. 2019 Sep 19;19(18):4046. doi: 10.3390/s19184046

Geometric Aberration Theory of Offner Imaging Spectrometers

Meihong Zhao 1,2, Yanxiu Jiang 1, Shuo Yang 1, Wenhao Li 1,*
PMCID: PMC6767683  PMID: 31546865

Abstract

A third-order aberration theory has been developed for the Offner imaging spectrometer comprising an extended source; two concave mirrors; a convex diffraction grating; and an image plane. Analytic formulas of the spot diagram are derived for tracing rays through the system based on Fermat’s principle. The proposed theory can be used to discuss in detail individual aberrations of the system such as coma, spherical aberration and astigmatism, and distortion together with the focal conditions. It has been critically evaluated as well in a comparison with exact ray tracing constructed using the commercial software ZEMAX. In regard to the analytic formulas, the results show a high degree of practicality.

Keywords: aberration theory, Offner imaging spectrometer, convex grating, spot diagrams

1. Introduction

An imaging spectrometer can provide a simultaneous collection of spatial and spectral information of targets with high resolution [1]. Currently, spectrometers have become an indispensable part of many fields including satellite remote sensing, space exploration, security, environment assessment, resource detection, agriculture, medicine, manufacturing, oceanography, and ecology [2,3,4,5,6].

The recent trend in imaging spectrometers is toward a simple set-up and a very compact configuration with high optical performance over the whole spectral range of the system [7]. This can be observed in the Offner imaging spectrometer with a concentric structure, using spherical optics. This spectrometer is obtained by replacing the convex secondary mirror of the Offner imaging system with a reflective convex diffraction grating [8]. It provides a high signal-to-noise ratio and small spot sizes together with low spatial and spectral distortions [7,8,9,10,11]. Because diffraction occurs at the grating, the perfect symmetry of the concentric configuration is altered, thereby increasing for example the coma and astigmatism. Although good optical performance is maintained with the rapid development of imaging spectrometers; more improvements need to be achieved to meet the dual demands for higher spatial and spectral resolution.

There have been various attempts to optimize and design an aberration-correct Offner imaging spectrometer. In 1999, Chrisp split the concave mirror into two concentric mirrors of different radii, increasing the degrees-of-freedom of the system designs [12]. By changing the off-axis parameters, tilting or decentering some elements, and making appropriate adjustments to the radii of the two spherical mirrors, the optical quality of the system was optimized. In 2001, Xiang and Mikes proposed an aberration-corrected spectrometer that included a convex diffraction grating having a number of nonparallel lines [13]. They believed the curves of the convex grating provided the correction for field aberrations. However, forming such a convex grating is difficult with the existing technology and theory. In 2006, Prieto-Blanco and coworkers presented an approach based on the calculation of both the meridional and the sagittal images of an off-axis object point [5]. Making the meridional and sagittal curves tangent to each other for a given wavelength results in a decrease in astigmatism. In 2007, Robert analyzed the out-of-plane dispersion in an Offner spectrometer. When the dispersion is perpendicular to the meridional plane, better performance is obtained for the system with a short entrance slit [14]. In 2014, Prieto-Blanco and coworkers proposed a Wynne-Offner layout consisting of a concave mirror and a concentric meniscus lens that included a diffraction grating at the center of one of its surfaces [15,16,17,18]. All the above methods have described the effect of aberrations such as astigmatism on the optical quality of the Offner spectrometer and how to optimize the system. However, these methods are relatively singular-use solutions and are not widely used in developing a system for different requirements [19,20].

In this paper, we propose a third-order geometric aberration theory of the Offner imaging spectrometer to provide an alternative aberration-correction method. This method is an extension and new application of Namioka’s theories [21,22,23,24,25,26]. Namioka and his team have shown aberration theories based on the light path function for a single grating or a double-element system that can correctly describe the individual aberrations and can be used to design an advanced optical system. Taking an extended source into consideration, analytic formulas of the spot diagram and the individual aberrations are derived for tracing rays through the system based on Fermat’s principle and Namioka’s theories. With these formulas, aberrations including coma, aberration, astigmatism, and distortion of the three-concentric-element (Offner) configuration are discussed in detail together with focal conditions. Finally, the theory is critically evaluated in a comparison with exact ray tracing constructed using the commercial software ZEMAX (Zemax software development company, bellevue, WS, USA). The results indicate a high degree of validity of the analytic formulas.

2. Three-Concentric-Element (Offner) Optical System

We consider an Offner optical system that comprises a planar light source S, two concave mirrors M1 and M2, a convex diffraction grating G, and an image plane Σ (Figure 1). In this system, the elements are arranged in such a way that the normal axes to S at A0, to M1 at O1, to G at O, and to M2 at O2 lie in a common plane called the meridional plane. The incident principal ray A0O1 is reflected by M1 toward O, and the reflected principal ray O1O of wavelength λ in the m1th order is diffracted by G toward O2. The diffracted principal ray OO2 is then further reflected by M2. This reflected principal ray of λ meets Σ at a point B0, which lies in the meridional plane as well. Here we assume that the principal ray of wavelength λ is designed to end up in the center of the image plane Σ; and we assume that the image plane Σ is perpendicular both to the reflected principal ray O2B0 and to the meridional plane as well. The distances A0O1, O1O, OO2, and O2B0 are denoted by r1, r, r′, and r2, respectively.

Figure 1.

Figure 1

Schematic diagram of an Offner configuration and its coordinate systems.

For convenience, we introduce five rectangular coordinate systems attached to S, M1, G, M2, and Σ (Figure 1). The origins are at A0, O1, O, O2, and B0, the XS, x1, x, x2, and X axes are the normal axes of the respective elements, and the YS, y1, y, y2, and Y axes lie in the meridional plane. A general ray originating from a source point A in S is reflected at a point Q1 on the surface of M1. The reflected ray meets G at a point P on the nth groove of G, and the diffracted ray of wavelength λ in m-th order meets M2 at a point Q2. The outgoing ray of wavelength λ from Q2 intersects Σ at a point B, forming a spot in the image plane Σ. We designate the coordinates of A, Q1, P, Q2, and B by (0, s, z), (ξ1, ω1, l1), (ξ, ω, l), (ξ2, ω2, l2), and (0, Y, Z) in the XSYSZS, x1y1z1, xyz, x2y2z2, and XYZ systems, respectively, and those of A and B as well by (x1, y1, z1) and (x2, y2, z2) in the x1y1z1 and x2y2z2 system, separately. Here, x1 and y1 are expressed as

x1=r1cosθ1+ssinθ1,    y1=r1sinθ1scosθ1. (1)

We assume as well that the zeroth groove of G passes through O, and that the groove number n is positive or negative according to whether the n-th groove passes through the y-axis on its positive or negative side. The groove number n of G expressed in a power series of ω and l is given as [26]:

nλ0=n10ω+12(n20ω2+n02l2+n30ω3+n12ωl2)+18(n40ω4+2n22ω2l2+n04l4)+, (2)

where λ0 is the recording wavelength of G.

In this system shown in Figure 1, both the concave mirrors M1 and M2 and the convex grating G are spherical in shape. The corresponding mathematical expression of the surface figure of Mi (or G) is given by

(ξiRi)2+ωi2+li2=Ri2, (3)

where Ri (i = 1, 2 for M1 and M2, no suffix for G) is the radius of Mi or G. Equation (3) expanded as a power series of ωi and li is:

ξi=12Riωi2+12Rili2+18Ri3ωi4+14Ri3ωi2li2+18Ri3li4+O(ωi6Ri3). (4)

The angles of incidence θi and reflection/diffraction θi′ of the principal ray at the vertices of Mi (or G) are considered as positive or negative depending on whether the relevant principal ray lies in the first or fourth quadrant of the xiyizi coordinate system. The angles θ and θ′ are related through the grating equation,

σ(sinθ+sinθ)=mλ, (5)

where σ is the effective grating constant obtained by:

σ1/(n/ω)ω=l=0=λ0/n10, (6)

which can be referred to [26].

3. Ray-Tracing Formulas

First, we denote the distances AQ1, Q1P, PQ2, and Q2B by q1, p1, q2, and p2, respectively. According to Namioka’s theory, the application of Fermat’s principle to the light-path function for Mi,

FMi=qi+pi, (7)

yields the direction cosines (Li′, Mi′, Ni′) of the reflected ray pi in terms of the direction cosines (Li, Mi, Ni) of the incident ray qi and given system parameters:

Li=Li+τi,Mi=Miτi(ξiωi),Ni=Niτi(ξili). (8)

where all the quantities are defined in the xiyizi coordinate system. In Equation (8), we have

τi=2[Li+Mi(ξi/ωi)+Ni(ξi/li)]1+(ξi/ωi)2+(ξi/li)2, (9)

where i = 1, 2 for M1 and M2. Li, Mi, and Ni are obtained from the definition of the direction cosines of the incident ray qi.

The intersecting point P (ξ, ω, l) is determined by solving simultaneously the equation of the ray Q1P in the xyz coordinate system

ξξ¯1L=ωω¯1M=ll¯1N, (10)

and Equation (3) with i = 1. In Equation (10) (ξ¯1, ω¯11, l¯1) and (L, M, N) are the coordinates of the point Q1 and the direction cosines of the ray Q1P, which are both defined in the xyz coordinate system. They are obtained by applying proper coordinate transformations to (ξ1, ω1, l1) and (L1′, M1′, N1′).

Different from the above calculation for Mi, the application of Fermat’s principle to the light-path function for G,

F=p1+q2+nmλ, (11)

yields the direction cosines (L′, M′, N′) of the diffracted ray PQ2 in terms of the direction cosines (L, M, N) of the incident ray Q1P and given system parameters:

L=L+τ,M=M+mλ(nω)τ(ξω),N=N+mλ(nl)τ(ξl). (12)

where all the quantities are defined in the xyz coordinate system. In Equation (12), we have

τ=1ρ(ν+ν2κρ),ρ=1+(ξω)2+(ξl)2,ν=L+(M+mλnω)ξω+(N+mλnl)ξl,κ=2mλ(Mnω+Nnl)+(mλ)2[(nω)2+(nl)2]. (13)

The intersecting point Q2 (ξ2, ω2, l2) is determined by solving simultaneously Equation (3) with i = 2 and the equation of ray PQ2 in the x2y2z2 coordinate system,

ξ2ξ¯L2=ω2ω¯M2=l2l¯N2, (14)

where (ξ¯, ω¯, l¯) and (L2, M2, N2) are the coordinates of point P and the direction cosines of ray PQ2, both defined in the x2y2z2 coordinate system. They are obtained by applying proper coordinate transformations to (ξ, ω, l) and (L′, M′, N′).

The image plane Σ is expressed in the x2y2z2 coordinate system as

x2cosθ2+y2sinθ2=r2. (15)

Then, the intersection B of the reflected ray Q2B with the image plane Σ is determined by solving the equation of the ray Q2B in the x

2y2z2 coordinate system,

x2ξ2L2=y2ω2M2=z2l2N2, (16)

from which we obtain:

x2=ξ2+p2L2,    y2=ω2+p2M2,    z2=l2+p2N2. (17)

By applying proper coordinate transformations to B (x2, y2, z2), the ray-traced spot B (0, Y, Z) in the XYZ coordinate system is expressed as

Y=(r2sinθ2y2)secθ2,​​​​    Z=z2. (18)

All the above equations presented in this section provide a complete set of ray-tracing formulas.

4. Analytic Expression of Spot Diagrams and Aberrations

The imaging characteristics of the three-concentric-element optical system may be analyzed numerically using ray tracing. Although ray tracing provides accurate spot diagrams with comparative ease, it lacks the ability to give explicit analytical expressions for the focal condition and individual aberrations of the system under consideration. According to Namioka’s theory, we express the relationship between the coordinates of a source point and its image by expanding the ray-tracing formulas given in Section 3 into power series of ω1, l1, and the coordinates of A0 in the XSYSZS system. In this way—although laborious—a third-order aberration theory is developed for the system, which has a high degree of validity.

Taking the expansion of the coordinates of point P as an example, we determine its position in the xyz coordinate system by finding the intersection of ray Q1P with the grating blank surface. We express ω and l in a power series of ω1hl1izjsk (h + i + j + k ≤ 3) under assumptions of:

ω=h+i+j+k=13Ahijkω1hl1izjsk,    l=h+i+j+k=13Bhijkω1hl1izjsk. (19)

To determine ω and l, we derive first the direction cosines (L, M, N) as power series of ω1, l1, s and z by expanding their definitions:

L=ξξ¯1p1,    M=ωω¯1p1,    N=ll¯1p1, (20)

as

L=h+i+j+k=13(Hhijk)Lω1hl1izjsk,     M=h+i+j+k=13(Hhijk)Mω1hl1izjsk,     N=h+i+j+k=13(Hhijk)Nω1hl1izjsk. (21)

where coefficients (Hhijk)L, (Hhijk)M, and (Hhijk)N are functions of R1, r, θ1, and θ only.

Next, we adopt another approach to expand the direction cosines of the ray Q1P in terms of L1′, M1′, and N1′:

L=h+i+j+k=13(Hhijk)Lω1hl1izjsk,    M=h+i+j+k=13(Hhijk)Mω1hl1izjsk,    N=h+i+j+k=13(Hhijk)Nω1hl1izjsk. (22)

where the coefficients (Hhijk)L′, (Hhijk)M′, and (Hhijk)N′ are functions of R1, r1, θ1, and θ only. We obtain coefficients Ahijk and Bhijk by equating coefficients (Hhijk)L, (Hhijk)M, and (Hhijk)N of Equation (21) to the corresponding ones, (Hhijk)L′, (Hhijk)M′, and (Hhijk)N′ of Equation (22), which determines the coefficients Ahijk and Bhijk uniquely. Therefore, the coordinates of the intersecting point P in terms of ω1, l1, s and z are

ω=A1000ω1+A0001s+A2000ω12+A1001ω1s+A0002s2+A0200l12       +A0110l1z+A0020z2+A3000ω13+A2001ω12s+A1200ω1l12+A1110ω1l1z       +A1020ω1z2+A1002ω1s2+A0201l12s+A0111l1zs+A0021z2s+A0003s3, (23)
l=B0100l1+B0010z+B1100ω1l1+B1010ω1z+B0101l1s+B0011zs+B2100ω12l1     +B2010ω12z+B1101ω1l1s+B1011ω1zs+B0300l13+B0210l12z+B0120l1z2     +B0102l1s2+B0012zs2+B0030z3. (24)

Explicit expressions of Ahijk and Bhijk that are applicable to spherical mirror M1 are given in [24].

This expansion method for the coordinates of P is used as well to derive power series expressions of the coordinates of Q2 and those of B in the x2y2z2 and XYZ coordinate system, respectively. Then, the coordinates (0, Y, Z) of the ray-traced spot B formed in the image plane Σ, which are determined through Equations (15) to (18), are finally expressed as power series in ω1hl1izjsk,

Y=E1000ω1+E0001s+E2000ω12+E1001ω1s+E0002s2+E0200l12+E0110l1z       +E0020z2+E3000ω13+E2001ω12s+E100ω1l12+E1110ω1l1z+E1020ω1z2       +E1002ω1s2+E0201l12s+E0111l1zs+E0021z2s+E0003s3+OE(κ4), (25)
Z=F0100l1+F0010z+F1100ω1l1+F1010ω1z+F0101l1s+F0011zs+F2100ω12l1      +F2010ω12z+F1101ω1l1s+F1011ω1zs+F0300l13+F0210l12z+F0120l1z2      +F0102l1s2+F0012zs2+F0030z3+OF(κ4). (26)

These two equations are the spot-diagram formulas for the three-concentric optical system. κ′4 represent the aberration terms ω1hl1izjsk with h + i + j + k ≥ 4. OE and OF denote the higher-order terms in the aberration coefficients. The coefficients Ehijk and Fhijk are the aberration coefficients, and we express them in terms of Ahijk, Ahijk, Ahijk, Bhijk, Bhijk, and Bhijk in Appendix A and Appendix B. Here Ahijk, Ahijk, Bhijk, and Bhijk are defined as:

Ahijk=Ahijk|r1r,rr,θθ2,RR2,ε1ε,Bhijk=Bhijk|r1r,rr,θθ2,RR2,ε1ε, (27)
Ahijk=Ahijk|rr,rr2,θ20,R2,εε2,Bhijk=Bhijk|rr,rr2,θ20,R2,εε2, (28)

where r1r, for example, indicates replacement of r1 in Ahijk and Bhijk by r. In Equation (27), ε1 represents all the parameters with a subscript 1, except r1, in coefficients Ahijk and Bhijk, and ε stands for the corresponding parameters with no subscript in Ahijk and Bhijk.

5. Analysis of Focal Conditions and Aberrations

For demonstrating various aberrations curves in the next section and evaluating the spot-diagram formulas, we adopt a well-designed and optimized Offner imaging spectrometer as a model. The values of the specific parameters are listed in Table 1; here, the signs of the values are determined by the sign convention.

Table 1.

Parameters of the model Offner imaging spectrometer.

Parameter Value
Spectral range/nm 380–900
Radius of M1/mm 220
Radius of G/mm 112.2
Radius of M2/mm 216.85
Dimension of slit/mm2 0.025 × 12
Aperture of M1/mm2 65 × 65
Aperture of G/mm2 30 × 30
Constant of G/mm−1 0.01
Diffraction order of G −1
∠O1OO2 50.66°

5.1. Focal Conditions

When the first-order aberration coefficients E1000 and F0100 are made zero, a configuration with the appropriate instrument parameters is obtained. In such a configuration, the paraxial rays in the meridional or sagittal plane are brought into focus, greatly reducing the aberration of the system. The conditions E1000 = 0 and F0100 = 0 give the meridional and sagittal focal curves, respectively.

5.1.1. Meridional Focal Condition

The meridional focal condition E1000 = C1000 A1000 + C0001 = 0 is expressed as

2(F1)20(F)20=cos2θ1cos2θr2,      2(F)20(F2)20=cos2θcos2θ2r2, (29)

where (F)20 is the value of (F)20 at r′ = (r″)M, and ()20 is the value of (F2)20 at r2 = (r2′)M. The focal distances r′ = (r″)M and r2 = (r2′)M that satisfy Equation (29) are called the meridional focal distances of G and M2 respectively. (F1)20, (F)20, and (F2)20 are defined as

(F1)20=cos2θ12r1+cos2θ12r2cosθ1R1,(F)20=cos2θ2r+cos2θ2rcosθ+cosθR+(n)20Π,(F2)20=cos2θ22r+cos2θ22r22cosθ2R2 (30)

Hence, Equation (29) reduces to:

2(F1)20=cos2θ1rcos2θ1(r1)M,      (F)20=cos2θrcos2θ(r)M, (31)
(F)20=cos2θrcos2θ(r)M,      2(F2)20=cos2θ2rcos2θ2(r2)M, (32)

where the meridional focal conditions for M1, G, and M2 are expressed, separately. In Equation (31), (r1′)M is the meridional focal distance of M1, giving the object distance of G in the meridional plane as (r)M = r − (r1′)M. Similarly, the object distance of M2 in the meridional plane is obtained from Equation (32) as (r2)M = r′ − (r″)M. We then obtain the meridional focal distance of the Offner optical system by solving Equations (30) to (32).

5.1.2. Sagittal Focal Condition

We present the sagittal condition F0100 = D0010 + D0100 B0100 = 0 as

2(F1)02(F)02=1r2,      2(F)02(F2)02=1r2, (33)

where (F*)02 is the value of (F)02 at r′ = (r″)S, and (F2*)02 is the value of (F2)02 at r2 = (r2′)S. The focal distances r′ = (r″)S and r2 = (r2′)S that satisfy Equation (33), are called the sagittal focal distances of G and M2, respectively. (F1)02, (F)02, and (F2)02 are defined by

(F1)02=12r1+12r2cosθ1R1,(F)02=12r+12rcosθ+cosθR+(n)02Π,(F2)02=12r+12r22cosθ2R2. (34)

Similar to obtaining the meridional focus, we resolve Equation (33) into:

2(F1)02=1r1(r1)S,      (F)02=1r1(r)S, (35)
(F)02=1r1(r)S,      2(F2)02=1r1(r2)S, (36)

which represent the sagittal focal conditions for the three elements of the system. Likewise, we obtain the object distances of G and M2 in the sagittal plane in the form (r)S = r − (r1′)S and (r2)S = r′ − (r″)S. Here (r1′)S in Equation (35) is the sagittal focal distance of G. Therefore, the sagittal focal distance of the system is given by solving Equations (34) and (35).

For a real point source and a real image, the system shown in Figure 1 is capable of making the tangential and sagittal focal points to coincide, yielding non-astigmatic image when (r2′)M = (r2′)S is satisfied. Failure to meet the condition leads to the astigmatic aberration.

5.2. Aberration Analysis

Next, we introduce the polar coordinates

ω1=rpcosα,      l1=rpsinα, (37)

in the entrance pupil centered at the vertex O1 of M1.

5.2.1. Spherical Aberration

In the ray-tracing formulas, spherical aberration is described by

Ysph=E3000ω13+E1200ω1l12,      Zsph=F0300l13+F2100ω12l1, (38)

which can be changed into:

Ysph=rp(E3000cos2α+E1200sin2α)cosα,      Zsph=rp(E3000sin2α+E1200cos2α)sinα. (39)

The spherical aberration curves of the model optical system for the center wavelength (Figure 2) are more complicated than common circular patterns of a centered lens system.

Figure 2.

Figure 2

Spherical aberration curves for λ = 700 nm of the model optical system in the meridional focal plane: (a) with rp = 30, 25, 20, 15 mm and (b) with rp = 30 mm (each inset is an enlargement of a central portion of the curve).

The spherical aberration curve is a circle of rp3E3000 only when E3000 = E1200 = F0300 = F2100 is met. This condition is satisfied by an axially symmetric centered Offner optical system, which is the same as both a single mirror and a centered double-mirror system, yielding a concentric circular pattern for various values of rp.

5.2.2. Coma

The coma of the concentric Offner optical system under consideration is expressed by:

Ycoma=E2000ω12+E2010ω12z+E0210l12z+E1101ω1l1s,Zcoma=F1100ω1l1+F2001ω12s+F1110ω1l1z+F0201l12s (40)

Substitution of Equation (37) into Equation (40) yields:

a{2Ycomar2[E2000+s(E2001+E0201)]}2+b{2Zcomar2z(F2010+F0210)}22h{2Ycomar2[E2000+s(E2001+E0201)]}{2Zcomar2z(F2010+F0210)}=c2, (41)

where

a=(F1100+sF1101)2+z2(F2010F0210)2,b=[E2000+s(E2001E0201)]2+z2E11102,c=(F1100+sF1101)[E2000+s(E2001E0201)]z2E1110(F2010F0210),h=z{E1110(F1100+sF1101)+(F2010F0210)[E2000+s(E2001E0201)]}. (42)

With Equation (41) describing an ellipse, the model optical system produces elliptical patterns for different values of rp (Figure 3).

Figure 3.

Figure 3

Coma curves of the model optical system in the meridional focal plane, illustrating changes in pattern with rp, for settings (a) z = 6 mm and s = 0.0125 mm and (b) z = −6 mm and s = −0.0125 mm.

5.2.3. Astigmatism

Astigmatism is an image defect caused by two mutually perpendicular line images, one at (r2′)M and the other at (r2′)S. Astigmatism of the concentric Offner optical system is represented by:

Yast=E0001s+E0200l12+E0110l1z+E0020z2,Zast=F0100l1+F0010z. (43)

which transforms to

Yast=E0200[ZastF0100+(E01102E0200F0010F0100)z]2+(E0020E011024E0200)z2+E0001s. (44)

The astigmatic curves obtained from the model optical system appear as crescent-shaped patterns (Figure 4).

Figure 4.

Figure 4

Astigmatic curves of the model optical system in the meridional focal plane.

5.2.4. Distortion

Distortion is the deviation between the actual image height and the ideal image height of the chief ray originating from a source point (0, s, z) and passing through the vertex O1. In the ray-tracing formulas, the distortion is expressed as

Ydist=E0001s+E0020z2+E0002s2+E0021z2s+E0003s3,Zdist=F0010z+F0011zs+F0030z3+F0012zs2. (45)

which manifests as a barrel-like structure from the model optical system (Figure 5). Because of the use of a very narrow slit illuminant, the distortions are visible in the Z direction and not in the Y direction.

Figure 5.

Figure 5

Distortion patterns of the model optical system for a mesh-like source of dimension 25 μm (s) × 12 mm (z) with line separations of Δs = 6.25 μm and Δz = 3 mm.

6. Analysis of Diagram and Discussion

In Section 5, the evaluation of individual aberrations in the model optical system was presented using aberration curves, facilitating a better understanding of the imaging properties of the Offner system. More importantly, such an evaluation method helps to design and optimize the Offner system with an aberration-correction convex grating for different requirements.

However, before adopting the spot diagram formulas (25) and (26) in the design of an Offner system and its grating, the equations need to be critically evaluated in a comparison with exact ray-tracing. Here we compare spot diagrams computed from Equations (25) and (26) with those determined by the exact ray tracing using ZEMAX configured with a model Offner system equipped with a holographic convex grating.

Because of the large z value and a relatively small s value in the model system, a portion of the spot diagram was constructed for various fields where we set z = 0, 0.6, and 6 mm with s = 0 without loss of generality. All the diagrams in Figure 6 and Figure 7 were constructed by generating 20000 rays of wavelength 700 nm covering the whole field of view. Spot diagrams in (a) and (b) were constructed for the selected point source presented in Figure 6, using the spot-diagram formulas and by ray tracing using ZEMAX, respectively. Clearly, the spot diagrams in (a) and (b) are similar in shape, but there are some deviations in size and position—especially in the Z direction; see Figure 6c.

Figure 6.

Figure 6

Spot diagrams constructed for the model optical system at λ = 700 nm with s = 0 from: (a) formulas (25) and (26); (b) ray tracing using ZEMAX; and (c) contrasted by overlaying (a) with (b).

Figure 7.

Figure 7

ΔY - ΔZ plots constructed for the model optical system at λ = 700 nm. (a) deviations, ΔY - ΔZ, of individual spots in Figure 6a from the corresponding ideal image points (0,0,0), (0,0,0.6) and (0,0,6); (b) deviations, ΔY - ΔZ, of individual spots in Figure 6b from the corresponding ideal image points (0,0,0), (0,0,0.6) and (0,0,6)7. Conclusions.

The standard deviations σY and σZ of the spots in the Y and Z directions (Figure 6a,b) illustrate the similarity in spot shape. The difference between the standard deviations of corresponding individual spots is smaller than 0.4 μm in the Y direction and 0.65 μm in the Z direction. Nearly the same dispersion tendency is seen depending on the system aberrations for the spot diagrams generated by both the present theoretical model and the simulation model of ZEMAX.

Figure 7a shows the deviations of individual spots obtained by the spot-diagram formulas from the corresponding ideal image points (0,0,0), (0,0,0.6) and (0,0,6). Figure 7b shows the deviations of individual spots generated by ray tracing using ZEMAX from the corresponding ideal image points. The deviations between ideal image points and spots from a simulation model (such as the theoretical simulation model or the ZEMAX simulation model) depend on both the system aberrations and model errors. The root-mean-squares RMSΔY and RMSΔZ of the deviations are given in the respective diagrams. Here, distinctions between the individual corresponding spots in Figure 7a,b—both in size and position—are mainly determined by different model errors. However, the difference between RMSΔY and RMSΔZ of the spots in Figure 7a,b is smaller than 0.4 μm in the Y direction and 0.7 μm in the Z direction. Therefore, the present theoretical model is similarly as useful as the ZEMAX model in designing and optimizing the Offner optical system. Certainly, supplemented by the fourth- and higher-order aberration terms into the spot diagram formulas (25) and (26), more exact theoretical model may be developed.

7. Conclusions

In this paper, a more practical method is adopted, comparing the theoretical simulation model with the simulation model provided by the commercial software ZEMAX, and offers great practicality.

A third-order aberration geometric theory was developed for tracing rays through the Offner imaging spectrometer comprising an extended source, two concave mirrors, a convex diffraction grating, and an image plane based on Fermat’s principle. The proposed theory provides analytic formulas for individual aberrations and spot diagrams. Following on from Namioka’s work, aberrations were analyzed and certain aberration curves were illustrated for a corresponding model optical system. The validity of the theoretical model was evaluated in a comparison with a simulation model provided by the commercial software ZEMAX and that of an actual model optical system. The results indicate the proposed theoretical model has great utility and practicality.

Appendix A. Aberration Coefficients Chijk and Dhijk

C1000=A1000A1000+A0001cosθ, (A1)
C0001=A0001A1000cosθ1, (A2)
C2000=A2000A1000+A10002A2000+A1000A1001cosθ+A0002cos2θA0001sinθ           ×(12RcosθrA1000cosθ2r), (A3)
C1001=A1001A1000cosθ1+2A0001A1000A2000cosθ1+A0001A1001cosθ1cosθ           +A0001A1000sinθ1cosθr+A0001A0001cosθ1sinθcosθ2r, (A4)
C0200=A0020+A0200A1000+A0110B0100+A0200B01002A0001sinθ2R, (A5)
C0110=A0110A1000+A0110B0010+2A0200B0010B0100, (A6)
C0020=A0020A1000+A0200B00102A0001A1000sinθ12R1, (A7)
C0002=A0002A1000cos2θ1+A0001A1000(sinθ1cosθ1rsinθ12R1)+A00012A2000cos2θ1, (A8)
C3000=A3000A1000+2A1000A2000A2000+A10003A3000+A2000A1001cosθ           +A10002A2001cosθ+A1000A1002cos2θ+A0003cos3θ(12Rcosθr)           ×(r3A1000cosθcos2θ2secθ2r2+A1000A1001sinθ+2A0002sinθcosθ           +A0001sin2θr)A10002A0001sinθsinθ22R2r+A0001cos2θ2Rr+sinθcosθ2r           ×(A2000A0001+A10002A1001)r2A1000r3(cos2θ2secθ2+sinθcosθcos3θ2           ×sec2θ2tanθ2)+A1000sinθsinθ2cosθ2r2(A1000A00012r2cosθsecθ2r), (A9)
C1110=A1110A1000+2A0110A1000A2000+A1000A1110B0010+2A1000A1200B0010B0100        ​​​    +A0110B1010+2A0200B0100B1010+2A0200B0010B1100+A0110A1001cosθ           +A0111B0010cosθ+2A0201B0010B0100cosθ+A0110A0001sinθcosθ2r           B0010sinθr(2A0020A0110+2A0110B0100+A0001B0100sinθ2R2), (A10)
C2001=A2001A1000cosθ1+2A1000A1001A2000cosθ1+2A0001A2000A2000cosθ1           +3A0001A10002A3000cosθ1+A1001A1001cosθ1cosθ+2A0001A1000A2001           ×cosθ1cosθ+A0001A1001sinθ1cosθcosθr+A1001A1000sinθ1cosθr           +2A0001A1000A2000sinθ1cosθr+A0001A1000sinθ1sinθr(12Rcosθr)           (r2A0001cosθ1cosθcos2θ2secθ2r2+A0001A1001cosθ1sinθ)(12R           cosθr)2A0001A1000A0001cosθ1sinθsinθ2r×(12R2cosθ2r)           +A0001A1002cosθ1cos2θ+sinθcosθ2r(A1001A0001cosθ1+2A0001A1000           ×A0001cosθ1)+A0001A0001sinθ1cosθsinθcosθ2rrr2A0001cosθ1r3           ×cosθ2secθ2(cosθ2+sinθcosθ(2sinθ2+cos2θ2secθ2tanθ2)), (A11)
C1200=A1200A1000+A1000A1020+2A0200A1000A2000+A1000A1110B0100           +A1000A1200B01002+A0110B1100+2A0200B0100B1100+A0200A1001cosθ           +A0111B0100cosθ+A0201B01002cosθ+A0021cosθA00012Rr           ×(1A1000cosθcosθ22cos2θ)+sinθr(2A00202A0020B0100           +A0110B0100A0110B01002+A0200A0001cosθ2)sinθ2R(A1000A1001           +2A0002cosθ)A0001B01002sinθsinθ22R2r, (A12)
C1020=A1020A1000+2A1000A0020A2000+A1000A1200B00102+2A0200B0010B1010           +A0020A1001cosθ+A0201B00102cosθA0001A1001sinθ1cosθ2R1           A1000A0001A2000sinθ1R1A1001A1000sinθ12R1+A0001A1000cosθ1cosθ2R1r           sinθr(A0001A0001sinθ1cosθ22R1+A0110B00102A0020A0001cosθ2           +A0001B00102sinθ22R2), (A13)
C0021=A0021A1000cosθ1+2A0001A0020A2000cosθ1+A0001A1200B00102cosθ1           sinθ1cosθ1R1(A00012A2000+A0002A1000)A0001A10002R1r(12cos2θ1)           +2A0020A1000sinθ1r+2A0200B0010(r2sinθ1r2+B0011cosθ1), (A14)
C1002=A1002A1000cos2θ1+A00012A2001cos2θ1cosθ+3A00012A1000A3000cos2θ1           +A0002A1001cos2θ1cosθ+2A1000A0002A2000cos2θ1+2A0001A1001A2000cos2θ1           (12R1cosθ1r)×(r2A1000cosθ1cos2θsecθsecθ3r2+A1001A1000sinθ1)           r2A1000secθsecθ2r3(cos2θ+sinθ1cosθ1(2sinθcosθ+Γ2cos3θsecθ))           +(2A1000A0001A2000+A0001A1001cosθ+A0001A0001sinθcosθ2r)×(sinθ1cosθ1r           sinθ12R1)+2A00012A2000sinθ1cosθ1cosθr+sinθcosθ2r(A0002A0001cos2θ1           +A00012A1001)A00012A0001cos2θ1sinθsinθ2r×(12R2cosθ2r), (A15)
C0201=A0201A1000cosθ1+A0001A1020cosθ1+2A0001A0200A2000cos2θ1+A0001A1110           ×B0100cosθ1+A0110B0101cosθ1+A0001A1200B01002cosθ1+2A0200B0100B0101           ×cosθ1A1000sinθ1r2(A0110+A0001sinθ2R)+r2sinθ1r2(A0110+2A0200B0100)           +A0001cosθ12R(A0001cosθcosθ2rA1001sinθ), (A16)
C0111=A0111A1000cosθ1+2A0001A0110A2000cosθ1+A0001A1110B0010cosθ1+2A0001           ×A1200B0010B0100cosθ1+A1000sinθ1r(2A0020+A0110)+2r2A0200B0010sinθ1r2           +(A0110+2A0200B0100)(r2sinθ1r2+B0011cosθ1)+2A0200B0010B0101cosθ1, (A17)
C0003=A0003A1000cos3θ1+A00013A3000cos3θ1+2A0001A0002A2000cos3θ1           +A0001A1000cos2θ12R1rA0001A1000sin2θ1r×(12R1cosθ1r)           +(12R1cosθ1r)×(2sinθ1cosθ1(A0002A1000+A00012A2000)), (A18)
D0100=B0100B0100r2r, (A19)
D0010=B0010B0100, (A20)
D1100=B1100B0100+A1000B1010+A1000B0100B1100+B0011cosθ           +B0100B0101cosθr2sinθr2(1B0100), (A21)
D1010=B1010B0100+A1000B0010B1100+B0010B0101cosθ+r2B0010sinθr2, (A22)
D0101=B0101B0100cosθ1+A0001B1010cosθ1+A0001B0100B1100cosθ1+rB0100sinθ1r2, (A23)
D0011=B0011B0100cosθ1+A0001B0010B1100cosθ1rB0100sinθ1r2, (A24)
D2100=B2100B0100+A2000B1010+A2000B0100B1100+A1000B1100B1100+B0100B0102cos2θ           +A10002B0100B2100+B1100B0101cosθ+B0012cos2θ+A1000B1011cosθ           +A10002B2010+A1000B0100B1101cosθ+r2B1100sinθr2B0100B0101sinθ           ×(12Rcosθr)r2(1B1100)r2×(sin2θr+A1000sinθsinθ2r+cosθ2R)           +A1000sinθr(B1010B0100B1010+B0100B0101cosθ2), (A25)
D2010=B2010B0100+A2000B0010B1100+A1000B1010B1100+A10002B0010B2100           +B1010B0101cosθ+A1000B0010B1101cosθ+B0010B0102cos2θ           B0010B0101sinθ2R+r2B0010sinθr3(sinθ+A1000sinθ2)           +r2B0010cosθ2Rr2+r2B1010sinθr2B0010sinθr(A1000B1010B0101           ×cosθA1000B0101cosθ2), (A26)
D1101=B1101B0100cosθ1+A1000B0101B1100cosθ1+A0001B0100B1101cosθ1cosθ            +A1001B0100B1100cosθ1+A0001B1100B1100cosθ1+2A0001A1000B2010cosθ1            +2A0001A1000B0100B2100cosθ1+A0001B1011cosθ1cosθ+A1001B1010cosθ1            +B0101B0101cosθ1cosθ+B0100sinθ1r(B1010+B0101cosθ+rsinθr2)            +A0001sinθ1cosθr(B1010+B0100B1100)+(A1000B1100+B0101cosθ)            ×rsinθ1r2+A0001cosθ1sinθr(B1010B0100B1010+B0100B0101cosθ2)            +sinθr2(rr2sinθ1r2+r2B0101cosθ1r2A0001cosθ1sinθ2r(1B0100)), (A27)
  D1011=A0001B1010B1100cosθ1+A1001B0010B1100cosθ1+A1000B0011B1100cosθ1            +B1011B0100cosθ1+B0011B0101cosθ1cosθ+A0001B0010B1101cosθ1cosθ            +2A0001A1000B0010B2100cosθ1+sinθ1r(B1010B0100+A0001B0010B1100cosθ)            rsinθ1r2(A1000B1100+B0101cosθ)rr3(sinθ+cos2θsecθtanθ2)            ×B0100sinθ1A0001B0010cosθ1sinθr(B1010B0101cosθ2)+r2sinθr2            ×(rsinθ1r2+B0011cosθ1+A0001B0010cosθ1sinθ2r), (A28)
D0300=B0300B0100+B0100B0120+B01002B0210+B01003B0300+A0200B0100B1100           +A0200B101012R(B0100B0120sinθ+r2cosθr2(1B0100)), (A29)
D0300=B0300B0100+B0100B0120+B01002B0210+B01003B0300+A0200B0100B1100           +A0200B101012R(B0100B0120sinθ+r2cosθr2(1B0100)), (A30)
D0120=B0120B0100+A0110B0010B1100+A0020B0100B1100+B00102B0210+A0020B1010           +3B0100B00102B0300sinθ12R1(B0101B0100+A0001(B1010+B0100B1100))           rB0100cosθ12R1r2, (A31)
D0102=A00012B2010cos2θ1+A00012B0100B2100cos2θ1+A0002B1010cos2θ1           +A0001B0101B1100cos2θ112R1(B0100B2100sinθ1rB0100cosθ1r2)           +B0102B0100+A0002B0100B1100cos2θ1+rA0001B1100sinθ1cosθ1r2           +(sinθ1cosθ1rsinθ12R1)(A0001B1010+A0001B0100B1100), (A32)
D0030=B0100(B0030rcosθ12R1r2)+B0010(B00102B0300+A0020B1100)A0001B0010B1100sinθ12R1, (A33)
D0012=B0012B0100cos2θ1+A0002B0010B1100cos2θ1+A00012B0010B2100cos2θ1+A0001B0011           ×B1100cos2θ1+rr2(B0100((12R12cosθ1r)sinθ1cosθsecθtanθ2cosθ12R1           sin2θ1r)A0002B1100sinθ1cosθ1)+A0001B0010B1100(sinθ1cosθ1rsinθ12R1). (A34)

Appendix B. Aberration Coefficients Ehijk and Fhijk

E1000=A1000C1000+C0001, (A35)
E0001=A0001C1000, (A36)
E2000=A2000C1000+A10002C2000+A1000C1001+C0002, (A37)
E1001=A1001C1000+2A0001A1000C2000+A0001C1001, (A38)
E0200=A0200C1000+B01002C0200+B0100C0110+C0020, (A39)
E0110=A0110C1000+2B0100B0010C0200+B0010C0110, (A40)
E0020=A0020C1000+B00102C0200, (A41)
E0002=A0002C1000+A00012C2000, (A42)
E3000=A3000C1000+2A1000A2000C2000+A2000C1001+A10003C3000+A10002C2001+A1000C1002+C0003, (A43)
E2001=A2001C1000+2A1000A1001C2000+2A0001A2000C2000+A1001C1001           +3A0001A10002C3000+2A1000A0001C2001+A0001C1002, (A44)
E1200=A1200C1000+2A1000A0200C2000+A0200C1001+2B1100B0100C0200+B1100C0110+A1000C1020           +A1000B01002C1200+A1000B0100C1110+B01002C0201+B0100C0111+C0021, (A45)
E1110=A1110C1000+2A1000A0110C2000+A0110C1001+2B1010B0100C0200+2B1100B0010C0200+B1010C0110           +2A1000B0010B0100C1200+A1000B0010C1110+2B0010B0100C0201+B0010C0111, (A46)
E1020=A1020C1000+2A1000A0020C2000+A0020C1001+2B0010B1010C0200+A1000B00102C1200+B00102C0201, (A47)
E1002=A1002C1000+2A1000A0002C2000+2A0001A1001C2000+A0002C1001+3A1000A00012C3000           +A00012C2001, (A48)
E0201=A0201C1000+2A0001A0200C2000+2B0101B0100C0200+B0101C0110           +A0001B01002C1200+A0001B0100C1110+A0001C1020, (A49)
E0111=A0111C1000+2A0110A0001C2000+2B0100B0011C0200+2B0010B0101C0200           +B0011C0110+2A0001B0100B0010C1200+A0001B0010C1110, (A50)
E0021=A0021C1000+2A0020A0001C2000+2B0010B0011C0200+A0001B00102C1200, (A51)
E0003=A0003C1000+2A0002A0001C2000+A00013C3000, (A52)
F0100=D0010+B0100D0100, (A53)
F0010=B0010D0100, (A54)
F1100=B1100D0100+A1000B0100D1100+A1000D1010+B0100D0101+D0011, (A55)
F1010=B1010D0100+A1000B0010D1100+B0010D0101, (A56)
F0101=B0101D0100+A0001B0100D1100+A0001D1010, (A57)
F0011=B0011D0100+A0001B0010D1100, (A58)
F2100=B2100D0100+A1000B1100D1100+A2000B0100D1100+A2000D1010+A10002B0100D2100           +B1100D0101+A10002D2010+A1000B0100D1101+A1000D1011+B0100D0102+D0012, (A59)
F2010=B2010D0100+A1000B1010D1100+A2000B0010D1100+B1010D0101+A10002B0010D2100           +A1000B0010D1101+B0010D0102, (A60)
F1101=B1101D0100+A1000B0101D1100+A0001B1100D1100+A1001B0100D1100+A0001B0100D1101           +B0101D0101+A1001D1010+2A1000A0001B0100D2100+2A1000A0001D2010+A0001D1011, (A61)
F1011=B1011D0100+A1000B0011D1100+A0001B1010D1100+A1001B0010D1100           +B0011D0101+2A1000A0001B0010D2100+A0001B0010D1101, (A62)
F0210=B0210D0100+A0110B0100D1100+A0200B0010D1100+A0110D1010+3B0010B01002D0300           +2B0100B0010D0210+B0010D0120, (A63)
F0300=B0300D0100+A0200B0100D1100+A0200D1010+B01003D0300+B01002D0210+B0100D0120+D0030, (A64)
F0120=B0120D0100+A0110B0010D1100+A0020B0100D1100+A0020D1010+3B0100B00102D0300+B00102D0210, (A65)
F0102=B0102D0100+A0001B0101D1100+A0002B0100D1100+A0002D1010+A00012B0100D2100+A00012D2010, (A66)
F0030=B0030D0100+A0020B0010D1100+B00103D0300, (A67)
F0012=B0012D0100+A0002B0010D1100+A0001B0011D1100+A00012B0010D2100, (A68)

Author Contributions

Conceptualization, W.L. and M.Z.; methodology, M.Z.; validation, M.Z.; software, M.Z.; writing―original draft preparation, M.Z.; writing―review and editing, W.L. and Y.J.; supervision, W.L.; funding acquisition, W.L. and S.Y.

Funding

The work presented in this paper was funded by National Key R and D Program of China (Number 2018YFF01011000); Scientific and Technological Developing Scheme of Jilin Province (Number 20170519019JH); Jlin Province Special Funds for High-tech Industrialization in cooperation with the Chinese Academy of Science (CAS) (Number 2017SYHZ0025); the National Youth Program Foundation of China (Grant Number 61805233).

Conflicts of Interest

The authors declare no conflict of interest.

References

  • 1.Huang H., Li X.T., Cen Z.F. Optical Design and Testing V. Volume 85570R. Society of Photo-Optical Instrumentation Engineers; Bellingham, WA, USA: 2012. Optical system design for a shortwave infrared imaging spectrometer; pp. 1–10. [Google Scholar]
  • 2.Jacob R., Aaron B., Kevin P.T., Jannick P.R. Freeform spectrometer enabling increased compactness. Light Sci. Appl. 2017;6:10. doi: 10.1038/lsa.2017.26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Liu C., Christoph S., Thomas F.P., Uwe D.Z., Herbert G. Optical Design and Tolerancing of a Hyperspectral Imaging Spectrometer. Volume 9947. Society of Photo-Optical Instrumentation Engineers; Bellingham, WA, USA: 2016. pp. 1–10. [Google Scholar]
  • 4.Pang Y.J., Zhang Y.X., Yang H.D., Liu Z.Y., Huang Z.H., Jin G.F. Compact high-resolution spectrometer using two plane gratings with triple dispersion. Opt. Express. 2018;26:6382–6391. doi: 10.1364/OE.26.006382. [DOI] [PubMed] [Google Scholar]
  • 5.Prieto-Blanco X., Montero-Orille C., Couce B., de la Fuente R. Analytical design of an Offner imaging spectrometer. Opt. Express. 2006;20:9156–9168. doi: 10.1364/OE.14.009156. [DOI] [PubMed] [Google Scholar]
  • 6.Tang T.J., Zhang Z., Wang B.H. Design of visible/infrared double-band spectral imager. J. Phys. Conf. Ser. 2016;680:1–6. [Google Scholar]
  • 7.Zhang D., Zheng Y.Z. Hyperspectral Imaging System for UAV. Volume 96780R. Society of Photo-Optical Instrumentation Engineers; Bellingham, WA, USA: 2015. pp. 1–6. [Google Scholar]
  • 8.Kwo D., Lawrence G., Chrisp M. Design of a Grating Spectrometer from a 1:1 Offner Mirror System. Volume 818. Society of Photo-Optical Instrumentation Engineers; Bellingham, WA, USA: 1987. pp. 275–279. [Google Scholar]
  • 9.Mouroulis P. Low-distortion imaging spectrometer designs utilizing convex gratings; Proceedings of the International Optical Design Conference, International Optical Design Conference; Kona, HI, USA. 8–12 June 1998; pp. 594–601. [Google Scholar]
  • 10.Huang Y.S., Pei Z.R., Hong R.J., Li B.C., Zhang D.W., Xu B.L., Ni Z.J., Zhuang S.L. Non-approximate method for designing Offner spectrometers. Optik. 2014;125:4578–4582. doi: 10.1016/j.ijleo.2013.12.063. [DOI] [Google Scholar]
  • 11.Seo H.K., Hong J.K., Soo C. Aberration analysis of a concentric imaging spectrometer with a convex grating. Opt. Commun. 2014;330:6–10. [Google Scholar]
  • 12.Chrisp M.P. Convex Diffraction Grating Imaging Spectrometer. [(accessed on 12 October 2018)]; Available online: http://www.freep-atentsonline.com/5880834.html.
  • 13.Xiang L.Q., Mikes T. Corrected Concentric Spectrometer. [(accessed on 18 April 2018)]; Available online: http://www.freepate-ntsonline.com/6266140.html.
  • 14.Robert L.L. Out-of-plane dispersion in an Offner spectrometer. Opt. Eng. 2007;073004:4. [Google Scholar]
  • 15.Prieto-Blanco X., Montero-Orille C., González-Núñez H., Mouriz M.D., Lago E.L., de la Fuente R. Imaging with classical spherical diffraction gratings: The quadrature configuration. Opt. Soc. Am. A. 2009;11:2400–2409. doi: 10.1364/JOSAA.26.002400. [DOI] [PubMed] [Google Scholar]
  • 16.Prieto-Blanco X., Montero-Orille C., González-Núñez H., Mouriz M.D., Lago E.L., de la Fuente R. The Offner imaging spectrometer in quadrature. Opt. Express. 2010;12:12756–12769. doi: 10.1364/OE.18.012756. [DOI] [PubMed] [Google Scholar]
  • 17.Prieto-Blanco X., González-Núñez H., de la Fuente R. Off-plane anastigmatic imaging in Offner spectrometers. Opt. Soc. Am. A. 2011;11:2332–2339. doi: 10.1364/JOSAA.28.002332. [DOI] [PubMed] [Google Scholar]
  • 18.Prieto-Blanco X., de la Fuente R. Compact Offner–Wynne imaging spectrometers. Opt. Commun. 2014;328:143–150. doi: 10.1016/j.optcom.2014.04.060. [DOI] [Google Scholar]
  • 19.Zhao M.H., Li W.H., Bayanheshig, Lv Q. Aberration correction technique of Offner imaging spectrometer. Opt. Precis. Eng. 2017;25:3001–3011. doi: 10.3788/OPE.20172512.3001. [DOI] [Google Scholar]
  • 20.Zhao X.L., Bayanheshig, Li W.H., Jiang Y.X., Yang S. Integrated design to complement aberrations of spherical focusing mirrors and concave holographic gratings in spectrometers. Acta Opt. Sin. 2016;36:10. [Google Scholar]
  • 21.Noda H., Namioka T., Seya M. Ray tracing through holographic gratings. Opt. Soc. Am. 1974;64:1037–1042. doi: 10.1364/JOSA.64.001037. [DOI] [Google Scholar]
  • 22.Namioka T. Analytical representation of spot diagrams and its application to the design of monochromators. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrometers Detect. Assoc. Equip. 1992;319:219–227. doi: 10.1016/0168-9002(92)90557-K. [DOI] [Google Scholar]
  • 23.Namioka T., Koike M., Content D. Geometric theory of the ellipsoidal grating. Appl. Opt. 1999;33:7261–7274. doi: 10.1364/AO.33.007261. [DOI] [PubMed] [Google Scholar]
  • 24.Masui S., Namioka T. Geometric aberration theory of double-element optical systems. Opt. Soc. Am. A. 1999;16:2253–2268. doi: 10.1364/JOSAA.16.002253. [DOI] [Google Scholar]
  • 25.Namioka T., Koike M., Masui S. Geometric theory for the design of multielement Optical system. Opt. Precis. Eng. 2001;9:458–466. [Google Scholar]
  • 26.Namioka T. Aspheric wave-front recording optics for holographic gratings. Appl. Opt. 1995;34:2180–2186. doi: 10.1364/AO.34.002180. [DOI] [PubMed] [Google Scholar]