Journal of Research of the National Bureau of Standards logoLink to Journal of Research of the National Bureau of Standards
. 1983 Mar-Apr;88(2):91–103. doi: 10.6028/jres.088.005

Ultrasonic Continuous-Wave Beam-Power Measurements; International Intercomparison

Carl E Tschiegg 1,*, Martin Greenspan 1,**, Donald G Eitzen 1,*
PMCID: PMC6768165  PMID: 34566097

Abstract

Some quartz transducers designed and fabricated at the National Bureau of Standards as transmitters of ultrasonic power appear to be sufficiently stable and linear to serve as standards. Therefore, an international intercomparison of measurements of the continuous-wave (cw) power emitted by these standards was arranged. Each of the seven participating laboratories performed such measurements using one or more methods representing its practice and reported the results to the National Bureau of Standards which served as the pilot laboratory. We present the results mostly in the form of tables. Some remarks on stability are appended.

Keywords: intercomparison of standards, ultrasonic power standards, ultrasonic transducers

1. Introduction

In 1974, Thomas L. Zapf, [1]1 of the then Electromagnetics Division of the National Bureau of Standards in Boulder, CO, U.S.A., described a method for the measurement of the radiation conductance of an ultrasonic transducer by means of high-accuracy impedance measurements made with a twin-T null circuit. Also described were some quartz transducers designed and constructed so that they could be expected to be stable over long intervals of time. By the spring of 1975, Helmut M. Altschuler of the Electromagnetics Division of NBS Boulder was actively arranging the international comparison of ultrasonic beam-power measurements utilizing the new standards; the technical direction of the project was to be Zapf’s responsibility. In the fall of 1976, when many of the arrangements with the participating laboratories had been completed, the responsibility for the project was transferred to Donald G. Eitzen, chief of the then Ultrasonic Standards Program Team in Washington, as a result of a management decision to consolidate some of the work in ultrasonics at NBS. Also transferred, besides some equipment, were 14 quartz transducers, having operating frequencies of 2, 3, 5, 10, and 15 MHz. These were calibrated by the modulated radiation-pressure method [2] over the ranges of ultrasonic power output from 1 to 8 mW up to 1 to 528 mW, and the transducers were found to be linear (i.e., power output proportional to square of voltage), so that each transducer could be characterized by a single radiation conductance (Gr). Furthermore, this Gr proved to be the same, within the estimated experimental errors, as that determined at very low power by the twin-T null method. Other transducers were also calibrated by a calorimetric method, using an instrument designed and built by Zapf et al. [3] at NBS Boulder and modified at NBS Washington2. Each transducer was calibrated (twice) at only one power level which varied, from case to case, over the range 50 to 750 mW. Again, the values of Gr agreed, within the estimated uncertainty, with those obtained by the other methods.

2. Procedure

Four ultrasonic transducers were selected for the intercomparison. They were intended to be operated at the fundamental series-resonance frequencies; these, together with the designations, follow:

Standard No. Operating Frequency
Nominal Actual
MHz MHz
3–16 2 1.995
3–18 2 1.995
3–20 5 5.046
3–22 5 5.015

The active element of each was a half-wave resonant, air-backed, x-cut quartz disc, having a “wrap-around” outer electrode to provide some electrostatic shielding. The discs were cut, polished, and plated (gold over chromium) by a commercial supplier. The transducers are shown in figure 1.

Figure 1.

Figure 1.

The four standard transducers.

The transducers were shipped to the participating laboratories and the measurements made in the following order:

  1. National Bureau of Standards Boulder, CO, U.S.A.

  2. National Bureau of Standards Washington, DC, U.S.A.

  3. Radiation Protection Bureau Ottawa, Ontario, Canada

  4. National Research Council Ottawa, Ontario, Canada

  5. NBS (Remeasurement)

  6. Bureau of Radiological Health Rockville, MD, U.S.A.

  7. NBS (Remeasurement)

  8. Ultrasonics Institute Sydney, Australia

  9. NBS (Remeasurement)

  10. Physikalisch-Technische Bundesanstaldt Braunschweig, FRG

  11. NBS (Remeasurement)

  12. National Physical Laboratory Teddington, Middlesex, U.K.

  13. NBS (Remeasurement)

The instructions to the participating laboratories are reproduced in the Appendix. Each laboratory used its own method or methods, a total of eight. The methods are listed together with their code designations in table 1.

Table 1.

Methods of ultrasonic cw beam power measurement and codes.

Method Code
Radiation balance, direct RD
Radiation balance, feedback RF
Radiation balance, modulated RM
Calorimeter C
Optical (Raman-Nath) O
Impedance, twin-T null T
Impedance, Q-meter Q
Reciprocity R

The voltage levels and tolerances specified in the Appendix were selected to cover ultrasonic power levels roughly as follows:

Low 2.5 — 25 mW
Medium 25 — 250 mW
High 250 — 2500 mW

3. Results

The results are given in tables 2, 3, 4, and 5 for transducers 3–16, 18, 20, and 22, respectively. In the column headed “Code” the arable numeral following the method designation (as in e.g., FLD-1) denotes one of the participating laboratories. In cases in which a laboratory used several methods, the associated numeral is different for each method. Each laboratory was furnished a key that enabled it to identify its own work, but not that of others. An exception is the pilot laboratory, NBS, for which RM-1 and T-l are used.

Table 2.

Transfer standard No. 3–16 nominal frequency 2 MHz*.

Codea Powerb Range No. of Data Applied Voltage Radiation Conductance

Min
Max
Uv
Min
Max
Avg.
Ub
V V ±% μS μS μS ±%
RD-1 low 5 30 30 0.25 9 12 11 22
med 5 100 100 0.25 11.5 11.9 11.6 7
high 5 180 180 0.25 12.0 12.2 12.1 5
overall 15 30 180 9 12.2 11.54
RD-2 low 3 35 35 1.3 11.7 11.8 11.8 10
med 12 56 71 1.3 11.2 11.9 11.5 10
high 0
overall 15 35 71 11.2 11.9 11.58
RD-3 low 10 30 30 0.5 24.9 38.5 29.9 ?c
med 10 100 100 0.5 10.4 12.6 10.9 7c
high 10 180 180 0.5 11.5 12.6 12.00 2.8c
overall 30 30 180 10.4 38.5 17.6d
RD-4 low 1 35e 35 2 9.3 9.3 9.3 19
med 2 53 71 2c 10.8 11.7 11.2 19
high 0
overall 3 35 71 9.3 11.7 10.6
RF-1 low 6 26 26 0.2 10.6 11.2 10.9 6.1c,f
med 6 81 82 0.2 10.9 11.2 11.0 5.6c,f
high 7 254 259 1.0 11.1 11.8 11.4 7.6c,f
overall 19 26 259 10.6 11.8 11.13
RF-3 low 3 35 35 1.3 11.7 11.8 11.7 10
med 6 56 70 1.3 11.9 12.0 11.9 10
high 0
overall 9 35 70 11.7 12.0 11.86
RF-4 low 9/20g 26 36 0.3 10.3 11.2 10.9 3.8
med 9/20 66 91 0.4 10.4 11.3 10.9 3.9
high 9/20 146 187 0.5 10.5 11.2 10.9 4.2
overall 27/60 26 187 10.3 11.3 10.87
RM-1 low 34 16 46 0.2 11.2 11.5 11.4 2.3
med 15 66 135 0.2 11.3 11.3 11.3 2.3
high 17 150 216 0.2 11.3 11.3 11.3 2.3
overall 66 16 216 11.2 11.5 11.34
C-3h i i i i i i 11.1 7
C-5 low 3 35 36 1.3 11.3 12.3 11.8 10
med 3 59 61 1.3 11.8 12.4 12.1 10
high 0
overall 6 35 61 11.3 12.4 11.94
0,1 low 0
med 9 70 141 1.3 10.6 12.1 11.3 10
high 5 176 211 1.3 10.9 11.7 11.3 10
overall 14 70 211 10.6 12.1 11.29
Q-4 j 10 j j i i i 10.56 i
R-2k low 6 30 38 i 10.9 11.5 11.1 17
med 0
high 0
overall 6 30 38 10.9 11.5 11.1
weighted meanp 11.44
*

Footnotes follow table 5.

Table 3.

Transfer standard No. 3–18 nominal frequency 2 MHz*.

Codea Powerb Range No. of Data Applied Voltage Radiation Conductance

Min
Max
Uv
Min
Max
Avg.
Ub
V V ±% μS μS μS ±%
RD-1 low 5 30 30 0.25 8 10 9 22
med 5 100 100 0.25 11.1 11.2 11.2 7
high 5 180 180 0.25 11.7 11.9 11.8 5
overall 15 30 180 8 11.9 10.66
RD-2 low 3 35 35 1.3 11.8 11.9 11.9 10
med 12 56 71 1.3 11.2 11.8 11.5 10
high 0
overall 15 35 71 11.2 11.9 11.57
RD-3 low 10 30 30 0.5 16.3 43.0 31.4 35c
med 10 100 101 0.5 12.6 14.0 13.2 6.7c
high 10 180 180 0.5 11.2 11.5 11.3 3.0
overall 30 30 180 11.2 43.0 18.64m
RD-4 low 1 35e 35 2.0 10.2 10.2 10.2 18
med 2 53 71 1.9c 10.5 11.4 11.0 19c
high 0
overall 3 35 71 10.2 11.4 10.7
RF-1 low 7 25 26 0.2 10.0 10.8 10.4 5.5c,f
med 7 81 82 0.2 10.0 10.7 10.3 5.6c,f
high 7 254 256 1 10.5 11.1 10.7 7.6c,f
overall 21 25 256 10.0 11.1 10.45
RF-3 low 3 35 35 1.3 11.0 11.3 11.2 10
med 6 56 70 1.3 11.4 11.5 11.4 10
high 0
overall 9 35 70 11.0 11.5 11.35
RF4.1n low 7/26g 21 36 0.3 11.2 12.5 11.7 3.8
med 7/26 72 97 0.4 11.1 11.7 11.4 3.9
high 6/22 160 296 0.5 11.8 12.2 11.9 4.2
overall 20/74 21 296 11.1 12.5 11.65
RF4.2n low 8/20g 22 32 0.3 10.3 11.0 10.7 3.8
med 8/20 70 97 0.4 10.3 10.9 10.6 3.9
high 8/20 140 191 0.5 10.4 10.9 10.7 4.2
overall 24/60 22 191 10.3 11.0 10.65
RM-1 low 32 16 36 0.2 11.0 11.3 11.2 2.3
med 13 46 150 0.2 11.0 11.2 11.1 2.3
high 19 151 232 0.2 11.0 11.4 11.2 2.3
overall 64 16 232 11.0 11.4 11.16
C-5 low 3 35 37 1.3 11.3 11.9 11.5 10
med 3 56 58 1.3 10.9 12.2 11.5 10
high 0
overall 6 35 58 10.9 12.2 11.50
0-1 low 0
med 9 70 141 1.3 10.8 11.9 11.4 10
high 6 176 211 1.3 10.8 11.5 11.3 10
overall 15 70 211 10.8 11.9 11.33
Q-4 j 10 j j i i i 10.24 i
R2.1k,n low 2 30 40 i 10.6 10.9 10.7 17
med 0
high 0
overall 2 30 40 10.6 10.9 10.7
R2.2k,n low 1 18 18 i 12.1 12,1 12.1 17
med 0
high 0
overall 1 18 18 12.1 12.1 12.1
weighted meanp 11.12
*

Footnotes follow table 5.

Table 4.

Transfer standard No. 3–20 nominal frequency 5 MHz*.

Codea Powerb Range No. of Data Applied Voltage Radiation Conductance

Min
Max
Uv
Min
Max
Avg.
Ub
V V ±% μS μS μS ±%
RD-1 low 5 12 12 0.25 49 67 58 22
med 5 40 40 0.25 78 81 78 7
high 5 100 100 0.25 69 69 69 5
overall 15 12 100 49 81 68.4
RD-2 low 0
med 6 35 57 1.3 69.9 72.8 71.1 10
high 0
overall 6 35 57 1.3 69.9 72.8 71.1 10
RD-3 low 10 12 12 0.5 54.9 110.7 93.7 46c
med 10 40 40 0.5 79.6 89.6 82.1 6.5c
high 10 124 126 0.5 72.0 74.6 73.6 1.6
overall 30 12 126 54.9 110.7 83.1
RD-4 low 1 18e 18 2.0 76.1 76.1 76.1 19
med 2 35 42 1.8c 67.5 71.8 69.7 12.5c
high 0
overall 3 18 42 67.5 76.1 71.8
RF-1 low 9 10 10 0.2 67.4 71.7 68.0 7.5c,f
med 11 32 33 0.2 68.3 70.4 68.8 7.4c,f
high 9 101 101 0.2 70.0 72.4 70.5 7.2c,f
overall 29 10 101 67.4 72.4 69.06
RF-3 low 0
med 12 35 57 1.3 69.3 73.6 72.0 10
high 6 70 71 1.3 69.2 69.8 69.5 10
overall 18 35 71 69.2 73.6 71.14
RF-4 low 11/24 9 14 0.41 66.1 72.8 69.4 4.5
med 12/26 36 48 0.25 67.2 71.6 69.3 4.1
high 12/26 68 87 0.59 67.8 71.9 70.2 4.8
overall 35/76 9 87 66.1 72.8 69.7
RM-1 low 25 7 19 0.2 69.3 70.8 70.4 2.8
med 16 20 58 0.2 70.3 70.7 70.5 2.8
high 17 61 112 0.2 69.3 71.0 70.7 2.8
overall 58 7 112 69.3 71.0 70.5
C-5 low 3 14 14 1.3 71.5 75.7 74.3 10
med 6 35 62 1.'3 66.5 70.7 69.5 10
high 0
overall 9 14 62 66.5 75.7 71.1
T-l j 1 j j na na na 69.9 1
Q-4 j 10 j j i i i 64.5 i
R-2k low 6 12 16 i 67 69 68 17
med 0
high 0
overall 6 12 16 67 69 68
weighted meanp 70.13
*

Footnotes follow table 5.

Table 5.

Transfer standard No. 3–22 nominal frequency 5 MHz.

Codea Powerb Range No. of Data Applied Voltage Radiation Conductance

Min
Max
Uv
Min
Max
Avg.
Ub
V V ±% μS μS μS ±%
RD-1 low 5 12 12 0.25 69 97 75 22
med 5 40 40 0.25 77 80 78 7
high 5 100 100 0.25 69.1 69.1 69.1 5
overall 15 12 100 69 97 74.2
RD-2 low 0
med 6 35 57 1.3 70.1 72.8 72.1 10
high 0
overall 6 35 57 1.3 70.1 72.8 72.1
RD-3 low 10 12 12 0.5 63.2 126.3 119.7 62c
med 10 40 40 0.5 75.6 93.1 80.9 6.6c
high 10 124 126 0.5 85.8 89.9 87.9 1.5
overall 30 12 126 63.2 126.3 96.2o
RD-4 low 1 18e 18 2.0 71.7 71.7 71.7 20
med 2 35 42 1.8 66.0 71.8 68.9 13
high 0
overall 3 18 42 66.0 71.8 69.8
RF-1 low 12 10 10 0.2 71.2 71.2 71.2 7c,f
med 15 32 33 0.2 71.6 72.6 71.7 7c,f
high 13 101 101 0.2 73.0 73.0 73.0 7.4c,f
overall 40 10 101 71.2 73.0 71.97
RF-3 low 0
med 6 35 56 1.3 72.7 74.5 73.6 10
high 3 70 70 1.3 72.4 72.5 72.5 10
overall 9 35 70 72.4 74.5 73.22
RF-4 low 17/44g 8 14 0.41 68.4 78.6 72.7 4.5
med 16/42 33 49 0.25 69.4 75.1 72.3 4.1
high 17/44 65 107 0.59 69.9 76.5 73.4 4.8
overall 50/130 8 107 68.4 78.6 72.81
RM-1 low 26 7 18 0.2 72.0 74.3 72.8 2.8
med 16 20 57 0.2 72.3 73.3 72.8 2.8
high 22 61 111 6.2 72.2 72.7 72.5 2.8
overall 64 7 111 72.0 74.3 72.68
C-3h i i i i i i 73 25
C-5 low 3 14 14 1.3 71.5 72.9 72.0 10
med 6 35 57 1.3 68.5 74.8 71.3 10
high 0
overall 9 14 57 68.5 74.8 71.52
T-l j 1 j j na na na 72.5 1
Q-4 j 10 j j i i i 67.6 i
R-2k low 2 12 17 i 68 69 69 17
med 2 32 42 i 64 71 67 17
high 0
overall 4 12 42 64 71 68.0
weighted meanp 72.22
a

See table 1.

b

See “Procedures and Instructions” in Appendix.

c

The error is not the same for all measurements; the median is given.

d

The average is badly biased by the poor accuracy of the low-power values. If these are disregarded as outliers the average value is 11.45.

e

All voltages in this group were reported as peak-to-peak and converted to rms by the pilot laboratory.

f

The estimated errors were not symmetrical, that is, the positive values were not quite equal to the negative.

g

For RF-4, the first number, e.g., 9, is the number of independent groups into which the second number, e.g., 20, which is the number of measurements, is divided. Furthermore, each of the (say) 20 measurements is the average of 4 power readings at the same voltage, two taken as the voltage was switched on, and two as it was switched off.

h

System described as “relatively unproven.”

i

Not given.

j

Not given but very low.

k

According to the investigator, “These results are included for the record but are not regarded as part of the principal measurements,”

m

Same as (d) but average is 12.25.

n

The investigator considers that the data form two independent groups, here designated–.1 and–.2. He suspects that transducer no, 3-18 suffered a physical change between runs.

o

Same as (d) but average is 84.4.

p

Calculated from the “overall” values and their ranges by the method given in the text.

In these tables, Uv is the fractional uncertainty in the measurement of input voltage, and Ub is that in the radiation conductance, taking into account that in the voltage. See the Appendix far details. Each investigator, acting in accordance with instructions, estimated these uncertainties by his own methods, so that the several laboratories have not reported on a common basis.

Inspection of tables 2, 3, 4, and 5 reveals that by and large, there is rather remarkable agreement in the measurement of ultrasonic beam power at the power levels specified among the several laboratories and methods (or cases)3. In order to see this better, it is convenient to look at the departure of each result from some sort of grand average of all results. The question is how to weight the individual averages of the radiation conductance G¯r, in the calculation of the grand average, G¯¯r. Our first thought was to use weights inversely proportional to the estimated errors, hut we discarded this because, as has been pointed out, the reported errors are not comparable. We eventually decided to use purely statistical weights, that is, weights equivalent to the reciprocal variances. However, it was easier to use an approximation to the variance calculated from the range (max-min in tables 2, 3, 4, and 5) and the mean value of the ratio of the range to the standard deviation (square root of the variance; this ratio depends on the number of data. Values of H are given in table 2.4.1 of Snedecor [4]. No deep significance is attached to the grand averages, G¯¯r; we present them as being perhaps good enough for the present purpose. Data for which the ranges were not given were not used in the calculation of the grand average, nor were those averages used which were calculated from fewer than three data. The deviations from the grand average were given in table 6.

Table 6.

Deviation of radiation conductance from grand average.

Codea Powerb Range No. ofc Data Deviation, (G¯rG¯¯r)/G¯¯r in percent

3-16
3-18
3-20
3-22
Min
Max
Avg.
Min
Max
Avg.
Min
Max
Avg.
Min
Max
Avg.
RD-1 low 5,5,5,5 −21 4.9 −3.8 −28 −10 −19 −30 −4.5 −17 −4.5 34 3.8
med 5,5,5,5 0.5 4.0 1.4 −0.2 0.7 0.7 11 15 11 6.6 11 8.0
high 5,5,5,5 4.9 6.6 5.8 5.2 7.0 −6.1 −1.6 −1.6 −1.6 −4.3 −4.3 −4.3
overall 15,15,15,15 −21 6.6 0.9 −28 7.0 −4.1 −30 15 −2.5 −4.5 34 2.7
RD-2 low 3,3,0,0 2.3 3.1 3.1 6.1 7.0 7.0
med 12,12,6,6 −2.1 4.0 0.5 0.7 6.1 3.4 −0.3 3.8 1.4 −2.9 0.8 −0.2
high 0,0,0,0
overall 15,15,6,6 −2.1 4.0 1.2 0.7 7.0 4.0 −3.0 3.8 1.4 −2.9 0.8 −0.2
RD-3 low 10,10,10,10 118 237 161 47 287 182 −22 58 34 −12 75 66
med 10,10,10,10 −9.1 10.1 −4.7 13 26 19 13 28 17 4.7 29 12
high 10,10,10,10 0.5 10.1 4.9 0.7 3.4 1.6 2.7 6.4 4.9 19 24 22
overall 30,30,30,30 −9.1 237 54d 0.7 287 68m −22 58 18 −12 75 33
RD-4 low 1,1,1.1 −19 −19 −19 −8.3 −8.3 .3 8.5 8.5 8.5 −0.7 −0.7 −0.7
med 2,2,2,2 −5.6 2.3 −2.1 5.6 2.5 1 −3.7 2.4 −0.6 −8.6 −0.6 −4.6
high 0,0,0,0
overall 3,3,3,3 −19 2.3 −7.3 −8.3 2.5 −3.8 −3.7 8.5 2.4 −8.6 −0.6 −3.4
RF-1 low 6,7,9,12 −7.3 −2.7 −4.7 −10 −2.9 −6.5 −3.9 2.2 −3.0 −1.4 −1.4 −1.4
med 6,7,11,15 −4.7 −2.1 −3.8 −10 −3.8 −7.4 −2.6 0.4 −1.9 −0.9 0.5 −0.7
high 7,7,9,13 −3.0 3.1 −0.3 −5.6 −0.2 −3.8 −0.2 3.2 0.5 1.1 1.1 1.1
overall 19,21,29,40 −7.3 3.1 −2.7 −10 −0.2 −6.0 −3.9 3.2 −1.5 −1.4 1.1 −0.3
RF-3 low 3,3,0,0 2.3 3.1 2.3 −1.1 1.6 0.7
med 6,6,12,6 4.0 4.9 4.0 2.5 3.4 2.5 −1.2 4.9 2.7 0.7 3.2 1.9
high 0,0,6,3 −1.3 −0.5 −0.9 0.2 0.4 0.4
overall 9,9,18,9 2.3 4.9 3.7 −1.1 3.4 2.1 −1.3 4.9 1.4 0.2 3.2 1.4
RF-4 low 20,0,24,44 −10 −2.1 −4.7 −5.7 3.8 −1.0 −5.3 8.8 0.7
med 20,0,26,42 −9.1 −1.2 −4.7 −4.2 2.1 −1.2 −3.9 4.0 0.1
high 20,0,26,44 −8.2 −2.1 −4.7 −3.3 2.5 0.1 −3.2 5.9 1.6
overall 60,0,76,130 −10 −1.2 −5.0 −5.7 3.8 −0.6 −5.3 8.8 0.8
RF-4.1n low 0,26,0,0 −0.7 12 5.2
med 0,26,0,0 −0.2 5.2 2.5
high 0,22,0,0 6.1 9.7 7.0
overall 0,74,0,0 −0.2 12 4.8
RF-4.2n low 0,20,0,0 −7.4 −1.1 −3.8
med 0,20,0,0 −7.4 −2.0 −4.7
high 0,20,0,0 −6.5 −2.0 −3.8
overall 0,60,0,0 −7.4 −1.1 −4.2
RM-1 low 34,32,25,26 −2.1 0.5 −0.3 −1.1 1.6 0.7 −1.2 1.0 0.4 −0.3 2.9 0.8
med 15,13,16,16 −1.2 −1.2 −1.2 −1.1 0.7 −0.2 0.2 0.8 0.5 0.1 1.5 0.8
high 17,19,17,22 −1.2 −1.2 −1.2 −1.1 2.5 0.7 0.1 1.2 0.8 0.0 0.7 0.4
overall 66,64,58,64 −2.1 0.5 −0.9 −1.1 2.5 0.4 −1.2 1.2 0.5 −0.3 2.9 0.6
C-3 h i,i,i,i, i i −3.0 i i 1.1
C-5 low 3,3,3,3 −1.2 7.5 3.1 1.6 7.0 3.4 2.0 7.9 5.9 −1.0 0.9 −0.3
med 3,3,6,6 3.1 8.4 5.8 −2.0 9.7 3.4 −5.2 0.8 −0.9 −5.2 3.6 −1.3
high 0,0,0,0
overall 6,6,9,9 −1.2 8.4 4.4 −2.0 9.7 3.4 −5.2 7.9 1.4 −5.2 3.6 −1.0
0-1 low 0,0,0,0
med 9,9,0,0 −7.3 5.8 −1.2 −2.9 7.0 2.5
high 5,6,0,0 −4.7 2.3 −1.2 −2.9 3.4 1.6
overall 14,15,0,0 −7.3 5.8 −1.3 −2.9 7.0 1.9
T-l 0,0,1,1 na na −0.3 na na 0.4
Q-4 10,10,10,10 i i −7.7 i i −7.9 i i −8.0 i i −6.4
R-2k low 6,0,6,2 −4.7 0.5 −3.0 −4.5 −1.6 −3.0 −5.8 −4.5 −4.5
med 0,0,0,2 −11 −1.7 −7.2
high 0,0,0,0
overall 6,0,6,4 −4.7 0.5 −3.0 −4.5 −1.6 −3.0 −11 −1.7 −5.8
R-2.1 low 0,2,0,0 −4.7 −2.0 −3.8
med 0,0,0,0
high 0,0,0,0
overall 0,2,0,0 −4.7 −2.0 −3.8
R-2.2 low 0,1,0,0 8.8 8.8 8.8
med 0,0,0,0
high 0,0,0,0
overall 0,1,0,0 8.8 8.8 8.8
a,b,k,h, and n

See footnotes to tables 2,3,4, and 5.

c

For 3–16, 18, 20, and 22 respectively.

d

This average is badly biased by the poor accuracy of the low-power values. If these are disregarded as outliers, the value becomes −0.1%.

m

Same as d but value becomes 10%.

o

Same as d but value is 17%.

We can consolidate the material given in table 6 in the form of table 7, in which the number of oases where the absolute value of the deviation, |(G¯rG¯¯r)/G¯¯r| is less than various amounts (in percent). A method of measurement by a particular laboratory forms a case, and the results are presented separately for the two nominal frequencies; the values for transducers 3–16 and 3–18 are lumped in one column and those for 3–20 and 3–22 in the other. We see that at 2 MHz the deviations are less than 1 percent in 11 percent of the cases and less than 3 percent in 33 percent of the cases. For 5 MHz the peak is sharper; the values are 40 and 80 percent of the cases, respectively. At both frequencies, more than 90 percent of the cases have G¯r within 9 percent of G¯¯r and if one case (RD-3) were ignored, the figure would be 100 percent.

Table 7.

Cumulative distribution of the deviations of tables 2, 3, 4, and 5.

No. of cases for which (G¯rG¯¯r)/G¯¯r<Δ
Frequency, MHz 2 5

Δ
No.
Fraction of total
No.
Fraction of total
% % %
1 3 11 10 40
2 6 22 17 68
3 9 33 20 80
4 14 52 21 84
5 19 70 21 84
6 20 74 22 88
7 21 78 23 92
8 24 89 23 92
9 25 93 23 92
27 100 25 100

We can only speculate as to why the data are better at 5 than at 2 MHz. Perhaps the increased absorption at 5 MHz eases the requirements on the anechoic materials, the performance of which is almost always less than one might hope.

It is worth noting that in each ease except one (RD-3) the deviation of the overall average for a case from the grand average is less than the experimenter’s estimate of the error. This would mean little if in most of the cases the results were all too high or all too low. But in consideration of the diversity of the methods employed, this is not likely, and it would seem that most of the experimenters have made conservative estimates of their errors.

The results are, on the whole, gratifying. They lead us to believe that one can really measure with acceptable accuracy the total cw sound power output of transducers in the frequency range 2–5 MHz and in the power range 2.5–2500 mW. Unfortunately, the same conclusion cannot be-extended into the fractional milliwatt range that is important to applications (such as medical diagnosis) characterized by high peak but low average power. The prospects for extending agreement to microwatt levels of average power are probably good.

4. Stability

As shown in table 3 and its footnote n, one of the participants noticed a substantial change, between runs, of the value of Gr for transducer 3–18. Extensive measurements on this transducer were made upon its return to the pilot Laboratory, but no significant changes from the pristine value were detected even after a three-day submersion of the operating face in water.

We have had a similar experience ourselves. A transducer of nominally identical construction as 3–18 (although a 3- rather than a 2-MHz unit) exhibited a large shift in the measured Gr. Upon autopsy it was found to have suffered a small crack in the epoxy seal and therefore a small leak. Upon being dried out and recemented the transducer yielded its original Gr.

One could speculate that a small leak occurred while transducer 3–18 was at the participant’s laboratory and that on the way back to the pilot laboratory the transducer dried out. Sixteen months later, when the last measurements were made, the transducer was still behaving properly, at least under the pilot laboratory conditions. While we were at it, we took additional measurements on the other transducers as well. All of the measurements made by the modulation radiation- pressure method at the pilot laboratory are summarized in table 8.

Table 8.

Summary of radiation conductance, Gr.

2 MHz
5 MHz
Transducer 3–16 3–18 3–20 3–22
n (over about 1233 days) 87 85 79 85
G¯¯r (Grand average) μS 11.352 11.150 70.42 72.36
Standard deviation
μS 0.075 0.106 0.21 0.40
% 0.51 0.95 0.31 0.55

From the material in table 8, we could conclude that whatever drifts occur in either the transducers or the apparatus itself are of no great consequence. However, close examination, using standard statistical tests, shows that the disparities are not entirely random. Indeed, the regressions of Gr on time show trends which are significant although not overwhelmingly so. To elucidate this question will require a carefully planned experimental design carried out over several years, and this is under serious consideration.

6. Appendix

International Intercomparison of cw Ultrasonic Beam Power Measurements — Procedures and Instructions

National Bureau of Standards Washington, DC 20234

This document provides instructions and a typical procedure for the intercomparison of ultrasonic beam power measurements. The intercomparison is being conducted by circulating to participating laboratories four transfer standards, namely, half-wave resonant, air-backed, quartz transducers.

6.1. Equipment Supplied by NBS

  1. Transfer standards. The resonance frequencies of the transducers have been measured at NBS.

  2. Matching circuit. A matching circuit may be needed to provide adequate voltage to the quartz transducers, which have a high electrical input impedance. It also helps to minimize distortion in the voltage applied to the transducers.

  3. Connectors. Adapters are supplied for use with the transducers.

6.2. Care of the Standard Transducers

  1. Avoid any severe mechanical shock.

  2. Voltage limitations on the transducers are 3 V unloaded, and 350 V water loaded. Transducer voltage should be monitored at all times when connected to a power amplifier. Experimental procedures should be used that will ensure that the test voltage is not applied when the transducer is out of water and that the voltage is removed before removing the transducer from the water.

  3. The transducer connectors are not waterproof, and should not be submerged.

  4. The front faces of the transducers are gold plated. They should be carefully wiped with lens tissue to remove water. Avoid excessive force.

6.3. Measurements To Be Made

The quantity to be measured is the total cw beam power radiated by a transducer (the supplied transfer standard) into a reflectionless water load at a specified temperature, frequency, and sinusoidal input voltage to the transducer. At the power levels to be used in the intercomparison, the beam power is proportional to the square of the applied voltage. Each participant is responsible for obtaining a suitable rf voltmeter and having it calibrated at thé voltages and frequencies required. The following table indicates the specified frequency, fs, and the specified voltage, Vs, at which each measurement is to be made. The actual frequency, fm, of the measurement should be within ±0.02% of fs. The measured voltage, Vm, should be within ±50% of Vs (subject to the 350 V maximum limitation). The voltages are arbitrary, but have been chosen to cover the range of interest.

6.4. Reporting

Please use the reporting form included herewith. Report the test data in the sequence in which the data were taken.

The pilot laboratory will accumulate and summarize the results from all participating laboratories. As discussed in the following paragraphs, numerical results from each participant should be reported in a manner to allow comparison on a common basis.

An error can result from attenuation in the water coupling medium between the transducer and the receiver target. The radiated beam power, Pr, equals the received power measured by the participant’s equipment, Pm, plus the power loss in the coupling medium. This loss should be determined by measurement or calculation, and applied as a correction as indicated below. To correct for the attenuation, the following formula may be used:

Pr=Pmexp44×1017fm2d,

where d is the water path distance in centimeters between the transducer and the point at which the power measurement is made, and fm is the measurement frequency in Hz.

The temperature of the water coupling medium should be as close as possible to 23 °C, and in the range of 20 °C to 26 °C. A correction to a common-basis temperature and voltage should be applied as follows:

Pb=Pr(Vs/Vm)2[1+0.0016(Tm23)],

where Pb is the “common basis power” that would be measured at 23 °C with the common basis voltage, Vs, applied to the transducer, Vm is the measured applied voltage, Tm is the measurement temperature in xC. The coefficient 0.0016 is related to the temperature dependence of the ϱc of water.

So that the fractional uncertainty, Ub, associated with the common-basis power is properly determined at each test point, the participant should estimate the fractional uncertainty, Ur, in the radiated power, taking into account the uncertainty in measuring or calculating the loss in the coupling medium. For example, if Pr is 0.137 watts and the uncertainty in Pr is ±0.004 watts, then Ur = ±0.004/0.137 = ±0.03, or ±3%. The common- basis power uncertainty, Ub, is then calculated as

Ub=Ur+2Uv

where Uv is the fractional uncertainty in the voltmeter calibration and the factor 2 is a result of the square-law relationship between voltage and power. Uncertainties associated with the application of the temperature correction will be negligible if temperatures are kept within the specified range. The estimate of Ur should be based on the participant’s experience with his equipment; it should not be influenced by the variations observed in the present intercomparison.

Provide a brief description of each set of equipment and method of measurement used in the intercomparison. If more than one method of measurement is used, please report each on a separate data sheet.

If the quantities Vm and Pm are inappropriate to the method of measurement used, then please supply at least the following information for each measurement: temperature of the measurement, Tm; frequency of the measurement, fm; the radiation conductance, Gr, measured for the transducer; and the estimated uncertainty, Ub, of the measured Gr.

6.5. Precautions

The transducer must be oriented so that the entire ultrasonic beam will be received by the participant’s measuring equipment. Reflections that would cause ultrasonic energy to return to the transducer must be reduced to a negligible level to prevent interference at the face of the transducer that may change the characteristics of the transfer standard.

Care should be exercised to avoid the presence or formation of bubbles on the transducer face during test. The use of degassed water may eliminate this problem.

6.6. Typical Procedure

The following information is included but may be disregarded if not applicable:

Typical equipment provided by participants

  1. Voltmeter. A radio-frequency voltmeter is needed to measure the voltage applied to the transducers. The voltmeter must be calibrated at the frequencies and voltages listed under Measurements to be made. The voltmeter should be calibrated with a short cable. This will minimize loading on the matching network. This cable then becomes part of the calibrated voltmeter. A calibration uncertainty of ±0.25% or better is desirable.

  2. Power amplifier. A power amplifier capable of sup plying 5 watts into a 50-ohm load should be adequate.

  3. Signal generator. A stable generator must drive the power amplifier with a cw sinusoidal waveform of low distortion (preferably less than 0.5%).

  4. Frequency counter. This is needed to set the frequency of the signal applied to the transducer.

Typical procedure

  1. Interconnect equipment as indicated in figure A-1, but do not turn on the signal generator and power amplifier before first setting the output level controls to minimum. NOTE: CONNECTORS ON THE TRANSDUCERS ARE NOT WATERPROOF AND MUST BE PROTECTED IF IMMERSED IN WATER.

  2. Prepare the measuring equipment for measurements of power at a temperature of 23 °C ±3°.

  3. Place the NBS transducer in the water coupling port of the measuring equipment. Brush away any bubbles adhering to the face of the transducer. CAUTION — DO NOT APPLY VOLTAGE TO TRANSDUCER UNLESS THE GOLD TRANSDUCER FACE IS IN THE WATER (i.e., THE TRANSDUCER IS LOADED). APPLICATION OF HIGH VOLTAGE TO AN UNLOADED QUARTZ TRANSDUCER (i.e., IN AIR) MAY DAMAGE IT.

  4. Set the signal generator controls to obtain a cw sinusoidal waveform. Set the frequency controls of the signal generator approximately to the specified resonance frequency (loaded) of the NBS transducer. Adjust the measurement frequency, fm, to within ±0.02% of the specified resonance frequency fs, of the transducer using the frequency counter as an indicator. Frequency should be monitored continuously during the test and adjusted, whenever necessary.

  5. Set the voltmeter range switch to a suitable low voltage range. Increase the signal generator output to obtain a reading of perhaps 3 to 10 volts. Adjust the controls in the matching-circuit to obtain maximum voltage, being careful not to exceed 3 volts unloaded to 350 volts water loaded (if necessary reduce the signal from the generator).

  6. The signal can now be increased to a desired test level. Table A-1 gives the test points (in terms of frequency and voltage) at which measurements are to be made by all participants. The number of measurements to be made at each point is not specified, although a number from 3 to 10 would seem practical.

Figure A-1.

Figure A-1.

Example of possible test assembly.

Table A-1.
Transducer No. Frequency fs, MHz Specified voltages, Vs
3–16 1.995 30–100–300
3–18 1.995 30–100–300
3–20 5.0457 12–40–125
3–22 5.0148 12–40–125

Footnotes

1

Figures in brackets denote literature references at the end of this paper.

2

The modifications, which were rather extensive, were made by Franklin R. Breckenridge and Carl E. Tschiegg.

3

Some of the methods used by the participating- laboratories, e.g,, steady radiation pressure and calorimetric, are in principal capable of measuring average pulsed power. In some applications, such as medical imaging, such average powers might be much lower than those used in these tests and the conclusion would not apply.

5. References

  • [1].Zapf Thomas L. Calibration of quartz transducers as ultrasonic power standards by an electrical method. IEEE Ultrasonics Symposium Proceedings: Cat. # 74 CHO ISU; 1974; 45–49. [Google Scholar]
  • [2].Greenspan M., Breckenridge F.R., Tschiegg C.E. Ultrasonic transducer power output by modulated radiation pressure. J. Acoust. Soc. Am. 63(4): 1031–1038; 1978. [Google Scholar]
  • [3].Zapf T.L., Harvey M.E., Larsen N.T., Stoltenberg R.T, Ultrasonic calorimeter for beam power measurements. Natl. Bur. Stand. (U.S.) Tech. Note 686; 1976. September 36 p. [Google Scholar]
  • [4].Snedecor George W. Statistical methods, 6 ed. Ames: Iowa State University Press; 1967. 593 p. [Google Scholar]

Articles from Journal of Research of the National Bureau of Standards are provided here courtesy of National Institute of Standards and Technology