (B) Graphical analysis of sensitivity and specificity curves in function of cut-off values. Both curves intersect at a Cpalb value of 22. At this Cpalb cut-off value sensitivity is 33.3% and specificity 66.6% (p<0.0001 and p<0.0001; Binomial tests for 95% theoretical sensitivity and specificity). For Pneumocystis qPCR (C), the maximal Youden’s index was too low to allow determining any Cpalb cut-off value. (D) Graphical profiles of sensitivity and curves in function of cut-off values. Both curves intersect at a Cpalb value of 22. At this Cpalb cut-off value sensitivity is 47.1% and specificity 62% (p<0.0001 and p<0.0001; Binomial tests for 95% theoretical sensitivity and specificity).

Fig 4. Correlation between leucocyte counts and Cpalb values in 1690 blood samples.

Fig 4

Semi-log scale graph where leucocyte count (x-axis) was plotted against Cpalb (orange and green dots corresponding to amplification control inhibited and not inhibited samples respectively). Linear model was used to determine trend line (Blue) with 95% confidence interval (Grey).

Another widely used method to search for inhibitors is the use of commercial “universal” controls. These commercial controls are made of exogenous DNA and are added either in the sample before extraction or with the DNA extract in the PCR mix. Differential susceptibility to inhibitors and efficiency discrepancies between PCR assays should also prove problematic in this approach. Indeed, the size and GC rate of the amplicons of the foreign DNA used will have an impact on the detection of the inhibition [13]. So, implementation of one of these controls should be avoided until their performances have been assessed in routine practice.

In conclusion, pathogen-specific amplification controls appear to be a method of choice for screening the presence of inhibitors in a PCR assay for infectious diseases as compared to the use of a human gene-based qPCR.

Supporting information

S1 Table. Spreadsheet of raw data.

(XLSX)

S1 Fig. Flow chart of data cleaning and tidying of Toxoplasma PCR results.

(DOCX)

S2 Fig. Flow chart of data cleaning and tidying of Pneumocystis PCR results.

(DOCX)

S3 Fig. Youden’s index, sensitivity and specificity of albumin PCR cut-off values for the most frequent matrixes analysed in Toxoplasma PCR.

(DOCX)

S4 Fig. Youden’s index, sensitivity and specificity of albumin PCR cut-off values for the most frequent matrixes analysed in Pneumocystis PCR.

(DOCX)

Acknowledgments

We are grateful to S. Douzou, B. Sanichanh and G. Serres for their technical help.

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

"RSI Assurance Maladie Professions Libérales - Provinces. C.A.M.P.L.P." has paid the LightCycler 480 (Roche®) real-time PCR equipment in 2009. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Hedman J, Radstrom P. Overcoming inhibition in real-time diagnostic PCR. Methods Mol Biol. 2013;943:17–48. Epub 2012/10/30. 10.1007/978-1-60327-353-4_2 . [DOI] [PubMed] [Google Scholar]
  • 2.Wallace PS, MacKay WG. Quality in the molecular microbiology laboratory. Methods Mol Biol. 2013;943:49–79. Epub 2012/10/30. 10.1007/978-1-60327-353-4_3 . [DOI] [PubMed] [Google Scholar]
  • 3.Schrader C, Schielke A, Ellerbroek L, Johne R. PCR inhibitors—occurrence, properties and removal. J Appl Microbiol. 2012;113(5):1014–26. Epub 2012/07/04. 10.1111/j.1365-2672.2012.05384.x . [DOI] [PubMed] [Google Scholar]
  • 4.Buckwalter SP, Sloan LM, Cunningham SA, Espy MJ, Uhl JR, Jones MF, et al. Inhibition controls for qualitative real-time PCR assays: are they necessary for all specimen matrices? J Clin Microbiol. 2014;52(6):2139–43. Epub 2014/04/18. 10.1128/JCM.03389-13 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Doskaya M, Caner A, Degirmenci A, Wengenack NL, Yolasigmaz A, Turgay N, et al. Degree and frequency of inhibition in a routine real-time PCR detecting Pneumocystis jirovecii for the diagnosis of Pneumocystis pneumonia in Turkey. J Med Microbiol. 2011;60(Pt 7):937–44. Epub 2011/04/05. 10.1099/jmm.0.030775-0 . [DOI] [PubMed] [Google Scholar]
  • 6.Bustin SA, Beaulieu JF, Huggett J, Jaggi R, Kibenge FS, Olsvik PA, et al. MIQE precis: Practical implementation of minimum standard guidelines for fluorescence-based quantitative real-time PCR experiments. BMC Mol Biol. 11. England2010. p. 74. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Sterkers Y, Varlet-Marie E, Marty P, Bastien P. Diversity and evolution of methods and practices for the molecular diagnosis of congenital toxoplasmosis in France: a 4-year survey. Clin Microbiol Infect. 2010;16(10):1594–602. 10.1111/j.1469-0691.2009.03101.x . [DOI] [PubMed] [Google Scholar]
  • 8.Rosenstraus M, Wang Z, Chang SY, DeBonville D, Spadoro JP. An internal control for routine diagnostic PCR: design, properties, and effect on clinical performance. J Clin Microbiol. 1998;36(1):191–7. Epub 1998/02/12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Maaroufi Y, de Bruyne JM, Duchateau V, Scheen R, Crokaert F. Development of a multiple internal control for clinical diagnostic real-time amplification assays. FEMS Immunol Med Microbiol. 2006;48(2):183–91. Epub 2006/10/27. 10.1111/j.1574-695X.2006.00125.x . [DOI] [PubMed] [Google Scholar]
  • 10.Brenier-Pinchart MP, Morand-Bui V, Fricker-Hidalgo H, Equy V, Marlu R, Pelloux H. Adapting a conventional PCR assay for Toxoplasma gondii detection to real-time quantitative PCR including a competitive internal control. Parasite. 2007;14(2):149–54. Epub 2007/07/25. 10.1051/parasite/2007142149 . [DOI] [PubMed] [Google Scholar]
  • 11.Daval S, Poirier P, Armenaud J, Cambon M, Livrelli V. [Development of a real-time PCR assay for quantitative diagnosis of Toxoplasma gondii after allogeneic bone marrow transplantation]. Pathol Biol (Paris). 2010;58(1):104–9. Epub 2009/11/07. 10.1016/j.patbio.2009.07.013 [DOI] [PubMed] [Google Scholar]
  • 12.Huggett JF, Novak T, Garson JA, Green C, Morris-Jones SD, Miller RF, et al. Differential susceptibility of PCR reactions to inhibitors: an important and unrecognised phenomenon. BMC Res Notes. 2008;1:70 Epub 2008/08/30. 10.1186/1756-0500-1-70 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Pionzio AM, McCord BR. The effect of internal control sequence and length on the response to PCR inhibition in real-time PCR quantitation. Forensic Sci Int Genet. 2014;9:55–60. Epub 2014/02/18. 10.1016/j.fsigen.2013.10.010 . [DOI] [PubMed] [Google Scholar]
  • 14.Chabbert E, Lachaud L, Crobu L, Bastien P. Comparison of two widely used PCR primer systems for detection of toxoplasma in amniotic fluid, blood, and tissues. J Clin Microbiol. 2004;42(4):1719–22. 173. 10.1128/JCM.42.4.1719-1722.2004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Hohlfeld P, Daffos F, Costa JM, Thulliez P, Forestier F, Vidaud M. Prenatal diagnosis of congenital toxoplasmosis with a polymerase-chain-reaction test on amniotic fluid. N Engl J Med. 1994;331(11):695–9. Epub 1994/09/15. 10.1056/NEJM199409153311102 . [DOI] [PubMed] [Google Scholar]
  • 16.Sterkers Y, Pratlong F, Albaba S, Loubersac J, Picot MC, Pretet V, et al. Novel interpretation of molecular diagnosis of congenital toxoplasmosis according to gestational age at the time of maternal infection. J Clin Microbiol. 2012;50(12):3944–51. Epub 2012/10/05. 10.1128/JCM.00918-12 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Delhaes L, Filisetti D, Brenier-Pinchart MP, Pelloux H, Yera H, Dalle F, et al. Freezing and storage at -20 degrees C provides adequate preservation of Toxoplasma gondii DNA for retrospective molecular analysis. Diagn Microbiol Infect Dis. 2014;80(3):197–9. Epub 2014/09/23. 10.1016/j.diagmicrobio.2014.08.007 . [DOI] [PubMed] [Google Scholar]
  • 18.Reischl U, Bretagne S, Kruger D, Ernault P, Costa JM. Comparison of two DNA targets for the diagnosis of Toxoplasmosis by real-time PCR using fluorescence resonance energy transfer hybridization probes. BMC Infect Dis. 2003;3:7 Epub 2003/05/06. 10.1186/1471-2334-3-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Fillaux J, Berry A. Real-time PCR assay for the diagnosis of Pneumocystis jirovecii pneumonia. Methods Mol Biol. 2013;943:159–70. Epub 2012/10/30. 10.1007/978-1-60327-353-4_11 . [DOI] [PubMed] [Google Scholar]
  • 20.Westgard JO, Barry PL, Hunt MR, Groth T. A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem. 1981;27(3):493–501. Epub 1981/03/01. . [PubMed] [Google Scholar]
  • 21.Mary C, Faraut F, Lascombe L, Dumon H. Quantification of Leishmania infantum DNA by a real-time PCR assay with high sensitivity. J Clin Microbiol. 2004;42(11):5249–55. Epub 2004/11/06. 10.1128/JCM.42.11.5249-5255.2004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Varlet-Marie E, Sterkers Y, Brenier-Pinchart MP, Cassaing S, Dalle F, Delhaes L, et al. Characterization and multicentric validation of a common standard for Toxoplasma gondii detection using nucleic acid amplification assays. J Clin Microbiol. 2014;52(11):3952–9. Epub 2014/09/05. 10.1128/JCM.01906-14 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Kralik P, Ricchi M. A Basic Guide to Real Time PCR in Microbial Diagnostics: Definitions, Parameters, and Everything. Front Microbiol. 2017;8:108 Epub 2017/02/18. 10.3389/fmicb.2017.00108 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.RCore-Team. R: a language and environment for statistical computing: R Foundation for Statistical Computing, Vienna, Austria; 2017. Available from: https://www.R-project.org. [Google Scholar]
  • 25.RStudio-Team. R Studio: Integrated Development for R: RStudio, Inc, Boston, MA; 2015. Available from: http://www.rstudio.com. [Google Scholar]
  • 26.Al-Soud WA, Radstrom P. Purification and characterization of PCR-inhibitory components in blood cells. J Clin Microbiol. 2001;39(2):485–93. Epub 2001/02/07. 10.1128/JCM.39.2.485-493.2001 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Antonella Marangoni

26 Jul 2019

PONE-D-19-16663

Original research. Inhibition of Polymerase Chain Reaction: Pathogen-Specific Controls are better than Human Gene Amplification

PLOS ONE

Dear Dr Sterkers,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

==============================

The paper is definetely worthy, but I feel that the statistics evaluation could be improved.

Follow the suggestions given by the referee and check the quality of your images.

==============================

We would appreciate receiving your revised manuscript by August, 14th. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

Antonella Marangoni, Ph.D.

Academic Editor

PLOS ONE

Journal requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following in the Financial Disclosure section:

""RSI Assurance Maladie Professions Libérales - Provinces. C.A.M.P.L.P." has paid the LightCycler 480 (Roche®) real-time PCR equipment in 2009

The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

We note that you received funding from a commercial source: RSI Assurance Maladie Professions Libérales - Provinces. C.A.M.P.L.P.

Please provide an amended Competing Interests Statement that explicitly states this commercial funder, along with any other relevant declarations relating to employment, consultancy, patents, products in development, marketed products, etc.

Within this Competing Interests Statement, please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests). If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please respond by return email with your amended Competing Interests Statement and we will change the online submission form on your behalf.

Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Authors present their work on inhibition of PCR. I found this work useful. Following are my few suggestions in general and few minor edits for the manuscript, that authors may consider.

Overall, it’ll be a good to provide a brief discussion of the work in the context of it’s utility to other type of pathogens as well. In addition, it’ll be helpful for others if authors could provide the analysis pipeline on GitHub or something similar. Since they used R, therefore it should be easy. Authors may also improve the quality of figures as I found them bit blurred.

Minor comments

Line 140: I think it should be pathogen- specific amplification.

Line 148-150: Is the p-value significant computed after correction or before?

Line 194: should be pathogen-specific control. Also, please check space issues elsewhere in the article.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2019 Sep 27;14(9):e0219276. doi: 10.1371/journal.pone.0219276.r002

Author response to Decision Letter 0


16 Aug 2019

We were able to addresses all the points raised during the review process, see the uploaded response to reviewer file

Attachment

Submitted filename: Roux_Inhibitors_ResponseToReviewers_PLoSOnev190816.docx

Decision Letter 1

Antonella Marangoni

26 Aug 2019

[EXSCINDED]

Original research. Inhibition of Polymerase Chain Reaction: Pathogen-Specific Controls are better than Human Gene Amplification

PONE-D-19-16663R1

Dear Dr. Sterkers,

We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements.

Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication.

Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

With kind regards,

Antonella Marangoni, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Acceptance letter

Antonella Marangoni

20 Sep 2019

PONE-D-19-16663R1

Inhibition of Polymerase Chain Reaction: Pathogen-Specific Controls are better than Human Gene Amplification.

Dear Dr. Sterkers:

I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

For any other questions or concerns, please email plosone@plos.org.

Thank you for submitting your work to PLOS ONE.

With kind regards,

PLOS ONE Editorial Office Staff

on behalf of

PhD Antonella Marangoni

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Table. Spreadsheet of raw data.

    (XLSX)

    S1 Fig. Flow chart of data cleaning and tidying of Toxoplasma PCR results.

    (DOCX)

    S2 Fig. Flow chart of data cleaning and tidying of Pneumocystis PCR results.

    (DOCX)

    S3 Fig. Youden’s index, sensitivity and specificity of albumin PCR cut-off values for the most frequent matrixes analysed in Toxoplasma PCR.

    (DOCX)

    S4 Fig. Youden’s index, sensitivity and specificity of albumin PCR cut-off values for the most frequent matrixes analysed in Pneumocystis PCR.

    (DOCX)

    Attachment

    Submitted filename: Roux_Inhibitors_ResponseToReviewers_PLoSOnev190816.docx

    Data Availability Statement

    All relevant data are within the paper and its Supporting Information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES