The effect of MLC leaf width in single-isocenter multi-target radiosurgery with volumetric modulated arc therapy

Zhanerke Abisheva 1, Scott R Floyd 2, Joseph K Salama 2, John Kirkpatrick 2, Fang-Fang Yin 2, Michael J Moravan 2, William Giles 2, Justus Adamson 2,
PMCID: PMC6774495  PMID: 31641549

Abstract

Purpose

Single-isocenter multi-target (SIMT) volumetric modulated arc therapy (VMAT) is primarily limited to linear accelerators utilizing 2.5 mm leaf width MLCs. We explore feasibility of applying this technique to linear accelerators utilizing MLCs with leaf width of 5 mm.

Methods

Twenty patients with 3-10 intracranial brain metastases originally treated with 2.5 mm MLCs were re-planned using 5 mm MLCs and relevant dosimetric indices were compared. We also evaluated various strategies of adding VMAT arcs to mitigate degradations of dose quality values.

Results

Wider MLCs caused small changes in total MUs (5827 ± 2334 vs 5572 ± 2220, p = 0.006), and conformity index (CI) (2.22% ± 0.05%, p = 0.045), but produced more substantial increases in brain V30%[%] and V50%[%] (27.75% ± 0.16% and 20.04% ± 0.13% respectively, p < 0.001 for both), and V12Gy[cc] (16.91% ± 0.12%, p < 0.001).

Conclusion

SIMT radiosurgery delivered via VMAT using 5 mm wide MLCs can achieve similar CI compared to that using 2.5 mm leaf width MLCs but with moderately increased isodose spill, which can be only partially mitigated by increasing the number of VMAT arcs.

Keywords: Intracranial brain metastases, SRS, SIMT, VMAT, 2.5 mm leaf width MLCs, 5 mm leaf width MLCs

Introduction

Brain metastases occur in 9% to 17% of cancer patients [1]. An autopsy of the brain revealed that 24% of patients have intracranial metastases and 20% of patients have intraparenchymal or leptomeningeal metastases [2]. Historically, whole-brain radiation therapy (WBRT) has been the standard treatment for brain metastases [1]. However, WBRT is gradually being superseded by stereotactic radiosurgery (SRS), due to improved local control [1,3,4] and improved cognition after treatment [5,6]. SRS combines multiple focused radiation beams to deliver a highly conformal dose to the target while still sparing surrounding normal tissue and can be delivered via multiple platforms including, multiple Cobalt sources [7-9], a linear accelerator mounted on a robotic arm [10,11], or isocentric linear accelerators [12-14].

Radiosurgery using isocentric linear accelerators has traditionally been carried out using dynamic conformal arcs (DCAs) to a single target [15,16], however a newer development is to treat multiple targets simultaneously using volumetric modulated arc therapy (VMAT). Clark et al. first demonstrated a single-isocenter VMAT radiosurgery technique which has the benefit of increased treatment efficiency [4,17] compared to traditional radiosurgery techniques. Single-isocenter multi-target (SIMT) VMAT radiosurgery has been further developed in other studies with increasing clinical use [18,19].

With few exceptions, prior studies using a SIMT VMAT radiosurgery technique have utilized a 2.5 mm leaf width MLCs within ±4 cm of the isocenter [4,17-19]. However, a more common MLC width in use in standard linear accelerators is 5 mm. The benefit of smaller MLC leaf widths is well-documented for single target conformal radiosurgery with DCAs and intensity-modulated radiation therapy (IMRT). Jin et al. showed that for DCAs, small leaf width MLCs provide better dose conformity than large leaf width MLCs especially for small target volumes. For IMRT technique, small leaf width MLCs have better sparing of small organs at risk (OARs) as compared to large leaf width MLCs [20]. Wu et al. also showed the dosimetric benefits of using of 2.5 mm leaf width MLCs with IMRT technique for the treatment of small lesions [21]. Serna et al. performed a treatment planning study comparing 2.5 mm and 5 mm MLCs and reported that for 3D DCA therapy, the usage of small MLC leaf width decreases dose to OAR in close proximity to the planned target volume (PTV) and improves dose conformity [22]. While these benefits are well documented for DCA for a single target, it is not known whether they also apply to SIMT using VMAT. Indeed, a number of differences between these two techniques may diminish the benefit of smaller MLCs. For instance, it is reasonable to expect conformality of DCA plans to be highly sensitive to the MLC leaf width because the MLCs conform to the outline of the PTV; in contrast, for multi-target VMAT the MLC trajectory dose is defined by the inverse optimization and thus may be less dependent on MLC leaf width. In this study, we examined the effect of MLC leaf width on dosimetric quality of SIMT VMAT radiosurgery plans; we also considered several methods of mitigating any dosimetric plan quality degradations caused by 5 mm leaf width MLCs.

Methods and Materials

Summary

We re-planned 20 single-isocenter multiple brain metastases VMAT treatment plans using 5 mm leaf width MLCs which were originally created with 2.5 mm leaf width MLCs. The dosimetric quality comparisons between the plans were made using the V30%[%], V50%[%], V12Gy[cc], and Conformity Index (CI) metrics. Then we investigated several treatment plan modifications to mitigate degradations in dosimetric quality. These included duplicating each VMAT arc from the original plan, delivered with reverse rotation, and adjusting collimator angles (altered by 10°, 15°, and/or 90°). The rationale for this approach was that the additional arcs did not increase the number of couch angles and therefore would not significantly increase the treatment time. The second approach was to add an additional 1-2 VMAT arcs at a new couch position, with all arcs equally spaced.

Patient selection and treatment planning

In a retrospective institutional review board approved study, 20 patient plans used to treat 3-10 intracranial brain metastases with a SIMT technique in 2016 were selected for our study cohort. Our radiosurgery technique using a frameless SIMT technique has been described in detail previously [4,17-19]. In brief, patients were immobilized using noninvasive thermoplastic masks designed specifically for radiosurgery. The patients subsequently underwent computer tomography (CT) based simulation with 1 mm slices, which was fused with diagnostic magnetic resonance imaging (MRI) for treatment planning. Normal structures were contoured by a physicist or dosimetrist (verified by an attending radiation oncologist) using the CT and MRI including the brain, chiasm, optic nerves, eyes, and brainstem. For each metastatic lesion, the gross tumor volume (GTV) was delineated by the attending radiation oncologist and was equal to the clinical target volume (CTV). The PTV was equal to CTV + 1 mm margin. The isocenter was generally placed at the geometric center of the combined PTV but was ultimately decided by the treatment planner. Treatment geometry included 3-5 VMAT non-coplanar arcs with flattening filter free (FFF) 6 MV photons. Single fraction dose was prescribed to sixteen cases and five fractions of dose was prescribed to the rest. All plans were normalized so that 99.5% of the combined PTV received the prescription dose. A Truebeam machine (Varian Medical Systems, Palo Alto CA) equipped with 2.5 mm leaf width MLCs within ±4 cm of the isocenter, and 5 mm leaf width for others was used for treatment planning and delivery of the clinical treatment. At each treatment fraction the patient was aligned prior to treatment using cone-beam CT (CBCT) with a 6-degree of freedom couch correction.

Each plan was re-optimized in the Eclipse treatment planning system (TPS) (Varian Medical Systems, Palo Alto, CA) using the same beam geometry, prescription dose, optimization parameters, and dose normalization and planned to be delivered using a Truebeam machine equipped with a 5 mm leaf width MLCs. The commissioned MLC parameters in the TPS included transmission factor and leaf gap; for the 2.5 mm leaf width MLCs these were 0.012 and 0.4 mm, respectively, and for the 5 mm leaf width MLCs were 0.016 and 1.2 mm, respectively. Aside from MLC parameters, the models for the two beam energies were identically matched, using the anisotropic analytical algorithm (AAA) version 13.6.23 dose calculation model with 1 mm dose grid size.

Plan evaluation

CI, as well as low and moderate isodose spills (brain V30%[%] and V50%[%]) were selected for analysis. We used the following definition of CI:

(1).

(1)

where V100%[%] is the volume receiving 100% of the prescribed dose and VPTV is the volume of the combined PTV. For perfectly conformal plan, CI = 1 and for less conformal plan CI is <1 or>1. We defined V30%[%] and V50%[%] as the volume of brain excluding PTV receiving greater than or equal to 30% and 50% of the prescribed dose, respectively. For plans treated with a single fraction, we also evaluated V12Gy[cc], which is defined as the volume of brain excluding PTV receiving greater than or equal to 12Gy.

Results

The combined PTV volume for all patients ranged from 0.68 to 49.18 cc (mean = 8.78 cc). 16 cases were prescribed 18-20Gy in single fraction and 4 cases were prescribed 5-5.5Gy in 5 fractions.

Re-planning with 2.5 mm leaf width MLC

The screenshot of the dose distributions generated with both 2.5 mm and 5 mm leaf width MLCs using the Eclipse TPS is shown in Figure 1.

Figure 1.

Figure 1

Dose distributions in axial views with both (a) 5 mm and (b) 2.5 mm leaf width MLCs.

Dosimetric parameters from reoptimized SIMT VMAT radiosurgery plans using the 5 mm leaf width MLCs were compared to those using the 2.5 mm MLCs. The average MU decreased from 5827 ± 2334 to 5572 ± 2220 (p = 0.006). The change in CI (mean ± standard deviation) was 2.22% ± 0.05% (p = 0.045). The V30%[%] and V50%[%] increased by 27.75% ± 0.16% and 20.04% ± 0.13% (p < 0.001 for both) respectively, while the V12Gy[cc] increased by 16.91% ± 0.12% (p < 0.001). Total V12Gy[cc] and V12Gy[cc] per target for both 5 and 2.5 mm leaf width MLCs are given in Table 1. The mean and standard deviation of V12Gy[cc] per target for 5 and 2.5 mm leaf width MLC plans are 2.83cc ± 1.17cc and 2.44cc ± 1.03cc respectively. The change in dose indices is summarized in Figure 2.

Table 1.

Total V12Gy[cc] and V12Gy[cc] per target for both 2.5 and 5 mm leaf width MLC plans

Cases Number of targets V12Gy[cc] V12Gy[cc] per target
5 mm leaf width MLC 2.5 mm leaf width MLC 5 mm leaf width MLC 2.5 mm leaf width MLC
1 4 8.75 5.93 2.19 1.48
2 3 7.36 6.59 2.45 2.20
3 7 14.09 12.97 2.01 1.85
4 4 5.99 5.02 1.50 1.26
5 4 7.7 6.56 1.93 1.64
6 4 7.82 7.98 1.96 2.00
7 3 10.36 8.36 3.45 2.79
8 6 9.1 8.4 1.52 1.40
9 4 12.72 11.92 3.18 2.98
10 3 5.54 4.78 1.85 1.59
11 4 11.21 8.42 2.80 2.11
12 3 17.54 15.36 5.85 5.12
13 6 23.73 19.48 3.96 3.25
14 4 13.67 11.12 3.42 2.78
15 3 12.54 11.45 4.18 3.82
16 3 9.22 8.27 3.07 2.76
Mean ± standard deviation 4.1 ± 1.2 11.08 ± 4.68 9.54 ± 3.98 2.83 ± 1.17 2.44 ± 1.03
Paired t-test p < 0.001 p < 0.001

Figure 2.

Figure 2

Percent differences of dose quality values between 2.5 and 5 mm leaf width MLC plans. Positive values indicate an increase in the dose index when the larger MLCs are utilized. The central marks of boxplots indicate median, the edges of boxplots indicate 25th and 75th quartiles, and whiskers extend to most extreme data points not considered outliers, with outliers plotted individually using the ‹+› symbol”.

Modified treatment planning to mitigate dosimetric quality degradation

a) Adding duplicate arcs with unique collimator angle

In an attempt to mitigate the increase in low and moderate isodose spill, we investigated adding duplicate arcs with altered collimator angles for all cases. The rationale for duplicating each arc is that doing so would not increase the number of couch angles in the plan; the duplicated arc could be delivered in reverse direction from the original, thus having negligible effect on treatment delivery time.

While this mitigation strategy resulted in lower CI, V50%[%], V30%[%], and V12Gy[cc] in select cases compared to 5 mm leaf width MLC plans, on average the improvement was modest or non-existent, and these values did not reach the values of the original 2.5 mm leaf width MLC plans. The percent differences of the dosimetric quality values are provided in Table 2.

Table 2.

The mean and standard deviation of the percent differences of V30%[%], V50%[%], V12Gy[cc], and CI values between original plans and re-planned plans using 5 mm leaf width MLCs and the plans with VMAT arcs duplicated and altered by 10°, 15°, and 90°.

Dosimetric values/Methods Re-planned Duplicated, altered by 10° Duplicated, altered by 15° Duplicated, altered by 90°
V30%[%] 27.75% ± 0.16% 24.11% ± 0.14% 29.51% ± 0.19% 35.84% ± 0.21%
V50%[%] 20.04% ± 0.13% 19.39% ± 0.14% 22.13% ± 0.15% 21.02% ± 0.14%
V12Gy[cc] 16.91% ± 0.12% 16.78% ± 0.1% 18.84% ± 0.11% 18.53% ± 0.11%
CI 2.22% ± 0.05% 0.3% ± 0.09% 0.17% ± 0.07% −0.18% ± 0.08%

b) Increasing the number of arcs

In a second attempt to mitigate increased low and moderate isodose spill, we added one and two more VMAT arcs and equally spacing all VMAT arcs. By comparing the percent differences of the dosimetric qualities of 5 mm leaf width MLC plans and 5 mm leaf width MLC plans with one and two added VMAT arc, we can see that as arcs were added all indices improved on average, with the exception of V30%[%]. While this strategy did improve most dosimetric indices, they still did not completely reach the values of the original 2.5 mm leaf width MLC plan. The percent differences of dosimetric quality values are shown in Table 3.

Table 3.

The mean and standard deviation of the percent differences of V30%[%], V50%[%], V12Gy[cc], and CI values between original plans and re-planned plans using 5 mm leaf width MLCs and the plans with additional one and two VMAT arcs.

Dosimetric values/Methods Re-planned +1 VMAT arc +2 VMAT arcs
V30%[%] 27.75% ± 0.16% 24.51% ± 0.18% 27.94% ± 0.28%
V50%[%] 20.04% ± 0.13% 18.24% ± 0.13% 17% ± 0.13%
V12Gy[cc] 16.91% ± 0.12% 16.18% ± 0.09% 15.57% ± 0.1%
CI 2.22% ± 0.05% 1.33% ± 0.08% 0.55% ± 0.07%

Figures 3, 4, 5, and 6 depict the percent differences between original 2.5 mm leaf width MLC plans and 5 mm leaf width MLC plans and other modified 5 mm leaf width MLC plans for dosimetric quality values V30%[%], V50%[%], V12Gy[cc], and CI. The central marks of boxplots indicate median, the edges of boxplots indicate 25th and 75th quartiles, and whiskers extend to most extreme data points not considered outliers, with outliers plotted individually using the ‘+’ symbol”.

Figure 3.

Figure 3

The percent differences of V30%[%] values between original 2.5 mm leaf width MLC plans and re-planned plans using 5mm leaf width MLCs, 5 mm leaf width MLC plans with inversely positioned and duplicated arcs with collimator angles altered by 10°, 15°, and 90°, 5 mm leaf width MLC plans with one or two additional equally spaced arcs.

Figure 4.

Figure 4

The percent differences of V50%[%] values between original 2.5 mm leaf width MLC plans and re-planned plans using 5 mm leaf width MLCs, 5mm leaf width MLC plans with inversely positioned and duplicated arcs with collimator angles altered by 10°, 15°, and 90°, 5 mm leaf width MLC plans with one or two additional equally spaced arcs.

Figure 5.

Figure 5

The percent differences of V12Gy[cc] values between original 2.5 mm leaf width MLC plans and re-planned plans using 5 mm leaf width MLCs, 5 mm leaf width MLC plans with inversely positioned and duplicated arcs with collimator angles altered by 10°, 15°, and 90°, 5 mm leaf width MLC plans with one or two additional equally spaced arcs.

Discussion

In the present study we have evaluated the dosimetric differences in using MLC leaf widths of 2.5 mm and 5 mm for SIMT radiosurgery with VMAT. Given the higher clinical prevalence of 5 mm leaves and increasing use of SIMT radiosurgery, we analyzed clinical plan quality of SIMT radiosurgery delivered via VMAT using 5 mm leaves. The dosimetric values CI, V50%[%], V30%[%], and V12Gy[cc] were used for single fraction cases for the comparison between two plans which were taken from the dose volume histograms. We also tested several ways to mitigate the dosimetric value degradations after the re-planning with 5 mm width MLCs. We found that 5 mm MLCs had plan degradation compared to 2.5 mm leaf width MLC delivery. By adding two more VMAT arcs and equally spacing them, we were able to improve the dose quality values, beginning to approach the plan quality of the original 2.5 mm leaf width MLC plan.

Figure 6.

Figure 6

The percent differences of CI values between original 2.5 mm leaf width MLC plans and re-planned plans using 5 mm leaf width MLCs, 5 mm leaf width MLC plans with inversely positioned and duplicated arcs with collimator angles altered by 10°, 15°, and 90°, 5 mm leaf width MLC plans with one or two additional equally spaced arcs.

In comparison, Huang et al in their study compared the plan qualities of single-isocenter dynamic conformal arcs (SIDCA) technique with multiple-isocenter dynamic conformal arcs (MIDCA) and VMAT [23]. They found that SIDCA and MIDCA have similar plan qualities, but MIDCA has longer delivery time. VMAT has faster delivery time and better conformity. However, SIDCA has lower peripheral dose compared to VMAT.

Prior studies have found that V12Gy[cc] is highly predictive for the development of radiation necrosis (RN) in patients receiving multi-isocenter, single-fraction SRS [24-26]. Therefore, part of our analysis included quantifying the absolute volume receiving V12Gy[cc] and the V12Gy[cc] per target. As expected, we found that using the larger leaves caused the V12Gy[cc] to increase from 9.54 ± 3.98cc to 11.08 ± 4.68cc (mean ± standard deviation), which corresponded to an increase in V12Gy[cc] per target from 2.44 ± 1.03cc to 2.83 ± 1.17cc. Minniti et al. investigated the factors affecting survival and toxicity in patients treated with SRS with no history of WBRT. They analyzed V10-V16Gy[cc] as predictors for brain RN and found out that V10Gy[cc] and V12Gy[cc] were the strongest predictors and that lesions with V12Gy[cc] >8.5 cc have a risk of RN >10% [24]. Korytko et al. performed a retrospective review of patients with non-arteriovenous malformation (non-AVM) intracranial tumors and found that the risk of symptomatic RN correlated with V12Gy[cc] and that the risk of RN increases significantly for V12Gy[cc]>10 cc per lesion [25]. Finally, Blonigen et al. performed a retrospective analysis of 63 patients treated with SRS for brain metastases to investigate patient and treatment factors for predictors of RN and found that V8Gy[cc]-V16Gy[cc] were the best predictors for symptomatic RN. They also found that patients with V12Gy[cc] >10.8 cc had a 68.8% rate of RN, however it should be noted that 63 % of patients received WBRT prior to SRS, which has been shown in other studies to increase the risk of RN [26]. To the author’s knowledge, there have been no trials to date that have reported on the incidence of RN following SIMT SRS and/or describing patient or treatment specific factors that are predictive of the development of RN. Additionally, the aforementioned studies primarily considered the probability of RN in the context of a single target volume, as a majority of the patients reported had only 1 lesion. It is doubtful that these same rates would apply to a total volume of V12Gy[cc] in the setting of multiple isolated targets and for this reason both total volume of V12Gy[cc] and volume per target V12Gy[cc] were reported.

It is important to note that in this study only Varian MLCs and their specific characteristics as well as the Varian Eclipse modeling of those MLCs was analyzed. MLCs from different vendors with different leakage characteristics and leaf edge design may have different results. Additionally, the results may be dependent on how well the leakage and leaf edges are modeled in the treatment planning system.

One of the purposes of this study was to broaden the applicability of SIMT radiosurgery using VMAT to include the greater number of linear accelerators with standard 5 mm MLCs. However, it should be noted that leaf size is not the only consideration as to whether a SIMT VMAT technique can and should be used. For instance, Stanhope et al in their study considered the challenges of the increased sensitivity to rotational uncertainties which results from distance of the targets from isocenter and potential for decreased plan quality from larger multi leaf collimator width N > 4 cm from isocenter with single-isocenter radiosurgery for multiple intracranial targets (SIRMIT). They showed the need for rotational corrections via image guidance for SIRMIT when a thermoplastic mask is used for immobilization [18]. Given this, a separate requirement for SIMT VMAT radiosurgery is the ability to apply 6D-couch corrections, which also greatly limits the number of linear accelerators to which it can be applied. However, many radiosurgery immobilization systems have the ability to adjust pitch and roll manually, which could potentially serve the same purpose. Additionally, 6-degree couches are now commonly offered on LINACs with 5 mm MLC width, making this combination commonly commercially available.

Conclusion

Using 5 mm MLCs for SIMT VMAT radiosurgery leads to minor increases in conformity index and moderate increases in low and moderate isodose spill. Duplicated VMAT arcs could not mitigate the changes in the dosimetric values of 5 mm leaf width MLC plans compared to 2.5 mm leaf width MLC plans. We were able to see the decrease trend of the percent differences of the dosimetric quality values when adding one more and two more VMAT arcs.

Acknowledgments

Authors disclosure of potential conflicts of interest

Fang-Fang Yin discloses research and licensing agreements with Varian Medical Systems, not related to this study and is also supported by NIH: 2R21CA218940, 1R01 CA184173-01A1, 5R21CA195317-02. John Kirkpatrick discloses research agreement with Varian Medical Systems and ownership of ClearSight RT LLC, both of which are unrelated to this study. Justus Adamson discloses ownership of ClearSight RT LLC, which is unrelated to this study. Other authors have nothing to disclose.

Author contributions

Conception and design: Scott R. Floyd, Joseph K. Salama, John Kirkpatrick, Fang-Fang Yin, Michael Moravan, William Giles, Justus Adamson

Data collection: Zhanerke Abisheva, William Giles, Justus Adamson

Data analysis and interpretation: Zhanerke Abisheva, William Giles, Justus Adamson

Manuscript writing: Zhanerke Abisheva, William Giles, Justus Adamson

Final approval of manuscript: Scott R. Floyd, Joseph K. Salama, John Kirkpatrick, Fang-Fang Yin, Michael Moravan, William Giles, and Justus Adamson

Nomenclature

SIMT single-isocenter multi-target
VMAT volumetric modulated arc therapy
MLC multi-leaf collimator
MU monitor unit
CI conformity index
V30%[%] volume of brain excluding PTV receiving greater than or equal to 30% of the prescribed dose
V50%[%] volume of brain excluding PTV receiving greater than or equal to 50% of the prescribed dose
V12Gy[cc] volume of brain excluding PTV receiving greater than or equal to 12Gy
WBRT whole-brain radiation therapy
SRS stereotactic radiosurgery
DCA dynamic conformal arc
IMRT intensity-modulated radiation therapy
OAR organ at risk
3D three-dimensional
PTV planned target volume
CT computer tomography
MRI magnetic resonance imaging
GTV gross tumor volume
CTV clinical target volume
FFF flattening filter free
MV megavolt
CBCT cone-beam CT
TPS treatment planning system
AAA anisotropic analytical algorithm
V100%[%] volume of brain receiving 100% of the prescribed dose
VPTV volume of the PTV
SIDCA single-isocenter dynamic conformal arcs
MIDCA multiple-isocenter dynamic conformal arcs
RN radiation necrosis
V10Gy[cc] volume of brain excluding PTV receiving greater than or equal to 10Gy
V8Gy[cc] volume of brain excluding PTV receiving greater than or equal to 8Gy
V16Gy[cc] volume of brain excluding PTV receiving greater than or equal to 16Gy
SIRMIT single-isocenter radiosurgery for multiple intracranial targets
6D six-dimensional
LINAC linear accelerator

References

  • 1. Badiyan SN, Regine WF, Mehta M. Stereotactic radiosurgery for treatment of brain metastases. J Oncol Pract [Internet]. 2016;12(8):703–12. http://ascopubs.org/doi/10.1200/JOP.2016.012922 [DOI] [PubMed] [Google Scholar]
  • 2. Lassman AB, DeAngelis LM. Brain metastases. Neurol Clin. 2003;21(1):1–23. [DOI] [PubMed] [Google Scholar]
  • 3. Lin X, DeAngelis LM. Treatment of brain metastases. J Clin Oncol. 2015;33(30):3475–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Clark GM, Popple RA, Young PE, Fiveash JB. Feasibility of single-isocenter volumetric modulated arc radiosurgery for treatment of multiple brain metastases. Int J Radiat Oncol Biol Phys. 2010;76(1):296–302. [DOI] [PubMed] [Google Scholar]
  • 5. Brown PD, Jaeckle K, Ballman KV, Farace E, Jane H, Anderson SK, et al. Effect of radiosurgery alone vs radiosurgery with whole brain radiation therapy on caognitive function in patients with 1 to 3 brain metastases: A randomized clinical trial. J Am Med Assoc. 2017;316(4):401–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Lippitz B, Lindquist C, Paddick I, Peterson D, O’Neill K, Beaney R. Stereotactic radiosurgery in the treatment of brain metastases: The current evidence. Cancer Treat Rev [Internet]. 2014;40(1):48–59. 10.1016/j.ctrv.2013.05.002 [DOI] [PubMed] [Google Scholar]
  • 7. Liew DN, Kano H, Kondziolka D, Mathieu D, Niranjan A, Flickinger JC, et al. Outcome predictors of Gamma Knife surgery for melanoma brain metastases. J Neurosurg [Internet]. 2011;114(3):769–79. http://thejns.org/doi/10.3171/2010.5.JNS1014 [DOI] [PubMed] [Google Scholar]
  • 8. Harner G, Foote L, Coffey RJ, Swanson JW, Beat-i-y CW, Kljne RW, et al. Stereotactic radiosurgery using the Gamma Knife for acoustic neuromas. Science (80- ). 1995;32(4):1153–60. [DOI] [PubMed] [Google Scholar]
  • 9. DiBiase SJ, Kwok Y, Yovino S, Arena C, Naqvi S, Temple R, et al. Factors predicting local tumor control after gamma knife stereotactic radiosurgery for benign intracranial meningiomas. Int J Radiat Oncol Biol Phys. 2004;60(5):1515–9. [DOI] [PubMed] [Google Scholar]
  • 10. Araki F. Monte Carlo study of a Cyberknife stereotactic radiosurgery system. Med Phys. 2006;33(8):2955–63. [DOI] [PubMed] [Google Scholar]
  • 11. Antypas C, Pantelis E. Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system. Phys Med Biol. 2008;53(17):4697–718. [DOI] [PubMed] [Google Scholar]
  • 12. Puataweepong P, Dhanachai M, Dangprasert S, Narkwong L, Sitathanee C, Sawangsilpa T, et al. Linac-based stereotactic radiosurgery and fractionated stereotactic radiotherapy for vestibular schwannomas: Comparative observations of 139 patients treated at a single institution. J Radiat Res. 2014;55(2):351–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Pokhrel D, Sood S, Mcclinton C, Saleh H, Badkul R, Jiang H, et al. Linac-based stereotactic radiosurgery (SRS) in the treatment of refractory trigeminal neuralgia: Detailed description SRS of procedure and reported clinical outcomes. J Appl Clin Med Phys. 2017;18(2):136–43. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Ruggieri R, Naccarato S, Mazzola R, Ricchetti F, Corradini S, Fiorentino A, et al. Linac-based VMAT radiosurgery for multiple brain lesions: Comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique. Radiat Oncol. 2018;13(1):1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Hazard LJ, Wang B, Skidmore TB, Chern SS, Salter BJ, Jensen RL, et al. Conformity of LINAC-based stereotactic radiosurgery using dynamic conformal arcs and micro-multileaf collimator. Int J Radiat Oncol Biol Phys. 2009;73(2):562–70. [DOI] [PubMed] [Google Scholar]
  • 16. Solberg TD, Boedeker KL, Fogg R, Selch MT, DeSalles AAF. Dynamic arc radiosurgery field shaping: A comparison with static field conformal and noncoplanar circular arcs. Int J Radiat Oncol Biol Phys. 2001;49(5):1481–91. [DOI] [PubMed] [Google Scholar]
  • 17. Clark GM, Popple RA, Prendergast BM, Spencer SA, Thomas EM, Stewart JG, et al. Plan quality and treatment planning technique for single isocenter cranial radiosurgery with volumetric modulated arc therapy. Pract Radiat Oncol. 2012;2(4):306–13. [DOI] [PubMed] [Google Scholar]
  • 18. Stanhope C, Chang Z, Wang Z, Yin FF, Kim G, Salama JK, et al. Physics considerations for single-isocenter, volumetric modulated arc radiosurgery for treatment of multiple intracranial targets. Pract Radiat Oncol [Internet]. 2016;6(3):207–13. 10.1016/j.prro.2015.10.010 [DOI] [PubMed] [Google Scholar]
  • 19. Morrison J, Hood R, Yin FF, Salama JK, Kirkpatrick J, Adamson J. Is a single isocenter sufficient for volumetric modulated arc therapy radiosurgery when multiple itracranial metastases are spatially dispersed? Med Dosim [Internet]. 2016;41(4):285–9. 10.1016/j.meddos.2016.06.007 [DOI] [PubMed] [Google Scholar]
  • 20. Jin JY, Yin FF, Ryu S, Ajlouni M, Jae HK. Dosimetric study using different leaf-width MLCs for treatment planning of dynamic conformal arcs and intensity-modulated radiosurgery. Med Phys. 2005;32(2):405–11. [DOI] [PubMed] [Google Scholar]
  • 21. Wu QJ, Wang Z, Kirkpatrick JP, Chang Z, Meyer JJ, Lu M, Huntzinger C, Yin FF. Impact of collimator leaf width and treatment technique on stereotactic radiosurgery and radiotherapy plans for intra- and extracranial lesions. Radiat Oncol. 2009;10:1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Serna A, Puchades V, Mata F, Ramos D, Alcaraz M. Influence of multi-leaf collimator leaf width in radiosurgery via volumetric modulated arc therapy and 3D dynamic conformal arc therapy. Phys Medica [Internet]. 2015;31(3):293–6. 10.1016/j.ejmp.2015.01.011 [DOI] [PubMed] [Google Scholar]
  • 23. Huang Y, Chin K, Robbins JR, Kim J, H Li, Amro H, et al. Radiosurgery of multiple brain metastases with single-isocenter dynamic conformal arcs (SIDCA). Radiother Oncol. 2014;112(1):128–32. [DOI] [PubMed] [Google Scholar]
  • 24. Minniti G, Clarke E, Lanzetta G, Osti MF, Trasimeni G, Bozzao A, et al. Stereotactic radiosurgery for brain metastases: analysis of outcome and risk of brain radionecrosis. 2011;1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Korytko T, Radivoyevitch T, Colussi V, Wessels BW, Pillai K, Maciunas RJ, et al. 12 gy gamma knife radiosurgical volume is a predictor for radiation necrosis in non-avm intracranial tumors. Int J Radiat Oncol Biol. Phys. 2006; 64(2): 419–424. [DOI] [PubMed] [Google Scholar]
  • 26. Blonigen BJ, Steinmetz RD, Levin L, Lamba MA, Warnick RE, Breneman JC. Irradiated volume as a predictor of brain radionecrosis after linear accelerator stereotactic radiosurgery. Int J Radiat Oncol Biol Phys. 2010;77(4):996–1001. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Radiosurgery and SBRT are provided here courtesy of Old City Publishing

RESOURCES