Abstract
Vibrational spectroscopy, comprised of infrared absorption and Raman scattering spectroscopy, is widely used for label-free optical sensing and imaging in various scientific and industrial fields. The two molecular spectroscopy methods are sensitive to different types of vibrations and provide complementary vibrational spectra, but obtaining complete vibrational information with a single spectroscopic device is challenging due to the large wavelength discrepancy between the two methods. Here, we demonstrate simultaneous infrared absorption and Raman scattering spectroscopy that allows us to measure the complete broadband vibrational spectra in the molecular fingerprint region with a single instrument based on an ultrashort pulsed laser. The system is based on dual-modal Fourier-transform spectroscopy enabled by efficient use of nonlinear optical effects. Our proof-of-concept experiment demonstrates rapid, broadband and high spectral resolution measurements of complementary spectra of organic liquids for precise and accurate molecular analysis.
Subject terms: Infrared spectroscopy, Raman spectroscopy, Atomic and molecular interactions with photons
Generally infrared and Raman spectroscopic methods are needed to study the symmetric and asymmetric molecular vibrational modes. Here the authors demonstrate complementary vibrational spectroscopy to organic molecules by simultaneously measuring their symmetric and anti-symmetric vibrations with one setup.
Introduction
Vibrational spectroscopy is a fundamental method for chemical analysis used in a variety of scientific fields such as organic/inorganic chemistry, geology, biomedical, material, food, environmental, and forensic science1–5. The label-free noninvasive molecular spectroscopy enables us to acquire bond-specific chemical information of specimen, and it is known that infrared (IR) absorption and Raman scattering spectroscopy provide complementary information of molecular vibrations: the former is active for anti-symmetric vibrations that alter the dipole moment, while the latter for symmetric vibrations that alter the polarizability1. IR absorption spectroscopy, which is active for polar bonds such as O–H or N–H, is often used for identification of functional groups of molecules, while Raman scattering spectroscopy, active for bonds such as C=C, S–S, or C–S4, is used for identification of skeletal structures. The group theory states that fundamental vibrational modes of molecules with the center of symmetry cannot be both IR and Raman active (which is known as the rule of mutual exclusion1,6, while there are some exceptions6,7). Therefore, to acquire the complete information of molecular vibrations for more accurate and precise chemical analysis, both the IR and Raman spectra must be measured. Measuring the complete information of molecular vibrations enables us to analyze complex molecular phenomena such as catalytic chemical reactions8–11.
Simultaneous measurement of IR and Raman spectra is a grand challenge in spectroscopy because wavelength regions of these two spectroscopy methods are largely separated, that is, mid-infrared (2.5–25 µm, corresponding to 400–4000 cm−1) for IR spectroscopy and visible to near-infrared (0.4–1 µm, corresponding to 10,000–25,000 cm−1) for Raman spectroscopy, respectively. Since this large wavelength discrepancy causes the difficulty of sharing light sources and optics, a primitive combination of conventional Fourier-transform infrared spectroscopy (FT-IR) and Raman spectrometers12,13 has never been a convincing approach. Such a system requires a complex instrument comprises different spectroscopy methods based on a Michelson interferometer and a dispersive spectrometer with two independent light sources such as an incoherent lamp source and a visible continuous-wave laser. Additionally, these conventional methods do not provide state-of-the-art sensitivity and data acquisition speed because of the low brightness of the lamp source for FT-IR (especially when spatial mode filtering is required to have a small focusing spot) and the inherent weakness of spontaneous Raman scattering. Meanwhile, the technical advancement of nonlinear optics based on ultrashort pulsed lasers has enabled us to have higher brightness of coherent IR sources and stronger Raman signals through the coherent Raman scattering5,14, and some approaches have been made towards IR/Raman dual-modal spectral acquisition with a single pulsed laser15,16. However, these techniques neither have capability of simultaneous acquisition of complementary IR/Raman spectra nor broadband and high-resolution spectral acquisition covering the molecular fingerprint region (800–1800 cm−1), where the richest vibrational modes exist.
Here, we propose and demonstrate a simple yet powerful technique, called complementary vibrational spectroscopy (CVS), that allows us to simultaneously measure broadband IR and Raman spectra covering the fingerprint region at the same position. CVS is dual-modal Fourier-transform spectroscopy (FTS) enabled by an ultrashort near-infrared (NIR) pulsed laser and a Michelson interferometer. The IR spectroscopy is implemented as FT-IR with a coherent mid-infrared (MIR) pulsed source generated via intra-pulse difference-frequency generation (IDFG) from the NIR pulses17–19, while the Raman spectroscopy as Fourier-transform coherent anti-Stokes Raman scattering spectroscopy (FT-CARS) with the same NIR pulses20. The former uses a second-order and the latter a third-order nonlinear phenomena, respectively. The system is simple and robust because it shares a single laser source and an interferometer. Note that our proposed method can be applied to advanced FTS techniques such as dual-comb spectroscopy21–23, empowering the emerging technique further in respect to data acquisition rate, spectral resolution, and frequency accuracy.
Results
Principle of CVS
The schematic representation of the system is shown in Fig. 1a. In CVS, both FT-IR and FT-CARS are implemented in a single FTS system that consists of a Michelson interferometer with a NIR femtosecond laser (10-fs Ti:Sapphire mode-locked laser at a repetition rate of 75 MHz in this study) as a light source. The pulses emitted from the laser are coupled in the interferometer and double NIR pulses are created from each pulse with an optical path length difference (OPD) set by the delay line in the interferometer. The NIR double pulses are focused onto a χ(2) nonlinear crystal (GaSe crystal in this case) and a portion of the NIR pulses are converted to MIR pulses through the IDFG process. The generated MIR and undepleted NIR pulses collinearly irradiate the sample. The MIR pulses are absorbed, while the NIR pulses are inelastically (Raman) scattered by molecules. The MIR and NIR pulses are spatially separated by a dichroic mirror after passing through the sample and are simultaneously detected by a HgCdTe (MCT) photodetector and a Si avalanche photodetector (APD), respectively. Here, the NIR pulses are optically filtered before the detector so that only the blue-shifted scattered photons reach the APD. The detected signals are A/D-converted by a digitizer and the digitized interferograms are Fourier-transformed. The full schematic of the CVS is depicted in Supplementary Fig. 1.
The working principle of CVS is shown in Fig. 1b. In CVS-IR process, the MIR double pulses are modulated by IR-active molecular absorptions and their optical interference is detected by the MIR detector. Since the delay between the first and second MIR pulses is determined by that of the NIR pulses, the MIR absorption interferogram is measured as a function of the OPD between the NIR double pulses. Fourier-transforming the IR interferogram shows a broadband IR spectrum. On the other hand, in CVS-Raman process, the first NIR pulse excites the molecular vibrations and the second NIR pulse probes them and generates blue-shifted photons via anti-Stokes Raman scattering. By scanning the OPD between the NIR pulses, optical frequency of the second NIR pulse can be shifted by the refractive index modulations caused by the Raman active molecular vibrations induced by the impulsive stimulated Raman scattering process. A CARS interferogram that represents the molecular vibrations is obtained as an intensity modulation of the blue-shifted part of the second NIR pulses, which can be separated out by the optical short-pass filter. Finally, a broadband Raman spectrum is obtained by Fourier-transforming the CARS interferogram.
Characterization of NIR and MIR pulses
We first characterize the NIR and MIR pulses. The spectrum of our 10-fs Ti:Sapphire laser spans over 10,870–14,490 cm−1 (690–920 nm) at the center wavelength of 12,500 cm−1 (800 nm), and its pulse duration is evaluated by autocorrelation measurement as 11.9 fs at the sample position. This ultrashort NIR pulses with the broadband spectrum spanning more than 3400 cm−1 allows us to measure broadband FT-CARS spectrum covering the molecular fingerprint region (800–1800 cm−1) and C–H stretching region (2800–3300 cm−1)24. The spectrum of the MIR pulses generated by the IDFG process in a 30-µm GaSe crystal is measured by a homemade FT-IR spectrometer and it spans from 790 to 1800 cm−1, which covers the fingerprint region. In this study, the lowest wavenumber of the IR spectrum is limited by the detection range of the MCT detector and the highest wavenumber is possibly limited by the phase-matching condition of the IDFG process. Note that the IR spectral region can be shifted by changing the angle of the crystal and also expanded by changing the crystal and/or laser system.
Complementary vibrational spectroscopy
As a proof of concept demonstration, we measure complementary vibrational spectra of liquid toluene. Figure 2a shows sequential CVS interferograms of toluene, where the IR and CARS interferograms are simultaneously detected. They show the synchronized bursts at zero-path-difference (ZPD) of the interferometer. The OPD is scanned over 2 cm at a rate of 0.8 Hz. Figure 2b shows 15-times coherently averaged IR/CARS interferograms, which clearly show signature of molecular vibrations. In the configuration where the MIR pulses are generated after the NIR interferometer, the raw IR temporal waveform contains components other than the desired IR interferogram. A detailed retrieval procedure of the IR interferogram is described in Supplementary Note 4 and Supplementary Fig. 3. The double-sided IR interferogram is apodized and Fourier-transformed, whereas the single-sided CARS interferogram is apodized and Fourier-transformed by omitting the center-burst caused by the non-resonant four-wave mixing process at ZPD.
The complementary vibrational spectra Fourier-transformed from the interferograms shown in Fig. 2 are displayed in Fig. 3 with the reference spectra individually measured by conventional spectrometers. The upper panel shows the CVS-IR spectrum together with the reference spectrum measured by a standard FT-IR spectrometer (FT/IR-6800, JASCO). The CVS-IR transmittance spectrum of toluene agrees well with the reference spectrum and clearly displays the vibrational modes of, for example, C–H bending at 896, 1179, and 1495 cm−1, ring stretching at 1030, 1082, and 1605 cm−1, CH3 rocking at 1042 cm−1, C–CH3 stretching at 1211 cm−1 and CH3 deformation at 1379 and 1462 cm−1 (ref. 25–28) with a triangular-apodized spectral resolution of 3.5 cm−1. The lower panel shows the CVS-Raman spectrum and the reference spectrum measured by a standard spontaneous Raman spectrometer (inVia, Renishaw). The CVS-Raman spectrum at the apodized spectral resolution of 5.5 cm−1 clearly shows the vibrational modes of ring stretching at 1003 cm−1 and 1030 cm−1, C–CH3 stretching at 1211 cm−1 and CH3 deformation at 1380 cm−125,27,28, which also agrees well with the reference spectrum. Note that the wavenumbers of the CVS spectra are calibrated with the same interferometer, so that we can compare the spectra in a precise manner. The sensitivity evaluation of the CVS is described in Supplementary Note 5 and Supplementary Figs. 4 and 5.
To show the applicability of this system for other samples, we measure the complementary spectra of three different kinds of liquid samples: benzene, chloroform and a 4:1 mixture of benzene and dimethyl sulfoxide (DMSO) (Fig. 4). Here, we show the CVS-Raman spectra up to around 3000 cm−1, showing its ultra-broadband measurement capability. This capability of measuring ultra-broadband spectra with high spectral resolution is a unique feature given by using Fourier-transform spectroscopy technique. Detailed assignment of these spectra are discussed in Supplementary Note 2. The small spikes at 500–550, 750, 1250, and 2500 cm−1 in the CVS-Raman spectra are attributed to the instrumental noise, which is discussed in Supplementary Note 2 and Supplementary Fig. 2.
Discussion
Our CVS technique can be improved further. The measurable spectral span of the IR absorption spectrum (limited to 1000 cm−1 in this study) can be largely expanded by using other nonlinear crystals or light sources29–32, and/or by implementing the EO-sampling technique for detecting the IR interferograms19,30,33. In addition, since the CVS is based on FTS, it can be more rapid, robust and compact by using dual-comb FTS21–23 or ultra-rapid-scan FTS24,34–36. The higher scan rate allows us to measure dynamically changing complex phenomena or hyper-spectral multi-modal wide-area images. Furthermore, we could use the second harmonic pulses generated from the nonlinear crystal, together with the IDFG pulses for adding other spectroscopic modalities in CVS.
The CVS have the potential to provide means of chemical analysis. The simultaneous measurement of IR and Raman broadband spectra at the same spot can be essential for studying dynamic change in structure of complex molecules11. In addition, the highly accurate and common frequency axis for IR and Raman spectra would allow us to have fundamental study on molecules via precise analysis on relative line position or peak intensity between the IR and Raman spectra, which is essential for, for example, determination of molecular symmetry37,38 or two-dimensional correlation spectroscopy39,40.
Methods
NIR interferometer
A detailed schematic of the CVS is shown in Supplementary Fig. 1. The system is based on dual-modal FTS equipped with an ultrashort NIR pulsed laser, a Michelson interferometer, a nonlinear crystal for MIR pulse generation and photodetectors. The NIR pulsed laser (Ti:Sapphire Kerr-lens mode-locked laser) generates ultrashort pulses centered at 800 nm with a pulse duration of 10 fs at a repetition rate of 75 MHz (Synergy Pro, Spectra-Physics). The NIR pulses are coupled into the Michelson interferometer after passing through a half-wave plate to generate double NIR pulses with an OPD between the first and second pulses, which is scanned by a motorized stage in the interferometer. The OPD is measured at interferometric precision by continuous-wave (CW) interferograms of a HeNe laser that monitor the motion of the scan mirror. The measured OPD is used for phase error correction of the interferograms. By using a polarization beamsplitter and quarter-wave plates to construct the Michelson interferometer, the double pulses generated from the Michelson interferometer are orthogonally polarized to each other. A long-pass filter at a cutoff wavelength of 700 nm (FELH0700, Thorlabs) and a chirped mirror pair (DCMP175, Thorlabs) tailors the NIR pulses spectrally and temporally. After the long-pass filter the spectral range of the NIR pulses spans from 10,870 to 14,340 cm−1.
MIR pulse generation
The NIR double pulses are focused onto the 30-µm GaSe crystal (EKSMA OPTICS) using an off-axis parabolic mirror (OAPM) with a focal length of 25.4 mm for generating MIR pulses. The polarization of the focused pulses is adjusted with a half-wave plate. The pulse energy of the NIR pulse is 2–5 nJ, and that of the generated MIR pulse is tens of fJ. The remaining NIR pulses and generated MIR pulses are collimated by another OAPM with a focal length of 25.4 mm. Since the NIR pulses are slightly chirped by passing through the GaSe crystal, it is compensated by another chirped mirror pairs by separating the NIR and MIR pulses with a NIR/MIR dichroic mirror. The NIR and MIR pulses are combined again by another dichroic mirror. Here, to avoid spurious nonlinear effects at the sample, we intentionally have the NIR and MIR pulses separated in time. A pulse duration of the NIR pulses at the sample position is 11.9 fs evaluated by fringe-resolved autocorrelation measurement.
Irradiation onto the sample
The NIR and MIR pulses are focused onto the sample by an OAPM with a focal length of 15 mm. The NIR pulse energies irradiated onto the sample are 1.1 nJ and 2.4 nJ for the first and second pulses, respectively. The liquid samples are contained in a cuvette made of 3-mm thick KBr windows, which is transparent in a wide spectral range covering both NIR and MIR. A teflon spacer with a thickness of 50 µm is inserted between the KBr windows for adjusting the sample thickness.
Acquisition of the interferograms
The focused light onto the sample is collected and collimated by another OAPM with a focal length of 15 mm. After the collimation, the NIR and MIR pulses are spatially separated by a NIR/MIR dichroic mirror. The MIR pulses are detected by a N2-cooled MCT detector (KLD-0.5-J1-3/11, Kolmar Technologies) after passing through a linear polarizer, while the optically filtered NIR pulses are detected by an APD (APD410A2/M, Thorlabs), respectively. To detect the anti-Stokes scattering only, two short-pass filters (FESH0700, Thorlabs) are inserted in front of the APD. The detectors’ signals are low-pass-filtered and A/D-converted by a digitizer (ATS9440, AlazarTech). The CW interferogram of the HeNe laser is also digitized simultaneously. A part of the system where the MIR pulses travel through is enclosed by a box and purged with N2 gas in order to suppress undesired absorptions of the ambient gases, especially H2O vapor.
Data processing
The sequentially measured interferograms are segmented into single interferograms and coherently averaged. The reference CW interferogram of HeNe laser is used for resampling the digitized interferograms. The MIR interferogram is processed as double-sided, while the CARS interferogram as single-sided. The MIR interferogram is retrieved numerically (See Supplementary Note 4 and Supplementary Fig. 3). A strong peak that appears at ZPD in the CARS interferogram caused by the non-resonant background is omitted from the Fourier-transform window. Both the interferograms are apodized by triangular function and Fourier-transformed. To show the IR transmittance, we measure spectra with and without the sample.
Supplementary information
Acknowledgements
We thank Makoto Kuwata-Gonokami, Junji Yumoto, Kuniaki Konishi, and Yusuke Morita for allowing us to use their equipment and giving helpful suggestions. V.R.B. thanks the financial support by JSPS. This work was financially supported by JST PRESTO (JPMJPR17G2), JSPS KAKENHI (17H04852, 17K19071, 17F17028).
Author contributions
T.I. conceived the idea of this work. K.H. and T.I. designed the CVS system. V.R.B. calculated the IDFG generation, built the N2-purge system and the crystal mount. A.K. measured the reference IR/Raman spectra and made the sample cuvette. K.H. conducted the CVS experiment and analyzed the data. T.I. supervised the work. K.H. and T.I. wrote the manuscript and all authors contributed to improvement of the manuscript.
Code availability
The Matlab code used for analyzing the data of this study are available from the corresponding author upon reasonable request.
Data availability
The data that support the findings of this study are available from the corresponding author upon reasonable request.
Competing interests
The authors declare no competing interests.
Footnotes
Peer review information Nature Communications thanks Frans Harren and other, anonymous, reviewers for their contributions to the peer review of this work.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Supplementary information is available for this paper at 10.1038/s41467-019-12442-9.
References
- 1.Larkin P. Infrared and Raman Spectroscopy: Principles and Spectral Interpretation. Waltham: Elsevier; 2011. [Google Scholar]
- 2.Schrader B. Infrared and Raman spectroscopy: Methods and Applications. Weinheim: VCH; 1995. [Google Scholar]
- 3.Griffiths PR, De Haseth JA. Fourier Transform Infrared Spectrometry. Hoboken: John Wiley & Sons; 2007. [Google Scholar]
- 4.Ferraro JR, Nakamoto K. Introductory Raman spectroscopy. San Diego: Academic Press; 1994. [Google Scholar]
- 5.Cheng, J. X. & Xie, X. S. Coherent Raman Scattering Microscopy. (CRC Press, Boca Raton, 2012).
- 6.Herzberg G. Molecular Spectra and Molecular Structure: Infrared and Raman of Polyatomic Molecules. Princeton: D. Van Nostrand Company; 1945. [Google Scholar]
- 7.Angus WR, et al. Structure of benzene. part VII. Discussion of coincidental frequencies in the infra-red and Raman spectra of benzene and hexadeuterobenzene. J. Chem. Soc. 1936;217:966–971. doi: 10.1039/jr9360000966. [DOI] [Google Scholar]
- 8.Tinnemans SJ, et al. Combining operando techniques in one spectroscopic-reaction cell: New opportunities for elucidating the active site and related reaction mechanism in catalysis. Catal. Today. 2006;113:3–15. doi: 10.1016/j.cattod.2005.11.076. [DOI] [Google Scholar]
- 9.Brandhorst M, et al. Catalytic oxidation of methanol on Mo/Al2O3 catalyst: An EPR and Raman/infrared operando spectroscopies study. Catal. Today. 2006;113:34–39. doi: 10.1016/j.cattod.2005.11.008. [DOI] [Google Scholar]
- 10.Li G, Hu D, Xia G, White JM, Zhang C. High throughput operando studies using Fourier transform infrared imaging and Raman spectroscopy. Rev. Sci. Instrum. 2008;79:074101. doi: 10.1063/1.2949389. [DOI] [PubMed] [Google Scholar]
- 11.Grabow K, Bentrup U. Homogeneous Catalytic processes monitored by combined in situ ATR-IR, UV–Vis, and Raman spectroscopy. ACS Catal. 2014;4:2153–2164. doi: 10.1021/cs500363n. [DOI] [Google Scholar]
- 12.Reffner JA, Wilks DK, Schreiber KC, Burch RV. A new approach to infrared microspectroscopy: adding FT-IR to a light microscope. Microsc. Microanal. 2002;8:1526–1527. doi: 10.1017/S1431927602104259. [DOI] [Google Scholar]
- 13.Crocombe RA. Portable spectroscopy. Appl. Spectrosc. 2018;72:1701–1751. doi: 10.1177/0003702818809719. [DOI] [PubMed] [Google Scholar]
- 14.Boyd, R. W. Nonlinear Optics. (Academic Press, San Diego, 2008).
- 15.Hanninen AM, Prince RC, Potma EO. Triple modal coherent nonlinear imaging with vibrational contrast. IEEE J. Sel. Top. Quantum Electron. 2019;25:1–11. doi: 10.1109/JSTQE.2018.2827780. [DOI] [Google Scholar]
- 16.Müller N, Brückner L, Motzkus M. Coherent Raman and mid-IR microscopy using shaped pulses in a single-beam setup. APL Photonics. 2018;3:092406. doi: 10.1063/1.5030062. [DOI] [Google Scholar]
- 17.Joffre M, Bonvalet A, Migus A, Martin J-L. Femtosecond diffracting Fourier-transform infrared interferometer. Opt. Lett. 1996;21:964–966. doi: 10.1364/OL.21.000964. [DOI] [PubMed] [Google Scholar]
- 18.Liu K, Xu J, Zhang X-C. GaSe crystals for broadband terahertz wave detection. Appl. Phys. Lett. 2004;85:863–865. doi: 10.1063/1.1779959. [DOI] [Google Scholar]
- 19.Kübler C, Huber R, Tübel S, Leitenstorfer A. Ultrabroadband detection of multi-terahertz field transients with GaSe electro-optic sensors: approaching the near infrared. Appl. Phys. Lett. 2004;85:3360–3362. doi: 10.1063/1.1808232. [DOI] [Google Scholar]
- 20.Ogilvie JP, Beaurepaire E, Alexandrou A, Joffre M. Fourier-transform coherent anti-Stokes Raman scattering microscopy. Opt. Lett. 2006;31:480–482. doi: 10.1364/OL.31.000480. [DOI] [PubMed] [Google Scholar]
- 21.Coddington I, Newbury NR, Swann W. Dual-comb spectroscopy. Optica. 2016;3:414–426. doi: 10.1364/OPTICA.3.000414. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Keilmann F, Gohle C, Holzwarth R. Time-domain mid-infrared frequency-comb spectrometer. Opt. Lett. 2004;29:1542–1544. doi: 10.1364/OL.29.001542. [DOI] [PubMed] [Google Scholar]
- 23.Ideguchi T, et al. Coherent Raman spectro-imaging with laser frequency combs. Nature. 2013;502:355–358. doi: 10.1038/nature12607. [DOI] [PubMed] [Google Scholar]
- 24.Hashimoto K, Omachi J, Ideguchi T. Ultra-broadband rapid-scan Fourier-transform CARS spectroscopy with sub-10-fs optical pulses. Opt. Express. 2018;26:14307–14314. doi: 10.1364/OE.26.014307. [DOI] [PubMed] [Google Scholar]
- 25.Wilmshurst JK, Bernstein HJ. The infrared and Raman spectra of toluene, toluene-α-d3, m-xylene, and m-xylene-αα’-d6. Can. J. Chem. 1957;35:911–925. doi: 10.1139/v57-123. [DOI] [Google Scholar]
- 26.Coates, J. Interpretation of infrared spectra, a practical approach. Encyclopedia of Analytical Chemistry 10815–10837 (2000).
- 27.Tasumi M, Urano T, Nakata M. Some thoughts on the vibrational modes of toluene as a typical monosubstituted benzene. J. Mol. Struct. 1986;146:383–396. doi: 10.1016/0022-2860(86)80306-4. [DOI] [Google Scholar]
- 28.Gardner AM, Wright TG. Consistent assignment of the vibrations of monosubstituted benzenes. J. Chem. Phys. 2011;135:114305. doi: 10.1063/1.3638266. [DOI] [PubMed] [Google Scholar]
- 29.Zhang J, et al. Multi-mW, few-cycle mid-infrared continuum spanning from 500 to 2250 cm−1. Light.: Sci. Appl. 2018;7:17180. doi: 10.1038/lsa.2017.180. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Kowligy AS, et al. Infrared electric field sampled frequency comb spectroscopy. Sci. Adv. 2019;5:eaaw8794. doi: 10.1126/sciadv.aaw8794. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Brückner L, Müller N, Motzkus M. Flexible and broadly tunable infrared light source based on shaped sub-10-fs pulses for a multimodal microscopy setup. Opt. Lett. 2018;43:2054–2057. doi: 10.1364/OL.43.002054. [DOI] [PubMed] [Google Scholar]
- 32.Vasilyev S, et al. Super-octave longwave mid-infrared coherent transients produced by optical rectification of few-cycle 2.5-μm pulses. Optica. 2019;6:111–114. doi: 10.1364/OPTICA.6.000111. [DOI] [Google Scholar]
- 33.Pupeza I, et al. High-power sub-two-cycle mid-infrared pulses at 100 MHz repetition rate. Nat. Photonics. 2015;9:721–724. doi: 10.1038/nphoton.2015.179. [DOI] [Google Scholar]
- 34.Griffiths PR, Hirsche BL, Manning CJ. Ultra-rapid-scanning Fourier transform infrared spectrometry. Vib. Spectrosc. 1999;19:165–176. doi: 10.1016/S0924-2031(99)00006-5. [DOI] [Google Scholar]
- 35.Süss B, Ringleb F, Heberle J. New ultrarapid-scanning interferometer for FT-IR spectroscopy with microsecond time-resolution. Rev. Sci. Instrum. 2016;87:063113. doi: 10.1063/1.4953658. [DOI] [PubMed] [Google Scholar]
- 36.Hashimoto K, Ideguchi T. Phase-controlled Fourier-transform spectroscopy. Nature. Communications. 2018;9:4448. doi: 10.1038/s41467-018-06956-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Ziffer H, Levin IW. Determining centrosymmetric dimers by infrared and Raman spectroscopy. J. Org. Chem. 1969;34:4056–4060. doi: 10.1021/jo01264a065. [DOI] [PubMed] [Google Scholar]
- 38.Barón M, Bain AD, Conde RS. A dynamic look on molecular symmetry. Can. J. Chem. 2017;95:736–743. doi: 10.1139/cjc-2016-0657. [DOI] [Google Scholar]
- 39.Muik B, Lendl B, Molina-Diaz A, Valcarcel M, Ayora-Canada MJ. Two-dimensional correlation spectroscopy and multivariate curve resolution for the study of lipid oxidation in edible oils monitored by FTIR and FT-Raman. Anal. Chim. Acta. 2007;593:54–67. doi: 10.1016/j.aca.2007.04.050. [DOI] [PubMed] [Google Scholar]
- 40.Park Y, Noda I, Jung YM. Two-dimensional correlation spectroscopy in polymer study. Front. Chem. 2015;3:14. doi: 10.3389/fchem.2015.00014. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The Matlab code used for analyzing the data of this study are available from the corresponding author upon reasonable request.
The data that support the findings of this study are available from the corresponding author upon reasonable request.