Abstract
An improved method tor the coulometric assay of uranium and uranium oxide has been developed based on the electrogeneration of Ti(III) in H2SO4, using Fe(II) as a catalyst. The endpoint is determined amperometrically. Hydrogen peroxide is used as the oxidant in the dissolution of the uranium to avoid interferences from nitrate. The precision of the method as indicated by the standard deviation of an individual observation ranged from 0.008 weight percent for the analysis of the metal to 0.02 weight percent for the analysis of the oxides.
Keywords: amperometry, coulometric titration, electrogeneration, high-precision coulometry, hydrogen peroxide, standard reference material, titanium, titanous ion, uranium, uranium oxide
1. Introduction
Coulometry is an absolute method of analysis based on Faraday’s Laws of Electrolysis which relate electric current (i), time (t), and chemical equivalent weight (mole) through the equation:
The constant of proportionality, F, is the Faraday, the exact value of which has been the subject of intense research over the years at the National Bureau of Standards (NBS) [1–4]1. The parameters which define the analytical system are based on measurements of physical quantities, the national standards of which are maintained by NBS, i.e., volt, ohm, second, and gram.
This paper is another from a series of investigations designed to develop highly precise and accurate analytical procedures based upon the coulometric technique. Earlier papers have described the coulometric titrations of acids and bases [5], halides [6], and potassium dichromate [7]. The method described here has been developed for high-precision analysis of uranium metal and its compounds, which are of great scientific, industrial, and commercial importance. Some portions of this investigation have been previously reported at the 1965 EURATOM Conference on High Precision Analysis of Substances of Interest to Nuclear Energy [8].
Among the multitude of existing methods for analysis of uranium, only a few are sufficiently precise for assaying relatively high purity materials. The classical methods involve preliminary reduction of uranyl ion to a mixture of tri- and tetravalent uranium by either passing it through a Jones reductor or reducing the uranyl ion at a mercury or gold amalgam cathode, followed by air oxidation of U(III) to U(IV) and finally oxidation of U(IV) to U(VI) with an oxidant such as K2Cr207 [9,10]. One variation involves potentiometric titration of a mixture of trivalent and tetravalent uranium (obtained from a Jones reductor) first to U(III) - U(IV) endpoint and then to U(IV) - U(VI) endpoint. The uranium content is calculated from the difference between the two endpoints [11]. In another variation, U(III) - U(IV) mixture is titrated potentiometrically to the first endpoint, then solid dichromate is added in excess and the excess dichromate is determined coulometrically [12], Another method involves the dissolution of uranium metal in orthophosphoric acid, followed by a titration with dichromate using no prereduction [13]. Recently a precise titrimetric method for uranium was reported by Leon Pszonicki [14]. In this method U(VI) is reduced to U(III) in hydrochloric acid solution, using an amalgamated cadmium reductor. Orthophosphoric acid is used to oxidize U(III) to U(IV) followed by quantitative oxidation with dichromate to U(VI). All of the aforementioned procedures involve the use of potassium dichromate as the quantitative oxidant converting U(IV) to U(VI). Most require pre-reduction and pre-titration steps to prepare the analyte for the actual assay.
An alternative to the oxidative assay of U(IV) to U(VI) is one based on a quantifiable reduction of U(VI) to U(IV). Such a procedure requires a strong reductant obtainable in a purified and stable form and deliverable in discrete quantities during the course of a titration. Of the possible chemical species meeting these requirements, titanous ion (Ti(III)) has proved to be the reductant of choice for this application. Several articles have been written describing conditions for the generation of Ti(III) [15–17]. The first successful coulometric titration method for uranium with the use of electrogenerated Ti(III) was reported by Lingane and Iwamoto [18]. This method uses a mercury-pool cathode at elevated temperatures to effect complete and efficient reaction.
Kennedy and Lingane proposed another coulometric procedure employing a platinum cathode and a catalyst. Their procedure was intended to circumvent the drawbacks associated with a mercury cathode and high temperatures [19,20]. In the present paper, a method based on the work of Kennedy and Lingane is described. The existing procedure has been refined to improve the precision and accuracy of the assay of uranium and to make it applicable to the certification of Standard Reference Materials (SRMs).
2. Summary of Method
The method is based on the coulometric reduction of hexavalent uranium, U(VI), to the tetravalent state, U(IV), using titanous ion, Ti(III), as the intermediate reductant and Fe(II) as a catalyst. This reduction takes place in a solution of 1 mol/L titanyl sulfate and 9 mol/L sulfuric acid at a platinum cathode in the controlled-current mode. The current density of the cathode is maintained at or below 2.5 mA/cm2. The standard potentials of the reduction reactions are:
At the beginning of the titration, direct reduction of U(VI) to U(IV) is the principal cathodic reaction. As the uranyl ion concentration is depleted, coreduction of titanyl ion begins. The generated titanous ion in turn acts as a reductant for the remaining uranyl ion in the bulk of the solution. Finally, as the endpoint is approached, titanous ion generation becomes the principal cathodic reaction. Under these conditions, the overall reduction of U(VI) to U(IV) occurs with 100 percent current efficiency.
To add further complexity to the reduction process, it has been reported by various authors, and verified in our laboratory, that the rate of reduction of U(VI) by Ti(III) is a slow process. Lingane and Kennedy [19] pointed out that in a sulfuric acid medium the reduction of U(VI) to U(V) proceeds rapidly, but further reduction to U(IV) is the slow and rate-determining step. They found that small amounts of ferrous ion catalyze the reduction of U(V) to U(IV) facilitating the titration at room temperature. Ferrous ion reduces U(V) to U(IV) more rapidly than does Ti(III) and in turn the ferric ion produced in this reaction is reduced relatively rapidly by the trivalent titanium.
Unfortunately, even in the presence of ferrous ion, the kinetics of this reaction are not sufficiently favorable for high precision work. The approach to equilibrium is slow. The rate expression derived by Kennedy and Lingane for this reaction in a sulfuric acid medium is:
It is apparent that since the rate of the reaction is of the first order with respect to U(VI) concentration and of the second order with respect to Ti(III) concentration, the equilibrium is established more rapidly in the presence of excess Ti(III). Accordingly, it is advantageous to over-titrate by generating a slight excess of Ti(III) and then, when equilibrium is established, to extrapolate back to the equivalence point after determining the response factor of the indicator electrode.
The principal modifications to the method which are described in the paper and which have resulted in increased precision and accuracy are:
a more stable endpoint detection system composed of a platinum versus saturated calomel electrode pair in the amperometric mode, rather than the dual platinum biamperometric technique;
highly stable and accurate current-controllers and timing circuitry to assure accurate integration of charge;
improved dissolution procedures for uranium metal and oxides;
overtitration to improve the kinetics at the endpoint.
3. Experimental
3.1. Apparatus
The apparatus used in this work is described in detail in previous publications [2,21–24]. The constant-current sources are of two types. The first employs a commercially-available power supply operated in the constant-current mode. The instrument has a maximum output of 60 V and a current rating of 0 to 2 A. The stability of this power supply is on the order of 0.001 percent per 8 hours and its reliability proven through its use in high-precision coulometric iodimetry [21] and on the determinations of the atomic weights of gallium [22] and zinc [23]. The second type of current-source, used in the later stages of this experiment, was designed and built at NBS in connection with a redetermination of the Faraday via 4-aminopyridine [24]. It is powered by batteries to minimize problems associated with ac ripple and ground loops and is stable to better than 0.001 percent per 8 hours.
The output of either current source is adjusted such that the current flowing through a standard resistor (which is in series with the coulometric titration cell) produces a potential equal to the emf of a Weston saturated cell. Balance is monitored with a null-point detector such as a microvoltmeter or galvanometer, and maintained manually.
The standard resistors are of the NBS-type, devoid of their metal containers and immersed in a large oil-bath to ensure temperature stability. The Weston saturated cells were enclosed in a thermostated box. The cells and resistors are periodically calibrated against the national working standards of voltage and resistance maintained at NBS. A commercially available electronic frequency meter was used as the timer, its operational mode set to count the cycles of the NBS 10 kHz standard frequency. The current-source and timer are integrated through a switching system, designed and built at NBS. In the standby position, the current flows through a surrogate load resistor comparable in resistance to that of the titration cell, and the timer gate is open; in the active position, the current is channeled to the titration cell, and the timer gate is closed.
The titration cell, similar to one used in previous high-precision coulometric research at NBS [2], is constructed of two 180 mL tail-form beakers connected by two intermediate chambers with fritted-glass separaters. A silicic acid gel plug was prepared in the last compartment of the titration H-cell by mixing sodium silicate solution with sulfuric acid directly in the compartment. The volumes of solutions used in gel preparation were selected such that the amount of gel produced would fill one-half of the 180 mL beaker, completely covering the glass frit of the connecting chamber. Sulfuric acid (2 mol/L) was poured over the silicic acid plug. The titration cell was further extended through a U tube, containing another silicic acid plug, to a 180 mL beaker filled with saturated KCl solution and serving as the anode compartment. The saturated calomel reference electrode of the indicator circuit was also dipped into this compartment.
A modified cage-type platinum electrode was used as the cathode. It consisted of a platinum foil cylinder concentrically located within a platinum strip cage. The apparent area of the former was 28 cm2, and of the latter 12 cm2, yielding a total area of 40 cm2. Early experiments showed that the platinum cathode will perform less than satisfactorily if its surface is not properly prepared. Consequently the conditioning procedure described below was scrupulously followed:
store the electrode each night in a solution of 1 mol/L K2Cr207 and 2 mol/L H2SO4
rinse with distilled water and soak in a solution of 0.1 mol/L ferrous ammonium sulfate and 2 mol/L H2SO4
rinse thoroughly with distilled water and install in titration cell.
After several titrations, it was sometimes necessary to strip the surface of the electrode by dipping it in aqua regia for 5 minutes and then reconditioning it according to the above procedure. The anode was a silver cylinder made from 2 mm thick foil and having a surface area of 100 cm2.
The amperometric indicator system consisted of a platinum foil electrode (1 cm2) and a saturated calomel electrode (SCE) with a flushable liquid junction. A polarograph was used as the source of the applied emf as well as for recording the indicator current. The value of applied emf was determined by running a current-voltage scan using the indicator electrode versus the SCE in the titration cell containing a solution similar to that expected at the equivalence point of an actual titration (Figure 1). The flat portion of this voltammogram typically extended from about +0.2 to +0.3 V versus SCE, which represents the useable window for the applied emf. Since this window is rather narrow (due to the oxidation of Fe(II) at voltages more positive than +0.3 V and the reduction of Ti(IV) at voltages below +0.2 V) and has a tendency to shift slightly from titration to titration, it is important that the potentiostat of the polarograph be very stable to achieve linearity in the endpoint current readings.
FIGURE 1.
Current-voltage curves. Platinum indicator electrode vs. SCE reference electrode in coulometric titration cell. Supporting electrolyte: 9 mol/L H2SO4, 1 mol/L Ti(IV), 0.1 g Fe(NH4)2(SO4)2, 100 μeq U. (A) at endpoint. (B) 20 μeq excess of Ti(III). (C) 50 μeq excess of Ti(III). (D) 80 μeq excess of Ti(III).
All weighings were performed on a 20 g capacity microbalance and were precise to 0.003 mg. The weights of all samples were corrected for air buoyancy.
Either nitrogen or argon was passed through the supporting electrolyte prior to and during the titration to keep atmospheric oxygen out of the cell. This purging gas was purified and conditioned by bubbling it through a chromous sulfate solution and then through a tower containing 9 mol/L H2SO4.
3.2. Reagents
The uranium which was assayed in this work was of two forms: the pure metal cast in dingots, one lot of which became Standard Reference Material 960; and the oxide of uranium of nominal stoichiometry U308 including samples of SRM 950a and SRM 950b.
The supporting electrolyte in the cathode compartment was a solution of 9 mol/L sulfuric acid and 1 mol/L titanyl sulfate. It was prepared according to the following procedure: purified liquid titanium tetrachloride is partially hydrolyzed (50% by weight) by slowly adding it to distilled water which is being well stirred and cooled in an ice-bath. The reaction is vigorous and exothermic. It must be carried out in a fume hood, because hydrogen chloride gas is liberated. The resulting solution is approximately 2.6 molal Ti(IV). This hydrolyzed titanium solution is then added slowly to concentrated (95%) sulfuric acid (ACS reagent-grade) (750 g Ti(IV) solution per liter of sulfuric acid). Once again HCl is evolved rather vigorously. When addition is complete, the solution is purged with nitrogen to remove all traces of HCl. A clear solution should result. Dilute with distilled water such that the final solution is 9 mol/L H2SO4 and 1 mol/L Ti(IV). [Note: After several weeks, a white precipitate of Ti02 will form due to the hydrolysis of the titanyl sulfate. Since this effectively depletes the Ti(IV) available for reaction, the solution should be discarded and fresh electrolyte prepared.]
The acids and the hydrogen peroxide used in the dissolution of uranium were ACS-reagent grade, as was the ferrous ammonium sulfate hexahydrate used to catalyze the reduction of U(VI).
3.3. Preparation and Dissolution of Uranium Metal
Using bolt-cutters, the dingot is cut into pieces weighing approximately 1 gram each. Surface oxide and impurities are removed by dipping the uranium sample in 8 mol/L HN03 for 10 minutes, rinsing in distilled water, etching in 3 mol/L HCl for 5 minutes, rinsing in distilled water, and drying in a vacuum desiccator. The sample is weighed on a calibrated analytical balance before the oxide reappears, correcting for the buoyancy of air (density of uranium, 19.05 g/cm3). The sample is placed into a 125 mL Erlenmeyer flask and dissolved slowly in 2–10 mL of 6 mol/L HCl, warming if necessary on a hot plate. It has been found most convenient to rest the flask in an inclined position in a small crystallizing dish throughout the dissolution process. An inverted 50 mL beaker is placed over the top of the flask to catch any possible spattering. After the metal has dissolved and only a small amount of black residue remains, 2 mL of 16 mol/L HN03 are slowly added to oxidize all of the uranium to the hexavalent state, U(VI). The black residue will also dissolve. The beaker serving as the splash guard is rinsed with 5 mL of distilled water, this water being added to the sample flask. The walls of the flask are then rinsed with 5 mL of distilled water. Next, 5 mL of concentrated H2SO4 (95%) are carefully added to the sample. The solution is then evaporated down to SO3 fumes three times with the addition of 5 mL of water after each fuming. It is imperative that all traces of chloride and nitrate be removed from the solution. (Nitrate is reduced at the cathode and as such would pose a serious interference in the subsequent coulometric titration.)
To avoid the repetitive fumings required to eliminate nitrate, hydrogen peroxide can be used instead of nitric acid to oxidize the uranium. With this procedure, after the metal has dissolved in HCl, 1 mL of H2O2 (30%) is added slowly. A light yellow precipitate (a peroxide of uranium) may form but is unstable and is rapidly reduced back to U(VI) upon heating, resulting in a clear brilliant yellow solution. To destroy the excess hydrogen peroxide the solution is evaporated nearly to dryness. The sides of the flask are then rinsed with distilled water and 5 mL of H2SO4 (95%) are added to the solution. The solution is then evaporated to SO3 fumes. If allowed to stand a day or two, a crystalline percipitate may appear. Distilled water should be added to redissolve the precipitate, with subsequent evaporation to SO3 fumes again before analyzing.
Blanks containing the same volumes of reagents used in the dissolution of the uranium samples were put through the same evaporation procedure. To some of these blank samples were added 30.31 μeq of uranium from a stock solution.
3.4. Preparation and Dissolution of Uranium Oxide (U3O8)
A sample of the uranium oxide (approximately 1.5 g) is placed in a platinum boat and ignited in a furnace for 1 hour at 900 °C (SRM 950b was fired at 800 °C). After the firing process, the sample is allowed to cool for 15 minutes, and stored in a desiccator. After cooling, the boat plus sample is weighed on a calibrated analytical balance, the sample is dumped from the boat into a 125 mL Erlenmeyer flask, and the empty boat is reweighed. The mass of the sample is calculated by difference, corrected for the buoyancy of air (density of uranium oxide 8.3 g/cm3). The sample is digested in 10 mL of 12 mol/L HCl on a warm hotplate overnight (8 to 12 hours). Care should be taken to avoid evaporating the solution to dryness. As described earlier, the flask is in an inclined position with a beaker over the top. After the uranium oxide sample is dissolved (some black residue may remain), the sample is carried through the oxidation and fuing procedures outlined in section 3.3. Either nitric acid or hydrogen perioxide may be used, with the latter being preferred.
3.5. Coulometric Titration
Prior to delivery of the sample, 75 mL of supporting electrolyte are added to the cathode chamber of the cell togetehr with 100 mg of ferrous ammonium sulfate hexahydrate. Nitrogen or argon is then passed through the compartment and into the chamber for one-half hour to remove air. After purging, the catholyte is permitted to flow into the intermediate compartments to the extent that it just covers the bottom of each compartment, thus establishing electrolytic connection between the anode and cathode chambers.
A small amount of uranyl sulfate solution, about 5μeq, is then added to the cathode chamber and is pre-titrated by passage of small increments of charge equivalent to 1 μeq using 3–10 mA. At the conclusion of each increment, the amperometric indicator current is recorded. This current, small and essentially constant up to the equivalence point, exhibits a curvature in the vicinity of the equivalence point and becomes a linear function of the concentration beyond the equivalence point. After the first excess of Ti(III) is noted, the solution is permitted to equilibrate for one-half hour, after which increment-wise generation is resumed. The linear portion of the indicator current is extrapolated graphically and its intersection with the zero-current line is taken as the endpoint.
After completion of the pre-titration, the intermediate cell chambers are rinsed by repeatedly emptying and filling the chambers with electrolyte, applying suction or nitrogen pressure as required. The final reading of the indicator current is recorded and this reading is used to determine the amount of overtitration of the pre-titration step.
Following the pre-titration, the intermediate compartments are completely filled with catholyte and the sample, which has been deaerated, is delivered into the cathode chamber by applying nitrogen pressure to a specially designed polyethylene siphon system.
The sample is titrated using 101 mA current for a precalculated period of time corresponding to a few microequivalents in excess of the stoichiometric amount. The middle compartments are again emptied into the cathode chamber using nitrogen pressure. The sample flask, which is still connected to the cathode chamber through the siphone tude, is rinsed several times with the catholyte by alternately applying vacuum and pressure to the siphon system.
Following the one-half hour equilibration period, small increments of charge are again passed as already described in the pre-titration procedure and the indicator current is recorded after each addition. The indicator current curve is again extrapolated to locate the virtual endpoint. The charge required to reduce the uranium is the amount to the endpoint plus the excess resulting from the preceding pre-titration.
4. Results and Discussion
As in any precise chemical analysis, it was necessary to determine the blank resulting from reagents used in the sample preparation. It is quite apparent that any of the ions which are reducible by Ti(III) will be titrated along with uranium, yielding high results. It should be noted, however, that reducible impurities in the supporting electrolyte do not contribute to the overall blank since they are removed in the pre-titration step.
To evaluate the bias arising from chemical treatment of the sample, a number of titrations were made of samples containing 30.31 μeq of uranium which had been processed with the same quantities of acids as in the larger sample preparation. The results for titrations of 11 aliquot samples gave a blank value of +0.27 μeq ± 0.17 μeq. In addition, a study was made of the relationship between the sample size and the assay. This study indicated a systematic increase of the assay value as the sample size decreased. The plot of the difference between the calculated and found number of equivalents verses sample size for 30 determinations of the uranium content in dingot metal yields a linear relationship that, upon extrapolation to zero sample size, gives an intercept close to 0.35 μeq (Fig. 2), well within the uncertainty of the value obtained in the blank titrations. On the basis of these two pieces of experimental evidence, a 0.27 μeq bias must be subtracted from all titrations.
FIGURE 2.
Analysis of titration data for bias.
A summary of the titration data of dingot uranium, corrected for the bias, is given in Table 1. The data were subdivided into four groups, encompassing different sample sizes: Group I — up to 100 mg of 0; Group II — 100 to 500 mg of U; Group III — 500 to 100 mg of U; and Group IV — higher than 1000 mg of U. The table summarizes individual results, along with the respective averages, and the corresponding standard deviation of a single determination for each group. It can be seen that for larger samples, the precision remains constant.
Table 1.
Assay in weight percent, of dingot uranium in method development.
Average Sample, Size, mg | ||||
---|---|---|---|---|
50 | 300 | 750 | 1010 | |
100.18 | 99.966 | 99.964 | 99.964 | |
100.15 | 99.957 | 99.954 | 99.979 | |
99.31 | 99.977 | 99.977 | 99.972 | |
99.95 | 99.970 | 99.964 | 99.913 | |
99.88 | 99.966 | 99.980 | 99.987 | |
100.18 | 99.971 | 99.948 | ||
99.91 | 99.982 | 99.985 | ||
99.96 | 99.982 | 99.962 | ||
99.978 | 99.947 | |||
99.983 | 99.984 | |||
99.974 | 99.976 | |||
99.968 | 99.974 | |||
99.975 | 99.993 | |||
99.969 | ||||
99.965 | ||||
99.982 | ||||
Average | 99.94 | 99.973 | 99.970 | 99.975 |
Std. Dev. | 0.29 | 0.008 | 0.015 | 0.009 |
It should be kept in mind that the assay reported here represents the reductometric value for the material, and as such would include iron and any other impurities in the uranium which would be reduced by trivalent titanium.
Uranium metal, issued by NBS as SRM 960, was analyzed independently by two analysts who used this method. The results are shown in table 2. Excellent agreement is obtained between the two analyses, indicating no operator bias. A correction of −0.0108 weight percent must be applied to this reductometric assay due to two electroactive impurities known to be present, 42.1 ppm iron and 4 ppm vandium. Thus this material is certified at 99.975 weight percent uranium.
Table 2.
Analysis of uranium metal, SRM 960. (uncorrected for impurities, see text)
Analyst | Number of Determinations | Atomic Weight of Uranium in Sample | Assay Weight Percent | Standard Deviation |
---|---|---|---|---|
A | 21 | 238.0289 | 99.9855 | 0.0081 |
B | 4 | 238.0289 | 99.9848 | 0.0038 |
The developed method of analysis was also used for the assay of various U3O8 preparations. The values obtained for these materials are summarized in table 3. SRM 950a and SRM 950b are of natural uranium isotopic composition and are issued by NBS as chemical standards for uranium analyses. The other materials are for use in the preparation of uranium isotopic standards. The atomic weights were calculated from the isotopic abundances of uranium in the various samples. All of the materials were “pure” in that the total of metallic impurities did not exceed 0.01 percent.
Table 3.
Analysis of Uranium Oxide, U3O8.
Material | Number of Determinations | Atomic Weight of Uranium in Sample | Assay Weight Percent | Standard Deviation |
---|---|---|---|---|
SRM | 3 | 238.03 | 99.923 | 0.023 |
SRM 950b | 6 | 238.03 | 99.958 | 0.022 |
Depleted Secondary | 3 | 238.050 | 99.929 | 0.027 |
Depleted Primary | 3 | 238.051 | 99.921 | 0.009 |
Enriched Primary | 3 | 235.047 | 99.922 | 0.006 |
5. Summary
An improved method for the coulometric assay of uranium and uranium oxide has been developed based on the electrogeneration of titanous ion in sufluric acid. Ferrous ion is used as a catalyst. The endpoint using an amperometric system is determined by extrapolation after a slight overtitration which increases the reaction rate and guarantees complete reaction. Hydrogen peroxide is employed as the oxidant in the dissolution procedure in lieu of nitric acid which if not completely destroyed would greatly interfere with the subsequent coulometric titration. The method has been applied to the analysis of several preparations of uranium metal and uranium oxide, including SRM’s. The demonstrated precision for the assay of the metal as represented by the standard deviation of the individual measurement ranges from 0.004 to 0.008 weight percent, and that for the assay of the oxide from 0.006 to 0.027 weight percent.
6. Acknowledgments
The authors wish to express their gratitude to L. A. Machlan and J. R. Moody for their help in the heat treatment of the uranium oxide samples, to E. J. Maienthal for her work in the polarographic analysis of the impurities in the metal, to J. K. Taylor for many meaningful discussions, and to the Office of Standard Reference Materials, NBS, for their support in this project.
Footnotes
Figures in brackets indicate literature references at the end of this paper.
7. References
- [1].Craig D. N.; Hoffman J. I.; Law C. A. Hamer W. J. Determination of the value of the Faraday with a silver-prechloric acid coulometer. J. Res. Natl. Bur. Stand. (U.S.). 64A(5): 381 −402; 1960. September-October. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].Marinenko G; Taylor J. K. Electrochemical equivalents of benzoic acid and oxalic acid. Anal. Chem. 40(11), 1645–1651; I960. September. [Google Scholar]
- [3].Koch W. F. The value of the Faraday via 4-aminopyridine In Atomic Masses and Fundamental Constants 6, Nolen J. A. Jr. and Benenson W., ed. New York, NY; Plenum Publishing Corp.; 1980, 167–172. [Google Scholar]
- [4].Bower V. E.; Davis R. S.; Murphy T. J.; Paulsen P. J.; Gramlich J. W.; Powell L. J. Recalculation of the Faraday constant due to a new value for the atomic weight of silver. J, Res. Natl. Bur. Stand. (U.S.). 87(1): 21–22; 1982. Jannary-February. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Taylor J. K.; Smith S. W. Precise coulometric titration of acids and bases. J. Res. Natl. Bur. Stand. (U.S.). 63A(2): 153–159; 1959. September-October. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Marinenko G.; Taylor J. K. Precise coulometric titration of halides. J. Res. Natl. Bur. Stand. (U.S.). 67A(1): 31–35; 1963. January-February. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Marinenko G.; Taylor J. K. Precise coulometric titrations of potassium dichromate. J. Res. Natl. Bur. Stand. (U.S.) 67A(5): 453–459; 1963. September-October. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [8].Taylor J.K.; Marinenko G. High-precision coulometric titrations with special reference to the determination of uranium. Proceedings of the conference on high-precision analysis of substances of interest to nuclear energy EURATOM; Brussels, Belgium 1965. 147–153. [Google Scholar]
- [9].Rodden C. J.; Warf J. C. Uranium, chapter 1 in Analytical Chemistry of the Manhattan Project. Rodden C. J., ed. New York, NY: McGraw-Hill Book Company, Inc.; 1950. 3–159. [Google Scholar]
- [10].Kramer H. Determination of uranium with high precision, New Brunswick Laboratory Report; 1960, May 20 p. [Google Scholar]
- [11].Voss F. S.; Green R. E. A precision potentiometric titration method for the determination of uranium. Atomic Energy Commission (U.S. ). AEC Report 4030; 1955. 11 p. [Google Scholar]
- [12].Vita O. A.; Trivisonno C. F.; Walker C. R. An improved titrimetric method for the precise determination of uranium. Goodyear Atomic Corp. Report. GAT-471 Chemistry; 1962. November 10 p. [Google Scholar]
- [13].Duckitt J. A.; Goode G. C. A method for high-precision assay of uranium metal. Analyst. 87(1031), 121–124; 1962. February. [Google Scholar]
- [14].Pszonicki L. Precise titrimetric determinations of uranium in high-purity uranium compounds. Talanta 13(3), 403–408; 1966. March. [DOI] [PubMed] [Google Scholar]
- [15].Arthur P.; Donahue J. F. Titanic chloride as an intermediate in coulometric analyses. Anal. Chem, 24(10), 1612–1614; 1952. October. [Google Scholar]
- [16].Malmstadt H. V.; Roberts C. B. Automatic spectrophotometric titration with coulometrically generated titanous ion. Anal. Chem. 27(5), 741–744; 1955. May. [Google Scholar]
- [17].Parsons J. S.; Seaman W. Coulometric titration of dyes with externally generated titanous ion. Anal. Chem. 27(2), 210–212; 1955. February. [Google Scholar]
- [18].Lingane J. J. ; Iwamoto R. T. Coulometric titration of uranium with electrogenerated titanous ion. Anal. Chim. Acta. 13(5), 465–472; 1955. November. [Google Scholar]
- [19|.Lingane J. J,; Kennedy, J. H. Coulometric titration of iron, cerium and vanadium with titanous ion. Anal. Chim. Acta. 15(5), 465–472; 1956. November. [Google Scholar]
- [20].Kennedy J. H.; Lingane J. J. Coulometric titration of uranium and uranium-vanadium mixtures with +3 titanium. Anal. Chim. Acta, 18(3), 240–244; 1958. March. [Google Scholar]
- [21].Marinenko G; Taylor J. K. High-precision coulometric iodimetry. Anal. Chem. 39(13), 1568–1571; 1967. November. [Google Scholar]
- [22].Marinenko G. On the atommic weight of gallium. J. Res. Natl. Bur. Stand. (U.S.). 81A(1): 1–4; 1977. January-February. [Google Scholar]
- [23].Marinenko G. Absolute determination of the electrochemical equivalent and the atomic weight of zinc. Method, apparatus, and preliminary experiments. J. Res. Natl. Bur. Stand. (U.S.). 79A(6): 737–745; 1975. November-December. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [24].Koch W. F.; Diehl H. Additional results on the value of the Faraday. Talanta 23(7), 509–512; 1976. July. [DOI] [PubMed] [Google Scholar]