Abstract
We describe in this work an advanced microfluidic chip for the capture of bioanalyte on the surface of droplets arranged in a dense array. We show the procedure for generating, functionalizing, and arranging the droplets inside the device for capturing a specific bioanalyte. Then, we demonstrate the capacity of the array to capture analyte from a cross-flowing liquid, using a biotin/streptavidin model. The paper also proposes to use the droplets array, after integration with acoustic detection, as a regenerable detection interface for bioanalyte sensing. We model the arrangement of droplet in dense array and show that they present a larger effective capture surface and shorter capture distance than standard flat surface biosensor of the same footprint. As the droplets can be easily evacuated and replaced inside the device analysis chamber, the proposed biosensor would allow biointerface regeneration and chain measurement without dismounting the device.
I. INTRODUCTION
Since its inception in the 1990s, the research on microfluidic systems has allowed creating devices for biological detection with regular improvement in compactness, functionality integration, and quantity of a biological sample, leading to the concept of lab-on-chip. These developments have resulted in dramatic changes in the biomedical field, opening the possibility to perform multiomics analysis (genomic, proteomic, etc.), to prepare patient targeted drugs (Advanced Therapy Medicinal Product), and to study cellular interactions or even organs, using organ-on-chip approach. An ultimate reduction in the sample volume has been reached by using droplet as a reaction chamber. This approach also offers an increase in speed of analysis by allowing large scale multiplexing, particularly interesting for drug research.
A particularly vibrant research direction in microfluidics focuses on the development of label-free biosensors for helping point-of-care diagnosis. These sensors are aimed at protein quantification or antibody detection without the need for attaching fluorescent or magnetic labels.3 Their principle of operation may be split into 5 elementary functions as shown in Fig. 1. Firstly, they require the preparation of a surface covered with bioreceptor targeting the bioanalyte of interest. We will call this operation the functionalization of the surface. Then, we need the bioanalyte carrying fluid to circulate close to the functionalized surface for increasing the probability of capture. Actually, the third function is the capture of the bioanalyte with the bioreceptors on the surface. Usually, captures with high specificity are required, and selective interactions have to be targeted, like antibody/antigen or ligand/protein bindings. Finally, to complete the sensor, we need a mechanism that could be used to measure the presence (or even the quantity) of bound bioanalyte on the surface. This can be accomplished using optical,21,23,36 electric,10 electrochemical,2 or acoustic transducers,14,20 that should ultimately deliver an electrical signal proportional to the quantity of bound bioanalyte on the surface.
FIG. 1.
Principle of operation of a biosensor with regenerable functionalized planar surface.
The main weakness of theses sensors lies in their need to replace the biodetection interface for performing a new measurement, which is obtained with a partially or totally disposable detection chamber. Then, cost issue does not allow one to integrate with the chamber some of the most advanced detection scheme, like acoustic probing. To circumvent this issue, some groups have recently tried to regenerate the bioreceptor layer on the detection interface using different chemical treatments.5 Alternatively, it has been proposed to use functionalized magnetic microparticles that are immobilized close to the transducer using a strong magnetic field.9 After capture on the particle surface and detection, the magnetic flux is removed, and the particles are flushed away. In this approach, the integration with an acoustic transducer is difficult, and the electrochemical sensing used presents relatively low sensitivity. Moreover, the small magnetic particles are mixed with the biological material, which may pose regulatory or safety issue.
We propose to develop a new concept of biosensor where the functionalized surface is no more planar but is composed by the surface of an array of functionalized droplets. This architecture has an obvious interest in simplifying the reusability issue raised above: it is sufficient to flush away the functionalized array and to replace it with a new array of droplet, potentially with a different functionalization, to reuse the biosensor. Moreover, we will show that this architecture increases both the effective surface and the efficiency of capture of the bioanalyte in the tested sample. We also suggest that this architecture should use acoustic transduction for allowing to probe the droplet surface dynamically and measure with high sensitivity the quantity of bioanalyte captured in the array.6,29,32,37 A summary of the different functions that would need to be integrated for such a biosensor are illustrated in Fig. 2.
FIG. 2.
Functions required for a droplet based biosensor with bioanalyte capture and regeneration.
In this paper, we only investigate the microfluidics aspects of the biosensor, detailing the strategy to generate functionalized droplets and arrange them in a regular array and then capture bioanalyte in the array and finally flush the droplet array for restarting from an empty chamber (the functions 1, 2, 3, 4, and 6 in Fig. 2). The acoustic detection (function 5) is still an ongoing work.
II. GENERATION OF FUNCTIONALIZED DROPLETS
The generation of monodisperse droplets in microfluidics is obtained by simultaneously injecting two immiscible fluids inside different microchannel structures. For this sensor, the bioanalyte being generally carried in water solution, we will need a continuous phase of water with droplets of oil. Hydrodynamics based generation uses two main configurations of channels: the T-junction35 and the coflowing junction.15 In these cases, the drag forces appearing in the zone where the two liquids mix, induce the breakup of the liquid stream and the generation of droplets. For helping the generation and stabilizing the generated emulsion, surfactants are generally mixed with the fluids before injection in the microchannels junction.
For the fabrication, we are using a well controlled micromachining process based on silicon and glass wafer assembly, as shown in Fig. 3.
FIG. 3.
(Left) Flowchart for fabrication using silicon/glass process. (Right) Zoomed view of a microfluidic chip with the generation of 80 m biotinylated soja oil droplets in DI water with a T-junction in dripping regime.
We are first patterning with S1813 photoresist (Fig. 3-1) and etching with a SPTS Rapier DRIE system the channels in a silicon wafer front side (Fig. 3-2). Then, we pattern with thick AZ9260 photoresist (Fig. 3-3) and etch from the silicon wafer backside the through holes for liquids access (Fig. 3-4). The channels on the silicon wafer are then closed by anodic bonding with a BF33 1 mm thick glass plate (Fig. 3-5). For increasing wetting of water with the silicon channels without changing wetting of oil, RCA1 solution (HO:NHOH:HO 2:1:1) is used to fill the channels by capillarity and rinsed 3 times with deionized (DI) water in the same manner. This process yields channels with controllable surface energy and low roughness, allowing to use the T-junction efficiently for the generation of oil droplet in water with the diameter ranging between 5 m and 150 m.33,34
However, if the generation of oil droplets is mastered, the ability to obtain functionalized oil droplets remains challenging. Actually, the standard techniques used for the functionalization of solid surfaces often makes the water/oil emulsion unstable, inducing coalescence of the droplets. In the literature we find only a handful of work dealing with droplets functionalization, with two main strategies:
-
•
either the droplet is generated using functionalized surfactant that will place itself at the oil/water interface. This can be obtained by using biotinylated phospholipids7,17,25 or using modified surfactants with different chemical terminations (amine, carboxylic acid, biotin, etc.) or22
-
•
the surfactant at the droplet surface is functionalized after the droplet is formed. This is obtained directly after activation of fatty acid13 or condensation of silanes at the droplet surface.1 Alternatively, an indirect strategy has also been demonstrated, which consists in decorating the droplet with gold nanoparticles27 allowing to later use standard thiol chemistry for the functionalization.
In order to demonstrate the capacity of the array of droplet to capture bioanalyte, we will be using the standard and very robust biotin/streptavidin interaction.11 Accordingly, we have tested both strategies to decorate the oil droplets with biotin. The second strategy is more versatile, but our trial, inspired by a protocol proposed by Fattaccioli group,13 did not yield droplets stable enough for forming droplet array. Thus, we based the generation of droplets on the first strategy, using biotinylated soy oil and DI water with Tween20 at 5 v/v% as shown in Fig. 3-right. The droplets generated in the T-junction could have any diameter between 58 m and 84 m by controlling oil and water flow rate as shown in Table I. This range of flow rate allows remaining in the dripping regime of generation, yielding high generation rate of small droplets with a low diameter variation.
TABLE I.
Parameter for biotinylated soja oil droplet generation.
| Parameter | Symbol | Value |
|---|---|---|
| Oil channel width | Wd | 20 μm |
| Water channel width | Wc | 100 μm |
| Channel depth | h | 40 μm |
| Oil flow rate | Qd | 3–4 μl/min |
| Water flow rate | Qc | 5–30 μl/min |
III. ARRANGEMENT OF DROPLETS IN ARRAY
Arrangement of soap bubbles in large and closely packed arrays at the free surface of an open reservoir has been demonstrated in the past, most famously in 1947 by Sir L. Bragg for studying crystals.8 For droplets generated in close channels, the need to drain the interstitial liquid for forming a closely packed array poses new problems. In fact, the generation of droplet by hydrodynamic forces in channel junctions16,35 requires a certain ratio of flow rate between the dispersed and the continuous phases that is not necessarily compatible with the production of a closely packed array. Accordingly, in the literature, we find essentially two strategies to overcome this issue: either use a large phase fraction 19,24 or drain the continuous phase from the array after generation.
The second strategy has the advantage to be mostly independent of the droplet generation mechanism and the ratio. It is based on the use of narrow lateral channels for draining the continuous phase interstitial liquid after generation, while at the same time retain the droplets in the main channel or chamber.18,26,28,34,38 We propose to use this strategy in our sensor for obtaining a closely packed array of monodisperse droplets, and we will thus combine a T-junction droplet generator with a chamber equipped with sieves and drain, as shown in the Fig. 4.
FIG. 4.
Layout of the chip with 8 standard ports (port 2 is unused) for the generation of oil droplets and their organization in the closely packed array.
The input ports 3 and 4 are used for generating the droplets, while the other ports are mostly used as purging channels at this step (ports 1 and 6 will be used for flowing the bioanalyte across the droplets array in Sec. IV).
A computerized control system allows one to propose a systematic method for filling the chamber with densely packed droplets, as shown in Fig. 5. In the first step, we stabilize the generation of droplet by using only the port 3 (oil), 4 (water), and 5 (dump) as in Sec. II. Then, we start filling the chamber using only the central channel dump (port 7). In the third step, we open the drains (ports 6 and 8) for densifying the droplet array by removing interstitial water. At the end of this step, the droplet gets compressed and deformed because of the continuous flow coming from the droplet generator. In the last step, we stop the droplet generator and the array relaxes back, letting the droplets get their round shape, and finally fully filling the chamber.
FIG. 5.
Chip ports control sequence for organizing droplets in a densely packed array in the detection chamber.
The flow behavior during this process has been partially simulated using COMSOL Multiphysics FEM simulation software and is presented in Fig. 6. In this chamber much more wide (3 mm) than deep (40 m), the behavior is close to Hele-Shaw flow and a 2D analysis is sufficient. Moreover, the chamber is symmetric and we simulated only half the chamber with one drain to decrease the computing time. We clearly see here that when the drains are closed (step 2), the liquid bypasses the chamber by the side, slowing down in the central section, which helps densifying the droplet array. However, the liquid is reinjected at the end of the chamber, finally lifting the droplet from the chamber side and disorganizing the droplet array as we observed previously. When we open the drain dump in step 3, we remove this issue and slow down the liquid until the end of the chamber allowing the droplets to form the closely packed array.
FIG. 6.
2D simulation of flow rate in cm/s in the (half) chamber during step 2 and step 3 of Fig. 5 (water flow is from left to right). Inset: SEM view of the forked sieves end.
We had to introduce sieves between the chamber and the drain for preventing the droplets from flowing out of the chamber. The shape of the sieves is crucial for preventing the droplets from deforming and squeezing out of the chamber, while at the same time keep an acceptable fluidic resistance. We observed that 500 m-long and 20 m-wide channels34 did not work well, and we studied the possibility to use 10 m and 5 m-wide channels instead. However, they suffer from a dramatic increase in hydraulic resistance (even taking into account that more could be packed by unit length), respectively, by 500% and 2500%, as obtained from simulation from COMSOL MultiphysicsTM. Instead of using uniform narrower channels for the sieve, we choose to design new forked channels with narrow but short branches at their end (5 m wide branch over 25 m), as shown in the inset in Fig. 6. This architecture showed an increase of the fluidic resistance by only 57% but proved to be much more efficient at preventing the droplet from escaping the chamber than the original uniform 20 m-wide channels.
From the results shown in step 4 of Fig. 5, it is clear that the strategy we developed allows us to fill the chamber with a densely packed array of droplets, even if we could observe some limited coalescence of droplets.
We have also verified that we could empty the chamber completely, by flushing water through port 4 while port 7 was left open, and the other ports closed. The chamber was then refilled with droplet array using the same strategy, and this process could be repeated multiple times, without observing any degradation in the formed droplet array. This is one of the advantages of our architecture, where the regeneration of the capture surface (function 5 in Fig. 1) is easily obtained for a fully reusable sensor.
IV. FLOW THROUGH THE DROPLET ARRAY
For studying the flow through the droplet array, we injected water with fluorescein (1 mg/ml) in the chip (Fig. 7) through port 1 and left port 7 open, while all the other ports were closed (see Fig. 4).
FIG. 7.
Flow of fluorescent liquid into the droplet array (zones 1 and 5 correspond to the two drains, zones 2 and 4 to the two sieves, and zone 3 to the droplet array).
A sequence of the image was recorded during the injection of fluid at a low flowrate (0.5 l/min). In each image, we took the intensity along the two lines shown in Fig. 7 and stacked the lines progressively atop each other in a diagram with the time as the vertical axis. This kinogram is showing the fluorescence front progression as a function of time and the slope of the black/white frontier gives access to the front speed in the different zones of the chip.
The flow in the sieves and in the dense droplet array has similarities with the flow in a porous medium and the total flow rate is then described by Darcy’s equation
with the permeability (linked with the pore characteristics) and the medium section. We may use this equation to obtain the velocity of the fluid in the pores, by noticing that , where we have introduced the porosity, to account for the fact that in porous media only a fraction of the cross-section area allows liquid circulation. We have
This equation may be used to describe the flow in the zones 2, 3, and 4 of the chamber.
We note that the difference of fluid velocity that is typically observed in the two sieves (zones 2 and 4) that have the same geometry, and hence the same porosity and permeability, indicates that the fluid path is not uniform. In fact, the fluid concentrates in some part of the exit sieve, which is traced back to defects in the droplet array as seen in Fig. 8: a bigger droplet (coalescence) in the chamber right is blocking the flow through a few pores of the sieve and a few missing droplets in the left of the array lower the fluidic resistance in this region. These combined effects concentrate the flow in the central and left part of the chamber. In fact, we see that the flow velocity increases by about 50% between the input sieve (zone 2) and the output sieve (zone 4), meaning that only about 2/3 of the pore in the exit sieve participates to the flow.
FIG. 8.
Kinograms of the flow of the fluorescent liquid in the chip along (top) the central line and (bottom) the left line shown in Fig. 7 (numbered zones correspond to the zone shown in the figure).
The porosity of the sieves may be directly estimated from the geometry (a channel 20 m wide every 40 m) and is . Accordingly, considering that the flow rate and the section (that is, the depth and width of the chamber) are approximately kept constant between the sieve and chamber zones, we may write
where is the porosity of the array of the droplet in the chamber. Thus, by using velocity in the top sieve, we obtain the porosity of the droplet array as
In the droplet array, the pores (i.e., the space between the oil droplets) are fully wet and the array porosity is, in fact, equal to the water fraction , meaning that water makes only % of the volume and oil is making 82%.
When the droplets are constrained by the chamber top and bottom surfaces they do not assume a spherical shape. We still refer to their radius (), by using the apparent radius of their projected shape seen from the top, as they remain circular from that direction when they do not touch the side walls. In fact, if the droplet top diameter is larger than the depth of the chamber (), the droplet assumes normally a barrel-like shape as shown in the top-right inset in Fig. 9. It spreads (and may wet) the top and the bottom of the chamber and shows an apparent contact angle (), whose magnitude depends on the fluids interfacial tension and on the droplet volume. By varying the apparent contact angle between and for droplets with constant top diameter, we may model different barrel droplet shapes, as shown in the insets in Fig. 9.
FIG. 9.
Evolution of droplet array porosity as a function of the normalized droplet diameter and for different values of the apparent contact angle . We show in the figure sketch of typical droplets for different values of normalized diameter and contact angle [top-right inset: the sketch of the oil droplets arrangement and cross section of a droplet inside the chamber when the droplet diameter is larger than the chamber depth ()].
From the geometry of the cross section of the constrained droplet in the chamber shown inFig. 9, the volume of the droplet4 may be expressed with the parameters , , and only and it is given by
| (1) |
where the tore radius has been left in the equation to make it easier to read.
Considering the dense arrangement of the droplet, that defines a hexagonal symmetry as shown in Fig. 9, we compute the water fraction as
where we used the difference between the volume of the circumscribed isosceles prism (), grayed out in the inset of Fig. 9, with the inscribed volume of the oil droplet () to find the volume of water. We find that this dimensionless porosity depends only on and . We plotted the function in Fig. 9 and we see that it varies between 9.3% and 39.5% for and corresponding to the range of meaningful values for the droplet array model.
For the droplets of radius m in a chamber of depth m (that is, ), corresponding to the case presented in Fig. 7, the porosity may vary between 9.3% and 26.5%, when the contact angle varies between 90 and 180 . By solving numerically the equation, we find that in the droplet array considered (), we have (this point is indicated with a cross in Fig. 9). This figure is compatible with measurements of contact angle using sessile drop experiment made separately, and with simple geometry measurements (by measuring the droplet radius and the contact radius ) although these measurements are not very precise.
The droplet array is now characterized in detail (the determination of the permeability coefficient of the array is not possible in this experiment as we cannot measure the pressure inside the chip) and we will now model and demonstrate its capability in capturing bioanalytes.
V. CAPTURE OF BIOANALYTE ON DROPLET SURFACE
The dense arrangement of functionalized droplets for bioanalyte capture present multiple advantages compared to the functionalization of a chamber flat surface:
-
•
it makes regeneration of functionalized surface straightforward,
-
•
it increases the bionalyte capture surface for the same chamber footprint, and
-
•
it provides shorter distance for complete bioanalyte capture.
If the first point has already been established, the two other points require more investigation.
The surface of the oil droplet in contact with circulating water (thus excluding the two discs at top and bottom) is obtained by computing the surface integral using the geometry shown in Fig. 9 and we obtained after verification with MapleTM
In the special case where , the droplet shape goes from a cylinder to a sphere of the same radius when goes from to . We verify easily that the side surface is the same in both cases, as Archimedes already noticed in 225 BC.
The ratio of the surface of the droplet () and the footprint of the array of droplet () has been estimated from geometric consideration as
where is the depth of the chamber, the radius of the droplets, and the contact angle of the droplet at the channel top and bottom surface. When the surface ratio becomes larger than 1, it means that we have a larger capture surface than what could be obtained with a flat sensor of the same footprint. Accordingly, is considered as a gain of surface for droplet based bioanalyte capture, that could be even more important if droplets are stacked-up in multiple layers and not only in a single layer. We have plotted in Fig. 10 this surface gain for different droplet diameter/height ratio as a function of the normalized droplet diameter, and we see that it is normally around 2 and may exceed 3.5.
FIG. 10.
Ratio of bioanalyte capture surface for a droplet sensor and a flat sensor of the same occupancy as a function of the normalized droplet diameter and for different values of the apparent contact angle . The cross represents the typical case shown in Fig. 7.
For comparing the capacity to capture bioanalyte between the array of the droplet and the flat surface, we start by establishing the length required for capturing all the analyte in a channel. We follow a reasoning inspired by Squires et al.31 that is illustrated in Fig. 11. We will only consider here the transport phenomena (advection and diffusion) and assume that the capture process is the same in the two situations and do not limit the complete process (in microfluidics terms, it means that we assume we work with large Damkohler number31 ).
FIG. 11.
Advective and diffusive transport of bioanalyte for reaching the surface of capture from the channel top.
Inside the microfluidic chip, the fluid flows parallel to the channel surfaces while carrying the bionalyte (advective flow). At microscale, we have laminar flow without turbulences and the bioanalyte particles would normally flow parallel to the surface and would have no chance of being captured. Hopefully, the bioanalyte particles are small and they are subject to diffusion (diffusive transport) that will help drive them toward the surface of capture. We will estimate the typical distance it takes for capturing all the bioanalyte particles present in the fluid by studying their motion under simultaneous advective (i.e., due to carrier liquid flow) and diffusive(i.e., due to diffusion) flows. The advection time it takes to move the bioanalyte by a distance is given by
where is the volumetric flowrate of the liquid carrier and the area of the channel.
Because we want to estimate the distance for full capture we will consider the bioanalyte that is the hardest to capture. This bioanalyte enters the channel the farthest from the surface and has to diffuse by the full channel height before it may be captured by the surface. The diffusion time needed for this bioanalyte to finally reach the surface is estimated by using typical diffusion length expression
where is the diffusion constant of the bioanalyte, and that can be related to bioanalyte size, temperature, and fluid properties using the Stokes-Einstein relationship.12 Thus, the maximum distance required for capturing this farthest bioanalyte, and consequently all the bioanalyte in the channel is obtained when , giving
The application of this expression to the case of a chamber of width and depth with a plane functionalized surface is straightforward and we have
If the chamber is filled with droplets, the flow may be approximated as if it went through an array of parallel channels in the interstices between the droplets. The depth of each of these equivalent channels is given by the distance between droplets (assumed here to be uniform), while their equivalent width is given by the chamber depth and the flowrate in these equivalent interstitial channels is noted , giving
Again, here, for showing the interest of our structure, we will compare the capture length between these two cases:
If we impose the same volumetric flow rate for the chamber with plane functionalized surface and with functionalized droplets, it means that , where is the number of parallel channels in the droplet array. Accordingly, we find that
where we have used the fact that in the chamber, the droplets are touching each other and thus with the droplet diameter. For a typical case where the droplet diameter is 80 m, m and the channel is 40 m deep, we find . It means that, for the same flow rate, the capture distance is 20 times shorter in the droplet array than in the planar configuration, ensuring complete capture in a more compact device. In practice, the capture will be even more efficient as the path between the droplets is not straight but zigzagging, forcing the bioanalyte toward the functionalized surface.
For demonstrating the capability of the functionalized droplet array to capture bioanalyte, we will be using the streptavidin/biotin interaction. Actually, the streptavidin protein and the biotin coenzyme have a very high affinity and make a strong bond, unaffected by solvents, detergent, heat, etc.11 After injecting the streptavidin solution through the sieves in the biotinylated droplet array for 1 min, we have disrupted the array of droplets by sending a high flowrate through ports 1 and 6 and observed the behavior of the droplets using a camera. We observed that the droplets, instead of gliding on each other as it is usually observed, are actually sticking to fixed points. This behavior may be observed in Fig. 12 on the sequence of images extracted with Fiji30 from a video where the lower “arm” appears to rotate. We also pushed the biotinylated droplets still present in the smaller stabilization chamber after the T-junction, and verified that these droplets, which had not seen the streptavidin, had retained their original behavior and did not stick to each other. The real behavior is dynamic and is better understood by directly comparing the video of biotinylated droplets after streptavidin injection (see Video 1 in the supplementary material) with the video of biotinylated droplets without streptavidin (see Video 2 in the supplementary material).
FIG. 12.
(Top) Sequence of image extracted from a video showing biotinylated droplets sticking to each other (rotation of lower “arm”) (see Video 1 in the supplementary material) and (bottom) the formation of streptavidin bridges between biotinylated droplets.
The visible stickiness of the droplets when they move around is attributed to the bridging of droplets by streptavidin. Actually, streptavidin has 2 binding sites on each side of the molecule, which could produce biotin-streptavidin-biotin bridges between droplets. Such a bridge (inset Fig. 12) has already been observed by Fattaccioli et al.13 using biotin decorated droplet with streptavidin.
A quantitative technique (our fluorescent setup is unfortunately not sensitive enough) will be needed for quantifying precisely the capture efficiency of the functionalized droplets array. However, this experiment validates for the first time the capacity of such droplet array to capture bioanalyte.
VI. CONCLUSION
We have designed and tested a microfluidic chip for capturing bioanalyte on the functionalized surface of droplets. We also proposed that this structure could be used for a new type of biosensor with regeneration capability after the integration of acoustic probing.
In this paper, we have focused on designing and demonstrating the microfluidic functions of the biosensor. We successfully demonstrated the generation of functionalized droplets, their arrangement in a dense array, the injection of bioanalyte through this array, and, finally, the binding of the bioanalyte to the surface of the droplets. Moreover, we have presented a model showing the advantage of our sensor architecture compared to a standard planar sensor: firstly, we increase by a factor of 2 or more the capture surface and we decrease by a factor of 20 or more the capture length of the bioanalyte for the same flowrate.
We are currently developing an integration strategy, based on AlN deposition on SOI wafer, for placing piezoelectric excitation and sensing below the chamber. The test of this future chip will establish the sensitivity of our method and we will investigate its capabilities to fully capture bioanalyte of interest in small samples.
It should be noted that without any integrated detection, the architecture proposed here can already be used as a regenerable bioanalyte filter. In fact, the microfluidic functions described in this paper allow one to collect targeted bioanalyte in a sample and extract them by flushing the droplets for further processing.
SUPPLEMENTARY MATERIAL
See in the supplementary material a first video presenting the motion of biotinylated oil droplet in water after streptavidin injection (the center of the video corresponds to the sequence of images shown in Fig. 12), where the droplets seem to be sticking to each other because of streptavidin bridges. The second video is showing the motion of biotinylated oil droplets just after generation before exposure to streptavidin, where droplets can be seen to move freely and glide on each other.
ACKNOWLEDGMENTS
We want to thank Jacques Fattacciolli and Lorraine Montel at IPGG, Institut Curie, France, for sharing their experience on droplet functionalization and providing the sample of biotinylated soy oil.
This work was partly supported by the French RENATECH Network and its FEMTO-ST technological facility. We acknowledge support of the CNRS/Université Bourgogne Franche-Comté CLIPP platform for biochemistry work and analysis and financial support of the LabEx ACTION under Contract No. ANR-11-LABX-0001-01 and the EUR (EIPHI) under No. 17-EUR-0002. Computations have been partly performed on the supercomputer facilities of the Mésocentre de Calcul de Franche-Comté.
REFERENCES
- 1.Attia M. F., Anton N., Bouchaala R., Didier P., Arntz Y., Messaddeq N., Klymchenko A. S., Mély Y., and Vandamme T. F., RSC Adv. 5, 74353 (2015). 10.1039/C5RA12676B [DOI] [Google Scholar]
- 2.Bakker E. and Telting-Diaz M., Anal. Chem. 74, 2781 (2002). 10.1021/ac0202278 [DOI] [PubMed] [Google Scholar]
- 3.Berthier A., Elie-Caille C., Lesniewska E., Delage-Mourroux R., and Boireau W., J. Mol. Recognit. 24, 429 (2011). 10.1002/jmr.1106 [DOI] [PubMed] [Google Scholar]
- 4.Berthier J. and Brakke K. A., The Physics of Microdroplets (John Wiley & Sons Inc., Hoboken, NJ, 2012). [Google Scholar]
- 5.Bienaime A., Leblois T., Lucchi G., Blondeau-Patissier V., Ducoroy P., Boireau W., and Elie-Caille C., Int. J. Nanosci. 11, 1240018 (2012). 10.1142/S0219581X12400182 [DOI] [Google Scholar]
- 6.Bienaime A., Liu L., Elie-Caille C., and Leblois T., Eur. Phys. J. Appl. Phys. 57, 21003 (2012). 10.1051/epjap/2011110111 [DOI] [Google Scholar]
- 7.Blankenburg R., Meller P., Ringsdorf H., and Salesse C., Biochemistry 28, 8214–8221 (1989). 10.1021/bi00446a037 [DOI] [PubMed] [Google Scholar]
- 8.Bragg L. and Nye J. F., Proc. R. Soc. Lond. A 190, 474 (1947). 10.1098/rspa.1947.0089 [DOI] [Google Scholar]
- 9.Choi J.-W., Oh K. W., Thomas J. H., Heineman W. R., Brian Halsall H., Nevin J. H., Helmicki A. J., Thurman Hendersona H., and Ahna C. H., Lab Chip 2, 27 (2002). 10.1039/B107540N [DOI] [PubMed] [Google Scholar]
- 10.Daniels J. S. and Pourmand N., Electroanalysis 19, 1239 (2007). 10.1002/elan.200603855 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Darst S. A., Ahlers M., Meller P. H., Kubalek E. W., Blankenburg R., Ribi H. O., Ringsdorf H., and Kornberg R. D., Biophys. J. 59, 387 (1991). 10.1016/S0006-3495(91)82232-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Einstein A., Ann. Phys. 322, 549 (1905). 10.1002/andp.19053220806 [DOI] [Google Scholar]
- 13.Fattaccioli J., Baudry J., Henry N., Brochard-Wyart F., and Bibette J., Soft Matter 4, 2434 (2008). 10.1039/b806635c [DOI] [Google Scholar]
- 14.Fu Y., Luo J., Nguyen N., Walton A., Flewitt A., Zu X., Li Y., McHale G., Matthews A., Iborra E., Du H., and Milne W., Prog. Mater. Sci. 89, 31 (2017). 10.1016/j.pmatsci.2017.04.006 [DOI] [Google Scholar]
- 15.Ga nán-Calvo A. M. and Gordillo J. M., Phys. Rev. Lett. 87, 274501 (2001). 10.1103/PhysRevLett.87.274501 [DOI] [PubMed] [Google Scholar]
- 16.Garstecki P., Fuerstman M. J., Stone H. A., and Whitesides G. M., Lab Chip 6, 437 (2006). 10.1039/b510841a [DOI] [PubMed] [Google Scholar]
- 17.Gizeli E., Liley M., Lowe C. R., and Vogel H., Anal. Chem. 69, 4808–4813 (1997). 10.1021/ac970519m [DOI] [PubMed] [Google Scholar]
- 18.Haliburton J. R., Kim S. C., Clark I. C., Sperling R. A., Weitz D. A., and Abate A. R., Biomicrofluidics 11, 034111 (2017). 10.1063/1.4984035 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Hashimoto M., Mayers B., Garstecki P., and Whitesides G., Small 2, 1292 (2006). 10.1002/smll.200600211 [DOI] [PubMed] [Google Scholar]
- 20.Kardous F., Fissi L. E., Friedt J.-M., Bastien F., Boireau W., Yahiaoui R., Manceau J.-F., and Ballandras S., J. Appl. Phys. 109, 094701 (2011). 10.1063/1.3576113 [DOI] [Google Scholar]
- 21.Krupin O., Wang C., and Berini P., “Optical plasmonic biosensor for leukemia detection,” Proc. SPIE (published online). 10.1117/2.1201512.006268 [DOI]
- 22.Li M., Jiang W., Chen Z., Suryaprakash S., Lv S., Tang Z., Chen X., and Leong K. W., Lab Chip 17, 635 (2017). 10.1039/C7LC00079K [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Listenberger L. L. and Brown D. A., Current Protocols in Cell Biology (John Wiley & Sons, Inc., 2001). [Google Scholar]
- 24.Mary P., Abate A. R., Agresti J. J., and Weitz D. A., Biomicrofluidics 5, 024101 (2011). 10.1063/1.3576934 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.M’Barek K. B., “Adhésion et phagocytose de gouttes d’émulsions fonctionnalisées,” Ph.D. thesis (Université Pierre et Marie Curie—Paris VI, 2015). [Google Scholar]
- 26.Niu X., Gulati S., Edel J. B., and deMello A. J., Lab Chip 8, 1837 (2008). 10.1039/b813325e [DOI] [PubMed] [Google Scholar]
- 27.Platzman I., Janiesch J.-W., and Spatz J. P., J. Am. Chem. Soc. 135, 3339 (2013). 10.1021/ja311588c [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Prat L., Sarrazin F., Tasseli J., and Marty A., Microfluid. Nanofluid. 2, 271 (2006). 10.1007/s10404-005-0066-3 [DOI] [Google Scholar]
- 29.Raimbault V., Rebiére D., Dejous C., Guirardel M., and Conedera V., Sensors Actuat. A 142, 160 (2008). 10.1016/j.sna.2007.05.026 [DOI] [Google Scholar]
- 30.Schindelin J., Arganda-Carreras I., Frise E., Kaynig V., Longair M., Pietzsch T., Preibisch S., Rueden C., Saalfeld S., Schmid B., Tinevez J.-Y., White D. J., Hartenstein V., Eliceiri K., Tomancak P., and Cardona A., Nat. Methods 9, 676 (2012). 10.1038/nmeth.2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Squires T. M., Messinger R. J., and Manalis S. R., Nat. Biotechnol. 26, 417 (2008). 10.1038/nbt1388 [DOI] [PubMed] [Google Scholar]
- 32.Tarbague H., Lachaud J.-L., Destor S., Vellutini L., Pillot J.-P., Bennetau B., Pascal E., Moynet D., Mossalayi D., Rebière D., and Dejous C., J. Integr. Circuits Syst. 5, 125 (2010). [Google Scholar]
- 33.Tarchichi N., Chollet F., and Manceau J.-F., Micro Nanosyst. 6, 145 (2014). 10.2174/187640290603150112122537 [DOI] [Google Scholar]
- 34.Tarchichi, Nathalie, T., “Tunable diphasic microfluidic,” Ph.D. thesis (Université de Franche-Comté, 2013). [Google Scholar]
- 35.Thorsen T., Roberts R. W., Arnold F. H., and Quake S. R., Phys. Rev. Lett. 86, 4163 (2001). 10.1103/PhysRevLett.86.4163 [DOI] [PubMed] [Google Scholar]
- 36.Tokel O., Yildiz U. H., Inci F., Durmus N. G., Ekiz O. O., Turker B., Cetin C., Rao S., Sridhar K., Natarajan N., Shafiee H., Dana A., and Demirci U., Sci. Rep. 5, 9152 (2015). 10.1038/srep09152 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Voiculescu I. and Nordin A. N., Biosens. Bioelectron. 33, 1 (2012). 10.1016/j.bios.2011.12.041 [DOI] [PubMed] [Google Scholar]
- 38.Wang J., Jin M., He T., Zhou G., and Shui L., Micromachines 6, 1331 (2015). 10.3390/mi6091331 [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
See in the supplementary material a first video presenting the motion of biotinylated oil droplet in water after streptavidin injection (the center of the video corresponds to the sequence of images shown in Fig. 12), where the droplets seem to be sticking to each other because of streptavidin bridges. The second video is showing the motion of biotinylated oil droplets just after generation before exposure to streptavidin, where droplets can be seen to move freely and glide on each other.